The HeroTracker Sense Device (HTS) is an add-on device which is attached by patients on a metered-dose inhaler (MDI). HTS is intended for single-patient multiple use in the home environment as an electronic data capture accessory for monitoring and recording actuation, inhaler shake, inhaler orientation, inhalation coordination, and inspiratory strength and duration for prescribed inhaler usage. HTS may be used in the following applications: in clinical practice or clinical trials, where specialists, general practitioners, nurses and educators need to know if a patient has used their prescribed medication or assess inhaler technique. HTS is designed to work only with the MDI and medication indicated on the HTS label. HTS is not intended to indicate the remaining quantity of medication in an inhaler and does not include a dose counting function. HTS is not intended to provide spirometry measurements. HTS is intended to be used by Metered Dose Inhaler (MDI) users aged 12 years and over.
Device Story
HTS is an add-on electronic accessory for MDIs; attaches to canister top. Sensors (force, acceleration, airflow) record date/time of usage, inhaler shake, orientation, and inspiratory parameters. Data transmits via Bluetooth Low Energy to mobile application; displayed to patients/clinicians. Used in home, clinic, or clinical trials by patients (12+ years) or healthcare providers. Provides objective data on medication adherence and technique; assists clinicians in assessing patient inhaler use. Does not count doses or provide spirometry. Powered by non-rechargeable CR2032 battery; 1-year service life.
Clinical Evidence
No clinical data was necessary to determine substantial equivalence. Evidence is based on non-clinical bench testing, including biocompatibility (ISO 10993-1, ISO 18562-1), electrical safety (IEC 60601 series), electromagnetic disturbance (IEC 60601-1-2), software verification/validation (IEC 62304), and transportation testing (ASTM 4169).
Technological Characteristics
Materials: ABS (cover), PC (light diffuser). Sensing: Force, acceleration, and airflow sensors. Connectivity: Bluetooth 4.0 (2.40-2.48 GHz, 1.0 mW). Power: CR2032 lithium coin cell. Form factor: MDI add-on. Sterility: Non-sterile. Software: Verified per IEC 62304.
Indications for Use
Indicated for MDI users aged 12+ for monitoring/recording inhaler usage (actuation, shake, orientation, coordination, inspiratory strength/duration). Used in home, clinical practice, or clinical trials to assess medication adherence and inhaler technique. Compatible only with specific MDI medications listed on the label.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 30, 2025
Voluntis
% Rachel Paul
Lead Quality and Regulatory Affairs Consultant
Emergo Global Consulting
LLC 2500 Bee Cave Road Building 1
Austin, Texas 78746
Re: K250022
Trade/Device Name: HeroTracker Sense
Regulation Number: 21 CFR 868.5630
Regulation Name: Nebulizer
Regulatory Class: Class II
Product Code: CAF
Dated: January 3, 2025
Received: January 3, 2025
Dear Rachel Paul:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250022 - Rachel Paul
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250022 - Rachel Paul
Page 3
Sincerely,
John S. Bender -S
2025.06.30 12:39:13 -04'00'
for Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250022
Device Name
HeroTracker Sense
Indications for Use (Describe)
The HeroTracker Sense Device (HTS) is an add-on device which is attached by patients on a metered-dose inhaler (MDI).
HTS is intended for single-patient multiple use in the home environment as an electronic data capture accessory for monitoring and recording actuation, inhaler shake, inhaler orientation, inhalation coordination, and inspiratory strength and duration for prescribed inhaler usage.
HTS may be used in the following applications: in clinical practice or clinical trials, where specialists, general practitioners, nurses and educators need to know if a patient has used their prescribed medication or assess inhaler technique.
HTS is designed to work only with the MDI and medication indicated on the HTS label.
HTS is not intended to indicate the remaining quantity of medication in an inhaler and does not include a dose counting function.
HTS is not intended to provide spirometry measurements.
HTS is intended to be used by Metered Dose Inhaler (MDI) users aged 12 years and over.
Compatible Medication - Strength and Dose Account
Ventolin 90 mcg HFA (60 count)
Advair 45 mcg / 21 mcg (120 count)
Advair 115 mcg / 21 mcg (120 count)
Advair 230 mcg / 21 mcg (120 count)
Albuterol Sulfate (Prasco) 90 mcg (200 count)
Fluticasone Propionate & Salmeterol 45 mcg / 21 mcg (120 count)
Fluticasone Propionate & Salmeterol 115 mcg / 21 mcg (120 count)
Fluticasone Propionate & Salmeterol 230 mcg / 21 mcg (120 count)
Albuterol Sulfate (Lupin) 90 mcg (200 count)
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}
K250022
Page 1 of 7
# 510(k) Summary
# HeroTracker Sense (HTS)
## 1. Submission Sponsor
- Voluntis
- 58 avenue de Wagram
- 75017 Paris
- France
- Sana Charnine, Regulatory Affairs and Quality Manager, Aptar Digital Health
## 2. Submission Correspondent
- Emergo by UL
- 2500 Bee Cave Road
- Building 1, Suite 300
- Austin, TX 78746
- Office Phone: (512) 327-9997
- Email: LST.US.EmergoFDASubmissions@ul.com
- Contact: Rachel Paul
- Title: Lead Quality and Regulatory Affairs Consultant
## 3. Date Prepared
30 June 2025
## 4. Device Identification
- Trade/Proprietary Name: HeroTracker® Sense
- Common/Usual Name: Nebulizer Accessory
- Regulation Number(s): 868.5630
- Product Code(s): CAF, Nebulizer (Direct Patient Interface)
- Class: Class 2
- Classification Panel: Anesthesiology
## 5. Legally Marketed Predicate Device(s)
- Device name: Hailie® Sensor NF0110
- 510(k) number: K222247
- Manufacturer: Adherium (NZ) Ltd
Page 1 of 7
{5}
K250022
Page 2 of 7
## 6. Device Description
The HeroTracker Sense (HTS) is a nebulizer accessory. It is an add-on device which is attached by patients on a metered dose inhaler (MDI) and is used to record and analyze data related to medication actuation and technique of use for prescribed inhaler usage. Data is transferred to a mobile application with appropriate settings and is displayed to end users.
## 7. Indication for Use Statement
The HeroTracker Sense Device (HTS) is an add-on device which is attached by patients on a metered-dose inhaler (MDI).
HTS is intended for single-patient multiple use in the home environment as an electronic data capture accessory for monitoring and recording actuation, inhaler shake, inhaler orientation, inhalation coordination, and inspiratory strength and duration for prescribed inhaler usage.
HTS may be used in the following applications: in clinical practice or clinical trials, where specialists, general practitioners, nurses and educators need to know if a patient has used their prescribed medication or assess inhaler technique.
HTS is designed to work only with the MDI and medication indicated on the HTS label.
HTS is not intended to indicate the remaining quantity of medication in an inhaler and does not include a dose counting function.
HTS is not intended to provide spirometry measurements.
HTS is intended to be used by Metered Dose Inhaler (MDI) users aged 12 years and over.
| Compatible Medication | Strength and Dose Account |
| --- | --- |
| Ventolin | 90 mcg HFA (60 count) |
| Advair | 45 mcg / 21 mcg (120 count) |
| Advair | 115 mcg / 21 mcg (120 count) |
| Advair | 230 mcg / 21 mcg (120 count) |
| Albuterol Sulfate (Prasco) | 90 mcg (200 count) |
| Fluticasone Propionate & Salmeterol | 45 mcg / 21 mcg (120 count) |
| Fluticasone Propionate & Salmeterol | 115 mcg / 21 mcg (120 count) |
| Fluticasone Propionate & Salmeterol | 230 mcg / 21 mcg (120 count) |
| Albuterol Sulfate (Lupin) | 90 mcg (200 count) |
## 8. Substantial Equivalence Discussion
The following table compares the HTS to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for
Page 2 of 7
{6}
K250022
Page 3 of 7
the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.
Page 3 of 7
{7}
K250022
Page 4 of 7
Comparison of Characteristics
| Attribute | Subject: HeroTracker Sense® v1.2 | | Predicate: Hailie® Sensor NF0110 K222247 | Comparison |
| --- | --- | --- | --- | --- |
| Product Code | CAF | | CAF | Same |
| Regulation Number | 868.5630 | | 868.5630 | Same |
| Intended for Use | Nebulizer accessory for recording nebulizer use and technique | | Nebulizer accessory for recording nebulizer use and technique | Same |
| Indications for Use | The HeroTracker Sense Device (HTS) is an add-on device which is attached by patients on a metered-dose inhaler (MDI).
HTS is intended for single-patient multiple use in the home environment as an electronic data capture accessory for monitoring and recording actuation, inhaler shake, inhaler orientation, inhalation coordination, and inspiratory strength and duration for prescribed inhaler usage.
HTS may be used in the following applications: in clinical practice or clinical trials, where specialists, general practitioners, nurses and educators need to know if a patient has used their prescribed medication or assess inhaler technique.
HTS is designed to work only with the MDI and medication indicated on the HTS label.
HTS is not intended to indicate the remaining quantity of medication in an inhaler and does not include a dose counting function.
HTS is not intended to provide spirometry measurements.
HTS is intended to be used by Metered Dose Inhaler (MDI) users aged 12 years and over. | | The Hailie sensor is intended for single-patient multiple use in the home environment as an electronic data capture accessory for monitoring and recording actuation, inspiratory flow and inhaler shake for prescribed inhaler usage.
The Hailie sensor may be used in the following applications: in clinical practice or clinical trials, where specialists, general practitioners, nurses and educators need to know if a patient has used their prescribed medication or assess inspiratory flow and inhaler technique; and in patient self-management.
The Hailie sensor is compatible only with the following TEVA MDI inhalers: ProAir® HFA & Albuterol Sulphate HFA. The Hailie sensor is not intended to indicate the remaining quantity of medication in an inhaler and does not include a dose counting function.
The Hailie sensor is not intended to provide spirometry measurements. | Similar; the differences do not constitute a new intended for use or bring up new questions related to safety or effectiveness. |
| | Compatible Medication | Strength and Dose Account | | |
| | Ventolin | 90 mcg HFA (60 count) | | |
Page 4 of 7
{8}
K250022
Page 5 of 7
| Attribute | Subject: HeroTracker Sense® v1.2 | | Predicate: Hailie® Sensor NF0110 K222247 | | Comparison |
| --- | --- | --- | --- | --- | --- |
| | Advair | 45 mcg / 21 mcg (120 count) | | | |
| | Advair | 115 mcg / 21 mcg (120 count) | | | |
| | Advair | 230 mcg / 21 mcg (120 count) | | | |
| | Albuterol Sulfate (Prasco) | 90 mcg (200 count) | | | |
| | Fluticasone Propionate & Salmeterol | 45 mcg / 21 mcg (120 count) | | | |
| | Fluticasone Propionate & Salmeterol | 115 mcg / 21 mcg (120 count) | | | |
| | Fluticasone Propionate & Salmeterol | 230 mcg / 21 mcg (120 count) | | | |
| | Albuterol Sulfate (Lupin) | 90 mcg (200 count) | | | |
| Design – Attachment to Medication Dispenser | Physically attached to dispenser without inhibiting patient use | | Physically attached to dispenser without inhibiting patient use | | Same |
| Principle of Operation | Attaches to the top of the medication canister and performs wireless uploading of usage history of the inhaler | | Attaches to the top of the medication canister and performs wireless uploading of usage history of the inhaler | | Same |
| Data Collection Technology | Records date and time of MDI usage and inhaler usage technique through sensors (force, acceleration, and airflow sensors). | | Records date and time of MDI usage and inhaler usage technique through sensors (optical, induction coil, motion and flow sensors). | | Similar; although the kind of sensors are not fully the same, the principle of operation is still the same. |
| Records Usage | Yes | | Yes | | Same |
| Dose Counter | No | | No | | Same |
| Patient Data Storage with Software | Yes | | Yes | | Same |
| Patient Data Report | Yes | | Yes | | Same |
Page 5 of 7
{9}
K250022
Page 6 of 7
| Attribute | Subject: HeroTracker Sense® v1.2 | Predicate: Hailie® Sensor NF0110 K222247 | Comparison |
| --- | --- | --- | --- |
| Generation with Software | | | |
| Mobile Platforms | iOS 13 and higher version (13, 14, 15) Android 8 and higher version (8, 9, 10 11, 12). | iOS 9 and above Android 5.1 and above | Similar; both devices are available with recent versions of iOS and Android. |
| Required Off-the Shelf Hardware | Apple smartphones or devices with Bluetooth | Apple smartphones or devices with Bluetooth | Same |
| Wireless Technology Bluetooth | Bluetooth® 4.0: 2.40 - 2.48 GHz, 1.0 mW Low Energy | Bluetooth® 4.0: 2.40 - 2.48 GHz, 1.0 mW Low Energy | Same |
| Power Source | Lithium coin cell CR2032 non-rechargeable battery | Lithium Coin Cells Non-rechargeable Battery | Same |
| Battery Life / Service Life | 1 year | 1 year | Same |
| Case Material – Patient Contact by Intact Skin (Hands) | Cover: ABS
Light diffuser: PC | Plastic cover | Similar; both devices are presented as a plastic external enclosure. |
| Electrical Safety | IEC 60601 | IEC 60601 | Same |
| Biocompatibility | ISO 10993-1 | ISO 10993-1 | Same |
| Sterility | Non-sterile | Non-sterile | Same |
Page 6 of 7
{10}
K250022
Page 7 of 7
## 9. Non-Clinical Performance Data
To demonstrate safety and effectiveness of HTS and to show substantial equivalence to the predicate device, Voluntis completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. The HTS passed the testing in accordance with internal requirements, national standards, and international standards shown below, supporting its safety and effectiveness, and its substantial equivalence to the predicate device:
- Biocompatibility evaluation per ISO 10993-1 - Materials can be considered as biocompatible.
- Electrical safety testing per IEC 60601-1, IEC 60601-1-11, IEC 60601-1-6 – Passed
- Electromagnetic Disturbance (EMD) testing per IEC 60601-1-2 – Passed
- Software verification and validation per IEC 62304/FDA Guidance – The correct functionality for the HTS software release, for all software modules and for association with the compatible mobile app. to display the data was verified
- Biocompatibility evaluation of breathing gas pathways in healthcare applications per ISO 18562-1: The biological safety of the gas pathway is supported by a risk-based evaluation and is considered adequately addressed.
- Performance Testing – Device was verified to meet specifications and the correct functionality and compatibility for the HTS with the concerned MDI inhalers was verified
- Service Life – Supports 1 year service life
- Shelf-Life Testing – Supports 2 years shelf life
- Transportation testing per ASTM 4169 – confirmed the transport has no impact on device.
## 10. Clinical Performance Data
No clinical data was necessary to determine the substantial equivalence of this device.
## 11. Statement of Substantial Equivalence
The HTS has the same intended use as the predicate device and similar technological characteristics. The differences in technological characteristics do not raise new or different questions of safety and effectiveness. Based on the results of the risk assessment and all applicable testing, the HTS is determined to perform as intended during normal anticipated usage and is at least as safe, as effective, and performs as well as or better than the legally marketed predicate device and therefore can be determined to be substantially equivalent to the predicate device.
Page 7 of 7
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.