Small Volume 0.2mL Syringe

K250138 · Prosum Medical Limited · QNQ · Apr 2, 2025 · General Hospital

Device Facts

Record IDK250138
Device NameSmall Volume 0.2mL Syringe
ApplicantProsum Medical Limited
Product CodeQNQ · General Hospital
Decision DateApr 2, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Small Volume 0.2mL Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.

Device Story

Small volume 0.2 mL manual piston syringe for injection or withdrawal of fluids. Comprised of PP barrel, ABS plunger, TPE stopper, and optional PP Luer protective cap and plunger shield. Features Luer lock connection for hypodermic needles. Low dead space design (≤ 0.0058 mL) minimizes residual liquid. Some models include Enhanced Dosing Technology (EDT) with colored reference lines on the plunger for improved visibility. Used in clinical settings by healthcare professionals. Device is single-use, sterile, and silicone-oil free. Benefits include precise delivery of small volumes and reduced medication waste.

Clinical Evidence

No clinical data. Bench testing only. Performance testing included ISO 7886-1 (syringe performance), ISO 80369-7/20 (Luer connector), and ASTM D4169-22 (shipping). Biocompatibility testing per ISO 10993-1 (cytotoxicity, irritation, sensitization, acute systemic toxicity, hemocompatibility, pyrogenicity). Sterility and shelf-life validated per ISO 11135, ISO 11607, and ASTM F1980.

Technological Characteristics

Manual piston syringe. Materials: PP barrel, ABS plunger, TPE stopper. Luer lock connection per ISO 80369-7. Dead space ≤ 0.0058 mL. Sterilization: EO gas (SAL 10^-6). Shelf-life: 3 years. Biocompatibility per ISO 10993-1. Performance testing per ISO 7886-1. Packaging validation per ISO 11607 and ASTM F1929/F88.

Indications for Use

Indicated for medical purposes to inject fluids into or withdraw fluids from the body. Prescription use only.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION April 2, 2025 Prosum Medical Limited % Scott Zawko Manager, Regulatory Affairs Kymanox 430 Davis Dr Ste #300 Morrisville, North Carolina 27560 Re: K250138 Trade/Device Name: Small Volume 0.2mL Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QNQ Dated: January 15, 2025 Received: January 17, 2025 Dear Scott Zawko: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250138 - Scott Zawko Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250138 - Scott Zawko Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Rong Guo -S Shruti Mistry Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K250138 | | | Device Name Small Volume 0.2mL Syringe | | | Indications for Use (Describe) The Small Volume 0.2mL Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} Page 1 of 5 510(k) Summary 1 Date Prepared March 31, 2025 2 Submitter Manufacturer: Prosum Medical Limited Address: Unit 14 Trade City Business Park, Cowley Mill Road, Uxbridge, UB8 2DB, United Kingdom Contact Person: Yanshen Sun Phone: +44 1895 913302 Email: info@prosummedical.co.uk 3 Device Trade name: Small Volume 0.2mL Syringe Common Name: Syringe Classification name: Low Dead Space Piston Syringe Regulation: 21 CFR 880.5860 Product code: QNQ Classification: Class II 4 Predicate Device Predicate device: K210443, PLPT LDV Sterile Syringe 5 Device Description The Small Volume 0.2mL Syringe is intended for manual, single use. It is a small volume 0.2 mL syringe. It is comprised of a PP barrel, ABS plunger, TPE plunger stopper, PP plunger shield and PP Luer protective cap. The device has a Luer lock connection type that enables connection to a hypodermic needle for liquid injection or withdrawal. The device has a plunger stopper to prevent any liquid residual from remaining in the syringe barrel, maximizing the efficiency of liquid usage. The feature is called Low Dead Space/Volume. The Luer protective cap serves to safeguard the small Luer connector from potential damage during transportation. Similarly, the plunger shield is utilized to ensure the small plunger remains undamaged during transportation. The Luer protective cap and the plunger shield are optional to be used. The device has been manufactured without any silicone oil. The device has six models, each equipped with a 0.2 mL Luer lock syringe. Among these models, three include an additional feature for enhanced dosing technology (EDT), which comprises a colored reference line on the plunger surface to improve visibility. These include identifying markings to distinguish the EDT models from the regular 0.2 mL syringe. {5} Table 1. Device Models | No. | Syringe Volume (mL) | Luer type | Description | | --- | --- | --- | --- | | 1 | 0.2 | Luer Lock | Small Volume 0.2mL Syringe, Luer Lock, EDT | | 2 | 0.2 | Luer Lock | Small Volume 0.2mL Syringe, Luer Lock, EDT, Luer protective cap | | 3 | 0.2 | Luer Lock | Small Volume 0.2mL Syringe, Luer Lock, EDT, Luer protective cap, Plunger shield | | 4 | 0.2 | Luer Lock | Small Volume 0.2mL Syringe, Luer Lock | | 5 | 0.2 | Luer Lock | Small Volume 0.2mL Syringe, Luer Lock, Luer protective cap | | 6 | 0.2 | Luer Lock | Small Volume 0.2mL Syringe, Luer Lock, Luer protective cap, Plunger shield | # 6 Indications for Use/Intended Use The Small Volume $0.2\mathrm{mL}$ Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body. # 7 Technological Characteristics The table below compares the technological characteristics of the subject device and those of the predicate device. Table 2. Device Comparison | | Subject Device (Small Volume 0.2mL Syringe) | Predicate Device (PLPT LDV (Low Dead Volume) Sterile Syringe, K210443) | Comments | | --- | --- | --- | --- | | Indications for Use / Intended use | The Small Volume 0.2mL Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body. | The PLPT LDV (Low Dead Volume) sterile syringe is intended to be used for medical purposes to inject fluid into or withdraw fluid from the human body. | Same | | Product Code | QNQ | QNQ | Same | | Rx Only or OTC | Prescription Use Only | Prescription Use Only | Same | {6} | | Subject Device (Small Volume 0.2mL Syringe) | Predicate Device (PLPT LDV (Low Dead Volume) Sterile Syringe, K210443) | Comments | | --- | --- | --- | --- | | Mechanism of Action | Manual use | Manual use | Same | | Components and Materials | - Barrel: PP - Plunger: ABS - Stopper: TPE - Luer protective Cap: PP - Plunger shield: PP | - Barrel: PP - Plunger: PP - Stopper: Rubber | See Comment #1 below | | Syringe Volume | 0.2 mL | 1 mL | See Comment #2 below | | Connection Type | Luer-lock | Luer-lock | Same | | Sterilization | EO gas SAL: 10-6 | EO gas SAL: 10-6 | Same | | Shelf-life | 3 years | 3 years | Same | | Biocompatibility | Complies with ISO 10993-1 - Cytotoxicity - Irritation - Sensitization - Acute systemic toxicity - Hemocompatibility - Pyrogenicity | Complies with ISO 10993-1 - Cytotoxicity - Irritation - Sensitization - Acute systemic toxicity - Hemocompatibility - Pyrogenicity | Same | | Syringe Performance Testing | Complies with the following standards - ISO 7886-1 - ISO 80369-7 | Complies with the following standards - ISO 7886-1 - ISO 80369-7 | Same | | Dead Space Specification | ≤ 0.0058 mL 95% confidence/ 95% reliability meets requirements | ≤ 0.023 mL 95% confidence/ 95% reliability meets requirements | See Comment #2 below | Page 3 of 5 {7} Comment #1, Components and Materials: The proposed device features an additional Luer protective cap and a plunger shield that safeguard the luer connector and the plunger from potential damage during transportation. The Luer protective cap and plunger shield are discarded by the user prior to use and do not alter the mechanism of action or functionality of the proposed device in comparison to the predicate. Protection of the device through transportation has been demonstrated through packaging validation per ISO 11607. The proposed device utilizes different polymer materials for the components in comparison to the predicate device. The difference of materials has been addressed through demonstration of biocompatibility in accordance with the ISO 10993 standards. Comment #2, Syringe Volume and Dead Space Specification: The proposed device features a lower syringe volume in comparison to the predicate device. In addition to the lower total syringe volume, a lower dead space volume is achieved. These differences have been assessed through syringe performance testing conducted in accordance with ISO 7886-1 that demonstrates that the proposed device meets all necessary requirements. The differences in syringe volume and dead volume specifications do not raise new questions of safety and effectiveness when compared to the predicate device. # 8 Non-clinical testing ## 8.1 Performance testing The proposed device was tested in compliance with ISO 7886-1:2017 for evaluating the overall non-clinical performance. In addition, Luer connector testing in compliance with ISO 80369-7:2016 and ISO 80369-20:2015 was conducted for evaluating the performance of connection to hypodermic needle. The performance and design testing results met the standards' requirements to demonstrate the device's substantial equivalence. Testing was conducted on unconditioned samples as well as samples aged according to ASTM F1980 and samples transit conditioned according to ASTM D4169-22. ## 8.2 Biocompatibility testing The proposed device was tested in compliance with FDA Guidance Use of International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process", as an Externally Communicating Device, Blood Path Indirect, Limited Contact (<24 h). The following endpoints were tested in accordance with ISO 10993-1: cytotoxicity, irritation, sensitization, acute systemic toxicity, hemocompatibility, and pyrogenicity. Particulate matter was tested in compliance with USP <788> and bacterial endotoxins were tested in compliance with USP <85> and met the acceptance criteria. ## 8.3 Sterility, packaging, and shelf-life The sterilization process of the proposed device has been validated in compliance with ISO 11135. The EO and ECH residuals do not exceed the limits according to ISO 10993-7. The shelf-life verification was conducted under accelerated aging conditions in compliance with ASTM F1980-16 to demonstrate device functionality through the claimed shelf-life of 3 years. Page 4 of 5 {8} Performance testing, package integrity testing, sterility testing, and particulate testing were conducted on samples accelerated aged to 3 years. Real time testing will be conducted to confirm device functionality through a shelf-life of 3 years. Package integrity testing under simulated shipping conditions was conducted to satisfy the requirements in ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems. Simulated shipping conditions consisted of Distribution Cycle 13, Assurance Level 2. All packaging was deemed acceptable for protection of product and sterility maintenance. Sterile barrier testing was conducted in compliance with the following FDA recognized consensus standards. - ASTM F88/F88-15 Standard Test Method for Seal Strength of Flexible Barrier Materials - ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration - USP <71> Sterility Tests A shelf life of 3 years is validated in accordance with ISO 11607-1 and ISO 11607-2. ## 9 Clinical testing Not applicable for this submission. ## 10 Conclusion The differences between the predicate and the proposed device do not raise any new or different questions of safety or effectiveness. The proposed device is substantially equivalent to the predicate device with respect to indications for use and technological characteristics. Page 5 of 5
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...