The CloudTMS Edge system is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medications in the current episode.
Device Story
CloudTMS Edge is a repetitive transcranial magnetic stimulation (rTMS) system; produces non-invasive, repetitive pulsed magnetic fields to induce neural action potentials in prefrontal cortex. Used in patient care institutions, diagnostics centers, and neurosurgical hospitals. Principle of operation: discharge of 1.8 kV high-voltage capacitor through stimulation coil; generates pulsed magnetic field (up to 10 kA) penetrating neuromuscular tissues to induce electrical currents in cortical neurons. Target: Left Dorsolateral Prefrontal Cortex. System includes TMS stimulator with software, electromagnetic coils (figure-of-eight design), and flexible positioning arm. Physician/clinician operates system; performs motor threshold determination, coil positioning, and delivers treatment using predefined parameters (iTBS, 19-min, 37-min protocols). Output affects clinical decision-making by providing non-pharmacological treatment for MDD. Benefits patient by offering alternative therapy for treatment-resistant depression.
Clinical Evidence
Bench testing only. Performance testing included SMT comparison, electric field strength decay, output waveform analysis, magnetic field strength linearity, spatial distribution, and gradient. Additional testing verified intensity variation of pulses within a burst (less than 1% variation). Compliance with IEC 60601-1 and IEC 60601-1-2 confirmed.
Technological Characteristics
rTMS system; 1.8 kV capacitor discharge; figure-of-eight air-core coils (liquid or forced-air cooled). Biphasic sinusoid waveform. Connectivity: standalone or PC-integrated. Standards: IEC 60601-1, IEC 60601-1-2, ISO 13485, ISO 14971. Software-controlled stimulation parameters. Coil temperature safety cutoff at 41°C.
Indications for Use
Indicated for treatment of Major Depressive Disorder in adults who failed to receive satisfactory improvement from prior antidepressant medications in current episode.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
Predicate Devices
Mag Vita TMS Therapy System w/Theta Burst Stimulation (K173620)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 14, 2025
TeleEMG, LLC
% Ashar Barry
President
Makromed, Inc.
88 Stiles Road
Salem, New Hampshire 03079
Re: K250689
Trade/Device Name: CloudTMS Edge
Regulation Number: 21 CFR 882.5805
Regulation Name: Repetitive transcranial magnetic stimulation system
Regulatory Class: Class II
Product Code: OBP
Dated: March 5, 2025
Received: March 6, 2025
Dear Ashar Barry:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250689 - Ashar Barry
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K250689 - Ashar Barry
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
PAMELA D. SCOTT -S
2025.09.14 00:54:38
-04'00'
Pamela D. Scott
Assistant Director
DHT5B: Division of Neuromodulation and Physical Medicine Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Form Approved
OMB N
0910-0120
Expiration
Date:
07/31/2026
See P
Statemen
below.
Submission Number (if known)
K250689
Device Name
CloudTMS Edge
Indications for Lise (Describe)
The CloudTMS Edge system is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medications in the current episode.
Type of use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(k) #: K250689 510(k) Summary Prepared on: 2025-09-11
Contact Details 21 CFR 807.92(a)(1)
Applicant Name TeleEMG, LLC
Applicant Address 167 Poinsettia Pl Los Angeles CA 90036 United States
Applicant Contact Telephone 6178403253
Applicant Contact Dr. Joe Jabre
Applicant Contact Email jjabre@teleemg.com
Correspondent Name Makromed, Inc.
Correspondent Address 88 Stiles Road Salem NH 03079 United States
Correspondent Contact Telephone 6036749074
Correspondent Contact Mr. Barry Ashar
Correspondent Contact Email bashar@makromed.com
Device Name 21 CFR 807.92(a)(2)
Device Trade Name CloudTMS Edge
Common Name Repetitive transcranial magnetic stimulation system
Classification Name Transcranial Magnetic Stimulator
Regulation Number 882.5805
Product Code(s) OBP
Legally Marketed Predicate Devices 21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K173620 | Mag Vita TMS Therapy System w/Theta Burst Stimulation | OBP |
| K173441 | CloudTMS | OBP |
TeleEMG, LLC, Traditional 510 (k) CloudTMS Edge
Page 1 of 13
{5}
Page 2 of 13
# Device Description Summary
21 CFR 807.92(a)(4)
The CloudTMS Edge is a repetitive transcranial magnetic stimulation (rTMS) system. This computerized medical device produces non-invasive, repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex for the treatment of Major Depressive Disorder.
It can be used in patient care institutions, diagnostics centers, neurosurgical hospitals
The CloudTMS Edge principle of operation is based on the discharge of high voltage capacitor (1.8 kV) through stimulation coil; the pulsed magnetic field generated by the discharge current (up to 10 kA) penetrates through neuromuscular tissues nearby to induce electrical currents in cortical neurons.
Area of the brain to be stimulated for MDD treatment is Left Dorsolateral Prefrontal Cortex.
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The CloudTMS Edge system is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medications in the current episode.
# Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject device of this 510(k) submission has the exact same Indication for Use as that of the predicate devices.
TeleEMG, LLC, Traditional 510 (k)
CloudTMS Edge
{6}
Technological Comparison 21 CFR 807.92(a)(6)
Introduction
The CloudTMS Edge system is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode. The subject device and the predicate devices (our own CloudTMS and MagVita TMS Therapy System) have identical intended use/indication for use, identical treatment target, and identical treatment parameters of 19-minutes, 37-minutes, and iTBS protocols.
Design
The design of the CloudTMS Edge is similar to that of the our CloudTMS and MagVita TMS Therapy System, as all systems are based on applying transcranial magnetic stimulation by means of repetitive pulse trains at a predetermined frequency. All systems use the same mechanism of action, i.e., an electromechanical instrument that produces and delivers brief duration, rapidly alternating (pulsed) magnetic fields to induce electrical currents in localized regions of the prefrontal cortex. Coils for the subject CloudTMS Edge and the predicate devices are of a similar nature and share the same transducer design (figure-of-eight).
Operational Characteristics
For the subject CloudTMS Edge and its predicate devices, their basic operational procedures including system setup, patient preparations, motor threshold determination, coil positioning and treatment with predefined treatment stimulation parameters are the same.
Technological Characteristics
The subject and predicate devices have similar components consisting of TMS stimulator with software, electromagnetic coils and a flexible arm for positioning of the treatment coil. The reliability of the positioning method used by the subject device is based on the direct relationship of the underlying cortical brain anatomy to the patient's scalp, as is the method used in the predicate device. The method for identifying the correct treatment position in the subject device is at least as effective as the method employed by the predicate device.
Non-clinical Performance Characteristics
Electrical/mechanical/thermal safety and electromagnetic compatibility of both the subject and predicate devices are in compliance with the standards IEC 60601-1:2005/(R)2012 and IEC 60601-1-2:2014.
Patient caps are the only accessory of this device that comes in contact with the patient's intact skin for a duration of less than 24 hours. As they are made of commonly used textile materials they are exempt from ISO 10993-1:2018 Biocompatibility testing per FDA's Guidance Document 'Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process', dated September 8, 2023.
The subject device meets the same requirement as the predicate device that the actual individual stimuli of a theta burst are of equal intensity to ensure that a constant dose of stimuli is delivered during iTBS treatment (maximum 1% drop in machine output among the first stimuli and the third stimuli in a burst during session).
Conclusion
The subject device is equivalent in its performance and does not introduce any new safety considerations in comparison to the predicate devices. There are no identified differences between the subject device and the predicates that impact on safety or efficacy.
TeleEMG, LLC, Traditional 510 (k)
CloudTMS Edge
Page 3 of 13
{7}
| Area | (Subject Device) CloudTMS Edge | (Primary Predicate) MagVita TMS Therapy K173620 | (Secondary Predicate) CloudTMS K173441 | Comments |
| --- | --- | --- | --- | --- |
| Indications for use | The CloudTMS Edge system is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode. | The MagVita TMS Therapy System w/Theta Burst Stimulation are indicated for Treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode. | The Cloud TMS system is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode. | Identical Indication for Use |
| Recommended Standard Treatment | | | | |
| Magnetic Field Intensity | 120% of the MT (iTBS, 19/37 mins protocols) | 120% of the MT (iTBS protocol) | 120% of the MT (19/37 mins protocols) | Identical recommended treatment parameters.
The subject device provides iTBS, 19-min and 37-min protocols.
The primary predicate device provides iTBS protocol.
The secondary predicate device provides 19-min and 37-min protocols. |
| Frequency | 50 Hz / 10 Hz (iTBS, 19/37 mins protocols) | 50 Hz (iTBS protocol) | 10 Hz (19/37 mins protocols) | |
| Train duration | 2 s / 4 s (iTBS, 19/37 mins protocols) | 2 s (iTBS protocol) | 4 s (19/37 mins protocols) | |
| Inter-train interval | 8 s / 11-26 s (iTBS, 19/37 mins protocols) | 8 s (iTBS protocol) | 11-26 s (19/37 mins protocols) | |
| Number of trains | 20 / 75 (iTBS, 19/37 mins protocols) | 20 (iTBS protocol) | 75 (19/37 mins protocols) | |
| iTBS Bursts | Burst Pulses: 3
No. of Bursts: 200
Inter-pulse: 20 msec | Burst Pulses: 3
No. of Bursts: 200
Inter-pulse: 20 msec | N/A | |
| Magnetic Pulses per Session | 600 / 3000 (iTBS, 19/37 mins protocols) | 600 (iTBS protocol) | 3000 (19/37 mins protocols) | |
| Treatment Session Duration | 3min 12s / 18.8min - 37.0 min (iTBS, 19/37 mins protocols) | 3min 12s (iTBS protocol) | 18.8min - 37.0 min (19/37 mins protocols) | |
TeleEMG, LLC, Traditional 510 (k)
CloudTMS Edge
{8}
| Area | (Subject Device)
CloudTMS Edge | (Primary Predicate)
MagVita TMS Therapy
K173620 | (Secondary Predicate)
CloudTMS
K173441 | Comments |
| --- | --- | --- | --- | --- |
| Sessions/wk | 5 | 5 | 5 | |
| Treatment Schedule | 5 daily sessions for 6 weeks | 5 daily sessions for 6 weeks | 5 daily sessions for 6 weeks | |
| Area of brain to be stimulated | Left Dorsolateral Prefrontal Cortex | Left Dorsolateral Prefrontal Cortex | Left Dorsolateral Prefrontal Cortex | |
| Coils | | | | |
| Configuration | FEC-02-100-C: Figure-of-eight coil
AFEC-02-100-C: Figure-of-eight coil
FEC-03-100-C: Figure-of-eight coil
AFEC-03-100-C: Figure-of-eight coil
FEC-04-100-AC: Figure-of-eight coil | Cool B70: Figure-of-eight coil | FEC-02-100-C: Figure-of-eight coil
AFEC-02-100-C: Figure-of-eight coil
FEC-02-100 (Optional): Figure-of-eight coil
AFEC-02-100 (Optional): Figure-of-eight coil | Same figure-of-eight configuration in all coils.
See Note A. |
| Core material | Air core | Air core | Air core | Same core material design. |
| Cooling | FEC-02-100-C: Liquid cooling
AFEC-02-100-C: Liquid cooling
FEC-03-100-C: Liquid cooling
AFEC-03-100-C: Liquid cooling
FEC-04-100-AC: Forced Air cooling | Cool B70: Liquid cooling | FEC-02-100-C: Liquid cooling
AFEC-02-100-C: Liquid cooling
FEC-02-100 (Optional): Air
AFEC-02-100 (Optional): Air | Equivalent.
The subject and secondary predicate devices offer liquid-cooled and forced air-cooled or uncooled coils.
The primary predicate |
TeleEMG, LLC, Traditional 510 (k)
CloudTMS Edge
{9}
| Area | (Subject Device)
CloudTMS Edge | (Primary Predicate)
MagVita TMS Therapy
K173620 | (Secondary Predicate)
CloudTMS
K173441 | Comments |
| --- | --- | --- | --- | --- |
| | | | | device offers a liquid-cooled coil. |
| Coil parameters | FEC-02-100-C:
ID - 50 mm
OD - 100 mm
N= 2 wings x 2 layers x 4 turns | Cool B70:
ID - 23 mm
OD - 97 mm
NW= 2 wings x 11 turns | FEC-02-100-C:
ID - 50 mm
OD - 100 mm
N= 2 wings x 2 layers x 4 turns | Equivalent, with no additional concerns for safety and effectiveness.
See Note A for equivalence explanation of coil parameters such as dimensions and windings. |
| | AFEC-02-100-C:
ID - 51 mm
OD - 106 mm
N= 2 wings x 2 layers x 4 turns | | AFEC-02-100-C:
ID - 51 mm
OD - 106 mm
N= 2 wings x 2 layers x 4 turns | |
| | FEC-03-100-C:
ID - 50 mm
OD - 105 mm
N= 2 wings x 2 layers x 4 turns | | FEC-02-100 (Optional):
ID - 23 mm
OD - 97 mm
N= 2 wings x 2 layers x 4 turns | |
| | AFEC-03-100-C:
ID - 52 mm
OD - 105 mm
N= 2 wings x 2 layers x 4 turns | | AFEC-02-100 (Optional):
ID - 51 mm
OD - 106 mm
N= 2 wings x 2 layers x 4 turns | |
| | FEC-04-100-AC:
ID - 55 mm
OD - 100 mm
N= 2 wings x 2 layers x 4 turns | | | |
| | Machine Output Parameters | | | |
TeleEMG, LLC, Traditional 510 (k)
CloudTMS Edge
{10}
| Area | (Subject Device) CloudTMS Edge | (Primary Predicate) MagVita TMS Therapy K173620 | (Secondary Predicate) CloudTMS K173441 | Comments |
| --- | --- | --- | --- | --- |
| Amplitude in Standard Motor Threshold (SMT) units | FEC-02-100-C: 0 – 2.0
50% Intensity Setting -> 1 SMT
AFEC-02-100-C: 0 – 2.3
43% Intensity Setting -> 1 SMT
FEC-03-100-C: 0 – 2.2
46% Intensity Setting -> 1 SMT
AFEC-03-100-C: 0 – 3.1
32% Intensity Setting -> 1 SMT
FEC-04-100-AC: 0 – 1.9
52% Intensity Setting -> 1 SMT | Cool B70: 0 – 1.9
52% Intensity Setting -> 1 SMT | FEC-02-100-C: 0 – 1.9
53% Intensity Setting -> 1 SMT
AFEC-02-100-C: 0 – 2.3
44% Intensity Setting -> 1 SMT
FEC-02-100 (Optional): 0 – 1.9
52% Intensity Setting -> 1 SMT
AFEC-02-100 (Optional): 0 – 2.3
42% Intensity Setting -> 1 SMT | Equivalent, with no additional concerns for safety and effectiveness. See Note B for equivalence explanation of Intensity/Amplitude. |
| Pulse Width (μs) | FEC-02-100-C: 280
AFEC-02-100-C: 265
FEC-03-100-C: 280
AFEC-03-100-C: 290
FEC-04-100-AC: 305 | Cool B70: 310 | FEC-02-100-C: 260
AFEC-02-100-C: 260
FEC-02-100 (Optional): 300
AFEC-02-100 (Optional): 290 | Equivalent, with no additional concerns for safety and effectiveness. See Note C for equivalence explanation of Pulse Width and Amplitude. |
TeleEMG, LLC, Traditional 510 (k)
CloudTMS Edge
{11}
TeleEMG, LLC, Traditional 510 (k)
CloudTMS Edge
Page 8 of 13
| Area | (Subject Device) CloudTMS Edge | (Primary Predicate) MagVita TMS Therapy K173620 | (Secondary Predicate) CloudTMS K173441 | Comments |
| --- | --- | --- | --- | --- |
| Max initial dB/dt (kT/s) near the coil surface | FEC-02-100-C: 25
AFEC-02-100-C: 28
FEC-03-100-C: 25
AFEC-03-100-C: 28
FEC-04-100-AC: 20 | Cool B70: 23 | FEC-02-100-C: 24
AFEC-02-100-C: 28 | Equivalent, with no additional concerns for safety and effectiveness. |
| Max initial dB/dt (kT/s) at 20mm from coil surface | FEC-02-100-C: 10
AFEC-02-100-C: 11
FEC-03-100-C: 10
AFEC-03-100-C: 14
FEC-04-100-AC: 9 | Cool B70: 8 | FEC-02-100-C: 9
AFEC-02-100-C: 11 | Equivalent, with no additional concerns for safety and effectiveness. |
| Waveform | Biphasic sinusoid | Biphasic sinusoid | Biphasic sinusoid | Identical waveform. |
| The system will automatically be disabled when the coil temperature exceeds: | 41 °C (106 °F) | 41 °C (106 °F) | 41 °C (106 °F) | Identical |
| Frequency range (Hz) | 0.1 – 30 (Stand-alone)
0.1 – 100 (with PC) | 0.1 - 30 or 0.1 - 100, depending on model | 0.1 – 30 (Stand-alone)
0.1 – 100 (with PC) | The differences in these parameters are simply |
{12}
| Area | (Subject Device) CloudTMS Edge | (Primary Predicate) MagVita TMS Therapy K173620 | (Secondary Predicate) CloudTMS K173441 | Comments |
| --- | --- | --- | --- | --- |
| Pulse train duration range (s) | 0.5 – 100 | Rep Rate: 0.1 ...100Hz
Pulses in Train: 1,2,3,4 ... 1000
Train duration = Pulses in Train / Rep Rate | 0.5 – 100 | the differences in the overall capabilities of these machines. These capabilities encompass the recommended treatment parameters for MDD listed above. In other words, these variations among different manufacturers’ models do not impact their ability to deliver the treatment parameters recommended for MDD. All machines use identical treatment parameters. |
| Inter-train interval range (s) | 0 – 300 | 1 - 120 | 0 – 300 | |
| Maximum trains per session | 4800 = 2400s [max session] / (0.5 s [min train] + 0 sec [min pause]) | 500 | 4800 = 2400s [max session] / (0.5 s [min train] + 0 sec [min pause]) | |
| Maximum # of pulses per session (cumulative exposure) | 72000(Stand-alone) =2400s [max session] *30Hz
240000(with PC) =2400s [max session] *100Hz | 500,000 = 1,000 (pulses max per train) x 500 (trains max per session) | 72000(Stand-alone) =2400s [max session] *30Hz
240000(with PC) =2400s [max session] *100Hz | |
| Standards | | | | |
| Electrical Safety and Electromagnetic Compatibility | IEC 60601-1
IEC 60601-1-2 | IEC 60601-1
IEC 60601-1-2 | IEC 60601-1
IEC 60601-1-2 | N/A |
| ISO Standards | ISO13485:2012
ISO 10993-1:2009
ISO 14971:2007 | ISO 13485:2012 | ISO13485:2012
ISO 10993-1:2009
ISO 14971:2007 | N/A |
TeleEMG, LLC, Traditional 510 (k)
CloudTMS Edge
{13}
Page 10 of 13
# Note A:
- Applicator Diameter: ID/OD differences are typical of variations among different manufacturers of TMS devices or even among different models of the same manufacturer. As shown in Section 18 Appendix 18-1 Performance Testing - Bench Top Study, they do not lead to significant differences in the electric and magnetic fields generated by these applicators.
- Applicator Windings: Both the subject and the predicate devices have the same number of windings (2 wings) and figure-of-eight, as illustrated below:

Layers and turns differences are typical of variations among different manufacturers of TMS devices or even among different models of the same manufacturer. As a comparison, a TMS device from another manufacturer (Rapid Therapy System, K143531) that is cleared for MDD treatment has a vastly different windings design of $3 \times 19 \times 2$ windings. As shown in Section 18 Appendix 18-1 Performance Testing - Bench Top Study, they do not lead to significant differences in the electrical and magnetic fields generated by these applicators.
# Note B:
- Subject Device (CloudTMS Edge, Coil FEC-02-100-C):
- Intensity/Amplitude Setting = 0% → 0 SMT
- Intensity/Amplitude Setting = 50% → 1 SMT
- Intensity/Amplitude Setting = 100% → 2.0 SMT
- Subject Device (CloudTMS Edge, Coil AFEC-02-100-C):
- Intensity/Amplitude Setting = 0% → 0 SMT
- Intensity/Amplitude Setting = 43% → 1 SMT
- Intensity/Amplitude Setting = 100% → 2.3 SMT
TeleEMG, LLC, Traditional 510 (k)
CloudTMS Edge
{14}
- Subject Device (CloudTMS Edge, Coil FEC-03-100-C):
- Intensity/Amplitude Setting = 0% → 0 SMT
- Intensity/Amplitude Setting = 46% → 1 SMT
- Intensity/Amplitude Setting = 100% → 2.2 SMT
- Subject Device (CloudTMS Edge, Coil AFEC-03-100-C):
- Intensity/Amplitude Setting = 0% → 0 SMT
- Intensity/Amplitude Setting = 32% → 1 SMT
- Intensity/Amplitude Setting = 100% → 3.1 SMT
- Subject Device (CloudTMS Edge, Coil FEC-04-100-AC):
- Intensity/Amplitude Setting = 0% → 0 SMT
- Intensity/Amplitude Setting = 52% → 1 SMT
- Intensity/Amplitude Setting = 100% → 1.9 SMT
- Primary Device (MagVita TMS, Coil Cool B70):
- Intensity/Amplitude Setting = 0% → 0 SMT
- Intensity/Amplitude Setting = 52% → 1 SMT
- Intensity/Amplitude Setting = 100% → 1.9 SMT
- Reference Device (CloudTMS, Coil FEC-02-100-C):
- Intensity/Amplitude Setting = 0% → 0 SMT
- Intensity/Amplitude Setting = 53% → 1 SMT
- Intensity/Amplitude Setting = 100% → 1.9 SMT
- Reference Device (CloudTMS, Coil AFEC-02-100-C):
- Intensity/Amplitude Setting = 0% → 0 SMT
- Intensity/Amplitude Setting = 44% → 1 SMT
- Intensity/Amplitude Setting = 100% → 2.3 SMT
TeleEMG, LLC, Traditional 510 (k)
CloudTMS Edge
{15}
These are comparable values and within the variations seen among previously cleared TMS devices by the FDA. These values are in slight favor of CloudTMS Edge as they indicate that the subject device will need a slightly lower intensity setting to achieve the same level of induced current as the predicate devices (K173620 and K173441).
**Note C:**
Since a patient is always treated at 120% MT for MDD on any TMS device, and MT is a function of both the intensity/amplitude and pulse width (higher pulse width will require lower intensity/amplitude to reach MT), differences in pulse widths are easily compensated by the changes in intensity settings needed to achieve 120% MT stimulation. Inherent differences in the design and construction of TMS devices of different manufacturers, or even different models of the same manufacturer, lead to variations in pulse widths.
Therefore, the pulse width is substantially equivalent to the predicate devices and this minor difference will not impact safety or effectiveness.
TeleEMG, LLC, Traditional 510 (k)
CloudTMS Edge
Page 12 of 13
{16}
Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The following special control is applicable to the subject device (Product Code: OBP) of this 510(k) submission:
"Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems."
The following performance tests specified in this guidance document were performed and the results were compared for a determination of substantial equivalence between the subject device (CloudTMS Edge) and the predicate devices (MagVita TMS Therapy System and CloudTMS):
Electric Field:
a. SMT Comparison
b. Electrical Field Strength Decay over z-axis distance from coil center (in the direction of head center)
Magnetic Field:
a. Output Waveform for pulse width and amplitude
b. Magnetic Field Strength Linearity as a function of machine intensity setting
c. Magnetic Field Spatial Distribution
d. Magnetic Field Gradient
Also, an additional test (Intensity Variation of Pulses within a Burst) not specified in the guidance document but essential for establishing substantial equivalence for iTBS protocol was performed.
CONCLUSION STATEMENT
The CloudTMS Edge and the predicate devices have identical intended use/indication for use, target population, treatment procedure, treatment position and all recommended standard treatment protocol parameters (intensity, frequency, number of pulses in a train, number of trains in a session, number of treatment sessions).
Both the subject device and the predicate devices share the same Figure-of-eight coil design. The tested magnetic properties of the CloudTMS Edge and the predicate devices are substantial equivalent for the coils. Variation in intensity of individual pulses in a burst for the subject device show less than 1% variation.
The reliability of the positioning method used by the CloudTMS Edge is based on the direct relationship of the underlying cortical brain anatomy to the patient's scalp, as is the method used in the predicate devices. The method for identifying the correct treatment position in the CloudTMS Edge is as effective as the method employed by the predicate devices.
The CloudTMS Edge does not introduce any new safety considerations in comparison to the predicate devices. All other identified differences between the subject device and the predicate devices are minor and without any known impact on safety or efficacy.
Based on the information and supporting documentation provided in the premarket notification, the CloudTMS Edge is substantially equivalent to the cited predicate devices. Testing demonstrates that the CloudTMS Edge fulfills prospectively defined design and performance specifications.
TeleEMG, LLC, Traditional 510 (k)
CloudTMS Edge
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.