K250233 · Measure Labs, Inc. (Dba Preemptiveai, Inc.) · DXH · Feb 13, 2026 · Cardiovascular
Device Facts
Record ID
K250233
Device Name
PreemptiveAI Clinical SDK
Applicant
Measure Labs, Inc. (Dba Preemptiveai, Inc.)
Product Code
DXH · Cardiovascular
Decision Date
Feb 13, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2920
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Pulse rate
Deep learning model with autocorrelation-based signal quality check
± 3 BPM ARMS
± 3 BPM ARMS
—
—
Clinical study: 111 participants
—
Indications for Use
The PreemptiveAI Clinical SDK is intended for noninvasive spot measurement of pulse rate. The measurement is based on video recording of the subject's fingertip using a smartphone camera. It is designed for use when the subject is still. It is indicated for individuals 18 years of age or older not requiring critical care or continuous vital signs monitoring. It is not intended for use in patients with known or suspected heart arrhythmias.
Device Story
Software as a Medical Device (SaMD) SDK for integration into third-party mobile apps; operates on Android/iOS. Captures 30-second fingertip video via smartphone back camera; extracts photoplethysmography (PPG) waveform. Proprietary locked deep learning algorithm processes signal in 10-second overlapping windows; performs autocorrelation-based signal quality check. Outputs single pulse rate value (average of passing windows) or 'Low Signal Quality' prompt. Used for noninvasive spot-check pulse rate measurement in non-critical care environments. Not for arrhythmia detection or alarm systems. Healthcare providers use output for pulse rate monitoring; device benefits patients by enabling convenient, non-invasive spot checks.
Clinical Evidence
Prospective clinical study of 111 participants (diverse age, sex, race, skin tone). Compared SDK pulse rate to reference device. Primary endpoint: Average Root Mean Square (ARMS) difference. Results confirmed SDK measures pulse rate within ±3 BPM ARMS. Sub-group analysis confirmed performance across demographics and skin tones (Fitzpatrick I-VI).
Technological Characteristics
Reflectance photoplethysmography (PPG) using smartphone camera/flash. Software-only SDK for Android/iOS. Locked deep learning model for pulse rate calculation. Operational range: 50-125 bpm. 30-second measurement window. Cybersecurity measures included.
Indications for Use
Indicated for individuals 18 years of age or older not requiring critical care or continuous vital signs monitoring. Not intended for use in patients with known or suspected heart arrhythmias.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 13, 2026
Measure Labs, Inc. (DBA PreemptiveAI, Inc.)
Matt Whitehill
Chief Technology Officer
2790 Mosside Blvd.
Monroeville, Pennsylvania 15146
Re: K250233
Trade/Device Name: PreemptiveAI Clinical SDK
Regulation Number: 21 CFR 870.2920
Regulation Name: Telephone Electrocardiograph Transmitter And Receiver
Regulatory Class: Class II
Product Code: DXH
Dated: February 10, 2026
Received: February 11, 2026
Dear Matt Whitehill:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250233 - Matt Whitehill
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250233 - Matt Whitehill
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JENNIFER W. SHIH -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250233
Device Name
PreemptiveAI Clinical SDK
Indications for Use (Describe)
The PreemptiveAI Clinical SDK is intended for noninvasive spot measurement of pulse rate. The measurement is based on video recording of the subject's fingertip using a smartphone camera. It is designed for use when the subject is still. It is indicated for individuals 18 years of age or older not requiring critical care or continuous vital signs monitoring. It is not intended for use in patients with known or suspected heart arrhythmias.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
PreemptiveAI
PreemptiveAI Clinical SDK Traditional 510(k) Summary
# 510(k) Summary
Company Address: Measure Labs, Inc. (DBA PreemptiveAI, Inc.)
2801 Alaskan Way, Suite 302, Seattle, WA 98121-1134
Primary Contact Person: Matt Whitehill
Chief Technology Officer
Date Prepared: February 10, 2026
Device Name: PreemptiveAI Clinical SDK
Classification: 21 CFR 870.2920 Telephone electrocardiograph transmitter and receiver
Product Code: DXH
Regulatory Class: II
Predicate Device: Fibricheck (K232804)
Indications for Use: The PreemptiveAI Clinical SDK is intended for noninvasive spot measurement of pulse rate. The measurement is based on video recording of the subject's fingertip using a smartphone camera. It is designed for use when the subject is still. It is indicated for individuals 18 years of age or older not requiring critical care or continuous vital signs monitoring. It is not intended for use in patients with known or suspected heart arrhythmias.
Device Description: The PreemptiveAI Clinical SDK is a Software as a Medical Device (SaMD) that utilizes a smartphone to noninvasively measure pulse rate (PR). Designed for integration into third-party mobile applications, the SDK operates on Android and iOS devices, enabling spot-check pulse rate measurements.
The PreemptiveAI Clinical SDK captures a 30-second fingertip video with the smartphone's back camera. The software extracts a photoplethysmography (PPG) waveform, partitions that 30-second
K250233
{5}
PreemptiveAI
PreemptiveAI Clinical SDK Traditional 510(k) Summary
signal into overlapping 10-second windows, and applies an autocorrelation-based signal quality check. A proprietary deep learning algorithm then calculates pulse rate for every window that passes the quality screen. The value shown to the user is the average of all passing windows, and if every window fails the screen, no rate is reported and a "Low Signal Quality" prompt is displayed.
The SDK outputs a single pulse-rate value; it is not intended for automated analysis or arrhythmia detection. The PreemptiveAI Clinical SDK is not intended for use with alarm systems.
# Technological Characteristics
The PreemptiveAI Clinical SDK uses photoplethysmography (PPG) technology to capture and analyze blood flow signals. The SDK comprises a mobile SDK for integration into a 3rd-party app, backend server, and locked deep learning model for pulse rate prediction.
Substantial Equivalence Comparison Summary:
| | Subject Device (K250233) | Predicate Device (K232804) |
| --- | --- | --- |
| Manufacturer | PreemptiveAI | Fibricheck |
| Device Name | PreemptiveAI Clinical SDK | Fibricheck |
| Indications for Use | The PreemptiveAI Clinical SDK is intended for noninvasive spot measurement of pulse rate. The measurement is based on video recording of the subject's fingertip using a smartphone camera. It is designed for use when the subject is still. It is indicated for individuals 18 years of age or older not requiring critical care or continuous vital signs monitoring. It is not intended for use in patients with known or suspected heart arrhythmias. | Fibricheck is indicated for self-testing by patients who have been diagnosed with, or are susceptible to developing, atrial fibrillation and who would like to monitor and record their heart rhythms on an intermittent basis. |
| Regulation names and numbers | 21 CFR 870.2920 Telephone electrocardiograph transmitter and receiver. | 21 CFR 870.2920 Telephone electrocardiograph transmitter and receiver. |
| Product codes and classifications | Code: DXH Class: II | Code: DXH Class: II |
K250233
{6}
PreemptiveAI
PreemptiveAI Clinical SDK
Traditional 510(k) Summary
| User Population | Adults 18 years or older not requiring critical care or continuous vital signs monitoring. | Adults. Patients, who have been diagnosed with, or are susceptible to developing, atrial fibrillation. |
| --- | --- | --- |
| Rx / OTC | Rx | Rx |
| Component | Software Only | Software Only |
| Measurement Type | Spot Check | Spot Check |
| Measurement Site | Fingertip | Fingertip |
| Single Patient Use | Yes | Yes |
| Outputs | Pulse Rate | Pulse Rate
Irregular Heart Rhythm Notification |
| Device Description | PreemptiveAI Clinical SDK uses reflectance photoplethysmography (PPG) to estimate a user’s pulse rate through the existing camera hardware on a smartphone. Specifically, the SDK analyzes optical signals captured from the fingertip, leveraging the phone’s camera and flash to detect changes in blood volume associated with each cardiac cycle. The measured pulse rate is displayed as a periodic spot check, enabling users to track their pulse rate in non-critical care environments. | FibriCheck is medical device software that determines heart rhythm conditions, with a primary focus on the detection of Atrial Fibrillation. It makes use of optical sensing from a mobile device to collect photoplethysmogram data (PPG data). The recordings can be shared optionally with a physician or monitoring service. |
| Technology Characteristics | The PreemptiveAI Clinical SDK measures pulse rate through an optical camera-based mechanism. It is intended to run on Android and iOS mobile devices, capturing video of the fingertip to compute the blood volume signal, then utilizing a locked deep learning ML model to calculate pulse rate from this signal. | FibriCheck obtains waveform via Mobile Platform camera and displays signal in real time on the Mobile Platform device with an arrhythmia index. Text display (green, amber, or blue) indicates heart regularity or irregularity. All tests are stored. |
| Measurement Window | 30 seconds | 1 minute |
| Pulse Rate Range | 50-125 bpm | Unknown |
K250233
{7}
PreemptiveAI
PreemptiveAI Clinical SDK Traditional 510(k) Summary
# Non-Clinical Software testing:
- Verification and Validation - Verification and validation testing of the PreemptiveAI Clinical SDK was conducted to ensure the software meets the design requirements and intended use. This testing included a full characterization of software technical parameters and algorithms.
- Hardware Compatibility - Testing was performed on 4 different smartphone models during Clinical Testing, confirming the SDK functions appropriately across a range of devices.
- Human Factors - A human factors study was conducted to evaluate potential risks associated with usability. The testing included evaluation of the device's response to user errors and the user's reaction to controls and mitigations.
- Bench Testing - Bench testing confirmed the device is accurate across its claimed 50-125 bpm operational range and behaves in a predictable, fail-safe manner by suppressing readings outside these boundaries. Further bench testing also demonstrated appropriate algorithm behavior in the presence of heart arrhythmias.
- Cybersecurity Testing - Cybersecurity testing was conducted including vulnerability analysis. Documentation was provided to show adequate cybersecurity measures have been taken and will be monitored and updated throughout the device life cycle.
# Summary of Clinical Testing:
A clinical study was conducted to evaluate the performance of the PreemptiveAI Clinical SDK for noninvasive pulse rate measurement and to demonstrate its substantial equivalence to the predicate device. The study enrolled 111 participants, aged 18 years and older, representing a diverse demographic distribution across sex, age, race, and skin tone. Participants were recruited from a general population, ensuring applicability to a broad user base.
The pulse rate obtained by the PreemptiveAI Clinical SDK was compared to the heart rate obtained by the reference device. The Average Root Mean Square (ARMS) and its $95\%$ confidence interval were calculated. The hypothesis that the PreemptiveAI Clinical SDK can measure pulse rate within $\pm 3$ BPM ARMS is accepted. Sub-group accuracy was calculated to ensure adequate performance. Based on the results of the study, the PreemptiveAI Clinical SDK is effective at determining pulse rate in the intended use population.
# Demographic characteristics of the clinical testing dataset
| Description | | Number |
| --- | --- | --- |
| Subjects | Total | 111 |
| Age | 18-29 | 21 |
| | 30-39 | 29 |
| | 40-49 | 17 |
| Gender | Male | 17 |
| Education | High school or lower | 17 |
| | College or university | 17 |
| Education level | High school or lower | 17 |
| | College or university | 17 |
| Education level | High school or lower | 17 |
| | College or university | 17 |
| Age | 18-29 | 21 |
| | 30-39 | 29 |
| Gender | Male | 17 |
| | Male | 17 |
K250233
{8}
PreemptiveAI
PreemptiveAI Clinical SDK
Traditional 510(k) Summary
| | 50-59 | 16 |
| --- | --- | --- |
| | 60-69 | 11 |
| | 70+ | 17 |
| Sex | Male | 78 |
| | Female | 33 |
| Race / Ethnicity (participants could select more than 1 option) | American Indian / Alaskan Native | 4 |
| | Hispanic / Latino | 7 |
| | South Asian | 10 |
| | East Asian | 13 |
| | Black African/American | 20 |
| | White | 58 |
| | Other | 4 |
| | Prefer not to answer | 3 |
| Skin Tone (Fitzpatrick scale) | Type I and II | 64 |
| | Type III and IV | 28 |
| | Type V and VI | 19 |
| Arrhythmia | No Arrhythmia | 77 |
| | Arrhythmia | 34 |
| Atrial Fibrillation | No Atrial Fibrillation | 100 |
| | Atrial Fibrillation | 11 |
Conclusion:
The PreemptiveAI Clinical SDK has been found to be substantially equivalent to the predicate device with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence to the predicate device.
K250233
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.