Pression Wave PRO External Counter-Pulsation System
K250756 · Pression, Inc. · DRN · Dec 22, 2025 · Cardiovascular
Device Facts
Record ID
K250756
Device Name
Pression Wave PRO External Counter-Pulsation System
Applicant
Pression, Inc.
Product Code
DRN · Cardiovascular
Decision Date
Dec 22, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5225
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pression Wave PRO External Counter-Pulsation System is indicated for the treatment of chronic stable angina (CSA) that is refractory to optimal anti-anginal medical therapy and without options for revascularization in adult patients that weigh greater than 40kg. In addition, it is intended for use in healthy adult patients to provide improvement in vasodilation, increased VO2, and increased blood flow. It is intended for use under the supervision of a healthcare professional.
Device Story
External counter-pulsation (ECP) system; applies sequential pneumatic compression to patient calves via inflatable sleeves. Input: real-time cardiac cycle phase data via finger-tip PPG sensor. Operation: Compression Control System synchronizes bladder inflation with diastole; sequential inflation from ankle to knee increases aortic retrograde flow; enhances coronary perfusion; provides systolic unloading. Used in clinical settings under healthcare professional supervision. Output: diastolic augmentation; improved blood flow/vasodilation. Benefits: non-invasive treatment for refractory CSA; improved cardiac perfusion; increased VO2 in healthy patients.
Clinical Evidence
WaveCRESST study: randomized crossover-control trial (N=51 mITT) comparing Wave PRO to ACS Model NCP-02. Population: adults ≥35 years with mild-to-moderate CAD. Primary endpoint: diastolic augmentation (DA) ratio at 8 minutes. Results: Wave PRO LS mean DA ratio 0.83 (95% CI: 0.77, 0.88) vs. predicate 0.57 (95% CI: 0.52, 0.62) (p<.0001). Wave PRO demonstrated non-inferiority. No adverse events reported.
Technological Characteristics
Pneumatic ECP system; calf-only inflatable sleeves. Sensing: finger-tip PPG. Energy: compressed air. Connectivity: GUI-based control system. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 10993-1 (biocompatibility), ISO 14971 (risk management). Software: microprocessor-controlled timing algorithm. Portable form factor (34 kg).
Indications for Use
Indicated for adult patients >40kg with chronic stable angina (CSA) refractory to optimal medical therapy and without revascularization options. Also indicated for healthy adults to improve vasodilation, VO2, and blood flow. Contraindications not specified.
Regulatory Classification
Identification
An external counter-pulsating device is a noninvasive, prescription device used to assist the heart by applying positive or negative pressure to one or more of the body's limbs in synchrony with the heart cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. The special controls for this device are:(i) Nonclinical performance evaluation of the device must demonstrate a reasonable assurance of safety and effectiveness for applied pressure, synchronization of therapy with the appropriate phase of the cardiac cycle, and functionality of alarms during a device malfunction or an abnormal patient condition;
(ii) Reliabilities of the mechanical and electrical systems must be established through bench testing under simulated use conditions and matched by appropriate maintenance schedules;
(iii) Software design and verification and validation must be appropriately documented;
(iv) The skin-contacting components of the device must be demonstrated to be biocompatible;
(v) Appropriate analysis and testing must be conducted to verify electrical safety and electromagnetic compatibility of the device; and
(vi) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) for the following intended uses: Unstable angina pectoris; acute myocardial infarction; cardiogenic shock; congestive heart failure; postoperative treatment of patients who have undergone coronary artery bypass surgery; peripheral arterial disease associated with ischemic ulcers rest pain or claudication, threatened gangrene, insufficient blood supply at an amputation site, persisting ischemia after embolectomy or bypass surgery, and/or pre- and post-arterial reconstruction to improve runoff; diabetes complicated by peripheral arterial disease or other conditions possibly related to arterial insufficiency including nocturnal leg cramps and/or necrobiosis diabeticorum; venous diseases, including prophylaxis of deep vein thrombophlebitis, edema (e.g., chronic lymphedema) and/or induration (e.g., stasis dermatitis) associated with chronic venous stasis, venous stasis ulcers, and/or thrombophlebitis; athletic injuries, including Charley horses, pulled muscles and/or edematous muscles; necrotizing cellulitis.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before March 31, 2014, for any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other external counter-pulsating device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
ACS Model NCP-2 External Counterpulsation System (K042413)
CardiAssist Counterpulsation System, Mark 3000 (K022107)
Renew NCP-5 External Counterpulsation System (K152115)
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 22, 2025
Pression, Inc.
% Julie Broderick
Principal Consultant
Broderick Regulatory Consulting, LLC
7 Kendall Street
Winchester, Massachusetts 01890
Re: K250756
Trade/Device Name: Pression Wave PRO External Counter-Pulsation System
Regulation Number: 21 CFR 870.5225
Regulation Name: External counter-pulsating device
Regulatory Class: Class II
Product Code: DRN
Dated: October 10, 2025
Received: October 14, 2025
Dear Julie Broderick:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}
K250756 - Julie Broderick
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K250756 - Julie Broderick
Page 3
Sincerely,
Nicole M. Gillette -S
Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
Pression Wave™ PRO External Counter-Pulsation System
Page 9 of 69
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250756 | ? |
| Please provide the device trade name(s). | | ? |
| Pression Wave™ PRO External Counter-Pulsation System | | |
| Please provide your Indications for Use below. | | ? |
| The Pression Wave PRO External Counter-Pulsation System is indicated for the treatment of chronic stable angina (CSA) that is refractory to optimal anti-anginal medical therapy and without options for revascularization in adult patients that weigh greater than 40kg. In addition, it is intended for use in healthy adult patients to provide improvement in vasodilation, increased VO2, and increased blood flow. It is intended for use under the supervision of a healthcare professional. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
Pression, Inc.
K250756: Pression Wave™ PRO External Counter-Pulsation System
510(k) Summary Revised
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92.
## I. Submitter Information
**Company:** Pression, Inc.
190 West Lincoln Highway
Coatesville, PA 19320 USA
**Contact:** Adam Salamon
President & CEO
Pression, Inc.
Phone No.: +1 215-326-9525
Email: adam@pression.com
**Date Prepared:** 22 December 2025
**510(k) No.:** K250756
## II. Name of Device
**Device Trade Name:** Pression Wave™ PRO External Counter-Pulsation System
**Model No.:** WP-01
**Classification Name:** Device, Counter-Pulsating, External
**Product Code:** DRN
**Regulation Number:** 21 CFR 870.5225, External counter-pulsating device
**Device Class:** II
**Panel Identification:** Cardiovascular
## III. Predicate Devices
All predicate and reference devices listed below are in the same Product Code:
**Classification Name:** Device, Counter-Pulsating, External
**Product Code:** DRN
**Regulation Number:** 21 CFR 870.5225, External counter-pulsating device
**Device Class:** II
**Panel Identification:** Cardiovascular
**Predicate Device**
**Manufacturer:** Applied Cardiac Systems, Inc. (ACS)
**Trade Name:** ACS Model NCP-2 External Counterpulsation System
**510(k) No.:** K042413
CONFIDENTIAL
{6}
Pression, Inc.
K250756: Pression Wave™ PRO External Counter-Pulsation System
510(k) Summary Revised
## Reference Devices
Manufacturer: Vamed Medical Instrument Co., Ltd
Trade Name: External Counterpulsation System, ECP MC-3
510(k) No.: K190683
Manufacturer: CardioMedics, Inc.
Trade Name: CardiAssist Counterpulsation System, Mark 3000
510(k) No.: K022107
Manufacturer: Renew Group Pte Ltd. / Stage 2 Innovations
Trade Name: Renew NCP-5 External Counterpulsation System
510(k) No.: K152115
## IV. Device Description
External counter-pulsation (ECP) as a general technology involves the application of pressure to the patient's legs using inflatable cuffs or sleeves to enhance perfusion at the heart. ECP using the Wave PRO involves wrapping the patient's calves in sleeves containing inflatable bladders, which compress the calves in sync with the patient's cardiac cycle. During diastole, a series of five bladders within the sleeves inflate sequentially from the patient's ankles towards their knees. This compression results in diastolic augmentation, which is observable from the pulse data obtained using a finger-tip PPG sensor. During the diastolic phase, the heart is relaxed, allowing for coronary perfusion to the cardiac muscle. By compressing the inflatable bladders during this portion of the cardiac cycle, aortic retrograde flow is increased, allowing for enhanced coronary perfusion.
The Pression Wave PRO System consists of three primary subsystems: 1) the Compression Control System, which controls the flow of compressed air to the sleeves, measures the phase of the cardiac cycle, and provides the graphical user interface (GUI); 2) Patient-Worn Calf Sleeves, which direct the compressive force of compressed air, delivered comfortably and consistently, to the patient's calf muscles; and 3) the Compression Timing Algorithm (Software), which determines when the patient's heart is in diastole, using a photoplethysmography (PPG) sensor, and controls the timed application of compression to the patient's calf muscles..
## V. Indications for Use
The Pression Wave PRO External Counter-Pulsation System is indicated for the treatment of chronic stable angina (CSA) that is refractory to optimal anti-anginal medical therapy and without options for revascularization in adult patients that weigh greater than $40\mathrm{kg}$. In addition, it is intended for use in healthy adult patients to provide improvement in vasodilation, increased VO2, and increased blood flow. It is intended for use under the supervision of a healthcare professional.
CONFIDENTIAL
{7}
Pression, Inc.
K250756: Pression Wave™ PRO External Counter-Pulsation System
510(k) Summary Revised
VI. Comparison of Technological Characteristics
| Characteristic | New Device
Wave PRO External Counter-Pulsation System | Predicate Device
Model NCP-2 External Counterpulsation System | Reference Device
Model ECP-MC3 External Counterpulsation System | Reference Device
CardiAssist Counterpulsation System, Mark 3000 | Reference Device
Renew NCP-5 External Counterpulsation System | Substantial Equivalence |
| --- | --- | --- | --- | --- | --- | --- |
| Manufacturer | Pression, Inc. | Applied Cardiac Systems, Inc. | Vamed Medical Instrument Co., Ltd. | CardioMedics, Inc. | Renew Group Pte Ltd. / Stage 2 Innovations, Inc. | N/A |
| 510(k) Number | K250756 | K042413 | K190683 | K022107 | K152115 | N/A |
| Regulation No. | 21 CFR 870.5225 | 21 CFR 870.5225 | 21 CFR 870.5225 | 21 CFR 870.5225 | 21 CFR 870.5225 | Substantially Equivalent
Same as predicates |
| Regulation Name | External counter-pulsating device | External counter-pulsating device | External counter-pulsating device | External counter-pulsating device | External counter-pulsating device | Substantially Equivalent
Same as predicates |
| Regulatory Class | Class II | Class II | Class II | Class II | Class II | Substantially Equivalent
Same as predicates |
| Product Classification Code | DRN | DRN | DRN | DRN | DRN | Substantially Equivalent
Same as predicates |
| Intended Use / Indications for Use | The Pression Wave PRO External Counter-Pulsation System is indicated for the treatment of chronic stable angina (CSA) that is refractory to optimal anti-anginal medical therapy and without options for revascularization in adult patients that weigh greater than 40 kg. In addition, it is intended for use in healthy adult patients to provide improvement in | ACS Model NCP-2 External Counterpulsation Device is a noninvasive external Counterpulsation device intended for the use in the treatment of patients with stable or unstable angina pectoris, acute myocardial infarction, cardiogenic shock or congestive heart failure. | The ECP-MC3 device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. In addition, it is intended for use in healthy patients to provide improvements in vasodilation, and increased blood flow. It is intended for use under the | The CardioMedics, Inc., CardiAssist Counterpulsation System, Mark 3000, is intended to provide external counterpulsation (ECP) for the treatment of ischemic heart disease by increasing perfusion during diastole in people with chronic angina pectoris, congestive heart failure, myocardial infarction and cardiogenic shock. Use of this | The Renew NCP-5 External Counterpulsation System is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. In addition, it is intended for use in healthy patients to provide improvement in vasodilation, increased VO2, and increased blood flow. It is intended for use | Substantially Equivalent
Substantially the same as predicates¹. |
¹ The ACS Model NCP-2 External Counterpulsation System and CardiAssist Counterpulsation System Mark 3000 were cleared prior to FDA’s reclassification of ECP systems for indications other than chronic stable angina to class III (PMA) in December 2013. Both device indication/intended use includes chronic or stable angina pectoris within the cleared indication, which is the same as the Pression Wave device (chronic stable angina).
CONFIDENTIAL
{8}
Pression, Inc.
K250756: Pression Wave™ PRO External Counter-Pulsation System
510(k) Summary Revised
| Characteristic | New Device
Wave PRO External Counter-Pulsation System | Predicate Device
Model NCP-2 External Counterpulsation System | Reference Device
Model ECP-MC3 External Counterpulsation System | Reference Device
CardiAssist Counterpulsation System, Mark 3000 | Reference Device
Renew NCP-5 External Counterpulsation System | Substantial Equivalence |
| --- | --- | --- | --- | --- | --- | --- |
| | vasodilation, increased VO_{2}, and increased blood flow.
It is intended for use under the supervision of a healthcare professional. | | oversight of a healthcare professional. | device may reduce pain and impairment associated with angina pectoris, congestive heart failure or myocardial infarction and may enhance coronary function. | under the oversight of a healthcare professional. | |
| Device Description / Operating Principle | The application of pressure to the patient’s legs using inflatable cuffs or sleeves surrounding the calves, inflated sequentially from the ankle to the knee and synchronized to the diastolic phase of the cardiac cycle, results in diastolic augmentation and systolic unloading, which is observable from the pulse data obtained using a finger-tip PPG sensor. During the diastolic phase, the heart is relaxed, allowing for coronary perfusion to the cardiac muscle. By compressing the inflatable sleeves during this portion of the cardiac cycle, aortic retrograde flow is increased, allowing for enhanced coronary perfusion and systolic unloading. | When the heart is in its relaxed state during the diastolic period, pressure is applied sequentially from the calves, to the lower thighs, to the upper thighs and buttocks, forcing blood back to the heart, increasing coronary perfusion pressure and coronary blood flow (diastolic augmentation), as well as venous return. Immediately before the heart begins to eject blood during the next systolic phase, the Cuffs are rapidly deflated and all externally applied pressure is eliminated. The vasculature in the lower extremities re-conforms and is able to receive the output of the heart with lessened resistance, thereby reducing systolic pressure and the workload of the heart (decreased after-load).
Although not specifically mentioned in the predicate 510(k) summary, systolic unloading is a known hemodynamic effect of ECP. | The ECP-MC3 system is computer-controlled system that inflates and deflates three pairs of air cuffs in synchronization with the patient’s cardiac cycle. The three pairs of cuffs are wrapped around the calves, lower thighs, and upper thighs/buttocks of the patients. As diastole begins, the cuffs inflate separately in sequence from the calves, to the lower thighs, to the upper thighs including the lower buttocks. This inflation sequence generates and impels a counter-pulsation wave, increasing diastolic blood pressure (diastolic augmentation), coronary perfusion pressure and coronary blood flow, and raising cardiac output.
Although not specifically mentioned in the predicate 510(k) summary, systolic unloading is a known hemodynamic effect of ECP | The CardioMedics CardiAssist ECP System is a non-invasive circulatory assist device which provides increased circulation via external counterpulsation (ECP) for the treatment of ischemic heart disease including congestive heart failure, chronic angina pectoris, acute myocardial infarction and cardiogenic shock. External counterpulsation therapy improves cardiac function by enhancing the perfusion of the coronary vasculature, the development of coronary collateral circulation, and by reducing the workload of the heart.
The CardiAssist ECP System consists of the portable console containing the computer and pumps with a touch screen for user interface, an integral strip chart recorder, and leg cuffs and hoses. Additional components provided with the System include a finger plethysmograph and 3-lead ECG cable and leads. This CardiAssist ECP System sequentially compresses the legs from the calves, thighs and | The Renew™ NCP-5 External Counterpulsation System (ECS) is comprised of the following major components, a switch panel, laptop PC computer, a foldable treatment bed, main unit, and a set of patient cuffs. The device is a microprocessor-controlled system that sequentially inflates then deflates three pairs of air cuffs, which compress vascular beds in the muscles of the calves, thighs, and buttocks in synchrony with the heart cycle to achieve the desired therapy. | **Substantially Equivalent**
Same as predicates except that the Wave PRO utilizes sleeves on the calves only, whereas the predicate devices utilize cuffs on the calves, thighs and buttocks/hips. Clinical data included in the 510(k) demonstrate that the Wave PRO System produces the same or better hemodynamic effect as the predicate device.
The Wave PRO is portable and substantially equivalent in this feature to the CardiAssist and Renew ECP reference devices. |
CONFIDENTIAL
{9}
Pression, Inc.
K250756: Pression Wave™ PRO External Counter-Pulsation System
510(k) Summary Revised
| Characteristic | New Device
Wave PRO External Counter-Pulsation System | Predicate Device
Model NCP-2 External Counterpulsation System | Reference Device
Model ECP-MC3 External Counterpulsation System | Reference Device
CardiAssist Counterpulsation System, Mark 3000 | Reference Device
Renew NCP-5 External Counterpulsation System | Substantial Equivalence |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | buttocks, 40 milliseconds apart, by inflating three sets of flexible fabric cuffs during diastole. This results in movement of blood from the legs to the heart and entire upper body. Pressure, up to 310 mmHg, is applied with the timing and duration of each pulse, synchronized with the patient’s ECG. When properly triggered, the pressure pulses applied to the vascular bed of the legs and buttocks transmit retrograde pressure through the entire vascular system. At the aorta, the aortic valve prevents retrograde flow into the left ventricle. Thus, a peak pulse of diastolic pressure occurs at or above systolic levels which increases the driving pressure in the coronary vasculature | | |
| Portability | Portable (34 kg) | Stationary | Stationary | Portable (~400 kg) | Portable (~250 kg) | Substantially Equivalent
Same as portable reference devices but much lighter |
| Treatment pressure range | 41-50 kPa (6-7.25 PSI)
(automatically set) | 13-41 kPa (1.9-6 PSI)
(manually set) | 0-50 kPa (0-7.25 PSI)
(manually set) | 11-41 kPa (1.6-6 PSI)
(manually set) | 0-41 kPa (0-6 PSI)
(manually set) | Substantially Equivalent
Same as ECP-MC3 reference predicate |
| Maximum safety pressure | 55.2 kPa (8 PSI) | 55.2 kPa (8 PSI) | 55 kPa (8 PSI) | Unreported | Unreported | Substantially Equivalent
Same as predicate and ECP-MC3 reference device |
CONFIDENTIAL
{10}
Pression, Inc.
K250756: Pression Wave™ PRO External Counter-Pulsation System
510(k) Summary Revised
| Characteristic | New Device
Wave PRO External Counter-Pulsation System | Predicate Device
Model NCP-2 External Counterpulsation System | Reference Device
Model ECP-MC3 External Counterpulsation System | Reference Device
CardiAssist Counterpulsation System, Mark 3000 | Reference Device
Renew NCP-5 External Counterpulsation System | Substantial Equivalence |
| --- | --- | --- | --- | --- | --- | --- |
| Software validation | Per FDA guidance, Content of Premarket Submissions for Device Software Functions, June 2023 | Per FDA guidance, General Principles of Software Validation, January 2002 | Per FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (version unknown) | Unknown | The software and firmware used in the system was validated per FDA’s guidance document. In all instances, the NCP-5 system functioned as intended. | Substantially Equivalent
Same as predicate |
| Electrical safety | IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance | Per IEC 601-1-4, Medical Electrical Equipment Part 1: General requirements for safety. Collateral Standard: Programmable electrical medical systems, 1996-05 | IEC60601-1, Medical Electrical Equipment – Part 1: General Requirements for Safety, 2005+A1:2012 | Unknown | Electrical safety and EMC compatibility testing per IEC 60601 was performed to ensure appropriate electrical performance. | Substantially Equivalent
Same as reference devices, using most recent standard |
| Electromagnetic compatibility (EMC) | IEC EN 60601-1-2, Edition 4.1:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | Unknown | IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests, 2014 | Unknown | Electrical safety and EMC compatibility testing per IEC 60601 was performed to ensure appropriate electrical performance. | Substantially Equivalent
Same as reference devices, using most recent standard |
| Usability | Per FDA Guidance for Applying Human Factors and Usability Engineering to Medical Devices, February 2016 | Unknown | IEC 60601-1-6 Edition 3.1 Medical Electrical Equipment - Part 1-6: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability | Unknown | Unknown | Similar
Similar to reference predicate, using most recent FDA guidance; achieves same goal |
| Clinical testing | A clinical investigation of the Wave PRO has demonstrated substantial equivalence to the predicate device. | Yes, 510(k) summary states: An Independent Review Board has overseen the clinical investigation of the External Counterpulsation Device. Results have demonstrated substantial equivalence to the predicate device. | Clinical performance is not deemed necessary. | Unknown | A literature review regarding the use of external counterpulsation in healthy patients was provided to support this additional indication. | Substantially Equivalent
Same as predicate |
CONFIDENTIAL
Page 6 of 9
{11}
Pression, Inc.
K250756: Pression Wave™ PRO External Counter-Pulsation System
510(k) Summary Revised
# VII. Performance or Clinical Testing
Bench performance testing to verify the design of the Pression Wave PRO System, all of which PASSED, included:
- Compression and Sensor Accuracy
- Product Safety & Electrical Testing
- EMC Testing
- Software Verification
- End of Line / Performance Test
- Label Verification
- Cycle Verification
- Service Life Verification (system and compression sleeves)
- Reprocessing Validation (compression sleeves)
- Summative Usability Testing
- Biocompatibility Assessment
The following standards were applied:
- IEC 60601-1, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 62304, Software lifecycle processes
- ISO 10993-1, Biological evaluation of medical devices – Part 1: evaluation and testing within a risk management process
- ISO 14971, Medical devices — Application of risk management to medical devices
- ISO 15223-1, Medical devices — Symbols to be used with information to be supplied by the manufacturer — Part 1: General requirements
The WaveCRESST clinical study demonstrated that the Wave PRO ECP System is non-inferior to the predicate device, ACS Model NCP-02 ECP System, with respect to the primary effectiveness endpoint, diastolic augmentation (DA) ratio. Details of the study are as follows:
- Level of Evidence: Randomized crossover-control study comparing Wave PRO ECP to ACS Model NCP-02 ECP (sequence of device use was randomized).
- Study Population: Volunteer participants at least 35 years of age with a history of mild to moderate coronary artery disease, but otherwise healthy.
- Location of Study: United States (one site)
- Safety Endpoints: Rates of device-related adverse events and device-related serious adverse events
CONFIDENTIAL
{12}
Pression, Inc.
K250756: Pression Wave™ PRO External Counter-Pulsation System
510(k) Summary Revised
- Primary Effectiveness Endpoint: DA ratio during the 8th minute of compressions, as calculated by the blinded Independent Waveform Evaluator. The difference of the average DA ratio attained using the Wave PRO and the predicate device was calculated for all randomized study participants with paired ECP data., with randomized assignment of ECP device sequence and unbiased (blinded) waveform scoring of primary effectiveness endpoint, DA ratio, and secondary endpoint systolic unloading.
- Patient Accountability:
| | | Randomization | |
| --- | --- | --- | --- |
| | Total | Wave PRO First | ACS First |
| Screened | 60 | | |
| Screen fail | 8 | | |
| Randomized (Intent to Treat) | 52 | 26 | 26 |
| Modified Intent to Treat Cohort & Safety Cohort (mITT) (received any ECP compressions) | 51 | 25 | 26 |
| Completed Study | 48 | 22 | 26 |
| Per Protocol (PP) Cohort (received any compressions from both ECP treatments with no major protocol deviations) | 48 | 22 | 26 |
- Baseline Demographic Information (mITT population, N=51):
- Sex (% female): 47.1%
- Age at procedure (mean ± SD, range): 60.9 ± 9.1 (41.0, 78.0)
- Race: 2% American Indian/Alaska Native; 2.0% Asian; 47.1% Black/African American; 45.1% White; 2.0% Other; 2.0% Unknown
- Ethnicity (% Hispanic or Latino): 2.0%
- BMI (kg/m^2) (mean ± SD, range): 32.6 ± 6.8 (17.7, 50.5)
- Fitzpatrick Skin Type: type I = 5.9%; type II = 19.6%; type III = 19.6%; type IV = 7.8%; type V = 33.3%; type VI = 13.7%
- Results, Primary Effectiveness Endpoint: The WaveCRESST study met its primary endpoint. In the mITT population, LS mean DA ratio was 0.83 (95% CI: 0.77, 0.88) for the Wave PRO device and 0.57 (95% CI: 0.52, 0.62) for the predicate ACS device (p<.0001). The Wave PRO device exceeded performance with the difference of 0.26 compared to the predicate device, far exceeding the non-inferiority margin of -0.15. The Wave PRO System was therefore determined to be non-inferior to the predicate ACS.
- Results, Safety: There were no adverse events, serious or otherwise, that occurred during the WaveCRESST study.
CONFIDENTIAL
{13}
Pression, Inc.
K250756: Pression Wave™ PRO External Counter-Pulsation System
510(k) Summary Revised
- Conclusions: The WaveCRESST study met its primary endpoint and demonstrated that the Wave PRO System was non-inferior to the predicate device with respect to effectiveness. There were no adverse events. The Wave PRO System was therefore shown to be safe and effective in this study.
## VIII. Special Controls
The Pression Wave PRO External Counter-Pulsation System meets the special controls specified in 21 CFR 870.5225, External counter-pulsating device.
## IX. Conclusions and Benefit/Risk Assessment
Based on the nonclinical and clinical testing summarized in this 510(k), the Pression Wave PRO External Counter-Pulsation System is as safe, performs as well as, and is at least as effective as the predicate device. Data from the WaveCRESST clinical study indicates that the Wave PRO System is non-inferior to the predicate device with respect to effectiveness based on DA ratio. There were no adverse events in the clinical study, and nonclinical testing and risk management have determined that the risk of the Wave PRO System has been reduced as far as possible. The Wave PRO System is at least as safe as the predicate device.
The clinical benefits of ECP treatment in patients with chronic stable angina are well documented in the literature. Given that the Wave PRO System has been shown safe and effective in nonclinical and clinical testing, the expected benefits of the device outweigh any risks.
Pression concludes that the Pression Wave PRO External Counter-Pulsation System is substantially equivalent to the predicate device, K042413, and the reference devices, K190683, K022107 and K152115. The Pression Wave PRO External Counter-Pulsation System does not raise any new issues or concerns of safety or effectiveness and is at least as safe and effective as the predicate device.
CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.