The Medi Lift Essential Eye Mask is intended for facial stimulation and indicated for over-the-counter cosmetic use.
Device Story
Medi Lift Essential Eye Mask is a wearable silicone mask for facial muscle stimulation. Device covers upper cheek and periocular region outside orbital rim. Two controllers with electrodes deliver electrical pulses to facial muscles; powered by rechargeable 3.7V Lithium-ion batteries. Includes film heater for upper cheek relaxation (non-medical purpose); stimulation and heating modes are mutually exclusive. Operated by user via buttons on controllers. Conductive media (water or electroconductive gel) required. Device provides cosmetic facial stimulation; intended for home use. Output parameters (voltage, current, frequency) are controlled by internal circuitry. Benefits include facial muscle stimulation for cosmetic purposes.
Clinical Evidence
Bench testing only. Performance data includes electrical output verification, thermal testing of electrodes, software verification and validation, and biocompatibility testing (ISO 10993-5, -10, -23). Safety testing conducted per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, and IEC 62133-2.
Technological Characteristics
Silicone rubber mask with two controllers; 3.7V Lithium-ion battery power. Symmetric pulsed quadphasic waveform; 5-100 Hz frequency range. Film heater with thermal sensor (limit <40 °C). Type BF applied part. Connectivity: USB charging. Sterilization: Not specified. Software: Embedded firmware for stimulation and heating control.
Indications for Use
Indicated for over-the-counter cosmetic facial stimulation in adults.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 29, 2025
YA-MAN Ltd.
% Lina Kontos
Partner
Hogan Lovells US LLP
555 Thirteenth Street NW
Washington, District of Columbia 20004
Re: K250618
Trade/Device Name: Medi Lift Essential Eye Mask
Regulation Number: 21 CFR 882.5890
Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief
Regulatory Class: Class II
Product Code: NFO
Dated: September 30, 2025
Received: September 30, 2025
Dear Lina Kontos:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}
K250618 - Lina Kontos
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250618 - Lina Kontos
Page 3
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250618
Device Name
Medi Lift Essential Eye Mask
Indications for Use (Describe)
The Medi Lift Essential Eye Mask is intended for facial stimulation and indicated for over-the-counter cosmetic use.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(k) Summary
YA-MAN's Medi Lift Essential Eye Mask
K250618
## Submitter
YA-MAN Ltd.
2-4-2 Toyo, Koto-ku,
Tokyo, 1350016 Japan
Phone: +81-3-5665-7321
Facsimile: +81-5665-7370
Contact Person: Jun Takada
Date Prepared: October 27, 2025
Name of Device: Medi Lift Essential Eye Mask
Common Name: Stimulator, transcutaneous electrical, aesthetic purposes
Regulatory Classification: Class II, 21 CFR 882.5890
Product Code: NFO (Stimulator, Transcutaneous Electrical, Aesthetic Purposes)
## Predicate Devices
Primary Predicate Device
510(k) Number: K220198
Trade Name: Medi Lift PLUS
Manufacturer: YA-MAN Ltd.
Product Code: NFO
Secondary Predicate Device
510(k) Number: K120511
Trade Name: Ageless Wonder
Manufacturer: Leto Enterprises Incorporation
Product Code: NFO
## Device Description
The Medi Lift Essential Eye Mask ("Medi Lift Eye") is composed of a mask made of silicone rubber which is worn on the upper part of the user's face, and covers the upper cheek and the periocular region outside the orbital rim. The mask contains two controllers with electrodes, which are attached to the mask. These controllers and electrodes deliver electrical pulses to stimulate facial muscles. The controllers are operated independently by pressing buttons on each controller. The mask also contains a heater to warm the upper cheek area for user relaxation, not for a medical purpose. The stimulation and heating functions are provided in separate modes and cannot be
{5}
used simultaneously. The controllers attached to the mask contain two charging pins which allow for the built-in Lithium-ion battery that powers the device to be charged using a USB charging cable that is provided as part of the device. The device is not operated during charging.
# Intended Use/Indications for Use
The Medi Lift Essential Eye Mask is intended for facial stimulation and indicated for over-the-counter cosmetic use.
# Summary of Technological Characteristics
The Medi Lift Eye device achieves its intended use of facial stimulation by energy delivered through electrodes placed on the user's face, exactly in the same manner as the predicate devices, Medi Lift Plus (K220198) and Ageless Wonder (K120511). The subject device is intended to stimulate the upper cheek and the periocular region outside the orbital rim, which is the same as the secondary predicate device, whereas the primary predicate device is intended to stimulate the cheeks only. Both Medi Lift Eye and the primary predicate device have a silicone mask as the set form factor, which standardizes the electrodes' placement and procedures for use. The secondary predicate device's eye-specific applicator is provided as a single-size electrode pad that is shaped to cover the intended treatment area. Both the primary predicate and subject device are powered by one rechargeable 3.7 V Lithium ion battery per signal generator, while the secondary predicate device is powered by non-rechargeable batteries.
The subject device is used with water or with an electroconductive gel provided with the device (Spectra 360 manufactured by Parker Laboratories, Inc.), depending on the user's preference, while both predicate devices use water as the conductive media.
The Medi Lift Eye device also includes a heater to warm the upper cheek, while neither predicate device has a heater. This feature is provided for user relaxation only, not for a medical purpose. Therefore, the device's therapeutic function (facial stimulation) is the same as the predicate devices.
Based on above, the differences between Medi Lift Eye device and the predicate devices do not raise new or different questions of safety or effectiveness. The table below provides a comparison of the technological characteristics for the subject device and the predicate device.
Table 1: Comparison of Technological Characteristics
| | Subject Device: Medi Lift Eye (K250618) | Primary Predicate Device: Medi Lift Plus (K220198) | Secondary Predicate Device: Ageless Wonder (K120511) | Comparison between the Medi Lift Eye and predicate devices |
| --- | --- | --- | --- | --- |
| Product code | NFO | NFO | NFO | Same |
| Regulation number | 21 CFR 882.5890 | 21 CFR 882.5890 | 21 CFR 882.5890 | Same |
| OTC or prescription | OTC | OTC | OTC | Same |
{6}
| | Subject Device: Medi Lift Eye (K250618) | Primary Predicate Device: Medi Lift Plus (K220198) | Secondary Predicate Device: Ageless Wonder (K120511) | Comparison between the Medi Lift Eye and predicate devices |
| --- | --- | --- | --- | --- |
| Indications for use | The Medi Lift Eye is intended for facial stimulation and indicated for over-the-counter cosmetic use. | The Medi Lift PLUS is intended for facial stimulation and indicated for over-the-counter cosmetic use. | Ageless Wonder Facial Toning Device is intended for facial stimulation and indicated for over-the-counter cosmetic use. | Same |
| Intended anatomical position in the face | The upper cheek and the periocular region outside the orbital rim | Cheeks | (Under eye mode) The upper cheek and the outer side of the eye; handheld and placed by users | Same as the secondary predicate device |
| Warming feature | Yes, for user comfort in a separate mode from EMS treatment. The upper cheek is warmed by a film heater inside the electrode. Includes a thermal sensor to keep the electrode surface <40 °C. | No | No | Different, but does not raise different questions because the heater mode is not for medical purposes. |
| Wearable or handheld | Wearable | Wearable | Wearable and handheld (under eye mode uses the handheld unit, which is held in place during treatment) | Same as the primary predicate, similar to secondary predicate. Device is stationary during treatment. |
| Conductive media | Conductive gel or Water | Water | Water | Similar, subject device has gel as an additional option for users. |
| Number of controllers (signal generator) | 2 (right and left) | 2 (right and left) | 1 | Same as primary predicate, similar to secondary predicate (applied to one side of the face at a time). |
| Number of electrodes per controller | 2 | 3 | 2 (under eye mode) | Same as secondary predicate; difference from primary predicate caused by |
{7}
| | Subject Device: Medi Lift Eye (K250618) | Primary Predicate Device: Medi Lift Plus (K220198) | Secondary Predicate Device: Ageless Wonder (K120511) | Comparison between the Medi Lift Eye and predicate devices |
| --- | --- | --- | --- | --- |
| | | | | different applied anatomic area. |
| Power source | One 3.7 V, 350 mAh Lithium ion battery per signal generator | One 3.7 V, 60 mAh Lithium ion battery per signal generator | Battery powered (two 1.5 V AAA batteries) | Similar, all are battery powered. |
| Patient leakage current | Protection method: Type BF applied part | Protection method: Type BF applied part | Protection method: Type BF applied part | Same |
| Number of output channels | 1 per controller | 2 per controller | 1 per controller | Same as secondary predicate, similar to primary predicate |
| Indicator display | LED indicator lights for on/off status of the device and the heater, battery status during charging | LED indicator lights for on/off status, battery status during charging | LCD display for on/off status, low battery, voltage/current level | Similar, all indicate on/off status, battery status, and mode. |
| Timer range (minutes) | Fixed 10 minutes. The user cannot change the default setting. | Fixed 10 minutes. The user cannot change the default setting. | Options for 5, 10, 15, or 20 minutes | Same as the primary predicate |
| Weight | Controller: 32 g each Silicone mask: 109 g Total: 173 g | Controller: 28 g each Silicone mask: 120 g Total: 176 g | 90 g total (including a headset) | Similar to the primary predicate |
| Dimensions [W x H x D] | Controller: 78 x 54 x 150 mm Mask: 620 x 230 mm | Controller: 90 x 55 x 20 mm Mask: 615 x 170 mm | 98.5 x 53 x 27.5 mm | Similar to the primary predicate |
| EMS Specifications: | | | | |
| Waveform | Symmetric Pulsed Quadphasic | Symmetric Pulsed Biphasic (1 kHz) and Quadphasic (under 1 kHz) | Symmetric Biphasic | Same as the primary predicate device for the applicable frequency range |
| Waveform shape | Sine pulses | Rectangular pulses | Sine pulses | Same as the secondary predicate |
| Maximum output voltage (at 500Ω) | 45.6 Vpp ± 20% | 39.3 Vpp ± 20% | Under eye mode: 4.08 - 29.6 V | Similar to primary predicate, within the range of the secondary predicate |
{8}
| | Subject Device: Medi Lift Eye (K250618) | Primary Predicate Device: Medi Lift Plus (K220198) | Secondary Predicate Device: Ageless Wonder (K120511) | Comparison between the Medi Lift Eye and predicate devices |
| --- | --- | --- | --- | --- |
| Maximum output current (at 500Ω) | 1.51 mA ± 20% | 3.22 mA ± 20% | 0 - 43.2 μA (0 to 1000Ω) | Similar, differences do not raise different questions |
| Pulse width | 208 μs ± 10% (<100Hz), 288 μs ± 10% (100Hz) | 152/168/192 μs ± 10% (2.5-100Hz) 52 μs ± 10% (1kHz) | 150 μs | Similar, differences do not raise different questions |
| Frequency | 5 Hz, 6 Hz, 7 Hz, 8 Hz, 9 Hz, 10 Hz, 20 Hz, 25 Hz, 33 Hz, 50 Hz, 100 Hz | 2.5 Hz, 5 Hz, 6 Hz, 7 Hz, 8 Hz, 9 Hz, 10 Hz, 20 Hz, 25 Hz, 33 Hz, 50 Hz, 66 Hz, 100 Hz, 1 kHz | 20 Hz | Similar, within the range of the primary and secondary predicate devices |
| For interferential modes only: - Beat frequency | NA | NA | Not publicly disclosed | Same |
| For multiphasic waveforms only: - Symmetrical phases? | Yes | Yes | Yes | Same |
| - Phase duration (state range, if applicable) (both phases, if asymmetrical) | 52 μs ±10% (<100 Hz) 72 μs ±10% (100 Hz) | 38/42/48 μs ±10% (2.5-100 Hz) 26 μs ±10% (1 kHz) | 75 μs | Similar to primary and secondary predicates |
| Maximum phase charge (at 500Ω) | 0.109 μC | 0.084 μC | Not publicly disclosed | Similar to primary predicate |
| Maximum current density (at 500Ω) | 0.70 mA/cm² = 1.51 mA / 2.16 cm² (area of smaller electrode) | 0.64 mA/cm² = 3.22 mA / 5.04 cm² (area of an electrode; all are the same size) | Not publicly disclosed | Similar to primary predicate |
| Maximum power density (at 500Ω) | 0.53 mW/cm² | 1.03 mW/cm² | Not publicly disclosed | Similar to primary predicate |
{9}
| | Subject Device: Medi Lift Eye (K250618) | Primary Predicate Device: Medi Lift Plus (K220198) | Secondary Predicate Device: Ageless Wonder (K120511) | Comparison between the Medi Lift Eye and predicate devices |
| --- | --- | --- | --- | --- |
| Burst mode (i.e., pulse trains) | None | None | Not publicly disclosed | Same |
| a. Pulses per burst | NA | NA | Not publicly disclosed | Same |
| b. Bursts per second | NA | NA | Not publicly disclosed | Same |
| c. Burst duration (seconds) | NA | NA | Not publicly disclosed | Same |
| d. Duty cycle [Line (b) x Line (c)] | NA | NA | Not publicly disclosed | Same |
| ON time (seconds) | 0.05 to 4 seconds (Depending on the pattern of output and the intensity level) | 0.2 to 16 seconds (Depending on the pattern of output and the intensity level) | Under eye mode: 3.174 seconds | Similar to primary predicate |
| OFF time (seconds) | 0.35 to 2 seconds (Depending on the pattern of output) | 1 to 4 seconds (Depending on the pattern of output) | Under eye mode: 1.148 seconds | Similar to primary predicate |
## Performance Data
The subject device was assessed in accordance with the following standards and performance tests:
- IEC 60601-1-2: Ed 4.1 2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
- ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021], Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD).
- IEC 62133-2 Edition 1.0 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems.
- IEC 60601-1-11: Ed. 2.1 2020, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
{10}
- IEC 60601-2-10:2012, AMD1:2016, Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators.
- ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.
- ISO 10993-10:2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization.
- ISO 10993-23:2021, Biological evaluation of medical devices - Part 23: Tests for irritation.
- Electrical output of the Medi Lift Essential Eye Mask
- Thermal testing of the electrodes
- Software verification and validation
## Conclusions
In conclusion, the Medi Lift Eye is substantially equivalent to the identified predicate devices, Medi Lift PLUS (K220198) and Ageless Wonder (K120511), because they have identical indications for use and similar technological characteristics. The minor differences in technological characteristics and procedures for use do not raise different questions of safety or effectiveness, because they do not alter the fundamental technology, purpose or method of use of the Medi Lift Eye as compared to its predicates. In addition, testing of the Medi Lift Eye with acceptable scientific methods evaluated the effects of the minor technological differences and demonstrated that the subject device is as safe and effective as the predicates.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.