RT Elements (4.5); (Elements) Multiple Brain Mets SRS; (Elements) Cranial SRS; (Elements) Spine SRS; (Elements) Cranial SRS w/ Cones; (Elements) RT Planning Platform; (Elements) Dose Review; (Elements) Retreatment Review; Elements Segmentation [Cranial , Basal Ganglia, Head & Neck, Pelvic, Spine, Thoracic & Spine, Extracranial] RT; Elements AI Tumor Segmentation RT; Elements SmartBrush [Angio, Spine] RT; Elements Object Management RT
K250440 · Brainlab AG · MUJ · Jun 17, 2025 · Radiology
Device Facts
Record ID
K250440
Device Name
RT Elements (4.5); (Elements) Multiple Brain Mets SRS; (Elements) Cranial SRS; (Elements) Spine SRS; (Elements) Cranial SRS w/ Cones; (Elements) RT Planning Platform; (Elements) Dose Review; (Elements) Retreatment Review; Elements Segmentation [Cranial , Basal Ganglia, Head & Neck, Pelvic, Spine, Thoracic & Spine, Extracranial] RT; Elements AI Tumor Segmentation RT; Elements SmartBrush [Angio, Spine] RT; Elements Object Management RT
Applicant
Brainlab AG
Product Code
MUJ · Radiology
Decision Date
Jun 17, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Cranial tumor segmentation
Anatomical Patient Model
Dice >= 0.7, Recall >= 0.8, Precision >= 0.8
Mean Dice 0.75, Mean Precision 0.86, Mean Recall 0.85
—
—
Validation study: 595 scans from 412 patients (1878 annotations) from multiple clinical sites in the US and Europe.
>1 (external/independent annotator team including US radiologists and non US radiologists)
Indications for Use
The device is intended for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment and indicated for cranial, head and neck and extracranial lesions.
Device Story
RT Elements 4.5 is a software suite for radiation therapy treatment planning and dose optimization. It processes co-registered DICOM image sets (CT, MRI, PET) to create 3D structures and generate RT dose plans for linear accelerators. Modules include automated segmentation (AI-based for cranial tumors), dose review, quality assurance, and retreatment analysis. Used in clinical settings by medical physicists, radiation oncologists, and dosimetrists. The software supports VMAT, SRS, SRT, and SBRT workflows. It integrates with third-party cloud-based contouring systems (Art-Plan) and utilizes an Anatomical Patient Model (APM) backend for image processing. Output is a DICOM RT dose plan used to control radiation delivery. Benefits include optimized planning efficiency, improved conformity, and reduced normal tissue dose via dynamic jaw tracking and advanced dose calculation algorithms (Pencil Beam, Monte Carlo).
Clinical Evidence
Bench testing only. Performance validated via software integration, GUI, and unit tests. Dosimetric accuracy verified against phantom measurements for Pencil Beam and Monte Carlo algorithms. AI Tumor Segmentation validated using 595 3D CE-T1 MR scans from 412 adult patients. Primary endpoints: Dice coefficient, precision, and recall. Results: Mean Dice 0.75 (95% CI: 0.74-0.76), Mean Precision 0.86 (95% CI: 0.85-0.88), Mean Recall 0.85 (95% CI: 0.83-0.87). All metrics met pre-defined acceptance criteria (Dice ≥0.7, Recall/Precision ≥0.8).
Technological Characteristics
Software-only medical device. Operates on Windows-based workstations/servers. Supports DICOM standards for input/output. Dose calculation algorithms: Pencil Beam, Monte Carlo, Circular Cone. Connectivity: Networked, supports cloud-based third-party integration (Art-Plan). AI/ML: Deep learning-based cranial tumor segmentation (Anatomical Patient Model).
Indications for Use
Indicated for radiation treatment planning for stereotactic, conformal, Linac-based radiation therapy of cranial, head and neck, and extracranial lesions in adult patients.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Contouring 5.0 (part of Brainlab Elements 7.0) (K243633)
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 17, 2025
Brainlab AG
Sadwini Suresh
QM Consultant
Olof-Palme-Str. 9
Munich, 81829
Germany
Re: K250440
Trade/Device Name: RT Elements (4.5); (Elements) Multiple Brain Mets SRS; (Elements) Cranial SRS; (Elements) Spine SRS; (Elements) Cranial SRS w/ Cones; (Elements) RT Planning Platform; (Elements) Dose Review; (Elements) Retreatment Review; Elements Segmentation [Cranial, Basal Ganglia, Head & Neck, Pelvic, Spine, Thoracic & Spine, Extracranial] RT; Elements AI Tumor Segmentation RT; Elements SmartBrush [Angio, Spine] RT; Elements Object Management RT
Regulation Number: 21 CFR 892.5050
Regulation Name: Medical Charged-Particle Radiation Therapy System
Regulatory Class: Class II
Product Code: MUJ, QIH
Dated: February 14, 2025
Received: May 19, 2025
Dear Sadwini Suresh:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{2}
K250440 - Sadwini Suresh
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K250440 - Sadwini Suresh
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250440
Device Name
RT Elements (4.5);
(Elements) Multiple Brain Mets SRS;
(Elements) Cranial SRS;
(Elements) Spine SRS;
(Elements) Cranial SRS w/ Cones;
(Elements) RT Planning Platform;
(Elements) Dose Review;
(Elements) Retreatment Review;
Elements Segmentation [Cranial, Basal Ganglia, Head & Neck, Pelvic, Spine, Thoracic & Spine, Extracranial] RT;
Elements Al Tumor Segmentation RT;
Elements SmartBrush [Angio, Spine] RT;
Elements Object Management RT
Indications for Use (Describe)
The device is intended for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment and indicated for cranial, head and neck and extracranial lesions.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}
BRAINLAB
K250440
# 510(k) Summary
June 17, 2025
| General Information | |
| --- | --- |
| Manufacturer | Brainlab AG; Olof-Palme Str.9; 81829, Munich, Germany |
| Establishment Registration | 8043933 |
| Trade Name | RT Elements 4.5 |
| Classification Name | System, planning, radiation therapy treatment |
| Product Code | MUJ; QIH |
| Regulation Number | 892.5050 |
| Regulatory Class | II |
| Panel | Radiology |
| Predicate Device | RT Elements (4.0); K223279 |
| Reference Device | Contouring 5.0 (part of Brainlab Elements 7.0 - K243633) |
| Contact Information | |
| Primary Contact | Alternate Contact |
| Sadwini Suresh
QM Consultant
Phone: +49 89 99 15 68 0
Email: regulatory.affairs@brainlab.com | Chiara Cunico
Senior Manager Regulatory Affairs
Phone: +49 89 99 15 68 0
Email: chiara.cunico@brainlab.com |
1. Indications for Use
The device is intended for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment and indicated for cranial, head and neck and extracranial lesions.
2. Device Description
RT Elements are computed-based software applications for radiation therapy treatment planning and dose optimization for linac-based conformal radiation treatments, i.e. stereotactic radiosurgery (SRS), fractionated stereotactic radiotherapy (SRT) or stereotactic ablative radiotherapy (SABR), also known as stereotactic body radiation therapy (SBRT) for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck, and extracranial lesions.
PAGE 1 OF 12
{6}
BRAINLAB
The device consists of the following software modules:
| Software | Description |
| --- | --- |
| Multiple Brain Mets SRS 4.5 | Multiple Brain Mets SRS provides optimized planning and display for cranial multi-metastases radiation treatment planning. |
| Cranial SRS 4.5 | Cranial SRS provides optimized planning and display for cranial radiation treatment planning. |
| Spine SRS 4.5 | Spine SRS provides optimized planning and display for spine metastases. |
| Cranial SRS w/ Cones 4.5 | Cranial SRS w/ Cones provides planning and display for functional diseases (e.g. trigeminal neuralgia) or cranial lesion radiation treatment. |
| RT Contouring 4.5 | RT Contouring contains features to outline, refine, combine and manipulate structures in patient image data for planning of cranial and extracranial radiotherapy treatments. The Cranial Tumor Segmentation feature uses AI/ML. The algorithm was trained on MRI image data with contrast-enhancing tumors from multiple clinical sites, including a wide variety of scanner models and patient characteristics. |
| RT QA 4.5 | RT QA contains features for patient specific quality assurance. Use RT QA to recalculate patient treatment plans on a phantom to verify that the patient treatment plan fulfills the planning requirements. |
| Dose Review 4.5 | Dose Review contains features for review of isodose lines, review of DVHs, dose comparison and dose summation. |
| Brain Mets Retreatment Review 4.5 | Retreatment Review allows the preparation and analysis of the next treatment by providing a 3D visualization of all previously treated and new metastases and review options for the summed dose of previous treatment plans. |
| Physics Administration 7.5 | The Physics Administration is a tool that allows to administer measured beam data and machine profiles for RT Elements software. The Software can be used to define and edit Hounsfield to electron density conversion tables. |
3. Substantial Equivalence
PAGE 2 OF 12
{7}
BRAINLAB
| Topic/ Feature | Predicate Device (RT Elements 4.0 - K223279) | Subject Device (RT Elements 4.5) |
| --- | --- | --- |
| Indications For Use | The device is intended for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment and indicated for cranial, head and neck and extracranial lesions. | The device is intended for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment and indicated for cranial, head and neck and extracranial lesions. |
| Applications/Elements included | • Multiple Brain Mets SRS
• Cranial SRS
• Spine SRS
• Cranial SRS w/ Cones
• RT QA – Recalculation
• RT QA – Patient Specific QA
• RT QA – Beam Model Verification
• Dose Review
• Brain Mets Retreatment Review
• Physics Administration
• Phantom Definition
• Tissue Model | • Multiple Brain Mets SRS
• Cranial SRS
• Spine SRS
• Cranial SRS w/ Cones
• RT QA – Recalculation
• RT QA – Patient Specific QA
• RT QA – Beam Model Verification
• Dose Review
• Brain Mets Retreatment Review
• Physics Administration
• Phantom Definition
• Tissue Model
• SmartBrush (RT; RT Spine; Angio)
• Object Management |
| Operating Principle | The device is composed by software only and can be controlled via mouse and keyboard. The software is displayed on a computer screen.
The input of the Brainlab RT Elements are co-registered DICOM image sets with segmented objects.
The output is a DICOM RT dose plan for further processing and finally to control the radiation delivery to a patient using a linear accelerator. | The device is composed by software only and can be controlled via mouse and keyboard. The software is displayed on a computer screen.
Based on co-registered DICOM image sets, the software can create 3D structures or objects which are further used for the creation of RT dose plans.
The output is a DICOM RT dose plan for further processing and finally to control the radiation delivery to a patient using a linear accelerator. |
PAGE 3 OF 12
{8}
BRAINLAB
| Topic/ Feature | Predicate Device (RT Elements 4.0 - K223279) | Subject Device (RT Elements 4.5) |
| --- | --- | --- |
| Display Resolution | • 1920 x 1200 (WUXGA)
• 2560 x 1440 (WQHD)
• 3840 x 2048 (4K) | • 1920 x 1080 (FHD)
• 1920 x 1200 (WUXGA)
• 2560 x 1440 (WQHD)
• 3840 x 2048 (4K) |
| Supported Operating Systems | • Windows 10 1507
• Windows 10 1607
• Windows 10 1809
• Windows 10 21H2
• Windows Server 2012 R2 64-bit
• Windows Server 2016
• Windows Server 2019
• Windows Server 2022 | • Windows 10 1607
• Windows 10 1809
• Windows 10 v 21H2
• Windows Server 2016
• Windows Server 2019
• Windows Server 2022 |
| Data Input | • DICOM Images
• DICOM Segmentation
• DICOM Registration
• DICOM Key Object Selection Document
• DICOM Raw (Brainlab Load/Save)
• DICOM RT Dose
• DICOM RT Structure Set
• Machine Profiles
• HuToEd-Tables | • DICOM Images
• DICOM Segmentation
• DICOM Registration
• DICOM Key Object Selection Document
• DICOM Raw (Brainlab Load/Save)
• DICOM RT Dose
• DICOM RT Structure Set
• Machine Profiles
• HuToEd-Tables |
| Intended User Profile | The intended users are medical professionals who perform radiation treatment planning (medical physicists, radiation oncologists, dosimetrists, physicians, etc.). | The intended users are medical professionals who perform radiation treatment planning (medical physicists, radiation oncologists, dosimetrists, physicians, etc.). |
PAGE 4 OF 12
{9}
BRAINLAB
| Topic/ Feature | Predicate Device (RT Elements 4.0 - K223279) | Subject Device (RT Elements 4.5) |
| --- | --- | --- |
| Data Output | • DICOM Images
• DICOM RT Struct
• DICOM RT Plan
• DICOM Key Object Selection Document
• DICOM RT Dose
• DICOM Raw (Brainlab Load/Save)
• DICOM Segmentation
• DICOM Registrations
• PDF Report
• ASCII Dose Export | • DICOM Images
• DICOM RT Struct
• DICOM RT Plan
• DICOM Key Object Selection Document
• DICOM RT Dose
• DICOM Raw (Brainlab Load/Save)
• DICOM Segmentations
• DICOM Registrations
• PDF Report
• ASCII Dose Export
• csv files |
| Supported Collimators | • Multileaf Collimators
• Circular Conical Collimators | • Multileaf Collimators
• Circular Conical Collimators |
| GUI Technology | • HTML5 (only RT Preparation & RT Analysis)
• WPF | • HTML5
• WPF (only Physics Administration) |
PAGE 5 OF 12
{10}
BRAINLAB
| Topic/ Feature | Predicate Device (RT Elements 4.0 - K223279) | Subject Device (RT Elements 4.5) |
| --- | --- | --- |
| Supported Image Modalities | • CT
• MRI
• PET | • CT
• MRI
• PET
Furthermore, it is possible to load the output of the Contrast Clearance Element which is labeled as OT (other) Image set.
*Background:* The Contrast Clearance Analysis calculation compares two 3D MRI data sets with injected contrast agents that were taken 60 to 105 minutes apart (early and late scans). When comparing the two 3D MRI data sets, Contrast Clearance Analysis performs a voxel-by-voxel subtraction of the early MR images from the late MR images resulting in a high-resolution color-coded Treatment Response Assessment Map (TRAM). This map provides effective separation between regions of contrast accumulation (positive value in red), and regions of contrast clearance (negative values in blue). Red regions show contrast accumulation, i.e. non-tumor regions and predominantly vessel necrosis. Dark blue regions show efficient contrast clearance, i.e. predominantly morphologically active tumor regions. |
| Dose Calculation Algorithm | • Pencil Beam Algorithm
• Monte Carlo Algorithm
• Circular Cone Algorithm | • Pencil Beam (PB) Algorithm
• Monte Carlo (MC) Algorithm
• Circular Cone Algorithm |
PAGE 6 OF 12
{11}
BRAINLAB
| Topic/ Feature | Predicate Device (RT Elements 4.0 - K223279) | Subject Device (RT Elements 4.5) |
| --- | --- | --- |
| Performance Characteristics | Pencil Beam/Monte Carlo: better than 3%
Circular Cone: 1%/1mm | Pencil Beam/Monte Carlo: better than 3%
Circular Cone: 1%/1mm |
| Object Changes | In the planning applications it is possible to add and remove review objects to/from a treatment plan.
In general object property changes are not possible during treatment planning. | Review objects can be added/removed in all planning applications as well as Dose Planning.
PTV (Planning Target Volume), OAR (Organ At Risk) and other objects' properties can be changed for treatment plans (e.g., name, color and comment). The optimization result remains valid after the change. |
| Separation of optimization and final dose calculation | The final dose calculation was automatically performed at the end of the optimization. | The user can trigger the final dose calculation separately from the plan optimization (VMAT (Volumetric Modulated Arc Therapy)):
→ Possible to optimize a plan using PB or MC first and then trigger final dose calculation using fine MC parameters
→ MC coarse and MC fine dose is cached of the current state and the MC dose of the last optimization result to enable the user to easily revert to the last plan state
→ The fine and coarse MC parameters visualized in the dropdown can be defined in the Clinical Protocol. The spatial resolution and grid size used during a MC based optimization can only be changed in the Clinical Protocol |
PAGE 7 OF 12
{12}
BRAINLAB
| Topic/ Feature | Predicate Device (RT Elements 4.0 - K223279) | Subject Device (RT Elements 4.5) |
| --- | --- | --- |
| Interface to Art-Plan (by TheraPanacea) for extracranial workflows | Not part of the Predicate Device. | For the following extracranial workflows an interface to the cloud based 3^{rd} party contouring system “Art-Plan” by TheraPanacea (TPC) was introduced.
Art-Plan provides an auto-segmentation of extracranial structures which are loaded into Object Management. Therefore, Art-Plan Connect (part of Art-Plan) is installed on the same HW as the RT Elements and works as a communication gateway between the main Art Plan cloud service and the RT Elements. |
| APM (Anatomical Patient Model) | Not part of the Predicate Device. | Component of the RT Elements with a gRPC API, which serves as an algorithmic backend for medical image data processing. The API can readily be used by client applications to have access to functionalities for automatic processing of 2D and 3D medical image data (e.g. segmentations, landmarks). To provide its functionality the APM service also includes interfaces to other components of the Universal Patient Model device like UniversalAtlasPerformer and UATransferPerformer. Additional functionality is also implemented directly in the APM service (like cranial tumor segmentations). The component is used by RT Contouring. |
PAGE 8 OF 12
{13}
BRAINLAB
## 4. Performance Data
The Subject Device has been verified and validated according to Brainlab processes for product design and development. A high level explanation of the testing provided in this submission for the Subject Device is provided below. For more details, please refer to the document Verification and Performance Bench Testing Summary – RT Elements 4.5.
### Software verification
Software verification has been performed on software level, verifying the software requirements through integration tests (see XToolPro Software protocols) as well as GUI tests and Unit tests. An incremental test strategy has been set up for the release candidates for changes with limited scope. In this case, an impact analysis of the modifications was performed and tests to be performed are identified and planned correspondingly.
### Usability Evaluation:
Only formative usability evaluation was carried out for the Subject Device. No summative usability evaluation has been performed for RT Elements 4.5.
### Performance Comparison Optimization Algorithm:
This test compared the performance of the new version to the predecessor version of Multiple Brain Mets SRS. The objective was to show that the dosimetric performance between the two versions is equivalent without compromising in treatment plan efficiency.
### Comparison Of Dose Calculations To Phantom Measurements:
This verifies that RT Elements 4.5 is able to accurately compute dose distributions for treatment machines equipped with Elekta or Varian multi-leaf collimators with both dose calculation algorithms: Pencil Beam (PB) and Monte Carlo (MC).
### Validation Of Reference Beam Models:
Reference Beam Models (RBM) contain Pencil Beam (PB) and Monte Carlo (MC) dose profiles that are pre-generated for different photon source sizes of a linear accelerator (linac). The test validates the dose calculations performed for Elekta and Varian linacs using 6 MV-STD, 6 MV-FFF, 10 MV-STD and 10 MV-FFF photon beams with respect to the accuracy of RBMB for PB and MC dose algorithms.
PAGE 9 OF 12
{14}
BRAINLAB
Small Field Dose Calculation With Scatter Model Enhanced Pencil Beam:
This test verifies that RT Elements is able to accurately calculate doses for targets with an equivalent square field size smaller than $10 \times 10 \, \text{mm}^2$ using its Scatter Model enhanced pencil beam algorithm.
Multiple Brain Mets SRS Calculation Time:
Multiple Brain Mets SRS enables fast automatic treatment plan generation for multiple brain metastases cases. The optimization problem is solved using an inverse stochastic optimization algorithm and requires several dose calculations. This test assesses the time to complete treatment plan generation for cases with up to 15 metastases.
Varian Dynamic Jaw Tracking:
Jaws in general are used to block radiation around the opening of the multi-leaf-collimator (MLC). With Static Jaws, the smallest possible box around all MLC positions is formed by the jaws. Dynamic Jaws enable a reshaping of the box at every control point. It is expected that this will result in reduced radiation of normal tissue as those boxes can be shaped more precisely around the positions of the leaves. The aim of this test is to confirm that Varian Dynamic Jaws result in reduced normal tissue dose compared to a treatment plan without Dynamic Jaws while maintaining similar planning target volume (PTV) conformity and coverage.
Elekta Dynamic Jaw Tracking:
Starting with RT Elements generation 2.5, Cranial SRS (1.5.1), Spine SRS (1.5.1) and Multiple Brain Mets SRS (2.0) support Dynamic Jaw Tracking for Elekta Agility machines. For those machines, the jaws are used to shape the form of irradiation targets in y direction, in addition to the leaves in x direction. With Dynamic Jaw Tracking, the jaws can be moved to arbitrary positions. Without Dynamic Jaw tracking, the jaws are limited to discrete positions at the leave edges. The aim of this test was to find out, whether Dynamic Jaw Tracking results in reduced normal tissue dose compared to the same plan without jaw tracking.
Collimator Optimization Spine SRS:
This test shows whether Collimator Optimization results in improved conformity and reduced normal tissue dose compared to plan generation without Collimator Optimization. The test was restricted to non-spherical targets for which a rotation of the collimator can have an effect.
AI/ML Performance Tests for Contouring 4.5:
Elements AI Tumor Segmentation can be used to semi-automatically segment supported cranial tumors (metastases, meningiomas, cranial and paraspinal nerve tumors, gliomas, glioneuronal
PAGE 10 OF 12
{15}
BRAINLAB
and neuronal tumors) on 3D medical Contrast Enhanced T1-weighted MR images powered by the Anatomical Patient Model. The approach used was quantitative validation - comparison to ground-truth annotations.
# Validation Summary:
Elements AI Tumor Segmentation algorithm was tested on de-identified 3D CE-T1 MR images from 412 patients (595 scans, 1878 annotations) from multiple clinical sites in the US and Europe. All data was acquired from adult patients with one or multiple contrast-enhancing tumors. The pool contained homogenous distribution by gender (female/male) and a diversity of ethnicity groups being White/Black/Latino/Asian with the highest representation. Most of data were obtained in patients who underwent stereotactic radiosurgery using a diverse acquisition of MR protocols (mainly containing 1.5T/3T MRI scans acquired in axial scan orientation). The test data pools also included control patients without cranial tumors. $\frac{1}{4}$ of the test pool corresponded to data from three independent sites in USA. The validation was conducted quantitatively by comparing the (manual) ground-truth segmentations, the so-called annotations with the respective automatically-created segmentations. The annotations involved external/independent annotator team including US radiologists and non US radiologists which followed a well-defined data curation process.
In accordance with best practices, measurement of recall and precision and object-specific DICE coefficient were performed to assess the similarity between both segmentations. Acceptance criteria were Dice $\geq 0.7$ , Recall $\geq 0.8$ and Precision $\geq 0.8$ for the lower bound of the respective $95\%$ confidence intervals. Successful validation has been completed based on images containing up to 30 cranial metastases, each showing a diameter of at least $3\mathrm{mm}$ , and images with primary cranial tumors that are at least $10\mathrm{mm}$ in diameter (for meningioma, cranial/paraspinal nerve tumors, gliomas, glioneuronal and neuronal tumors). See performance metrics in Table 1.
Table 1 Summary of test statistics
| Diagnostic Characteristics | Mean Dice | 95% Confidence Interval | Mean Precision | 95% Confidence Interval | Mean Recall | 95% Confidence Interval |
| --- | --- | --- | --- | --- | --- | --- |
| All | 0.75 | [0.74; 0.76] | 0.86 | [0.85; 0.88] | 0.85 | [0.83; 0.87] |
| Metastases to the CNS | 0.74 | [0.73; 0.75] | 0.85 | [0.83; 0.87] | 0.84 | [0.82; 0.86] |
{16}
BRAINLAB
| Meningiomas | 0.76 | [0.73; 0.79] | 0.89 | 0.84; 0.94] | 0.90 | [0.85; 0.94] |
| --- | --- | --- | --- | --- | --- | --- |
| Cranial and paraspinal nerve tumors | 0.89 | [0.88; 0.90] | 0.97 | [0.93; 1.0] | 0.97 | [0.93; 1.0] |
| Gliomas and glio-/neuronal tumors | 0.81 | [0.76; 0.86] | 0.95 | [0.88; 1.0] | 0.85 | [0.74; 0.94] |
# 5. Conclusion
The comparison of the Subject Device with the predicate device shows that RT Elements 4.5 has similar functionality, intended use and technological characteristics as the predicate device. Based on the comparison to the predicate and the performance testing conducted, the Subject Device is considered substantially equivalent to the predicate device.
PAGE 12 OF 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.