BD Phoenix™ Automated Microbiology System

K250344 · Becton, Dickinson and Company · LON · Mar 6, 2025 · Microbiology

Device Facts

Record IDK250344
Device NameBD Phoenix™ Automated Microbiology System
ApplicantBecton, Dickinson and Company
Product CodeLON · Microbiology
Decision DateMar 6, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 866.1645
Device ClassClass 2

Indications for Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria. The BD Phoenix™ System provides rapid results for most aerobic and facultative anaerobic Gram-positive bacteria as well as most aerobic and facultative anaerobic Gramnegative bacteria of human origin. The BD Phoenix System is also intended for the rapid identification of yeast and yeast-like organisms.

Device Story

BD Phoenix M50 Automated Microbiology System performs antimicrobial susceptibility testing. Modification introduces BDXpert 2.0, enabling connectivity to BD Synapsys Informatics Solution and codifying previously custom user-defined rules into the system. System operates in clinical microbiology laboratories; used by laboratory personnel to automate susceptibility reporting. Output provides antimicrobial susceptibility results to clinicians for therapeutic decision-making. Modifications are limited to the M50 instrument configuration.

Clinical Evidence

Bench testing only. Software verification and validation activities conducted in accordance with ISO 13485:2016, IEC 62304:2015, and ISO 14971:2019 demonstrate that the BD Phoenix instrument connected to Synapsys performs as intended.

Technological Characteristics

Automated microbiology system for antimicrobial susceptibility testing. Connectivity via BD Synapsys Informatics Solution. Software-based rule engine (BDXpert 2.0) for automated interpretation of susceptibility data. System architecture supports integration with laboratory informatics.

Indications for Use

Indicated for in vitro rapid identification and quantitative antimicrobial susceptibility determination (MIC) of Gram-positive bacteria (Staphylococcus, Enterococcus, other cocci, bacilli) and Gram-negative aerobic/facultative anaerobic bacteria (Enterobacteriaceae, Non-Enterobacteriaceae) from pure culture. For prescription use only.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION # SPECIAL 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K250344 B Applicant Becton, Dickinson and Company C Proprietary and Established Names BD Phoenix Automated Microbiology System D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | LON | Class II | 21 CFR 866.1645 - Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System | MI - Microbiology | ## II Review Summary: This 510(k) submission contains information/data on modifications made to the submitter's own CLASS II device requiring a 510(k). The following items are present and acceptable: 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. 2. Submitter's statement that the INDICATIONS FOR USE/INTENDED USE of the modified device as described in its labeling HAS CHANGED along with the proposed labeling which includes user manuals. Labeling changes are considered minor and do not affect the intended use/indications for use of the original or modified device. The changes in the Indications for Use statement of the modified device (K250344) aim to: a) Apply minor clarifying changes to the text to simplify the language and remove redundancies. Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {2} 3. A description of the device **MODIFICATION(S)**, including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the **FUNDAMENTAL SCIENTIFIC TECHNOLOGY** of the modified device **has not changed**. Changes were made to the BDXpert System to: a) Allow for connectivity to the BD Synapsys Informatics Solution and b) Add utility to BDXpert that codifies functionality previously created by customers using custom rules. The new configuration of BDXpert is referred to as BDXpert 2.0, which is a separate product to BDXpert on standalone Phoenix and on EpiCenter (referred to as Legacy BDXpert). The device modifications under K250344 are only applicable to the BD Phoenix M50 Automated Microbiology System Instrument. The changes in the labeling of the modified device (K250344) aim to provide the End User/Customer with a set of updated operational documents (i.e., BDXpert User’s Manual, BD Phoenix M50 Instrument User’s Manual, and Synapsys User’s Manual). 4. Comparison Information (i.e., similarities and differences) to the submitter's legally marketed predicate device including, labeling, intended use, and physical characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared device. K250344 - Page 2 of 2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...