K250263 · Olympus Medical Systems Corp. · KTI · Mar 31, 2025 · Ear, Nose, Throat
Device Facts
Record ID
K250263
Device Name
Disposable Grasping Forceps FG-52D/FG-54D
Applicant
Olympus Medical Systems Corp.
Product Code
KTI · Ear, Nose, Throat
Decision Date
Mar 31, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Disposable Grasping Forceps FG-52D/54D are intended to be used to retrieve foreign bodies, calculus or tissue specimens from the tracheobronchial tree in combination with an endoscope.
Device Story
Disposable grasping forceps (FG-52D spiral basket; FG-54D three-nail type) designed for endoscopic retrieval of foreign bodies, calculus, or tissue from tracheobronchial tree. Device consists of proximal handle with ring mechanism and distal stainless steel grasping portion. Operated by clinician via compatible endoscope. User manipulates handle to open/close distal grasping tip. Mechanical retrieval of objects facilitates airway clearance. Provided sterile (EtO), single-use.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including biocompatibility (ISO 10993), sterilization validation (ISO 11135), packaging/shelf-life (ISO 11607, ASTM F1980), and mechanical performance testing (insertion/withdrawal force, handle operation, grasping performance, connection strength, and device reliability).
Technological Characteristics
Materials: Stainless steel grasping portion. Energy: Mechanical (manual). Form factor: Insertion portion diameter 0.8mm (FG-52D) or 1.05mm (FG-54D); working length 1150mm. Connectivity: None. Sterilization: Ethylene Oxide (EtO).
Indications for Use
Indicated for retrieval of foreign bodies, calculus, or tissue specimens from the tracheobronchial tree in patients requiring endoscopic intervention.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
Microvasive Zero Tip Airway Retrieval Basket, Models M00513200 and M00513210 (K020765)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 31, 2025
Olympus Medical Systems Corp. % Roshana Ahmed Regulatory Affairs Consultant Olympus Corporation of the Americas 800 West Park Drive Westborough, Massachusetts 01581
Re: K250263
Trade/Device Name: Disposable Grasping Forceps FG-52D/FG-54D Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: KTI Dated: January 28, 2025 Received: January 29, 2025
Dear Roshana Ahmed:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ethan L. Nyberg -S
Ethan Nyberg, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K250263
Device Name Disposable Grasping Forceps FG-52D/54D
### Indications for Use (Describe)
The Disposable Grasping Forceps FG-52D/54D are intended to be used to retrieve foreign bodies, calculus or tissue specimens from the tracheobronchial tree in combination with an endoscope.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are all capitalized and evenly spaced. There is a thin, blurred, yellow line underneath the word. There is a registered trademark symbol to the right of the letter "S".
# 510(k) Summary
#### I. Submitter
Olympus Medical Systems Corporation 2951 Ishikawa-cho Hachioji-shi Tokyo 192-8507 Japan
Contact Person: Seiko Yunoki Phone: +81 42-642-2111
Date Prepared: March 28, 2025
#### II. Device
| Device Proprietary Name | Disposable Grasping Forceps FG-52D/FG-54D |
|-------------------------|--------------------------------------------------|
| Common or Usual Name | Grasping Forceps |
| Classification Name | Bronchoscope (flexible or rigid) and accessories |
| Regulation Number | 874.4680 |
| Product Code | KTI |
| Device Classification | Class II |
#### III. Predicate Device
Substantial equivalence is claimed to the following device:
- Microvasive Zero Tip Airway Retrieval Basket, Models M00513200 and ● M00513210, K020765, Boston Scientific Corp.
#### Device Description IV.
The Disposable Grasping Forceps FG-52D/FG-54D are designed to retrieve foreign bodies, calculus, or tissue specimens endoscopically within the tracheobronchial tree when used in conjunction with a compatible endoscope.
{6}------------------------------------------------
Each device is provided sterile and consists of a handle and an insertion portion. The handle (proximal portion) of the Disposable Grasping Forceps includes a ring which can be pushed or pulled to open or close the grasping portion of the device. The distal end of the insertion portion consists of the grasping portion of the device. The grasping portion, manufactured from stainless steel, is provided in two shapes: spiral basket type (model FG-52D) and three nail type (model FG-54D) and has limited contact with the patient. The user will select the device model depending on procedural needs.
#### V. Indications for Use
The Disposable Grasping Forceps FG-52D/54D are intended to be used to retrieve foreign bodies, calculus or tissue specimens from the tracheobronchial tree in combination with an endoscope.
#### VI. Comparison of Technological Characteristics
There are slight differences between the subject device and predicate device indications for use statements; however, these differences do not create a new intended use with respect to the subject device. The subject and predicate devices are both used in conjunction with scopes to remove materials from the tracheobronchial tree.
The subject device and predicate devices are sterile (via EtO), single use devices which are used with compatible endoscopes. A comparison of the subject and predicate device technological characteristics is provided below.
| | Disposable Grasping Forceps<br>FG-52D/FG-54D | Microvasive® Zero Tip™<br>Airway Retrieval Basket | Analysis |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Manufacturer | Olympus Medical Systems Corp. | Boston Scientific Corp. | N/A |
| Indications for<br>Use Statement | The Disposable Grasping Forceps<br>FG-52D/54D are intended to be<br>used to retrieve foreign bodies,<br>calculus or tissue specimens from<br>the tracheobronchial tree in<br>combination with an endoscope. | The Microvasive® Zero Tip™<br>Airway Retrieval Basket is used<br>to access the airway tree via a<br>bronchoscope for the purpose of<br>removing foreign bodies. | Substantially<br>equivalent |
| Models | FG-52D<br>FG-54D | M00513200 | N/A |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly stylized, with a modern and clean font. There is a thin, horizontal, yellow line underneath the word, adding a subtle accent to the logo. The registered trademark symbol is present to the right of the word.
| | Disposable Grasping Forceps<br>FG-52D/FG-54D | Microvasive® Zero Tip™<br>Airway Retrieval Basket | Analysis |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|-----------------------------|
| Maximum<br>Insertion<br>Portion<br>Diameter | FG-52D: 0.8 mm<br>FG-54D: 1.05 mm | M00513200: 0.8 mm | Substantially<br>equivalent |
| Working<br>Length | 1150 mm | 1200 mm | Substantially<br>equivalent |
| Opening<br>Width | FG-52D: 8 mm<br>FG-54D: 10 mm | M00513200: 12 mm | Substantially<br>equivalent |
| Shape of the<br>Grasping<br>Portion | FG-52D: Spiral Basket Type<br>FG-54D: Three Nail Type | M00513200: Basket Type | Substantially<br>equivalent |
| Compatible<br>Endoscopes | Model and length: working length<br>less than 700 mm (BF)<br>Channel inner diameter: Ø 1.2,<br>2.0, 2.2, 2.6, 2.8, 3.2 | Unknown | Substantially<br>equivalent |
| Sterilization | Ethylene Oxide | Ethylene Oxide | Identical |
| Single-Use/<br>Reusable | Single Use | Single Use | Identical |
The subject and predicate devices differ as follows:
- . The subject device (model FG-54D) has a larger maximum insertion portion diameter than the predicate device.
- The working length of the subject device is slightly shorter than the working length . of the predicate device.
- . The opening width of the subject device is smaller than the opening width of the predicate device.
- . The shape of the grasping portion of the subject device (model FG-54D) is different than the grasping portion of the predicate device.
- . The subject devices are compatible with endoscopes featuring a working length of 700 mm (BF) and a channel inner diameter ranging from 1.2 - 3.2.
The differences in insertion portion diameter, working length, opening width, shape of the grasping portion, and compatible endoscopes do not raise different questions of safety or effectiveness when comparing the subject device to the predicate device. These differences are addressed through bench testing of the subject and predicate device.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly shadowed, giving them a three-dimensional effect. A thin, horizontal gold line is underneath the word. The registered trademark symbol is to the right of the S.
#### Performance Data VII.
The following performance data were provided to demonstrate substantial equivalence:
- Biocompatibility testing per ISO 10993-1:2018 including cytotoxicity (ISO 10993-. 5:2009), sensitization (ISO 10993-10:2021), irritation (ISO 10993-23:2021), acute systemic toxicity (ISO 10993-11:2017), and material mediated pyrogenicity (USP <151>)
- Sterilization validation per ISO 11135:2014 .
- . Ethylene oxide residuals per ISO 10993-7:2008
- . Packaging validation and shelf life testing in accordance with ISO 11607-1:2019 and ASTM F1980-16
- . Mechanical testing and comparative testing to verify device performance:
- o Insertion force/ Withdrawal force
- Handle Operation o
- Grasping Performance о
- Strength Testing of the Connection and Distal Tip O
- Device Reliability O
- Human Factors Testing o
Clinical data is not required to demonstrate substantial equivalence.
# VIII. Conclusion
The non-clinical data demonstrate that the subject device is as safe, as effective, and performs as well as or better than the identified predicate device. Therefore, it is concluded that the Disposable Grasping Forceps FG-52D/FG-54D are substantially equivalent to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.