K250629 · Hathor Professional Skincare , Ltd. · NUC · May 2, 2025 · Obstetrics/Gynecology
Device Facts
Record ID
K250629
Device Name
Sutil Rich Personal Lubricant
Applicant
Hathor Professional Skincare , Ltd.
Product Code
NUC · Obstetrics/Gynecology
Decision Date
May 2, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
Sutil Rich Personal Lubricant is a personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. The product is not compatible with polyurethane condoms.
Device Story
Sutil Rich Personal Lubricant is a non-sterile, water-based personal lubricant intended for over-the-counter use. It is applied topically to penile and/or vaginal tissues to provide lubrication during intimate sexual activity. The device is supplied in LDPE tubes with HDPE flip-top caps. It functions by providing a physical barrier and moisture to reduce friction. The healthcare provider or user evaluates the product based on its physical properties (viscosity, pH, osmolality) and compatibility with specific condom types. It benefits the patient by enhancing comfort during sexual activity. The device is not intended for use with polyurethane condoms.
Clinical Evidence
Bench testing only. Performance data included shelf-life testing (real-time aging), biocompatibility testing (cytotoxicity per ISO 10993-5, skin irritation/sensitization via HRIPT, acute systemic toxicity per ISO 10993-11), and condom compatibility testing per ASTM D7661-18. Results confirmed the device is non-cytotoxic, non-irritating, non-sensitizing, and compatible with natural rubber latex and polyisoprene condoms.
Indicated for penile and/or vaginal application to moisturize and lubricate, enhancing ease and comfort of intimate sexual activity and supplementing natural lubrication. Compatible with natural rubber latex and polyisoprene condoms; not compatible with polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 2, 2025
Hathor Professional Skincare Ltd.
% Rob Packard
President
Medical Device Academy, Inc.
345 Lincoln Hill Rd
Shrewsbury, Vermont 05738
Re: K250629
Trade/Device Name: Sutil Rich Personal Lubricant
Regulation Number: 21 CFR 884.5300
Regulation Name: Condom
Regulatory Class: II
Product Code: NUC
Dated: March 2, 2025
Received: March 3, 2025
Dear Rob Packard:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250629 - Rob Packard
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K250629 - Rob Packard
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250629 | |
| Device Name Sutil Rich Personal Lubricant | |
| Indications for Use (Describe) Sutil Rich Personal Lubricant is a personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. The product is not compatible with polyurethane condoms. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Hathor Professional Skincare Ltd.
K250629
Page 1 of 4
# 510(k) SUMMARY (K250629)
## I. SUBMITTER
Company Name: Hathor Professional Skincare Ltd.
Address: 2618 Roseberry Ave.
City, State, Zip: Victoria, BC V8R 3T7 Canada
Telephone: +1.778.679.0037
Contact Person: Carolyn (Mellta) Swift
Date Prepared: April 30, 2025
## II. DEVICE
Device Trade Name: Sutil Rich Personal Lubricant
Common Name: Personal Lubricant
Regulation Name: Condom
Regulation Number: 21 CFR 884.5300
Regulatory Class: II
Product Code: NUC (lubricant, personal)
## III. PREDICATE DEVICE
Predicate Manufacturer: Good Clean Love, Inc.
Predicate Trade Name: Desnuda Reflect
Predicate 510(k): K211998
The predicate device has not been subject to a design-related recall.
## IV. DEVICE DESCRIPTION
Sutil Rich Personal Lubricant is a non-sterile, water-based personal lubricant that is compatible with natural rubber latex and polyisoprene condoms. It is not compatible with polyurethane condoms. Its formulation consists of purified water, propanediol, hyaluronic acid, gluconolactone, sodium benzoate, Nelumbo Nucifera Root Water, Horny Goat (Epimedium Sagittatum) Herb, Siberian Ginseng (Eleuthero) Std. Extract, Jujube (Zizyphus Jujuba) Fruit Powder Extract. The device is provided in tube containers composed of low-density polyethylene (LDPE) with flip-top caps made of high-density polyethylene (HDPE). The packaging configurations range from 0.34 FL OZ (10 mL) to 8 FL OZ (240 mL), depending on the device. The personal lubricant is non-sterile and intended for over-the-counter (OTC) use.
The device specifications are listed in Table 1 below:
Table 1: Device Specifications for Sutil Rich Personal Lubricant
| Parameter | Test Method | Specification |
| --- | --- | --- |
| Appearance | Visual | Viscous liquid |
| Color | Visual | Clear |
| Odor | Organoleptic | No scent |
| pH | USP <791> | 5.8 – 6.2 |
510(k) Summary
{5}
Hathor Professional Skincare Ltd.
K250629
Page 2 of 4
| Viscosity (cps) | USP <911> | 30,000 – 40,000 cps |
| --- | --- | --- |
| Osmolality (mOsm/kg) | USP <785> | 550 - 700 (1:5 dilution) |
| Antimicrobial Effectiveness | USP <51> | Meets USP <51> criteria for category 2.
Bacteria: No less than 2.0 log reduction from initial count at 14 days, and no increase from the 14-day count at 28 days.
Yeast and Molds: No increase from the initial calculated count at 14 days and 28 days. |
| Total Microbial Count (TAMC) | USP <61> | < 100 cfu/g |
| Fungal/Yeast/Mold Limits (TYMC) | USP <61> | < 10 cfu/g |
| Absence of Pathogenic Organisms (Pseudomonas aeruginosa, Staphylococcus aureus, Salmonella, Escherichia coli, Candida albicans) | USP <62> | Absent |
## V. INDICATIONS FOR USE
Sutil Rich Personal Lubricant is a personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. The product is not compatible with polyurethane condoms.
## VI. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The table below compares the intended use and technological characteristics of the subject and predicate devices:
| | Subject Devices (K241443) | Predicate Device (K211998) |
| --- | --- | --- |
| Indications for Use | Sutil Rich Personal Lubricant is a personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene | Desnuda Reflect is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber |
510(k) Summary
{6}
K250629
Page 3 of 4
Hathor Professional Skincare Ltd.
| | condoms. The product is not compatible with polyurethane condoms. | latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. |
| --- | --- | --- |
| Rx/OTC | OTC | OTC |
| Water-based | Yes | Yes |
| Ingredients | Water, Propanediol (bio 1,3-propanediol), Hyaluronic Acid, D-Glucono-1,5-lactone, Sodium Benzoate, Nelumbo Nucifera Root Water, Horny Goat (Epimedium Sagittatum) Herb, Organic Siberian Ginseng (Eleuthero) Std. Extract, Jujube (Zizyphus Jujuba) Fruit Powder Extract | Water, Hyaluronic Acid, Hydroxyethylcellulose, Xanthan Gum, Zemea Propanediol, Sodium Benzoate, Potassium Sorbate, DL Lactic Acid |
| Sterile | No | No |
| Condom Compatibility | Natural Rubber Latex and Polyisoprene Condoms | Natural Rubber Latex and Polyisoprene Condoms |
| Shelf Life | 2 years | 1 year |
| Physical Characteristics Tested (appearance, color, odor) | Yes | Yes |
| pH Tested | Yes | Yes |
| Osmolality Tested | Yes | Yes |
| Biocompatibility Tested | Yes | Yes |
| Microbiology Tested (absence of pathogenic organisms, TAMC, and TYMC) | Yes | Yes |
| Antimicrobial Effectiveness Tested | Yes | Yes |
The subject device and predicate device have the same intended use (i.e., provides lubrication during intimate sexual activity). The subject device and predicate device have different technological characteristics, including variations in formulations, specifications, and shelf-life durations. The differences in the technological characteristics of the subject and predicate do not raise different questions of safety and effectiveness.
## VII. SUMMARY OF PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Shelf-life Testing
The subject device has a 2-year shelf-life based on the results of real-time aging. Testing on samples showed that the subject device met all device specifications listed in Table 1 above across the device shelf-life.
### Biocompatibility Testing
Biocompatibility testing was performed in accordance with the 2023 FDA guidance document “Use of
510(k) Summary
{7}
Hathor Professional Skincare Ltd.
K250629
Page 4 of 4
International Standard ISO 10993-1, “Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process”, as follows:
- Cytotoxicity per ISO 10993-5:2009
- Skin Irritation and Sensitization – Human Repeat Insult Patch Test (HRIPT), an alternative test method to ISO 10993-10:2021 and ISO 10993-23:2021
- Acute Systemic Toxicity per ISO 10993-11:2017
The results of this testing demonstrate that the subject lubricants are non-cytotoxic, non-irritating, non-sensitizing, and not systemically toxic.
## Condom Compatibility
The compatibility of the subject devices with condoms was evaluated in accordance with ASTM D7661-18, “Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms” and were determined to be compatible with natural rubber latex and polyisoprene condoms and not compatible with polyurethane condoms.
## VIII. CONCLUSIONS
The results of the performance testing described above demonstrate that Sutil Rich Personal Lubricant is as safe and effective as the predicate device and support a determination of substantial equivalence.
510(k) Summary
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.