K250172 · Steritec Products, Inc. (A Getinge Company) · JOJ · Feb 20, 2025 · General Hospital
Device Facts
Record ID
K250172
Device Name
Green Card Bowie-Dick Test (BD115)
Applicant
Steritec Products, Inc. (A Getinge Company)
Product Code
JOJ · General Hospital
Decision Date
Feb 20, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The Green Card Bowie-Dick Test Card is designed to detect the presence of residual air in prevacuum steam sterilizers operating at 270°F (132°C) and 273°F (134°C) for 3.5 minutes. The indicator will demonstrate a uniform color change from purple to green when the proper sterilization conditions are met, and no air is present. If enough air is present to create a 2°C or greater temperature difference in a standard Bowie-Dick test towel pack, the indicator will demonstrate a non-uniform color change.
Device Story
The Green Card Bowie-Dick Test is a chemical indicator card used for daily monitoring of pre-vacuum steam sterilizers. The device consists of a 3.5" x 1.91" card with indicator ink, laminated with polyester film. It is placed inside a holder assembly within the sterilizer chamber. During a sterilization cycle, the indicator ink undergoes a color change from purple to green. A uniform green color indicates successful air removal and proper sterilization conditions. A non-uniform color change, characterized by a purple center, indicates the presence of residual air sufficient to cause a 2°C temperature depression. Healthcare personnel interpret the color change visually to verify sterilizer performance. The device helps ensure the effectiveness of the sterilization process, thereby benefiting patients by confirming that instruments are processed under conditions sufficient to achieve sterility.
Clinical Evidence
Bench testing only. Performance verified against ANSI/AAMI/ISO 11140-5 standards for Class 2 indicators. Testing included uniform color change (pass), non-uniform color change (fail), dry heat testing, color density, color stability, and shelf-life stability. All tests passed.
Technological Characteristics
Chemical indicator card (3.5" x 1.91"). Materials: proprietary indicator ink, polyester film laminate. Sensing principle: chemical color change (purple to green) in response to steam and temperature conditions. Standalone device; no energy source or software. Complies with ANSI/AAMI/ISO 11140-5.
Indications for Use
Indicated for daily monitoring of pre-vacuum steam sterilizers to detect residual air in healthcare facilities. Used in cycles at 132°C and 134°C for 3.5 minutes.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font in blue.
February 20, 2025
SteriTec Products, Inc. (a Getinge Company) % Barb Smith Principal Regulatory Affairs Specialist Getinge 75 Barbour Pond Wayne, New Jersey 07470
Re: K250172
Trade/Device Name: Green Card Bowie-Dick Test (BD115) Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: January 17, 2025 Received: January 21, 2025
Dear Barb Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Christopher K. Dugard -S
Christopher K. Dugard, M.S. Division Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
# K250172
Device Name
Green Card Bowie-Dick Test (BD115)
#### Indications for Use (Describe)
The Green Card Bowie-Dick Test is designed to detect the presence of residual air in pre-vacuum steam sterilizers operating at 270°F (132°C) and 273°F (134°C) for 3.5 minutes. The indicator will demonstrate a uniform color change from purple to green when the proper sterilization conditions are met and no air is present. If enough air is present to create a 2°C or greater temperature difference in a standard Bowie-Dick test towel pack, the indicator will demonstrate a non-uniform color change.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "SterTec" in a bold, sans-serif font. The letters "Ster" are larger than the letters "Tec". A horizontal line underlines the entire word. A small triangle is located to the right of the word.
SteriTec Products, Inc. 74 Inverness Drive East Englewood, CO 80112 (303) 660-4201
510(k) Summary for the Green Card Bowie-Dick Test
Page 1 of 5
# 510(k) Summary K250172
Date Prepared: February 17, 2025
Applicant Information
SteriTec Products Manufacturing CO INC (a Getinge Group company) 74 Inverness Drive East Englewood, CO 80112 U.S. Ph: 303-660-4201
Contact: Barb Smith Principal Requlatory Affairs Specialist Ph: 585-370-6101 e-mail: barb.smith@getinge.com
- 1. Device Name
| Trade Name: | Green Card Bowie-Dick Test (BD115) |
|------------------------|---------------------------------------------------------------------------|
| Device Classification: | Class II |
| Common/Usual Name: | Indicator, physical/chemical sterilization process |
| Classification Name: | Indicator, physical/chemical sterilization process (21 CFR 880.2800, JOJ) |
- 2. Predicate Device
SteriTec Bowie-Dick Mini Pak (K024293). SteriTec Products, Manufacturing Co. Inc.
# 3. Description of Device
The Green Card Bowie-Dick Test is designed specifically for daily monitoring of pre-vacuum steam sterilizers to detect the presence of residual air as recommended by ANSI/AAMI ST79: Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities. The Green Card Bowie-Dick Test is a 3.5" X 1.91" card with the indicator ink printed on one side. Both sides of the card are laminated with a polyester film and an identification/record label is placed on the back side. The Green Card Bowie-Dick Test duplicates the resistance of a Bowie-Dick towel pack as described in ANSI/AAMI ST8 Hospital Steam Sterilizers.
The Green Card Bowie-Dick Test indicator areas will uniformly change color from purple to green when a passing Bowie-Dick test is performed. If a significant amount of air is present inside the sterilizer chamber during the sterilization cycle, it will be detected by the Green Card and K250172
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Steritec" in a bold, black font. The word is written in a stylized manner, with the letters overlapping each other. A horizontal line is present below the word, with a small triangle shape at the end of the line.
#### 510(k) Summary for the Green Card Bowie-Dick Test
displayed as a purple area on the indicator card, whereas the remainder of the indicator will have turned green. The non-uniform color change demonstrates the presence of air that would have created a 2°C temperature depression in a standard cloth Bowie-Dick pack.
# 4. Intended Use/Indications for Use:
The Green Card Bowie-Dick Test Card is designed to detect the presence of residual air in prevacuum steam sterilizers operating at 270°F (132°C) and 273°F (134°C) for 3.5 minutes.
The indicator will demonstrate a uniform color change from purple to green when the proper sterilization conditions are met, and no air is present. If enough air is present to create a 2°C or greater temperature difference in a standard Bowie-Dick test towel pack, the indicator will demonstrate a non-uniform color change.
# 5. Technological Characteristics
The chart below compares the technological characteristics of the subject device, to that of the predicate device:
| Feature | Predicate Device | Subject/Modified Device | Comparison |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | SteriTec Bowie-Dick Mini Pak<br>(K024293) | Green Card Bowie-Dick Test | |
| Intended<br>Use/<br>Indications<br>for Use | SteriTec Bowie-Dick Mini Pak is<br>designed to detect the presence<br>of residual air in pre-vacuum<br>steam sterilizers operating at 132°<br>C for 3.5 minutes.<br><br>The indicator will demonstrate a<br>uniform color change from purple<br>to green when the proper<br>sterilization conditions are met,<br>and no air is present. If enough<br>air is present to create a 2° C or<br>greater temperature difference in<br>a standard Bowie-Dick test towel<br>pack, the indicator will<br>demonstrate a non-uniform color<br>change. | The Green Card Bowie-Dick Test<br>Card is designed to detect the<br>presence of residual air in pre-<br>vacuum steam sterilizers<br>operating at 270°F (132°C) and<br>273°F (134°C) for 3.5 minutes.<br><br>The indicator will demonstrate a<br>uniform color change from purple<br>to green when the proper<br>sterilization conditions are met,<br>and no air is present. If enough<br>air is present to create a 2°C or<br>greater temperature difference in<br>a standard Bowie-Dick test towel<br>pack, the indicator will<br>demonstrate a non-uniform color<br>change. | The Intended Use of the<br>modified device is the same<br>as the predicate.<br><br>This submission proposes<br>the monitoring of an<br>additional exposure<br>temperature and time to the<br>Indications for Use of:<br>• 134°C 3.5 minutes<br><br>In addition, the name was<br>changed from Bowie-Dick<br>Mini Pak to Green Card<br>Bowie-Dick Test. |
| Device<br>Design | Card measuring 3.5"x1.85" with<br>indicator ink printed on one side<br>and laminated with a polyester<br>film on both sides.<br><br>Identification/record label is on<br>the back side opposite the<br>indicator side.<br><br>Indicator area will turn uniformly<br>from purple to green when a | Card measuring 3.5"x1.91" with<br>indicator ink printed on one side<br>and laminated with a polyester<br>film on both sides.<br><br>Identification/record label is on<br>the back side opposite the<br>indicator side.<br><br>Indicator area will turn uniformly<br>from purple to green when a | Minor change in card<br>dimension. |
| successful Bowie-Dick test is<br>performed. If a significant amount<br>of air is present in the sterilizer<br>chamber during the test, the<br>indicator will have a non-uniform<br>color change with purple still<br>present.<br>The card is placed inside a holder<br>assembly prior to placement in<br>the sterilizer. The holder is open<br>on both sides allowing the<br>unrestricted flow of steam and air<br>removal around the card. The<br>holder acts as a weight to keep<br>the indicator card from blowing<br>around during the test. The holder<br>does not have any functionality<br>regarding the test itself. | successful Bowie-Dick test is<br>performed. If a significant amount<br>of air is present in the sterilizer<br>chamber during the test, the<br>indicator will have a non-uniform<br>color change with purple still<br>present.<br>The card is placed inside a holder<br>assembly prior to placement in<br>the sterilizer. The holder is open<br>on both sides allowing the<br>unrestricted flow of steam and air<br>removal around the card. The<br>holder acts as a weight to keep<br>the indicator card from blowing<br>around during the test. The holder<br>does not have any functionality<br>regarding the test itself. | | |
| Sterilization<br>methods<br>and cycles | Steam sterilizer Bowie-Dick test<br>132°C 3.5 minutes | Steam sterilizer Bowie-Dick test<br>132°C 3.5 minutes and 134°C 3.5<br>minutes | This submission proposes<br>the monitoring of additional<br>exposure temperature and<br>time to the Indications for<br>Use of:<br>• 134°C 3.5 minutes |
| Indicator<br>Agent | Proprietary | Proprietary | Same |
| Color<br>Change | The indicator area will change<br>from purple to green when<br>exposed to steam sterilization<br>cycle with acceptable vacuum<br>cycles. | The indicator area will change<br>from purple to green when<br>exposed to steam sterilization<br>cycle with acceptable vacuum<br>cycles. | Same |
| Interpretati<br>on | If no significant amount of<br>residual air is present, the<br>indicator will show a uniform<br>green color change. If the<br>indicator shows a purple color in<br>the center, this indicates the<br>presence of residual air. | If no significant amount of<br>residual air is present, the<br>indicator will show a uniform<br>green color change. If the<br>indicator shows a purple color in<br>the center, this indicates the<br>presence of residual air. | Same |
| Shelf Life | 2 Years | 2 Years | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for SteriTec. The logo is in black and white, with the word "SteriTec" in a bold, sans-serif font. A horizontal line runs underneath the word, with a small triangle at the end. The logo is simple and modern.
SteriTec Products, Inc.
74 Inverness Drive East Englewood, CO 80112 (303) 660-4201
510(k) Summary for the Green Card Bowie-Dick Test
Page 3 of 5
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Steritec" in a bold, sans-serif font. The word is black and is set against a white background. A horizontal line runs beneath the word, adding a sense of grounding to the design. The overall design is simple and modern.
Page 4 of 5
#### 510(k) Summary for the Green Card Bowie-Dick Test
# 6. Non-Clinical Performance Testing
Performance testing was conducted to verify that the proposed Green Card Bowie-Dick Test meets the applicable portions of ANSI/AAMI/ISO 11140-5 Sterilization of health care products — Chemical indicators Part 5: Class 2 indicators for Bowie and Dick-type air removal tests.
The following table summarizes the performance testing that was completed, with acceptance criteria and results to demonstrate that performance of the Green Card Bowie-Dick Test is as safe and effective as the predicate device. This testing confirms and demonstrates that the proposed device meets the requirements of its pre-determined acceptance criteria in its claimed intended steam sterilization cycles.
| Performance Testing of 3<br>manufactured lots of Green<br>Card Bowie-Dick Test | Pre-Determined Acceptance Criteria | Results |
|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Uniform Color Change (Pass<br>Test)-ANSI/AAMI/ISO 111140-5<br>Annex B | The Standard Test Pack must demonstrate<br>no greater than a 0.5° C difference between<br>the inside of the Standard Test Pack and the<br>drain temperature and shall remain so for the<br>duration of the exposure time exclusive of a<br>15 second equilibration time. This should be<br>demonstrated with a thermocouple graph.<br>The SteriTec Green Card Bowie-Dick Test<br>Card indicator cards must demonstrate a<br>uniform color change to green | PASS |
| Non-Uniform Color Change (Fail<br>Test)- ANSI/AAMI/ISO 11140-5<br>Annex F | The Standard Test Pack must demonstrate a<br>2° C (+1°/-0°C) temperature difference<br>between the towel pack and the drain<br>temperatures for the last one minute of the<br>three and half minute sterilization cycle. This<br>should be demonstrated with a thermocouple<br>graph.<br>The SteriTec Green Card Bowie-Dick Test<br>Card indicator cards must demonstrate a non-<br>uniform color change with a purple area in the<br>center and the rest of the indicator green. | PASS |
| Dry Heat Testing - ANSI/AAMI/ISO<br>11140-5 Annex C | The SteriTec Green Card Bowie-Dick Test<br>Card shall show either no change or a change<br>that is markedly different from the change<br>occurring after exposure to a steam<br>sterilization process. | PASS |
| Color Density Test-<br>ANSI/AAMI/ISO 11140-5 Annex A | Difference in color density between the<br>substrate and indicator should be not less<br>than 0.3. | PASS |
| Indicator Color Stability | Verify that the Green Card Indicator sample<br>end color has not changed to any other color<br>and has remained green after 18 months or<br>longer from steam sterilization exposure. | PASS |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the SteriTec logo in black and white. The logo features the word "SteriTec" in a bold, stylized font. A horizontal line extends from the bottom of the "S" to below the "ec" portion of the word, creating a subtle underline effect.
SteriTec Products, Inc. 74 Inverness Drive East Englewood, CO 80112 (303) 660-4201
#### 510(k) Summary for the Green Card Bowie-Dick Test
Page 5 of 5
| | Verify that the Green Card Indicator sample<br>fail point color has not changed to any other<br>color and has remained purple in center of<br>card after 18 months or longer from steam<br>sterilization exposure. | |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Shelf-Life Stability | The following tests will be performed at 24<br>months of shelf life or longer.<br>● The pass cycle samples shall exhibit<br>complete color change to green.<br>● The fail cycle samples shall exhibit<br>incomplete color change. This<br>presents as a purple spot in the<br>center of the test card. | PASS |
# 7. Conclusion:
The conclusions drawn from the nonclinical tests demonstrate that the subject device in 510(k) submission K250172, the modified Green Card Bowie-Dick Test, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared in K024293, Class II Indicator, physical/chemical sterilization process (21 CFR 880.2800), JOJ.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.