← Product Code [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT) · K250171

# IPL Hair Removal Device (BE965A, BE965B, BE965C) (K250171)

_Shenzhen Yang WO Electronic Co., Ltd. · OHT · Mar 19, 2025 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K250171

## Device Facts

- **Applicant:** Shenzhen Yang WO Electronic Co., Ltd.
- **Product Code:** [OHT](/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT.md)
- **Decision Date:** Mar 19, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/ or facial hair.

## Device Story

Handheld IPL hair removal device; utilizes xenon arc flashlamp to emit light pulses (600-1200nm) to target hair follicles; operates via external power adapter (100-240V); features sapphire cooling for skin comfort; includes skin contact sensor to prevent accidental light emission; intended for home use by consumers; user places flash window against skin to trigger pulses; device inhibits hair growth; provides non-invasive hair removal.

## Clinical Evidence

Bench testing only; no clinical data. Device verified against IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-83, 62471, 60601-1-6, and IEC TS 60601-4-2. Biocompatibility evaluated per ISO 10993-1, 10993-5, 10993-10, and 10993-23.

## Technological Characteristics

Light source: Xenon Arc Flashlamp; Wavelength: 600-1200nm; Energy density: 2.75-6.56 J/cm²; Pulse mode: Double-pulse (0.82-5.32ms); Power: External adapter 100-240V, 50/60Hz; Features: Sapphire cooling, skin contact sensor; Form factor: Handheld; Software: Embedded firmware for pulse control and safety interlocks.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Ice Cooling IPL Hair Removal Device ([K241998](/device/K241998.md))
- IPL Hair Removal Device ([K241834](/device/K241834.md))
- Ice Cooling IPL Hair Removal Device ([K241205](/device/K241205.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date March 19, 2025. The month is spelled out, followed by the day and year. The text is in a simple, sans-serif font and is horizontally aligned. The date is likely intended to represent a specific event or deadline.

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol features a stylized depiction of a human figure, while the text is written in a clear, sans-serif font, with "FDA" in a larger, bolder typeface.

Shenzhen Yang Wo Electronic Co., Ltd. % Riley Chen Registration Engineer Feiying Drug & Medical Consultting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90, Oianhai Road Shenzhen, Guangdong 518052 China

Re: K250171

Trade/Device Name: IPL Hair Removal Device (BE965A, BE965B, BE965C) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: January 20, 2025 Received: January 21, 2025

Dear Riley Chen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

YAN FU -S
Digitally signed by YAN FU
-S
Date: 2025.03.19 08:49:13
-04'00'

for Tanisha Hithe

Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

510(k) Number (if known) K250171

Device Name

IPL Hair Removal Device (BE965A, BE965B, BE965C)

Indications for Use (Describe)

The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/ or facial hair.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

| 510(k) #:K250171                                          | 510(k) Summary                                                                                                                               |                                                          |                     |  | Prepared on: 2025-01-20 |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|---------------------|--|-------------------------|
| Contact Details                                           |                                                                                                                                              |                                                          | 21 CFR 807.92(a)(1) |  |                         |
| Applicant Name                                            | Shenzhen Yang Wo Electronic Co., Ltd                                                                                                         |                                                          |                     |  |                         |
| Applicant Address                                         | 101# 3Bldg, The third industry, Shangshi jia Village, Shijia Community,<br>Matian Street, GuangMing District Shenzhen Guangdong 518106 China |                                                          |                     |  |                         |
| Applicant Contact Telephone                               | +86-13058187625                                                                                                                              |                                                          |                     |  |                         |
| Applicant Contact                                         | Mr. Ken Wang                                                                                                                                 |                                                          |                     |  |                         |
| Applicant Contact Email                                   | wangkun@beautigo.com                                                                                                                         |                                                          |                     |  |                         |
| Correspondent Name                                        | FEIYING DRUG & MEDICAL CONSULTING TECHNICAL SERVICE                                                                                          |                                                          |                     |  |                         |
| Correspondent Address                                     | Rm 2401 Zhenye International Business Center, No. 3101-90,<br>Qianhai Road Shenzhen Guangdong 518052 China                                   |                                                          |                     |  |                         |
| Correspondent Contact Telephone                           | +86 13660660449                                                                                                                              |                                                          |                     |  |                         |
| Correspondent Contact                                     | Ms. Riley Chen                                                                                                                               |                                                          |                     |  |                         |
| Correspondent Contact Email                               | c3714930@gmail.com                                                                                                                           |                                                          |                     |  |                         |
| Device Name<br>21 CFR 807.92(a)(2)                        |                                                                                                                                              |                                                          |                     |  |                         |
| Device Trade Name                                         | IPL Hair Removal Device (BE965A, BE965B, BE965C)                                                                                             |                                                          |                     |  |                         |
| Common Name                                               | Laser surgical instrument for use in general and plastic surgery and in<br>dermatology                                                       |                                                          |                     |  |                         |
| Classification Name                                       | Light Based Over-The-Counter Hair Removal                                                                                                    |                                                          |                     |  |                         |
| Regulation Number                                         | 878.4810                                                                                                                                     |                                                          |                     |  |                         |
| Product Code(s)                                           | OHT                                                                                                                                          |                                                          |                     |  |                         |
| Legally Marketed Predicate Devices<br>21 CFR 807.92(a)(3) |                                                                                                                                              |                                                          |                     |  |                         |
| Predicate #                                               |                                                                                                                                              | Predicate Trade Name (Primary Predicate is listed first) |                     |  | Product Code            |
| K241998                                                   |                                                                                                                                              | Ice Cooling IPL Hair Removal Device                      |                     |  | OHT                     |
| K241834                                                   |                                                                                                                                              | IPL Hair Removal Device                                  |                     |  | OHT                     |
| K241205                                                   |                                                                                                                                              | Ice Cooling IPL Hair Removal Device                      |                     |  | OHT                     |
| Device Description Summary                                |                                                                                                                                              |                                                          |                     |  | 21 CFR 807.92(a)(4)     |

{5}------------------------------------------------

IPL Hair Removal Device is an over-the-counter, light based device for unwanted hair removal by using Intense Pulsed Light (IPL) technology. The device works below the skin's surface and does not involve any culling, removing hair growth with minimal pain.

The device includes three models with the only difference in enclosure color. It is only power adapter and its PL emission activation is by a switch or auto light emission, and it contains a skin sensor to detect appropriate skin contact, if the flash window is not in full contact with the skin, the device cannot emit the treatment light pulses. The device has a sapplire cooling function for a better hair removal experience.

### Intended Use/Indications for Use

The IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/ or facial hair.

## Indications for Use Comparison

The subject device has the same indications for use as the predicate devices, all are mainly used for removal of unwanted hair.

## Technological Comparison

The technical characteristic of PL Hair Removal Device (BE965A, BE965C) are substantially equivalent to the predicate and reference devices in the following aspects:

1. The same intended use, mode of action;

- 2. The same source energy and power supply: supplied by external adapter with 100-240V~, 50/60Hz;
3. The same light source: Intense Pulsed Light;

4. The same energy medium: Xenon Arc Flashlamp;

5. The same wavelength: 600-1200nm (same as the secondary predicate and within the range of the primary predicate).

The difference between the subject device and the predicate and reference devices mainly includes the following:

1. Different dimensions: Though the dimensions are different from the predicate devices and reference is insignificant and do not raise any safety or effectiveness problems.

2. Different energy density: The energy density (2.75~6.56)/cm²) of the subject device are a little different from that of the primary predicate device, but they are within the minimum value of the primary predicate device (1.67~6.67//cm²), and also very close to the value of the secondary predicate device (2.25~6.37J/cm²).

3. Different pulse duration: The subject device is double-pulse with a range in 0.82ms~5.32ms. Though the subject device is a little different from the primary predicate device (0.88~3.20ms), it is the same pulse output mode as the primary predicate device. And the maximum pulse duration of the subject device is within the range of the reference device K241205 (0.88~6.6ms). Thus the subject device is determined to be substantially equivalent to the predicate devices.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

21 CFR 807.92(a)(6)

21 CFR 807.92(a)(5)

# 21 CFR 807.92(a)(5)

{6}------------------------------------------------

Non-clinical testing have been conducted to verify that the PL Hair Removal Device meets all design specifications which supports the conclusion that it's substantially equivalent to the predicate devices. The testing results demonstrate that the subject device complies with the following standards:

IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION Medical electrical equipment - Part 1: General requirements for basic safety and essential performance;

IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests;

IEC 60601-1-11 Edition 2.1 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment;

IEC 60601-2-83 Edition 1.1 2022-12 CONSOLIDATED VERSION Medical electrical equipment - Part 2-83: General requirements for basic safety and essential performance of home light therapy equipment;

IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems;

IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability

IEC TS 60601-4-2 Edition 1.0 2024-03 Medical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems

The device has been evaluated for biocompatibility as per ISO 10993-1 Fifth edition 2018-08 and tested as per ISO 10993-5 Third edition 2009-06-01, ISO 10993-10 Fourth edition 2021-11 and ISO 10993-23 First edition 2021-01.

The device software has been evaluated as per FDA quidance " Content of Premarket Submissions for Device Software Functions". In order to verify and assure the performance, function and quality of the producted performance verification. The clinical test is not applicable, there's no clinical data.

Based on the above analysis and tests, the subject device is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K250171](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/OHT/K250171)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
