K244041 · Xemis Medical Technology (Shenzhen) Co., Ltd. · IMG · Oct 7, 2025 · Physical Medicine
Device Facts
Record ID
K244041
Device Name
Ultrasound Therapy Workstation (XMS-UET2)
Applicant
Xemis Medical Technology (Shenzhen) Co., Ltd.
Product Code
IMG · Physical Medicine
Decision Date
Oct 7, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended for use only in adult patients. Therapeutic Ultrasound Application of therapeutic deep heat for the treatment of selected sub-chronic and chronic medical conditions such as: 1) Pain relief, muscle spasms and joint contractures. 2) Relief of pain, muscle spasms and joint contractures that may be associated with: - Adhesive capsulitis - Bursitis with slight calcification - Myositis - Soft tissue injuries - Shortened tendons due to past injuries and scar tissues 3) Relief of sub-chronic, chronic pain and joint contractures resulting from: - Capsular tightness - Capsular scarring For TENS, Interferential, premodulated (IFC), NMS and Microcurrent: 1) Symptomatic relief of chronic intractable pain 2) Post-traumatic acute pain 3) Post-surgical acute pain Additionally for NMS, NMS Burst, Hi-Volt and Russian: 1) Relaxation of Muscle spasms 2) Prevention or retardation of disuse atrophy 3) Increasing local blood circulation 4) Muscle re-education 5) Maintaining or increasing range of motion 6) Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis For DC Continuous Mode 1) Relaxation of muscle spasm
Device Story
Ultrasound Therapy Workstation (XMS-UET2) is an AC-powered device for clinical use (hospitals, clinics, rehab centers) by healthcare professionals. It provides electrotherapy (TENS, IFC, NMS, Microcurrent, Hi-Volt, Russian, DC) and therapeutic ultrasound. Device features color touchscreen and control knob for parameter selection. Software controls all indicators and outputs. Ultrasound module uses piezoelectric crystals to deliver deep heat; electrotherapy module delivers electrical stimulation via electrodes. Device includes pre-set protocols and manual adjustment modes. Output parameters (voltage, current, frequency, pulse width) are software-controlled. Safety features include automatic overload/no-load trips and shut-off. Output affects muscle tissue and nerves to provide pain relief, muscle relaxation, and rehabilitation benefits. Patient safety is managed via operator monitoring and labeling warnings regarding skin irritation/burns.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), electrical safety (ANSI AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance testing per IEC 60601-2-5 and IEC 60601-2-10.
Indicated for adult patients requiring therapeutic deep heat for sub-chronic/chronic pain, muscle spasms, joint contractures, adhesive capsulitis, bursitis, myositis, soft tissue injuries, or shortened tendons. Also indicated for symptomatic relief of chronic intractable, post-traumatic, or post-surgical pain via TENS, IFC, NMS, or Microcurrent. NMS, Hi-Volt, and Russian modes indicated for muscle spasm relaxation, atrophy prevention, circulation improvement, muscle re-education, range of motion maintenance, and post-surgical venous thrombosis prevention.
Regulatory Classification
Identification
An ultrasound and muscle stimulator for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies. The device also passes electrical currents through the body area to stimulate or relax muscles.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 7, 2025
Xemis Medical Technology (Shenzhen) Co., Ltd.
% Jie Yang
Consultant
Chonconn Consulting Co., Ltd.
Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District
Shenzhen, Guangdong 518067
China
Re: K244041
Trade/Device Name: Ultrasound Therapy Workstation (XMS-UET2)
Regulation Number: 21 CFR 890.5860
Regulation Name: Ultrasound And Muscle Stimulator
Regulatory Class: Class II
Product Code: IMG, GZI, GZJ, IPF, LIH
Dated: December 30, 2024
Received: August 25, 2025
Dear Jie Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K244041 - Jie Yang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K244041 - Jie Yang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Amber T. Ballard -S
Amber Ballard, PhD
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K244041
Device Name
Ultrasound Therapy Workstation (XMS-UET2)
Indications for Use (Describe)
This device is intended for use only in adult patients.
Therapeutic Ultrasound
Application of therapeutic deep heat for the treatment of selected sub-chronic and chronic medical conditions such as:
1) Pain relief, muscle spasms and joint contractures.
2) Relief of pain, muscle spasms and joint contractures that may be associated with:
- Adhesive capsulitis
- Bursitis with slight calcification
- Myositis
- Soft tissue injuries
- Shortened tendons due to past injuries and scar tissues
3) Relief of sub-chronic, chronic pain and joint contractures resulting from:
- Capsular tightness
- Capsular scarring
For TENS, Interferential, premodulated (IFC), NMS and Microcurrent:
1) Symptomatic relief of chronic intractable pain
2) Post-traumatic acute pain
3) Post-surgical acute pain
Additionally for NMS, NMS Burst, Hi-Volt and Russian:
1) Relaxation of Muscle spasms
2) Prevention or retardation of disuse atrophy
3) Increasing local blood circulation
4) Muscle re-education
5) Maintaining or increasing range of motion
6) Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
For DC Continuous Mode
1) Relaxation of muscle spasm
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}
510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
Prepared Date: 2025-10-07
1. Submission sponsor
Name: Xemis Medical Technology (Shenzhen) Co., Ltd.
Address: 1101/1102/1201/1202, Unit 1, Building 2, Hongpeng Building, Tiegang Community, Xixiang Street, Baoan, Shenzhen, China
Contact person: Yang Yongfen
Title: RA Engineer
E-mail: yangyongfen@ximisihn.com
Tel: 86-755-23316735
2. Submission correspondent
Name: Chonconn Consulting Co., Ltd.
Address: Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P.R. China
Contact person: Yang Jie
E-mail: yangjie@chonconn.com
Tel: +86-755 33941160
3. Subject Device Information
| Trade/Device Name | Ultrasound Therapy Workstation |
| --- | --- |
| Model | XMS-UET2 |
| Common Name | Ultrasound Therapy Workstation |
| Regulatory Class | Class II |
| Classification Regulation | 21 CFR 890.5860 (Ultrasound and muscle stimulator) |
| Product codes | IMG, IPF, GZI, GZJ, LIH |
| Submission type | Traditional 510(K) |
4. Predicate Device
| | Device name and model | K number | Manufacturer |
| --- | --- | --- | --- |
| Predicate device | ComboRehab | K150436 | Shenzhen Dongdixin Technology Co. Ltd. |
5. Device Description
The Ultrasound Therapy Workstation (model: XMS-UET2) is an AC-powered device designed to use electrical stimulation and therapeutic ultrasound. Key features include ultrasound output and electrical stimulation output. This device is intended for use in healthcare environments such as hospitals, rehabilitation centers, community health service centers, and clinics.
1 / 16
{6}
The device has a color touchscreen display and a control knob with light ring. The screen display shows system information to the user. Both display screen and control knob can assist the user in selecting or adjusting parameters. All controls and indicators are controlled by software.
Three non-invasive therapeutic methods are available: electrotherapy, ultrasound therapy and combination therapy (ultrasound plus electrical stimulation). For each method, the device is configured with protocols and manual operation, which can facilitate users to choose pre-set protocols or adjust parameters to suit individual needs.
## 6. Indication for use
This device is intended for use only in adult patients.
## Therapeutic Ultrasound
Application of therapeutic deep heat for the treatment of selected sub-chronic and chronic medical conditions such as:
1) Pain relief, muscle spasms and joint contractures.
2) Relief of pain, muscle spasms and joint contractures that may be associated with:
- Adhesive capsulitis
- Bursitis with slight calcification
- Myositis
- Soft tissue injuries
- Shortened tendons due to past injuries and scar tissues
3) Relief of sub-chronic, chronic pain and joint contractures resulting from:
- Capsular tightness
- Capsular scarring
## For TENS, Interferential, premodulated (IFC), NMS and Microcurrent:
1) Symptomatic relief of chronic intractable pain
2) Post-traumatic acute pain
3) Post-surgical acute pain
## Additionally for NMS, NMS Burst, Hi-Volt and Russian:
1) Relaxation of Muscle spasms
2) Prevention or retardation of disuse atrophy
3) Increasing local blood circulation
4) Muscle re-education
5) Maintaining or increasing range of motion
6) Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
## For DC Continuous Mode
1) Relaxation of muscle spasm
2 / 16
{7}
7. Comparison to the Predicate Device
| Features | Subject Device: | Predicate Device | Comparison |
| --- | --- | --- | --- |
| 1.510(k) number | K244041 | K150436 | / |
| 2.Device name, Model | Ultrasound Therapy Workstation XMS-UET2 | ComboRehab | / |
| 3.Manufacturer | Xemis Medical Technology (Shenzhen) Co., Ltd. | Shenzhen Dongdixin Technology Co. Ltd | / |
| Product Code | IMG,GZI,GZJ,IPF,LIH | IMG, IPF, GZJ, GZI, LIH, HCC | Same, See indications for use. |
| Regulation Number | 21 CFR 890.5860 | 21 CFR 890.5860 | Same |
| Classification | Class II | Class II | Same |
| Type of use | Prescription | Prescription | Same |
| Indications for Use | This device is intended for use only in adult patients Therapeutic Ultrasound Application of therapeutic deep heat for the treatment of selected sub-chronic and chronic medical conditions such as: 1) Pain relief, muscle spasms and joint contractures. 2) Relief of pain, muscle spasms and joint contractures that may be associated with: - Adhesive capsulitis - Bursitis with slight calcification - Myositis - Soft tissue injuries - Shortened tendons due to past injuries and scar tissues 3) Relief of sub-chronic, chronic pain and joint contractures resulting from: | Therapeutic Ultrasound Application of therapeutic deep heat for the treatment of selected sub-chronic and chronic medical conditions such as: 1. Pain relief, muscle spasms and joint contractures. 2. Relief of pain, muscle spasms and joint contractures that may be associated with: • Adhesive capsulitis • Bursitis with slight calcification • Myositis • Soft tissue injuries • Shortened tendons due to past injuries and scar tissues 3. Relief of sub-chronic, chronic pain and joint contractures resulting from: • Capsular tightness • Capsular scarring For TENS, Interferential, | Same, The indications of the predicate device cover the subject device. The subject device does not have EMG and EMG triggered Stim indications compared with predicate. |
3 / 16
{8}
4 / 16
| | - Capsular tightness
- Capsular scarring
For TENS, Interferential, premodulated (IFC), NMS and Microcurrent:
1) Symptomatic relief of chronic intractable pain
2) Post-traumatic acute pain
3) Post-surgical acute pain
Additionally for NMS, NMS Burst, Hi-Volt and Russian:
1) Relaxation of Muscle spasms
2) Prevention or retardation of disuse atrophy
3) Increasing local blood circulation
4) Muscle re-education
5) Maintaining or increasing range of motion
6) Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
For DC Continuous Mode
1) Relaxation of muscle spasm | premodulated(IFC), NMS and Microcurrent:
1. Symptomatic relief of chronic intractable pain
2. Post-traumatic acute pain
3. Post-surgical acute pain
Additionally for NMS, NMS Burst, Hi-Volt and Russian:
1. Relaxation of Muscle spasms
2. Prevention or retardation of disuse atrophy
3. Increasing local blood circulation
4. Muscle re-education
5. Maintaining or increasing range of motion
6. Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
For DC Continuous Mode
Relaxation of muscle spasm
For EMG
To determine the activation timing of muscles for:
1. Retaining of muscle activation
2. Coordination of muscle activation
An indication of the force produced by muscle for control and maintenance of muscle contractions
1. Relaxation muscle training
2. Muscle re-education
For EMG triggered Stim
1. Stroke rehab by muscle re-education
2. Relaxation of muscle spasms
3. Prevention or retardation of disuse atrophy
4. Increase local blood circulation
5. Muscle re-education |
| --- | --- | --- |
{9}
| | | 6. Maintaining or increasing range of motion | |
| --- | --- | --- | --- |
| Electrotherapy Module | | | |
| Electrotherapy mode | NMS (Pulsed mode, burst mode)
TENS(Symmetrical, Asymmetrical, Alternating, Symmetrical Burst, Asymmetrical Burst, Alternating Burst, Symmetrical RAAS, Asymmetrical RAAS)
high volt
Russian
interferential (4-pole), premodulated (2-pole interferential), Microcurrent
Continuous DC | NMS (Pulsed mode, burst mode)
TENS(Symmetrical, Asymmetrical, Alternating, Symmetrical Burst, Asymmetrical Burst, Alternating Burst)
high volt
interferential (4-pole), premodulated (2-pole interferential),
Russian, Microcurrent
Direct current | Similar |
| EMG (Biofeedback) | no | yes | Different,
See indications for use. |
| Basic unit Characteristics | | | |
| 4. Power Source(s) | AC line | AC line | Same |
| -Method of Line Current Isolation | Transformer | Transformer | Same |
| -Leakage current
– Normal condition
- Single fault condition | 75 μA
130 μA | d.c:0. a.c.:76.5
d.c:0. a.c.:132.1 | Same |
| 5. Number of output modes | 10 | 10 | Same |
| 6. Number of output channel | 6 | 4 | Similar |
| - Synchronous | 1&2 or 3&4 or 5&6 | 1 &2 or 3&4 | Similar |
<br>
5 / 16</br>
{10}
| or Alternation? | | | |
| --- | --- | --- | --- |
| -Method of Channel Isolation | Through transformers and isolators | Not publicly available | Similar |
| 7.Regulated Current or Regulated Voltage? | Optional
Optional | Optional
Optional | Same |
| 8. Software/Firmware/Microprocessors or control? | Yes | Yes | Same |
| 9. Automatic Overload trip | Yes | Yes | Same |
| 10. Automatic no-load trip | Yes | Yes | Same |
| 11. Automatic Shut Off | Yes | Yes | Same |
| 12. Patient override control method | Yes | Yes
Patient interrupt switch | Same |
| 13. Indicator display
-On/Off status
-Low battery
-Voltage/Current Level? | Yes
N/A
Yes | Yes
Yes
Yes | Same |
| 14. Timer range (minutes) | 0-60 minutes | 0-60 minutes | Same |
| 15.Compliance with voluntary standards | ISO 14971,
IEC 60601-1;
IEC 60601-1-2;
IEC 60601-2-5; | ISO14971,
IEC 60601-1,
60601-1-2,
IEC 60601-2-10, | Same |
6 / 16
{11}
| | IEC 60601-2-10 | IEC60601-2-5, MDD 93/42/EEC,Annex II | |
| --- | --- | --- | --- |
| 16. Compliance with 21CFR 898 | Yes | Yes | Same |
| 17. Weight(lbs.) | 9.48 | 10.36 | Similar |
| 18. Dimensions (in.) H*W * L | 9.5in*7.5in*16.2 in | 10 x 7x15.5 | Similar |
| 19. Housing material and construction | Plastic | Plastic | Same |
| Output specifications | | | |
| Waveform | NMS (Pulsed mode, burst mode): Biphasic TENS: Biphasic high volt: Monophasic Russian: Biphasic interferential (4-pole): Biphasic premodulated (2-pole interferential): Biphasic Microcurrent: Monophasic Continuous DC: DC | NMS (Pulsed mode, burst mode): Biphasic TENS: Biphasic high volt: Monophasic Russian: Biphasic interferential (4-pole): Biphasic premodulated (2-pole interferential): Biphasic Microcurrent: Monophasic Continuous DC: DC | Same |
| Shape | NMS (Pulsed mode, burst mode): Square TENS: Square high volt: Twin spike Russian: Sinusidal interferential (4-pole): Sinusidal premodulated (2-pole interferential): Sinusidal Microcurrent: Square Continuous DC:DC | NMS (Pulsed mode, burst mode): Square TENS: Square high volt: Twin spike Russian: Sinusidal interferential (4-pole): Sinusidal premodulated (2-pole interferential): Sinusidal Microcurrent: Square Continuous DC:DC | Same |
| Maximum output voltage(V)± 20% | NMS (Pulsed mode, burst mode): 100@500Ω TENS: 100@500Ω | Not publicly available | different for i.NMS mode at 2k Ω |
{12}
8 / 16
| | high volt: 462@500 Ω
Russian: 50.2@500Ω
interferential (4-pole): 50.2@500Ω
premodulated (2-pole interferential): 50.3@500Ω
Microcurrent: 0.5@500Ω
Continuous DC: 38@500Ω
NMS (Pulsed mode, burst mode): 220@2KΩ
TENS: 220@2KΩ
high volt: 487@2KΩ
Russian: 202@2KΩ
interferential (4-pole): 202@2KΩ
premodulated (2-pole interferential): 202@2KΩ
Microcurrent: 2.2@2KΩ
Continuous DC: 150@2KΩ
Microcurrent: 11@10k Ω
high volt: 487@10KΩ
NMS, TENS, high volt, Russian, interferential (4-pole), premodulated (2-pole interferential) and Continuous DC are unable to output@10kΩ, indicating no load | | ii.Russian mode at 2k Ω
iii.Inferential mode at 2k Ω
iv.Premodulated mode at 2k Ω
v.Continuous DC mode at 2k Ω |
| --- | --- | --- | --- |
| Maximum output current(mA) ±20% | NMS (Pulsed mode, burst mode): 200@500Ω
TENS: 200@500Ω
Russian: 100.4@500Ω
interferential (4-pole): 100.4@500Ω
premodulated (2-pole interferential): 100.6@500Ω
Microcurrent: 1@500Ω
Continuous DC: 76@500Ω
high volt: 927@500 Ω | Not publicly available | Different for
i.Continuous DC mode at 500 Ω
ii.Russian mode at 2k Ω
iii.Inferential mode at 2k Ω |
{13}
| | NMS (Pulsed mode, burst mode): 110@2KΩ
TENS: 110@2KΩ
Russian: 101@2KΩ
interferential (4-pole): 101@2KΩ
premodulated (2-pole
interferential): 101@2KΩ
Microcurrent: 1.1@2KΩ
Continuous DC: 75@2KΩ
high volt: 243.5@2k Ω
high volt: 48.7@10k Ω
Microcurrent: 1.1@10k Ω
NMS, TENS, Russian, interferential (4-pole), premodulated (2-pole
interferential) and Continuous DC are unable to output@10kΩ, indicating no load | | iv.Premodulated mode at 2k Ω
v.Continuous DC mode at 2k Ω |
| --- | --- | --- | --- |
| Pulse Width | NMS (Pulsed mode, burst mode): 20~400μs
TENS: 20~1000μs
high volt: 10μs
Russian: 50~250μs
interferential (4-pole): 50~250μs
premodulated (2-pole
interferential): 50~250μs
Microcurrent: 0.5ms~5s | Not publicly available | Different |
| Frequency(H z) | NMS (Pulsed mode, burst mode): 1~250Hz
TENS: 1~250Hz
high volt: 1~120Hz
Russian: 2k~10kHz
interferential (4-pole): 2k~10kHz
premodulated (2-pole
interferential): 2k~10kHz
Microcurrent: 0.1~1000Hz
Continuous DC: N/A | NMS (Pulsed mode, burst mode): 1~250Hz
TENS: 1~250Hz
high volt:1~120Hz
Russian: Not publicly available
interferential (4-pole):2k~10kHz
premodulated (2-pole
interferential): 2k~10kHz
Microcurrent: 0.1~1000Hz
Continuous DC: N/A | Different for Russian mode |
9 / 16
{14}
| For interferential modes only: - Beat Frequency(H z) | Interferential, 4-pole: 1-200Hz Interferential, 2-pole, Premodulated: 1-150Hz | Interferential, 4-pole:1-200Hz Interferential, 2-pole, Premodulated: 1-200Hz | Similar |
| --- | --- | --- | --- |
| For multiphasic waveforms only: - Symmetrical phases? | NMS (Pulsed mode, burst mode): Yes TENS: Yes high volt: No Russian: Yes interferential (4-pole): Yes premodulated (2-pole interferential): Yes Microcurrent: No | NMS (Pulsed mode, burst mode): Yes TENS: Yes high volt: No Russian: Yes interferential (4-pole): Yes premodulated (2-pole interferential): Yes Microcurrent: No | Same |
| - Phase Duration (include units) (state range, if applicable) (both phases, if asymmetrical) | NMS (Pulsed mode, burst mode): 20~400μs TENS: 20~1000μs high volt: 100μs Russian: 50~250μs interferential (4-pole): 50~250μs premodulated (2-pole interferential): 50~250μs Microcurrent: 0.5ms~5s Continuous DC:N/A | NMS (Pulsed mode, burst mode): 20us-1000us TENS: 20~1000μs high volt:100μs Russian: Not publicly available interferential (4-pole):Not publicly available premodulated (2-pole interferential): Not publicly available Microcurrent: Not publicly available Continuous DC: N/A | Similar |
| Net Charge (per pulse) | CC mode: Microcurrent: 5000μC @500Ω CV mode: high volt: 5μC @500Ω The rest: Zero | Not publicly available | Different for Microcurrent mode |
| Symmetry | Symmetric and asymmetric | Symmetric and asymmetric | Same |
| Method | Balanced | Not publicly available | Same |
| Maximum Phase Charge(μC)(500Ω) | NMS (Pulsed mode, burst mode): 60 TENS: 100 high volt:5 Russian:15.92 interferential (4-pole): 15.92 premodulated (2-pole | Not publicly available | Different for Microcurrent mode |
10 / 16
{15}
| | interferential): 15.92
Microcurrent: 5000 | | |
| --- | --- | --- | --- |
| Maximum current density(mA/cm²,500Ω) | NMS (Pulsed mode, burst mode): 1.8,0.9
TENS: 2.0,1.4,1.5,1.1
high volt: 0.92
Russian:2.0
interferential (4-pole): 2.8
premodulated (2-pole interferential): 2.0
Microcurrent: 0.028
Continuous DC: 3.2 | Not publicly available | Different for i.Russian mode
ii.Microcurrent mode |
| Maximum power density (W/cm²,500 Ω) | NMS (Pulsed mode, burst mode): 0.04,0.01
TENS:0.049,0.025,0.030,0.015
high volt: 0.01
Russian:0.050
interferential (4-pole): 0.101
premodulated (2-pole interferential): 0.050
Microcurrent: 0.00001
Continuous DC:0.128 | Not publicly available | Different for i.NMS mode
ii.TENS mode
iii.High volt mode
iv.Russian mode |
| Burst Mode
a. Pulses per burst | NMS (Pulsed mode): N/A
NMS (burst mode): 7
TENS: 7
high volt: N/A
Russian: N/A
interferential (4-pole): N/A
premodulated (2-pole interferential): N/A
Microcurrent: N/A
Continuous DC:N/A | Not publicly available | Same |
| b. Bursts per second | NMS (Pulsed mode): N/A
NMS (burst mode): 1~9
TENS: 1~9
high volt: N/A
Russian: 20~100
interferential (4-pole): N/A
premodulated (2-pole interferential): N/A
Microcurrent: N/A | Not publicly available | Same |
11 / 16
{16}
| | Continuous DC:N/A | | |
| --- | --- | --- | --- |
| c. Burst duration (ms) | NMS (Pulsed mode): N/A NMS (burst mode): 28~984 ms TENS: 28~984 ms high volt: N/A Russian: 2~25 ms interferential (4-pole): N/A premodulated (2-pole interferential): N/A Microcurrent: N/A Continuous DC:N/A | Not publicly available | Same |
| d. Duty Cycle [Line (b) x Line (c)] | NMS (Pulsed mode): N/A NMS (burst mode): 2.8%~98.4% TENS: 2.8%~98.4% high volt: N/A Russian: 10%~50% interferential (4-pole): N/A premodulated (2-pole interferential): N/A Microcurrent: N/A Continuous DC:N/A | Not publicly available | Same |
| ON Time (seconds) | NMS (Pulsed mode): 0~60s TENS(not include burst mode): 0~60s high volt: 0~60s Russian:0~60s interferential (4-pole): N/A premodulated (2-pole interferential): 0~60s Microcurrent: 0~60s Continuous DC:N/A | NMS (Pulsed mode): 0~60s TENS(not include burst mode): 0~60s high volt: 0~60s Russian: 0~60s interferential (4-pole): N/A premodulated (2-pole interferential): 0~60s Microcurrent: 0~60s Continuous DC:N/A | Same |
| OFF Time (seconds) | NMS (Pulsed mode): 0~60s TENS(not include burst mode): 0~60s high volt: 0~60s Russian:0~60s interferential (4-pole): N/A premodulated (2-pole interferential): 0~60s Microcurrent: 0~60s Continuous DC:N/A | NMS (Pulsed mode): 0~60s TENS(not include burst mode): 0~60s high volt: 0~60s Russian: 0~60s interferential (4-pole): N/A premodulated (2-pole interferential): 0~60s Microcurrent: 0~60s Continuous DC:N/A | Same |
{17}
| Ultrasound Module | | | |
| --- | --- | --- | --- |
| Power Source | AC line | AC line | Same |
| Maximum Treatment Time: | 30 minutes | 30 minutes | Same |
| Frequency | 1 MHz or 3 MHz | 1 MHz or 3 MHz | Same |
| Modes | Continuous and Pulsed | Continuous and Pulsed | Same |
| Pulse Repetition Rate | 100Hz, 48Hz, 16Hz | 16Hz, 48Hz, 100Hz | Same |
| Pulse Duration | 1.0ms~56.3ms (±5%) | Not publicly available | Different |
| Temporal Peak/average intensity ration | 1~10 (±5%) | 1~10 | Same |
| Maximum intensity | 2 W/cm2for continuous mode3 W/cm2for pulsed mode | 3 W/cm2 | Same |
| Indication accuracy | ±20 % | ±20 % | Same |
| Piezoelectric discs | Ultrasound transducer attached to a metal surface and patient contact through the metal | Ultrasound transducer attached to a metal surface and patient contact through the metal | Same |
| Frequency | 1 MHz or 3 MHz | 1 MHz or 3 MHz | Same |
| Effective radiating area | A9050: 4cm2A8040: 4cm2A3010: 1cm2 | 5 cm21 cm2 | Similar |
| Maximum beam non-uniformity ratio | 5:1 maximum | 5:1 maximum | Same |
| Beam type | collimated | collimated | Same |
| Applicator material | AL6061 | Not publicly available | Similar |
| Crystal Material | piezocrystal | piezocrystal | Same |
| Temporal Max Power (W) for | A3010 (1cm2): 3W±20%A8040 (3cm2): 9W±20%A9050 (4cm2): 12W±20% | 1cm2: 2W±20%5cm2: 15W±20% | Similar |
| Maximum Time | 100s | 100s | Same |
{18}
| pulsed mode | | | |
| --- | --- | --- | --- |
| Temporal Average Power (W) for continuous mode | A3010 (1cm2): 2W±20%A8040 (3cm2): 6W±20%A9050 (4cm2): 8W±20% | Not publicly available | Similar |
| Instantaneous Peak Power (W) | A3010 (1cm2): 3W±20%A8040 (3cm2): 9W±20%A9050 (4cm2): 12W±20% | 1cm2: 2W±20%5cm2: 15W±20% | Similar |
| Temporal Max Effective Intensity (W/cm2), peak P/ERA | 3W/cm2 | 3 W/cm2±20 % | Same |
| Temporal Average Intensity(W/cm2) | 0~2.1W/cm2, ±20% | Not publicly available | Similar |
| Temporal Peak to Average Ratio (Rtpa) for pulsed mode | 10.00:1 ±5 % at 10 % Duty Cycle5.00:1 ±5 % at 20 % Duty Cycle3.33:1 ±5 % at 30 % Duty Cycle2.50:1 ±5 % at 40 % Duty Cycle2.00:1 ±5 % at 50 % Duty Cycle1.67:1 ±5 % at 60 % Duty Cycle1.43:1 ±5 % at 70 % Duty Cycle1.25:1 ±5 % at 80 % Duty Cycle1.11:1 ±5 % at 90 % Duty Cycle | 10.00:1 ±5 % at 10 % Duty Cycle5.00:1 ±5 % at 20 % Duty Cycle3.33:1 ±5 % at 30 % Duty Cycle2.50:1 ±5 % at 40 % Duty Cycle2.00:1 ±5 % at 50 % Duty Cycle1.67:1 ±5 % at 60 % Duty Cycle1.43:1 ±5 % at 70 % Duty Cycle1.25:1 ±5 % at 80 % Duty Cycle1.11:1 ±5 % at 90 % Duty Cycle | Similar |
| Error uncertainties for the ultrasonic frequency | ±10% | ±10% | Same |
{19}
There are differences in output parameters between the subject and predicate devices. First, the subject device has additional TENS output modes (i.e., Symmetrical RAAS, Asymmetrical RAAS). These output modes do not raise new questions of safety and effectiveness nor a new intended use.
There are minor differences in maximum output voltage, maximum output current, frequency, and phase duration which do not raise new questions of safety and effectiveness.
The net charge and phase charge for the Microcurrent mode are very large which could increase the risk of skin irritation and burns. Additionally, the device is capable of delivery current densities greater than $2\mathrm{mA/cm^2}$ (i.e., the limit set by IEC 60601-2-10). However, this risk associated with these parameters has been mitigated with labeling revisions which 1) warn users of the risk of skin irritation and burns, 2) instruct device operators to monitor patients for adverse events throughout the duration of treatment, and 3) instruct that the treatment site should be examined after each treatment session to monitor for signs of irritation and/or burns.
Minor differences in the power density do not raise questions of safety and effectiveness since they remain under the $0.25\mathrm{W/cm^2}$ limit set within the FDA guidance document "Guidance Document for Powered Muscle Stimulator 510(k)s."
Lastly, there are minor differences in the ultrasound output parameters which do not raise new questions of safety and effectiveness.
## 8. Non-clinical tests
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
Biocompatibility of the subject device was evaluated in accordance with the FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.
## Electrical safety, electromagnetic compatibility (EMC) and performance
Electrical safety, EMC and performance testing were performed to, and passed, the following standards:
- ANSI AAMI ES60601-1 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: electromagnetic compatibility Requirements and tests
- IEC 60601-2-5 Particular requirements for the basic safety and essential
15 / 16
{20}
performance of ultrasonic physiotherapy equipment
- IEC 60601-2-10 Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
9. Clinical study
Not applicable.
10. Conclusion
Substantial equivalence comparisons, performance testing and compliance with voluntary standards demonstrate that the proposed subject device is substantially equivalent to the predicate device.
16 / 16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.