← Product Code [QAS](/productcode/QAS) · K250685

# Methinks NCCT Stroke (K250685)

_Methinks Software, S.L · QAS · Jun 16, 2025 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K250685

## Device Facts

- **Applicant:** Methinks Software, S.L
- **Product Code:** [QAS](/productcode/QAS.md)
- **Decision Date:** Jun 16, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2080
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Software as a Medical Device, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K250685 · Jun 16, 2025 | Methinks NCCT Stroke | Methinks Software, S.L | Retrospective clinical NCCT head images | The sponsor conducted a retrospective, multicenter, multinational study to evaluate the software's performance in identifying intracranial hemorrhage (ICH) and large vessel occlusion (LVO) in clinical images. | Retrospective study; Multicenter; Multinational; NCCT images; Performance validation |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective, blinded, multicenter, multinational study; Retrospective, blinded, multicenter, multinational study | Adult patients undergoing NCCT head imaging; Sample Size: 358 cases for ICH; 335 cases for LVO; Number of Sites: Multicenter | Not applicable for this study | Sensitivity and specificity for ICH and LVO detection |

## Indications for Use

Methinks NCCT Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of (1) non-contrast head CT (NCCT) images. The device is intended to assist hospital networks and trained physicians in workflow triage by flagging and communicating suspected positive findings of (1) Intracranial Hemorrhage (ICH) and (2) Large Vessel Occlusion (LVO) of the ICA, MCA-M1 and MCA-M2. Methinks NCCT Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with suspected (1) ICH and (2) LVO in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected ICH or LVO findings via PACS and/or notifications. Notifications include preview images that are meant for informational purposes only, and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Methinks NCCT Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care. Methinks NCCT Stroke is for adults only.

## Device Story

Software-as-a-Medical-Device (SaMD) for radiological triage; receives NCCT images; uses AI/ML algorithm to analyze images in the cloud; identifies suspected Intracranial Hemorrhage (ICH) and Large Vessel Occlusion (LVO) of ICA, MCA-M1, and MCA-M2; operates in parallel to standard-of-care interpretation; sends notifications with preview images to PACS/mobile devices; assists trained physicians in prioritizing patient workflow; does not replace primary diagnostic review; does not alter original images; benefits include reduced time to notification for suspected stroke cases.

## Clinical Evidence

Retrospective, blinded, multicenter, multinational study (N=358 for ICH; N=335 for LVO). ICH performance: Sensitivity 94.7%, Specificity 99.5%. LVO performance: Sensitivity 76.4%, Specificity 91.1%. Reader study (N=4 experts, N=4 non-experts) demonstrated device superiority in LVO sensitivity compared to both expert (50.0%) and non-expert (37.7%) radiologists. Mean time to notification: 1.43 min (ICH) and 1.42 min (LVO). Subgroup analyses performed across gender, age, slice thickness, and vendor machines.

## Technological Characteristics

AI/ML-based software; cloud-deployed; processes NCCT DICOM images; single algorithm with dual outputs (ICH/LVO); cybersecurity controls include secure design, vulnerability assessment, SBOM, and penetration testing; conforms to ISO 62304:2015 and EN ISO 14971:2019.

## Regulatory Identification

Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.

## Special Controls

Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.

## Predicate Devices

- Rapid NCCT Stroke ([K222884](/device/K222884.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

June 16, 2025

Methinks Software S.L.
% Claudia Carbonell
Senior QA/RA Manager
Methinks Software, S.L.
Pier 01, Plaça Pau Vila 1, Office 2D2
Barcelona, 08039, Spain

Re: K250685
Trade/Device Name: Methinks NCCT Stroke
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological Computer Aided Triage And Notification Software
Regulatory Class: Class II
Product Code: QAS
Dated: May 12, 2025
Received: May 12, 2025

Dear Nima Akhlaghi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K250685 - Nima Akhlaghi
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

{2}

K250685 - Nima Akhlaghi
Page 3

the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb, Ph.D.
Imaging Software Team
Assistant Director
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)

K250685

Device Name

Methinks NCCT Stroke

Indications for Use (Describe)

Methinks NCCT Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of (1) non-contrast head CT (NCCT) images. The device is intended to assist hospital networks and trained physicians in workflow triage by flagging and communicating suspected positive findings of (1) Intracranial Hemorrhage (ICH) and (2) Large Vessel Occlusion (LVO) of the ICA, MCA-M1 and MCA-M2.

Methinks NCCT Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with suspected (1) ICH and (2) LVO in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected ICH or LVO findings via PACS and/or notifications. Notifications include preview images that are meant for informational purposes only, and are not intended for diagnostic use beyond notification.

The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Methinks NCCT Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care. Methinks NCCT Stroke is for adults only.

**Cautions:**

- All patients should get adequate care for their symptoms including CTA and/or other appropriate care per the standard clinical practice, irrespective of the device output.
- The device is not intended to be a rule-out device and for cases that have been processed by the device without notification for "LVO Suspected" should not be viewed as indicating that LVO is excluded. All cases should undergo CTA, per the standard stroke workup.

**Limitations**

- The device does not replace the need for CTA in ischemic stroke workup, it provides workflow prioritization and notification only.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}

N

K250685

# 510(k) Summary

The content included in the 510(k) Summary of the Methinks NCCT Stroke device is based on the 21 CFR 807.92 requirements.

# I. SUBMITTER

Name: Methinks Software, SL

Address: Pier 01, Office 2D2, Plaça de Pau Vila
08039 Barcelona, Spain

Contact Person: Claudia Carbonell

Date Prepared: May 7, 2025

# II. DEVICE

Name of Device: Methinks NCCT Stroke

Common or Usual Name: Methinks NCCT Stroke

Classification Name: Radiological computer aided triage and notification software

Regulatory Class: Class II

Product Code: QAS

Regulation Number: 21 CFR §892.2080

# III. PREDICATE DEVICE

The Methinks NCCT Stroke device is claimed to be substantially equivalent to the following legally marketed predicate device: iSchemaView, Inc, Rapid NCCT Stroke, K222884.

Name of the device: Rapid NCCT Stroke

Classification Name: Radiological computer aided triage and notification software

Regulatory Class: Class II

Product Code: QAS

Regulation Number: 21 CFR §892.2080

# IV. DEVICE DESCRIPTION

Methinks NCCT Stroke is a radiological computer-assisted triage and notification software device. The device receives Non-Contrast Computed Tomography (NCCT) images and processes them to provide triage and notification prioritization of suspected Intracranial Hemorrhage (ICH) and Large Vessel Occlusion (LVO) of the ICA, MCA-M1 and MCA-M2. The Methinks NCCT Stroke device is an AI/ML Software as a Medical Device. The outputs of the device are intended to be used by trained clinicians in the prioritization of patients with suspected ICH and/or LVO.

{5}

2

# V. INDICATIONS FOR USE

Methinks NCCT Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of (1) non-contrast head CT (NCCT) images. The device is intended to assist hospital networks and trained physicians in workflow triage by flagging and communicating suspected positive findings of (1) Intracranial Hemorrhage (ICH) and (2) Large Vessel Occlusion (LVO) of the ICA, MCA-M1 and MCA-M2.

Methinks NCCT Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with suspected (1) ICH and (2) LVO in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected ICH or LVO findings via PACS and/or notifications. Notifications include preview images that are meant for informational purposes only, and are not intended for diagnostic use beyond notification.

The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Methinks NCCT Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care. Methinks NCCT Stroke is for adults only.

## Cautions:

- All patients should get adequate care for their symptoms including CTA and/or other appropriate care per the standard clinical practice, irrespective of the device output.
- The device is not intended to be a rule-out device and for cases that have been processed by the device without notification for "LVO Suspected" should not be viewed as indicating that LVO is excluded. All cases should undergo CTA, per the standard stroke workup.

## Limitations:

- The device does not replace the need for CTA in ischemic stroke workup, it provides workflow prioritization and notification only.

# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Methinks NCCT Stroke does not raise new questions of safety or effectiveness compared to the previously cleared Rapid NCCT Stroke (K222884). Both devices are radiological computer-aided triage and notification software applications for determining suspicion of ICH and LVO. Both devices use NCCT images for the determination of suspected ICH and LVO.

Compared to the predicate device, the subject device identifies an additional type of occlusion MCA-M2 (the predicate identifies ICA and MCA-M1 only) which does not raise a substantial difference in the indications for use nor different questions of safety or effectiveness.

The benefits include the shorter time to a suspected LVO notification by assessing non-contrast imaging immediately, then CTA follow-up per standard of care for stroke workup. Thus, the Methinks NCCT Stroke is substantially equivalent to the Rapid NCCT Stroke device.

{6}

3

The following table summarizes and compares Methinks NCCT Stroke and the Rapid NCCT Stroke device (predicate device). A table comparing the key features of the subject and the predicate devices is provided below:

Table 1. Substantial Equivalence Table.

|  Substantial Equivalence Table  |   |   |
| --- | --- | --- |
|  Comparison Feature | Predicate Device: Rapid NCCT Stroke | Subject Device: Methinks NCCT Stroke  |
|  Product Code | QAS | QAS  |
|  Regulation | 21 CFR §892.2080 | 21 CFR §892.2080  |
|   | Rapid NCCT Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of (1) nonenhanced head CT (NCCT) images. The device is intended to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of (1) head CT images for Intracranial Hemorrhage (ICH) and (2) NCCT large vessel occlusion (LVO) of the ICA and MCA-M1.

Rapid NCCT Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected (1) ICH or (2) NCCT LVO on the Rapid server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected ICH or LVO findings via PACS, email or mobile device. Notifications include compressed preview images that are meant for informational purposes only, and are not intended for diagnostic use beyond notification.

The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Rapid NCCT Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care. Rapid NCCT Stroke is for Adults only.

Cautions:
• All patients should get adequate care for their symptoms including CTA and/or other appropriate care per the standard clinical practice, irrespective of the device output
• The device is not intended to be a rule-out device and for cases that have been processed by the device without notification for “Suspected LVO” should not be viewed as indicating that LVO is excluded. All cases should undergo CTA, per the standard stroke workup.

Limitations:
• Rapid NCCT Stroke does not replace the need for CTA or MRA in ischemic stroke workup, it provides workflow prioritization and notification only. | Methinks NCCT Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of (1) non-contrast head CT (NCCT) images. The device is intended to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of (1) Intracranial Hemorrhage (ICH) and (2) Large Vessel Occlusion (LVO) of the ICA, MCA-M1 and MCA-M2.

Methinks NCCT Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected (1) ICH and (2) LVO in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected ICH or LVO findings via PACS and/or notifications. Notifications include preview images that are meant for informational purposes only, and are not intended for diagnostic use beyond notification.

The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Methinks NCCT Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care. Methinks NCCT Stroke is for adults only.

Cautions:
• All patients should get adequate care for their symptoms including CTA and/or other appropriate care per the standard clinical practice, irrespective of the device output.
• The device is not intended to be a rule-out device and for cases that have been processed by the device without notification for “LVO Suspected” should not be viewed as indicating that LVO is excluded. All cases should undergo CTA, per the standard stroke workup.

Limitations:
• The device does not replace the need for CTA in ischemic stroke workup, it provides workflow prioritization and notification only.  |

{7}

4

|   | • Rapid ICH has been shown to reliably identify hemorrhages of ≥ 0.4ml.
Contraindications/Exclusions
• Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate.
• Hemorrhagic Transformation, Hematoma
• Very thin or no Ventricles. |   |
| --- | --- | --- |
|  User | Trained Clinicians | Trained Physicians  |
|  Anatomy | Head | Head  |
|  Input Data | NCCT images | NCCT images  |
|  Technology | AI/ML/Neural Network | AI/ML/Neural Network  |
|  Segmentation of ROI | The device does not highlight or direct a user's attention to a specific location in the image file. | The device does not highlight or direct a user's attention to a specific location in the image file.  |
|  Preview Images | Presentation of a preview of the study for initial assessment not meant for diagnostic purposes.
The device operates in parallel with the standard of care. | Presentation of a preview of the study for initial assessment not meant for diagnostic purposes.
The device operates in parallel with the standard of care.  |
|  Annotation/Localization | Device does not mark, highlight, or direct users' attention to a specific location in the original image. | Device does not mark, highlight, or direct users' attention to a specific location in the original image.  |
|  Prioritization Notification | Yes | Yes  |
|  Clinical SoC Workflow | In parallel to | In parallel to  |
|  Technical Pipeline | Two cascaded functions (ICH then LVO) using three integrated algorithms. | Single algorithm with two different outputs (ICH and LVO).  |
|  Removal of Cases from SoC review | No | No  |

VII. PERFORMANCE STANDARDS AND DATA

Performance Data

The Methinks NCCT Stroke device underwent rigorous non-clinical testing to confirm that it meets all specified design requirements under clinically relevant conditions. Software performance, validation and verification testing demonstrated that the Methinks NCCT Stroke device meets all design requirements and specifications. Design Verification and Validation according to 21 CFR Part 820.30 passed. Cybersecurity has been considered and addressed as part of the software verification process, in line with Section 524B of the FD&amp;C Act as a Cyber device. Methinks implemented a risk-based cybersecurity strategy including secure design principles, vulnerability assessment, Software Bill of Materials (SBOM), and penetration testing. These activities ensure the software's safety, effectiveness, and robustness against potential cybersecurity threats.

Methinks conducted a retrospective, blinded, multicenter, multinational study with the Methinks NCCT Stroke device with the primary endpoint to evaluate the software's performance in identifying NCCT head images containing intracranial hemorrhage (ICH) and

{8}

5

Large Vessel Occlusion (LVO) findings in 358 (ICH Positive: 132, ICH Negative: 226) for the ICH and 335 cases for LVO (LVO Positive: 110, LVO Negative: 225).

Sensitivity and specificity exceed the pre-specified performance goals for ICH and LVO. Specifically, ICH performance was observed at Se: 94.7% (95% CI: 89.3% - 97.8%), Sp: 99.5% (95% CI: 97.5% - 99.9%) and LVO was observed at Se: 76.4% (95% CI: 67.3% - 83.9%) and Sp: 91.1% (95% CI: 86.6% - 94.5%).

In addition, a reader study was conducted to compare NCCT LVO sensitivity of the device to that of radiologists. Secondary endpoints of expert non-inferiority and non-expert superiority were used for sensitivity. Methinks NCCT Stroke passed both conditions. LVO sensitivity (Se) for expert neuroradiologists was 50.0% (40.1% - 59.9%), with a difference in Se between Methinks NCCT Stroke and experts of 23.6% (95%CI: 8.5% - 38.7%), showing superiority of the Methinks device; Se for general radiologists (non-experts) was 37.7% (95% CI: 28.5% - 47.7%), and with a difference in Se between Methinks NCCT Stroke and non-experts of 35.9% (95% CI: 16.0% - 42.9%), also showing superiority of the Methinks device. The results of the reader study are included in the following table.

Table 2. Performance of the readers vs Methinks NCCT-LVO in the reader study.

|   | Sensitivity (95% CI) | Specificity (95% CI)  |
| --- | --- | --- |
|  Methinks NCCT-LVO | 73.6% (59.7% - 84.7%) | 89.1% (82.0% - 94.1%)  |
|  R1 + R2 (Experts) | 50.0% (40.1% - 59.9%) | 92.0% (87.8% - 95.1%)  |
|  R3 + R4 (Non-experts) | 37.7% (28.5% - 47.7%) | 87.4% (82.5% - 91.3%)  |
|  R1 + R2 + R3 + R4 (All) | 43.9% (37.1% - 50.8%) | 89.7% (86.6% - 92.3%)  |

The Methinks NCCT Stroke time to notification analysis includes the time to get the DICOM exam, de-identify it (if required), analyze and send a notification back to the PACS and notification clients. Methinks NCCT Stroke time to notification has been documented for all cases with results shown in the following table.

Table 3. Time to notification of Methinks NCCT Stroke.

|  Parameter | Mean | 95% CI Lower | 95% CI Upper  |
| --- | --- | --- | --- |
|  Time to notification of NCCT-ICH (minutes) | 1.43 | 1.36 | 1.50  |
|  Time to notification of NCCT-LVO (minutes) | 1.42 | 1.36 | 1.48  |

Performance across subgroups

Subgroup analyses were conducted for both the LVO and ICH indication, to support the generalizability of the device across intended patient population and device input.

Performance metrics by subgroups for NCCT-ICH

Table 4. performance metrics NCCT-ICH by Gender.

|  Gender | Measure | N | Estimate | Lower 95% CI | Upper 95% CI  |
| --- | --- | --- | --- | --- | --- |
|  Male | Sensitivity | 76 | 97.4% | 90.8% | 99.7%  |

{9}

6

|   | Specificity | 119 | 99.2% | 95.4% | 100.0%  |
| --- | --- | --- | --- | --- | --- |
|  Female | Sensitivity | 55 | 90.9% | 80.0% | 97.0%  |
|   |  Specificity | 107 | 100.0% | 96.6% | 100.0%  |
|  Unspecified | Sensitivity | 1 | 100.0% | 2.5% | 100.0%  |
|   |  Specificity | 0 | - | - | -  |

Table 5. performance metrics NCCT-ICH by Age.

|  Age | Measure | N | Estimate | Lower 95% CI | Upper 95% CI  |
| --- | --- | --- | --- | --- | --- |
|  < 50 years | Sensitivity | 17 | 94.1% | 71.3% | 99.9%  |
|   |  Specificity | 43 | 97.7% | 87.7% | 99.9%  |
|  50 - 70 years | Sensitivity | 53 | 98.1% | 89.9% | 100.0%  |
|   |  Specificity | 98 | 100.0% | 96.3% | 100.0%  |
|  > 70 years | Sensitivity | 62 | 91.9% | 82.2% | 97.3%  |
|   |  Specificity | 85 | 100.0% | 95.8% | 100.0%  |

Table 6. performance metrics NCCT-ICH by Slice Thickness.

|  Slice Thickness | Measure | N | Estimate | Lower 95% CI | Upper 95% CI  |
| --- | --- | --- | --- | --- | --- |
|  1mm ≤ Slice Thickness < 2.5mm | Sensitivity | 37 | 97.3% | 85.8% | 99.9%  |
|   |  Specificity | 77 | 98.7% | 93.0% | 100.0%  |
|  2.5mm ≤ Slice Thickness < 4mm* | Sensitivity | 30 | 83.3% | 65.3% | 94.4%  |
|   |  Specificity | 36 | 97.2% | 85.5% | 99.9%  |
|  4mm ≤ Slice Thickness ≤ 5.5mm | Sensitivity | 95 | 93.7% | 86.8% | 97.6%  |
|   |  Specificity | 149 | 100.0% | 97.6% | 100.0%  |

2.5mm ≤ Slice Thickness &lt; 4mm* are additional cases. This is why if you sum all N's it will exceed the sample size of the study.

Table 7. performance metrics NCCT-ICH by Vendor Machine.

|  Vendor Machine | Measure | N | Estimate | Lower 95% CI | Upper 95% CI  |
| --- | --- | --- | --- | --- | --- |
|  Siemens | Sensitivity | 83 | 92.8% | 84.9% | 97.3%  |
|   |  Specificity | 146 | 100.0% | 97.5% | 100.0%  |
|  General Electric* | Sensitivity | 48 | 93.8% | 85.7% | 97.5%  |
|   |  Negatives | 22 | 100.0% | 84.6% | 100.0%  |
|  Philips | Sensitivity | 47 | 97.9% | 88.7% | 99.9%  |
|   |  Specificity | 80 | 98.8% | 93.2% | 100.0%  |
|  Toshiba / Canon* | Sensitivity | 12 | 100.0% | 73.5% | 100.0%  |
|   |  Negatives | 7 | 85.7% | 42.1% | 99.6%  |

{10}

7

General Electric and Toshiba/Canon* are additional cases. This is why if you sum all N's it will exceed the sample size of the study.

Table 8. performance metrics NCCT-ICH by hemorrhage volume.

|  Hemorrhage volume | Measure | N | Estimate | Lower 95% CI | Upper 95% CI  |
| --- | --- | --- | --- | --- | --- |
|  Smaller than 1 ml | Sensitivity | 10 | 80.0% | 44.4% | 97.5%  |
|   |  Specificity | 0 | - | - | -  |
|  Between 1 ml and 5 ml | Sensitivity | 50 | 92.0% | 80.8% | 97.8%  |
|   |  Specificity | 0 | - | - | -  |
|  Bigger than 5 ml | Sensitivity | 95 | 96.8% | 91.0% | 99.3%  |
|   |  Specificity | 0 | - | - | -  |

Table 9. performance metrics NCCT-ICH by hemorrhage sub-type.

|  Hemorrhage sub type | Measure | N | Estimate | Lower 95% CI | Upper 95% CI  |
| --- | --- | --- | --- | --- | --- |
|  Epidural hemorrhage* | Sensitivity | 3 | 100.0% | 29.2% | 100.0%  |
|   |  Specificity | 0 | - | - | -  |
|  Subdural hemorrhage* | Sensitivity | 21 | 85.7% | 63.7% | 97.0%  |
|   |  Specificity | 0 | - | - | -  |
|  Subarachnoid hemorrhage | Sensitivity | 13 | 92.3% | 64.0% | 99.8%  |
|   |  Specificity | 0 | - | - | -  |
|  Intraparenchymal hemorrhage | Sensitivity | 112 | 97.3% | 92.4% | 99.4%  |
|   |  Specificity | 0 | - | - | -  |
|  Intraventricular hemorrhage | Sensitivity | 27 | 100.0% | 87.2% | 100.0%  |
|   |  Specificity | 0 | - | - | -  |

Epidural hemorrhage and subdural hemorrhage* are additional cases. This is why if you sum all N's it will exceed the sample size of the study.

Performance metrics by subgroups for NCCT-LVO

Table 10. performance metrics NCCT-LVO by Gender.

|  Gender | Measure | N | Estimate | Lower 95% CI | Upper 95% CI  |
| --- | --- | --- | --- | --- | --- |
|  Female | Sensitivity | 56 | 76.8% | 63.6% | 87.0%  |
|   |  Specificity | 109 | 93.6% | 87.2% | 97.4%  |
|  Male | Sensitivity | 54 | 75.9% | 62.4% | 86.5%  |
|   |  Specificity | 116 | 88.8% | 81.6% | 93.9%  |

{11}

8

Table 11. performance metrics NCCT-LVO by Age.

|  Age | Measure | N | Estimate | Lower 95% CI | Upper 95% CI  |
| --- | --- | --- | --- | --- | --- |
|  < 50 | Sensitivity | 7 | 85.7% | 42.1% | 99.6%  |
|   |  Specificity | 38 | 100.0% | 90.7% | 100.0%  |
|  50 - 70 | Sensitivity | 43 | 74.4% | 58.8% | 86.5%  |
|   |  Specificity | 103 | 91.3% | 84.1% | 95.9%  |
|  > 70 | Sensitivity | 60 | 76.7% | 64.0% | 86.6%  |
|   |  Specificity | 84 | 86.9% | 77.8% | 93.3%  |

Table 12. performance metrics NCCT-LVO by Slice Thickness.

|  Slice Thickness | Measure | N | Estimate | Lower 95% CI | Upper 95% CI  |
| --- | --- | --- | --- | --- | --- |
|  1mm ≤ Slice Thickness < 2.5mm | Sensitivity | 25 | 72.0% | 50.6% | 87.9%  |
|   |  Specificity | 78 | 91.0% | 82.4% | 96.3%  |
|  2.5mm ≤ Slice Thickness < 4mm | Sensitivity | 35 | 77.1% | 59.9% | 89.6%  |
|   |  Specificity | 45 | 84.4% | 70.5% | 93.5%  |
|  4mm ≤ Slice Thickness ≤ 5.5mm | Sensitivity | 50 | 78.0% | 64.0% | 88.5%  |
|   |  Specificity | 102 | 94.1% | 87.6% | 97.8%  |

Table 13. performance metrics NCCT-LVO by Vendor Machine.

|  Vendor Machine | Measure | N | Estimate | Lower 95% CI | Upper 95% CI  |
| --- | --- | --- | --- | --- | --- |
|  Siemens | Sensitivity | 36 | 75.0% | 57.8% | 87.9%  |
|   |  Specificity | 86 | 96.5% | 90.1% | 99.3%  |
|  General Electric | Sensitivity | 19 | 73.7% | 48.8% | 90.9%  |
|   |  Specificity | 28 | 82.1% | 63.1% | 93.9%  |
|  Philips | Sensitivity | 34 | 76.5% | 58.8% | 89.3%  |
|   |  Specificity | 82 | 86.6% | 77.3% | 93.1%  |
|  Toshiba / Canon | Sensitivity | 21 | 81.0% | 58.1% | 94.6%  |
|   |  Specificity | 29 | 96.6% | 82.2% | 99.9%  |

Table 14. performance metrics NCCT-LVO by LVO subgroups.

|  LVO Subgroups | Measure | N | Estimate | Lower 95% CI | Upper 95% CI  |
| --- | --- | --- | --- | --- | --- |
|  ICA | Sensitivity | 36 | 77.8% | 60.8% | 89.9%  |
|   |  Specificity | 0 | - | - | -  |
|  MCA-M1 | Sensitivity | 44 | 88.6% | 75.4% | 96.2%  |
|   |  Specificity | 0 | - | - | -  |

{12}

1

|  ICA + MCA-M1 (as predicate) | Sensitivity | 77 | 83.1% | 72.9% | 90.7%  |
| --- | --- | --- | --- | --- | --- |
|   |  Specificity | 0 | - | - | -  |
|  MCA-M2 | Sensitivity | 25 | 68.0% | 46.5% | 85.1%  |
|   |  Specificity | 0 | - | - | -  |
|  LVO without ischemia | Sensitivity | 33 | 42.4% | 25.5% | 60.8%  |
|   |  Specificity | 0 | - | - | -  |

Out of the 33 LVO without ischemia, 18 were also tagged by radiologists only reading the NCCT image. In this subgroup where we can compare our algorithm with them the results were:

Table 15. Performance of the Methinks NCCT-LVO and Radiologists in the LVO without ischemia subgroup.

|  Output | Sensitivity (95% CI)  |
| --- | --- |
|  Methinks NCCT-LVO | 38.8% (17.3% - 64.3%)  |
|  Radiologist 1 | 33.3% (13.3% - 59.0%)  |
|  Radiologist 2 | 22.2% (6.4% - 47.6%)  |
|  Radiologist 3 | 16.6% (3.6% - 41.4%)  |
|  Radiologist 4 | 11.1% (1.4% - 34.7%)  |

Table 16. performance metrics NCCT-LVO by non-LVO subgroups.

|  Non-LVO Subgroups | Measure | N | Estimate | Lower 95% CI | Upper 95% CI  |
| --- | --- | --- | --- | --- | --- |
|  Tumors | Sensitivity | 0 | - | - | -  |
|   |  Specificity | 2 | 50.0% | 1.3% | 98.7%  |
|  Stenosis > 51% | Sensitivity | 0 | - | - | -  |
|   |  Specificity | 28 | 85.7% | 67.3% | 96.0%  |
|  Intracranial Hemorrhage (ICH) | Sensitivity | 0 | - | - | -  |
|   |  Specificity | 3 | 100.0% | 29.2% | 100.0%  |
|  Other Acute Ischemic Events | Sensitivity | 0 | - | - | -  |
|   |  Specificity | 62 | 79.0% | 66.8% | 88.3%  |
|  Chronic Lesions | Sensitivity | 0 | - | - | -  |
|   |  Specificity | 104 | 89.4% | 81.9% | 94.6%  |
|  No Findings (none of the above) | Sensitivity | 0 | - | - | -  |
|   |  Specificity | 160 | 95.6% | 91.2% | 98.2%  |

In summary, the performance validation data demonstrated that the proposed device provides accurate detection of acute ICH and LVO under a range of clinically relevant variables

{13}

10

associated with the intended use of the device. The notification functionality of the device also met the acceptance criteria.

## Safety and Effectiveness

Methinks NCCT Stroke has been designed, verified and validated in compliance with 21 CFR Part 820.30 requirements. The device has been designed to meet the requirements associated with EN ISO 14971:2019 (risk management) and the software development process conforms to ISO 62304:2015. The Methinks NCCT Stroke performance has been validated through retrospective case data based on expert reader truthing of the data and reader testing.

## Substantial Equivalence Discussion

Both devices have identical intended use and similar technological characteristics/features and roles within a clinical workflow relative to triage and notification. Both devices are intended to automatically process and analyze non-contrast CT scans to provide a notification to users in case of a suspected LVO or ICH being identified.

The subject device does not introduce any new risks when compared to the predicate and both include similar mitigation strategies for reducing the risk of off label use of preview images which are shared via mobile.

## VIII. CONCLUSIONS

In conclusion, Methinks NCCT Stroke is substantially equivalent in technological characteristics, safety, and performance characteristics to the legally marketed predicate device, Rapid NCCT Stroke (K222884).

---

**Source:** [https://fda.innolitics.com/device/K250685](https://fda.innolitics.com/device/K250685)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
