← Product Code [FED](/productcode/FED) · K250128

# Single Use Suction-Evacuation Ureteral Access Sheath (K250128)

_Anhui Happiness Workshop Medical Instruments Co., Ltd. · FED · Sep 19, 2025 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K250128

## Device Facts

- **Applicant:** Anhui Happiness Workshop Medical Instruments Co., Ltd.
- **Product Code:** [FED](/productcode/FED.md)
- **Decision Date:** Sep 19, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Single Use Suction-Evacuation Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. It is designed to establish a conduit for the treatment of urinary stones during endoscopic procedures.

## Device Story

Single-use ureteral access sheath; creates conduit to urinary tract for endoscope/instrument passage; facilitates stone cleaning via fluid injection and vacuum pressure. Components include dilator, sheath, sheath hub, sealing cap, dilator hub, and pressure control vent. Outer surfaces feature hydrophilic coating for lubrication; distal 10cm is flexible to accommodate endoscope movement; radiopaque for fluoroscopic visualization. Used in clinical settings by urologists during endoscopic procedures. Device provides access and suction-evacuation capabilities to assist in stone removal, potentially improving procedural efficiency and patient outcomes.

## Clinical Evidence

Bench testing only. Performance testing included X-ray radiodetectability, luer connector integrity, functional testing, transportation simulation (ASTM D4169), biocompatibility (ISO 10993), sterilization validation (USP 71, ISO 11135), hydrophilic coating performance, and shelf-life (ASTM F1980). All tests passed.

## Technological Characteristics

Materials: PVP, Pebax, SUS304, PTFE, PE. Features: hydrophilic coating, radiopaque markers, pressure control vent. Dimensions: 10-12 Fr. Sterilization: EO. Connectivity: None. Software: None.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- ClearPetra Suction-Evacuation Sheath ([K161110](/device/K161110.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

September 19, 2025

Anhui Happiness Workshop Medical Instruments Co., Ltd
% Daguang Sun
Consultant
PureID Medical Technology Co., Ltd.
33F, Building A, Guanzhou Life Sci Innov Centre,
Guangzhou Int'l Bio Island, Huangpu District
Guangzhou,
China

Re: K250128
Trade/Device Name: Single Use Suction-Evacuation Ureteral Access Sheath
Regulation Number: 21 CFR§ 876.1500
Regulation Name: Endoscope and accessories
Regulatory Class: II
Product Code: FED
Dated: August 8, 2025
Received: August 8, 2025

Dear Daguang Sun:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K250128 - Daguang Sun
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

{2}

K250128 - Daguang Sun
Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MARK J. ANTONINO -S

Mark J. Antonino, M.S.
Assistant Director
DHT3B: Division of Reproductive, Gynecology and Urology Devices
OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K250128

Device Name
Single Use Suction-Evacuation Ureteral Access Sheath

Indications for Use (Describe)
The Single Use Suction-Evacuation Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. It is designed to establish a conduit for the treatment of urinary stones during endoscopic procedures.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}

PureID Medical Technology Co., Ltd.
2022/08/09
Page 1 of 3
K250128

# 510(k) Summary

|  Prepared on: | 2025-08-05  |
| --- | --- |
|  Contact Details [21 CFR 807.92(a)(1)] |   |
|  Applicant Name | ANHUI HAPPINESS WORKSHOP MEDICAL INSTRUMENTS CO., LTD  |
|  Applicant Address | 5/F, Building 4, Bengshan Science And Technology Innovation Park, 1600 Yan'an South Road, 233000 Bengbu, Anhui Province, China (+86) 15313775083  |
|  Applicant Contact Telephone |   |
|  Applicant Contact | Wang Shuanghong  |
|  Applicant Contact Email | bryant.wang@hwmeds.com  |
|  Correspondent Name | PureID Medical Technology Co., Ltd.  |
|  Correspondent Address | 33F, Building A, Guanzhou Life Sci Innov Centre, Guangzhou Int'l Bio Island, Guangzhou China 86-13776067922  |
|  Correspondent Contact Telephone |   |
|  Correspondent Contact | Daguang Sun  |
|  Correspondent Contact Email | raq@pureglobal.cn  |
|  Device Name [21 CFR 807.92(a)(2)] |   |
|  Device Trade Name | Single Use Suction-Evacuation Ureteral Access Sheath  |
|  Common Name | Endoscopic Access Overtube, Gastroenterology-Urology  |
|  Device Classification Name | Endoscope and Accessories  |
|  Device Classification | Class II  |
|  Regulation Number | 21 CFR 876.1500  |
|  Product Code | FED  |
|  Legally Marketed Predicate Devices [21 CFR 807.92(a)(3)] |   |
|  Predicate Device 510(k) # | K161110  |
|  Predicate Trade Name (Primary Predicate is listed first) | ClearPetra Suction-Evacuation Sheath  |
|  Predicate Device Product Code | FED and 21 CFR 876.1500  |
|  Device Description Summary [21 CFR 807.92(a)(4)] |   |
|  The Single Use Suction-Evacuation Ureteral Access Sheath is a single-use ureteral access sheath, which can create access to the urinary tract by provide ureteral dilation and a continuous working channel for the introduction of endoscope and instruments during ureteral access procedures. The  |   |

Page 1 of 3
33F, Building A, Guanzhou Life Sci Innov Centre, Guangzhou Int'l Bio Island, Huangpu District, Guangzhou +86 400-0933-301 | info@cn.purefda.com | www.purefda.com

{5}

PureID Medical Technology Co., Ltd. 2022/08/09

K250128

Page 2 of 3

device can inject fluids and has vacuum pressure design which can facilitate with stone cleaning procedures.

The device consists of Dilator, Sheath, Sheath hub, Sealing cap, Dilator hub and Pressure control vent; Both dilator and sheath outer surfaces are coated with hydrophilic coatings, which provide long-lasting lubrication for better expansion and passage. The distal end of the sheath (10cm) can bend with the endoscopes, and the tip is soft which provides seamless transition between sheath core and sheath tube. The dilator and sheath are designed with radiopaque which is visible under X-Ray, can be visualized under X-ray(fluoroscopy) during placement to confirm location.

# Intended Use/Indications for Use [21 CFR 807.92(a)(5)]

The Single Use Suction-Evacuation Ureteral Access Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. It is designed to establish a conduit for the treatment of urinary stones during endoscopic procedures.

# Technological Comparison [21 CFR 807.92(a)(6)]

There are minor differences of technological characteristic between the subject device and predicate device.

|  Attribute | Subject: Single Use Suction-Evacuation Ureteral Access Sheath | Predicate: K161110 ClearPetra Suction-Evacuation Sheath  |
| --- | --- | --- |
|  Manufacturer | ANHUI HAPPINESS WORKSHOP MEDICAL INSTRUMENTS CO., LTD | Well Lead Medical Co., LTD.  |
|  Product Code | Same | FED, FAJ, FGA  |
|  Regulation Number | Same | 21 CFR§ 876.1500  |
|  Intended Use | Same |   |
|  Indications for Use | The Single-Use Suction-Evacuation Ureteral Access Sheath is intended to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. It is designed to establish a conduit for the treatment of urinary stones during endoscopic procedures. | The ClearPetra Suction-Evacuation Sheath is used to establish a conduit during endoscopic urological procedures facilitating the passage of endoscopes and other instruments into the urinary tract. It is designed to establish a conduit for the treatment of urinary stones or other urinary diseases during endoscopic procedures.  |
|  Mechanism of Action | Same | For the passage of endoscopes  |

33F, Building A, Guanzhou Life Sci Innov Centre, Guangzhou Int'l Bio Island, Huangpu District, Guangzhou +86 400-0933-301 | info@cn.purefda.com | www.purefda.com

{6}

PureID Medical Technology Co., Ltd.
2022/08/09
K250128
Page 3 of 3

|   |  |   | and other urological devices for the purpose of performing diagnostic and surgical procedures such as, nephroscopy, ureteroscopy, or cystoscopy.  |
| --- | --- | --- | --- |
|  Patient Population | Same |   | Patients undergoing endoscopic procedures, through percutaneous passages  |
|  Patient Contacting Materials | Sheath | PVP, Pebax, SUS304, PTFE | PE, PA  |
|   |  Dilator | PE  |   |
|  Sterilization | Same |   | EO Sterilized  |
|  Size | 10, 10.5, 11, 11.5 and 12 Fr, 3.33mm, 3.5, 3.67, 3.83 and 4mm respectively |   | 10Fr to 22Fr  |
|  OD of Obturator |  |   | Compatible with 10Fr-26Fr sheath  |
|  Length | Same |   | 13cm to 55cm  |
|  Components | Dilator
Sheath
Sheath hub
Sealing cap
Dilator hub
Pressure control vent |   | Sheath
Obturator
Connector
Rubber Cap  |
|  Single Use | Same |   | Yes  |
|  Biocompatibility | Same |   | Compliance with ISO 10993-1  |

## Non-Clinical and/or Clinical Tests Summary &amp; Conclusions [21 CFR 807.92(b)]

The performance testing has been conducted on the device on the following:

- Device Performance Study including X-Ray Radiodetectability, Luer Connector and Functional testing – Passed.
- Transportation Simulation Testing per ASTM D4169 – Passed.
- Biocompatibility Study per ISO 10993 Standards – Passed.
- Sterilization Validation per USP 71 and ISO 11135 – Passed.
- Hydrophilic Coating – Passed.
- Shelf-Life per ASTM F1980 – Passed.

## Conclusion

It can be concluded that, based on the obtained results from the nonclinical and clinical testing, the device is substantially equivalent to the predicate device.

Page 3 of 3
33F, Building A, Guanzhou Life Sci Innov Centre, Guangzhou Int'l Bio Island, Huangpu District, Guangzhou
+86 400-0933-301 | info@cn.purefda.com | www.purefda.com

---

**Source:** [https://fda.innolitics.com/device/K250128](https://fda.innolitics.com/device/K250128)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
