← Product Code [FMF](/productcode/FMF) · K250192

# Verisafe Safety Retractable Insulin Syringes (K250192)

_Promisemed Hangzhou Meditech Co., Ltd. · FMF · Dec 10, 2025 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K250192

## Device Facts

- **Applicant:** Promisemed Hangzhou Meditech Co., Ltd.
- **Product Code:** [FMF](/productcode/FMF.md)
- **Decision Date:** Dec 10, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

Verisafe Safety Retractable Insulin Syringes are intended for aspiration and subcutaneous injection of insulin. After injection, the needle can be retracted directly from the patient's skin into the syringe barrel to avoid exposure to a contaminated needle, thereby minimizing accidental needle stick injuries.

## Device Story

Sterile, single-use safety retractable insulin syringe; consists of calibrated barrel, plunger, piston, push button, retraction cylinder, sealing ring, spring, and needle. Used in professional healthcare facilities, home, and transport environments. Operator (clinician or patient) fills syringe via plunger; performs subcutaneous injection. Post-injection, user continues pressing push-button to depress plunger fully; triggers internal safety mechanism. Spring force retracts needle tube, hub, and cylinder into barrel, shielding contaminated needle to prevent accidental sharps injury. Device incorporates color-coded needle caps (red for U40, orange for U100) per ISO 8537:2016. Benefits include reduced risk of needle stick injuries during insulin administration.

## Clinical Evidence

No clinical data. Bench testing only. Verification and validation included performance testing (freedom from defects, cleanliness, dimensions, stiffness, corrosion resistance, lubricant, needle point, patency, leakage, dead space, safety mechanism) and biocompatibility (cytotoxicity, skin irritation/sensitization, acute systemic toxicity, hemolysis, pyrogenicity, subchronic toxicity, USP <788> particulate matter).

## Technological Characteristics

Materials: plastic, stainless steel, silicone, PE (needle cap). Principle: manual plunger-driven fluid delivery with spring-loaded needle retraction. Dimensions: 0.5ml/1.0ml volumes; 28G-32G needles; 5mm-12mm lengths. Sterilization: EO (EN ISO 11135). Standards: ISO 9626, ISO 7864, ISO 8537, ISO 23908, ISO 10993-1. Single-use. No software/electronics.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- VeriSafe Safety sterile syringes ([K231792](/device/K231792.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

December 10, 2025

Promisemed Hangzhou Meditech Co., Ltd.
Zearou Yang
Regulatory Affairs Manager
No. 1388 Cangxing St.,
Cangqian Community, Yuhang District
Hangzhou City, 311121
China

Re: K250192
Trade/Device Name: Verisafe Safety Retractable Insulin Syringes
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston Syringe
Regulatory Class: Class II
Product Code: FMF, MEG
Dated: November 17, 2025
Received: November 17, 2025

Dear Zearou Yang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K250192 - Zearou Yang
Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K250192 - Zearou Yang
Page 3

Sincerely,

Shruti N. Mistry -S

Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K250192  |   |
|  Device Name Verisafe Safety Retractable Insulin Syringes  |   |
|  Indications for Use (Describe) Verisafe Safety Retractable Insulin Syringes are intended for aspiration and subcutaneous injection of insulin. After injection, the needle can be retracted directly from the patient's skin into the syringe barrel to avoid exposure to a contaminated needle, thereby minimizing accidental needle stick injuries.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

Promisemed

Promisemed Hangzhou Meditech Co., Ltd.

No. 1388 Cangxing Street, Cangqian Community, Yuhang District,

Hangzhou City, 311121 Zhejiang, P. R. China.

# K250192 510(k) Summary

Date prepared: 2025-12-10

|  1. Manufacturer[21 CFR 807.92 (a) (1)]  |   |
| --- | --- |
|  Name | Promisemed Hangzhou Meditech Co., Ltd.  |
|  Address | No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou City, 311121 Zhejiang, P.R.China.  |
|  Contact Person | Zearou Yang  |
|  Phone | +86 571 88772985  |
|  Fax | +86 571 88772985  |
|  Email | zearou.yang@promisemed.ca  |
|  2. Device [21 CFR 807.92 (a) (2)]  |   |
|  Name | Verisafe Safety Retractable Insulin Syringes  |
|  Common Name | Safety Retractable Insulin Syringes  |
|  Classification Name | Piston Syringe  |
|  Regulation Number | 21 CFR 880.5860  |
|  Class | Class II  |
|  Product Code | FMF, MEG  |
|  3. Legally Marketed Predicate Device[21 CFR 807.92 (a) (3)]  |   |
|  Predicate Name | VeriSafe Safety sterile syringes  |
|  510(k) Number | K231792  |
|  Product Code | FMF, MEG  |
|  Reference Devices | No reference devices were used in this submission.  |
|  4. Device Description [21 CFR 807.92 (a) (4)]  |   |
|  Safety Retractable Insulin Syringes is a sterile device consisting of a calibrated barrel, plunger, piston, push button, retraction cylinder, sealing ring, spring and a needle. It is made of plastic, stainless steel and silicone materials. Type 5 is a syringe with a fixed needle. The needle cap color is used to indicate insulin concentration that red for U40 and orange for U100. Safety Retractable Insulin Syringes is using plunger to fill syringe as well as discharge the fluid. After use, by continuing to press the push-button down to depress the plunger all of the way within the barrel, the safety mechanism will be triggered, then the needle tube, needle hub and retraction cylinder will be retracted into the barrel completely under the force of the spring to minimize the risk of accidental sharps injury. Product is delivered sterile. Sterilization process is validated according to EN ISO 11135. This is a single-use device. Safety Sterile Syringes is intended to be used in Professional Healthcare Facility,  |   |

Page 1 of 5

{5}

Promisemed

Promisemed Hangzhou Meditech Co., Ltd.

No. 1388 Cangxing Street, Cangqian Community, Yuhang District,

Hangzhou City, 311121 Zhejiang, P. R. China.

Home Environment and Transport (Ambulatory) Environment.

# 5. Indication for use [21 CFR 807.92 (a) (5)]

Verisafe Safety Retractable Insulin Syringes are intended for aspiration and subcutaneous injection of insulin. After injection, the needle can be retracted directly from the patient's skin into the syringe barrel to avoid exposure to a contaminated needle, thereby minimizing accidental needle stick injuries.

# 6. Indications for Use Comparison [21 CFR 807.92 (a) (5)]

Type 5 is added which its indication is only for insulin and is not affected the safety and effectiveness of the subjective device.

# 7. Comparison of Technological Characteristic [21 CFR 807.92 (a) (6)]

The subject device has the same technological characteristics as the predicate device. The similarities above between the subject device and predicate device do not affect the basic design principle, usage, effectiveness and safety of the subject device. And there are no new or different questions of effectiveness and safety when compared to the predicate device. We concluded the subject device is substantially equivalent to the identified predicate device.

At a high level, the subject and predicate devices are based on the following same technological elements:

|  Item of description | Predicate device (K231792) | Subject device (K250192) | Comments on Similarities/ Differences | Safety/Effectiveness Statement  |
| --- | --- | --- | --- | --- |
|  Common name | VeriSafe Safety sterile syringes | Safety Retractable Insulin Syringe | N/A | No new concerns. The common name was to be distinguished for difference indication.  |
|  Classification name | Piston Syringe (21 CFR 880.5860) | No change | N/A | No new concerns. The classification name remains consistent.  |
|  Device class | Class II | No change | N/A | No new concerns. The device remains in Class II, indicating no change in risk level.  |
|  Product Code | FMF, MEG | No change | N/A | No new concerns. The product code remains the same, reflecting no change in device functionality.  |
|  General description | Safety Sterile Syringes is a sterile device consisting of a calibrated barrel, plunger, piston, push button, retraction cylinder, sealing ring, spring and a needle. It is made of plastic, stainless steel and silicone materials. Safety Sterile Syringes have two types: Type F and type D. Type F is a syringe with a fixed needle. Type D is a syringe with a detachable needle. The detachable needle is secured by rotating it through the front needle hub into the back needle hub located on the syringe. Safety Sterile Syringes is using plunger to fill syringe as well as discharge the fluid. After use, by continuing to press the push-button down to depress the plunger all of the way within the barrel, the safety mechanism will be triggered, then the needle tube, needle hub and retraction cylinder will be retracted into the barrel completely under the force of the syringe to minimize the risk of accidental sharps injury. | Safety Sterile Syringes is a sterile device consisting of a calibrated barrel, plunger, piston, push button, retraction cylinder, sealing ring, spring and a needle. It is made of plastic, stainless steel and silicone materials. Type 5 is a syringe with a fixed needle. The needle cap color is used to indicate insulin concentration that red for U40 and orange for U100. Safety Sterile Syringes is using plunger to fill syringe as well as discharge the fluid. After use, by continuing to press the push-button down to depress the plunger all of the way within the barrel, the safety mechanism will be triggered, then the needle tube, needle hub and retraction cylinder will be retracted into the barrel completely under the force of the syringe to minimize the risk of accidental sharps injury. | Type 5 is added and needle cap color for insulin concentration was added in accordance with ISO 8537:2016. | No new concerns. Type 5 is added which its indication is only for insulin and needle cap color for insulin concentration was added in accordance with ISO 8537:2016.  |

{6}

Promisemed

Promisemed Hangzhou Meditech Co., Ltd.

No. 1388 Cangxing Street, Cangqian Community, Yuhang District,

Hangzhou City, 311121 Zhejiang, P. R. China.

|   | spring to minimize the risk of accidental sharps injury. Product is delivered sterile. Sterilization process is validated according to EN ISO 11135. This is a single-use device. Safety Sterile Syringes is intended to be used in professional healthcare facility/hospital and transport environment. | Product is delivered sterile. Sterilization process is validated according to EN ISO 11135. This is a single-use device. Safety Sterile Syringes is intended to be used in professional healthcare facility/hospital and transport environment. |  |   |
| --- | --- | --- | --- | --- |
|  Principle of operation | Safety Sterile Syringes is using plunger to fill syringe as well as discharge the fluid. After use, by continuing to press the push-button down to depress the plunger all of the way within the barrel, the safety mechanism will be triggered, then the needle tube, needle hub and retraction cylinder will be retracted into the barrel completely under the force of the spring to minimize the risk of accidental sharps injury. | No change | N/A | No new concerns. The principles of operation remain unchanged.  |
|  Indication of use | It is intended to inject fluids into the body or withdraw blood for medical purposes. After injection, the needle tube will be retracted into the barrel when the safety features are manually activated to minimize the risk of accidental sharps injury. | Verisafe Safety Retractable Insulin Syringes are intended for aspiration and subcutaneous injection of insulin. After injection, the needle can be retracted directly from the patient's skin into the syringe barrel to avoid exposure to a contaminated needle, thereby minimizing accidental needle stick injuries. | The indication is changed only for insulin. | No new concerns. Type 5 is added which its indication is only for insulin. It has narrowed the indication scope compared to the legally marketed predicate device and does not affect the safety and effectiveness of the subject device.  |
|  Type | Type D Type F | Type 5 | Type 5 is added which its indication is only for insulin. | No new concerns. Type 5 is added. This does not introduce new risks as the fundamental technology and operation remain the same.  |
|  Syringe volume | 1ml, 2ml, 3ml,5ml,10ml | 0.5ml, 1.0ml | Volume is only 0.5ml and 1.0ml. | No new concerns. The narrower volume scope does not introduce new risks as the fundamental technology and operation remain the same.  |
|  Needle size | 18G, 20G, 22G, 23G, 25G, 27G, 28G, 29G, 30G, 31G, 32G | 28G, 29G, 30G, 31G, 32G | Needle gauge is reduced to 28G, 29G, 30G, 31G, 32G. | No new concerns. The narrower gauge scope does not introduce new risks as the fundamental technology and operation remain the same.  |
|  Needle length | 6mm, 8mm, 12mm, 25mm, 38mm | 5mm, 6mm, 8mm, 12mm | Needle length is reduced to 5mm, 6mm, 8mm, 12mm. | No new concerns. The narrower needle length scope does not introduce new risks as the fundamental technology and operation remain the same.  |
|  Schematic diagram | Type D, Type F: Remain no change | Type 5: No change | N/A | No new concerns. The change to type  |

Page 3 of 5

{7}

Promisemed

Promisemed Hangzhou Meditech Co., Ltd.

No. 1388 Cangxing Street, Cangqian Community, Yuhang District,

Hangzhou City, 311121 Zhejiang, P. R. China.

|   |  |  |  | name and a narrower type does not introduce new risks as the fundamental technology and operation remain the same.  |
| --- | --- | --- | --- | --- |
|  Material | Type D, Type F: Remain no change | Type 5: Needle cap is PE; | Needle cap material is changed from PP to PE; | No new concerns. The change to needle cap has been confirmed to biocompatibility evaluation and doesn't introduce new risk.  |
|  Performance requirements | Type D, Type F: Complied with ISO 9626, ISO 7864 and ISO 23908. | Type 5: Complied with ISO 9626, ISO 7864, ISO 8537 and ISO 23908. | Type 5 is intended used for insulin injection. | No new concerns. Performance requirements of type 5 are in accordance with ISO 9626, ISO 7864, ISO 8537 and ISO 23908 ensuring consistent functionality.  |
|  Sterilization method | EO Sterilization | EO Sterilization | N/A | No new concerns. The change of type 5 is used the same sterilization method.  |
|  Sterility | SAL of 10-6 | SAL of 10-6 | N/A | No new concerns. The change of type 5 is required the same SAL.  |
|  Shelf-life | 3 years | 3 years | N/A | No new concerns. The change of type 5 is required the same shelf-life.  |
|  Single use | Single use | Single use | N/A | No new concerns. The change of type 5 is required of single use.  |
|  Biocompatibility | Complied with ISO10993 series standards, and the following tests are performed - Cytotoxicity: No cytotoxicity - Skin Irritation: No evidence of skin irritation - Skin Sensitization: No evidence of sensitization - Acute Systemic Toxicity: No systemic toxicity - Hemolysis: No evidence of hemolysis - Pyrogen: Non pyrogenic - USP <788> Particulate matter | Complied with ISO10993 series standards, and the following tests are performed: - Cytotoxicity: No cytotoxicity - Skin Irritation: No evidence of skin irritation - Skin Sensitization: No evidence of sensitization - Acute Systemic Toxicity: No systemic toxicity - Hemolysis: No evidence of hemolysis - Pyrogen: Non pyrogenic - USP <788> Particulate matter; - Subchronic test: No evidence of toxicity; | Subchronic test was added that no toxicity. | The subchronic testing is added of type 5 that consideration of insulin injection of accumulated using. The testing result is indicated that no toxicity.  |
|  Labeling | Type D, Type F: Remain no change | Type 5: Marking of insulin concentration on barrel, unit packaging, user packaging and IFU were updated in accordance with ISO 8537:2016. | Marking of insulin concentration on barrel, unit packaging, user packaging and IFU were updated in accordance with ISO 8537: 2016. | No new concerns. Labeling updates meet ISO 8537:2016 and enhance the device's safety and usability.  |

Page 4 of 5

{8}

Promisemed

Promisemed Hangzhou Meditech Co., Ltd.

No. 1388 Cangxing Street, Cangqian Community, Yuhang District,

Hangzhou City, 311121 Zhejiang, P. R. China.

|  8. Nonclinical test [21 CFR 807.92 (b) (1)]  |
| --- |
|  To support substantial equivalence to the predicate device, Promisemed Hangzhou Meditech Co., Ltd. completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met.
Verification and Validation Activities - Design Changes:
• Performance requirements: full performance testing was conducted using the worst-case representative model, including freedom from defects, cleanliness, tube dimensions, stiffness, resistance to corrosion, lubricant, needle point, fragment, patency, leakage, dead space, color code, package, safety mechanism, chemical and biological in accordance with ISO 9626, ISO 7864, ISO 8537 and ISO 23908.
• Material of needle cap (PE): Biocompatibility of type 5 needle cap (PE) has been confirmed that the material has been used in legally marketed device (K193273, manufactured by PROMISEMED) in the same contact method (intact skin) and duration (limited exposure-A) according to ISO 10993-1:2018.
All verification and validation tests passed without deviations, confirming that the subject device meet the necessary design specifications and regulatory requirements. The tests demonstrated that the product modifications did not introduce any new risks related to safety or effectiveness when compared to the predicate device.  |
|  9. Clinical test [21 CFR 807.92 (b) (2)]  |
|  No clinical testing is included in this submission.  |
|  10. Conclusion [21 CFR 807.92 (b) (3)]  |
|  Based on the information provided within this 510(k) submission, the subject device is substantially equivalent to the predicate device and is as safe, as effective and performs as well as the legally marketed predicate device.  |

Page 5 of 5

---

**Source:** [https://fda.innolitics.com/device/K250192](https://fda.innolitics.com/device/K250192)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
