Validation Set: 615 characterized samples (153 APS patients and 462 control samples).
—
Indications for Use
The Aptiva APS IgA Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-cardiolipin (aCL) and anti-beta 2 glycoprotein 1 (aβ2GPI) IgA autoantibodies in human serum as an aid in the diagnosis of primary and secondary antiphospholipid syndrome (APS), when used in conjunction with other laboratory and clinical findings. The Aptiva APS IgA Reagent is intended for use with the Aptiva System.
Device Story
Aptiva APS IgA Reagent is a solid-phase immunoassay using particle-based multi-analyte technology (PMAT) on the Aptiva System; a fully automated, random-access analyzer. The device uses paramagnetic microparticles coated with cardiolipin/β2GPI antigens and a control IgA capture bead. The system dilutes human serum samples, incubates with reagent beads, adds PE-conjugated anti-human IgA, and performs wash cycles. The optical module generates images to identify unique particle signatures and measure median fluorescent intensity (MFI), which is proportional to the concentration of bound IgA antibodies. Results are calculated via lot-specific master curves stored on the reagent cartridge RFID. The device is used in clinical laboratories by trained personnel to assist physicians in diagnosing APS. Output is a semi-quantitative result (FLU) used alongside clinical findings to support diagnostic decisions.
Clinical Evidence
Clinical validation involved 615 serum samples (153 APS, 462 differential diagnosis controls). For aCL IgA, clinical sensitivity was 31.4% and specificity 97.8%. For aβ2GPI IgA, clinical sensitivity was 32.0% and specificity 98.5%. Method comparison with predicates (n=332 for aCL IgA, n=362 for aβ2GPI IgA) showed high agreement, with sensitivity and specificity metrics consistent with the predicate devices.
Technological Characteristics
Particle-based multi-analyte technology (PMAT) using paramagnetic microparticles coated with native cardiolipin and β2GPI antigens. Fluorescent immunoassay principle using PE-conjugated anti-human IgA. Fully automated, random-access analyzer platform. Reagent cartridge includes RFID for master curve storage. Sterilization/materials not specified beyond protein stabilizers and sodium azide.
Indications for Use
Indicated for semi-quantitative determination of anti-cardiolipin (aCL) and anti-beta 2 glycoprotein 1 (aβ2GPI) IgA autoantibodies in human serum to aid in diagnosis of primary and secondary antiphospholipid syndrome (APS) in conjunction with other clinical findings.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K243979
B Applicant
Inova Diagnostics, inc.
C Proprietary and Established Names
Aptiva APS IgA Reagent
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| MID
MSV | Class II | 21 CFR 866.5660 – Multiple Autoantibodies Immunological Test System | IM – Immunology |
## II Submission/Device Overview:
A Purpose for Submission:
New Device
B Measurand:
Anti-cardiolipin IgA autoantibodies (aCL IgA)
Anti-beta 2 glycoprotein 1 IgA autoantibodies (aβ2GPI IgA)
C Type of Test:
Semi quantitative, Particle-based multi-analyte assay
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K243979 - Page 2 of 13
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
The Aptiva APS IgA Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-cardiolipin (aCL) and anti-beta 2 glycoprotein 1 (aβ2GPI) IgA autoantibodies in human serum as an aid in the diagnosis of primary and secondary antiphospholipid syndrome (APS), when used in conjunction with other laboratory and clinical findings.
The Aptiva APS IgA Reagent is intended for use with the Aptiva System.
### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
### D Special Instrument Requirements:
Inova Diagnostics Aptiva System (K193604)
## IV Device/System Characteristics:
### A Device Description:
The Aptiva APS IgA Reagent contains one cartridge with the following reagents for 200 determinations:
- APS IgA Beads (1×0.5 mL): Paramagnetic beads coated with native Cardiolipin (CL) plus β2GPI antigens, native β2GPI antigen, AffiniPure goat polyclonal anti-human IgA, and stabilizer
- Assay Buffer (1×17 mL): Containing protein stabilizers and preservatives
- PE Tracer IgA (1×17 mL): Phycoerythrin (PE) labeled goat anti-human IgA antibody with stabilizer and preservative
- Rehydration Buffer (1×6.5 mL): Containing protein stabilizer and preservative
### Materials Required but Not Provided
- Aptiva APS IgA Calibrators: 3 Calibrators (2×0.3mL/each) of human antibodies, each contains human aCL IgA and aβ2GPI IgA in stabilizers and preservatives
{2}
- Aptiva APS IgA Controls: 2 Controls (2×0.5 mL/each) of human antibodies, each contains human aCL IgA and aβ2GPI IgA in stabilizers and preservatives
## B Principle of Operation:
The Aptiva APS IgA reagent utilize particle based multi-analyte technology (PMAT) in a cartridge format and contains two different populations of particles; one particle population coated with cardiolipin, and another particle population coated with human purified β2GPI. The two analyte microparticles, along with the control microparticle, are stored in the reagent cartridge for the Aptiva instrument.
The Aptiva System dilutes the sample 1:8, then combines an aliquot of diluted sample, and reagent into a cuvette. The mixture is incubated at 37°C. After a wash cycle, conjugated anti-human IgA antibodies are added to the particles and this mixture is incubated at 37°C. Excess conjugate is removed in another wash cycle, and the particles are re-suspended in system fluid. Multiple images are generated by the system to identify and count the two (2) unique analyte particles, as well as determine the amount of conjugate on each particle. A third particle, coated with goat anti-human IgA antibodies, is present in the reagent as a control to flag low concentrations of IgA in the sample as an assay verification step. The median fluorescent intensity (MFI) for each analyte is proportional to the concentration of conjugate bound to human IgA, which is proportional to the concentration of IgA antibodies bound to the corresponding particle population. The system uses the MFI from at least 50 particles of each population. The identity of the particles is determined by the unique signature of the particles.
Each analyte in the Aptiva APS IgA Reagent is assigned a predefined lot specific master curve. The analyte specific master curve is stored on the reagent cartridge RFID label. Based on results obtained by running calibrators (supplied separately), the system creates individual working curves. Working curves are used by the software to calculate Fluorescent Light Units (FLU) for each analyte from the MFI values obtained for each sample.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
QUANTA Flash aCL IgA
QUANTA Flash β2GP1 IgA
### B Predicate 510(k) Number(s):
K120817
### C Comparison with Predicate(s):
K243979 - Page 3 of 13
{3}
| Device & Predicate Device(s): | K243979
(Candidate Device) | K120817
(Predicate Device) |
| --- | --- | --- |
| Device Trade Name | Aptiva APS IgA Reagent | QUANTA Flash aCL IgA,
QUANTA Flash β2GP1 IgA |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The Aptiva APS IgA Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-cardiolipin (aCL) and anti-beta 2 glycoprotein I (aβ2GPI) IgA autoantibodies in human serum as an aid in the diagnosis of primary and secondary antiphospholipid syndrome (APS), when used in conjunction with other laboratory and clinical findings.
The Aptiva APS IgA Reagent is intended for use with the Aptiva System. | Fully automated chemiluminescent immunoassay for the semi-quantitative measurement of anti-cardiolipin (aCL)/anti-β2 glycoprotein-I (β2GPI) IgA antibodies in human citrated plasma and serum on the BIO-FLASH instrument, as an aid in the diagnosis of thrombotic disorders related to primary and secondary antiphospholipid syndrome (APS), when used in conjunction with other laboratory and clinical findings. |
| Type of Test | Semi-quantitative | Same |
| Quality Control | Two Control Levels | Same |
| General Device Characteristic Differences | | |
| Instrumentation | Aptiva System | BIO-FLASH instrument |
| Technology | Fluorescent immunoassay | Chemiluminescent immunoassay |
| Clinical Cut-off | 5.00 FLU | 20.0 CU |
| Analytical Measuring Interval (AMI) | 0.41 – 68.00 FLU (aCL IgA)
0.66 – 80.00 FLU (aβ2GPI IgA) | 1.4 – 351.6 CU (aCL IgA)
4.0 – 512 CU (β2GP1 IgA) |
| Calibrator | Three Calibrator Levels | Two Calibrator Levels |
| Conjugate | Phycoerythrin conjugated polyclonal anti-human IgA antibody | Isoluminol conjugated monoclonal anti-human IgA antibody |
| Sample Type | Serum | Serum or Citrated Plasma |
VI Standards/Guidance Documents Referenced:
The following Clinical and Laboratory Standards Institute (CLSI) guidelines were used:
- CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures, Approved Guideline – Third Edition.
K243979 - Page 4 of 13
{4}
- CLSI EP06-Ed2: Evaluation of the Linearity of Quantitative Measurement Procedures – Second Edition
- CLSI EP07-A3: Interference Testing in Clinical Chemistry – Third Edition.
- CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures, Approved Guideline – Second Edition.
- CLSI EP25-Ed2: Evaluation of Stability of In Vitro Medical Laboratory Test Reagents – Second Edition.
- CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory, Approved Guideline – Third Edition.
- CLSI EP37: Supplemental Tables for Interference Testing in Clinical Chemistry
- CLSI EP39: A Hierarchical Approach to Selecting Surrogate Samples for the Evaluation of In Vitro Medical Laboratory Tests
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
#### 1. Precision/Reproducibility:
Precision and reproducibility of the Aptiva APS IgA Reagent was evaluated in accordance with the CLSI guideline EP05-A3.
**Within-Laboratory Precision:**
Within-lab precision of the Aptiva APS IgA reagent was performed by testing six serum samples containing various concentrations of aCL IgA and aβ2GPI IgA autoantibodies. All samples were run in duplicate, twice a day, for 20 days using single lot of reagents on a single instrument. Data was analyzed for repeatability (within-run), between-run, between-day and within-laboratory precision. Results are summarized below.
| aCL IgA | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | N | Mean (FLU) | Repeatability | | Between-Run | | Between-Day | | Within-Laboratory | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 80 | 1.49 | 0.07 | 5.0 | 0.05 | 3.3 | 0.08 | 5.6 | 0.12 | 8.2 |
| 2 | 80 | 3.56 | 0.19 | 5.5 | 0.05 | 1.4 | 0.23 | 6.5 | 0.31 | 8.6 |
| 3 | 80 | 5.50 | 0.24 | 4.3 | 0.12 | 2.2 | 0.32 | 5.8 | 0.42 | 7.6 |
| 4 | 80 | 11.64 | 0.30 | 2.6 | 0.27 | 2.3 | 0.51 | 4.4 | 0.66 | 5.6 |
| 5 | 80 | 24.94 | 0.93 | 3.7 | 1.26 | 5.1 | 1.15 | 4.6 | 1.94 | 7.8 |
| 6 | 80 | 45.39 | 1.32 | 2.9 | 1.50 | 3.3 | 2.16 | 4.8 | 2.94 | 6.5 |
K243979 - Page 5 of 13
{5}
| aβ2GPI IgA | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | N | Mean (FLU) | Repeatability | | Between-Run | | Between-Day | | Within-Laboratory | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 80 | 2.05 | 0.14 | 6.7 | 0.14 | 6.9 | 0.12 | 5.7 | 0.23 | 11.2 |
| 2 | 80 | 4.82 | 0.30 | 6.2 | 0.13 | 2.7 | 0.45 | 9.3 | 0.55 | 11.5 |
| 3 | 80 | 7.43 | 0.35 | 4.7 | 0.22 | 2.9 | 0.68 | 9.2 | 0.80 | 10.7 |
| 4 | 80 | 15.58 | 0.49 | 3.2 | 0.42 | 2.7 | 0.83 | 5.3 | 1.05 | 6.7 |
| 5 | 80 | 28.44 | 1.08 | 3.8 | 1.23 | 4.3 | 1.87 | 6.6 | 2.49 | 8.7 |
| 6 | 80 | 56.85 | 1.49 | 2.6 | 2.06 | 3.6 | 3.12 | 5.5 | 4.03 | 7.1 |
Lot-to-Lot Imprecision:
The lot-to-lot imprecision of the Aptiva APS IgA Reagent was evaluated by testing five serum samples for aCL IgA and a $\beta$ 2GPI IgA with three lots of the Aptiva APS IgA Reagent. Samples were tested in replicates of five, once a day, for five days using one instrument to generate a total of 75 data points for each sample. The results are summarized in the tables below.
| aCL IgA | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | N | Mean (FLU) | Within-Day | | Between-Day | | Between-Lot | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 75 | 1.97 | 0.17 | 8.4 | 0.09 | 4.6 | 0.02 | 1.0 | 0.19 | 9.6 |
| 2 | 75 | 5.01 | 0.33 | 6.6 | 0.22 | 4.4 | 0.00 | 0.0 | 0.40 | 7.9 |
| 3 | 75 | 5.09 | 0.22 | 4.4 | 0.15 | 2.8 | 0.05 | 0.9 | 0.27 | 5.3 |
| 4 | 75 | 17.42 | 0.90 | 5.2 | 0.39 | 2.2 | 0.42 | 2.4 | 1.07 | 6.1 |
| 5 | 75 | 50.12 | 1.67 | 3.3 | 0.85 | 1.7 | 4.01 | 8.0 | 4.43 | 8.8 |
| aβ2GPI IgA | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | N | Mean (FLU) | Within-Day | | Between-Day | | Between-Lot | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 75 | 2.12 | 0.19 | 9.0 | 0.12 | 5.5 | 0.11 | 5.1 | 0.25 | 11.7 |
| 2 | 75 | 5.53 | 0.29 | 5.3 | 0.22 | 4.0 | 0.34 | 6.1 | 0.50 | 9.0 |
| 3 | 75 | 5.63 | 0.28 | 4.9 | 0.10 | 1.8 | 0.25 | 4.5 | 0.39 | 6.9 |
| 4 | 75 | 26.53 | 1.21 | 4.6 | 0.61 | 2.3 | 1.15 | 4.3 | 1.78 | 6.7 |
| 5 | 75 | 60.69 | 1.95 | 3.2 | 3.63 | 6.0 | 2.99 | 4.9 | 5.09 | 8.4 |
Site-to-Site Precision (Reproducibility):
The reproducibility of the Aptiva APS IgA Reagent was conducted at three different sites/instruments using seven serum samples for aCL IgA and a $\beta$ 2GPI IgA. Samples were tested in replicates of five, once a day, for five days using one reagent lot to generate a total of 75 replicates per sample. The results are summarized below.
K243979 - Page 6 of 13
{6}
| aCL IgA | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | N | Mean (FLU) | Within-Day | | Between-Day | | Between-Site | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 75 | 2.21 | 0.11 | 5.1 | 0.04 | 1.7 | 0.13 | 5.8 | 0.18 | 7.9 |
| 2 | 75 | 4.85 | 0.19 | 3.8 | 0.15 | 3.2 | 0.24 | 5.0 | 0.34 | 7.1 |
| 3 | 75 | 5.45 | 0.23 | 4.3 | 0.19 | 3.5 | 0.28 | 5.1 | 0.41 | 7.6 |
| 4 | 75 | 7.84 | 0.35 | 4.4 | 0.24 | 3.1 | 0.50 | 6.3 | 0.65 | 8.3 |
| 5 | 75 | 16.51 | 0.56 | 3.4 | 0.60 | 3.6 | 1.39 | 8.4 | 1.61 | 9.8 |
| 6 | 75 | 37.16 | 1.27 | 3.4 | 0.97 | 2.6 | 3.91 | 10.5 | 4.23 | 11.4 |
| 7 | 75 | 54.80 | 1.95 | 3.6 | 0.91 | 1.7 | 2.06 | 3.8 | 2.98 | 5.4 |
| aβ2GPI IgA | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | N | Mean (FLU) | Within-Day | | Between-Day | | Between-Site | | Total | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 75 | 2.50 | 0.13 | 5.1 | 0.08 | 3.2 | 0.16 | 6.5 | 0.22 | 8.8 |
| 2 | 75 | 5.10 | 0.20 | 3.9 | 0.17 | 3.4 | 0.30 | 5.9 | 0.40 | 7.8 |
| 3 | 75 | 6.41 | 0.24 | 3.7 | 0.22 | 3.4 | 0.41 | 6.4 | 0.52 | 8.2 |
| 4 | 75 | 9.36 | 0.42 | 4.5 | 0.17 | 1.8 | 0.65 | 7.0 | 0.80 | 8.5 |
| 5 | 75 | 19.77 | 0.56 | 2.8 | 0.65 | 3.3 | 1.63 | 8.3 | 1.84 | 9.3 |
| 6 | 75 | 35.86 | 1.06 | 2.9 | 1.06 | 2.9 | 3.13 | 8.7 | 3.47 | 9.7 |
| 7 | 75 | 64.91 | 1.95 | 3.0 | 2.98 | 4.6 | 1.91 | 2.9 | 4.04 | 6.2 |
## 2. Linearity:
The linearity of the Aptiva APS IgA Reagent according to CLSI EP06-Ed2.
For the Aptiva APS IgA Reagent, multiple dilution series were prepared to cover the AMI of each of the aCL IgA and a $\beta 2\mathrm{GPI}$ IgA assays by diluting human serum samples with various antibody concentrations in human negative serum. Four unique specimens were used for aCL IgA and three for a $\beta 2\mathrm{GPI}$ IgA, with dilutions made in $10\%$ increments that overlapped other samples' dilution series. Each sample was tested in duplicate using two lots of the Aptiva APS IgA Reagent on the Aptiva Multi-Analyte Instrument. The overall testing range for the linearity study was from 0.41–301.63 FLU for the aCL IgA assay, and 0.57–81.66 FLU for the a $\beta 2\mathrm{GPI}$ IgA assay. The percent deviation from the weighted least squares regression analysis was used to assess the fit of the regression for each sample and analyte.
| aCL IgA | | | | | |
| --- | --- | --- | --- | --- | --- |
| Sample | Test Range (FLU) | Slope (95% CI) | R² | % Deviation from Linearity | FLU of Deviation from Linearity# |
| 1 | 30.16–301.63 | 1.07 (1.03–1.12) | 0.99 | -6.8% to 10.4%* | N/A |
| 2 | 5.51–55.11 | 1.00 (0.97–1.03) | 1.00 | -7.9% to 5.0% | N/A |
| 3 | 2.72–27.20 | 1.02 (1.00–1.03) | 1.00 | -1.8% to 6.6% | 0.48 FLU |
| 4 | 0.41–4.05 | 1.00 (0.95–1.05) | 1.00 | -14.1%** to 6.4% | -0.19 to -0.15 FLU |
| Combined | 301.63 – 0.41 | 0.93 (0.91 – 0.95) | 1.00 | -14.1% to 10.4% | -0.19 to 0.48 FLU |
# For sample below 5 FLU.
* % Deviation from linearity of $10.4\%$ for sample level with aCL IgA concentration of 60.33 FLU
** % Deviation from linearity of -14.1% for sample level with aCL IgA concentration of 1.62 FLU
K243979 - Page 7 of 13
{7}
K243979 - Page 8 of 13
| aβ2GPI IgA | | | | | |
| --- | --- | --- | --- | --- | --- |
| Sample | Test Range (FLU) | Slope (95% CI) | R² | % Deviation from Linearity | FLU of Deviation from Linearity# |
| 1 | 1.26–12.57 | 0.94 (0.90–0.98) | 0.99 | -11.7%* to 6.7% | -0.48 FLU |
| 2 | 8.17–81.66 | 1.00 (0.97–1.03) | 1.00 | -13.9%** to 5.1% | N/A |
| 3 | 0.57–5.71 | 0.99 (0.94–1.04) | 0.99 | -4.2% to 6.1% | -0.44 to -0.18 FLU |
| Combined | 81.66 – 0.57 | 0.99 (0.97 – 1.00) | 1.00 | -13.9% to 6.7% | -0.48 to -0.18 FLU |
# For sample below 5 FLU.
* % Deviation from linearity of -11.7% for sample level with aβ2GPI IgA concentration of 3.77 FLU
** % Deviation from linearity of -13.9% for sample level with aβ2GPI IgA concentration of 16.33 FLU
The data summarized in the tables below support the linearity of AMI of 0.41–68.00 FLU for the aCL IgA assay and the AMI of 0.66–80.00 FLU for the aβ2GPI IgA assay, as part of the Aptiva APS IgA Reagent.
## 3. Hook effect
The hook effect for Aptiva APS IgA Reagent on the Aptiva System was evaluated by testing high positive serum samples for each device: three high aCL IgA samples and three high aβ2GPI IgA samples for the Aptiva APS IgA Reagent. No antigen excess hook effect was observed up to 98.19 FLU for aCL IgA and 133.48 FLU for aβ2GPI IgA.
## 4. Analytical Specificity/Interference:
An interference study was performed according to CLSI EP07-A3 and CLSI EP37-A1 for the Aptiva APS IgA Reagent. A set of human serum specimens—one positive, one around the cut-off (±25%), and one negative sample for each assay were used to prepare interferent spiked samples or corresponding control samples without the interfering substance. All samples were tested in five replicates using the Aptiva APS IgA Reagent for both aCL IgA and aβ2GPI IgA analytes. The percent recovery for each sample spiked with the potential interfering substance was calculated by comparing its result to that of the corresponding control sample without the interfering substance. Based on the results, the following table displays the highest concentration at which no interference was observed for both analytes in the Aptiva APS IgA Reagent assay.
| Aptiva APS IgA Reagent (aCL IgA / aβ2GPI IgA) | |
| --- | --- |
| Endogenous Interfering Substances | Concentration without Interference |
| Bilirubin, Conjugated | 100.0 mg/dL |
| Hemoglobin | 10.0 g/L |
| Triglyceride | 1000.0 mg/dL |
| Cholesterol | 332.5 mg/dL |
| RF IgM | 196.5 IU/mL |
| Human IgG | 2000.0 mg/dL |
{8}
| Aptiva APS IgA Reagent (aCL IgA / aβ2GPI IgA) | |
| --- | --- |
| Exogenous Interfering Substances | Concentration without Interference |
| Ibuprofen | 21.9 mg/dL |
| Warfarin | 7.5 mg/dL |
| Prednisone | 0.0099 mg/dL |
| Acetaminophen | 15.6 mg/dL |
| Aspirin | 3.0 mg/dL |
| Hydroxychloroquine | 0.465 mg/dL |
| Omeprazole | 0.84 mg/dL |
| Simvastatin | 0.168 mg/dL |
| Heparin | 330.0 units/dL |
# 5. Assay Reportable Range:
The reportable range for the Aptiva APS IgA Reagent is the same as the analytical measuring interval (AMI) for both measurands. The AMI for each measurand is shown in the table below:
| Aptiva APG IgA Reagent | Analytical Measuring Interval |
| --- | --- |
| aCL IgA | 0.41 – 68.00 FLU |
| aβ2GPI IgA | 0.66 – 80.00 FLU |
# 6. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Traceability:
There are no international reference materials for anti-cardiolipin antibodies and anti- $\beta 2$ -GPI antibodies. Calibrator and control values are directly traceable to in-house reference materials that are used to create the master curves for each analyte of the Aptiva APS IgA Reagent.
Stability:
Reagent shelf-life stability: An on-going real-time stability and an accelerated stability study were conducted to establish shelf-life stability for the Aptiva APS IgA Reagent. The accelerated study was performed by testing three reagent lots with incubation at $37^{\circ}\mathrm{C}$ and $47^{\circ}\mathrm{C}$ . Five serum samples spanning the analytical measuring range for both aCL IgA and a $\beta$ 2GPI IgA analytes were tested at multiple timepoints. The collected data showed the initial shelf-life of 11 months for the Aptiva APS IgA Reagent when stored at $2 - 8^{\circ}\mathrm{C}$ .
Reagent in-use (on-board) stability: An in-use stability study for the Aptiva APS IgA Reagent was conducted by testing multiple samples with different assay levels at different time points using one lot of reagent stored on board. The in-use (on-board) stability of the Aptiva APS IgA Reagent was set at 28 days, with a 14-day recalibration.
Reagent shipping stability: A transport simulation study was performed to assess the shipping stability of the Aptiva APS IgA Reagent kits during transport. The study was performed under simulated conditions, including high-temperature exposure and thermal
K243979 - Page 9 of 13
{9}
cycling, that the products may potentially be exposed to during transport. The collected data support that shipping conditions at 2–8°C mitigate the potential impact of prolonged or repeated temperature stressing during transport.
**Sample stability:** Serum samples with different antibody levels were stored at various storage temperatures and tested with the Aptiva APS IgA Reagent at various timepoints. The collected data showed that the sample can be stored up to 14 days at 2–8°C, up to 48 hours at room temperature (20–26°C), and up to four freeze/thaw cycles when stored at -20°C or lower.
## 7. Detection Limit:
Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) were assessed for each analyte in the Aptiva APS IgA Reagent following CLSI EP17-A2.
LoB: To determine the LoB, four blank samples were run in five replicates, once a day for three days using two reagent lots, for a total of 60 data points per lot. The LoB was determined for each analyte, on each reagent lot separately, at the 95th percentile using the non-parametric method for all analyses.
LoD: To determine the LoD for each analyte, five low level samples were assayed in five replicates, twice per day for three days using two reagent lots, to generate a total of 150 data points for each analyte on each reagent lot. The LoD was calculated using the equation $$ \text{LoD} = \text{LoB} + \text{SD}_\text{L} \times c_p $$ where $$ \text{SD}_\text{L} $$ is the pooled standard deviation of the low-level samples and $$ c_p $$ is a multiplier for the 95th percentile.
LoQ: To determine the LoQ for each analyte, five low level samples were run in five replicates, twice per day for three days using two reagent lots, to generate 150 data points for each assay on each reagent lot. The LoQ was determined separately for each assay on each reagent lot by calculating the total imprecision of each sample. The LoQ was defined as the lowest concentration level that meets the within-laboratory imprecision of ≤20% for each lot.
The claimed LoB, LoD, and LoQ for the Aptiva APS IgA Reagent are summarized in the following table:
| | LoB (FLU) | LoD (FLU) | LoQ (FLU) |
| --- | --- | --- | --- |
| Aptiva APS IgA Reagent | | | |
| aCL IgA | 0.06 | 0.11 | 0.41 |
| aβ2GPI IgA | 0.38 | 0.48 | 0.66 |
## 8. Assay Cut-Off:
The following cut-off is used for the aCL IgA and aβ2GPI IgA assays in the Aptiva APS IgA Reagent:
Positive: ≥5.00 FLU
Negative: <5.00 FLU
K243979 - Page 10 of 13
{10}
B Comparison Studies:
1. Method Comparison with Predicate Device:
The samples for the Aptiva APS IgA method comparison analysis were serum samples from the clinical validation study (see VII.C) and additional samples known to be around the cutoff that were within the analytical measuring interval of the assay and the respective predicate devices (QUANTA Flash aCL IgA and QUANTA Flash β2GP1 IgA). Positive percent agreement (PPA), negative percent agreement (NPA), and total percent agreement (TPA) with 95% confidence intervals (95%CI) were calculated for each analyte comparison, excluding values that were outside of the measuring ranges of either assay. The results are summarized in the following tables:
| aCL IgA | | | | |
| --- | --- | --- | --- | --- |
| | QUANTA Flash aCL IgA | | | |
| | | Positive | Negative | Total |
| Aptiva APS IgA Reagent (aCL IgA) | Positive | 31 | 6 | 37 |
| | Negative | 4 | 205 | 209 |
| | Total | 35 | 211 | 246 |
| PPA: 88.6% (31/35) (95% CI: 74.0–95.5%)
NPA: 97.2% (205/211) (95% CI: 93.9–98.7%)
TPA: 95.9% (236/246) (95% CI: 92.7–97.8%) | | | | |
| aB2GPI IgA | | | | |
| --- | --- | --- | --- | --- |
| | QUANTA Flash aCL IgA | | | |
| | | Positive | Negative | Total |
| Aptiva APS IgA Reagent (aB2GPI IgA) | Positive | 37 | 9 | 46 |
| | Negative | 3 | 51 | 54 |
| | Total | 40 | 60 | 100 |
| PPA: 92.5% (37/40) (95% CI: 80.1–97.4%)
NPA: 85.0% (51/60) (95% CI: 73.0–91.9%)
TPA: 88.0% (88/100) (95% CI: 80.2–93.0%) | | | | |
2. Matrix Comparison:
Not applicable. Only human serum specimens are intended use sample type for the Aptiva APS IgA Reagent.
C Clinical Studies:
1. Clinical Sensitivity and Specificity:
The clinical validation study for the Aptiva APS IgA Reagent involved 615 characterized serum samples. This cohort included 70 samples from patients with primary antiphospholipid syndrome (pAPS), 83 samples from patients with secondary antiphospholipid syndrome (sAPS), and 462 samples from patients with various autoimmune and infectious diseases that could be considered in the differential diagnosis of APS. All samples were tested using the Aptiva APS IgA Reagent and the clinical performance of the Aptiva APS IgA Reagent, as an aid in the diagnosis of APS, is summarized in the following tables.
K243979 - Page 11 of 13
{11}
| aCL IgA | | | | |
| --- | --- | --- | --- | --- |
| | Clinical Diagnosis | | | |
| | | APS | Non-APS | Total |
| Aptiva APS IgA Reagent (aCL IgA) | Positive (≥5.0 FLU) | 48 | 10 | 58 |
| | Negative (<5.0 FLU) | 105 | 452 | 557 |
| | Total | 153 | 462 | 615 |
| Clinical Sensitivity: 31.4% (48/153) (95% CI: 24.6–39.1%)
Clinical Specificity: 97.8% (452/462) (95% CI: 96.1–98.8%) | | | | |
| aβ2GPI IgA | | | | |
| --- | --- | --- | --- | --- |
| | Clinical Diagnosis | | | |
| | | APS | Non-APS | Total |
| Aptiva APS IgA Reagent (aβ2GPI IgA) | Positive (≥5.0 FLU) | 49 | 7 | 56 |
| | Negative (<5.0 FLU) | 104 | 455 | 559 |
| | Total | 153 | 462 | 615 |
| Clinical Sensitivity: 32.0% (49/153) (95% CI: 25.2–39.8%)
Clinical Specificity: 98.5% (455/462) (95% CI: 96.9–99.3%) | | | | |
Distribution of target and differential disease samples and antibody positive rates are shown in the table below:
| Aptiva APS IgA Reagent | | | | | |
| --- | --- | --- | --- | --- | --- |
| Diagnostic Group | N=615 | aCL IgA | | aβ2GPI IgA | |
| | | n | (%) | n | (%) |
| Target diagnosis | 153 | 48 | 31.4% | 49 | 32.0% |
| pAPS | 70 | 14 | 20.0% | 17 | 24.3% |
| sAPS | 83 | 34 | 41.0% | 32 | 38.6% |
| Differential diagnostic controls | 462 | 10 | 2.2% | 7 | 1.5% |
| Viral Infectious Disease | 53 | 2 | 3.8% | 0 | 0.0% |
| Bacterial Infectious Disease | 45 | 2 | 4.4% | 0 | 0.0% |
| Myositis | 24 | 0 | 0.0% | 0 | 0.0% |
| Preeclampsia or Placental Insufficiency (PREPI) | 36 | 0 | 0.0% | 0 | 0.0% |
| Systemic Lupus Erythematosus (without APS) | 34 | 1 | 2.9% | 1 | 2.9% |
| Inflammatory Bowel Disease (IBD) | 19 | 0 | 0.0% | 0 | 0.0% |
| Rheumatoid Arthritis (RA) | 20 | 0 | 0.0% | 0 | 0.0% |
| Fetal loss (without APS) | 13 | 0 | 0.0% | 0 | 0.0% |
| Thrombosis (without APS) | 1 | 0 | 0.0% | 0 | 0.0% |
| ANCA-Associated Vasculitis | 18 | 0 | 0.0% | 0 | 0.0% |
| Autoimmune Thyroiditis | 49 | 1 | 2.0% | 2 | 4.1% |
| Celiac Disease (CeD) | 44 | 1 | 2.3% | 0 | 0.0% |
| COVID-19 Related Thrombosis | 22 | 0 | 0.0% | 0 | 0.0% |
| Hematologic Malignancies | 20 | 1 | 5.0% | 1 | 5.0% |
| Idiopathic thrombocytopenic purpura (ITP) | 14 | 1 | 7.1% | 1 | 7.1% |
| Solid Tumor Malignancies | 20 | 0 | 0.0% | 1 | 5.0% |
| Deep Vein Thrombosis | 18 | 1 | 5.6% | 1 | 5.6% |
| Myocardial Infarction (MI) | 12 | 0 | 0.0% | 0 | 0.0% |
K243979 - Page 12 of 13
{12}
K243979 - Page 13 of 13
2. Comparison of the Aptiva APS IgA and Predicates
Clinical Performance of the Aptiva APS IgA was compared with the predicates using the same samples for aCL IgA (n=332) and aβ2GPI IgA (n=362). The clinical sensitivity and specificity are summarized in the following table.
| Assay | n= | Sensitivity | Specificity |
| --- | --- | --- | --- |
| aCL IgA | | | |
| Aptiva APS IgA | 332 | 25.4% (18.3 - 34.1%) | 97.2% (94.1 - 98.7%) |
| QUANTA Flash aCL IgA | | 26.3% (19.1 - 35.1%) | 98.6% (96.0 - 99.5%) |
| aβ2GPI IgA | | | |
| Aptiva APS IgA | 362 | 33.8% (26.3 - 42.2%) | 97.4% (94.4 - 98.8%) |
| QUANTA Flash β2GP1 IgA | | 30.8% (23.6 - 39.1%) | 98.3% (95.6 - 99.3%) |
D Clinical Cut-Off:
Refer to Assay Cut-off
E Expected Values/Reference Range:
To determine the reference range of the Aptiva APS IgA, a panel of 126 apparently healthy blood donors (ages 18–70 years, with a mean age of 38 years and median age of 35 years) were tested and analyzed following CLSI C28-A3c. For Aptiva APS IgA aCL IgA, with a cut-off of 5.00 FLU, two samples (1.6%) were positive. For aβ2GPI IgA, with a cut-off of 5.00 FLU, no samples were positive. The results are summarized in the following table:
| | N | Min (FLU) | Max (FLU) | Median (FLU) | 97.5^{th} Percentile (FLU) |
| --- | --- | --- | --- | --- | --- |
| aCL IgA | 126 | 0.41 | 9.13 | 0.41 | 2.83 |
| aβ2GPI IgA | 126 | 0.66 | 1.52 | 0.66 | 0.77 |
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.