SelectSite C304 Deflectable Catheter System (C304); C315 Delivery System (C315)

K250558 · Medtronic, Inc. · DQY · Mar 27, 2025 · Cardiovascular

Device Facts

Record IDK250558
Device NameSelectSite C304 Deflectable Catheter System (C304); C315 Delivery System (C315)
ApplicantMedtronic, Inc.
Product CodeDQY · Cardiovascular
Decision DateMar 27, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

C315: This non-therapeutic delivery system is intended to aid in the introduction and placement of cardiac leads into the right chambers of the heart. The leads are implanted in patients who are indicated for a cardiac implantable electronic device (CIED) for treatment of heart rhythm disorders. Refer to the respective CIED instructions for use for details about the types of heart rhythm abnormalities or patient conditions treated by each type of CIED. C304: This non-therapeutic delivery system is intended to aid in the introduction and placement of cardiac leads into the right chambers of the heart. The leads are implanted in patients who are indicated for a cardiac implantable electronic device (CIED) for treatment of heart rhythm disorders. Refer to the respective CIED instructions for use for details about the types of heart rhythm abnormalities or patient conditions treated by each type of CIED.

Device Story

Non-therapeutic delivery systems (C315 and C304) used to facilitate introduction/placement of cardiac leads into right heart chambers; used during CIED implantation procedures. C315 system includes catheter and dilator; C304 system includes deflectable catheter, dilator, universal slitter, valve, guidewire, needle, and syringe. C315 features hemostatic valve and radiopaque distal tip for fluoroscopic visualization; slittable design allows removal post-placement. C304 deflectable catheter enables access to coronary sinus and heart chambers. Operated by physicians in clinical settings. Output is physical access/guidance for lead placement; assists in successful CIED implantation, benefiting patients with heart rhythm disorders.

Clinical Evidence

No clinical data. Substantial equivalence supported by verification activities related to labeling updates for ISO 80369-7 compliance.

Technological Characteristics

C315 constructed of Polyether Block Amide and Polyethylene. C315 models: 5.4Fr inner/7.0Fr outer diameter, 20-43cm lengths. C304 models: C304-S59, C304-L69, C304-XL74. Features include hemostatic valves, radiopaque tips for fluoroscopy, and slittable designs. No changes to materials, design, or sterilization from predicates.

Indications for Use

Indicated for patients requiring cardiac implantable electronic device (CIED) lead placement in the right heart chambers for treatment of heart rhythm disorders.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below that. March 27, 2025 Medtronic, Inc. Samiksha Tiwari Senior Regulatory Affairs Specialist 8200 Coral Sea Street NE Mounds View, Minnesota 55112 Re: K250558 Trade/Device Name: SelectSite C304 Deflectable Catheter System (C304); C315 Delivery System (C315) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: February 24, 2025 Received: February 25, 2025 Dear Samiksha Tiwari: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/2/Picture/3 description: The image shows a digital signature. The signature is from Sara M. Royce. The signature was created on March 27, 2025 at 08:59:37 -04'00'. Sara Royce Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K250558 #### Device Name SelectSite C304 Deflectable Catheter System (C304); C315 Delivery System (C315) #### Indications for Use (Describe) C315: This non-therapeutic delivery system is intended to aid in the introduction and placement of cardiac leads into the right chambers of the heart. The leads are implanted in patients who are indicated for a cardiac implantable electronic device (CIED) for treatment of heart rhythm disorders. Refer to the respective CIED instructions for use for details about the types of heart rhythm abnormalities or patient conditions treated by each type of CIED. C304: This non-therapeutic delivery system is intended to aid in the introduction and placement of cardiac leads into the right chambers of the heart. The leads are implanted in patients who are indicated for a cardiac implantable electronic device (CIED) for treatment of heart rhythm disorders. Refer to the respective CIED instructions for use for details about the types of heart rhythm abnormalities or patient conditions treated by each type of CIED. | Type of Use (Select one or both, as applicable) | | | | |-------------------------------------------------|--|--|--| | | | | | X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary # Submitter Information | Date Prepared: | February 25, 2025 | |--------------------|--------------------------------------------------------------------------------------------------------------------| | Submitter: | Medtronic, Inc.<br>Cardiac Rhythm Management<br>8200 Coral Sea Street NE<br>Mounds View, MN 55112<br>ERN: 2182208 | | Contact Person: | Samiksha Tiwari<br>Sr Regulatory Affairs Specialist<br>Phone: 404-429-7275<br>email: Samiksha.tiwari@medtronic.com | | Alternate Contact: | Dustin Hawbaker<br>Sr Regulatory Affairs Manager<br>Phone: 952-256-3711<br>email: dustin.hawbaker@medtronic.com | # General Information | Trade Name: | SelectSite C304 Deflectable Catheter System<br>C315 Delivery System | |-----------------------|----------------------------------------------------------------------------------------------------| | Common Name: | Catheter Delivery System | | Regulation Number: | 21 CFR 870.1250 | | Product Code: | DQY | | Classification: | Class II | | Classification Panel: | Cardiovascular | | Predicate Device: | SelectSite C304 Deflectable Catheter System<br>(SelectSite C304<br>Deflectable Catheter<br>System) | | Predicate Device: | C315 Delivery Catheter<br>(C315 Delivery System) | {5}------------------------------------------------ # Device Description ### C315 Delivery System The Medtronic C315 Delivery System contains one catheter and one dilator constructed of Polyether Block Amide and Polyethylene respectively. It is designed to aid in the introduction of various types of pacing or defibrillator leads and catheters. There are seven models in the Medtronic C315 Delivery Catheter product family, all of which have the same inner and outer diameter (5.4Fr and 7.0Fr respectively). The models differ in useable length, which varies from 20cm to 43cm. Proximally, the C315 is equipped with a hemostatic valve, and the distal tip is radiopaque to facilitate imaging under fluoroscopy. The C315 is designed to be slittable, thereby allowing its removal after device placement. A variety of curves are available to accommodate various anatomies and different lead locations. #### SelectSite C304 Deflectable Catheter System The SelectSite C304 Deflectable Catheter System contains a single deflectable catheter, deflectable catheter dilator, universal slitter, valve, guidewire, needle, and syringe. The SelectSite C304 Deflectable Catheter System is designed to access the coronary sinus and the chambers of the heart. The percutaneous needle and syringe are used to access to venous insertion site, the guidewire to access the vein, the introducer valve to reduce blood loss during the implant procedure, the deflectable catheter to introduce a transvenous device, the deflectable catheter dilator to facilitate deflectable catheter passage and the guide catheter slitter to remove the deflectable catheter. The SelectSite C304 Deflectable Catheter System is available in three models which are the C304-S59, C304-L69, and C304-XL74. All components except the deflectable catheter and dilator are identical in each model. ### Indications for Use ### C315 Delivery System This non-therapeutic delivery system is intended to aid in the introduction and placement of cardiac leads into the right chambers of the heart. The leads are implanted in patients who are indicated for a cardiac implantable electronic device (CIED) for treatment of heart rhythm disorders. Refer to the respective CIED instructions for use for details about the types of heart rhythm abnormalities or patient conditions treated by each type of CIED. ### SelectSite C304 Deflectable Catheter System This non-therapeutic delivery system is intended to aid in the introduction and placement of cardiac leads into the right chambers of the heart. The leads are implanted in patients who are indicated for a cardiac implantable electronic device (CIED) for treatment of heart rhythm disorders. Refer to the respective CIED instructions for use for details about the types of heart rhythm abnormalities or patient conditions treated by each type of CIED. {6}------------------------------------------------ # Technological Characteristics The technology of the subject devices is identical to the respective predicates. There are no changes to the design, physical characteristics, materials, packaging, or sterilization presented in this submission. The subject devices have identical indications for use to the respective predicate devices. The intended use of the catheters, to introduce transvenous devices to the heart, remains the same. When compared to the predicate devices, the subject C315 Delivery System and subject SelectSite C304 Deflectable Catheter System presented in this submission have the same: - Intended use ● - Operating principle ● - Design features ● - Device functionality ● - Biological safety ● - Packaging materials ● - Shelf life ● The subject C315 Delivery System and SelectSite C304 Deflectable Catheter System devices differ from the respective predicates in that the subject devices have updated instructions for use (IFU) to indicate warning for potential small-bore misconnection. # Substantial Equivalence and Conclusion The labeling modification is supported by verification activities and makes the products compliant with clause 7f of ISO 80369-7. The subject devices are substantially equivalent to the specified predicate devices based on comparisons of the indications for use, intended use, device functionality, and technological characteristics.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...