Arthrex FiberTak Suture Anchor

K250596 · Arthrex, Inc. · MBI · Mar 28, 2025 · Orthopedic

Device Facts

Record IDK250596
Device NameArthrex FiberTak Suture Anchor
ApplicantArthrex, Inc.
Product CodeMBI · Orthopedic
Decision DateMar 28, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Arthrex FiberTak Suture Anchor is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures: Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Hand/Wrist: Scapholunate Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers. Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty) Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Joint Capsule Closure Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair

Device Story

All-suture knotless anchor for soft tissue-to-bone fixation; used in orthopedic procedures (shoulder, knee, hip, etc.). Device consists of hollow braided polyester sheath with preloaded UHMWPE/nylon repair and passing sutures. Provided sterile, single-use, on disposable inserter. Surgeon deploys anchor into bone; suture secures soft tissue. Benefits include secure fixation for various orthopedic repairs. No electronic or software components.

Clinical Evidence

Bench testing only. Cyclic pull-out testing demonstrated performance statistically equivalent to the predicate. Bacterial endotoxin testing per EP 2.6.14/USP <85> confirmed compliance with pyrogen limit specifications.

Technological Characteristics

All-suture knotless anchor. Materials: polyester yarn sheath, UHMWPE/nylon braided sutures. Form factor: preloaded on disposable inserter. Sterilization: Ethylene Oxide. No energy source or software.

Indications for Use

Indicated for fixation of soft tissue to bone in shoulder, foot/ankle, knee, hand/wrist, elbow, and hip procedures including tendon reattachments, ligament reconstructions, and joint repairs.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. The logo is simple and professional, and it is easily recognizable. March 28, 2025 Arthrex Inc. Ruth Segall Regulatory Affairs Specialist II 1370 Creekside Blvd Naples, Florida 34108 Re: K250596 Trade/Device Name: Arthrex FiberTak Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: February 27, 2025 Received: February 28, 2025 Dear Ruth Segall: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Thomas Mcnamara -S For: Christopher Ferreira, MS Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use Submission Number (if known) K250596 Device Name Arthrex FiberTak Suture Anchor #### Indications for Use (Describe) The Arthrex FiberTak Suture Anchor is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures: - Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction - · Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction · Hand/Wrist: Scapholunate Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers. Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty) · Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction • Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Joint Capsule Closure · Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Arthrex. The logo consists of the word "Arthrex" in a sans-serif font, with the "x" stylized to resemble a surgical instrument. The logo is simple and modern, and it is likely used to promote the company's products and services. ## 510(k) Summary | Date Prepared | 02/27/2025 | |------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter | Arthrex Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945 | | Contact Person | Name: Ruth Segall<br>Title: Regulatory Affairs Specialist II<br>Phone: 239-598-4302 x 71764<br>Email: Ruth.Segall@arthrex.com | | Trade Name | Arthrex FiberTak Suture Anchor | | Common Name | Suture Anchor | | Product Code | MBI | | Classification Name | 21 CFR 888.3040: Smooth or threaded metallic bone<br>fixation fastener | | Regulatory Class | Class II | | Primary Predicate Device | K221396: Arthrex FiberTak Suture Anchor | | Additional Predicate Devices | N/A | | Reference Devices | K230433: Arthrex Double Loaded Knotless FiberTak®<br>Suture Anchor<br>K193157: Arthrex 3.9 Corkscrew Suture Anchor | | Purpose of Submission | This Special 510(k) premarket notification is submitted<br>to obtain clearance for the subject Arthrex FiberTak<br>Suture Anchor as a line extension to the Arthrex<br>FiberTak Suture Anchor cleared via K221396. | | Device Description | The Arthrex FiberTak Suture Anchor is an all-suture<br>knotless anchor intended to be used for fixation of<br>suture (soft tissue) to bone in the shoulder, foot/ankle,<br>knee, hand/wrist, elbow, and hip.<br><br>The anchor is constructed from a hollow braid sheath<br>made of polyester yarns with Arthrex suture<br>components assembled to the sheath. The suture<br>components include a repair suture and a passing<br>suture which are braided polyester sutures made of<br>ultra-high molecular weight polyethylene (UHMWPE) | | | where the suture may also include nylon yarns. The<br>anchor is preloaded on a disposable inserter and<br>is provided sterile (Ethylene Oxide), single-use, and is<br>packaged in a 5-pack. | | Indications for Use | The Arthrex FiberTak Suture Anchor is intended for<br>fixation of suture (soft tissue) to bone in the shoulder,<br>foot/ankle, knee, hand/wrist, elbow, and hip in the<br>following procedures:<br>• Elbow: Biceps Tendon Reattachment, Ulnar or Radial<br>Collateral Ligament Reconstruction<br>• Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP<br>Lesion Repair, Biceps Tenodesis, Acromio-Clavicular<br>Separation Repair, Deltoid Repair, Capsular Shift or<br>Capsulolabral Reconstruction<br>• Hand/Wrist: Scapholunate Ligament Reconstruction,<br>Repair/Reconstruction of collateral ligaments, Repair of<br>Flexor and Extensor Tendons at the PIP, DIP and MCP<br>joints for all digits, digital tendon transfers, Carpal<br>Ligament Reconstruction and Carpometacarpal joint<br>arthroplasty (basal thumb joint arthroplasty)<br>• Foot/Ankle: Lateral Stabilization, Medial Stabilization,<br>Achilles Tendon Repair, Metatarsal Ligament Repair,<br>Hallux Valgus reconstruction, digital tendon transfers,<br>Mid-foot reconstruction<br>• Knee: Medial Collateral Ligament Repair, Lateral<br>Collateral Ligament Repair, Patellar Tendon Repair,<br>Posterior Oblique Ligament Repair, Iliotibial Band<br>Tenodesis, Joint Capsule Closure<br>• Hip: Capsular Repair, Acetabular Labral Repair, Gluteal<br>Tendon Repair | | Performance Data | Cyclic pull-out testing was conducted on the subject<br>Arthrex FiberTak Suture Anchor submitted in this<br>Special 510(k). The test data demonstrates that the<br>subject device performs statistically equivalent to the<br>primary predicate device.<br>Bacterial endotoxin per EP 2.6.14/USP <85> was<br>conducted to demonstrate that the device meets<br>pyrogen limit specifications. | | Technological Comparison | The subject Arthrex FiberTak Suture Anchor is a line<br>extension to the primary predicate device K221396. The<br>subject and primary predicate devices have identical<br>basic design, indications for use, intended use,<br>packaging, shelf-life, biocompatibility profile, overall<br>manufacturing, and sterilization processes. In<br>comparison to the predicate device, the subject device<br>contains a modification to the pre-loaded sutures. Any<br>differences between the subject and predicate devices<br>are considered minor and do not raise any questions<br>concerning safety or effectiveness. | | Conclusion | The Arthrex FiberTak Suture Anchor is a line extension<br>and substantially equivalent to the predicate device<br>(K221396) in which the basic design features,<br>indications for use and intended use are identical. Any<br>differences between the subject device and the primary<br>predicate devices are considered minor and do not<br>result in new or different questions concerning safety or<br>effectiveness. Based on the indications for use,<br>technological characteristics, and the summary of data<br>submitted, Arthrex has determined that the subject<br>device is substantially equivalent to the currently<br>marketed primary predicate device. | {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Arthrex. The logo consists of the word "Arthrex" in a bold, sans-serif font. To the right of the text is an image of a surgical instrument, possibly a type of clamp or forceps. The instrument is stylized and simplified, with a handle and a grasping end. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for Arthrex. The logo consists of the word "Arthrex" in a bold, sans-serif font. To the right of the word is an image of a surgical tool, possibly a type of clamp or scissor. The logo is simple and professional, and it is likely used to represent the company's brand.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...