Diode laser system (Night Universe; Predator)

K250206 · Ivylaser (Beijing) Technology Co.,Ltd · GEX · Apr 24, 2025 · General, Plastic Surgery

Device Facts

Record IDK250206
Device NameDiode laser system (Night Universe; Predator)
ApplicantIvylaser (Beijing) Technology Co.,Ltd
Product CodeGEX · General, Plastic Surgery
Decision DateApr 24, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

Diode laser system is used for hair removal, HR mode is for professional hair removal, and FHR is for rapid hair removal. The indications for use for the 810nm Diode Laser Module include: The HR mode and FHR mode are intended for hair removal and permanent hair reduction. Use on all skin types (Fitzpatrick I-VI) including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Device Story

Diode laser system (Night Universe; Predator) is a powered laser surgical instrument for hair removal. Device consists of a main unit (CPU, power supply, control/service panels, cooling system), a handlepiece (diode laser, TEC, cable/plug), and a foot switch. Operated by professionals in a clinical setting. Device emits 810nm laser energy to target hair follicles. HR mode (1-3Hz) and FHR mode (5-10Hz) provide varying energy densities (2-60 J/cm² and 2-10 J/cm² respectively) to achieve permanent hair reduction. Clinicians use the device to treat unwanted body hair across all Fitzpatrick skin types. Benefits include long-term, stable reduction in hair regrowth.

Clinical Evidence

Bench testing only. Compliance verified via IEC 60601-1, IEC 60601-1-2, IEC/TR 60601-4-2, IEC 60601-2-22, IEC 60825-1, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation/hypersensitivity).

Technological Characteristics

810nm diode laser; HR (1-3Hz) and FHR (5-10Hz) modes; 2cm² spot size; 5-200ms pulse width; 2-60 J/cm² energy density. Cooling: Water+air. Dimensions: Night Universe (350x450x1000mm), Predator (500x600x500mm). Weight: 33kg-60kg. Standards: IEC 60601-1, IEC 60601-1-2, IEC/TR 60601-4-2, IEC 60601-2-22, IEC 60825-1, ISO 10993-5, ISO 10993-10.

Indications for Use

Indicated for hair removal and permanent hair reduction in patients of all skin types (Fitzpatrick I-VI), including tanned skin.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION April 24, 2025 Ivylaser (Beijing) Technology Co., Ltd % Owen He Consultant Microkn Medical Technology Service (Shanghai) Co., Ltd. Room 901, No.889, Pinglu Road, Jing'an District Shanghai, China Re: K250206 Trade/Device Name: Diode laser system (Night Universe; Predator) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 7, 2024 Received: January 24, 2025 Dear Owen He: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250206 - Owen He Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250206 - Owen He Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, TANISHA L. HITHE -S Digitally signed by TANISHA L. HITHE -S Date: 2025.04.24 14:52:59 -04'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | Submission Number (if known) | | --- | | K250206 | | Device Name | | Diode laser system (Night Universe; Predator) | | Indications for Use (Describe) | | Diode laser system is used for hair removal, HR mode is for professional hair removal, and FHR is for rapid hair removal. The indications for use for the 810nm Diode Laser Module include: The HR mode and FHR mode are intended for hair removal and permanent hair reduction. Use on all skin types (Fitzpatrick I-VI) including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov > "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} | 510(k) #: K250206 | 510(k) Summary | Prepared on: 2025-01-24 | | --- | --- | --- | | Contact Details | | 21 CFR 807.92(a)(1) | | Applicant Name | Ivy laser (Beijing) Technology Co., Ltd | | | Applicant Address | Room 411, Building 3, No. 2 Shengfang Road, Daxing District, Beijing Beijing eijing China | | | Applicant Contact Telephone | 010-62258470 | | | Applicant Contact | Ms. Yuan Xiujuan | | | Applicant Contact Email | yuanxiujuan@oriental-laser.com | | | Correspondent Name | Microkn Medical Technology Service (Shanghai) Co., Ltd. | | | Correspondent Address | Room 901, No.889, Pinglu Road, Jing'an District, Shanghai Shanghai China | | | Correspondent Contact Telephone | +86 17521186001 | | | Correspondent Contact | Mr. Owen He | | | Correspondent Contact Email | fda@microkn.com | | | Device Name | | 21 CFR 807.92(a)(2) | | Device Trade Name | Diode laser system (Night Universe; Predator) | | | Common Name | Laser surgical instrument for use in general and plastic surgery and in dermatology | | | Classification Name | Powered Laser Surgical Instrument | | | Regulation Number | 878.4810 | | | Product Code(s) | GEX | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K222064 | The Alma Soprano Titanium | GEX, ILY | | K123483 | Diode Laser | GEX | | Device Description Summary | | 21 CFR 807.92(a)(4) | | The product is intended for the removal of unwanted body hair. It is suitable for all skin types, including tanned skin. The Diode laser system consists of the following main components. 1) The main unit, including the CPU main board, power supply module, control panel, service panel and cooling system. 2) Handlepiece, including the diode laser, TEC, handle switch, signal, cable and plug. 3) Foot switch. | | | {5} Intended Use/Indications for Use 21 CFR 807.92(a)(5) Diode laser system is used for hair removal, HR mode is for professional hair removal, and FHR is for rapid hair removal. The indications for use for the 810nm Diode Laser Module include: The HR mode and FHR mode are intended for hair removal and permanent hair reduction. Use on all skin types (Fitzpatrick I-VI) including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. Indications for Use Comparison 21 CFR 807.92(a)(5) The indication for use is the same with predicate devices. They are intended for the removal of unwanted body hair. Technological Comparison 21 CFR 807.92(a)(6) The device has the same technological characteristics as predicate devices in wavelength, energy density, operation model, spot size, laser type, pulse width and frequency. And there is a little difference in cooling ways, weight and size, while it is considered would not raise any issues in safety and effectiveness and they can be considered substantially equivalent in safety and effectiveness. We choose K222064 and KI23483 as the equivalent products. The device has the same technological characteristics as predicate devices in wavelength: 810nm, energy density HR: 2J/cm - 60J/cm, FHR: 2J/cm-10J/cm, operation model: HR and FHR, spot size: 2 cm2, laser type: Diode Laser, pulse width: 5ms - 200 ms, and frequency HR: 1-3Hz, FHR: 5-10Hz. And there is a little difference in cooling ways (Water+air), weight: 60Kg; 33kg and size: Night Universe: 350mmX450mmX1000mm and Predator: 500mmX600mmX500mm. However, such differences will not affect actual safety and effectiveness of this product, and we have also made a detailed analysis of this in the Substantial Equivalence Discussion in the document, Meanwhile, although there are slight differences in this product, However, all comply with the requirements of IEC 60601-1, IEC 60601-2-22, IEC 60601-1-2 and IEC/TR 60601-4-2. The proposed device is as safe and effective as a legally marketed predicate devices, and does not raise new safety or effectiveness issues and the proposed device is substantially identical to the predicate devices. Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: - IEC 60601-1:2005+A1:2012+A2:2020 Medical electrical device Part1: General requirements for basic safety and essential performance - IEC 60601-1-2:2014+A1:2020 Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests - IEC/TR 60601-4-2:2016 Medical electrical equipment - Par4.2: Guidance and interpretation - electromagnetic immunity: performance of medical electrical equipment and medical electrical systems - IEC 60601-2-22:2007+A1:2012 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment - IEC 60825-1:2014 Safety of laser products -Part 1: Equipment classification and requirements - ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity - ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity. The subject device does not raise different types of safety or effectiveness questions. The conclusions drawn from these testing, and previously conducted testing of the predicate and legacy devices, demonstrate that the subject device is as safe, as effective, and performed as well as the predicates. Thus, the subject device is substantially equivalent to its predicates.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%