The Extension Feeding Set with ENFit™ connectors is intended for use as an extension set for a Low-Profile Gastrostomy Replacement Tube (G. Button). It is intended to provide delivery of Enteral Nutrition from administration media with ENFit™ connector and or medication into a single patient, for a maximum of two weeks. Feeding via an Extension Feeding Set is advocated when nutrition cannot be provided to the patient by mouth and therefore cannot consume an adequate number of calories because of an underlying condition. Intended patient population: adult, children older than 1 month.
Device Story
Extension Feeding Set acts as a bridge between enteral nutrition administration media and a low-profile gastrostomy replacement tube (G. Button). Device consists of a polyvinylchloride (PVC) tube with an ENFit™ connector (ISO 80369-3 compliant) at the proximal end and a secure lock connector (straight or right-angle) at the distal end. Used in clinical or home settings to deliver nutrition or medication to patients unable to consume adequate calories orally. Single-patient use; maximum duration of two weeks. Healthcare providers or patients/caregivers connect the set to the G. Button to facilitate gravity or pump-assisted feeding. Benefits include secure, misconnection-resistant delivery of enteral nutrition.
Clinical Evidence
No clinical testing submitted. Substantial equivalence supported by non-clinical bench testing, including visual/dimensional inspection, small-bore connector testing, leakage, tensile strength, clamp functionality, flow rate, gravity flow, packaging, and connector performance testing. Biocompatibility testing was successfully completed.
Technological Characteristics
Materials: Polyvinylchloride (PVC) tube, ABS secure lock connector, Tritan ENFit™ connector. Design: Hollow tube with proximal ENFit™ (ISO 80369-3) and distal secure lock connector. Lengths: 5 cm, 30.5 cm, or 61 cm. Configuration: Straight or right-angle secure lock. Sterilization: Sterile. Connectivity: Mechanical connection only; no electronic or software components.
Indications for Use
Indicated for adults and children older than 1 month requiring enteral nutrition or medication delivery via a low-profile gastrostomy replacement tube (G. Button) when oral intake is inadequate due to underlying conditions.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 18, 2025
Sky Medical, a.s.
Yulia Surinova
General Manager
Vicenzy 24/2058
Samorin, 931 01
Slovakia
Re: K250481
Trade/Device Name: Extension Feeding Set with ENFit™ Connectors
Regulation Number: 21 CFR 876.5980
Regulation Name: Gastrointestinal Tube And Accessories
Regulatory Class: Class II
Product Code: PIF
Dated: February 19, 2025
Received: February 19, 2025
Dear Yulia Surinova:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250481 - Yulia Surinova
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K250481 - Yulia Surinova
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anthony Lee -S
Anthony Lee, Ph.D., M.B.A.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250481
Device Name
Extension Feeding Set with ENFit™ connectors
Indications for Use (Describe)
The Extension Feeding Set with ENFit™ connectors is intended for use as an extension set for a Low-Profile Gastrostomy Replacement Tube (G. Button). It is intended to provide delivery of Enteral Nutrition from administration media with ENFit™ connector and or medication into a single patient, for a maximum of two weeks.
Feeding via an Extension Feeding Set is advocated when nutrition cannot be provided to the patient by mouth and therefore cannot consume an adequate number of calories because of an underlying condition.
Intended patient population: adult, children older than 1 month.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
K250481 - Page 1/2
| 510(k) #: | 510(k) Summary | Prepared on: 2025-04-15 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Sky Medical, a.s. | |
| Applicant Address | Vicenzy 24/2058 Samorin 931 01 Slovakia | |
| Applicant Contact Telephone | +421 903 503562 | |
| Applicant Contact | Ms. Yulia Surinova | |
| Applicant Contact Email | yulia@med-sky.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Extension Feeding Set with ENFit™ connectors | |
| Common Name | Gastrointestinal tube and accessories | |
| Classification Name | Gastrointestinal Tubes With Enteral Specific Connectors | |
| Regulation Number | 876.5980 | |
| Product Code(s) | PIF | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K141631 | AQUARIUS™ Extension Feeding set | KNT |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| Description: The Extension Feeding Set with ENFit™ connectors (Extension Feeding Set), comprises a hollow tube with two different connectors at each end: an ENFit™ connector connecting to an administration media, the other is a secure lock connector for connecting to a gastric feeding device. The Extension Feeding Set is sterile, single patient use device, consisting of a polyvinylchloride tube, ABS secure lock connector, Tritan ENFit™ connector. The connectors on the proximal end of the extension sets are ENFit™ ISO 80369-3 compliant. The ENFit™ connector allows for connections of enteral specific applications while reducing the likelihood of misconnections to non-enteral devices. Extension Feeding Sets are available in three lengths: 5 cm (2"), 30.5 cm (12") or 61 cm (24"). All Extension Feeding Sets are available with either a right-angle secure lock connector or a straight secure lock connector. Both connectors are compatible with G. Button. Indications for use: The Extension Feeding Set with ENFit™ connectors is intended for use as an extension set for a Low-Profile Gastrostomy Replacement Tube (G. Button). It is intended to provide delivery of Enteral Nutrition from administration media with ENFit™ connector and or medication into a single patient, for a maximum of two weeks. Feeding via an Extension Feeding Set is advocated when nutrition cannot be provided to the patient by mouth and therefore cannot consume an adequate number of calories because of an underlying condition. Intended patient population: adult, children older than 1 month. Shelf life: 3 years. | | |
{5}
K250481 - Page 2/2
Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The Extension Feeding Set with ENFit™ connectors is intended for use as an extension set for a Low-Profile Gastrostomy Replacement Tube (G. Button). It is intended to provide delivery of Enteral Nutrition from administration media with ENFit™ connector and or medication into a single patient, for a maximum of two weeks.
Feeding via an Extension Feeding Set is advocated when nutrition cannot be provided to the patient by mouth and therefore cannot consume an adequate number of calories because of an underlying condition.
Intended patient population: adult, children older than 1 month.
Indications for Use Comparison
21 CFR 807.92(a)(5)
Both the subject device and the predicate device are intended to be used as extension sets for gastric feeding devices to provide delivery of feeding media from nutrition reservoir into a patient feeding tube.
Slight difference in the wording of the indications for use are not critical to the intended use of the device, the differences do not affect the safety and effectiveness of the device when used as labeled.
Technological Comparison
21 CFR 807.92(a)(6)
The subject device and the predicate device have the same technological characteristics (i.e., design, material, chemical composition) as the predicate device.
There are slight differences in the material used in the predicate device in ENFit™ connectors, tube and clamp. These differences do not have effect on patient and do not increase risks associated with the use of the device.
There are no significant differences in the design, construction and technological processes used during manufacturing of the subject device and the predicate.
Both the subject device and the predicate device properties and have been successfully tested per applicable standards and passed also relevant testing for biocompatibility.
The identified small differences will not raise new question on the new safety and effectiveness of the proposed device as evidence by the provided Substantial Equivalence Comparison.
Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
The following nonclinical testing was conducted to support this submission and determine substantial equivalence:
- Visual and dimensional inspection
- Small-bore connectors testing
- Leakage test
- Tensile strength testing
- Extension Feeding Set clamp functionality testing
- Flow rate testing
- Gravity flow test
- Packaging testing
- Performance testing of connectors
- Biocompatibility testing was also completed successfully.
No clinical testing is submitted.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.