AtriCure's cryoICE® cryoXT™ cryoablation probes are intended for use to temporarily block pain by ablating peripheral nerves performed by freezing target tissues, creating an inflammatory response (cryonecrosis).
Device Story
The cryoICE cryoXT is a sterile, single-use cryoablation probe designed for peripheral nerve ablation. It features a malleable shaft and a blunt, prong-shaped distal tip. The device connects to an AtriCure cryoICE BOX (ACM) console, which circulates nitrous oxide through the probe. Operation relies on the Joule-Thomson effect to freeze target tissues, creating cryonecrosis to temporarily block pain. The device is used by surgeons in a clinical setting; the surgeon manually positions the prong-shaped tip to surround the exposed peripheral nerve. The console controls the gas flow, initiated by the surgeon via a footswitch or handle button. By creating an inflammatory response, the device interrupts nerve conduction, providing a temporary analgesic effect for the patient.
Clinical Evidence
Bench testing only. No clinical data was provided. Performance was validated through cryogenic performance, iceball formation, mechanical reliability, pressure cycle, drop, dimensional, biocompatibility, and sterility testing. All tests met predetermined acceptance criteria.
Indicated for temporary pain block via peripheral nerve ablation in patients requiring cryonecrosis of target tissues.
Regulatory Classification
Identification
A cryogenic surgical device is a device used to destroy nervous tissue or produce lesions in nervous tissue by the application of extreme cold to the selected site.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 10, 2025
AtriCure, Inc.
Ronit Shah
Regulatory Affairs Specialist
7555 Innovation Way
Mason, Ohio 45040
Re: K250371
Trade/Device Name: cryoICE cryoXT cryoablation probe (cryoXT)
Regulation Number: 21 CFR 882.4250
Regulation Name: Cryogenic Surgical Device
Regulatory Class: Class II
Product Code: GXH
Dated: February 10, 2025
Received: February 10, 2025
Dear Ronit Shah:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250371 - Ronit Shah
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250371 - Ronit Shah
Page 3
Sincerely,
Adam D. Pierce -S
Digitally signed by
Adam D. Pierce -S
Date: 2025.04.10
13:41:24 -04'00'
Adam D. Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250371
Device Name
cryoICE® cryoXT™ cryoablation probe (cryoXT)
Indications for Use (Describe)
AtriCure’s cryoICE cryoXT cryoablation probes are intended for use to temporarily block pain by ablating peripheral nerves performed by freezing target tissues, creating an inflammatory response (cryonecrosis).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
K250371
# 510(k) Summary
## Applicant Information
Manufacturer: AtriCure, Inc.
7555 Innovation Way
Mason Ohio 45040
P: 513-755-4100
Contact Person: Ronit Shah
Regulatory Affairs Specialist
P: 409-434-7267
Date: April 10, 2025
## Device Information
Proprietary Name: cryoICE® cryoXT™ cryoablation probe
Common Name: Cryosurgical Probe
Classification: Cryogenic surgical device
Regulatory Class: Class II; per 21 CFR 882.4250
Product Code: GXH
Classification Panel: Neurodiagnostic and Neurosurgical Devices
Predicate Device: cryoICE® cryoablation probe (CRYO2) (cleared via 510(k) K200697, March 17, 2020)
Reference Device: cryoICE® cryoSPHERE® cryoablation probe (CRYOS) (cleared via 510(k) K200697, March 17, 2020)
## Device Description
The cryoICE® cryoXT™ probe is a sterile, single-use device designed to enable cryoablation of target nerves by allowing surgeons to surround exposed peripheral nerves with its blunt prong-shaped distal tip, and in conjunction with an AtriCure cryoICE BOX (ACM), temporarily freeze the tissue in contact with the probe tip by circulating a cryogenic agent, nitrous oxide, through the device. cryoXT is offered in one single probe length of approximately 11" (28 cm) long. The device has a flexible tubeset that consists of a gas supply tube and exhaust tube, a handle, a malleable probe which is covered by an insulating sheath and has a prong-shaped distal tip at the distal end.
## Intended Use / Indications for Use
AtriCure's cryoICE® cryoXT™ cryoablation probes are intended for use to temporarily block pain by ablating peripheral nerves performed by freezing target tissues, creating an inflammatory response (cryonecrosis).
Page 1 of 3
{5}
K250371
# Comparison of Technological Characteristics
- The devices have the same intended use, and;
- No changes were made in operating principle, or specifications of performance, and;
- The results of the verification and validation testing:
- Demonstrated equivalency in performance, and
- Did not raise any new issues of safety.
The cryoICE® cryoXT™ cryoablation probe's blunt prong-shaped distal tip is being introduced based on physicians' feedback to provide a distal tip design to gently and efficiently surround the targeted exposed peripheral nerves.
| # | Feature | Predicate Device – CRYO2 per K200697 | Reference Device – CRYOS per K200697 | Proposed Device – cryoXT | Equivalence |
| --- | --- | --- | --- | --- | --- |
| 1 | Marketed Product Name | cryoICE cryoablation probes | cryoICE cryoSPHERE cryoablation probe | cryoICE cryoXT cryoablation probe | N/A |
| 2 | Intended Use | FOR ADULT PATIENTS
AtriCure's Cryo2 cryoICE cryoablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway.
The Cryo2 cryoICE cryo-ablation probes are also intended for use to temporarily block pain by ablating peripheral nerves.
FOR ADOLESCENT PATIENTS
The Cryo2 cryoICE cryo-ablation probes are intended for use to temporarily block pain by ablating intercostal nerves under direct visualization¹ in adolescent patients of at least 12 years of age.
¹Direct visualization, in this context, requires that the surgeon is able to see the targeted tissue for cryoablation directly or with assistance from a camera, endoscope or other similar optical technology. | FOR ADULT PATIENTS
AtriCure's cryoICE cryoSPHERE cryoablation probes are sterile, single use devices intended for use performed by freezing target tissues, creating an inflammatory response (cryonecrosis) for blocking pain by temporarily ablating peripheral nerves.
FOR ADOLESCENT PATIENTS
The cryoICE cryoSPHERE cryo-ablation probes are intended for use to temporarily block pain by ablating intercostal nerves under direct visualization¹ in adolescent patients of at least 12 years of age.
¹Direct visualization, in this context, requires that the surgeon is able to see the targeted tissue for cryoablation directly or with assistance from a camera, endoscope or other similar optical technology. | AtriCure's cryoICE cryoXT cryoablation probes are intended for use to temporarily block pain by ablating peripheral nerves performed by freezing target tissues, creating an inflammatory response (cryonecrosis). | Equivalent
The pain block indication statement is the same as the predicate device.
The cryoXT is not intended for cardiac arrhythmias or adolescents.
Note: The 'what' and 'how' of the indications for use was rearranged to make the sentence flow better. The 'what', blocking pain, was moved to read before the 'how', by ablating peripheral nerves performed by freezing target tissues, creating an inflammatory response (cryonecrosis). The key aspects all remain within the indications for use between the predicate and proposed devices.
Elimination of 'sterile, single use devices' as this information is identified within multiple areas of the Instructions For Use and is not a part of the device's indication or intended use. |
| 3 | Operating Principle | Joule-Thompson Effect | Joule-Thompson Effect | Joule-Thompson Effect | Same |
| 4 | Technology | The system consists of cryoprobes that are used for freezing target tissue. A console is used to control the supply of gas to the cryoprobe. | The system consists of cryoprobes that are used for freezing target tissue. A console is used to control the supply of gas to the cryoprobe. | The system consists of cryoprobes that are used for freezing target tissue. A console is used to control the supply of gas to the cryoprobe. | Same |
| 5 | Energy Used | Nitrous Oxide | Nitrous Oxide | Nitrous Oxide | Same |
Page 2 of 3
{6}
K250371
| # | Feature | Predicate Device – CRYO2 per K200697 | Reference Device – CRYOS per K200697 | Proposed Device – cryoXT | Equivalence |
| --- | --- | --- | --- | --- | --- |
| 6 | Operating Temperature | Below -40°C | Below -40°C | Below -40°C | Same |
| 7 | Human Factors | Hand-held device connected to a console which circulates the cryogen through the device in a closed loop system via the activation button or footswitch. | Hand-held device connected to a console which circulates the cryogen through the device in a closed loop system via the activation button or footswitch. | Hand-held device connected to a console which circulates the cryogen through the device in a closed loop system via the activation button or footswitch. | Same |
| 7 | Distal Tip | Linear, cylindrical probe, Aluminum | Ball-shaped probe, Aluminum | Linear, prong-shaped probe with cylindrical arms, Aluminum | Equivalent |
| 8 | Exposed Shaft Length | 28cm (11") | 28cm (11") | 28cm (11") | Same |
| 9 | Biocompatibility | Biocompatible patient contacting materials | Biocompatible patient contacting materials | Biocompatible patient contacting materials | Same |
| 10 | Packaging | Sterile - single use, disposable cryoablation probe. | Sterile - single use, disposable cryoablation probe. | Sterile - single use, disposable cryoablation probe. | Same |
| 11 | Sterilization | Gamma Irradiation | Gamma Irradiation | Gamma Irradiation | Same |
| 12 | Forming Tool | Forming Tool Supplied | Forming Tool Supplied | No Forming tool will be supplied | Equivalent |
## Performance Data
The following bench testing was conducted for design and performance elements deemed appropriate to demonstrate equivalence to the previously cleared cryoICE cryoablation probes (CRYO2, CRYOS). The cryoICE® cryoXT™ cryoablation probe met the predetermined acceptance criteria ensuring substantial equivalence to the previously cleared cryoICE cryoablation probes (CRYO2, CRYOS). No new safety or performance issues were raised during testing.
## Non-clinical Bench Testing:
- Cryogenic Performance Test
- Iceball Performance Test
- Mechanical Reliability Test
- Pressure Cycle Withstand Test
- Drop Test
- Dimensional Verification Test
- Biocompatibility Test
- Sterility Test
## Conclusions
AtriCure has demonstrated that the cryoICE® cryoXT™ cryoablation probe is substantially equivalent in fundamental design, technology, function, device materials, packaging, sterilization, operating principle, and intended use / indication for use as the previously cleared cryoICE cryoablation probes (CRYO2) and cryoICE cryoSPHERE cryoablation probes (CRYOS) cleared via K200697.
Page 3 of 3
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.