LoFric Origo is indicated for short- and long-term bladder management with intermittent urinary catheterization.
Device Story
LoFric Origo is a sterile, single-use, hydrophilic-coated urinary catheter for intermittent bladder drainage. Designed for male and pediatric anatomy, it features Nelaton, Tiemann (Coudé), or Flexible (straight ball-shaped) tips. The device includes an integrated sterile water sachet for hydrophilic coating activation, plus an adjustable insertion grip or extendable protective sleeve to facilitate a non-touch technique. Used in clinical or home settings by patients or caregivers. Operation involves breaking the water sachet to wet the catheter, inserting through the urethra into the bladder, draining urine, and disposing of the device. The non-touch aids reduce direct handling of the catheter shaft, potentially lowering contamination risk. Output is the drainage of urine from the bladder. Clinical benefit includes effective bladder management for patients requiring intermittent catheterization.
Clinical Evidence
No clinical data included. Substantial equivalence supported by non-clinical bench testing, including biocompatibility, sterile barrier integrity, and mechanical performance (flow rate, connector security, kink stability, coating quality, tensile strength, and insertion force).
Technological Characteristics
Materials: Polyolefin-based elastomer (POBE) tubing, PVP/NaCl hydrophilic coating, medical-grade TPS funnel. Dimensions: 30cm and 40cm lengths, French sizes CH8-CH18. Sterilization: E-beam. Packaging: Multilayer cast coextruded film with resealable label. Standards: ISO 10993-1 (biocompatibility), ISO 11607-1/2 (packaging), ISO 11137-1/2 (sterilization), ISO 20696 (catheter performance).
Indications for Use
Indicated for short- and long-term bladder management via intermittent urinary catheterization in male and pediatric patients.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 26, 2025
Wellspect AB
Emad Ramzi
Sr. Regulatory Affairs Manager
Amniogatan 1, P.O.Box 14
SE-431 51 Sweden
Mölndal
SWEDEN
Re: K250659
Trade/Device Name: LoFric Origo
Regulation Number: 21 CFR 876.5130
Regulation Name: Urological Catheter and accessories
Regulatory Class: II
Product Code: EZD
Dated: March 5, 2025
Received: August 28, 2025
Dear Emad Ramzi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K250659 - Emad Ramzi
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{3}
K250659 - Emad Ramzi
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica K. Nguyen -S
Jessica K. Nguyen
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250659
Device Name
LoFric Origo
Indications for Use (Describe)
LoFric Origo is indicated for short- and long-term bladder management with intermittent urinary catheterization.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.**
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{5}
K250659
Page 1 of 7
510(k) SUMMARY for LoFric Origo™
1. Submitter Information:
Wellspect AB
Aminogatan 1
P.O. Box 14
431 21 Mölndal
Contact Person: Emad Ramzi
Telephone Number: +46 31 376 32 41
Date Prepared: September 26, 2025
2. Device Name:
- Proprietary Name: LoFric Origo™
- Common Name: Intermittent urinary catheter
- Classification Name: Urological catheter and accessories
- CFR Number: 21 CFR 876.5130
- Device Class: Class II
- Product Code: EZD, Catheter Straight
3. Predicate and Reference Devices:
| Type | Device Name | 510(k) | Company Name |
| --- | --- | --- | --- |
| Predicate device | LoFric® Primo™ | K122078 | Wellspect HealthCare
(formerly Astra Tech AB) |
4. Description of Device:
The subject device: LoFric® Origo™ is a range of sterile, single-use, urinary catheters designed as an intermittent pathway for drainage of the bladder. The catheters, designed to accommodate the individual anatomy of male users, are available in various lengths and French sizes, and come in three tip designs: Nelaton (straight tip), Flexible (straight ball-shaped tip), and Tiemann (Coudé tip). Each catheter is individually packaged together with an adjustable insertion grip, or an extendable protective sleeve attached to the catheter along with a sterile water sachet (packet). The adjustable insertion grip or the extendable protective sleeve are provided to minimize touching the catheter shaft directly during insertion and retraction. The water sachet is provided for wetting of the hydrophilic surface and is to be broken immediately before use in order to soak the tubing.
Page 1 of 7
{6}
K250659
Page 2 of 7
5. Indications for Use:
LoFric® Origo™ is indicated for short- and long-term bladder management with intermittent urinary catheterization.
6. Intended use population:
LoFric® Origo™ catheters are intended for male, and pediatric patients (adolescents and children).
7. Comparison of Technological Characteristics:
The purpose of this Traditional 510(k) is to gain U.S. premarket clearance for the modification to the predicate device LoFric® Primo™ (K122078).
- The subject and predicate devices have the same intended use for intermittent urinary catheterization.
- The subject and predicate devices are sterile, single use, hydrophilic coated catheters, use the same operational principle and incorporating the same tubing and coating material.
- The subject and predicate devices have similar packaging configurations, each including a separate water sachet for activation, and are sterilized using the same sterilization method (e-beam).
The key differences between the predicate device (K122078) and the subject device are:
- Minor adjustment to the device’s indication for use phrase, without affecting the substance or meaning of the indication previously cleared under (K122078).
- Replacing the previously cleared optional insertion aid with an adjustable insertion grip or an extendable protective sleeve, assembled on the catheter body to facilitate a non-touch technique.
- Introduction of a 30 cm Nelaton tip catheter in French sizes: CH.8, 10 and 12.
- Introduction of a 40 cm Flexible, straight ball-shaped tip catheter in French sizes: CH 12, 14 and 16.
- Modification in the single package material and design.
- Enhancing the previously cleared funnel material to a medical-grade version, incorporating FDA-listed color additives.
- Use of a different dip-coating solvent.
- Updates made to Instructions for Use to include pictorial instructions, include device warnings/precautions and adverse reaction information for patient awareness.
An overview of the similarities and differences between the subject device and predicate devices is given in Table FS-1.
8. Non-Clinical Performance Data
Non-clinical testing data that was submitted, referenced, or relied upon to demonstrate substantial equivalence includes:
a) Qualification of biological safety assessment according to ISO 10993-1:2018, Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process and FDA Guidance “Use of International Standard ISO 10993-1”. Testing performed included Chemical characterization, Cytotoxicity, Irritation, Sensitization, Genotoxicity, and Pyrogenicity.
Page 2 of 7
{7}
K250659
Page 3 of 7
b) Test methods for confirmation of sterile barrier systems and packaging validation testing following requirements in:
- ISO 11607-1:2019, Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems.
- ISO 11607-2:2019, Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes.
- ASTM D4169-22, Standard Practice for Performance Testing of Shipping Containers and Systems.
- ASTM D4332-14, Standard Practice for Conditioning Containers Packages or Packaging Components for Testing.
- ASTM F1929-15, Standard test method for detecting seal leaks in porous medical packaging by dye penetration.
- ASTM F88/F88M-23, Standard Test Method for Seal Strength of Flexible Barrier Materials.
c) Sterilization validation was performed according to
- ISO 11137-1:2006/(R)2015, Sterilization of health care products – Radiation – Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- ISO 11137-2:2013, Sterilization of health care products – Radiation – Part 2: Establishing the sterilization dose.
d) Shelf life and Performance bench testing was performed per applicable sections of ISO 20696:2018, Sterile urethral catheters for single use, and internal test methods to document the following properties of the catheters:
- Flow rate was verified by the test method in Annex E of ISO 20696:2018.
- Connector security was verified by the test method in Annex B of ISO 20696:2018.
- Kink stability was verified by the test method in Annex G of ISO 20696:2018.
- Coating quality was verified by internal test method.
- Catheter strength was verified by the test method in Annex A of ISO 20696:2018.
- Peak tensile force was verified by the test method in Annex H of ISO 20696:2018.
- Water retention was verified by internal test method.
- Flexible tip bending stiffness was verified by internal test method.
- Flexible tip insertion force was verified by internal test method.
- Catheter Sleeve insertion aid strength was verified by internal test method.
- Packaging sterile integrity was verified by internal test method.
- Primary package seal strength was verified by internal test method.
All tests met the pre-determined acceptance criteria.
The results of the testing show substantial equivalence in performance of the subject device when compared to the predicate device (K122078).
Page 3 of 7
{8}
K250659
Page 4 of 7
9. Clinical Performance Data
No data from human clinical studies has been included to support the substantial equivalence of the subject LoFric® Origo™
10. Conclusion
The subject device and the predicate device (K122078) share the same intended use and have similar indications for use. Both are sterile, single-use catheters that incorporate the same basic design, use the same or similar materials, and have a similar packaging configuration, i.e., a urinary catheter with a sterile water sachet for activation.
Performance data have been included to address the safety and effectiveness of the subject devices. The results of non-clinical bench testing, combined with the design, biocompatibility, and intended use comparison to the predicate device (K122078), demonstrate that the subject devices are as safe, as effective, and perform as well as the legally marketed device.
Page 4 of 7
{9}
K250659
Table FS-1: Comparison of technological characteristics between subject and predicate devices
| Element | Subject Device | Predicate Device | Discussion |
| --- | --- | --- | --- |
| | LoFric® Origo™ | LoFric® Primo™ (K122078) | |
| Product Code | EZD | EZD | Same |
| Device for Prescription use | Yes | Yes | Same |
| Indications for Use | LoFric Origo is indicated for short- and long-term bladder management with intermittent urinary catheterization. | LoFric Primo is indicated for intermittent urinary catheterization. | Both the subject and predicate devices are used for intermittent urinary catheterization. The additional phrase "short- and long-term bladder management" in the subject device indication clarifies that the device is suitable for patients who use intermittent catheterization for both short and long periods. However, it does not imply that the catheter itself is meant for long-term indwelling use. |
| Anatomical site | Bladder through the Urethra | Bladder through the Urethra | Same |
| Patient population | Intended for male, and pediatric patients. | Intended for male, female and pediatric patients. | The patient population for the subject device excludes the female population. |
| Principal of operation | • Squeeze water pocket • Peel pack open • Insert catheter • Empty bladder • Withdraw catheter • Dispose device | • Squeeze water pocket • Peel pack open • Insert catheter • Empty bladder • Withdraw catheter • Dispose device | Same |
| Catheter tip Configuration | Nelaton (Straight) tip, Tiemann (Coudé) tip & Flexible (Straight ball-shaped) tip | Nelaton (Straight) tip & Tiemann (Coudé) tip | The subject device includes an additional tip configuration, i.e., a Flexible (straight ball-shaped) tip. |
{10}
K250659
Page 6 of 7
| Element | Subject Device | Predicate Device | Discussion |
| --- | --- | --- | --- |
| | LoFric® Origo™ | LoFric® Primo™ (K122078) | |
| | | | The Flexible tip insertion force and bending stiffness was evaluated to support substantial equivalence. |
| Type/Sizes | Nelaton with insertion grip
CH.8, 10 & 12
in 30cm length | Nelaton
CH.8, 10, 12, & 14
in 15cm length | The subject LoFric Origo catheters are available in 30 cm and 40 cm lengths. |
| | Nelaton with insertion grip
CH.10, 12, 14, 16 & 18
in 40cm length | Nelaton
CH.6, 8, 10, 12, 14, 16, & 14
in 20cm length | The 30 cm catheter has been introduced to the LoFric Origo line to enhance handling for male users. |
| | Flexible with insertion grip
CH.12, 14 & 16
in 40cm length | Nelaton
CH.8, 10, 12, 14, 16, & 18
in 40cm length | Both the 30 cm and 40 cm catheters, available in French sizes CH8 to CH18, come with an insertion grip to facilitate the non-touch technique. |
| | Tiemann with insertion grip
CH.10, 12, 14, 16 & 18
in 40cm length | Tiemann
CH.10, 12, 14, 16 &18
in 40cm length | The protective sleeve is included only with the 40 cm catheters in French sizes CH12, CH14, and CH16. |
| | Nelaton with protective Sleeve
CH.12, 14 & 16
in 40cm length | | |
| | Flexible with protective Sleeve
CH.12, 14 & 16
in 40cm length | | |
| Catheter tube Material | Polyolefin based elastomer (POBE) | Polyolefin based elastomer (POBE) | Same |
| Hydrophilic surface coating | Polyvinylpyrrolidone (PVP) & NaCl | Polyvinylpyrrolidone (PVP) & NaCl | Same |
| Hydrophilic coating activation | Integrated water sachet with sterile water & NaCL | Integrated water sachet with sterile water & NaCL | Same |
| Catheter funnel | Color coded end-funnel | Color coded end-funnel | Same base material (TPS), different material grade as predicate device. |
Page 6 of 7
{11}
K250659
Page 7 of 7
| Element | Subject Device | Predicate Device | Discussion |
| --- | --- | --- | --- |
| | LoFric® Origo™ | LoFric® Primo™ (K122078) | |
| | | | Biocompatibility testing was performed to address material formulation differences between the subject and predicate device in support of substantial equivalence. |
| **Catheter insertion guide** | Adjustable Insertion Grip
or
an Extendable Protective Sleeve for no-touch functionality. | Optional Insertion Aid, for no-touch functionality. | The subject device insertion grip or protective sleeve both have the same functionality as the predicate device insertion aid to prevents direct contact with the catheter shaft during use. |
| **Single Package** | Multilayer cast coextruded film with:
a polyolefin/polyethylene bottom foil,
a polyester/ polyethylene top foil and a
polyester resealable label | Three-layer co-extruded polyethylene (PE) film | Different packaging materials and designs have been used. Packaging testing has been conducted and included to address material formulation and sterility integrity, supporting substantial equivalence. |
| **Resealable label** | Yes | No | The subject device features a resealable label for accessing the catheter. |
| **Sterilization method** | E-beam | E-beam | Same |
| **Condition of Use** | Single Use | Single Use | Same |
| **Shelf Life** | 2 years | 2 years | Same |
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.