K250435 · Distalmotion SA · SDD · May 27, 2025 · General, Plastic Surgery
Device Facts
Record ID
K250435
Device Name
Dexter L6 System
Applicant
Distalmotion SA
Product Code
SDD · General, Plastic Surgery
Decision Date
May 27, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Distalmotion Dexter L6 System is intended to assist in the accurate control of endoscopes as well as endoscopic instruments for endoscopic manipulation of tissue, including grasping, suturing, dissecting, coagulating and cutting, with or without high frequency functionality. The Distalmotion Dexter L6 System is intended for use in laparoscopic inguinal hernia repair and cholecystectomy. The system is indicated for adult use, defined as 22 years old and older. It is intended for use by trained laparoscopic or robotic surgeons in an operating room environment in accordance with the representative and specific procedures set forth in the Instructions for Use.
Device Story
Electromechanical surgical system enabling minimally invasive surgery; comprises surgeon console, patient carts, and robotic instruments. Surgeon operates via handle grips and foot pedals to control endoscope and instruments (needle holder, graspers, dissectors, scissors, hooks). System translates surgeon inputs into precise instrument movements for tissue manipulation (grasping, suturing, dissecting, coagulating, cutting). Used in operating rooms by trained surgeons. Provides robotic assistance for laparoscopic procedures; enhances surgical precision and control; facilitates complex maneuvers in hernia repair and cholecystectomy.
Clinical Evidence
Prospective, multicenter, open-label NEST study (n=51) evaluated Dexter-assisted cholecystectomy. Primary endpoint: successful procedure completion without conversion to open/laparoscopic surgery achieved in 98% (50/51) of subjects (95% CI: 89.6%, 100%). Safety profile: 4% serious adverse event rate (2/51), none device-related. No reoperations required. Results demonstrate safety and effectiveness for cholecystectomy.
Technological Characteristics
Electromechanical surgical system; includes surgeon console, patient carts, and sterile/reusable instruments. Compatible with standard 10mm+ laparoscopic trocars. Connectivity: integrated firmware-based control system. Sterilization: single-use sterile accessories/instruments. No changes to hardware/software from predicate.
Indications for Use
Indicated for adult patients (≥22 years) undergoing laparoscopic inguinal hernia repair or cholecystectomy. Intended for use by trained laparoscopic or robotic surgeons in an operating room.
Regulatory Classification
Identification
An electromechanical surgical system with transient sterile field presence of both surgeon and primary control interface is a software-controlled electromechanical system with fully positionable patient/device interfaces which allows a qualified user to perform surgical techniques during minimally invasive surgical procedures. The device allows for both the surgeon and primary control interface to have presence inside the sterile field as needed within the clinical context of use.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 27, 2025
Distalmotion SA
% Lina Kontos
Regulatory Counsel
Hogan Lovells US LLP
Columbia Square
555 Thirteenth Street, NW
Washington, District of Columbia 20004
Re: K250435
Trade/Device Name: Dexter L6 System
Regulation Number: 21 CFR 878.4965
Regulation Name: Electromechanical surgical system with transient sterile field presence of both surgeon and primary control interface
Regulatory Class: Class II
Product Code: SDD
Dated: February 14, 2025
Received: April 17, 2025
Dear Lina Kontos:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K250435 - Lina Kontos
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{3}
K250435 - Lina Kontos
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark Trumbore -5
Digitally signed by Mark Trumbore
Date: 2025.05.27 13:16:59
04'00"
Mark Trumbore Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250435
Device Name
Dexter L6 System
Indications for Use (Describe)
The Distalmotion Dexter L6 System is intended to assist in the accurate control of endoscopes as well as endoscopic instruments for endoscopic manipulation of tissue, including grasping, suturing, dissecting, coagulating and cutting, with or without high frequency functionality. The Distalmotion Dexter L6 System is intended for use in laparoscopic inguinal hernia repair and cholecystectomy. The system is indicated for adult use, defined as 22 years old and older. It is intended for use by trained laparoscopic or robotic surgeons in an operating room environment in accordance with the representative and specific procedures set forth in the Instructions for Use.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}
D
Distalmotion
K250435
# 510(K) SUMMARY
# DISTALMOTION DEXTER L6 SYSTEM
Submitter: Distalmotion SA
Route de la Corniche 3b
1066 Epalinges
Switzerland
Contact: Larry Carrier
VP Regulatory Affairs & Quality Assurance
Phone: +1-707-239-6086
Email: larry.carrier@distalmotion.com
Date Prepared: May 23, 2025
Trade/Device Name: Dexter L6 System
Regulation Number: 21 CFR 878.4965
Regulation Name: Electromechanical surgical system with transient sterile field presence of both surgeon and primary control interface
Regulatory Class: Class II
Product Code: SDD
Review Panel: General & Plastic Surgery
Predicate Devices: Dexter L6 System (DEN230084)
# Device Description:
The Dexter L6 System is designed to enable complex surgery using a minimally invasive approach. It is composed of the Robot, the single-use accessories, fully articulated instruments, as well as reusable accessories. The Dexter L6 System is compatible with standard single-use laparoscopic trocars measuring 10 [mm] in diameter and larger.
The Robot consists of a Surgeon Console, with which the surgeon controls the movement of the instruments and of the Endoscope Arm using two Handle Grips (reusable), a Clutching Foot Pedal and an Endoscope Foot Pedal; two Patient Carts positioned at the operating room table in which the instruments are inserted and removed through the Hub during surgery; and the Dexter L6 Software installed in the Robot firmware.
Page 1 of 4
{6}
K250435
Page 2 of 4
The single-use, sterile instruments consist of the Needle Holder, Bipolar Johann Grasper, Bipolar Maryland Dissector, Monopolar Scissors, and Monopolar Hook.
The single-use accessories consist of the Sterile Interface, Endoscope Arm and Sterile Drapes.
The reusable accessories consist of the Accessory tray, Incision Pointer, Emergency Release Tool and Handle Grips.
# Intended Use / Indications for Use:
The Distalmotion Dexter L6 System is intended to assist in the accurate control of endoscopes as well as endoscopic instruments for endoscopic manipulation of tissue, including grasping, suturing, dissecting, coagulating and cutting, with or without high frequency functionality. The Distalmotion Dexter L6 System is intended for use in laparoscopic inguinal hernia repair and cholecystectomy. The system is indicated for adult use, defined as 22 years old and older. It is intended for use by trained laparoscopic or robotic surgeons in an operating room environment in accordance with the representative and specific procedures set forth in the Instructions for Use.
The Dexter L6 System is for prescription use only.
The indications for use statement are consistent with the previously cleared Dexter L6 system, except for the additional of the term "suturing" to the statement and the addition of use of the system in performing cholecystectomy procedures.
# Summary of Technological Characteristics:
The subject device has the same technological characteristics as the predicate device, the Dexter L6 System (DEN230084). The labeling has been changed to address use of the device in cholecystectomy procedures, to clarify the contraindications and that the device can be used for suturing.
# Performance Data:
There have been no changes to the device since the previous clearance (DEN230084), other than the modification to expand the indications for use to include cholecystectomy indication. Extensive bench testing was conducted on the previously cleared Dexter L6 System, and these data remain applicable to support the safety and effectiveness of the subject device.
{7}
Distalmotion
K250435
# Clinical Data:
The NEST (DEXTER-assisted Non-Emergent and acute cholecystectomy) Study was a prospective, multicenter, open-label, clinical investigation to confirm the perioperative and early postoperative safety and effectiveness of the Dexter Robotic System (Dexter) in patients undergoing cholecystectomy.
The NEST Study included 51 subjects who were enrolled and on whom robotic-assisted surgery with Dexter was at least started (mITT population). Subject data was collected at baseline, intraoperatively, discharge, and postoperatively at 30 days.
The mean age of the mITT population was 55.1 years. Female subjects accounted for $62.7\%$ (32/51). Most subjects were classified as ASA II status $84.3\%$ (43/51). A demographic and comorbidity comparison of the NEST cohort and the EU and US literature is included in the table below.
| Demographic/ Comorbidity | NEST (n=51) | EU Literature | US Literature | Discussion |
| --- | --- | --- | --- | --- |
| Age (years) | 55.1 ± 14.7 | 42.1-80 | 44.3-61.1 | NEST cohort in range |
| Gender | 64.2% (Female) | 71-76.6 (Female) | 55-77% Female | NEST cohort in range |
| Ethnicity | 58.8% (White) 2% (African American) 3.9% (Other) 35.3% (Unknown or Not Reported) | Not reported | 53.7-89.2 (White) 4-14.7% (African American) 0.4-21.2 (Hispanic) | NEST cohort in range for most common White ethnicity NOTE: Due to patient privacy regulations in France, the reported ethnicity of 35.3% (Unknown/ Not Reported) also includes additional subjects with White ethnicity. |
| BMI (kg/m2) | 27.7 | 23.4-29 | 28.8-35.6 | NEST cohort similar to US population. See Obesity comorbidity below. |
| ASA Status | 92% (ASA I/II) 7.8% (ASA III+) | 63 - 97% (ASA I/II) 3-37% (ASA III+) | 54.4 - 80.2% (ASA I/II) 19.7-54.6% (ASA III+) | NEST cohort higher proportion of ASA I/II EU population and US population in similar range |
| Nassar Difficulty Scale Rating | 21.6% (I) 45.1% (II) 27.5% (III) 5.9% (IV) | Not reported | Not reported | Nassar Scale only reported for NEST cohort |
| Obesity (>30 kg/m2) | 21.6% (n=11/51) | Not reported | 5.7-54.2% | NEST cohort in range of US population |
| Hypertension | 25.5% (13/51) | Not reported | 27-59.7% | NEST cohort similar to US population |
| COPD | Not reported | 8.4% | 2.05-8.8% | EU population in range of US population |
| Cardiac Disease | 23.5% (N=12/51)* | 6.4% | 0.2-17.3 | EU population in range of US population |
| Diabetes | 13.7 (7/51) | 4.3-20.1% | 9.2-24.4% | NEST cohort in range of US population |
| Hypothyroidism | 5.8% (3.51) | Not reported | Not reported | Hypothyroidism only reported for NEST cohort |
| Smoking | Not reported | 9-35% | 13.1-28.9% | EU population in range of US population |
The operation was successfully performed in 50 out of 51 subjects of the mITT population using robotic-assisted surgery (98.0%, 50/51), (95% CI:89.6%, 100.0%).
The primary performance endpoint, defined as successful completion of the Dexter-assisted procedure without conversion to an open or fully laparoscopic surgical approach, was confirmed in 50 of 51 procedures from the mITT group (98.0%). One subject was converted to laparoscopic mode due to time constraints in which the surgeon decided to complete the procedure laparoscopically to shorten the operative time.
For the primary safety endpoint, there was one postoperative (Clavien-Dindo grades III-V) adverse
Page 3 of 4
{8}
K250435
event perioperatively up to 30 days. There were 10 adverse events and two serious adverse events in the perioperative phase to 30 days after the procedure.
Two serious adverse events were reported (4.0%):
- Post-operative agranulocytosis and pancytopenia as an adverse reaction to metamizole therapy. The event was resolved at the time of study completion and classified as Clavien-Dindo II, not related to the procedure and not related to the device.
- Re-hospitalization due to the migration of a pre-existing choledocholithiasis. The event was resolved at the time of study completion and classified as Clavien-Dindo III, not related to the procedure and not related to the device.
Two intraoperative adverse events were reported (4.0%):
- Gallbladder perforation; the event was treated with prophylactic antibiotics. The event was resolved at the time of study completion and was classified as Clavien-Dindo II, procedure related and not related to the device.
- Injury of the posterior wall of the gallbladder; the event was treated with prophylactic antibiotics. The event was resolved at the time of study completion and was classified as Clavien-Dindo II, procedure related and not related to the device.
None of the subjects required reoperation after the index procedure. In conclusion, the results of the NEST Study demonstrate the Dexter L6 System is safe and effective for subjects undergoing cholecystectomy.
## Conclusions:
The subject Dexter L6 System is as safe and effective as the predicate device. The subject Dexter L6 System has the same intended uses and similar indications, same technological characteristics, and principles of operation as its predicate device. The expanded specific indications do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. Analysis and clinical performance data of the device with cholecystectomy procedures demonstrate that the Dexter L6 System is as safe and effective as the predicate device. Thus, the subject Dexter L6 System is substantially equivalent.
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.