Retrospective clinical study used to establish the clinical performance (Positive Percent Agreement and Negative Percent Agreement) of the miDiagnostics HSV-1&2 CSF Test by comparing results against a comparator method.
Retrospective Clinical Study (Cohort 2); Retrospective clinical performance study; Study Period: August 2024 - December 2024
Patients presenting with signs and symptoms of HSV CNS infection; Sample Size: 152 specimens; Number of Sites: 4 participating sites
Standard of Care PCR testing
Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA)
Indications for Use
The miDiagnostics HSV-1&2 CSF Test is an in vitro polymerase chain reaction (PCR) assay intended for use with the miDiagnostics PCR Platform for the qualitative detection and differentiation of HSV-1 and HSV-2 DNA in cerebrospinal fluid (CSF) samples from patients suspected of Herpes Simplex Virus (HSV) infection of the central nervous system (CNS). This test is intended as an aid in the diagnosis of HSV-1 and HSV-2 infection in the CNS. Negative results do not preclude HSV-1 or HSV-2 infection and should not be used as the sole basis for treatment or other patient management decisions. The test is not intended for use as a donor screening test. The test is for professional use only.
Device Story
The miDiagnostics HSV-1&2 CSF Test is a molecular diagnostic assay for qualitative detection and differentiation of HSV-1 and HSV-2 DNA in CSF specimens. The process involves manual sample preparation (dilution, concentration, heating) followed by an automated PCR run on the miDiagnostics PCR Platform. The platform uses a single-use fluidic chip (PCR Card 2.0) and a reader to perform thermocycling and real-time fluorescence detection of viral targets (Glycoprotein B for HSV-1, Glycoprotein D for HSV-2) and a human housekeeping gene (RPP30) as a sample processing control. The miDiagnostics Software interprets fluorescence signals against a predetermined cut-off to report 'Detected', 'Not Detected', or 'Invalid' results. Used in clinical settings by professionals, the device aids in diagnosing CNS HSV infections. Results assist clinicians in patient management decisions, though negative results do not rule out infection.
Clinical Evidence
Clinical performance was established using 325 CSF specimens (prospective and retrospective cohorts). For HSV-1, PPA was 100% (16/16) and NPA was 97.79% (133/136) in retrospective samples. For HSV-2, PPA was 100% (29/29) and NPA was 98.37% (121/123) in retrospective samples. Prospective study NPA was 98.84% for HSV-1 and 99.42% for HSV-2. Discrepant analysis indicated some false positives were clinically diagnosed with HSV infection, suggesting high sensitivity.
Technological Characteristics
Real-time PCR assay; targets HSV-1 (Glycoprotein B), HSV-2 (Glycoprotein D), and human RPP30 (SPC). Uses a disposable fluidic chip (miDiagnostics PCR Card 2.0) and a dedicated reader (miDiagnostics PCR Reader 2.0). Connectivity via USB, Ethernet, or WiFi. Software-based automated analysis of fluorescence signals. Manual sample preparation required.
Indications for Use
Indicated for qualitative detection and differentiation of HSV-1 and HSV-2 DNA in CSF samples from patients of all ages/sexes suspected of CNS HSV infection. Intended as an aid in diagnosis. Not for donor screening. Professional use only.
Regulatory Classification
Identification
A herpes simplex virus nucleic acid-based assay for central nervous system infections is a qualitative in vitro diagnostic device intended for the detection and differentiation of HSV-1 and HSV-2 in cerebrospinal fluid (CSF) samples from patients suspected of Herpes Simplex Virus (HSV) infections of the central nervous system (CNS). This test is intended as an aid in the diagnosis of HSV-1 and HSV-2 infections of the CNS. Negative results do not preclude HSV-1 or HSV-2 infection and should not be used as the sole basis for treatment or other patient management decisions.
Special Controls
(b) Classification. Class II (special controls). The special controls for this device are:
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) Detailed documentation for the device description, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including primer design and selection.
(ii) Detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (limit of detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carryover, and cross contamination. Documentation must include reagent and sample stability recommendations.
(iii) Detailed documentation from a clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to the results of two polymerase chain reaction methods followed by bidirectional sequencing.
(iv) Documentation of an appropriate end user device training program that will be offered as part of efforts to mitigate the risk of failure to correctly operate the instrument.
(v) Quality assurance protocols and detailed documentation for device software, including standalone software applications and hardware-based devices that incorporate software.
(2) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed explanation of the interpretation of results and acceptance criteria.
(ii) A limiting statement indicating that negative results do not preclude HSV-1 or HSV-2 infection and should not be used as the sole basis for treatment or other patient management decisions.
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K250050
B Applicant
miDiagnostics nv
C Proprietary and Established Names
miDiagnostics HSV-1&2 CSF Test
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PGH | Class II | 21 CFR 866.3307 - Herpes Simplex Virus Nucleic Acid-Based Assay For Central Nervous System Infections | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
Clearance of a new device
B Measurand:
Target DNA sequences from conserved regions of the herpes simplex virus type 1 (HSV-1) and herpes simplex virus type 2 (HSV-2) genes.
C Type of Test:
Molecular diagnostic test using real-time PCR (Polymerization Chain Reaction) technology for the qualitative detection and differentiation of HSV 1 and HSV 2 DNA present in cerebrospinal fluid (CSF) specimens, obtained via lumbar puncture from individuals with signs and/or symptoms of HSV-1 or HSV-2 infection in the central nervous system (CNS).
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
K250050 - Page 2 of 18
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
The miDiagnostics HSV-1&2 CSF Test is an in vitro polymerase chain reaction (PCR) assay intended for use with the miDiagnostics PCR Platform for the qualitative detection and differentiation of HSV-1 and HSV-2 DNA in cerebrospinal fluid (CSF) samples from patients suspected of Herpes Simplex Virus (HSV) infection of the central nervous system (CNS). This test is intended as an aid in the diagnosis of HSV-1 and HSV-2 infection in the CNS.
Negative results do not preclude HSV-1 or HSV-2 infection and should not be used as the sole basis for treatment or other patient management decisions.
The test is not intended for use as a donor screening test. The test is for professional use only.
### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
### D Special Instrument Requirements:
miDiagnostics PCR Platform
## IV Device/System Characteristics:
### A Device Description:
The miDiagnostics HSV-1&2 CSF Test is a real-time polymerase chain reaction (PCR) assay that enables the direct amplification, detection, and differentiation of Herpes Simplex Virus 1 (HSV-1) and Herpes Simplex Virus 2 (HSV-2) DNA from cerebrospinal fluid (CSF) specimens, obtained via lumbar puncture, from individuals with signs and/or symptoms of HSV-1 or HSV-2 infection in the central nervous system (CNS).
The test consists of three major steps: (1) manual sample preparation, (2) manual PCR reaction preparation and (3) automated PCR run on the miDiagnostics PCR Platform, which includes targets amplification and detection by fluorescent probes for 1 target specific to HSV-1, 1 target specific to HSV-2 and a human DNA control target, the Sample Processing Control (SPC).
The miDiagnostics HSV-1&2 Test includes the following components:
1. miDiagnostics HSV-1&2 CSF Detection Mix - contains a freeze-dried cake, stored in a tube, containing the PCR components for amplification and fluorescent detection of the HSV-1, HSV-2 and the Sample Processing Control (SPC). Each tube is individually packed in a pouch. The device consists of 24 pouches in a box.
2. miDiagnostics HSV-1&2 CSF Dilution Buffer - contains 300 µL HSV CSF Dilution Buffer, including the sample processing control SPC (RPP30, a human housekeeping
{3}
gene) that acts as an internal control, stored in a tube, for diluting the CSF specimen. The device consists of 24 tubes in the box.
3. miDiagnostics Concentration Device - consists of one bag with 24 concentration columns and 48 collection tubes to concentrate the viral material.
4. miDiagnostics HSV-1&2 CSF Algorithm - The software system interprets the fluorescence signals and reports the test results. The algorithm is embedded in the miDiagnostics Software.
The miDiagnostics PCR Platform includes the following components:
1. The miDiagnostics PCR Reader 2.0 - receives the miDiagnostics PCR Card 2.0 and is responsible for performing consecutive PCR cycles through the thermocycler.
2. The miDiagnostics PCR Card 2.0 is a single-use, disposable test accessory. The miDiagnostics PCR Card 2.0 is manually loaded with the sample mixed with PCR components of the selected miDiagnostics test (called reaction mixture). Once loaded, the miDiagnostics PCR Card 2.0 is manually inserted in the miDiagnostics PCR Reader 2.0. The miDiagnostics PCR Card 2.0 includes a fluidic chip with the reaction chamber that enables the polymerase chain reaction driven by the miDiagnostics PCR Reader 2.0. Each miDiagnostics PCR Card 2.0 is individually packed in a pouch. The device consists of 24 pouches in a box.
3. The miDiagnostics Computer connects to the miDiagnostics PCR Reader 2.0 via two USB-cables and can connect to the customer's network via ethernet or WiFi. It interacts with a user via a touchscreen, has an integrated barcode scanner that allows the user to scan the sample consumable identifiers, and it runs the miDiagnostics Software.
The miDiagnostics Software is the software of the miDiagnostics PCR Platform, that is installed on the miDiagnostics Computer. It controls the miDiagnostics PCR Reader 2.0, retrieves the camera captures, hosts the test specific algorithm that interprets the fluorescence signals, reports the test results, and provides a user interface for the operator to interact with the miDiagnostics PCR Platform.
## B Principle of Operation:
The miDiagnostics HSV-1&2 CSF Test is a real-time polymerase chain reaction (PCR) assay that enables the direct amplification, that detects and differentiates HSV-1 and HSV-2 genomic DNA present in human CSF specimen. The miDiagnostics HSV-1&2 Dilution Buffer contains the Sample Processing Control SPC (RPP30 - Internal Control) which acts as a control for the entire process (specimen processing, amplification, and detection). Identification of HSV-1, HSV-2 and the SPC occurs using target-specific primers and fluorescent-labeled probes that hybridize to conserved regions in the viral genomes.
The SPC is used to monitor the integrity of the miDiagnostics HSV-1&2 CSF Test process, instrument, or reagent failures and PCR inhibition. The miDiagnostics HSV-1&2 CSF Algorithm compares the fluorescence signal to a predetermined cut-off to produce a qualitative result for the presence or absence of the analyte.
K250050 - Page 3 of 18
{4}
Table. Primer and Probe Targets
| Analyte | Gene Targeted |
| --- | --- |
| HSV-1 | Glycoprotein B |
| HSV-2 | Glycoprotein D |
| SPC | Ribonuclease P/MRP subunit p30 RPP30 |
As described under the device description above, the miDiagnostics HSV-1&2 CSF test consists of three major steps: (1) manual sample preparation, (2) manual PCR reaction preparation and (3) automated PCR run on the miDiagnostics PCR Platform which is an automated in vitro diagnostic system for the amplification and detection of the target nucleic acid(s) using real-time polymerase chain reaction (PCR).
The manual sample preparation starts out with adding $200~\mu \mathrm{L}$ CSF to the miDiagnostics HSV-1&2 CSF Dilution Buffer. The diluted specimen is then transferred to the miDiagnostics Concentration Device and centrifuged at high speed for 5 minutes. The remaining volume in the miDiagnostics Concentration Column is washed with molecular grade water (DNase & RNase free) and centrifuged again at high speed for 5 minutes. The sample is then heated for 2 minutes at $95^{\circ}\mathrm{C}$ . Thereafter, an aliquot of the sample is mixed with the miDiagnostics HSV-1&2 CSF Detection Mix and loaded on the miDiagnostics PCR Card 2.0, to fill the reaction chamber of the fluidic chip.
The miDiagnostics PCR Card 2.0 is manually inserted in the miDiagnostics PCR Reader 2.0 for amplification and detection of HSV-1, HSV-2 and the SPC. The miDiagnostics PCR Reader 2.0 rapidly heats and cools the chip to drive a polymerase chain reaction inside the reaction chamber of the fluidic chip. Via the transparent optical window in the miDiagnostics PCR Card 2.0's bottom, the progress of the reaction is measured by a camera system in the miDiagnostics PCR Reader 2.0, that sends the raw images to the miDiagnostics Computer for analysis by the miDiagnostics Software. The miDiagnostics HSV-1&2 CSF Algorithm (hosted by miDiagnostics Software) automatically interprets the fluorescence signals and reports the test results.
Interpretation of Results by the miDiagnostics HSV-1&2 Algorithm:
The miDiagnostics Software automatically analyses, interprets and displays the test results on the screen as: "Detected", "Not Detected" for HSV-1 and HSV-2 and "invalid", as shown in below in the Table 1.
Table 1: miDiagnostics HSV-1&2 CSF Test results interpretation
| Results | Interpretation | Action to be taken |
| --- | --- | --- |
| HSV-1 Not Detected | HSV-1 Not Detected in the specimen | Report results |
| HSV-2 Not Detected | HSV-2 Not Detected in the specimen | |
| HSV-1 Detected | HSV-1 Detected in the specimen | Report results |
| HSV-2 Not Detected | HSV-2 Not Detected in the specimen | |
| HSV-1 Not Detected | HSV-1 Not Detected in the specimen | Report results |
K250050 - Page 4 of 18
{5}
K250050 - Page 5 of 18
| HSV-2 Detected | HSV-2 Detected in the specimen | |
| --- | --- | --- |
| HSV-1 Detected | HSV-1 Detected in the specimen | Report results |
| HSV-2 Detected | HSV-2 Detected in the specimen | |
| Invalid* | Presence or absence of HSV-1 and HSV-2 cannot be established of a test failure or due to the absence or amplification of each target (HSV-1, HSV-2 and SPC) after completion of the PCR run | A retest is required starting from ‘Step 1: Prepare the sample’ of the test procedure. |
## C Instrument Description Information:
1. Instrument Name: miDiagnostics PCR Platform
2. Specimen Identification: Manual
3. Specimen Sampling and Handling: Manual
4. Calibration: NA
5. Quality Control:
Sample Processing Control (Internal Control): The miDiagnostics HSV-1&2 CSF Test contains a sample processing control (SPC) which is introduced into a sample with the miDiagnostics HSV-1&2 CSF Test process, instrument, or reagent failures and PCR inhibition. Positive and/or negative results for HSV-1 and HSV-2 are considered valid and consequently reported when a SPC result is positive for the HSV-1 and HSV-2. The SPC is required to be detected for a valid negative test result of HSV-1 and HSV-2. The SPC is human genomic DNA (RPP30) purified from buffy coat of blood, purchased from a commercial source.
Recommended External Controls: External controls, Positive Sample Control (PSC) and Negative Sample Control (NSC), are not provided with the miDiagnostics HSV-1&2 CSF Test but are required to run the test (NSC should be pooled human CSF material from healthy donors).
## V Substantial Equivalence Information:
A Predicate Device Name(s):
Simplexa HSV1 & 2 Direct
B Predicate 510(k) Number(s):
K133621
## C Comparison with Predicate(s):
| Device & Predicate Device(s): | K250050
(Candidate) | K133621
(Predicate) |
| --- | --- | --- |
| Device Trade Name | miDiagnostics HSV-1&2 CSF Test | Simplexa HSV-1& 2 Direct |
{6}
K250050 - Page 6 of 18
| General Device Characteristic Similarities | | |
| --- | --- | --- |
| Intended Use/Indications For Use | The miDiagnostics HSV-1&2 CSF Test is an in-vitro polymerase chain reaction (PCR) assay intended for use with the miDiagnostics PCR Platform for the qualitative detection and differentiation of HSV-1 and HSV-2 DNA in cerebrospinal fluid (CSF) samples from patients suspected of Herpes Simplex Virus (HSV) infection of the central nervous system (CNS). This test is intended as an aid in the diagnosis of HSV-1 and HSV-2 infection in the CNS.
Negative results do not preclude HSV-1 or HSV-2 infection and should not be used as the sole basis for treatment or other patient management decisions.
The test is not intended for use as a donor screening test.
The test is for professional use only. | The Focus Diagnostics Simplexa HSV 1 & 2 Direct is intended for use on the 3M Integrated Cycler instrument for the qualitative detection and differentiation of HSV-1 and HSV-2 DNA in cerebrospinal fluid (CSF) samples from patients suspected of Herpes Simplex Virus (HSV) infections of the central nervous system (CNS). This test is intended as an aid in the diagnosis of HSV-1 and HSV-2 infections of the CNS.
Negative results do not preclude HSV-1 or HSV-2 infection and should not be used as the sole basis for treatment or other patient management decisions.
The assay is not intended for use as a donor screening test.
The assay is for professional use only.
The Positive Control is intended to be used as a control with the Simplexa HSV 1 & 2 Direct. This control is not intended for use with other assays or systems. |
| Test Principle/Technology | Real -Time Polymerase Chain Reaction | Same |
| Qualitative | Yes | Same |
{7}
| Analyte detected | Viral DNA from HSV-1 and HSV-2 | Same |
| --- | --- | --- |
| Specimen Type | Cerebrospinal fluid (CSF) | Same |
| Automated Analysis | Yes | Same |
| Quality Control (Internal control) | Assay contains Sample Processing Control (SPC) to detect reagent failure/or inhibition | Same |
| General Device Characteristic Differences | | |
| Instrument | miDiagnostics PCR Platform | 3M Integrated Cycler Studio system |
| External Controls (Positive and Negative) | Required but not provided | Provided |
VI Standards/Guidance Documents Referenced:
Not applicable
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
1a. Within-Laboratory Precision:
The miDiagnostics HSV-1&2 CSF Test within-laboratory Precision was evaluated including intermediate precision (within lab) and within run repeatability. The intermediate study included 3 panel members 2x LoD (low positive), 5x LoD (moderate positive) and negative samples. The positive panel members were prepared by spiking HSV-1 (HF strain) and HSV-2 (MS strain) virus stocks into pooled negative human CSF matrix.
Each panel member was tested in 2 replicates by 2 operators on 2 miDiagnostic PCR platforms for 12 days, with 2 runs per day per operator, using 2 different lots of reagents for a total of 96 replicates of each panel member (2 replicates x 2 runs per day x 2 operators x 12 days = 96 total replicates).
The within-laboratory precision results for HSV-1 and HSV-2 are summarized below in Table 2.
K250050 - Page 7 of 18
{8}
Table 2. miDiagnostics HSV-1&2 CSF Test - Within-laboratory Precision
| Analyte | Sample | Percent Agreement with Expected Results | Mean Cq | Within Run | | Between Run | | Between Day | | Within laboratorya | | Between Operator | | Totalb | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| HSV-1 HF strain | Low Positive 2x LoD | 96/96 (100%) | 34.84 | 0.61 | 1.75 | 1.06 | 3.05 | 0.00 | 0.00 | 1.23 | 3.52 | 0.00 | 0.00 | 1.23 | 3.52 |
| | Moderate Positive 5x LoD | 96/96 (100%) | 32.94 | 0.51 | 1.54 | 0.98 | 2.97 | 0.00 | 0.00 | 1.10 | 3.34 | 0.00 | 0.00 | 1.10 | 3.34 |
| | Negative | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| HSV-2 MS strain | Low Positive 2x LoD | 95/96 (98.96%) | 35.92 | 0.79 | 2.19 | 0.99 | 2.75 | 0.00 | 0.00 | 1.26 | 3.51 | 0.00 | 0.00 | 1.26 | 3.51 |
| | Moderate Positive 5x LoD | 96/96 (100%) | 34.53 | 0.71 | 2.07 | 1.11 | 3.21 | 0.00 | 0.00 | 1.32 | 3.82 | 0.00 | 0.00 | 1.32 | 3.82 |
| | Negative | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
aWithin Laboratory includes Within Run, Between Run, and Between Day Components
bTotal includes Within-Run, Between-Run, Between-Day, and Between-Instrument/Lot/Operator Components.
# 1b. Reproducibility Study (multi-site precision):
The reproducibility of miDiagnostics HSV-1&2 CSF Test was evaluated by testing a 4 panel member panel, including a 2x LoD (low positive), 5x LoD (moderate positive), 10x LoD (high positive) and 1x LoD (high negative). The positive panel members were prepared by spiking HSV-1 (HF strain) and HSV-2 (MS strain) virus stocks into pooled negative human CSF matrix. Each panel member was tested with 3 replicates at 3 sites, 2 operators for 5 days (3 replicates x 3 sites x 2 operators x 5 days = 90 total replicates). The study was carried out at three sites (two external and one internal) with 2 operators, 2 lots of reagents, and 3 miDiagnostics PCR Platforms per site across 5 non-consecutive days. Each panel member was tested with 3 replicates at 3 sites, 2 operators for 5 days (3 replicates x 3 sites x 2 operators x 5 days = 90 total replicates). The reproducibility results of the miDiagnostics HSV-1&2 CSF Test for HSV-1 and HSV-2 are summarized below in the Table 3.
Table 3. miDiagnostics HSV-1&2 CSF Test - Reproducibility
| Analyte | Panel Member | Percent Agreement with Expected Results | Mean Cq | Within Run | | Between Run | | Between Day | | Between Lot | | Between Site | | Totala | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| HSV-1 HF strain | Low Positive 2x LoD | 90/90 (100%) | 35.16 | 1.06 | 3.02 | 0.58 | 1.65 | 0.00 | 0.00 | 0.30 | 0.86 | 0.51 | 1.44 | 1.35 | 3.83 |
| | Moderate Positive 5x LoD | 90/90 (100%) | 33.41 | 0.96 | 2.87 | 0.00 | 0.00 | 0.52 | 1.56 | 0.52 | 1.56 | 0.24 | 0.72 | 1.23 | 3.69 |
| | High Positive | 88/88 (100%) | 31.48 | 0.83 | 2.63 | 0.00 | 0.00 | 0.53 | 1.69 | 0.11 | 0.35 | 0.19 | 0.61 | 1.01 | 3.20 |
K250050 - Page 8 of 18
{9}
K250050 - Page 9 of 18
| Analyte | Panel Member | Percent Agreement with Expected Results | Mean Cq | Within Run | | Between Run | | Between Day | | Between Lot | | Between Site | | Total* | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| | 10x LoDb | | | | | | | | | | | | | | |
| | High Negative | 50/90 (55.56%) | 38.16 | 1.34 | 3.50 | 1.20 | 3.15 | 0.00 | 0.00 | 0.40 | 1.05 | 0.00 | 0.00 | 1.84 | 4.82 |
| HSV-2 MS strain | Low Positive 2x LoD | 90/90 (100%) | 36.36 | 1.62 | 4.46 | 0.00 | 0.00 | 0.33 | 0.91 | 0.29 | 0.80 | 0.26 | 0.72 | 1.70 | 4.68 |
| | Moderate Positive 5x LoD | 90/90 (100%) | 34.56 | 0.79 | 2.28 | 0.95 | 2.74 | 0.17 | 0.49 | 0.27 | 0.78 | 0.23 | 0.66 | 1.30 | 3.75 |
| | High Positive 10x LoD | 88/88 (100%) | 32.74 | 0.85 | 2.59 | 0.00 | 0.00 | 0.61 | 1.88 | 0.00 | 0.00 | 0.21 | 0.63 | 1.07 | 3.26 |
| | High Negative | 33/90 (36.67%) | 38.72 | 2.11 | 5.45 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 2.11 | 5.45 |
*Total includes Within-Run/Day, Between-Run/Day, Between-Lot, and Between-Site/Instrument Components.
bFor site 1 and site 2, one relicate of the high positive sample type was excluded from the analysis due to operator errors
## 2. Linearity:
Not applicable
## 3. Analytical Specificity/Interference:
3a. Potential Cross-reactivity: A study was performed to evaluate the performance of the miDiagnostics HSV-1&2 CSF Test in the presence of 46 potential cross-reactivity microorganisms that are closely related, cause similar clinical symptoms, or may be present in CSF. The potential cross-reactants were spiked into negative human CSF at high concentrations and tested in triplicate. The microorganisms were tested at $10^{5}$ Units/mL for viruses and fungi, and $10^{6}$ Units/mL for bacteria or the highest titer available. No cross-reactivity was observed in any of the microorganisms.
Table 4. miDiagnostics HSV-1&2 CSF Test – Potential Cross-reactivity
| Potential Cross-Reactant Organism | Tested Concentration | Qualitative Result (#Detected / #Total) | |
| --- | --- | --- | --- |
| | | HSV-1 | HSV-2 |
| Adenovirus (type 1) | 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| Adenovirus (type 7A) | 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| BK polyomavirus | 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| Citrobacter freundii | 10^{6} CFU/mL | 0/3 | 0/3 |
| Citrobacter koseri | 10^{6} CFU/mL | 0/3 | 0/3 |
| Cryptococcus neoformans | 10^{6} CFU/mL | 0/3 | 0/3 |
{10}
K250050 - Page 10 of 18
| Cytomegalovirus (AD169) | 3.12 x 10^{3} TCID_{50}/mL | 0/3 | 0/3 |
| --- | --- | --- | --- |
| Dengue virus | 10^{5} copies/mL | 0/3 | 0/3 |
| Enterobacter aerogenes | 10^{6} CFU/mL | 0/3 | 0/3 |
| Enterovirus 71 | 7.78 x 10^{3} TCID_{50}/mL | 0/3 | 0/3 |
| Epstein Barr virus (B95-8) | 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| Escherichia coli | 10^{6} CFU/mL | 0/3 | 0/3 |
| Haemophilus influenzae | 10^{6} CFU/mL | 0/3 | 0/3 |
| Haemophilus influenzae type b (MinnA) | 10^{6} CFU/mL | 0/3 | 0/3 |
| Hepatitis B | 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| Hepatitis C1 | 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| HIV1 (type IIIB) | 10^{4} IU/mL | 0/3 | 0/3 |
| Human herpesvirus 6 | 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| Human herpesvirus 7 | 9.36 x 10^{3} TCID_{50}/mL | 0/3 | 0/3 |
| Human herpesvirus 8 | 1.45 x 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| Influenza B | 2.80 x 10^{4} TCID_{50}/mL | 0/3 | 0/3 |
| Influenza A H1N1 | 1.46 x 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| John Cunningham virus | 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| Klebsiella pneumoniae | 10^{6} CFU/mL | 0/3 | 0/3 |
| Listeria monocytogenes | 10^{6} CFU/mL | 0/3 | 0/3 |
| Measles | 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| Mumps | 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| Mycobacterium tuberculosis | 10^{6} copies/mL | 0/3 | 0/3 |
| Naegleria fowleri | 10^{6} copies/mL | 0/3 | 0/3 |
| Neisseria meningitides (serogroup A) | 10^{6} CFU/mL | 0/3 | 0/3 |
| Parainfluenza Virus 1 | 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| Parainfluenza Virus 2 | 8.34 x 10^{4} TCID_{50}/mL | 0/3 | 0/3 |
| Parainfluenza Virus 3 | 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
| Parvovirus B19 | 1.28 x 10^{2} TCID_{50}/mL | 0/3 | 0/3 |
| Proteus mirabilis (Z050) | 10^{6} CFU/mL | 0/3 | 0/3 |
| Pseudomonas aeruginosa | 10^{6} CFU/mL | 0/3 | 0/3 |
| Rabies virus | Not available * | 0/3 | 0/3 |
| Rhinovirus (Type 1A) | 2.82 x 10^{4} TCID_{50}/mL | 0/3 | 0/3 |
| Rotavirus (Type Wa) | 10^{5} TCID_{50}/mL | 0/3 | 0/3 |
{11}
| Rubella | 2.58 x 103TCID50/mL | 0/3 | 0/3 |
| --- | --- | --- | --- |
| St. Louis Encephalitis | 105copies/mL | 0/3 | 0/3 |
| Staphylococcus aureus COL | 106CFU/mL | 0/3 | 0/3 |
| Streptococcus agalactiae | 106CFU/mL | 0/3 | 0/3 |
| Streptococcus pneumoniae Z022; 19F | 106CFU/mL | 0/3 | 0/3 |
| Varicella zoster virus | 105TCID50/mL | 0/3 | 0/3 |
| West Nile Virus | 105TCID50/mL | 0/3 | 0/3 |
*No official concentration was provided by the supplier. RNA extracted from the maximum volume of pathogen was spiked for testing without exceeding 20% of the total sample volume.
Cross-reactivity with Polio virus and La Crosse encephalitis virus was evaluated by an in silico analysis using viral genome RNA sequences of these two viruses available in GenBank. There was no homology found between the primers and probes of the miDiagnostics HSV 1 &2 CSF test to either of the virus.
## 3b. Microbial Interference Study:
The susceptibility of the miDiagnostics HSV-1&2 CSF Test to microbial interference was evaluated using the same microorganisms listed above in Table 4. A total of 46 microorganisms were spiked either individually or as a pool into negative human CSF along with HSV-1 and HSV-2 at 3x LoD levels and tested in triplicate. The interfering microorganisms were tested at 10⁵ Units/mL for viruses and fungi, and 10⁶ Units/mL for bacteria or the highest titer available. All samples spiked with HSV-1 and HSV-2 spiked were reported positive for HSV-1 and HSV-2, in the presence of the potentially interfering microorganisms. There was no interference observed for either HSV-1 or HSV-2 due to the presence of the 46 microorganisms tested with the miDiagnostics HSV-1&2 CSF Test.
## 3c. Interfering Substances:
The potential interference of substances that may be present in CSF specimens was evaluated by testing eleven substances at the concentrations listed (above normal clinical concentrations) in the table below. A panel composed of eleven (11) substances were spiked individually into negative human CSF and were tested alone and in the presence of HSV-1 (HF strain), HSV-2 (MS strain) at 3x LoD levels. There was no evidence of interference (false positive or false negative results) caused by the substances tested at the concentrations shown below in the Table 5 below.
Table 5. miDiagnostics HSV-1&2 CSF Test - Interference
| Interfering Substance | Interferent Concentration | Clinically relevant Concentration |
| --- | --- | --- |
| Albumin | 10 mg/mL | 0.32 - 2.63 mg/mL |
| Lactate | 2.2 mg/mL | 0 - 0.6 mg/mL |
| Glucose | 9.9 mg/mL | 0.32 – 0.98 mg/mL |
K250050 - Page 11 of 18
{12}
| Casein | 5 mg/mL | 0.15 – 40 mg/mL |
| --- | --- | --- |
| Hemoglobin | 0.625 g/mL | <0.03 mg/mL |
| Immunoglobulin (IgG) | 10 mg/mL | 0.8 mg/mL |
| White Blood Cell (WBC) | 2x10^6ells/mL | 0.009 - 1x 10^6 cells/mL |
| Whole Blood | 5% (v/v) | 1% (v/v) |
| Topical Antiseptic – Betadin | 1% (v/v) | 0.25% (v/v) |
| Antiviral Drug (Acyclovir) | 2.5 mg/mL | 0.022 mg/mL |
| Acylovir Triphosphate | 0.5 mg/mL | Metabolized from up to 0.022 mg/mL |
# 3d. Competitive Interference:
A competitive interference study was performed to evaluate the performance of the miDiagnostics HSV-1&2 CSF Test in the presence of the 2 HSV-1 and HSV-2 target analytes. The interference between HSV-1 (HF strain) and HSV-2 (MS strain) targets with the miDiagnostics HSV-1&2 CSF Test was evaluated by spiking one of the targets at high concentration $(10^{5}\mathrm{TCID}50 / \mathrm{mL})$ and the other target at low concentration (3x LoD) into pooled negative human CSF. No interference between the targets was observed.
Table 6. miDiagnostics HSV-1&2 CSF Test - Competitive Interference
| Low conc. (3x LoD) | High Conc. | HSV-1 Positivity | HSV-1 Mean Cq (SD) | HSV-2 Positivity | HSV-2 Mean Cq (SD) |
| --- | --- | --- | --- | --- | --- |
| HSV-1 HF strain 60.96 TCID50/mL | HSV-2 HF strain 105TCID50/mL | (3/3) 100.0% | 36.76 (2.44) | (3/3) 100.0% | 18.35 (0.11) |
| HSV-2 MS strain 1.68 TCID50/mL | HSV-1 MS strain 105TCID50/mL | (3/3) 100.0% | 22.47 (0.19) | (3/3) 100.0% | 36.41 (1.30) |
# 4. Assay Reportable Range:
Not Applicable
# 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Not Applicable
# 6. Detection Limit:
The Limit of Detection (LoD) was determined for the miDiagnostics HSV-1&2 CSF Test in 2 phases:
Phase 1: LoD Range Finding - The Limit of Detection (LoD) for the miDiagnostics HSV-1&2 CSF Test was determined using two HSV 1 strains (MacIntyre and HF strains) and two HSV 2 strains (MS and G strains). The serial dilutions of each strain were made in pooled negative human CSF matrix (negative CSF). For each viral strain, tests were performed over
K250050 - Page 12 of 18
{13}
the course of 4 days, with 2 reagent batches and 9 miDiagnostics PCR Platforms. Ten replicates at each dilution series were tested per viral strain per reagent batch with a total of 20 replicates per strain. The concentration at which $\geq 95\%$ of replicates were detected, was the preliminary LOD for that strain.
Phase 2: LoD Confirmation - The preliminary LOD for each strain was subsequently confirmed with 20 replicates diluted at the preliminary LOD concentration. The confirmed LOD per strain are included in Table 7 below.
Table 7. miDiagnostics HSV-1&2 CSF Test - LoD
| Virus strain | LoD (TCID50/mL) | LoD (copies/mL) | Number of detected samples (% of detection) | Mean Cq (SD) (from detetected replicates) |
| --- | --- | --- | --- | --- |
| HSV-1 MacIntyre | 8.84 | 800 | 20/20 (100%) | 37.06 ± 0.98 |
| HSV-1 HF | 20.32 | 1000 | 20/20 (100%) | 36.68 ± 0.84 |
| HSV-2 MS | 0.56 | 400 | 19/20 (95%) | 37.99 ± 1.55 |
| HSV-2 G | 0.43 | 800 | 20/20 (100%) | 36.63 ± 0.86 |
# 7. Analytical Reactivity (Inclusivity):
The analytical reactivity of the miDiagnostics HSV-1&2 CSF Test was evaluated with four HSV-1 strains and three HSV-2 strains in addition to the strains tested for limit of detection (i.e., analytical sensitivity in Table 7 above). Viral material with known concentration was spiked into negative CSF at near LOD concentration and tested in triplicates at the LoD determined for the HSV-1 and HSV-2. The miDiagnostics HSV-1&2 CSF Test was able to detect all HSV strains: at 1x or 2x LOD concentration for HSV-1 strain F, HSV-1 strain KOS, HSV-1 ATCC-2011-1, HSV-1 strain 95 (95/1906), HSV-2 strain 131596 and HSV-2 ATCC-2011-2, and at 6x LOD for HSV-2 strain HG52.
Table 8. miDiagnostics HSV-1&2 CSF Test - Analytical reactivity
| Analyte | Strain/Isolate | Spiked Concentration | Inclusive (Yes/No) |
| --- | --- | --- | --- |
| HSV-1 | F | 20.32 TCID50/mL | Yes |
| | KOS | 40.64 TCID50/mL | Yes |
| | ATCC 2011-1 | 20.32 TCID50/mL | Yes |
| | Strain 95 (95/1906) | 20.32 TCID50/mL | Yes |
| HSV-2 | 131596 | 1605 copies/mL | Yes |
| | ATCC 2011-2 | 3.36 TCID50/mL | Yes |
| | HG52 | 0.56 TCID50/mL | Yes |
K250050 - Page 13 of 18
{14}
K250050 - Page 14 of 18
8. Assay Cut-Off:
Initially, the cut-off of the miDiagnostics HSV-1&2 Test was determined using the large datasets from clinical studies, in addition to natural and contrived samples. All studies were performed to a maximum of 46 Cq cycles of amplification during feasibility study.
The clinical cut-off was further validated using a total of 22 positive samples (11 spiked with HSV-1 and 11 with HSV-2) that were prepared by spiking either HSV-1 or HSV-2 into 11 individual negative clinical CSF specimens from symptomatic patients at or near the LoD levels. In addition, 10 negative samples from same individual CSF specimens were tested without spiking. Out of the 22 contrived samples, 10 resulted in positive call for HSV-1 while 1 resulted in a negative call, and 10 resulted in a positive call for HSV-2 while 1 resulted in a negative call. Based on the results of the verification study, HSV-1 was detected between Cq of 35.05 and 39.55 and HSV-2 was detected between Cq of 34.22 and 40.08.
The combined data analysis from clinical studies and contrived studies demonstrated that the cut-off algorithm implemented for the HSV CSF test achieved a positive concordance 96.3% for HSV-1 and 97.6% for HSV2 and a negative concordance of 98.5% for HSV-1 and 99.0% for HSV-2, therefore confirming its adequacy of the cut-off.
9. Accuracy (Instrument):
Not applicable
10. Carry-Over:
The risk of sample carry-over contamination was evaluated with the miDiagnostics HSV-1&2 Test by alternately testing high positive samples (HSV-1 and HSV-2 at ≥ 10⁵ pfu/mL) followed by negative sample on a single Reader. A total of 5 runs were performed, each by different operators. Within each run, eight cycles of a high positive sample followed by a negative sample were measured using a single instrument per operator per run. A total of 16 data points, including 8 measurements for high positive samples and 8 measurements for negative samples, were generated for each operator. All high positive samples resulted in "Detected" calls for both HSV-1 and HSV-2 and all negative samples resulted in "Not detected" calls. No carry-over contamination was observed throughout the study.
11. Specimen Stability:
The specimen stability study was executed by spiking HSV-1 and HSV-2 in pooled human CSF material obtained from deidentified healthy donors. The material was confirmed negative for HSV-1 (MacIntyre) and HSV-2 (MS). The positive specimens were prepared by spiking HSV-1 and HSV-2 at the concentration of 1-2x LoD and 3-4x LoD in pooled negative CSF matrix. The data supports the following CSF specimen storage conditions for testing with the miDiagnostics HSV-1&2 Test:
- Up to 24 hrs at room temperature (between 15°C to 30°C).
- Up to 7 days at 2-8 °C.
- Up to 30 days at -70°C.
12. Comparison Studies:
{15}
a. Method Comparison with Predicate Device:
See Section 13.C below.
b. Matrix Comparison:
Not applicable.
## 13. Clinical Studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical Specificity:
Not applicable
c. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Performance of the miDiagnostics HSV-1&2 CSF Test was established in both prospective (Cohort 1) and retrospective (Cohort 2) clinical studies at 4 participating testing sites across the United States and United Kingdom. The CSF specimens were either collected prospectively (Cohort 1) or originating from a biorepository, and thus retrospective (Cohort 2), and each CSF specimen originated from a unique subject. The specimens were tested with the miDiagnostics HSV-1&2 CSF Test and the comparator method.
A total of 362 cerebrospinal fluid (CSF) specimens were collected from patients presenting with signs and symptoms of Herpes Simplex Virus (HSV) central nervous system (CNS) infection at 4 participating sites between August 2024 and December 2024. A total of 325 specimens were included for the clinical performance of the miDiagnostics HSV-1&2 CSF Test. Thirty-seven (37) specimens did not meet the inclusion criteria and were excluded from the studies. The CSF specimens of individuals of all sexes, all ethnicities, and all ages, with signs and/or symptoms of HSV-1 or HSV-2 infection in the CNS, that were eligible according to the inclusion and exclusion criteria, were enrolled.
The gender and age demographic of individuals for prospective and retrospective studies are presented in Table 9 and Table 10 below.
Table 9. Prospective Study - Age and Gender Distribution of individuals
| Age Category | Female | | | | Male | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | HSV-1 | HSV-2 | Not detected | All | HSV-1 | HSV-2 | Not detected | All |
| 18-24 | 12 | 1 | 1 | 10 | 1 | 0 | 0 | 1 |
| 25-34 | 12 | 1 | 1 | 10 | 1 | 0 | 0 | 1 |
| 35-44 | 12 | 1 | 1 | 10 | 1 | 0 | 0 | 1 |
| 45-54 | 12 | 1 | 1 | 10 | 1 | 0 | 0 | 1 |
| 55-64 | 12 | 1 | 1 | 10 | 1 | 0 | 0 | 1 |
| 65-74 | 12 | 1 | 1 | 10 | 1 | 0 | 0 | 1 |
| 75-84 | 12 | 1 | 1 | 10 | 1 | 0 | 0 | 1 |
| 85-94 | 12 | 1 | 1 | 10 | 1 | 0 | 0 | 1 |
K250050 - Page 15 of 18
{16}
K250050 - Page 16 of 18
| From birth to 1 year | 0 | 0 | 4 | 4 | 0 | 0 | 3 | 3 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| >1 to 2 years | 0 | 0 | 2 | 2 | 0 | 0 | 0 | 0 |
| >2 to 12 years | 0 | 0 | 3 | 3 | 0 | 0 | 2 | 2 |
| >12 to 21 years | 0 | 0 | 4 | 4 | 0 | 0 | 4 | 4 |
| >21 to 60 years | 1 | 1 | 41 | 43 | 1 | 1 | 35 | 37 |
| >60 years | 0 | 0 | 35 | 35 | 0 | 1 | 35 | 36 |
| Total | 1 | 1 | 89 | 91 | 1 | 2 | 79 | 82 |
Table 10. Retrospective Study - Age and Gender Distribution of individuals
| Age Category | Female | | | | Male | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | HSV-1 | HSV-2 | Not detected | All | HSV-1 | HSV-2 | Not detected | All |
| From birth to 1 year | 0 | 0 | 2 | 2 | 0 | 0 | 2 | 2 |
| >1 to 2 years | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| >2 to 12 years | 0 | 0 | 2 | 2 | 0 | 0 | 1 | 1 |
| >12 to 21 years | 0 | 1 | 2 | 3 | 0 | 1 | 2 | 3 |
| >21 to 60 years | 1 | 19 | 34 | 54 | 4 | 5 | 24 | 33 |
| >60 years | 4 | 3 | 16 | 23 | 10 | 2 | 17 | 29 |
| Total | 5 | 23 | 56 | 84 | 14 | 8 | 46 | 68 |
Performance of the miDiagnostics HSV-1&2 Test is presented as Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) with respect to the comparator method in the following tables below.
{17}
Table 11. HSV-1 Prospective Samples
| Candidate Device | Comparator Method | | Total |
| --- | --- | --- | --- |
| | Detected | Not Detected | |
| Detected | 0 | 2* | 2 |
| Not Detected | 0 | 171 | 171 |
| Total | 0 | 173 | 173 |
| PPA | PPA: N/A | | |
| NPA | NPA: 171/173 (98.84%); (95% CI: 95.88% -99.68%) | | |
*Two false positive specimens were reported to be HSV-1 negative by Standard of Care PCR testing. No clinical diagnoses were available.
N/A: Not Applicable.
Table 12. HSV-1 Retrospective Samples
| Candidate Device | Comparator Method | | Total |
| --- | --- | --- | --- |
| | Detected | Not Detected | |
| Detected | 16 | 3* | 19 |
| Not Detected | 0 | 133 | 133 |
| Total | 16 | 136 | 152 |
| PPA | PPA: 16/16 (100.00%); (95% CI: 80.64% - 100.00%) | | |
| NPA | NPA: 133/136 (97.79%); (95% CI: 93.72% -99.25%) | | |
*All 3 false positive specimens were reported to be HSV-1 negative by Standard of Care PCR testing. Two of them were clinically diagnosed with HSV-1 infection, and one subject was clinically diagnosed with Herpes Meningoencephalitis
Table 13. HSV-2 Prospective Samples
| Candidate Device | Comparator Method | | Total |
| --- | --- | --- | --- |
| | Detected | Not Detected | |
| Detected | 2 | 1* | 3 |
| Not Detected | 0 | 170 | 171 |
| Total | 2 | 171 | 173 |
| PPA | PPA: 2/2 (100.00%); (95% CI: 34.24% -100.00%) | | |
| NPA | NPA: 170/171 (99.42%); (95% CI: 96.76% -99.90%) | | |
*One false positive specimen was reported to be HSV-2 negative by Standard of Care PCR testing. No clinical diagnosis was available.
Table 14. HSV-2 Retrospective Samples
| Candidate Device | Comparator Method | | Total |
| --- | --- | --- | --- |
| | Detected | Not Detected | |
| Detected | 29 | 2* | 31 |
| Not Detected | 0 | 121 | 121 |
K250050 - Page 17 of 18
{18}
| Total | 29 | 123 | 152 |
| --- | --- | --- | --- |
| PPA | PPA: 29/29 (100.00%); (95% CI 88.31% - 100.00%) | | |
| NPA | NPA: 121/123 (98.37%); (95% CI: 94.27% -99.55%) | | |
*The 2 false positive specimens were reported to be HSV-2 positive by Standard of Care PCR testing. One subject was clinically diagnosed with Herpes Meningitis. No clinical diagnosis was available for the other subject.
14. Clinical Cut-Off:
Not applicable.
15. Expected Values/Reference Range:
Not applicable
16. Other Supportive Instrument Performance Characteristics Data:
Not applicable
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K250050 - Page 18 of 18
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.