The SPROTTE® STANDARD (LUER/ NRFit®) lumbar puncture needles are intended to gain entry into or puncture the spinal cavity permitting injection/ withdrawal of fluids for purposes of diagnostic lumbar puncture, myelography/ discography.
Device Story
Single-use anesthesia conduction needle; stainless-steel tube with atraumatic, non-cutting, symmetrical pencil-point tip and side-hole opening. Available with LUER (ISO 80369-7) or NRFit (ISO 80369-6) hub; includes stylet and optional introducer. Used by clinicians to access spinal cavity for fluid injection/withdrawal. Device facilitates diagnostic procedures; pencil-point tip design intended to reduce trauma. Sterilized via Ethylene Oxide; supplied sterile. Packaging options include Tyvek/foil or medical paper/foil. Shelf life 60 months.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via ISO 7864 (cleanliness, tube integrity, bond strength), ISO 9626 (stiffness, breakage, corrosion resistance), and ISO 80369-6/7 (leakage, stress cracking, axial load). Sterilization validated per ISO 11135 (SAL 10^-6). Biocompatibility per ISO 10993-1.
Technological Characteristics
Stainless steel 1.4301 (tube/introducer) and 1.4310 (stylet). Hub materials: PC (Lexan 164R-112) and Copolyester (Tritan MX731). Bonding: Epoxy resin (Araldite 2011) or UV adhesive (Dymax 1406-M). Connectivity: ISO 80369-6 (NRFit) or ISO 80369-7 (LUER). Sterilization: Ethylene Oxide (SAL 10^-6). Dimensions: 18G-22G diameter, 90mm-200mm length.
Indications for Use
Indicated for adult and pediatric patients requiring entry into or puncture of the spinal cavity for diagnostic lumbar puncture, myelography, or discography, including injection or withdrawal of fluids.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 25, 2025
PAJUNK GmbH Medizintechnologie
Chiara Meyer
Specialist Regulatory Affairs
Karl-Hall-Str. 1
Pajunkstr. 1
Geisingen, BW 78187
Germany
Re: K250774
Trade/Device Name: SPROTTE® STANDARD (LUER/ NRFit®) Lumbar Puncture
Regulation Number: 21 CFR 868.5150
Regulation Name: Anesthesia Conduction Needle
Regulatory Class: Class II
Product Code: BSP
Dated: March 14, 2025
Received: March 14, 2025
Dear Chiara Meyer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250774 - Chiara Meyer
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250774 - Chiara Meyer
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250774 | |
| Device Name SPROTTE® STANDARD (LUER/ NRFit®) Lumbar Puncture | |
| Indications for Use (Describe) The SPROTTE® STANDARD (LUER/ NRFit®) lumbar puncture needles are intended to gain entry into or puncture the spinal cavity permitting injection/ withdrawal of fluids for purposes of diagnostic lumbar puncture, myelography/ discography. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
PAJUNK®
USA
08 510(k) Summary
# USA
# 08 510(k) Summary
# Registrations USA
510(k) Summary as required by 21 CFR 807.92(c).
Date of Preparation: 2025-06-18
Document Control Number: K250774
510(k) owner:
PAJUNK® GmbH Medizintechnologie
Karl-Hall-Str. 1/ Pajunkstr. 1
78187 Geisingen
Baden-Wuerttemberg, Germany
Fon: +49(0)7704-9291-586
Fax: +49(0)7704-9291-605
Establishment Registration Number: 9611612
Submitter Information/ production site:
PAJUNK® GmbH Medizintechnologie
Karl-Hall-Str. 1/ Pajunkstr. 1
78187 Geisingen
Baden-Wuerttemberg, Germany
Fon: +49(0)7704-9291-586
Fax: +49(0)7704-9291-605
Establishment Registration Number: 9611612
Manufacturer/ Submitter Contact:
Christian G. H. Quass
Director Regulatory Affairs, Safety Officer
Fon: +49(0)7704-9291-586
Fax: +49(0)7704-9291-602
E-Mail: christian.quass@pajunk.com
Alternative Manufacturer/ Submitter Contact:
Chiara Meyer
Regulatory Affairs Specialist
Fon: +49(0)7704-9291-627
Fax: +49(0)7704-9291-602
E-Mail: chiara.meyer@pajunk.com
USA Contact:
PAJUNK MEDICAL SYSTEMS
4575 Marconi Dr.
Alpharetta, GA 30005
Registration Number: 3004076349
FEI Number*: 3004076349
US-Official Correspondent
Marco Wohnig
President
PAJUNK Medical Systems
4575 Marconi Dr.
Alpharetta, GA 30005
{5}
PAJUNK®
USA
08 510(k) Summary
Phone: 001-770-4936832
E-Mail: Marco.Wohnig@pajunk-usa.com
Contract Sterilizer:
Sterigenics Germany GmbH
Kastelerstraße 45
65203 Wiesbaden
Germany, Hessen
Establishment Registration Number: 3002807090
2nd Contract Sterilizer:
HA2 MEDIZINTECHNIK GMBH
Am Bahndamm 11
Halberstadt Saxony-Anhalt, DE 38820
Registration Number: 3009039068
FEI Number*: 3009039068
Device Name and Classification
| Device Name: | SPROTTE® STANDARD (LUER/ NRFit®) lumbar puncture needles |
| --- | --- |
| Premarket Notification Number: | K250774 |
| Classification Name: | Anesthesia Conduction Needle |
| Classification Reference: | 21 CFR § 868.5150 |
| Product Code: | BSP |
| Establishment Registration Number: | 9611612 |
| Regulatory Class: | II |
| Panel: | Anesthesiology |
| Sterilization Method: | Ethylene Oxide disposable device, supplied sterile to the end user and non-sterile intended to be sterilized prior to use to re-packagers/medical device manufacturers |
| Contract Sterilizer: | Sterigenics Germany GmbH
Kastelerstraße 45
65203 Wiesbaden
Germany, Hessen
Establishment Registration Number: 3002807090 |
| Additional Contract Sterilizer: | HA2 MEDIZINTECHNIK GMBH
Am Bahndamm 11
Halberstadt Saxony-Anhalt, DE 38820 |
2 | Page
{6}
PAJUNK®
USA
08 510(k) Summary
| Registration Number: 3009039068 |
| --- |
| FEI Number*: 3009039068 |
## Changes to Device: Predicate Device
| Device Name: | Sprotte NRFit™, Quincke NRFit™ |
| --- | --- |
| Premarket Notification Number: | K160294 |
| Classification Name: | Anesthesia Conduction Needle |
| Classification Reference: | 21 CFR § 868.5150 |
| Product Code: | BSP |
| Establishment Registration Number: | 9611612 |
| Regulatory Class: | II |
| Panel: | Anesthesiology |
| Sterilization Method: | Ethylene Oxide disposable device, supplied sterile to the end user and non-sterile intended to be sterilized prior to use to re-packagers/medical device manufacturers |
| Contract Sterilizer: | Sterigenics Germany GmbH
Kastelerstraße 45
65203 Wiesbaden
Germany, Hessen
Establishment Registration Number: 3002807090 |
## Changes to Device: Reference Device
| Device Name: | SPROTTE® STANDARD (LUER/ NRFit®) |
| --- | --- |
| Premarket Notification Number: | K241953 |
| Classification Name: | Anesthesia Conduction Needle |
3 | Page
{7}
PAJUNK®
USA
08 510(k) Summary
| Classification Reference: | 21 CFR § 868.5150 |
| --- | --- |
| Product Code: | BSP |
| Establishment Registration Number: | 9611612 |
| Regulatory Class: | II |
| Panel: | Anesthesiology |
| Sterilization Method: | Ethylene Oxide
disposable device, supplied sterile to the end user and non-sterile intended to be sterilized prior to use to re-packagers/ medical device manufacturers |
| Contract Sterilizer: | Sterigenics Germany GmbH
Kasteler straße 45
65203 Wiesbaden
Germany, Hessen
Establishment Registration Number: 3002807090 |
| Additional Contract Sterilizer: | HA2 MEDIZINTECHNIK GMBH
Am Bahndamm 11
Halberstadt Saxony-Anhalt, DE 38820
Registration Number: 3009039068
FEI Number*: 3009039068 |
4 | Page
{8}
PAJUNK
USA
08 510(k) Summary
# 1 Narrative Device Description
The subject device, the SPROTTE® STANDARD (LUER/ NRFit®) anaesthesia conduction needle is a single-use anaesthesia conducting intended for administer anesthetic agent to the spinal space.
The SPROTTE® cannulas are equipped with a stylet as well as optional with an Introducer. It is available with a LUER respectively with a NRFit® hub. The LUER is either a standard hub, a magnifying hub or a 2.G hub. The cannula tube is straight. The distal connection of the hub is either equipped with a LUER Connector according to ISO 80369-7 or a NRFit-Connector according to ISO 80369-6.

# 1.1 Differences among models
## SPROTTE® STANDARD
The SPROTTE® cannulas consist of a stainless-steel tube, a Tritan hub and are equipped with a stylet. The cannulas have a LUER hub. The LUER hub is either a standard hub or magnifying. The cannula has a atraumatic, noncutting, symmetrical pencil point tip with a side-hole opening. The cannula tube is straight.
The different technical specifications are the diameter, the length and if it comes with an introducer. The diameter has a range from 18G-22G. The length has a range from 90mm-200mm. The introducer is optional. It comes in a diameter range from 1,0mm-1,65mm and a length range from 30mm-40mm
## SPROTTE® STANDARD NRFit®
The SPROTTE® cannulas consist of a stainless steel tube, a Tritan hub and are equipped with a stylet. The cannulas have a NRFit® hub. The NRFit® hub is a magnifying hub. The cannula has a atraumatic, noncutting, symmetrical pencil point tip with a side-hole opening. The cannula tube is straight.
{9}
PAJUNK
USA
08 510(k) Summary
The different technical specifications are the diameter, the length and if it comes with an introducer. The diameter has a range from 20-22G. The length has a range from 103mm-150mm. The introducer is optional. It comes in a diameter range from 1,00mm-1,20mm and a length range from 30mm-40mm
## 2 Determination of Substantial Equivalence
### Intended Use Predicate Device (510(K) Predicate)
The SPROTTE® NRFit™, Quincke NRFit™ lumbar puncture needles are intended to gain entry into or puncture the spinal cavity permitting injection (including anesthesia) / withdrawal of fluids for purposes of diagnostic lumbar puncture, myelography/ discography procedures.
The device is intended for adult and pediatric patients.
### Intended Use Subject Device
The SPROTTE® STANDARD (LUER/ NRFit®) lumbar puncture needles are intended to gain entry into or puncture the spinal cavity permitting injection/ withdrawal of fluids for purposes of diagnostic lumbar puncture, myelography/ discography.
### Discussion
The Intended Use of the Predicate device and of the subject device is substantially equivalent.
Conclusion: substantially equivalent
## 3 Technical Description
Substantial Equivalence: Tabular Outline
{10}
PAJUNK®
USA
08 510(k) Summary
| Characteristics | Subject Device
Sprotte Standard (Luer/NRFit)
Manufacturer: PAJUNK® GmbH
Medizintechnologie, Geisingen | Predicate Device K160294
Sprotte NRFit
Manufacturer: PAJUNK® GmbH
Medizintechnologie, Geisingen | Reference Device K241953
Sprotte Standard (Luer/NRFit)
Manufacturer: PAJUNK® GmbH Medizintechnologie, Geisingen |
| --- | --- | --- | --- |
| Picture | | | |
| Needle Tubing | Stainless steel 1.4301 | Stainless steel 1.4301 | Stainless steel 1.4301 |
| Stylet | Stainless steel 1.4310 | Stainless steel 1.4310 | Stainless steel 1.4310 |
| Introducer Tubing | Stainless steel 1.4301 | Stainless steel 1.4301 | Stainless steel 1.4301 |
| Tip Cannula | Sprotte tip | Sprotte tip | Sprotte tip |
| Tip Introducer | Facet tip | Facet tip | Facet tip |
| Cannula Hub | PC (Lexan 164R-112)
Copolyester (Tritan MX731) | PC (Lexan 164R-112)
Copolyester (Tritan MX731) | PC (Lexan 164R-112)
Copolyester (Tritan MX731) |
| Introducer hub | Copolyester (Tritan MX731) | PC (Lexan 164R-112) | Copolyester (Tritan MX731) |
| Stylet Knob | PC (Lexan 164R-112) | PC (Lexan 164R-112) | PC (Lexan 164R-112) |
| Graduation | For cannulas 100mm length and longer | For cannulas 100mm length and longer | For cannulas 100mm length and longer |
7 | Page
{11}
PAJUNK®
USA
08 510(k) Summary
| Bonding Technology Needle-to-Hub Cannula | Adhesive bonding | Adhesive bonding | Adhesive bonding |
| --- | --- | --- | --- |
| Bonding Technology Needle-to-Hub Introducer | Adhesive bonding Overmolding | Adhesive bonding | Adhesive bonding Overmolding |
| Glue Cannula | Epoxy resin (Araldite 2011) UV Adhesive (Dymax 1406-M) | Epoxy resin (Araldite 2011) UV Adhesive (Dymax 1406-M) | Epoxy resin (Araldite 2011) UV Adhesive (Dymax 1406-M) |
| Glue Stylet | Epoxy resin (Araldite 2011) | Epoxy resin (Araldite 2011) | Epoxy resin (Araldite 2011) |
| Glue Introducer | Epoxy resin (Araldite 2011) UV Adhesive (Dymax 1406-M) | Epoxy resin (Araldite 2011) | Epoxy resin (Araldite 2011) UV Adhesive (Dymax 1406-M) |
| Diameter | 18-22G | 18G-29G | 22G-29G |
| Length | 90mm-200mm | 50-150 | 25-200mm |
| Connectivity | ISO 80369-6 (NRFit) ISO 80369-7 (LUER) | ISO 80369-6 (NRFit) | ISO 80369-6 (NRFit) ISO 80369-7 (LUER) |
| Packaging | Tyvek, Foil, chromo-duplex board Alternatively: Paper, Foil | Tyvek, Foil, chromo-duplex board | Tyvek, Foil, chromo-duplex board Alternatively: Paper, Foil |
| Biocompatibility | Complies with ISO10993-series as required by « Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" », Docket Number : FDA-2013-D-0350 | | |
8 | Page
{12}
PAJUNK
USA
08 510(k) Summary
| Sterilization Method | Ethyleneoxide Sterilization | Ethyleneoxide Sterilization | Ethyleneoxide Sterilization |
| --- | --- | --- | --- |
| Sterility Assurance Level | SAL=10-6 | SAL=10-6 | SAL=10-6 |
| Sterilization Service Provider | Serigenics, Wiesbaden
HA2 Medizintechnik, Halberstadt | Serigenics, Wiesbaden | Serigenics, Wiesbaden |
| Shelf Life | 60 month from sterilization | 60 month from sterilization | 60 month from sterilization |
# 4 Substantial Equivalence Discussion
The predicate device and the subject device have some characteristic differences:
a) diameter -> All offered diameters are within commonly used clinical ranges and comply with ISO 9626 standards for stainless steel needles. The variation in diameter allows customization for specific anatomical approaches but does not introduce new risks.
b) length -> The availability of multiple needle lengths allows the clinician to select the appropriate configuration based on patient size, depth of target, and anatomical location. All lengths are within clinically accepted ranges for peripheral and musculoskeletal nerve access and follow design precedents established in similar legally marketed needles.
The differences between the two predicate devices result from the area of application. In lumbar puncture, cerebrospinal fluid (CSF) is primarily taken or pressure is measured. A slightly larger inner diameter enables a faster and stronger CSF flow so that samples can be obtained in a reasonable amount of time. spinal anesthesia, local anesthetic is injected into the subarachnoid space. Because only relatively small volumes need to be administered, needles with a much finer outer diameter can be used.
The subject device can be either directly injection moulded or glued. The predicate device technology for hub-to-needle bonding is glueing. Both technologies - direct molding as well as glueing - are standard technologies.
Also the connectivity of the subject device can be either LUER or NRFit. The connectivity of the predicate device is NRFit. Both connectivity's - LUER as well as NRFit - are state of the art.
The predicate devices, the Sprotte NRFit and Quincke NRFit needles, are packed with Tyvek and foil and sterilized with Ethyleneoxide at Sterigenics, Germany.
9 | Page
{13}
PAJUNK
USA
08 510(k) Summary
The subject devices, the Sprotte STANDARD (LUER/ NRFit) needles, are packed with Tyvek and foil as well as optionally with Medical Paper and foil.
Furthermore, the subject devices are sterilized at Sterigenics in Wiesbaden as well as optionally at HA2 Medizintechnik in Halberstadt.
The main differences which are subject to this premarket submission are
a) the addition of a sterilization facility and
b) the addition of an alternative packaging material.
Both differences have been cleared in K241953 Sprotte STANDARD (LUER/ NRFit).
In order to verify substantial equivalence of the sterilization process a sterilization validation according ISO 11135 has been performed and successfully accomplished. The sterility assurance level of 10⁻⁶ has been successfully validated at both facilities: Sterigenics and HA2.
In order to verify substantial equivalence of the packaging exhaustive packaging and shelf-life tests have been performed. Both, medical paper and Tyvek have been validated and verified to allow sterilization using EtO and to maintain sterility after a shelf life of 60 month after sterilization.
# 5 Performance Testing
## 5.1 ISO 7864
The following sections have been tested:
| section | Pass / Fail |
| --- | --- |
| 4.3 Cleanliness | Passed |
| 4.4 Limits for acidity or alkalinity | Passed |
| 4.5 Limits for extractable metals | Passed |
| 4.10 Needle Tube | Passed |
| 4.12 Bond between hub and needle tube | Passed |
| 4.13 Patency of lumen | Passed |
## 5.2 ISO 9626
The following sections have been tested:
{14}
PAJUNK®
USA
08 510(k) Summary
| section | Pass / Fail |
| --- | --- |
| 5.2 Surface finish and visual appearance | Passed |
| 5.3 Cleanliness | Passed |
| 5.4 Limits for acidity and alkalinity | Passed |
| 5.5 Size designation | Passed |
| 5.6 Dimensions | Passed |
| 5.7 Sample size | Passed |
| 5.8 Stiffness | Passed |
| 5.9 Resistance to breakage | Passed |
| 5.10 Resistance to corrosion | Passed |
## 5.3 ISO 80369-6
The following sections been tested:
| section | |
| --- | --- |
| 6.1 Fluid Leakage | Passed |
| 6.2 Air Leakage | Passed |
| 6.3 Stress Cracking | Passed |
| 6.4 Separation Axial Load | Passed |
| 6.5 Unscrewing | Passed |
| 6.6 Overriding | Passed |
## 5.4 ISO 80369-7
The following sections been tested:
{15}
PAJUNK®
USA
08 510(k) Summary
| section | |
| --- | --- |
| 7.1 Fluid Leakage | Passed |
| 7.2 Air Leakage | Passed |
| 7.3 Stress Cracking | Passed |
| 7.4 Separation Axial Load | Passed |
| 7.5 Unscrewing | Passed |
| 7.6 Overriding | Passed |
# 6 Sterilization
The contract sterilizer and the sterilizing process are identical to the process and sterilizer used for all PAJUNK® - manufactured devices which are already cleared for market or exempt.
Sterilization parameters are:
| SAL | 10^{-6} |
| --- | --- |
| Type of gas | Ethylene Oxide 99,99% |
Sterilization has been validated according to ISO 11135-1 Overkill Approach (1 sublethal cycle, 2 half cycle, 1 full cycle)
Residuals of EO and ECH are in compliance with ISO 10993-7.
Cleaning and Sterilization method, which ensures an SAL of 10-6 as well as compliance with limits for chemical burden, bioburden, pyroburden (i.e. LAL) and EtO-residuals as well as shelf life have been validated and are safe and effective.
The limits listed below are met by each device:
Limits for Residuals: 25ppm = 25μg/(g/device) of Ethyleneoxide (EO); 25ppm = 25μg/(g/device) Ethylene chlorhydrine
12 | Page
{16}
PAJUNK
USA
08 510(k) Summary
# 7 Shelf Life
Efficiency of sterile product's lifecycle has been validated using process most challenging worst case devices.
Sterility tests have been performed using process most challenging worst case devices with similar characteristics made from the same material after 5 years. The devices were found to be sterile after 5 years, the sterile barrier system is efficient.
Performance of the device (LUER/ NRFit connection, stability of bonding connections, needle's bending rigidity) has been tested. There is no decrease in performance after 5 years.
Shelf-life is set to 5 years.
# 8 Biocompatibility
All products comply with ISO 10993-1 and with FDA's guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
Therefore, based on sterilization validation and residuals validation the kits also are considered to be biocompatible.
# 9 Technology Characteristics
Both, the subject device and the predicate device, consist of identical components.
# 10 Conclusion
The comparison between the predicate device and the subject device of this submission focused on the validated sterilization process and the shelf life testing demonstrates that the subject device can be optionally in addition packed using medical paper and can be optionally sterilized at the alternative sterilization service provider at an SAL of 10-6.
Based on current in-process testing regularly conducted, safe performance of the SPROTTE® STANDARD (LUER/ NRFit®) needle is demonstrated.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.