K250588 · Beckman Coulter, Inc. · LFX · Nov 17, 2025 · Microbiology
Device Facts
Record ID
K250588
Device Name
Access Rubella IgG
Applicant
Beckman Coulter, Inc.
Product Code
LFX · Microbiology
Decision Date
Nov 17, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3510
Device Class
Class 2
Indications for Use
The Access Rubella IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to the rubella virus in human serum using the Access Immunoassay Systems. The Access Rubella IgG assay aids in the diagnosis of rubella infection and the determination of immunity.
Device Story
Access Rubella IgG is a paramagnetic-particle, chemiluminescent immunoassay; detects IgG antibodies to rubella virus in human serum. Operates on DxI 9000 Access Immunoassay Analyzer; utilizes two-step immunoenzymatic assay principle. Capture reagent: paramagnetic particles coated with rubella (strain HPV 77) antigen; detection antibody: mouse monoclonal anti-human IgG conjugated to alkaline phosphatase. Instrument automates sample processing, incubation, and chemiluminescent signal measurement. Output: qualitative/quantitative results (IU/mL) interpreted as Non-Reactive, Equivocal, or Reactive. Used in clinical laboratories by trained personnel. Results assist clinicians in diagnosing rubella infection and assessing patient immune status.
Clinical Evidence
Bench testing only. Precision studies (within-laboratory and between-instrument) were conducted over 20 and 5 days respectively, showing acceptable variability. Linearity was confirmed across the 5.0–500.0 IU/mL range. A method comparison study (n=162) between the DxI 9000 and the predicate Access 2 system demonstrated a PPA of 97.47% and NPA of 96.55%. No new clinical studies were performed; clinical agreement relies on the predicate device.
Indicated for qualitative and quantitative determination of IgG antibodies to rubella virus in human serum to aid in diagnosis of rubella infection and determination of immunity.
Regulatory Classification
Identification
Rubella virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to rubella virus in serum. The identification aids in the diagnosis of rubella (German measles) or confirmation of a person's immune status from past infections or immunizations and provides epidemiological information on German measles. Newborns infected in the uterus with rubella virus may be born with multiple congenital defects (rubella syndrome).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) National Committee for Clinical Laboratory Standards':
(i) 1/LA6 “Detection and Quantitation of Rubella IgG Antibody: Evaluation and Performance Criteria for Multiple Component Test Products, Speciment Handling, and Use of the Test Products in the Clinical Laboratory, October 1997,”
(ii) 1/LA18 “Specifications for Immunological Testing for Infectious Diseases, December 1994,”
(iii) D13 “Agglutination Characteristics, Methodology, Limitations, and Clinical Validation, October 1993,”
(iv) EP5 “Evaluation of Precision Performance of Clinical Chemistry Devices, February 1999,” and
(v) EP10 “Preliminary Evaluation of the Linearity of Quantitive Clinical Laboratory Methods, May 1998,”
(2) Centers for Disease Control's:
(i) Low Titer Rubella Standard,
(ii) Reference Panel of Well Characterized Rubella Sera, and
(3) World Health Organization's International Rubella Standard.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K250588
B Applicant
Beckman Coulter Inc.
C Proprietary and Established Names
Access Rubella IgG
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LFX | Class II | 21 CFR 866.3510 - Rubella Virus Serological Reagents | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
Modification of a previously cleared device, Access Rubella IgG assay to add the DxI 9000 Access Immunoassay Analyzer.
B Measurand:
IgG antibodies to Rubella virus.
C Type of Test:
Chemiluminescent Immunoassay.
## III Intended Use/Indications for Use:
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K250588 - Page 2 of 9
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The Access Rubella IgG assay is a paramagnetic particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to the rubella virus in human serum using the Access Immunoassay Systems. The Access Rubella IgG assay aids in the diagnosis of rubella infection and the determination of immunity.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
DxI 9000 Access Immunoassay Analyzer.
IV Device/System Characteristics:
A Device Description:
The Access Rubella IgG assay is a paramagnetic-particle chemiluminescent immunoassay for the qualitative and quantitative detection of Rubella virus-specific IgG antibody in adult human serum using the Access Immunoassay Systems.
The Access Rubella IgG assay consists of the reagent pack, calibrators, and quality controls (QCs) packaged separately. Other items needed to run the assay but not supplied with reagent kit include substrate and wash buffer.
B Principle of Operation:
The Access Rubella IgG assay is an enzyme immunoassay using an indirect technique. A sample is added to a reaction vessel with paramagnetic particles coated with Rubella viral membrane antigen. Specific antibodies present in the sample bind to the antigen. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Alkaline phosphatase-conjugated monoclonal anti-human IgG antibody is then added and attaches to the IgG antibodies captured on the particles. A second separation and wash step remove unbound conjugate. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of IgG antibodies to the Rubella virus in the sample. The amount of analyte in the sample is determined from a stored, multi-point calibration curve standardized against the Second International Standard material (2nd ISP). Test results are determined automatically by the system software.
The Access Rubella IgG assay results interpretation is as follows:
{2}
| Results | Result Interpretation | Comments |
| --- | --- | --- |
| < 10 IU/mL | Non-reactive | Samples are considered non-reactive for the presence of Rubella IgG antibodies. A non-reactive result may indicate that an individual has not been infected with Rubella or vaccinated and is thus susceptible to infection. A non-reactive result, however, does not rule out an acute infection. If exposure to the pathogen is suspected despite an initial non-reactive result, a second sample should be collected and tested at least two to three weeks later. |
| ≥ 10 IU/mL to < 15 IU/mL | Equivocal | Samples are considered equivocal for the presence of Rubella IgG antibodies. Specimens that are equivocal may contain low levels of anti-Rubella IgG. As any cutoff is a statistically determined value, it is recommended that results close to the cutoff are interpreted cautiously and further investigated. If exposure to the pathogen is suspected despite an initial equivocal result, a second sample should be collected and tested at least two to three weeks later. |
| ≥ 15 IU/mL | Reactive | Samples are considered reactive for the presence of Rubella IgG antibodies and indicate acute or past infection or vaccination. |
C Instrument Description Information:
1. Instrument Name:
DxI 9000 Access Immunoassay Analyzer.
2. Specimen Identification:
Human serum.
3. Specimen Sampling and Handling:
The instrument performs all handling procedures automatically.
4. Calibration:
Calibration is required every 28 days. The calibrators are formulated with equine serum and human defibrinated plasma. There is one negative and five positive calibrators containing approximately 10, 25, 50, 200, and 500 IU/mL of human anti-Rubella IgG. They are traceable to the Second International Standard Preparation for Anti-Rubella Serum (2nd ISP).
5. Quality Control:
Access Rubella IgG Controls are required to run the assay and are available separately. One negative and one positive (target mean of 22 – 43 IU/mL) controls formulated in human defibrinated plasma are traceable to the Second International Standard Preparation for Anti-Rubella Serum (2nd ISP).
K250588 - Page 3 of 9
{3}
# V Substantial Equivalence Information:
A Predicate Device Name(s):
Access Rubella IgG
B Predicate 510(k) Number(s):
K954687
C Comparison with Predicate(s):
| Device & Predicate Device(s): | Candidate K250588 | Predicate K954687 |
| --- | --- | --- |
| Device Trade Name | Same | Access Rubella IgG |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Same | The Access Rubella IgG assay is a paramagnetic particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to the rubella virus in human serum using the Access Immunoassay Systems. The Access Rubella IgG assay aids in the diagnosis of rubella infection and the determination of immunity. |
| Analyte | Same | IgG antibody to rubella virus |
| Technology | Same | Two-step immunoenzymatic assay |
| Format | Same | Chemiluminescent |
| Method | Same | Automated |
| Calibration | Same | Utilizes a stored calibration curve |
| Calibration Frequency | Same | 28 days |
| Sample Type | Same | Serum |
| Results Interpretation | Same | <10.0 IU/mL: Non-Reactive ≥10 - <15 IU/mL: Equivocal ≥15.0 IU/mL: Reactive |
| Capture Reagent | Same | Paramagnetic particles coated with rubella (strain HPV 77) sucrose gradient purified antigen |
| Detection Antibody | Same | Mouse monoclonal anti-human IgG antibody (clone 125 A 15) - alkaline phosphatase (bovine) conjugate |
K250588 - Page 4 of 9
{4}
| Device & Predicate Device(s): | Candidate K250588 | Predicate K954687 |
| --- | --- | --- |
| Stability | Same | 28 days after opening, 2 - 10°C |
| General Device Characteristic Differences | | |
| Analytical Measuring Interval | 5.0 – 500 IU/mL | 10.0 – 500 IU/mL |
| Substrate | Lumi-Phos PRO substrate | Access Substrate |
| Instrument | DxI 9000 Access Immunoassay Analyzer | Access 2 Immunoassay System |
# VI Standards/Guidance Documents Referenced:
- CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition (Reaffirmed: September 2019).
- CLSI EP06 Evaluation of the Linearity of Quantitative Measurement Procedures - 2nd Edition (2020).
- CLSI EP09c-Ed 3 Measurement Procedure Comparison and Bias Estimation Using Patient Samples (2018).
- CLSI EP12-Ed3 Evaluation of Qualitative, Binary Output Examination Performance, Third Edition (2023).
- CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - $2^{\text{nd}}$ Edition (2012).
# VII Performance Characteristics (if/when applicable):
# A Analytical Performance:
# 1. Precision/Reproducibility:
a) Within-Laboratory Precision: A 20-day within-laboratory precision study was conducted using 3 DxI 9000 Access Immunoassay Analyzer systems. Six human serum samples were tested in 2 replicates at 2 runs per day over 20 days using 3 reagent lot/calibrator lot combinations, where a unique reagent lot and calibrator lot are paired. The data were analyzed for repeatability (within-run), between-run, between-day, and within-laboratory precision. Within-laboratory precision data summary is shown in Table 1.
Table 1. Access Rubella IgG 20-day Within-Laboratory Precision
K250588 - Page 5 of 9
{5}
| Sample | n | Mean (IU/mL) | Repeatability (Within-Run) | | Between-Run | | Between-Day | | Between-Instrument/Reagent Lot/Calibrator Lota | | Overall Precisionb | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Sample 1 | 240 | 10.6 | 0.38 | 3.6 | 0.29 | 2.7 | 0.47 | 4.4 | 0.33 | 3.2 | 0.75 | 7.0 |
| Sample 2 | 240 | 19.2 | 0.78 | 4.0 | 0.69 | 3.6 | 0.91 | 4.7 | 0.19 | 1.0 | 1.39 | 7.3 |
| Sample 3 | 240 | 59.9 | 3.09 | 5.2 | 2.43 | 4.0 | 1.99 | 3.3 | 1.10 | 1.8 | 4.54 | 7.6 |
| Sample 4 | 240 | 120.7 | 4.69 | 3.9 | 3.89 | 3.2 | 6.69 | 5.5 | 2.82 | 2.3 | 9.48 | 7.9 |
| Sample 5 | 240 | 164.5 | 5.99 | 3.6 | 6.96 | 4.2 | 0.00 | 0.0 | 1.35 | 0.8 | 9.29 | 5.6 |
| Sample 6 | 240 | 379.4 | 14.36 | 3.8 | 8.75 | 2.3 | 19.23 | 5.1 | 9.75 | 2.6 | 27.35 | 7.2 |
a Access Rubella IgG reagent lot, Access Rubella IgG calibrator lot, and DxI 9000 instrument are confounded, and the confounding effect is represented by Between-Instrument/Reagent Lot/Calibrator Lot.
b Overall within-laboratory variability includes within-run, between-run, between-day, and between-instrument/lot variance components.
b) Reproducibility (between-Instrument Precision): An additional precision study was conducted over 5-day by testing six serum samples on 3 DxI 9000 Access Immunoassay Analyzers in an internal site. The samples were tested with 3 lots of Access Rubella IgG reagents, and 1 lot of Access Rubella IgG calibrator on each instrument with 5 replicates per run and 1 run per day over 5 days. Between-instrument reproducibility data is shown in Table 2.
Table 2. Access Rubella IgG Assay Reproducibility (Between-Instrument Precision)
| Sample | n | Mean (IU/mL) | Repeatability (Within-Run) | | Between-Day/Run\(^a\) | | Between-Instrument | | Between-Lot | | Reproducibility\(^b\) | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Sample 1 | 225 | 10.5 | 0.40 | 3.8 | 0.55 | 5.2 | 0.04 | 0.3 | 0.31 | 3.0 | 0.75 | 7.1 |
| Sample 2 | 225 | 18.4 | 0.82 | 4.4 | 0.76 | 4.1 | 0.43 | 2.3 | 0.14 | 0.8 | 1.20 | 6.5 |
| Sample 3 | 225 | 58.3 | 2.85 | 4.9 | 3.82 | 6.6 | 2.16 | 3.7 | 3.70 | 6.3 | 6.41 | 11.0 |
| Sample 4 | 225 | 119.7 | 5.18 | 4.3 | 7.64 | 6.4 | 0.00 | 0.0 | 3.55 | 3.0 | 9.90 | 8.3 |
| Sample 5 | 225 | 162.8 | 6.44 | 4.0 | 9.74 | 6.0 | 4.26 | 2.6 | 5.53 | 3.4 | 13.61 | 8.4 |
| Sample 6 | 225 | 370.9 | 13.38 | 3.6 | 18.13 | 4.9 | 5.76 | 1.6 | 7.21 | 1.9 | 24.35 | 6.6 |
a Days and runs are confounded.
b Reproducibility includes within-run, between-run, between-day, and between-lot variance components.
c) Reproducibility (between-site precision): Refer to K954687.
2. Linearity:
Linearity of the Access Rubella IgG on the DxI 9000 Access Immunoassay Analyzer was evaluated following CLSI Guideline EP06, 2nd Edition "Evaluation of Linearity of Quantitative Measurement Procedures". A tiled approach was used to evaluate the linearity of six segments of the measuring range. Six sample panels were generated by diluting unique pairs of high and low Rubella IgG antibody positive serum samples in various ratios. Each panel consisted of 7 - 9 dilutions in addition to high and low concentration samples. The
K250588 - Page 6 of 9
{6}
table below indicates the number of dilutions and concentration ranges covered by each sample panel.
| Panel | Concentration Range (IU/mL) | Number of Samples per Panel |
| --- | --- | --- |
| 1 | 3.87 – 140.97 | 9 |
| 2 | 9.42 – 110.56 | 11 |
| 3 | 92.92 – 200.53 | 9 |
| 4 | 188.02 – 340.47 | 9 |
| 5 | 332.23 – 422.55 | 9 |
| 6 | 405.11 – 548.77 | 9 |
This study was run on two DxI 9000 Access Immunoassay Analyzers, using one reagent lot and one calibrator lot. Low sample was run in replicates of eight, and all other samples were run in replicates of four or six. Two quality controls were run in replicates of two. The linearity of the Access Rubella IgG assay was considered acceptable if the data set was determined to be linear and if the deviation from linearity within or equal to $\pm 10\%$ for samples with concentrations $\geq 5.0~\mathrm{IU / mL}$ . This assay is linear across the analytical measuring interval of 5.0 to $500.0~\mathrm{IU / mL}$ .

Regression equation: $y = 0.9825x - 0.0819$
3. Analytical Specificity/Interference:
Refer to K954687.
4. Assay Reportable Range:
5.0 IU/mL - 500.0 IU/mL. See linearity above.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
K250588 - Page 7 of 9
{7}
Refer to K954687. The Access Rubella IgG calibrators were standardized against the 2nd International Standard Preparation for Anti-Rubella Serum 2nd ISP.
## 6. Detection Limit:
The limit of blank (LoB), limit of detection (LoD) and Lower limit of quantitation (LLoQ) for the Access Rubella IgG assay on DxI 9000 Immunoassay Analyzer were established following CLSI guideline EP17-A2 (Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition). In order to obtain at least 60 replicates of each sample level, the LoB was established with 5 zero-analyte serum samples tested in replicates of 5 over 3 days; the LoD with 6 low-analyte serum samples tested in replicates of 9 over 5 days; and the LLoQ with 13 low-analyte serum samples tested in replicates of 9 over 5 days. Detection Limit for the Access Rubella IgG assay on the DxI 9000 Access Immunoassay Analyzer are:
LoB = 0.1 IU/mL.
LoD = 0.2 IU/mL.
LLoQ = 5.0 IU/mL.
## 7. Assay Cut-Off:
Refer to K954687.
## 8. Carry-Over:
Not applicable.
## B Comparison Studies:
1. Comparison of Access Rubella IgG assay on the Access 2 Immunoassay System to the DxI 9000 Access Immunoassay Analyzer:
A method comparison study was conducted to compare the performance of the Access Rubella IgG assay on DxI 9000 Access Immunoassay Analyzer, to the Access Rubella IgG assay on the Access 2 Immunoassay System. A total of 162 native serum samples were tested on both instruments. PPA and NPA between the Rubella IgG assay when run on both systems are presented in Table 3.
Table 3. Performance Agreement of the Access Rubella IgG on the Access 2 Immunoassay System and the DxI 9000 Access Immunoassay Analyzer (n=162)
| Access Rubella IgG | Access 2 Immunoassay System | | | |
| --- | --- | --- | --- | --- |
| | | Reactive | Equivocal | Non-reactive |
| DxI 9000 Access Immunoassay Analyzer | Reactive | 77 | 0 | 0 |
| | Equivocal | 0 | 25 | 2 |
| | Non-Reactive | 0 | 2 | 56 |
| | Total | 77 | 27 | 58 |
| | PPA | 97.47% (77/79), 95% CI = 91.23% to 99.30% | | |
| | NPA | 96.55% (56/58), 95% CI = 88.27% to 99.05% | | |
K250588 - Page 8 of 9
{8}
2. **Matrix Comparison:**
Not applicable.
**C Clinical Studies:**
1. **Clinical Sensitivity:**
Not applicable.
2. **Clinical Specificity:**
Not applicable.
3. **Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):**
For clinical agreement study, refer to K954687.
**D Clinical Cut-Off:**
Not applicable.
**E Expected Values/Reference Range:**
Not applicable.
**VIII Proposed Labeling:**
The labeling supports the finding of substantial equivalence for this device.
**IX Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K250588 - Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.