EXTRON Series are a mobile fluoroscopic X-ray system with high output capacity, high thermal capacity and high resolution image processing system, which provides X-ray images of the patient's anatomy during surgery or treatment. This device plays an important role in emergency injury treatment, orthopedic surgery, neurosurgery surgery, bone surgery, etc. This device has a function to save important a specific images as records, so you can easily search for the images and transmit it to the PACS system in the hospital to help the medical staff in diagnosis. Examples of a clinical application may include: Neurosurgery, Orthopedics, Anesthesiology, Urology, Gynecology, Internal Medicine (※ This device is not intended for mammography applications.)
Device Story
Mobile fluoroscopic X-ray system; C-arm main body and monitor cart. Inputs: X-ray photons generated by tube, collimated, and captured by flat panel detector. Process: Two-step conversion (photons to light, light to electrical signals); signals digitized, processed, and displayed on monitor. Operated by healthcare professionals in clinical settings (OR, emergency). Output: Real-time X-ray images for surgical/diagnostic guidance; image storage and PACS transmission. Benefits: Real-time anatomical visualization during interventional procedures; improved diagnostic workflow via digital archiving.
Clinical Evidence
No clinical studies performed. Evidence based on bench testing and comparative phantom image analysis. Anthropomorphic phantoms representative of indicated anatomies were imaged using subject and predicate devices; images reviewed by qualified clinical experts, confirming equivalent image quality.
Indicated for patients requiring X-ray radiography or fluoroscopy during surgery or treatment, including emergency, orthopedic, neurosurgery, and bone surgery. Applicable to adults and pediatrics. Not intended for mammography.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 10, 2025
DRTECH Corporation
% Minjeong Kim
Official Correspondent
Suite No.1, 2 Floor /Suite No. 2, 3 Floor, 29, Dunchon-Daero
541 Beon-Gil, Jungwon-Gu
SEONGNAM-SI, GYEONGGI 13216
SOUTH KOREA
Re: K250010
Trade/Device Name: Extron 3; Extron 5; Extron 7
Regulation Number: 21 CFR 892.1650
Regulation Name: Image-Intensified Fluoroscopic X-Ray System
Regulatory Class: Class II
Product Code: OWB, OXO, JAA
Dated: December 31, 2024
Received: May 9, 2025
Dear Minjeong Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250010 - Minjeong Kim
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250010 - Minjeong Kim
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Gabriela M. Rodal - for
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250010
Device Name
EXTRON 3;
EXTRON 5;
EXTRON 7
Indications for Use (Describe)
EXTRON Series are a mobile fluoroscopic X-ray system with high output capacity, high thermal capacity and high resolution image processing system, which provides X-ray images of the patient's anatomy during surgery or treatment. This device plays an important role in emergency injury treatment, orthopedic surgery, neurosurgery surgery, bone surgery, etc. This device has a function to save important a specific images as records, so you can easily search for the images and transmit it to the PACS system in the hospital to help the medical staff in diagnosis.
Examples of a clinical application may include: Neurosurgery, Orthopedics, Anesthesiology, Urology, Gynecology, Internal Medicine
(※ This device is not intended for mammography applications.)
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K250010
DRTECH
# 510(k) Summary
[As required by 21 CFR 807.92]
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92
# 1. Date Prepared [21 CFR 807.92(a) (1)]
12/31/2024
# 2. Submitter's Information [21 CFR 807.92(a) (1)]
- Name of Sponsor: DRTECH Corporation
- Address: Suite No.1, 2 Floor / Suite No. 2, 3 Floor, 29, Dunchon-daero541 beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, 13216, Republic of Korea
- Contact Name: Minjeong Kim
- Telephone No.: +82-31-779-7783
- Fax No.: +82-31-779-7790
- Email Address: drtechra@drtech.com
- Registration Number: 3005172103
- Name of Manufacturer: Same as Sponsor
# 3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]
- Trade Name: EXTRON 3, EXTRON 5, EXTRON 7
- Common Name: Mobile Fluoroscopic X-ray System
- Classification Name: Image-intensified fluoroscopic x-ray system
- Classification Panel: Radiology
- Classification Regulation: 21 CFR 892.1650
- Product Code: OWB, OXO, JAA
- Device Class: II
510(k) Summary
{5}
DRTECH
4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
| | Predicate Device | Reference 1 | Reference 2 |
| --- | --- | --- | --- |
| 510(k) Number: | K142708 | K122234 | K230871 |
| Applicant: | PHILIPS MEDICAL SYSTEMS NETHRLANDS BV | GE HEALTHCARE SURGERY GE OEC MEDICAL SYSTEMS, INC | DRTECH Corporation |
| Trade Name: | VERADIUS UNITY | OEC 9900 ELITE | EXTRON 5, EXTRON 7 |
| Classification Name: | Image-intensified fluoroscopic x-ray system | | |
| Classification Panel: | Radiology | | |
| Classification Regulation: | 21 CFR 892.1650 | | |
| Product Code: | OWB, JAA, OXO | | |
| Device Class: | II | | |
5. Description of the Device [21 CFR 807.92(a) (4)]
EXTRON Series are mobile fluoroscopic X-ray systems with high output capacity, high thermal capacity, and high-resolution image processing systems that provide X-ray images of the patient's anatomical structures during surgery or treatment. This device plays an important role in emergency injury treatment, orthopedic surgery, neurosurgery surgery, bone surgery, etc. This device has a function to save important a specific images as records, so you can easily search for the images and transmit it to the PACS system in the hospital to help the medical staff in diagnosis.
The EXTRON Series are composed of a C-arm main body and a monitor cart. The C-arm main body is composed of an X-ray tube, a flat panel detector, a collimator, a generator, a touch panel, foot switch, hand switch and an XConsoleOP program, while the monitor cart is composed of a monitor, a thermal transfer printer, a mouse, a keyboard and an XConsole program.
The operating principle of the device is designed to expose the patient to X-ray beams. The range of X-ray irradiation are adjusted by the collimator.
X-rays can penetrate into the human body through a two-step conversion process.
X-ray photons are converted into light. The light is then converted into electrical signals through the sensor. The electrical charges are transmitted as the sensor output and converted into signals. These signals are digitized and captured by memory. The captured images are processed and displayed on the monitor. The displayed images can be saved or transmitted to an external storage device, such as a network printer.
6. Indications for Use [21 CFR 807.92(a)(5)]
EXTRON Series are a mobile fluoroscopic X-ray system with high output capacity, high thermal capacity and high resolution image processing system, which provides X-ray images of the patient's anatomy during surgery or treatment. This device plays an important role in emergency injury treatment, orthopedic surgery, neurosurgery surgery, bone surgery, etc. This device has a function to save important a specific images as records, so you can easily search for the images and transmit it to the PACS system in the hospital to help the medical staff in diagnosis.
Examples of a clinical application may include: Neurosurgery, Orthopedics, Anesthesiology, Urology, Gynecology, Internal Medicine
(※ This device is not intended for mammography applications.)
510(k) Summary
{6}
DRTECH
# 7. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device [21 CFR 807.92(a)(6), 21 CFR 807.92(b)]
The EXTRON Series are designed as a combination of components, including a C-arm, X-ray tube and housing, flat panel detector, digital imaging system, collimator, generator, and monitor cart. Its purpose is to provide X-ray images of a patient's anatomy during surgery or treatment.
The subject device has the same fundamental scientific technologies as the predicate devices. The technological comparison table below demonstrates the comparability of the technological characteristics of the new device and the currently cleared predicate devices. The Technological differences do not affect the intended use of the device.
The table 1 below compares the main performance data of the subject device with the predicate devices to substantiate equivalence of the subject device and predicates.
Table 1. Comparison of the Subject Device (EXTRON 3/5/7) with Predicate and Reference Devices
| Parameter | Subject Device | Predicate Device | Reference 1 | Reference 2 | Discussion |
| --- | --- | --- | --- | --- | --- |
| 510(K) Number | K250010 | K142708 | K122234 | K230871 | - |
| Manufacturer | DRTECH Corporation | Philips Medical Systems Nederland B.V. | GE HEALTHCARE SURGERY GE OEC MEDICAL SYSTEMS, INC | DRTECH Corporation | - |
| Model Name | EXTRON 3, EXTRON 5, EXTRON 7 | Veradius Unity | OEC 9900 ELITE | EXTRON 5, EXTRON 7 | - |
| Classification Name | Image-intensified fluoroscopic x-ray system | Image-intensified fluoroscopic x-ray system | Image-intensified fluoroscopic x-ray system | Image-intensified fluoroscopic x-ray system | Equivalent |
| Classification Panel | Radiology | | | | Equivalent |
| Classification Regulation | 21 CFR 892.1650 | 21 CFR 892.1650 | 21 CFR 892.1650 | 21 CFR 892.1650 | Equivalent |
| Product Code | OWB, OXO, JAA | OWB, OXO, JAA | OWB, JAA, OXO | OWB, OXO, JAA | Equivalent |
| Device Class | Class II | Class II | Class II | Class II | Equivalent |
| Indications for Use | EXTRON Series are a mobile fluoroscopic X-ray system with high output capacity, high thermal capacity and high resolution image processing system, which provides X-ray images of the patient's | The proposed Veradius Unity device is intended to be used and operated by: adequately trained, qualified, and authorized health care professionals such as physicians, surgeons, cardiologists, | The OEC® 9900 Elite Mobile Fluoroscopy System is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. | EXTRON5 and EXTRON 7 are a mobile fluoroscopic X-ray system with high output capacity, high thermal capacity and high resolution image processing system, which provides X-ray images of the patient's | Equivalent |
510(k) Summary
{7}
DRTECH
| Parameter | Subject Device | Predicate Device | Reference 1 | Reference 2 | Discussion |
| --- | --- | --- | --- | --- | --- |
| | anatomy during surgery or treatment. This device plays an important role in emergency injury treatment, orthopedic surgery, neurosurgery surgery, bone surgery, etc. This device has a function to save important a specific images as records, so you can easily search for the images and transmit it to the PACS system in the hospital to help the medical staff in diagnosis.
Examples of a clinical application may include: Neurosurgery, Orthopedics, Anesthesiology, Urology, Gynecology, Internal Medicine
(※ This device is not intended for mammography applications.) | radiologists, and radiographers, who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional, and surgical procedures on all patients except neonates (birth to one month), within the limits of the device. The device is to be used in healthcare facilities both inside and outside the operating room, in sterile as well as non-sterile environments, in a variety of procedures.
Applications
• Orthopedic
• Neuro
• Abdominal
• Vascular
• Thoracic
• Cardiac | Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. | anatomy during surgery or treatment. This device plays an important role in emergency injury treatment, orthopedic surgery, neurosurgery surgery, bone surgery, etc. This device has a function to save important a specific images as records, so you can easily search for the images and transmit it to the PACS system in the hospital to help the medical staff in diagnosis.
Examples of a clinical application may include: Neurosurgery, Orthopedics, Anesthesiology, Urology, Gynecology, Internal Medicine
(※ This device is not intended for mammography applications.) | |
| Target population | Patients who need X-ray radiography or fluoroscopy images
Adults and Pediatrics | Patients who need diagnostic, interventional and surgical procedures, except neonates (birth to one month) | Patients who need fluoroscopic and spot-film images | Patients who need X-ray radiography or fluoroscopy images
Adults and Pediatrics | Equivalent |
| Mobile Platform | Yes | Yes | Yes | Yes | Equivalent |
| X-ray tube | EXTRON 3 : Stationary Anode | Rotating Anode | Rotating Anode | Rotating Anode | Equivalent : X-ray tubes and |
510(k) Summary
{8}
DRTECH
| Parameter | Subject Device | Predicate Device | Reference 1 | Reference 2 | Discussion |
| --- | --- | --- | --- | --- | --- |
| | EXTRON 5 : Rotating Anode
EXTRON 7: Rotating Anode | | | | systems verified according to the IEC 60601-2-28 and IEC 60601-1 series meet strict international safety and performance standards. Therefore, differences in X-ray tubes do not raise new concerns regarding safety and effectiveness. |
| Radiographic Mode | kV Range : 40 to 120kV mA Range
EXTRON 7 - Up to 150mA
EXTRON 5 - N/A
EXTRON 3 - Up to 100mA | kV Range : 40 to 120kV mA Range : Up to 125mA | kV Range : 40 to 120kV mA Range : Up to 150mA | kV Range : 40 to 120kV mA Range : Up to 150mA
※ EXTRON 7 only | Equivalent : Alteration in the mA does not give rise to any novel concerns regarding safety and effectiveness. |
| Fluoroscopic Mode | kV Range : 40 to 120kV mA Range :
EXTRON 3 - Up to 30mA
EXTRON 5 - Up to 40mA
EXTRON 7 - Up to 60mA | kV Range : 40 to 120kV mA Range : 1 to 60mA | kV Range : 40 to 120kV mA Range :
- Normal : 0.2- 10 mA
- High Level Fluoro : 1.0 - 40 mA | kV Range : 40 to 120kV mA Range :
EXTRON 5 - Up to 40mA
EXTRON 7 - Up to 60mA | Equivalent : Alteration in the mA does not give rise to any novel concerns regarding safety and effectiveness. |
| Dimension | Immersion Depth
EXTRON 3 - 73cm
EXTRON 5 - 74cm
EXTRON 7 - 74cm
Free Space : 80cm
Orbital movement : 165° | Immersion Depth : 73cm
Free Space : 77cm
Orbital movement : 145° | Immersion Depth : 84cm
Free Space : 79cm
Orbital movement : 145° | Immersion Depth : 74cm
Free Space : 80cm
Orbital movement : 165° | Equivalent : Alteration in the dimension does not give rise to any novel concerns regarding safety and effectiveness.
Additionally, due to the greater scope of movement, the Subject device offers |
510(k) Summary
{9}
DRTECH
| Parameter | | Subject Device | Predicate Device | Reference 1 | Reference 2 | Discussion |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | | | a higher degree of convenience compared to the Predicate device. |
| Laser Guide | | Yes | Yes | Yes | Yes | Equivalent |
| Foot Switch | | Wired Foot Switch Wireless Foot Switch | Wired Foot Switch Wireless Foot Switch | Wired Foot Switch Wireless Foot Switch | Wired Foot Switch Wireless Foot Switch | Equivalent |
| Detector | Pixels | EXPD 2121P : 1500 x 1500 pixels
EXPD 2323P : 1536 x 1536 pixels
EXPD 2323PB:1536 x 1536 pixels
EXPD 3030P : 2048 x 2048 pixels | 1560 x 1420 pixels | 21 cm FPD : 1536 x 1496 pixels
31 cm FPD : 1548 x 1524 pixels | EXPD 2121P : 1500 x 1500 pixels
EXPD 3030P : 2048 x 2048 pixels | Equivalent : Alteration in the detector pixels do not give rise to any novel concerns regarding safety and effectiveness. |
| | DQE | 55% @1lp/mm | 70% @ 0p/mm | 21 cm FPD : Typical 63% @0.5lp/mm
31 cm FPD : Typical 62% @0.5lp/mm | 55% @1lp/mm | Equivalent : Similarly, while there is a difference in DQE values, the Subject Device demonstrated clinically comparable image quality to the Predicate and Reference Devices during clinical image comparison evaluations. Thus, no novel concerns regarding safety and effectiveness are introduced. |
| | MTF | 55% @1lp/mm | 59% @ 1lp/mm | - | 55% @1lp/mm | Equivalent : Although there is a difference in MTF values, actual clinical image comparison |
510(k) Summary
{10}
510(k) Summary
7 / 9
# DRTECH
| Parameter | Subject Device | Predicate Device | Reference 1 | Reference 2 | Discussion |
| --- | --- | --- | --- | --- | --- |
| | | | | | evaluations
confirmed that the
Subject Device
provides image
resolution that is
equivalent to or better
than that of the
Predicate and
Reference Devices.
Therefore, this
difference does not
give rise to any novel
concerns regarding
safety and
effectiveness. |
{11}
DRTECH
# 8. Summary of Non-Clinical Data [21 CFR 807.92(b)(1)]
The EXTRON Series comply with the following international and FDA-recognized consensus standards list in Table 2.
Table 2. International and FDA-recognized consensus standards
| Standards development organization, reference number, and date | Standard name |
| --- | --- |
| ISO 14971 Third Edition 2019-12 | Medical devices - Application of risk management to medical devices |
| IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - Note: This standard is recognized with relevant US national differences applied, see references #1 and #2 in the Relevant FDA Guidance and/or Supportive Publication section below. |
| ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)] |
| IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
| IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION | Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability |
| IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION | Medical devices - Part 1: Application of usability engineering to medical devices |
| IEC 60601-1-3 Edition 2.2 2021-01 CONSOLIDATED VERSION | Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment |
| IEC 60601-2-28 Edition 3.0 2017-06 | Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis |
| IEC 60601-2-54 Edition 2.0 2022-09 | Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy |
| IEC 60601-2-43 Edition 3.0 2022-12 | Medical electrical equipment - Part 2-43: Particular requirements for the safety and essential performance of X-ray equipment for interventional procedures |
| IEC 60601-2-43 Edition 2.2 2019-10 CONSOLIDATED VERSION | Medical electrical equipment - Part 2-43: Particular requirements for the safety and essential performance of X-ray equipment for interventional procedures |
| IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION | Medical device software - Software life cycle processes |
| ANSI UL 2900-1 First Edition 2017 | Standard for Safety, Standard for Software Cybersecurity Network-Connectable Products, Part 1: General Requirements |
510(k) Summary
{12}
D R T E C H
IEC 81001-5-1 Edition 1.0 Health software and health IT systems safety, effectiveness and security 2021-12 - Part 5-1: Security - Activities in the product life cycle
And EXTRON Series comply with the FDA guidance documents listed in Table 3.
Table 3. FDA Guidance Documents
| Title of Guidance Document | Issue Date |
| --- | --- |
| Guidance for Industry and Food and Drug Administration Staff: The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] | July 28, 2014 |
| Guidance for Industry and FDA Staff: Content of Premarket Submissions for Device Software Functions | June 14, 2023 |
| Guidance for Industry and Food and Drug Administration Staff: Medical X-Ray Imaging Devices Conformance with IEC Standards | May 8, 2019 |
| Guidance for Industry and FDA Staff: Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices | July 11, 2016 |
| Guidance for Industry and FDA Staff: Applying Human Factors and Usability Engineering to Medical Devices | February 3, 2016 |
| Guidance for Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions | September 27, 2022 |
Sample clinical images using anthropomorphic phantoms representative of the indicated anatomies and populations have been taken for both the proposed devices (EXTRON 3/5/7) and the predicate devices (Veradius Unity and OEC 9900 ELITE). These images have been reviewed and compared by qualified clinical experts. As a result of this evaluation, it has been confirmed that the EXTRON Series devices provide equivalent image quality to the predicate devices.
## 9. Summary of Clinical Data [21 CFR 807.92(b)(2)]
Not Applicable
Clinical studies were not performed, but phantom images were taken and reviewed by qualified clinicians to support SE (see Section 8 for more details).
## 10. Conclusion [21 CFR 807.92(b)(3)]
The EXTRON Series are substantially equivalent to the currently marketed predicate devices (Veradius Unity (K142708), EXTRON 5/7(K230871), OEC 9900 ELITE(K122234)) in terms of design, fundamental scientific technology, and indications for use, safety, and effectiveness.
Substantial equivalence for Mobile fluoroscopic X-ray System was demonstrated through the non-clinical performance in compliance with the requirements specified in the international and FDA recognized consensus standards, ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, IEC 60601-2-43, IEC 62304, ANSI UL 2900-1, and IEC 81001-5-1.
The comparison of technological characteristics, non-clinical performance data and safety testing demonstrate that the EXTRON Series are substantially equivalent to the predicate devices.
510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.