Arthrex Humeral Plating System and Cerclage Button

K243995 · Arthrex, Inc. · HRS · Mar 26, 2025 · Orthopedic

Device Facts

Record IDK243995
Device NameArthrex Humeral Plating System and Cerclage Button
ApplicantArthrex, Inc.
Product CodeHRS · Orthopedic
Decision DateMar 26, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Humeral Plating System is indicated for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone. The Cerclage Button is intended for use with the humeral plating system and FiberTape® cerclage suture, to augment fracture stabilization with humeral plates in long bone fixation. The cerclage button is designed for use with the humeral plating system and may not be used alone.

Device Story

System consists of straight humeral plates and 3.5 threaded cerclage button; used for proximal humerus fracture repair. Plates feature suture holes for soft tissue refixation; range 90.5–309 mm. Cerclage button augments fracture stabilization when used with humeral plates and FiberTape® cerclage suture; not for standalone use. Devices provided sterile (gamma) or non-sterile for single use. Operated by surgeons in clinical/OR settings. Output is physical fixation of bone fragments; aids in fracture healing and stabilization. Benefits include improved structural support in osteopenic bone.

Clinical Evidence

Bench testing only. 4-point bend testing (ASTM F382-17) demonstrated statistical equivalence to predicate. MRI safety testing (force, torque, heating, artifacts) performed per FDA guidance and ASTM standards (F2052, F2119, F2182, F2213). Bacterial Endotoxins Test (BET) performed per ANSI/AAMI ST72, USP <161>, USP <85>, and EP 2.6.14 confirmed compliance with pyrogen limits.

Technological Characteristics

Materials: Titanium alloy (ASTM F136). Design: Straight humeral plates (90.5–309 mm) and 3.5 threaded cerclage button. Principle: Mechanical bone fixation. Sterilization: Gamma radiation. Compatibility: Arthrex 3.5 mm and 4.0 mm screws.

Indications for Use

Indicated for patients with proximal humerus fractures, fracture dislocations, osteotomies, and non-unions, including those with osteopenic bone. The Cerclage Button is indicated as an adjunct for fracture stabilization in long bone fixation when used with the humeral plating system and FiberTape® cerclage suture.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ March 26, 2025 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Arthrex Inc. Ivette Galmez Principal Regulatory Affairs Specialist 1370 Creekside Blvd. Naples, Florida 34108 Re: K243995 Trade/Device Name: Arthrex Humeral Plating System and Cerclage Button Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, JDQ Dated: December 23, 2024 Received: December 26, 2024 Dear Ivette Galmez: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {3}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # CHRISTOPHER FERREIRA -S Christopher Ferreira, M.S. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Indications for Use Submission Number (if known) K243995 Device Name Arthrex Humeral Plating System and Cerclage Button Indications for Use (Describe) The Humeral Plating System is indicated for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone. The Cerclage Button is intended for use with the humeral plating system and FiberTape® cerclage suture, to augment fracture stabilization with humeral plates in long bone fixation. The cerclage button is designed for use with the humeral plating system and may not be used alone. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ ## K243995 510(k) Summary | Date Prepared | 12/23/2024 | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter | Arthrex Inc. | | | 1370 Creekside Boulevard | | | Naples, FL 34108-1945 | | Contact Person | Name: Ivette Galmez | | | Title: Principal Regulatory Affairs Specialist | | | Phone: 239-643-5553, ext. 71263 | | | Email: Ivette.galmez@arthrex.com | | Trade Name | Arthrex Humeral Plating System and Cerclage Button | | Common Name | Plate, Fixation, Bone | | | Cerclage, Fixation | | Product Code | HRS, JDQ | | Classification Name | 21 CFR 888.3030: Single/Multiple Component Metallic Bone Fixation | | | Appliances and Accessories | | | 21 CFR 888.3010: Bone fixation cerclage | | Regulatory Class | II | | Primary Predicate Device | K180310: Depuy Synthes LCP Proximal Humerus Plate | | Additional Predicate | K172975: Depuy Synthes Cerclage Positioning Pin | | Reference Devices | K242079: Arthrex Elbow Fracture Plating System | | Purpose of Submission | This Traditional 510(k) premarket notification is submitted to obtain clearance | | | for the Arthrex Humeral Plating System and Cerclage Button. | | Device Description | The Arthrex Humeral Plating and Cerclage Button consist of straight humeral | | | plates and a 3.5 threaded cerclage button for use in the repair of proximal | | | humerus fractures. The devices are provided sterile (gamma) and non-sterile, | | | for single use. The proposed plates range in length from 90.5 - 309 mm and | | | include suture holes for soft tissue refixation. The proposed devices are | | | manufactured from titanium alloy conforming to ASTM F136. The proposed | | | devices are compatible with Arthrex 3.5 mm screws (K241592 and K203294)<br>and Arthrex 4.0 mm screws (K150456, K143614, and K103705). | | | | | Indications for Use | The Arthrex Humeral Plating System is indicated for fractures and fracture | | | dislocations, osteotomies, and non-unions of the proximal humerus, | | | particularly in osteopenic bone. | | | The Cerclage Button is intended for use with the humeral plating system and<br>FiberTape® cerclage suture, to augment fracture stabilization with humeral | | | plates in long bone fixation. The cerclage button is designed for use with the | | | humeral plating system and may not be used alone. | | Performance Data | Arthrex conducted 4-point bend (ASTM F382-17) testing to demonstrate that | | | the Arthrex Humeral Plating System perform statistically equivalent to the<br>primary predicate device Depuy-Synthes LCP Plates (K180310).<br>MRI force, torque, and image artifact testing were conducted in accordance<br>with FDA guidance Testing and Labeling Medical Devices for Safety in the<br>Magnetic Resonance (MR) Environment, ASTM F2052 Standard Test Method<br>for Measurement of Magnetically Induced Displacement Force on Medical<br>Devices in the Magnetic Resonance Environment, ASTM F2119 Standard Test | | | Method for Evaluation of MR Image Artifacts from Passive Implants, ASTM<br>F2182 Standard Test Method for Measurement of Measurement of Radio<br>Frequency Induced Heating Near Passive Implants During Magnetic Resonance<br>Imaging and ASTM F2213 Standard Test Method for Measurement of<br>Magnetically Induced Torque on Medical Devices in the Magnetic Resonance<br>Environment.<br>Bacterial Endotoxins Test (BET) was performed on the Arthrex devices utilizing<br>the Kinetic Chromogenic Method in accordance with ANSI/AAMI<br>ST72:2011/(R)2016, USP <161>, USP <85>, EP 2.6.14. The testing conducted<br>demonstrates that the sterile devices meet pyrogen limit specifications.<br>Assessment of the physical product attributes including product, design, size,<br>and materials has determined that the Arthrex Humeral Plating System and<br>Cerclage Button do not introduce additional risks or concerns regarding<br>sterilization and shelf-life. | | Technological<br>Comparison | The main differences with the predicates are provided below:<br>• The indications for use for the proposed Cerclage Button are equivalent to<br>the Depuy-Synthes Cerclage Positioning Pin (K172975), except that the<br>indications for the proposed device have been updated to indicate that the<br>cerclage button is an adjunct device for use only with the proposed Arthrex<br>Humeral Plating System, and to include the recommended Arthrex suture<br>for use.<br>• The proposed humeral plates are manufactured from a titanium alloy that<br>is a different from the predicate humeral plates (K180310).<br>• The proposed humeral plates will be offered in lengths outside the<br>predicate humeral plates (K180310) size range. | | Conclusion | The Arthrex Humeral Plating System and Cerclage Button is substantially<br>equivalent to the primary Depuy Synthes LCP Proximal Humerus Plate<br>(K180310) and Depuy Synthes Cerclage Positioning Pin (K172975) in which the<br>basic design features and intended use are the same. The proposed devices<br>have the same fundamental scientific technology, and are made with<br>equivalent materials, design configuration, and sterility availability. Any<br>differences between the Arthrex Humeral Plating System and Cerclage Button,<br>and the predicate devices are considered minor and do not raise different<br>questions of safety or effectiveness.<br>Based on the indications for use, technological characteristics, and the<br>summary of data submitted, Arthrex Inc. has determined that the proposed<br>devices are substantially equivalent to the currently marketed predicate<br>devices. | {6}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...