The NEUROVENT is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP). The NEUROVENT IFD-S is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP). The NEUROVENT IFD-R is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP). The NEUROVENT-P is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP). The NEUROVENT-PX is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP). The NEUROVENT-P-TEMP is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature. The NEUROVENT-TEMP is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature. The NEUROVENT-TEMP IFD-S is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature. The NEUROVENT-TEMP-IFD-R is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature. The NEUROVENT-PTO is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature. Additional measurement of the oxygen partial pressure is an adjunct monitor of trends indicating the perfusion status of cerebral tissue local to sensor placement. The measured values are relative within an individual and should not be used as the sole basis for decisions as to diagnosis or therapy. The NEUROVENT-PTO-2L is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature. Additional measurement of the oxygen partial pressure is an adjunct monitor of trends indicating the perfusion status of cerebral tissue local to sensor placement. The measured values are relative within an individual and should not be used as the sole basis for decisions as to diagnosis or therapy. The BOLT(-DRILL) KITs are indicated to provide a cranial access for RAUMEDIC neurosurgical precision pressure catheters of the RAUMEDIC NEUROMONITORING-SYSTEM. The DRILL KITs are indicated to provide a cranial access for RAUMEDIC neurosurgical precision pressure catheters of the RAUMEDIC NEUROMONITORING-SYSTEM. The Tunneling KITs are indicated to provide a cranial access for catheters of the RAUMEDIC NEUROMONITORING-SYSTEM.
Device Story
System comprises probes/catheters for intracranial pressure (ICP) monitoring, CSF drainage, brain temperature, and partial tissue oxygen pressure (ptiO2). Inputs: physiological signals from brain tissue/ventricles via sensors (ICP, thermistors, fiber optics). Operation: catheters implanted via bolt/drill/tunneling kits; connected to external monitors (EASY logO, MPR2, NPS3/NPS2 X) for signal processing and display. Output: real-time pressure, temperature, and oxygen trends. Used in clinical/neurosurgical settings by physicians. Output aids clinical decision-making regarding cerebral perfusion and intracranial status. MR conditional status allows use under specific constraints. Benefits: enables localized monitoring of cerebral parameters to guide therapy.
Clinical Evidence
Bench testing only. MR safety and compatibility verified per ASTM F2052-15, ASTM F2213-06, ASTM F2119-07, and ISO/TS 10974:2018 (gradient/RF-induced heating/malfunction, combined fields). No clinical data presented.
Technological Characteristics
Catheters/probes with ICP sensors, thermistors, and fiber optic sensors for ptiO2. Materials substantially equivalent to predicates. MR conditional (1.5T/3.0T). Connectivity via external dataloggers/monitors. Sterilization method not specified.
Indications for Use
Indicated for patients requiring intracranial pressure (ICP) monitoring, cerebrospinal fluid (CSF) drainage, brain temperature monitoring, or cerebral tissue oxygen partial pressure monitoring. Used in neurosurgical settings for parenchymal or ventricular access.
Regulatory Classification
Identification
An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 29, 2025
Raumedic Ag
Hannes Engelhardt
Head of Regulatory Affairs
Hermann-Staudinger-Straße 2
Helmbrechts, Bavaria 95233
Germany
Re: K250285
Trade/Device Name: NEUROVENT-P (092946-003); NEUROVENT (092956-003); NEUROVENT-IFD-R (095317-003); NEUROVENT-IFD-S (091678-003); NEUROVENT-P-TEMP (094268-003); NEUROVENT-TEMP (094278-003); NEUROVENT-TEMP-IFD-R (095327-003); NEUROVENT-TEMP-IFD-S (094288-003); NEUROVENT-PTO (095008-003); NEUROVENT-PTO 2L (095108-003); NEUROVENT-PX (091580-003); BOLT-DRILL KIT CH5 (091888-003); BOLT-DRILL KIT CH9 (091898-003); BOLT-DRILL KIT PTO (092380-003); BOLT-DRILL KIT VP 16 (092969-003); BOLT KIT CH5 (091868-003); BOLT KIT CH9 (091688-003); BOLT KIT PTO (096026-003); DRILL KIT CH5 (091878-002); DRILL KIT CH9 (091668-002); Tunneling KIT CH8 (090516-001); Tunneling KIT CH12 (090727-001)
Regulation Number: 21 CFR 882.1620
Regulation Name: Intracranial pressure monitoring device
Regulatory Class: Class II
Product Code: GWM
Dated: July 31, 2025
Received: July 31, 2025
Dear Hannes Engelhardt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250285 - Hannes Engelhardt
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K250285 - Hannes Engelhardt
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce -S
Digitally signed by
Adam D. Pierce -S
Date: 2025.08.29
13:52:37 -04'00'
Adam D. Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250285
Device Name
NEUROVENT-P
NEUROVENT
NEUROVENT-IFD-R
NEUROVENT-IFD-S
NEUROVENT-P-TEMP
NEUROVENT-TEMP
NEUROVENT-TEMP-IFD-R
NEUROVENT-TEMP-IFD-S
NEUROVENT-PTO
NEUROVENT-PTO 2L
NEUROVENT-PX
BOLT-DRILL KIT CH5
BOLT-DRILL KIT CH9
BOLT-DRILL KIT PTO
BOLT KIT CH5
BOLT KIT CH9
BOLT KIT PTO
DRILL KIT CH5
DRILL KIT CH9
Tunneling KIT CH8
Tunneling KIT CH12
Indications for Use (Describe)
The NEUROVENT is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP).
The NEUROVENT IFD-S is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP).
The NEUROVENT IFD-R is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP).
The NEUROVENT-P is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP).
The NEUROVENT-PX is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP).
The NEUROVENT-P-TEMP is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature.
The NEUROVENT-TEMP is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral
{4}
tissue temperature.
The NEUROVENT-TEMP IFD-S is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature.
The NEUROVENT-TEMP-IFD-R is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature.
The NEUROVENT-PTO is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature. Additional measurement of the oxygen partial pressure is an adjunct monitor of trends indicating the perfusion status of cerebral tissue local to sensor placement. The measured values are relative within an individual and should not be used as the sole basis for decisions as to diagnosis or therapy.
The NEUROVENT-PTO-2L is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature. Additional measurement of the oxygen partial pressure is an adjunct monitor of trends indicating the perfusion status of cerebral tissue local to sensor placement. The measured values are relative within an individual and should not be used as the sole basis for decisions as to diagnosis or therapy.
The BOLT(-DRILL) KITs are indicated to provide a cranial access for RAUMEDIC neurosurgical precision pressure catheters of the RAUMEDIC NEUROMONITORING-SYSTEM.
The DRILL KITs are indicated to provide a cranial access for RAUMEDIC neurosurgical precision pressure catheters of the RAUMEDIC NEUROMONITORING-SYSTEM.
The Tunneling KITs are indicated to provide a cranial access for catheters of the RAUMEDIC NEUROMONITORING-SYSTEM.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
---
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}
510(k) #: K250285
510(k) Summary
Prepared on: 2025-08-28
# Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | RAUMEDIC AG |
| --- | --- |
| Applicant Address | Hermann-Staudinger-Straße 2 Helmbrechts Bavaria 95233 Germany |
| Applicant Contact Telephone | 004992523591530 |
| Applicant Contact | Dr. Hannes Engelhardt |
| Applicant Contact Email | hannes.engelhardt@raumedic.com |
# Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | NEUROVENT-P
NEUROVENT
NEUROVENT-IFD-R
NEUROVENT-IFD-S
NEUROVENT-P-TEMP
NEUROVENT-TEMP
NEUROVENT-TEMP-IFD-R
NEUROVENT-TEMP-IFD-S
NEUROVENT-PTO
NEUROVENT-PTO 2L
NEUROVENT-PX
BOLT-DRILL KIT CH5
BOLT-DRILL KIT CH9
BOLT-DRILL KIT PTO
BOLT KIT CH5
BOLT KIT CH9
BOLT KIT PTO
DRILL KIT CH5
DRILL KIT CH9
Tunneling KIT CH8
Tunneling KIT CH12 |
| --- | --- |
| Common Name | Intracranial pressure monitoring device |
| Classification Name | Device, Monitoring, Intracranial Pressure |
| Regulation Number | 882.1620 |
| Product Code(s) | GWM |
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K103206 | RAUMEDIC® ICP-MONITORING-SYSTEM | GWM |
| K112017 | RAUMEDIC® ICP-MONITORING-SYSTEM ventricular | GWM |
{6}
K120252
K130529
K153347
RAUMEDIC® ICP-TEMP-MONITORING-SYSTEM
RAUMEDIC® PTO-MONITORING-SYSTEM
Codman® Codman Microsensor Basic Kit
GWM
GWM
GWM
# Device Description Summary
21 CFR 807.92(a)(4)
The RAUMEDIC® NEUROMONITORING-SYSTEM consists of several different models of probes and probe catheters capable of performing one or several different functions::
- Models with a dedicated lumen can be used for drainage of cerebrospinal fluid (CSF).
- Models equipped with ICP sensors can determine the level and change in intracranial pressure (ICP).
- Models equipped with temperature thermistors can monitor intracranial temperature.
- Models equipped with fiber optic sensors can monitor partial tissue oxygen pressure (ptiO2).
The RAUMEDIC® NEUROMONITORING-SYSTEM is intended to be used in conjunction with previously cleared RAUMEDIC® EASY logO Monitor (K130529), RAUMEDIC® MPR2 logO DATALOGGER (K171666), RAUMEDIC® NPS3 (K103206) or RAUMEDIC® NPS2 X (Brand name for NPS2 cleared in K103206).
The RAUMEDIC® NEUROMONITORING-SYSTEM includes components needed to facilitate the surgical implantation of NEUROVENT® catheters.
The RAUMEDIC® NEUROMONITORING-SYSTEM can be used in MR environment under specific constraints (MR conditional). Those constraints vary by device type, implantation method (bolting or tunneling), and magnetic field strength (1.5 or 3.0 Tesla).
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The NEUROVENT is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP).
The NEUROVENT IFD-S is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP).
The NEUROVENT IFD-R is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP).
The NEUROVENT-P is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP).
The NEUROVENT-PX is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP).
The NEUROVENT-P-TEMP is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature.
The NEUROVENT-TEMP is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature.
The NEUROVENT-TEMP IFD-S is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature.
The NEUROVENT-TEMP-IFD-R is indicated for use in ventricular pressure monitoring and cerebrospinal fluid drainage applications. It can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature.
{7}
The NEUROVENT-PTO is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature. Additional measurement of the oxygen partial pressure is an adjunct monitor of trends indicating the perfusion status of cerebral tissue local to sensor placement. The measured values are relative within an individual and should not be used as the sole basis for decisions as to diagnosis or therapy.
The NEUROVENT-PTO-2L is indicated for use in parenchymal pressure monitoring and can be used for the measurement of the intracranial pressure (ICP). Additional measurement of the brain temperature allows the direct measurement of the cerebral tissue temperature. Additional measurement of the oxygen partial pressure is an adjunct monitor of trends indicating the perfusion status of cerebral tissue local to sensor placement. The measured values are relative within an individual and should not be used as the sole basis for decisions as to diagnosis or therapy.
The BOLT(-DRILL) KITs are indicated to provide a cranial access for RAUMEDIC neurosurgical precision pressure catheters of the RAUMEDIC NEUROMONITORING-SYSTEM.
The DRILL KITs are indicated to provide a cranial access for RAUMEDIC neurosurgical precision pressure catheters of the RAUMEDIC NEUROMONITORING-SYSTEM.
The Tunneling KITs are indicated to provide a cranial access for catheters of the RAUMEDIC NEUROMONITORING-SYSTEM.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The RAUMEDIC® NEUROMONITORING-SYSTEM is substantially equivalent to the legally marketed predicate devices: RAUMEDIC®-ICP-MONITORING-SYSTEM (K103206), RAUMEDIC®-ICP-Monitoring-System ventricular (K112017), RAUMEDIC®-ICP-TEMP-MONITORING-SYSTEM (K120252), RAUMEDIC®-PTO-Monitoring-System (K130529), and Codman Microsensor Basic Kits (K153347).
Based on performance testing and the available information concerning the referenced comparison devices, the RAUMEDIC® NEUROMONITORING-SYSTEM is equivalent in that:
- The devices have the same intended use and indication for use.
- Performance characteristics are suitable for designated indications for use.
Based on this, the anticipated clinical performance of the RAUMEDIC® NEUROMONITORING-SYSTEM is substantially equivalent to the referenced systems and does not raise new issues of safety or effectiveness.
## Technological Comparison
21 CFR 807.92(a)(6)
The RAUMEDIC® NEUROMONITORING-SYSTEM is substantially equivalent to the legally marketed predicate devices: RAUMEDIC®-ICP-MONITORING-SYSTEM (K103206), RAUMEDIC®-ICP-Monitoring-System ventricular (K112017), RAUMEDIC®-ICP-TEMP-MONITORING-SYSTEM (K120252), RAUMEDIC®-PTO-Monitoring-System (K130529), and Codman Microsensor Basic Kits (K153347).
Based on performance testing and the available information concerning the referenced comparison devices, the RAUMEDIC® NEUROMONITORING-SYSTEM is equivalent in that:
- The devices are made of the same materials or substantially equivalent materials.
- The devices have equivalent form, function, procedures and features.
- Performance characteristics are suitable for designated indications for use.
Based on this, the anticipated clinical performance of the RAUMEDIC® NEUROMONITORING-SYSTEM is substantially equivalent to the referenced systems and does not raise new issues of safety or effectiveness.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
MR safety requirements of the RAUMEDIC® NEUROMONITORING-SYSTEM have been tested and confirmed as MR conditional by the
{8}
following tests:
- Magnetically induced displacement force (ASTM F 2052-15)
- Magnetically induced torque (ASTM F2213-06)
- Image artifacts (ASTM F2119-07)
- Gradient-induced heating (ISO/TS 10974:2018, Clause 9)
- Gradient-induced malfunction (ISO/TS 10974:2018, Clause 16)
- RF-induced heating (ISO/TS 10974:2018, Clause 8, Tier 3)
- RF-induced malfunction (ISO/TS 10974:2018, Clause 15)
- Combined Fields (ISO/TS 10974:2018, Clause 17)
Based on this, the anticipated clinical performance of the RAUMEDIC® NEUROMONITORING-SYSTEM is substantially equivalent to the referenced systems and does not raise new issues of safety or effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.