ENLIGHT 2100 is a non-invasive, radiation free medical device that provides information from impedance variation from a cross-section of a patient's thorax. This information is presented to the clinician user as an adjunctive tool to other clinical information in order to support the user's assessment of variations in regional air content within a cross section of a patient's thorax. ENLIGHT 2100 also provides respiratory parameters based on spirometric monitoring. It is intended for mechanically ventilated patients, from adults to infants, older than 29 days, whose thorax perimeter is within the range of 37.5 -134 cm. ENLIGHT 2100 does not measure regional ventilation of the lungs.
Device Story
ENLIGHT 2100 is a ventilatory electrical impedance tomograph (EIT) used in hospital settings for mechanically ventilated patients. It utilizes 16-32 electrodes placed around the thorax to measure local impedance variations (LIV) during respiratory cycles, which correlate to regional air content. The device processes these voltage measurements to generate dynamic images and distribution maps of the thorax cross-section. It also provides respiratory parameters derived from spirometric monitoring. Clinicians use the output—including A/P and R/L ratios, ΔEEZ, and compliance loss metrics (HPCzL, LPCzL)—as an adjunctive tool to support clinical assessment of lung air content. The device automates calculations previously performed manually by clinicians using the predicate device. It does not measure regional ventilation directly.
Clinical Evidence
Bench testing only. Performance testing verified the accuracy of automated calculations for ΔEEZ, Percentage of Higher Pressure Compliancez Loss (HPCzL), and Percentage of Lower Pressure Compliancez Loss (LPCzL). Results showed uncertainty of +/- (0.1 + 10% of reading) for ΔEEZ and differences not exceeding 10% for compliance loss metrics compared to reference values.
Technological Characteristics
Electrical impedance tomography (EIT) using 16-32 electrodes. Operates on Yocto-based Linux OS. Non-sterile. Reusable electrode belt and reference cable. Provides dynamic imaging and distribution maps based on voltage measurements. Software includes automated calculation of respiratory parameters and compliance loss metrics.
Indications for Use
Indicated for mechanically ventilated adult and pediatric patients (older than 29 days) with thorax perimeters between 37.5 and 134 cm. Used as an adjunctive tool for assessing regional air content variations in the thorax.
Regulatory Classification
Identification
A ventilatory electrical impedance tomograph is a prescription non-invasive, non-radiological ventilatory device that provides an assessment of local impedance variation within a cross-section of a patient's thorax.
Special Controls
In combination with the general controls of the FD&C Act, the ventilatory electrical impedance tomograph is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including the following:
(i) Characterization of device parameters, including signal-to-noise ratio, voltage accuracy, drift, reciprocity accuracy, amplitude response, position error, and ringing;
(ii) Real time evaluation of local impedance variation;
(iii) Plethysmogram accuracy testing; and
(iv) Use life testing of reusable components.
(3) Performance data must validate reprocessing instructions for any reusable components of the device.
(4) Performance data must demonstrate the electrical, thermal, and mechanical safety and the electromagnetic compatibility of the device.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Labeling must include the following:
(i) Guidance for interpretation of the images generated;
(ii) A warning that the device should be removed before use of a defibrillator, or defibrillator interaction information based on defibrillator performance testing with the device;
(iii) A use life for any reusable components; and
(iv) Instructions for reprocessing any reusable components.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 10, 2025
Timpel S.A.
% Paul Dryden
Consultant
ProMedic Consulting LLC
131 Bay Point Dr NE
Saint Petersburg, Florida 33704
Re: K250464
Trade/Device Name: Enlight 2100 (TPL-E2103-0)
Regulation Number: 21 CFR 868.1505
Regulation Name: Ventilatory Electrical Impedance Tomograph
Regulatory Class: Class II
Product Code: QEB, BZK
Dated: February 16, 2025
Received: February 18, 2025
Dear Paul Dryden:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
2K250464 - Paul Dryden
Page
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
3K250464 - Paul Dryden
Page
Sincerely,
Ethan L. Nyberg -S
Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250464 | |
| Device Name ENLIGHT 2100 | |
| Indications for Use (Describe) ENLIGHT 2100 is a non-invasive, radiation free medical device that provides information from impedance variation from a cross-section of a patient's thorax. This information is presented to the clinician user as an adjunctive tool to other clinical information in order to support the user's assessment of variations in regional air content within a cross section of a patient's thorax. ENLIGHT 2100 also provides respiratory parameters based on spirometric monitoring. It is intended for mechanically ventilated patients, from adults to infants, older than 29 days, whose thorax perimeter is within the range of 37.5 -134 cm. ENLIGHT 2100 does not measure regional ventilation of the lungs. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
K250464
510(k) Summary
8-Sep-25
Page 1 of 9
Sponsor: TIMPEL S.A.
Rua Simão Álvares, 356 Cj. 41/42
Pinheiros, São Paulo/SP - Brazil
T - +55-113088-0305
Rafael Holzhacker - CEO
Submission Correspondent: Paul Dryden
ProMedic Consulting, LLC
Proprietary or Trade Name: ENLIGHT 2100
Regulation Number: 868.1505
Regulation Name: Ventilatory electrical impedance tomograph
Product code: QEB
Predicate: ENLIGHT 2100 - K222897
Regulation Number: 868.1505
Regulation Name: Ventilatory electrical impedance tomograph
Product code: QEB
Device Description:
ENLIGHT 2100 is a Ventilatory electrical impedance tomograph that uses several electrodes (between 16 and 32) placed around the patient's thorax to assess regional impedance variation in a lung slice (tomography). It provides a relative measurement, so it only provides information on variations in local impedance.
ENLIGHT 2100 estimates Local Impedance Variation, occurring in a cross section of the thorax during a respiratory cycle, and which are linearly related to Variations in Regional Air Content within the lung.
Principle of Operation:
Electrical impedance tomography (EIT) provides information on Local Impedance Variation (LIV) within a cross section of a patient's thorax.
Indications for Use:
ENLIGHT 2100 is a non-invasive, radiation free medical device that provides information from impedance variation from a cross-section of a patient's thorax.
This information is presented to the clinician user as an adjunctive tool to other clinical information in order to support the user's assessment of variations in regional air content within a cross section of a patient's thorax.
ENLIGHT 2100 also provides respiratory parameters based on spirometric monitoring.
It is intended for mechanically ventilated patients, from adults to infants, older than 29 days, whose thorax perimeter is within the range of 37.5 -134 cm.
ENLIGHT 2100 does not measure regional ventilation of the lungs.
{6}
K250464
# 510(k) Summary
8-Sep-25
Page 2 of 9
## Patient Population:
Mechanically ventilated adult and pediatric patients whose thorax perimeter is within the range of 37.5 -134 cm.
## Environments of use:
Hospital setting.
## Substantial Equivalence Discussion
**Indications** – There are no differences in the indication for use.
**Patient Population** – There is no difference in the patient population.
**Environment of Use** – There is no difference in the environment of use.
**Technological Characteristics** – There are no differences in the Technological Characteristics. There is no hardware change, and the algorithms related to EIT data are not modified.
**Software Change** – The Processing and Interface Module (PIM) is responsible for the interface to the user. The differences do not impact on the software structure. Software verification and validation were executed to ensure the device’s safety and performance.
**Parameters** – The subject device will include parameters related to the already cleared Distribution Change Map, normalized by Driving Pressure, to complement the information of local impedance variation.
There are no standard values or expectations related to the parameters shown, as well as clinical significance for the absolute values. Variations of the same parameter shall be used as adjunctive tool to other clinical information in order to support the user’s assessment of variations in regional air content within a cross section of a patient’s thorax.
**Cleaning methods** – Electrode belt and reference cable are now reusable. These reusable parts are not intended to contact the patient. There is no change in material or hardware. Cleaning instructions are described in the instructions for use.
## Non-clinical testing
**Biocompatibility** – There is no difference in the patient contact or materials between the proposed device and the predicate. Prior testing has been leveraged to support biocompatibility.
**Human Factors** – There are no changes in the primary functions of the device or risk control measures that require new human factors validation.
**Bench Testing** – We have performed performance tests to check the automatic calculation of the parameters obtained at the Trend Screen.
- EIT Parameters
- $\Delta$ EEZ
- Percentage of Higher Pressure Compliance$_z$ Loss (HPCzL)
- Percentage of Lower Pressure Compliance$_z$ Loss (LPCzL)
{7}
K250464
510(k) Summary
8-Sep-25
Page 3 of 9
## Substantial Equivalence Conclusion
The ENLIGHT 2100 has the same indications for use as the predicate device, and there are no differences in technological characteristics which would raise different questions of safety and effectiveness. Therefore, these devices are substantially equivalent.
{8}
K250464
510(k) Summary
8-Sep-25
Page 4 of 9
Table 1 – Comparison – Subject vs. Predicate – EIT Data
| Attributes | Subject ENLIGHT 2100 | Predicate ENLIGHT 2100 | Explanation of Differences |
| --- | --- | --- | --- |
| 510(k) | TBD | K222897 | New 510(k) number |
| Product Classification 21 CFR | 868.1505 – QEB | 868.1505 – QEB | Same product code. |
| Indications for Use | ENLIGHT 2100 is a non-invasive, radiation free medical device that provides information from impedance variation from a cross-section of a patient’s thorax.
This information is presented to the clinician user as an adjunctive tool to other clinical information in order to support the user’s assessment of variations in regional air content within a cross section of a patient’s thorax.
ENLIGHT 2100 also provides respiratory parameters based on spirometric monitoring.
It is intended for mechanically ventilated patients, from adults to infants, older than 29 days, whose thorax perimeter is within the range of 37.5 -134 cm.
ENLIGHT 2100 does not measure regional ventilation of the lungs. | ENLIGHT 2100 is a non-invasive, radiation free medical device that provides information from impedance variation from a cross-section of a patient’s thorax.
This information is presented to the clinician user as an adjunctive tool to other clinical information in order to support the user’s assessment of variations in regional air content within a cross section of a patient’s thorax.
ENLIGHT 2100 also provides respiratory parameters based on spirometric monitoring.
It is intended for mechanically ventilated adults and pediatric patients in a hospital setting, whose thorax perimeter is within the range of 37.5 – 134 cm.
ENLIGHT 2100 does not measure regional ventilation of the lungs. | Similar |
| Patient Population | Adult and Pediatric patients, whose thorax perimeter is within the range of 37.5 -134 cm. | Adult and Pediatric patients, whose thorax perimeter is within the range of 37.5 -134 cm. | Similar |
| Patient type | Mechanically ventilated patients | Mechanically ventilated patients | Similar |
| Prescriptive | Yes | Yes | Similar |
{9}
K250464
# 510(k) Summary
8-Sep-25
Page 5 of 9
| Attributes | Subject ENLIGHT 2100 | Predicate ENLIGHT 2100 | Explanation of Differences |
| --- | --- | --- | --- |
| Principle of Operation | Electrical Impedance Tomography based on voltage measures to estimate local impedance variation within a cross-section of a patient’s thorax. | Electrical Impedance Tomography based on voltage measures to estimate local impedance variation within a cross-section of a patient’s thorax. | Similar |
| Contraindications | No changes, presented in Instructions for use | No changes, presented in Instructions for use | Similar |
| Environment of Use | Hospital Setting | Hospital Setting | Similar |
| Duration of Use | Up to 30 days, with Addere Change each 48 hours. | Up to 30 days, with Addere Change each 48 hours. | Similar |
| Service-life | ENLIGHT 2100 has 7 years of useful life. Electrode Belt and Reference Cable have 2 years of useful life. | ENLIGHT 2100 has 7 years of useful life. Electrode Belt and Reference Cable have 2 years of useful life. | Similar |
| Shelf life | Addere has 2 year of shelf life. Electrode Belt has 1 year of shelf life. | Addere has 2 year of shelf life. Electrode Belt has 1 year of shelf life. | Similar |
| Non-sterile | There are no sterile components or accessories. | There are no sterile components or accessories. | Similar |
| Cleaning methods | Cleaning and disinfection for the main device, electrode belt and reference cable.
Shaper and Addere (which are the accessories that get in contact with the patient) remain as single patient use. | Cleaning and disinfection only for the main device.
Accessories are single patient use (electrode belt, reference cable, Addere and Shaper). | Electrode belt and reference cable are now reusable. |
| Operating System | Yocto based Linux OS for the Processing and Interface Module (PIM). | Yocto based Linux OS for the Processing and Interface Module (PIM) | Similar |
| Features | | | |
{10}
K250464
# 510(k) Summary
# 8-Sep-25
Page 6 of 9
| Attributes | Subject ENLIGHT 2100 | Predicate ENLIGHT 2100 | Explanation of Differences |
| --- | --- | --- | --- |
| Accessories / Available Sizes | Electrode Belt sizes P0, P1, P2, 4S, 5S, XXS, XS, S, M, L, XL
Addere sizes P0, P1, P2, 4S, 5S, XXS, XS, S, M, L, XL
Shaper No 1 and Shaper No 2
Reference Cable (Single size) | Electrode Belt sizes P0, P1, P2, 4S, 5S, XXS, XS, S, M, L, XL
Addere sizes P0, P1, P2, 4S, 5S, XXS, XS, S, M, L, XL
Shaper No 2
Reference Cable (Single size) | Adding a Shaper size does not raise any new risks compared to the predicate. |
| Patient Contact per ISO 10993-1 | Surface contact
Intact skin
Prolonged duration (Up to 30 days) | Surface contact
Intact skin
Prolonged duration (Up to 30 days) | Similar |
| Image Screen | Change of the screen name to Distribution Screen: | Screen Name - Images Screen: | Change of the screen name from Image Screen to Distribution Screen. Distribution Screen displays the same “Dynamic Image” and “Distribution Map” as the Images Screen. |
| Distribution Screen | Distribution Screen displays Distribution Map (updated at every breath), Dynamic Image (updated continuously), and A/P and R/L ratios, updated every 30 seconds. | Images Screen displays Distribution Map (updated at every breath) and Dynamic Image (updated continuously).
The Trend Screen displays A/P and R/L ratios, updated every 30 seconds. | The subject device will display on the Distribution Screen the A/P and R/L ratios that users can already see on the predicate device’s Trends Screen |
| Distribution Screen | Distribution Screen displays Distribution map (updated at every breath), Dynamic Image (updated continuous), as well as respiratory parameters based on spirometric measurements (updated every 30 seconds). | Images Screen displays Distribution map (updated at every breath) and Dynamic Image (updated continuous).
The Trend Screen displays respiratory parameters, updated every 30 seconds. | The subject device will display on the Distribution Screen the same respiratory parameters that users can already see in the predicate device’s Trends Screen. |
{11}
510(k) Summary
8-Sep-25
Page 7 of 9
K250464
| Attributes | Subject ENLIGHT 2100 | Predicate ENLIGHT 2100 | Explanation of Differences |
| --- | --- | --- | --- |
| Trends Screen (Layout 1 - Two Graphs and Distribution Change Map) | Data from the device is presented to the user in graphic format and with possibility to compare two moments (reference and cursor) to assist the user in the patient’s assessment:
Data in the trend is kept for 48 hours of continuous monitoring.
The user can select two time points for comparison: reference and cursor. Time stamp of the selected moments (reference and cursor) by the user are shown | Data from the device is presented to the user in graphic format and with possibility to compare two moments (reference and cursor) to assist the user in the patient’s assessment:
Data in the trend is kept for 48 hours of continuous monitoring.
The user can select two time points for comparison: reference and cursor. Time stamp of the selected moments (reference and cursor) by the user are shown. | Similar |
| Trends Screen (Layout 2 – Multi-Step) | The Compliance_{z} Change Map at the MultiStep Layout represents the comparison of more than two Compliance_{z} Maps, selected by the user.
At the Compliance_{z} Change Map, regions classified as Lower Pressure Compliance_{z} Loss (LPC_{z}L) are shown in blue, while regions classified as Higher Pressure Compliance_{z} Loss (HPC_{z}L) are shown in white. The percentages of LPC_{z}L and HPC_{z}L observed in a Compliance_{z} Change Map are displayed. | Distribution Change Map represents the comparison between two Distribution Maps selected by the user, normalized by driving pressure.
At the Distribution Change Map, regions with increased impedance variation normalized by driving pressure (increased Compliance_{z}) are shown in gold, while regions with decreased impedance variation normalized by driving pressure (decreased Compliance_{z}) are shown in silver. | The predicate device enables users to compare two Maps, while the MultiStep Layout of the subject device allows users to compare multiple Maps at the same time. The results of the comparison among more than two maps at the subject device are shown in blue and white (instead of gold and silver as the predicate device). The percentages of regions identified as LPC_{z}L and HPC_{z}L are also displayed. |
| Trends Screen
End Expiratory Impedance (EEZ) | The subject device automatically identifies the baseline of the plethysmogram and displays the EEZ in a graph.
In addition to the graph, the device displays ΔEEZ between REF and CUR as the change in EEZ divided by the TVz at the REF. | The user can obtain the baseline of the plethysmogram (equivalent to EEZ) by placing the cursors (reference and cursor) on the baseline of the plethysmogram.
The user can calculate the ΔEEZ by positioning the reference and cursor on the baselines of the plethysmogram, and dividing the difference by the TVz value read at REF on the TVz graph. | The subject device will automate a calculation that could be done manually using the predicate device. |
{12}
K250464
# 510(k) Summary
8-Sep-25
Page 8 of 9
| Attributes | Subject ENLIGHT 2100 | Predicate ENLIGHT 2100 | Explanation of Differences |
| --- | --- | --- | --- |
| Compliance_{z} Change Display | The Compliance_{z} Change Map at the Compliance_{z} Change Display represents the comparison of more than two Compliance_{z} Maps, acquired by the device according to configurations defined by the user.
At the Compliance_{z} Change Map, regions classified as Lower Pressure Compliance_{z} Loss (LPC_{z}L) are shown in blue, while regions classified as Higher Pressure Compliance_{z} Loss (HPC_{z}L) are shown in white. The percentages of LPC_{z}L and HPC_{z}L observed in a Compliance_{z} Change Map are displayed and plotted in a graph. | Distribution Change Map represents the comparison between two Distribution Maps selected by the user, normalized by driving pressure.
At the Distribution Change Map, regions with increased impedance variation normalized by driving pressure (increased Compliance_{z}) are shown in gold, while regions with decreased impedance variation normalized by driving pressure (decreased Compliance_{z}) are shown in silver. | The predicate device enables users to compare two Maps, while the Compliance_{z} Change Display of the subject device allows users to compare multiple Maps at the same time.
The results of the comparison among more than two maps at the subject device are shown in blue and white (instead of gold and silver as the predicate device). The percentages of regions identified as LPC_{z}L and HPC_{z}L are also displayed.
The results of the comparison among more than two maps at the subject device are shown in blue and white (instead of gold and silver as the predicate device). The percentages of regions identified as LPC_{z}L and HPC_{z}L are also displayed and plotted. |
| Performance Characteristics – Bench Test | | | |
| ΔEEZ | For the range between -15.0 and +15.0, uncertainty of +/- (0.1 + 10% of reading) | Same | The predicate didn’t present the numeric parameter, but as the hardware is the same and the EIT algorithm is the same, the performance is equivalent for this characteristic. |
{13}
K250464
# 510(k) Summary
8-Sep-25
Page 9 of 9
| Attributes | Subject ENLIGHT 2100 | Predicate ENLIGHT 2100 | Explanation of Differences |
| --- | --- | --- | --- |
| Percentage of Higher Pressure Compliance_{z} Loss (HPC_{z}L) | Difference between reference and actual values of HPCzL (%) and of LPCzL (%) is not larger than 10 (%) | Same | The predicate didn't present the numeric parameter, but as the hardware is the same and the EIT algorithm is the same, the performance is equivalent for this characteristic. |
| Percentage of Lower Pressure Compliance_{z} Loss (LPC_{z}L) | Difference between reference and actual values of HPCzL (%) and of LPCzL (%) is not larger than 10 (%) | Same | The predicate didn't present the numeric parameter, but as the hardware is the same and the EIT algorithm is the same, the performance is equivalent for this characteristic. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.