The cuff is compatible with non-invasive blood pressure monitors to measure blood pressure. It is intended for use by healthcare professionals (prescription use) and by laypersons in home settings (over-the-counter use). The cuff is designed for adult patients. It is non-sterile and reusable.
Device Story
Device is an inflatable arm cuff with bladder and hook-and-loop fastener; used with non-invasive blood pressure monitors. Operates via oscillometric method: cuff inflates/deflates to occlude/release arterial blood flow. Used by healthcare professionals or laypersons in clinical or home settings. Output is pressure data processed by the monitor to determine systolic/diastolic values. Benefits include accurate, reliable blood pressure monitoring for adults. Device is non-sterile and reusable.
Clinical Evidence
Clinical validation performed per ISO 81060-2:2018 across five cuff models (15–52 cm range). Compared device measurements against simultaneous auscultatory reference measurements using a mercury sphygmomanometer. Met Criterion 1 (mean difference ±5 mmHg, SD ≤8 mmHg) and Criterion 2 (per-subject distribution).
Technological Characteristics
Inflatable bladder with outer fabric and hook-and-loop fastener. Five sizes (15-52 cm range). Pressure range 0-330 mmHg. Max leakage < 4 mmHg/min. Biocompatibility per ISO 10993-5, -10, -23. Durability tested to 10,000 inflation cycles and 3,000 hook-and-loop closures. Non-sterile.
Indications for Use
Indicated for adult patients requiring non-invasive blood pressure measurement using compatible monitors. Suitable for both clinical (prescription) and home (OTC) settings.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 3, 2025
AViTA Corporation
Rex Hsieh
Regulatory Affairs
9F, No.78, Sec.1, Kwang-Fu Road, Sanchong Dist.
New Taipei, 24158
Taiwan
Re: K250765
Trade/Device Name: AViTA Arm Type Blood Pressure Monitor Cuff
Regulation Number: 21 CFR 870.1120
Regulation Name: Blood Pressure Cuff
Regulatory Class: Class II
Product Code: DXQ
Dated: August 4, 2025
Received: August 4, 2025
Dear Rex Hsieh:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250765 - Rex Hsieh
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250765 - Rex Hsieh
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh S. Deoras -S
for
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250765 | ? |
| Please provide the device trade name(s). | | ? |
| AViTA Arm Type Blood Pressure Monitor Cuff | | |
| Please provide your Indications for Use below. | | ? |
| The cuff is compatible with non-invasive blood pressure monitors to measure blood pressure.
It is intended for use by healthcare professionals (prescription use) and by laypersons in home settings (over-the-counter use).
The cuff is designed for adult patients. It is non-sterile and reusable. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
AViTA Arm Type Blood Pressure Monitor Cuff
{4}
K250765
# 510(k) Summary
This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92.
## 1 APPLICANT INFORMATION
Applicant: AViTA Corporation
Address: 9F, No.78, Sec.1, Kwang-Fu Rd., San-Chung District, New Taipei City 24158, Taiwan
Applicant Establishment Number: 9617543
Applicant Contact: Rex hsieh
Email: rex_hsieh@avita.com.tw
Phone Number: +886-2-8512-1568 ext.5306
Fax Number: +886-2-8512-1347
Date of Submission: Feb 24, 2025
## 2 SUBJECT DEVICE
Trade/Proprietary Name: AViTA Arm Type Blood Pressure Monitor Cuff
Common Name: AViTA Arm Type Blood Pressure Monitor Cuff
Review Panel: Cardiovascular
Classification Product Code: DXQ
Regulation Number: 870.1120
Device Class: II
## 3 PREDICATE DEVICE
510(k) Number: K231961
Trade Name: Reusable NIBP Cuff
Manufacturer: Shenzhen SINO-K Medical Technology Co., Ltd
Page 1 of 7
{5}
K250765
# 4 DEVICE DESCRIPTION
The primary components of the AViTA blood pressure cuff include an inflatable bladder and an outer fabric layer with a hook-and-loop fastening system. The cuff is designed to wrap securely around the patient's upper arm, allowing for accurate measurement of blood pressure when used with AViTA blood pressure monitors.
The AViTA blood pressure cuffs are available in five different sizes to accommodate a range of arm circumferences:
| Model. | Limb Range (Unit: cm) |
| --- | --- |
| 6NRUI-K | 22-42 cm |
| 6NRLI-K | 33-42 cm |
| 6NRXI-K | 42-52 cm |
| 6NRSI-K | 15-22 cm |
| 6NRMI-K | 22-33 cm |
During operation, the cuff inflates and deflates to occlude and release arterial blood flow, utilizing the oscillometric method to detect systolic and diastolic pressure. The cuff is constructed with durable materials to withstand repeated inflation cycles and ensure long-term performance.
The AViTA blood pressure cuff is designed for compatibility with AViTA blood pressure monitors, ensuring optimal performance and measurement accuracy.
# 5 INDICATIONS FOR USE
The cuff is compatible with non-invasive blood pressure monitors to measure blood pressure.
It is intended for use by healthcare professionals (prescription use) and by laypersons in home settings (over-the-counter use).
The cuff is designed for adult patients. It is non-sterile and reusable.
# 6 IDENTIFICATION OF THE PREDICATE DEVICE
| NO. | Device Name | Limb Range | Manufacture | Classification and Code | Classification regulation | 510K number |
| --- | --- | --- | --- | --- | --- | --- |
{6}
K250765
| Subject Device | AViTA Arm Type Blood Pressure Monitor Cuff | 15-52cm | AViTA Corporation | Class II DXQ | 870.1120 | TBD |
| --- | --- | --- | --- | --- | --- | --- |
| Predicate Device | Reusable NIBP Cuff | 6-66cm | Shenzhen SINO-K Medical Technology Co., Ltd | Class II DXQ | 870.1120 | K231961 |
# 7 NON-CLINICAL TESTING
A series of safety, essential performance, and biocompatibility tests were conducted to evaluate the safety and effectiveness of the AViTA Arm Type Blood Pressure Monitor Cuff. The tests listed below were conducted in accordance with:
- ISO 81060-1: Non-invasive sphygmomanometers – Part 1: Requirements and test methods for non-automated measurement type.
- ISO 10993-1: Biological Evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
- ISO 10993-5: Biological Evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity.
- ISO 10993-10: Biological evaluation of medical devices – Part 10: Tests for Sensitization.
- ISO 10993-23: Biological evaluation of medical devices – Part 23: Tests for irritation.
# 8 CLINICAL PERFORMANCE TESTING
Clinical validation of the subject device was conducted in accordance with ISO 81060-2:2018 requirements for non-invasive sphygmomanometers. Validation studies were performed across a broad population with arm circumferences spanning the full intended range.
The following cuff models were included in the validation:
{7}
K250765
- 6NRSI-K Cuff (15–22 cm arm circumference)
- 6NRMI-K Cuff (22–33 cm arm circumference)
- 6NRLI-K Cuff (33–42 cm arm circumference)
- 6NRXI-K Cuff (2–52 cm arm circumference)
- 6NRUI-K Cuff (22–42 cm arm circumference)
All clinical testing followed the ISO 81060-2:2018 protocol, comparing device measurements against simultaneous auscultatory reference measurements obtained by trained observers using a mercury sphygmomanometer.
Results:
For each cuff model, the subject device met the accuracy requirements of ISO 81060-2:2018:
- Criterion 1 (mean difference and standard deviation): Within the allowable limits of ±5 mmHg (mean) and ≤8 mmHg (standard deviation).
- Criterion 2 (per-subject distribution of differences): Variability across subjects was within acceptance criteria.
Conclusion:
The validation studies confirmed that the subject device provides accurate and reliable blood pressure measurements across the full intended range of arm circumferences. All five cuff models listed above have been clinically validated in compliance with ISO 81060-2:2018.
9 DEVICE COMPARISON TABLE
| Item | Proposed Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| Product Name | AViTA Arm Type
Blood Pressure Monitor Cuff | Reusable NIBP Cuff | -- |
| K number | TBD | K231961 | |
Page 4 of 7
{8}
K250765
| Applicant | AViTA Corporation | Shenzhen SINO-K Medical Technology Co., Ltd | |
| --- | --- | --- | --- |
| 510(k) Information | | | |
| Regulation Number | 870.1120 | 870.1120 | Identical |
| Classification | Class II | Class II | Identical |
| Product Code | DXQ | DXQ | Identical |
| Indication for Use | | | |
| Statement | The cuff is compatible with non-invasive blood pressure monitors to measure blood pressure. It is intended for use by healthcare professionals (prescription use) and by laypersons in home settings (over-the-counter use). The cuff is designed for adult patients. It is non-sterile and reusable. | The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated. | Substantially Equivalent. (Note 1) |
| Population | adults | adults/pediatrics | Identical |
| Tube number | one | one | Identical |
| Principles of operation | Bladder is wrapped around the patient's limb and secured by hook and loop closure Air hose is connected to the noninvasive blood pressure measurement systems | Bladder is wrapped around the patient's limb and secured by hook and loop closure Air hose is connected to the noninvasive blood pressure measurement systems | Identical |
| Limb Circumference (Range in cm) | 6NRMI-K (22-33cm)6NRUI-K (22-42cm)6NRLI-K (33-42cm) | Neonatal (6-11cm)Infant (10-19cm)Child (18-26cm) | Substantially Equivalent*Note 2 |
{9}
K250765
| | 6NRXI-K(42-52cm)
6NRSI-K (15-22cm) | Small Adult (20-28cm)
Adult (25-35cm)
Adult Long (25-35cm)
Large Adult (33-47cm)
Adult Thigh (44-66cm) | |
| --- | --- | --- | --- |
| Pressure Range | 0~330mmHg | 0~330mmHg | Identical |
| Max. Leakage | < 4mm Hg/Min | < 4mm Hg/Min | Identical |
| Biocompatibility | Comply with
ISO 10993-5
ISO 10993-10
ISO 10993-23 | Comply with
ISO 10993-5
ISO 10993-10 | Substantially
Equivalent
*Note 3 |
| Sterility | Non-sterile | Non-sterile | Identical |
| Repeated inflation | 10,000 inflations
3,000 hook and loop closures | Unknown | *Note 4 |
*Note 1
Both devices are intended to be used with non-invasive BP monitors. The subject device expands labeling to include OTC use for adults, but this does not impact fundamental technology or performance.
*Note 2
Although the predicate device includes cuffs intended for neonatal, infant, child, and adult populations, the subject device is intended only for adult patients, with five cuff sizes covering the full adult arm circumference range (15–52 cm). Clinical validation has been performed in accordance with ISO 81060-2:2018 across all five subject cuff sizes. Therefore, while the intended user population is narrower (adults
Page 6 of 7
{10}
K250765
only), the differences do not raise new questions of safety or effectiveness, and the subject device remains substantially equivalent to the predicate device.
*Note 3*
Although the subject device has undergone additional biocompatibility testing (ISO 10993-23, in addition to ISO 10993-5 and ISO 10993-10), it remains substantially equivalent to the predicate device. The additional compliance with ISO 10993-23 ensures further safety validation regarding irritation and skin sensitization, but does not introduce new risks. Therefore, these differences do not negatively impact the overall safety and effectiveness of the device.
*Note 4*
The subject device has been tested for durability, undergoing 10,000 inflation cycles and 3,000 hook-and-loop closure tests to ensure long-term performance and reliability. The predicate device does not specify repeated inflation test results in publicly available documentation. However, this difference does not impact substantial equivalence, as both devices are intended for the same use and expected to perform reliably over their lifespan
10 SUBSTANTIAL EQUIVALENCE CONCLUSIONS
Based on the non-clinical and clinical performance data, the AViTA Arm Type Blood Pressure Monitor Cuff has been shown to be substantially equivalent to the legally marketed predicate device.
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.