← Product Code [QJI](/productcode/QJI) · K250798

# Control-IQ+ technology (K250798)

_Tandem Diabetes Care, Inc. · QJI · May 21, 2025 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K250798

## Device Facts

- **Applicant:** Tandem Diabetes Care, Inc.
- **Product Code:** [QJI](/productcode/QJI.md)
- **Decision Date:** May 21, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1356
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Software as a Medical Device, Therapeutic, PCCP, Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K250798 · May 21, 2025 | Control-IQ+ technology | Tandem Diabetes Care, Inc. | T1D Exchange registry | The sponsor compared the incidence of severe hypoglycemia and diabetic ketoacidosis (DKA) observed in their prospective clinical study against historical event rates reported in the T1D Exchange registry to demonstrate safety. | T1D Exchange; Safety benchmarking; Registry data |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| T1D Exchange registry comparison; Registry-based comparative analysis; Follow-up/Duration: 3-month period; Study Period: 2016-2018 | Participants with Type 1 Diabetes matched by age to the pivotal study population | Prospective clinical study (TL1) of Control-IQ+ with Lyumjev | Severe hypoglycemia (SH) and Diabetic Ketoacidosis (DKA) event rates |

## Indications for Use

Control-IQ+ technology is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and suspend delivery of basal insulin based on iCGM readings and predicted glucose values. It can also deliver correction boluses when the glucose value is predicted to exceed a predefined threshold. Control-IQ+ technology is intended for the management of Type 1 diabetes mellitus in persons 2 years of age and greater and of Type 2 diabetes mellitus in persons 18 years of age and greater. Control-IQ+ technology is intended for single patient use and requires a prescription.

## Device Story

Software-only device (Control-IQ+) installed on compatible ACE insulin pumps; manages Type 1 and Type 2 diabetes. Inputs: iCGM glucose readings, user-entered carbohydrate intake, exercise, sleep schedule, weight, and total daily insulin requirements. Operation: Closed-loop algorithm automatically adjusts basal insulin delivery (increase, decrease, suspend) and delivers correction boluses based on predicted glucose values. Modes: Normal (target 112.5-160 mg/dL), Sleep (112.5-120 mg/dL), Exercise (140-160 mg/dL). Output: Automated insulin pump commands. Used in home/outpatient settings by patients. Healthcare providers establish baseline insulin requirements. Output enables automated glycemic control, reducing hypoglycemia/hyperglycemia risk.

## Clinical Evidence

Prospective, multi-center, single-arm study (N=173 completers) evaluated safety of Lyumjev U-100 in the t:slim X2 pump with Control-IQ+. Participants (Type 1 diabetes, ages 6-81) underwent 16-day Humalog lead-in followed by 13-week Lyumjev treatment. Primary endpoints: severe hypoglycemia (SH) and diabetic ketoacidosis (DKA) rates. Results: 1.7% SH incidence (vs 6.1% in T1D Exchange, p=0.07); 0% DKA incidence (vs 2.8% in T1D Exchange, p=0.04). Secondary metrics (Time in Range, hypoglycemia event rates) showed no significant degradation compared to Humalog. Clinical literature review supported extrapolation to Type 2 diabetes (≥18 years) and children (2-5 years).

## Technological Characteristics

Software-only iAGC; interoperable with ACE pumps and iCGM. Compatible with NovoLog U-100, Humalog U-100, and Lyumjev U-100 insulins. Features three modes (Normal, Sleep, Exercise) with fixed target ranges. Basal rate attenuation and automatic correction bolus logic based on predicted glucose values. User-configurable weight (20-440 lbs) and total daily insulin (5-200 units/day) settings.

## Regulatory Identification

An interoperable automated glycemic controller is a device intended to automatically calculate drug doses based on inputs such as glucose and other relevant physiological parameters, and to command the delivery of such drug doses from a connected infusion pump. Interoperable automated glycemic controllers are designed to reliably and securely communicate with digitally connected devices to allow drug delivery commands to be sent, received, executed, and confirmed. Interoperable automated glycemic controllers are intended to be used in conjunction with digitally connected devices for the purpose of maintaining glycemic control.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) An appropriate, as determined by FDA, clinical implementation strategy, including data demonstrating appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use.
(A) The clinical data must be representative of the performance of the device in the intended use population and in clinically relevant use scenarios and sufficient to demonstrate appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use.
(B) For devices indicated for use with multiple therapeutic agents for the same therapeutic effect (
*e.g.,* more than one type of insulin), data demonstrating performance with each product or, alternatively, an appropriate, as determined by FDA, clinical justification for why such data are not needed.(C) When determined to be necessary by FDA, the strategy must include postmarket data collection to confirm safe real-world use and monitor for rare adverse events.
(ii) Results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use.
(iii) A detailed and appropriate, as determined by FDA, strategy to ensure secure and reliable means of data transmission with other intended connected devices.
(iv) Specifications that are appropriate, as determined by FDA, for connected devices that shall be eligible to provide input to (
*e.g.,* specification of glucose sensor performance) or accept commands from (*e.g.,* specifications for drug infusion pump performance) the controller, and a detailed strategy for ensuring that connected devices meet these specifications.(v) Specifications for devices responsible for hosting the controller, and a detailed and appropriate, as determined by FDA, strategy for ensuring that the specifications are met by the hosting devices.
(vi) Documentation demonstrating that appropriate, as determined by FDA, measures are in place (
*e.g.,* validated device design features) to ensure that safe therapy is maintained when communication with digitally connected devices is interrupted, lost, or re-established after an interruption. Validation testing results must demonstrate that critical events that occur during a loss of communications (*e.g.,* commands, device malfunctions, occlusions, etc.) are handled and logged appropriately during and after the interruption to maintain patient safety.(vii) A detailed plan and procedure for assigning postmarket responsibilities including adverse event reporting, complaint handling, and investigations with the manufacturers of devices that are digitally connected to the controller.
(2) Design verification and validation documentation must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following:
(i) Risk control measures to address device system hazards;
(ii) Design decisions related to how the risk control measures impact essential performance; and
(iii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design.
(3) The device shall include appropriate, as determined by FDA, and validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following:
(i) Secure authentication (pairing) to connected devices;
(ii) Secure, accurate, and reliable means of data transmission between the controller and connected devices;
(iii) Sharing of necessary state information between the controller and any connected devices (
*e.g.,* battery level, reservoir level, sensor use life, pump status, error conditions);(iv) Ensuring that the controller continues to operate safely when data is received in a manner outside the bounds of the parameters specified;
(v) A detailed process and procedures for sharing the controller's interface specification with connected devices and for validating the correct implementation of that protocol; and
(vi) A mechanism for updating the controller software, including any software that is required for operation of the controller in a manner that ensures its safety and performance.
(4) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices, and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the controller must, at a minimum, include:
(i) Commands issued by the controller, and associated confirmations the controller receives from digitally connected devices;
(ii) Malfunctions of the controller and malfunctions reported to the controller by digitally connected devices (
*e.g.,* infusion pump occlusion, glucose sensor shut down);(iii) Alarms and alerts and associated acknowledgements from the controller as well as those reported to the controller by digitally connected devices; and
(iv) Connectivity events (
*e.g.,* establishment or loss of communications).(5) The device must only receive glucose input from devices cleared under § 862.1355 (integrated continuous glucose monitoring system), unless FDA determines an alternate type of glucose input device is designed appropriately to allow the controller to meet the special controls contained within this section.
(6) The device must only command drug delivery from devices cleared under § 880.5730 of this chapter (alternate controller enabled infusion pump), unless FDA determines an alternate type of drug infusion pump device is designed appropriately to allow the controller to meet the special controls contained within this section.
(7) An appropriate, as determined by FDA, training plan must be established for users and healthcare providers to assure the safety and performance of the device when used. This may include, but not be limited to, training on device contraindications, situations in which the device should not be used, notable differences in device functionality or features compared to similar alternative therapies, and information to help prescribers identify suitable candidate patients, as applicable.
(8) The labeling required under § 809.10(b) of this chapter must include:
(i) A contraindication for use in pediatric populations except to the extent clinical performance data or other available information demonstrates that it can be safely used in pediatric populations in whole or in part.
(ii) A prominent statement identifying any populations for which use of this device has been determined to be unsafe.
(iii) A prominent statement identifying by name the therapeutic agents that are compatible with the controller, including their identity and concentration, as appropriate.
(iv) The identity of those digitally connected devices with which the controller can be used, including descriptions of the specific system configurations that can be used, per the detailed strategy submitted under paragraph (b)(1)(iii) of this section.
(v) A comprehensive description of representative clinical performance in the hands of the intended user, including information specific to use in the pediatric use population, as appropriate.
(vi) A comprehensive description of safety of the device, including, for example, the incidence of severe hypoglycemia, diabetic ketoacidosis, and other relevant adverse events observed in a study conducted to satisfy paragraph (b)(1)(i) of this section.
(vii) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device.
(viii) For any controller with hardware components intended for multiple patient reuse, instructions for safely reprocessing the hardware components between uses.

## Predicate Devices

- Control-IQ+ technology ([K243823](/device/K243823.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

## I Background Information:

A 510(k) Number

K250798

B Applicant

Tandem Diabetes Care, Inc.

C Proprietary and Established Names

Control-IQ+ technology

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  QJI | Class II | 21 CFR 862.1356 - Interoperable Automated Glycemic Controller | CH - Clinical Chemistry  |

## E. Purpose for Submission:

The purpose of this submission is to add Lyumjev U-100 insulin as a compatible insulin to the labeling of the subject device.

## II Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

B Indication(s) for Use:

Control-IQ+ technology is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and suspend delivery of basal insulin based on iCGM readings and predicted glucose values. It can also deliver correction boluses when the glucose value is predicted to exceed a predefined threshold.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

Control-IQ+ technology is intended for the management of Type 1 diabetes mellitus in persons 2 years of age and greater and of Type 2 diabetes mellitus in persons 18 years of age and greater.

Control-IQ+ technology is intended for single patient use and requires a prescription.

## C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

Only use U-100 insulin analogs that have been tested and found to be compatible for use in the pump. Use of insulin with lesser or greater concentration can result in under delivery or over delivery of insulin. This can cause hypoglycemia (low BG) or hyperglycemia (high BG) events.

When the CGM reading is automatically populated into the bolus calculator, only the current CGM reading is used to calculate the correction bolus. The trend arrow is not used in the dose calculation. Speak with your healthcare provider for recommendations on how best to utilize the arrows for your correction bolus dosing.

Control-IQ+ should not be used in anyone under the age of two years old. Control-IQ+ should also not be used in patients who require less than a total daily insulin dose of 5 units per day or who weigh less than 20 pounds, as those are the required minimum values needed in order for Control-IQ technology to operate safely.

The pump is magnetic resonance (MR) unsafe. You must take off your pump and leave it outside the procedure room.

DO NOT use Dexcom CGM readings to make diabetes treatment decisions or assess glucose control when taking hydroxyurea, a medication used in the treatment of diseases including cancer and sickle cell anemia. Hydroxyurea is known to interfere with glucose readings from the Dexcom sensor. Relying on sensor glucose values while taking hydroxyurea could result in missed hypoglycemia alerts or errors in diabetes management, such as giving a higher dose of insulin than necessary to correct falsely high sensor glucose values.

## III Device Description:

The Subject Device, Control-IQ+ technology ("Control-IQ+") is a software-only device intended for the management of type 1 and type 2 diabetes mellitus. The device controls insulin delivery from a compatible alternate controller enabled insulin pump (ACE pump) based on inputs provided by a compatible integrated continuous glucose monitor (iCGM) and inputs provided by the user (e.g., carbohydrate intake, exercise, and sleep schedule). Control-IQ+ technology is meant to be installed on a compatible ACE pump.

Control-IQ+ technology has three different modes: Normal, Sleep, and Exercise. The glucose targets are not individually customizable in these modes but can change based on the mode selected. During Normal mode, Control-IQ+ technology aims to control glucose within a target range of $112.5 - 160\mathrm{mg / dL}$, during Sleep mode the target range is $112.5 - 120\mathrm{mg / dL}$, and during Exercise mode the target range is $140 - 160\mathrm{mg / dL}$.

K250798 - Page 2 of 14

{2}

Control-IQ+ technology includes an integrated feature whereby iCGM values are automatically populated into the glucose field of the integrated bolus calculator when Control-IQ+ technology is active (i.e., the device is operating in closed-loop mode). This feature is disabled when Control-IQ is turned off.

Control-IQ+ technology requires users to input their weight and their total daily insulin requirement, which should be established with the help of a health care provider before using the device.

## IV Substantial Equivalence Information:

A Predicate Device Name(s):

Control-IQ+ technology

B Predicate 510(k) Number(s):

K243823

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K250798 | K243823  |
| --- | --- | --- |
|  Device Trade Name | Control-IQ+ technology | Same  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | Control-IQ+ technology is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and suspend delivery of basal insulin based on iCGM readings and predicted glucose values. It can also deliver correction boluses when the glucose value is predicted to exceed a predefined threshold.
Control-IQ+ technology is intended for the management of Type 1 diabetes mellitus in persons 2 years of age and greater and of Type 2 diabetes mellitus in persons 18 years of age and greater. | Same  |

K250798 - Page 3 of 14

{3}

K250798 - Page 4 of 14

|   | Control-IQ+ technology is intended for single patient use and requires a prescription. |   |
| --- | --- | --- |
|  General Device Characteristic Differences |  |   |
|  Compatible Insulins | For Type 1 diabetes mellitus in persons 2 years of age and greater and Type 2 diabetes mellitus in persons 18 years of age and greater: NovoLog U-100 Insulin Humalog U-100 Insulin Lyumjev U-100 Insulin | For Type 1 diabetes mellitus in persons 2 years of age and greater and Type 2 diabetes mellitus in persons 18 years of age and greater: NovoLog U-100 Insulin Humalog U-100 Insulin  |

## V Standards/Guidance Documents Referenced:

- Special controls established under 21 CFR 862.1356
- ANSI AAMI ISO 14971:2019 - “Medical devices - Applications of risk management to medical devices”

## VI Performance Characteristics:

### A Analytical Performance:

For the purposes of analytical and clinical validation testing, the Control-IQ algorithm was installed on the t:slim X2 Insulin Pump with Interoperable Technology ACE pump, which was paired with the Dexcom G6 Continuous Glucose Monitoring (CGM) system.

### B Other Supportive Instrument Performance Characteristics Data:

#### Summary of Clinical Testing:

The sponsor conducted a prospective, multi-center, single-arm study to evaluate the safety of Lyumjev (insulin lispro-aabc) U-100 insulin use in the Tandem t:slim X2 insulin pump with Control-IQ+ technology in adult and pediatric participants with type 1 diabetes in an outpatient setting. The Lyumjev Treatment Period was 13 weeks in duration.

A summary of the pivotal clinical study is provided in the following table:

|  Study Feature | Description  |
| --- | --- |
|  Title | Safety Evaluation of an Advanced Hybrid Closed Loop System Using Lyumjev with the Tandem t:slim X2 with Control-IQ in Adults, Adolescents and Children with Type 1 Diabetes (TL1)  |
|  Summary | The study design was a single-arm trial of existing pediatric and adult users of the Tandem t:slim X2 insulin pump with Control-IQ+ technology. Participants completed a ~16 day Humalog Lead-In  |

{4}

|   | Period using Humalog in the pump followed by a 13-week period using Lyumjev in the pump. At-home exercise and meal challenges were performed in both periods.  |
| --- | --- |
|  Investigational Device | t:slim X2 insulin pump with Control-IQ+ technology  |
|  Objectives | To evaluate the safety of Lyumjev (insulin lispro-aabc) use in the Tandem t:slim X2 insulin pump with Control-IQ+ technology in adult and pediatric participants with type 1 diabetes in an outpatient setting to support system labeling.  |
|  Study Design | Single-arm prospective safety trial  |
|  Number of Sites | 14 clinical sites  |
|  Population | 173 participants completed the study. The mean ± SD age at enrollment was 24 ± 17 years, mean HbA1c at the start of Lyumjev was 7.1% ± 0.9%, 50% of participants were female, and 95% were White.

Key Inclusion Criteria:
• Age 6 to <81 years
• Diagnosis of Type 1 Diabetes for at least 1 year
• Currently using Control-IQ technology for at least 3 months
• Total daily insulin dose at least 2 U/day
• HbA1c <10.5%

Key Exclusion Criteria:
• More than 1 episode of diabetic ketoacidosis or severe hypoglycemia (needing assistance) in the past 6 months
• Concurrent use of any non-insulin glucose-lowering agent other than metformin  |
|  Sample Size | 193 participants were enrolled in the study, 179 completed the Humalog run-in and initiated Lyumjev insulin, and 173 completed the 13 week visit post-Lyumjev initiation.  |
|  Treatment Groups | The study had 2 periods: Humalog Lead-In Period (~16 days) and Lyumjev Treatment Period (13 weeks).  |
|  Study Duration | Duration of treatment was about 15 weeks (~16 days of Humalog lead-in, and 13 weeks of Lyumjev treatment)  |
|  Protocol Overview/Synopsis | Informed consent was signed, and eligibility was assessed. Eligible participants who were using Control-IQ for at least 3 months initiated a Humalog Lead-In Period of ~16 days. Participants who successfully completed the Humalog Lead-In Period, indicated by 85% of active closed-loop use during this period, were trained on the use of the study pump with Lyumjev and initiated the Lyumjev Treatment Period for 13 weeks.

Participants were asked to perform 1 missed meal bolus and 1 exercise challenge at home during the Humalog Lead-in Period, then 3 meal and 3 exercise at-home challenges during the Lyumjev Treatment Period.  |

K250798 - Page 5 of 14

{5}

Participant Demographics

|   | Overall
N=179 | Pediatrics
N=109 | Adults
N=70  |
| --- | --- | --- | --- |
|  Age at Enrollment (years)
mean ± standard deviation (SD)
Range | 24 ± 17
6 to 75 | 12 ± 3
6 to 17 | 43 ± 14
18 to 75  |
|  Sex – Female (n (%)) | 90 (50%) | 57 (52%) | 33 (47%)  |
|  Weight (kg)
mean ± SD
Range | 65 ± 26
21 to 138 | 51 ± 18
21 to 102 | 87 ± 21
56 to 138  |
|  Body Mass Index at Enrollment (kg/m²)
mean ± SD
Range | NA | NA | 28.9 ± 5.9
19.1 to 47.8  |
|  Race (n (%))^{a}
White
Black/African American
Asian
Native Hawaiian/Pacific Islander | 166 (95%)
1 (<1%)
2 (1%)
1 (<1%) | 97 (92%)
1 (<1%)
2 (2%)
0 (0%) | 69 (99%)
0 (0%)
0 (0%)
1 (1%)  |

K250798 - Page 6 of 14

{6}

|   | Overall N=179 | Pediatrics N=109 | Adults N=70  |
| --- | --- | --- | --- |
|  American Indian/Alaskan Native | 1 (<1%) | 1 (<1%) | 0 (0%)  |
|  More than One Race | 4 (2%) | 4 (4%) | 0 (0%)  |
|  Ethnicity – (n (%))a |  |  |   |
|  Hispanic | 12 (7%) | 10 (9%) | 2 (3%)  |
|  Non-Hispanic | 165 (93%) | 97 (91%) | 68 (97%)  |
|  Highest Level of Education (n (%))a,b |  |  |   |
|  <Bachelor’s Degree | 49 (28%) | 29 (27%) | 20 (29%)  |
|  Bachelor’s Degree | 74 (42%) | 43 (40%) | 31 (44%)  |
|  Advanced Degree | 55 (31%) | 36 (33%) | 19 (27%)  |
|  Annual Household Income (n (%))a |  |  |   |
|  <50K | 14 (9%) | 7 (7%) | 7 (11%)  |
|  50K to <100K | 43 (27%) | 22 (23%) | 21 (32%)  |
|  ≥100K | 105 (65%) | 68 (70%) | 37 (57%)  |
|  Health Insurance (n (%))a |  |  |   |
|  Private | 143 (81%) | 84 (78%) | 59 (86%)  |
|  Medicare | 8 (5%) | 2 (2%) | 6 (9%)  |
|  Medicaid | 18 (10%) | 17 (16%) | 1 (1%)  |
|  Other Government | 8 (5%) | 5 (5%) | 3 (4%)  |
|  None | 0 (0%) | 0 (0%) | 0 (0%)  |

a 2 pediatric participants were missing ethnicity, 4 pediatric participants were missing race, 1 pediatric participant was missing education, 12 pediatric and 5 adult participants were missing income, and 1 pediatric and 1 adult participant were missing insurance.
b For pediatric participants, highest education of parent/guardian reported.

Participant Diabetes History

|   | Overall N=179 | Pediatrics N=109 | Adults N=70  |
| --- | --- | --- | --- |
|  Diabetes Duration at Enrollment (years) |  |  |   |
|  mean ± SD | 14 ± 13 | 6 ± 3 | 27 ± 12  |
|  Range | 1 to 57 | 1 to 15 | 3 to 57  |
|  Baseline HbA1c at End of Humalog Lead-In (%) |  |  |   |
|  mean ± SD | 7.2 ± 0.9 | 7.2 ± 0.9 | 7.1 ± 0.9  |
|  Range | 5.5 to 10.3 | 5.5 to 10.3 | 5.5 to 8.9  |
|  Humalog Period Total Daily Insulin (U) |  |  |   |

K250798 - Page 7 of 14

{7}

K250798 - Page 8 of 14

|   | Overall
N=179 | Pediatrics
N=109 | Adults
N=70  |
| --- | --- | --- | --- |
|  mean ± SD | 56 ± 32 | 52 ± 29 | 61 ± 36  |
|  Range | 14 to 222 | 14 to 167 | 22 to 222  |
|  Humalog Period Total Daily Insulin (U/kg) |  |  |   |
|  mean ± SD | 0.86 ± 0.32 | 0.99 ± 0.30 | 0.67 ± 0.26  |
|  Range | 0.32 to 2.39 | 0.38 to 2.39 | 0.32 to 1.62  |

## Safety Results

The primary endpoint evaluated the total number of safety events (severe hypoglycemia, diabetic ketoacidosis, unanticipated adverse device effects, adverse drug reactions, and other serious adverse events), number and proportion of participants who experienced at least one safety event, and the rate of events per 100 person years. The proportion of severe hypoglycemia (SH) and diabetic ketoacidosis (DKA) events were compared to proportions of events occurring in a similar time frame among participants enrolled in the T1D exchange. All participants who initiated the Lyumjev treatment and provided at least 24 hours of CGM data were included in the analyses.

During the Lyumjev Treatment Period there were 3 severe hypoglycemic events, 0 DKA events, 3 infusion site reactions reported as adverse events, 2 other adverse drug reactions reported as adverse events, and 2 serious adverse events other than severe hypoglycemia and DKA. In the Lyumjev Treatment Period, the overall percentage of participants with at least 1 severe hypoglycemia event requiring third party assistance was 1.7% and the percentage of participants with severe hypoglycemia associated with seizure or loss of consciousness was 0.6% compared with 6.1% with at least one severe hypoglycemia event associated with seizure or loss of consciousness during a 3-month period in the T1D Exchange registry when matched on age (P=0.07 and P=0.02, respectively, favoring Lyumjev). For DKA, the overall percentage of participants with at least 1 DKA event was 0% in the Lyumjev treatment period compared with 2.8% during a 3-month period in the T1D Exchange registry (P=0.04 favoring Lyumjev).

|   | Pediatrics (6-17 Years) |   | Adults (18+ Years)  |   |
| --- | --- | --- | --- | --- |
|   | Humalog Lead-In Period (N=109)
# Events/# Participants with ≥1 Event | Lyumjev Treatment Period (N=109)
# Events/# Participants with ≥1 Event | Humalog Lead-In Period (N=70)
# Events/# Participants with ≥1 Event | Lyumjev Treatment Period (N=70)
# Events/# Participants with ≥1 Event  |
|  All Adverse Events | 25/19 | 124/64 | 8/7 | 61/37  |
|  Severe Hypoglycemia Events | 0/0 | 2/2 | 0/0 | 1/1  |
|  Diabetic Ketoacidosis Events | 0/0 | 0/0 | 0/0 | 0/0  |
|  Other Serious Adverse Events | 0/0 | 1/1 | 0/0 | 1/1  |

{8}

K250798 - Page 9 of 14

|   | Pediatrics (6-17 Years) |   | Adults (18+ Years)  |   |
| --- | --- | --- | --- | --- |
|   | Humalog Lead-In Period (N=109) # Events/# Participants with ≥1 Event | Lyumjev Treatment Period (N=109) # Events/# Participants with ≥1 Event | Humalog Lead-In Period (N=70) # Events/# Participants with ≥1 Event | Lyumjev Treatment Period (N=70) # Events/# Participants with ≥1 Event  |
|  Other Adverse Events (Non-serious) | 25/19 | 121/64 | 8/7 | 59/37  |
|  Hyperglycemia with or without ketosis related to study device | 7/6 | 23/20 | 1/1 | 17/12  |
|  Hyperglycemia with or without ketosis not related to study device | 2/2 | 4/4 | 2/2 | 2/1  |
|  Nonsevere hypoglycemia | 0/0 | 0/0 | 0/0 | 0/0  |
|  Nonglycemic adverse events | 16/15 | 94/52 | 5/4 | 40/29  |
|   | Overall N=179 | Pediatrics (6-17 Years) N=109 | Adults (18+ Years) N=70  |
| --- | --- | --- | --- |
|  Severe Hypoglycemic (SH) Events  |   |   |   |
|  Number of SH Events per Participant |  |  |   |
|  0 | 176 | 107 | 69  |
|  1 | 3 | 2 | 1  |
|  2 | 0 | 0 | 0  |
|  Incidence Rate per 100 Person-Years | 6.6 | 7.1 | 5.8  |
|  % of Participants with ≥1 SH event in Lyumjev Treatment Period | 1.7% | 1.8% | 1.4%  |
|  % of Participants with ≥1 SH event in T1D Exchange a | 6.1% | 4.8% | 7.4%  |
|  P-Value Comparing % of Participants with ≥1 SH event in Current Study vs. T1D Exchange | P = 0.07 | P = 0.17 | P = 0.11  |
|  Diabetic Ketoacidosis (DKA) Events  |   |   |   |
|  Number of DKA Events per Participant |  |  |   |
|  0 | 179 | 109 | 70  |

{9}

K250798 - Page 10 of 14

|  1 | 0 | 0 | 0  |
| --- | --- | --- | --- |
|  2 | 0 | 0 | 0  |
|  Incidence Rate per 100 Person-Years | 0.0 | 0.0 | 0.0  |
|  % of Participants with ≥1 DKA event: Control IQ+Lyumjev | 0.0% | 0.0% | 0.0%  |
|  % of Participants with ≥1 DKA event: T1D Exchange^{a} | 2.8% | 3.2% | 2.5%  |
|  P-Value Comparing % of Participants with ≥1 DKA event in Control IQ+Lyumjev vs. T1D Exchange | P = 0.04 | P = 0.09 | P = 0.25  |
|  Lyumjev Adverse Drug Reactions (Not Including Infusion Site Reactions)  |   |   |   |
|  Total Events | 2 | 1 | 1  |
|  Incidence Rate per 100 Person-Years | 4.4 | 3.6 | 5.8  |
|  % of Participants with ≥1 Event | 1.1% | 0.9% | 1.4%  |
|  Infusion Site Reactions during Lyumjev Treatment Period  |   |   |   |
|  Total Infusion Site Reactions
(# events/# participants with ≥1 event) | 445 / 101 | 153 / 55 | 292 / 46  |
|  Infusion Site Reactions Resulting in Adverse Event | 3 (1%) | 2 (1%) | 1 (<1%)  |
|  Infusion Site Reactions Associated with Study Discontinuation | 2 (<1%) | 1 (1%) | 1 (<1%)  |
|  Infusion Site Reactions per Participant |  |  |   |
|  0 | 78 (44%) | 54 (50%) | 24 (34%)  |
|  1 | 37 (21%) | 24 (22%) | 13 (19%)  |
|  ≥2 | 64 (36%) | 31 (28%) | 33 (47%)  |

a T1D Exchange frequency reported from the following published article: Foster et al. State of type 1 diabetes management and outcomes from the T1D Exchange in 2016-2018. Diabetes Technol Ther 2019; 21: 66-72. In the T1D Exchange, an SH event required loss of consciousness or seizure and a DKA event required a diagnosis by a doctor and an overnight hospitalization. The overall percentage of participants with an event in the T1D Exchange was derived weighting by age group in the TL1 study.

Secondary safety endpoints included CGM hypoglycemia outcomes percentage of time &lt;54 mg/dL, percentage of time &lt;70 mg/dL, and rate of hypoglycemia events. Secondary safety CGM metrics comparing the Humalog lead-in period and Lyumjev treatment period by age group are shown below.

{10}

|   | Pediatrics (6-17 Years) |   | Adults (18+ Years)  |   |
| --- | --- | --- | --- | --- |
|   | Humalog Lead-In Period (N=109) Mean± SD | Lyumjev Treatment Period (N=109) Mean± SD | Humalog Lead-In Period (N=70) Mean± SD | Lyumjev Treatment Period (N=70) Mean± SD  |
|  24 Hours  |   |   |   |   |
|  % Time < 54 mg/dL | 0.28% ± 0.34% | 0.31% ± 0.28% | 0.26% ± 0.38% | 0.19% ± 0.21%  |
|  % Time < 70 mg/dL | 1.4% ± 1.3% | 1.4% ± 1.0% | 1.5% ± 1.3% | 1.0% ± 1.0%  |
|  CGM Hypoglycemic Event Rate per Weeka | 0.46 ± 0.78 | 0.56 ± 0.59 | 0.46 ± 0.72 | 0.37 ± 0.45  |
|  Daytime (6:00 AM – 11:59 PM)  |   |   |   |   |
|  % Time < 54 mg/dL | 0.27% ± 0.35% | 0.32% ± 0.30% | 0.24% ± 0.40% | 0.18% ± 0.23%  |
|  % Time < 70 mg/dL | 1.6% ± 1.5% | 1.5% ± 1.1% | 1.3% ± 1.2% | 1.1% ± 1.0%  |
|  CGM Hypoglycemic Event Rate per Weeka | 0.30 ± 0.58 | 0.44 ± 0.49 | 0.34 ± 0.61 | 0.28 ± 0.36  |
|  Nighttime (12:00 AM – 5:59 AM)  |   |   |   |   |
|  % Time < 54 mg/dL | 0.21% ± 0.46% | 0.26% ± 0.31% | 0.28% ± 0.51% | 0.17% ± 0.21%  |
|  % Time < 70 mg/dL | 1.0% ± 1.3% | 0.9% ± 0.9% | 1.6% ± 1.8% | 0.8% ± 0.8%  |
|  CGM Hypoglycemic Event Rate per Weeka | 0.10 ± 0.26 | 0.10 ± 0.15 | 0.09 ± 0.23 | 0.08 ± 0.12  |
|  Postprandial % Time < 54 mg/dLb | N=108 | N=108 | N=66 | N=66  |
|  ≤1 Hour | 0.16% ± 0.37% | 0.42% ± 0.52% | 0.14% ± 0.47% | 0.16% ± 0.27%  |
|  ≤2 Hours | 0.31% ± 0.55% | 0.48% ± 0.52% | 0.15% ± 0.38% | 0.22% ± 0.33%  |
|  >1-≤2 Hour | 0.39% ± 0.71% | 0.50% ± 0.57% | 0.12% ± 0.34% | 0.26% ± 0.45%  |
|  >2-≤4 Hour | 0.18% ± 0.40% | 0.22% ± 0.29% | 0.28% ± 0.66% | 0.19% ± 0.30%  |
|  ≤4 Hours | 0.27% ± 0.42% | 0.36% ± 0.37% | 0.23% ± 0.48% | 0.21% ± 0.33%  |
|  Postprandial % Time < 70 mg/dLb | N=108 | N=108 | N=66 | N=66  |
|  ≤1 Hour | 1.2% ± 1.6% | 1.9% ± 1.8% | 0.7% ± 1.6% | 1.0% ± 1.3%  |
|  ≤2 Hours | 1.7% ± 1.9% | 2.4% ± 1.9% | 1.0% ± 1.5% | 1.3% ± 1.5%  |
|  >1-≤2 Hour | 2.1% ± 2.5% | 2.8% ± 2.3% | 1.3% ± 1.9% | 1.7% ± 1.8%  |
|  >2-≤4 Hour | 1.5% ± 1.7% | 1.3% ± 1.3% | 1.3% ± 1.7% | 1.1% ± 1.2%  |
|  ≤4 Hours | 1.6% ± 1.6% | 1.9% ± 1.4% | 1.2% ± 1.5% | 1.3% ± 1.3%  |

a Analytic Definition of a CGM-Measured Hypoglycemic Event: A hypoglycemic event was defined as 15 consecutive minutes with a sensor glucose value  $&lt; 54\mathrm{mg / dL}$ . The end of the

K250798 - Page 11 of 14

{11}

hypoglycemic event was defined as a minimum of 15 consecutive minutes with a sensor glucose concentration ≥70 mg/dL. When a hypoglycemic event ended, the study participant became eligible for a new event.

b Postprandial periods denote the time from a non-zero carb entry. 1 pediatric and 4 adult participants did not have sufficient postprandial CGM data to be analyzed.

Exploratory analyses included various CGM-based measures of glycemic control, HbA1c, and insulin delivery metrics, and patient-reported outcome questionnaires. Exploratory CGM metrics calculated during the daytime (06:00 AM to 11:59 PM), nighttime (12:00 AM to 5:59 AM), and overall, are summarized by age group below.

|   | Pediatrics (6-17 Years) |   | Adults (18+ Years)  |   |
| --- | --- | --- | --- | --- |
|   | Humalog Lead-In Period (N=109) Mean± SD | Lyumjev Treatment Period (N=109) Mean± SD | Humalog Lead-In Period (N=70) Mean± SD | Lyumjev Treatment Period (N=70) Mean± SD  |
|  24 Hours  |   |   |   |   |
|  Time in Range 70-180 mg/dL (%) | 62% ± 15% | 64% ± 12% | 69% ± 13% | 70% ± 13%  |
|  Time in Range 70-140 mg/dL (%) | 39% ± 15% | 40% ± 13% | 43% ± 12% | 45% ± 13%  |
|  Time >180 mg/dL (%) | 36% ± 15% | 34% ± 13% | 29% ± 14% | 28% ± 13%  |
|  Time >250 mg/dL (%) | 14% ± 10% | 13% ± 9% | 8% ± 7% | 8% ± 7%  |
|  Mean Glucose (mg/dL) | 172 ± 28 | 169 ± 24 | 159 ± 20 | 159 ± 20  |
|  Glucose SD (mg/dL) | 64 ± 16 | 64 ± 14 | 55 ± 12 | 53 ± 12  |
|  Glucose CV (%) | 37% ± 6% | 37% ± 5% | 34% ± 5% | 33% ± 4%  |
|  CGM Hyperglycemic Event Rate per Weeka | 2.1 ± 2.2 | 2.0 ± 1.8 | 1.2 ± 1.2 | 1.0 ± 1.1  |
|  Daytime (6:00 AM – 11:59 PM)  |   |   |   |   |
|  Time in Range 70-180 mg/dL (%) | 59% ± 15% | 61% ± 13% | 68% ± 14% | 69% ± 13%  |
|  Time in Range 70-140 mg/dL (%) | 36% ± 14% | 37% ± 13% | 42% ± 12% | 44% ± 13%  |
|  Time >180 mg/dL (%) | 40% ± 16% | 38% ± 14% | 31% ± 14% | 29% ± 13%  |
|  Time >250 mg/dL (%) | 16% ± 11% | 15% ± 10% | 8% ± 7% | 8% ± 7%  |
|  Mean Glucose (mg/dL) | 177 ± 29 | 174 ± 26 | 161 ± 21 | 160 ± 20  |
|  Glucose SD (mg/dL) | 66 ± 16 | 66 ± 14 | 55 ± 11 | 54 ± 11  |
|  Glucose CV (%) | 37% ± 5% | 38% ± 5% | 34% ± 5% | 33% ± 4%  |

a: 1 pediatric and 4 adult participants were eligible for a new event.
b: Postprandial period 12:00 AM to 5:59 AM was not included in the study.

K250798 - Page 12 of 14

{12}

K250798 - Page 13 of 14

|   | Pediatrics (6-17 Years) |   | Adults (18+ Years)  |   |
| --- | --- | --- | --- | --- |
|   | Humalog Lead-In Period (N=109) Mean± SD | Lyumjev Treatment Period (N=109) Mean± SD | Humalog Lead-In Period (N=70) Mean± SD | Lyumjev Treatment Period (N=70) Mean± SD  |
|  CGM Hyperglycemic Event Rate per Weeka | 1.9 ± 2.1 | 1.8 ± 1.7 | 0.9 ± 1.1 | 0.8 ± 0.9  |
|  Nighttime (12:00 AM – 5:59 AM)  |   |   |   |   |
|  Time in Range 70-180 mg/dL (%) | 73% ± 18% | 75% ± 14% | 72% ± 17% | 73% ± 16%  |
|  Time in Range 70-140 mg/dL (%) | 47% ± 21% | 50% ± 18% | 47% ± 18% | 48% ± 18%  |
|  Time >180 mg/dL (%) | 26% ± 19% | 24% ± 14% | 26% ± 17% | 26% ± 16%  |
|  Time >250 mg/dL (%) | 9% ± 10% | 8% ± 7% | 8% ± 9% | 7% ± 7%  |
|  Mean Glucose (mg/dL) | 158 ± 30 | 155 ± 23 | 156 ± 25 | 156 ± 23  |
|  Glucose SD (mg/dL) | 51 ± 19 | 53 ± 15 | 52 ± 17 | 50 ± 14  |
|  Glucose CV (%) | 32% ± 8% | 34% ± 6% | 33% ± 8% | 32% ± 6%  |
|  CGM Hyperglycemic Event Rate per Weeka | 0.4 ± 0.6 | 0.4 ± 0.4 | 0.3 ± 0.6 | 0.3 ± 0.5  |

a Analytic definition of a CGM-Measured Hyperglycemic Event: A hyperglycemic event &gt; 300 mg/dL is defined as spending a cumulative 90 minutes or more &gt; 300 mg/dL in a 120-minute window. The end of the hyperglycemic event was defined as a minimum of 15 consecutive minutes with a sensor glucose concentration &lt; 180 mg/dL. When a hyperglycemic event ended, the study participant became eligible for a new event.

The change in central lab HbA1c from the end of the Humalog lead-in period to the end of the Lyumjev treatment period is summarized by age group below.

|   | Pediatrics (6-17 Years) |   | Adults (18+ Years)  |   |
| --- | --- | --- | --- | --- |
|   | End of Humalog Lead-In Period (N=109) Mean± SD | End of Lyumjev Treatment Period (N=109) Mean± SD | End of Humalog Lead-In Period (N=70) Mean± SD | End of Lyumjev Treatment Period (N=70) Mean± SD  |
|  HbA1c (%) | 7.2 ± 0.9 | 7.1 ± 0.8 | 6.9 ± 0.8 | 6.7 ± 0.7  |

## Challenge Results

During this clinical study, participants completed one meal challenge in the Humalog Lead-in Period and three meal challenges in the Lyumjev Treatment Period. Participants were instructed

{13}

to consume at least 50 grams of carbohydrates and either no bolus, a half bolus, or a full bolus was delivered. There were no CGM-measured hypoglycemic events (defined as at least 15 consecutive minutes &lt;54 mg/dL) in the first three hours after the challenge in either the Humalog or Lyumjev periods. For the no bolus challenge, there were no significant differences in hypoglycemic event rates or any CGM metrics between periods in the first 3 hours post-meal, 4-5 hours post-meal, or overnight periods. For the full bolus and half bolus meal challenges, the percentage of hypoglycemic events was low during each time period.

One exercise challenge was completed by participants during the Humalog lead-in period and three exercise challenges were completed during the Lyumjev treatment period. Exercise challenges were 2 hours with at least 1 hour of moderate intensity exercise (remainder could be lower intensity). One percent of participants during the Humalog period and 2%, 1%, and 0% of participants during the 3 Lyumjev exercise challenges had at least 1 CGM hypoglycemic event &lt; 54 mg/dL in the first 2 hours after starting exercise.

## Clinical Literature Review

Based on existing literature of clinical studies assessing the safety and effectiveness of Lyumjev insulin, Control-IQ technology, and Control-IQ+ technology, the sponsor provided a justification for extrapolating the results of the pivotal clinical study of Control-IQ+ with Lyumjev described above (which included persons with Type 1 Diabetes ages 6 years and older) to support that Lyumjev U-100 insulin can be safely used with Control-IQ+ technology by adults 18 years and older with Type 2 Diabetes and children 2-5 years old with Type 1 Diabetes. The justification provided supports the proposed extrapolation.

## Predetermined Change Control Plan (PCCP)

A predetermined change control plan (PCCP) for implementing the proposed labeling modifications to the device User Guide and Quick Start Guide upon clearance of a compatible ACE pump was provided in this submission. The PCCP includes a description of the labeling modifications, a modification protocol, and an impact assessment, and was determined to be adequate to support and clearly specify expectations, requirements, and specifications for updating the subject device labeling upon the clearance of a compatible ACE pump. Control-IQ+ Technology, with the proposed updated labeling, must not be distributed until the pre-specified acceptance criteria in the PCCP are met. Following implementation, new users will receive the updated Quick Start Guide upon receipt of their corresponding device, and the updated User Guide will be available electronically, or physically upon request. Existing users will be notified via email about the additional insulin compatible with the device and will have the option to either electronically download the updated User Guide and Quick Start Guide or request a physical version.

## VII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

## VIII Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K250798 - Page 14 of 14

---

**Source:** [https://fda.innolitics.com/device/K250798](https://fda.innolitics.com/device/K250798)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
