Electrically Powered Wheelchair

K243962 · Kunshan Hi-Fortune Health Products Co., Ltd. · ITI · Feb 12, 2025 · Physical Medicine

Device Facts

Record IDK243962
Device NameElectrically Powered Wheelchair
ApplicantKunshan Hi-Fortune Health Products Co., Ltd.
Product CodeITI · Physical Medicine
Decision DateFeb 12, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 890.3860
Device ClassClass 2

Intended Use

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Device Story

Electrically powered wheelchair (HP558) provides mobility for disabled or elderly users; operates via user-controlled joystick; inputs include joystick directional commands; transforms inputs into motor drive signals for rear wheels; features DC brushless motors, electromagnetic braking, and magnesium alloy frame; foldable design for transport; used in indoor/outdoor environments; output is physical movement of wheelchair; assists users in daily mobility; healthcare provider uses output to facilitate patient independence.

Clinical Evidence

Bench testing only. No clinical data. Performance verified through testing against ISO 7176 series (stability, braking, energy consumption, dimensions, speed, strength, climatic, obstacle climbing, power/control systems, EMC, batteries/chargers) and ISO 16840-10 (ignition resistance). Biocompatibility testing performed per ISO 10993-5, -10, and -23.

Technological Characteristics

Magnesium alloy frame; DC brushless motors (24Vdc, 170W); M7084 controller; solid PU tires; Li-ion battery (25.2VDC, 10.4Ah); automatic electromagnetic brake system; foldable form factor; max load 120kg. Complies with ISO 7176 series and IEC 60601-1-2.

Indications for Use

Indicated for disabled or elderly persons limited to a seated position requiring indoor and outdoor mobility assistance.

Regulatory Classification

Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The Department of Health and Human Services logo is a stylized representation of a human figure, while the FDA acronym and name are written in blue, with the word "FDA" in a square. February 12, 2025 Kunshan Hi-Fortune Health Products Co., Ltd % Ariel Xiang Consultant Shanghai Sungo Management Consulting Co. Ltd. 14th Floor, 1500# Century Avenue Shanghai, Shanghai 200122 China Re: K243962 Trade/Device Name: Electrically Powered Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: December 20, 2024 Received: December 23, 2024 Dear Ariel Xiang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/2/Picture/3 description: The image shows the name "Tushar Bansal -S" in a large, sans-serif font. The text is black and stands out against a white background. The letters are evenly spaced and easy to read. The overall impression is clean and professional. for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K243962 Device Name Electrically Powered Wheelchair Indications for Use (Describe) It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Kunshan Hi-Fortune Health Products Co., Ltd. No.625 Juxiang Road zhangpu town , Kunshan City ,Jiangsu Province, China ## 510(K) Summary K243962 ## Document Prepared Date: 2025/02/11 #### A. Applicant Kunshan Hi-Fortune Health Products Co., Ltd Address:NO.625 Juxiang Road, Zhangpu Town,Kunshan City, Jiangsu, 215321, CHINA Contact Person: zhanli.Tian Email:zhanli.tian@hi-fortune.net Submission Correspondent: Primary contact: Ms.Ariel Xiang Title: Senior Consultant Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: shouqiu.xiang@sungoglobal.com Secondary contact: Mr. Raymond Luo Title: Technical Director Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda@sungoglobal.com #### Device B. Trade Name: Electrically powered wheelchair Common Name: Powered wheelchair Models: HP558 Regulatory Information Classification Name: Powered wheelchair Classification: Class II. Product code: ITI Regulation Number: 890.3860 Review Panel: Physical Medicine {5}------------------------------------------------ #### C. Predicate device K212503 Electrically powered wheelchair,HP458E Kunshan Hi-Fortune Health Products Co., Ltd. #### D. Indications for use of the device lt is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. #### ய் Device Description This product consists of frame, wheels, seat, armrest, lithium battery, motor and controller with a lightweight and compact design. The whole wheelchair can be folded and it can be easily carried or rolled after folding. The seat cushion is detachable. The armrest can be flipped upside down, which is convenient for the elderly to move. Users can drive the wheelchair by themselves through the control device. The wheelchair uses lithium batteries as its power source. The controls the drive left/right motor to realize the wheelchair forward, backward and turn functions. The frame of the device is Magnesium alloy. The front wheels are driven wheels suitable for rotation, acceleration, retrograde and other actions of the wheelchair. The front wheels movement will be achieved by thrust generated from the rear wheels are driving wheels to control the speed and direction. The wheels are Solid PU tires.When in use, the operator drives the motor of the rear wheel by operating the controller joystick to achieve the rear wheels movement. The DC brushless motor and brake system are fixed on the rear wheels. The max loading of the device is 120KG. Only for one person sit. #### F. Comparison to predicate device #### a) There are three changes to the proposed device: | changes | Proposed Device | Predicate Device (K212503) | |-------------------------------------------|----------------------------------------|------------------------------------| | 1.Change Motor | DC Brushless Motor,<br>24Vdc 170W;2pcs | DC Brush Motor,<br>24Vdc 120W;2pcs | | 2. Change controller | M7084 | PG Drives Technology Ltd., VR2 | | 3. Change the Main frame<br>material size | Magnesium alloy | Aluminium alloy | #### Technical comparison b) {6}------------------------------------------------ # Table 1 General Comparison | Elements of<br>Comparison | Proposed<br>Device(K243962) | Predicate Device<br>(K212503) | Remark | |---------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | Kunshan Hi-Fortune<br>Health Products Co.,<br>Ltd | Kunshan Hi-Fortune<br>Health Products Co., Ltd | -- | | Common or<br>Usual name | Electrically powered<br>wheelchair | Electric wheelchair | -- | | Model(s) | HP558 | HP458E | -- | | Classification | Class II | Class II | Same | | Classification<br>regulation | 21 CFR890.3860 | 21 CFR890.3860 | Same | | Product code | ITI | ITI | Same | | Indications for<br>use | It is a motor<br>driven, indoor<br>and outdoor<br>transportation<br>vehicle with the<br>intended use to<br>provide mobility<br>to a disabled or<br>elderly person<br>limited to a<br>seated position. | It is a motor driven,<br>indoor and outdoor<br>transportation vehicle<br>with the intended use to<br>provide mobility to a<br>disabled or elderly<br>person limited to a<br>seated position. | Same | | Use condition | indoor and<br>outdoor use | indoor and outdoor use | Same | | Number of<br>wheels | 6, including two<br>front wheels and<br>two rear<br>wheels, two anti-<br>tip wheels | 6, including two<br>front wheels and<br>two rear wheels,<br>two anti-tip wheels | Same | | Function of<br>wheels | Front wheels:<br>driven wheels<br>suitable for<br>rotation,a<br>cceleratio<br>n,<br>retrograde | Front wheels:<br>driven wheels<br>suitable for<br>rotation,acce<br>leration,<br>retrograde<br>Rear wheels: driving | Same | | | Rear wheels: driving wheels<br>to control the speed and<br>direction<br>Anti-tip wheels:<br>preventing the<br>wheelchair from<br>tipping turning<br>over when driving<br>on the slope. Non-<br>adjustable | wheels<br>to control the speed and<br>direction<br>Anti-tip wheels:<br>preventing the<br>wheelchair from<br>tipping turning over<br>when driving on the<br>slope. Non-<br>adjustable | | | Product<br>structure | Consists of two<br>foldable armrests, a<br>backrest, a seat<br>cushion, a foldable<br>frame, two rear<br>driving wheels with<br>hub<br>motor/electromagnet<br>ic brake assemblies,<br>two pivoting casters,<br>a Li-ion batteries, an<br>off-board battery<br>charger, a control<br>panel,<br>and an electric motor<br>controller | Consists of two foldable<br>armrests, a backrest, a<br>seat cushion, a foldable<br>frame, two rear driving<br>wheels with<br>hub<br>motor/electromagnetic<br>brake<br>assemblies, two pivoting<br>casters, a Li-ion<br>batteries, an off-board<br>battery charger, a<br>control panel,<br>and an electric motor<br>controller | Same | | Movement<br>co<br>ntrol<br>method | By Joystick control | By Joystick control | Same | | Driving system | Direct drive on the<br>rear<br>wheels | Direct drive on the rear<br>wheels | Same | | Brake system | Automatic<br>electromagnetic<br>brake system | Automatic<br>electromagnetic<br>brake system | Same | | Armrest | PU | PU | Same | | Back cushion | Terylen<br>(polyester fabrics) | Terylene | Same | | Seat cushion | Terylene<br>(polyester fabrics) | Terylene | Same | | Main frame<br>material | Magnesium alloy | Aluminium alloy | Different | | Max loading<br>weight | 120Kg | 264lbs(120Kg) | Same | | Min<br>obstacle<br>climbing | 35mm | 50mm | Minor difference will not<br>cause different performance. | | Total mass | 18KG | 41.9lbs/19KG | Similar | | Overall<br>Dimension<br>(length*width*<br>height) | 1100mmX600mmX9<br>40mm<br>(43.3" × 23.6″ ×<br>37.0") | 39.0" × 24.4" × 34.6" | Similar | | Folded<br>Dimension<br>(length*width*<br>height) | 690mmX380mmX72<br>0mm<br>(27.0" X 15.0" X<br>28.3") | 28.3" × 13.4" × 28.0" | Minor difference will not<br>cause different performance. | | Front wheel<br>size/type | 5.5" PU tire | 6" PU tire | Similar | | Rear wheel<br>size/type | 9.5" PU tires | 10" PU tires | Similar | | Max speed<br>forward | 5.0km/h(1.4m/s) | 6km/h | Similar | | Max speed<br>backward | 2.88km/h(0.8m/s) | 2.88km/h | Same | | Static stability | Forward:20.0°<br>Rear ward:17.0°<br>Sideways:19.2° | Forward:16.5°<br>Rear ward:16.5°<br>Sideways:17° | Minor difference on safe<br>operational incline degree<br>will not cause new safety and<br>effectiveness concerns, as<br>both the static and dynamic<br>stability under specific<br>inclining degree have been<br>evaluated according to | | Maximum safe operational in | 10° | 6° | Minor difference on safe<br>operational incline degree<br>will not cause new safety and<br>effectiveness concerns, as<br>both the static and dynamic<br>stability under specific<br>inclining degree have been<br>evaluated according to | | cline<br>degree | | | standard ISO 7176 series. | | Braking distance | 1.1m | 1 m | Minor difference on braking distance will not cause different performance.<br>Shorter distance for braking will be more safety. | | Maximum<br>distance of<br>travel on the<br>fully charged<br>battery | 12.6km | 15km | Minor difference will not cause different performance. | | Min.Turning<br>Radius | 900mm | 1000mm | Minor difference in turning radius is caused by different size of wheelchair ,it will not cause different performance. | | Battery | li-ion battery pack;rechargeable,<br>25.2VDC,<br>10.4Ah,2pcs | li-ion battery pack;rechargeable,<br>25.2VDC, 10.4Ah,2pcs | Same | | Battery<br>charger | High Power TechnologyInc.<br>HP0060W(L2)<br>Input: 100-240 VAC<br>Output: DC 24V, 2 Amp | High Power TechnologyInc.<br>HP0060W(L2)<br>Input: 100-240 VAC<br>Output: DC 24V, 2 Amp | Same | | Motor | DC Brushless Motor,<br>24Vdc 170W;2pcs | DC Brush Motor,<br>24Vdc 120W;2pcs | Analysis,The design principles of the controller and Driving system are the same, and both | | controller | M7084 | PG Drives Technology<br>Ltd., VR2 | meet the requirements of the ISO 7176-14:2008. Software validation is carried out on<br>both control systems. | {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ # Table 2 safety comparison | Item | Proposed Device | Predicate Device | Results | |------------------|------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|---------| | Biocompatibility | All user directly contacting<br>materials are compliance with<br>ISO10993-5、<br>ISO10993-10 and ISO10993-23<br>requirements. | All user directly contacting<br>materials are compliance with<br>ISO10993-5 and<br>ISO10993-10 requirements. | SE | | EMC | ISO7176-21& IEC 60601-1-2 | ISO7176-21 | SE | | Performance | ISO7176 series | ISO7176 series | SE | {10}------------------------------------------------ | Label and labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | SE | |--------------------|----------------------------|----------------------------|----| |--------------------|----------------------------|----------------------------|----| #### Substantial Equivalence Discussion G. The proposed device and predicate device are complying to the same ISO standards, ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 16840-10, ISO 7176-21, ISO7176-22,ISO 7176-25, IEC 60601-1-2 and FDA guidance Submission for Power Wheelchair. The indications for use, design, structure, composition, principle of operation and energy source of proposed device and predicate device are identical.The dimensions and performance parameters are similar.The main differences are motor ,controller and main frame appearance and materials. The design principles of the controller and Driving system are the same, and both meet the requirements of the ISO 7176-14. Software validation is carried out on both control systems. Brake system and speed control are designed in the same way as well, and both meet the requirements of the ISO 7176-3. Turning radius, Maximum obstacle climbing and Maximum safe operational incline are slightly different while such differences will not impact the safety and effectiveness of the subject device or raise new safety and effectiveness concerns as well as both meet the requirements of the ISO 7176-2 and ISO 7176-10. The main frame is not in contact with human tissue, and the material difference does not cause a security risk.The materials in contact with human tissue are identical for both proposed device and predicate device.The biocompatibility of the proposed device and predicate device meet the requirements of the ISO 10993-5:2009 & ISO 10993-10:2021 & ISO 10993-23: 2021. The flame-retardant test of the seat cushion/ backrest of both subject device and predicate device is carried out according to the ISO 16840-10 test. Therefore, both devices are assured to be under the same safety level. In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. #### H. Product Performance ### 2.1 Product Material Safety To ensure the safety of the wheelchair material, the bio-compatibility testing was done to the products. The testing was conducted according to the following standards: ISO10993-5 Biological evaluations of medical devices -- Part 5: Tests for In Vitro {11}------------------------------------------------ cytotoxicity ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for skin sensitization ISO 10993-23 Biological evaluation of medical devices - Part 23: Tests for irritation From the test result, we can find the material are safety and can meet the requirements. 2.2 Performance of the products The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination. Risk Analysis developed in accordance with ISO 14971: 2019. Software validation ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strength ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces. ISO 7176-14:2022 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling. ISO 16840-10:2021 Wheelchair seating - Part 10: Resistance to ignition of postural support devices - Requirements and test method ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters ISO 7176-22: 2014 Wheelchairs-Part 22: Set-up proceduresd ISO 7176-25: 2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs. Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2020 {12}------------------------------------------------ #### Conclusion - The subject device and the predicate products have the same technological characteristics, features, specifications, materials, mode of operation, and intended use of the device. To ensure all the key characteristic of the products can meet the requirements, the testing was done and all the results indicate the positive conclusion. Based on the analysis above, we confirm these two products are substantially equivalent.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...