K243903 · Ray Co., Ltd. · OAS · Mar 12, 2025 · Radiology
Device Facts
Record ID
K243903
Device Name
RCT600
Applicant
Ray Co., Ltd.
Product Code
OAS · Radiology
Decision Date
Mar 12, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
RCT600 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus and TMJ structure for adult and pediatric patients. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified heath care professionals.
Device Story
RCT600 is a CBCT and panoramic X-ray imaging system with optional cephalometric capability. It uses a rotating C-arm housing a high-voltage generator, X-ray tube, and detector to capture dental, maxillofacial, sinus, and TMJ images. The device is operated by dentists or qualified healthcare professionals in a clinical setting. It produces 3D CBCT, panoramic, and cephalometric images, which are viewed via RayScan software. These images support diagnostic decision-making for dental and orthodontic treatment planning. The system includes a 'deadman' exposure switch and supports standing or wheelchair-bound patients.
Clinical Evidence
Clinical evidence consisted of sample images for each dental modality (CBCT, panoramic, cephalometric) reviewed by an independent, licensed practitioner (DDS/ABOMR). The practitioner confirmed the images were of acceptable diagnostic quality for the intended use. No formal clinical trial with statistical endpoints was required; performance was validated through bench testing per IEC 61223-3-4 and IEC 61223-3-7 standards.
Indicated for radiographic examination of dento-maxillofacial, sinus, and TMJ structures in adult and pediatric patients (recommended >5 years old). Includes cephalometric imaging of the wrist for growth and maturity assessment in orthodontic treatment.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 12, 2025
RAY Co., Ltd. % Soo Ji Huh Official Correspondent 1F~3F, 4F(Part), 5F, 265, Daeji-Ro, Suji-gu Yongin-si, Gyeonggi-do 16882 SOUTH KOREA
Re: K243903
Trade/Device Name: RCT600 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: OAS Dated: December 19, 2024 Received: December 19, 2024
Dear Soo Ji Huh:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K243903
Device Name RCT600
#### Indications for Use (Describe)
RCT600 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus and TMJ structure for adult and pediatric patients. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified heath care professionals.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary - K243903
## 1. 510(k) Summary
The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.
## 2. Date
December 19, 2024
| Applicant | Ray Co., Ltd. | |
|----------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Address | 1F~3F, 4F(Part), 5F, 265, Daeji-ro, Suji-gu, Yongin-si,<br>16882, Korea | |
| Manufacturer | Name | Ray Co., Ltd. |
| | Address | 1F~3F, 4F(Part), 5F, 265, Daeji-ro, Suji-gu, Yongin-si,<br>16882, Korea |
| | Tel | +82-31-605-1000 |
| | Fax | +82-2-6280-5534 |
| Contact Person | Name | Sooji Huh |
| | Email | sooji.huh@raymedical.co.kr |
## 3. Administrative Information
## 4. Device Information
| Trade/Proprietary Name | RCT600 | |
|------------------------|----------------------------------|------------------------------------|
| Common Name | Computed Tomography X-Ray System | |
| Classification Name | Device | X-ray tomography, computed, dental |
| | Regulation Number | 21 CFR 892.1750 |
| | Class | 2 |
| | Product Code | OAS |
| | Review Panel | Radiology |
{5}------------------------------------------------
### 5. Predicate device
| Parameter | Predicate Device | Parameter | | Proposed Device | Predicate Device |
|---------------------------|----------------------------------|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | RCT700 | Manufacturer | | Ray Co., Ltd. | Ray Co., Ltd. |
| Manufacturer | Ray Co., Ltd | Device name | | RCT600 | RCT700 |
| 510(K) Number | K213226 | 510(K) Number | | (Traditional 510K) | K213226<br>(Traditional 510K) |
| Classification<br>name | Computed tomography x-ray system | Common Name | | Computed Tomography X-Ray<br>System | Dental panoramic/tomography and<br>cephalometric x-ray system |
| Regulation<br>number | 892.1750 | Indications for use | | RCT600 is CBCT and panoramic x-<br>ray imaging system with<br>cephalometric. Which is intended to<br>radiographic examination of the<br>dento-maxillofacial, sinus and TMJ<br>structure for diagnostic support for<br>adult and pediatric patients.<br>Cephalometric image is also<br>includes wrist to obtain carpus<br>images for growth and maturity<br>assessment for orthodontic<br>treatment.<br>The device is to be operated and<br>used by dentists or other legally<br>qualified heath care professionals | RCT700 is CBCT and panoramic x-<br>ray imaging system with<br>cephalometric. Which is intended to<br>radiographic examination of the<br>dento-maxillofacial, sinus, TMJ,<br>Airway and ENT structure for<br>diagnostic support for adult and<br>pediatric patients. And a model scan<br>is included as an option.<br>Cephalometric image is also<br>includes wrist to obtain carpus<br>images for growth and maturity<br>assessment for orthodontic<br>treatment. The device is to be<br>operated and used by dentists or<br>other legally qualified health care<br>professionals. |
| Primary product<br>code | OAS | Mode of Operation | | Same as predicate device #1 | Continuous operation with<br>intermittent, stated permissible<br>loading |
| 3D technology | | Same as predicate device #1 | CBCT Cone beam<br>Computed Tomography | | |
| Performance Specification | | 1) CBCT Computed tomography<br>2) Panoramic<br>3) Cephalometric(optional)<br>- Scan type | 1) CBCT Computed tomography<br>- Patient<br>2) Panoramic<br>3) Cephalometric(optional)<br>- One shot type<br>- Scan type | | |
| Functional Option | | Base<br>CT + PANO<br>Option(CEPH)<br>CT + PANO + SCAN CEPH | Base<br>CT+PANO<br>Option(CEPH)<br>CT + PANO + SCAN CEPH<br>CT + PANO + One shot(One shot,<br>Standard Type)<br>CT + PANO + One shot(One shot,<br>Large Type). | | |
| Detector | CT | N/A | FXDD-0606CA | | |
### 6. Device Description
RCT600 provides 3D computed tomography for scanning hard tissues such as bone and teeth. By rotating the C-arm, which houses a high-voltage generator, an X-ray tube and a detector on each end, CBCT images of dental maxillofacial structures are obtained by recombining data scanned from the same level at different angles. Functionalities include panoramic image scanning for obtaining images of whole teeth, and a cephalometric option for obtaining cephalometric images.
#### 7. Indication for use
RCT600 is CBCT and panoramic x-ray imaging system with cephalometric. Which is intended to radiographic examination of the dento-maxillofacial, sinus and TMJ structure for diagnostic support for adult and pediatric patients. Cephalometric image is also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment. The device is to be operated and used by dentists or other legally qualified heath care professionals
#### 8. Patient population
The patient population that may receive X-ray diagnostic radiation exposure includes individuals of any ethnic group, gender, weight, health status, or condition.
We recommend that patients who undergo X-ray diagnostic radiation exposure be over 5 years old.
{6}------------------------------------------------
## 9. Comparison with predicate device
The following table provides the summary of the technological characteristics of RCT600 compared to the predicate device.
{7}------------------------------------------------
| Type | | | |
|------------------------|-----------------------------|------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| | PANO | Same as predicate device #1 | Jupi0606X1 |
| | | N/A | FXDD-0606CA |
| | PANO | Same as predicate device #1 | Jupi0606X1 |
| | Ceph (Scan) | Pluto0900X | XID-C24DC |
| | Ceph (One shot) | N/A | FXRD-1717VA |
| | | N/A | FXDD-1012CA |
| Exposure switch Type | | Same as predicate device #1 | "Deadman" Button type |
| | | Same as predicate device #1 | Ceph Apparatus |
| | | Same as predicate device #1 | Vertical Carriage |
| | | Same as predicate device #1 | Rotator |
| | | Same as predicate device #1 | X-RAY Generator |
| | | Same as predicate device #1 | X-ray tube |
| | | Same as predicate device #1 | High Frequency Generator |
| Main Components | | Same as predicate device #1 | Column |
| | | Control panel | Touch monitor (panel) |
| | | Detector | Detector |
| | | -CT | -CT |
| | | Jupi0606X1 | FXDD-0606CA |
| | | -PANO | Jupi0606X1 |
| | | Jupi0606X1 | -PANO |
| | | -Ceph | FXDD-0606CA |
| | | Pluto0900X(Scan) | Jupi0606X1 |
| | | Same as predicate device #1 | -Ceph |
| | | Temple support | XID-C24DC(Scan)<br>FXRD-1717VA(One shot, Large<br>Size)<br>FXDD-1012CA(One shot, Standard<br>Size) |
| | | Same as predicate device #1 | Chinrest |
| | | Temple support | Head rest |
| | | Same as predicate device #1 | Automatic Collimator |
| | | Same as predicate device #1 | Exposure switch |
| | | Same as predicate device #1 | Emergency stop switch |
| Automatic Collimator | | N/A | Console PC set |
| | | Same as predicate device #1 | CT exams<br>Panoramic exams<br>Cephalometric exams |
| Class | Same as predicate device #1 | Class I with type B applied parts according to IEC 60601-1 | |
| Focal size | Same as predicate device #1 | 0.5 | |
| Field of View(CT) | Same as predicate device #1 | Max. 160x100 mm | |
| X-ray Voltage | Same as predicate device #1 | 60~100kVp | |
| X-ray Current | Same as predicate device #1 | 1~17mA | |
| Total Filtration | Same as predicate device #1 | Min. 2.8 mm Al equivalent | |
| Detector<br>Pixel size | CT | N/A | FXDD-0606CA: 119µm |
| | | Same as predicate device #1 | Jupi0606X1 : 100µm |
| | PANO | N/A | FXDD-0606CA: 119µm |
| | | Same as predicate device #1 | Jupi0606X1 : 100µm |
| | Ceph (Scan) | Pluto0900X: 100µm | XID-C24DC: 100µm |
| | Ceph(One<br>shot) | N/A | FXRD-1717VA : 140µm |
| | | N/A | FXDD-1012CA : 124µm |
| Magnification | CT | 1.55 | 1.44 |
| | PANO | 1.44 | 1.31 |
| | Ceph (Scan) | Same as predicate device #1 | 1.11 |
| | Ceph(One<br>shot) | N/A | Large Type: 1.13 |
| | | N/A | Standard Type: 1.12 |
| Scan time | | CT : below 20sec | CT : below 14sec |
| | | Same as predicate device #1 | Pano : below 14sec |
| | | Ceph[Scan size] : below 17.6sec | Ceph[Scan size]: below 20sec |
| | | N/A | Ceph [One shot size]: below 2 sec |
| Format compatible | | Same as predicate device #1 | DICOM 3.0 Format compatible |
| Image Viewing Software | | SMARTScam | RayScan |
| Image acquisition | | Same as predicate device #1 | Giga-Ethernet Network |
| Total Height | | Max 2,288mm | Max 2,296mm |
{8}------------------------------------------------
{9}------------------------------------------------
| Weight | 1) Computed Tomography (CT) +<br>Panoramic(PANO)=196kg(432.1lb) ±<br>10%<br>2) Computed Tomography (CT) +<br>Panoramic(PANO) + Ceph (Scan<br>type)= 217kg (478.4lb) ± 10% | 1) Computed Tomography (CT) +<br>Panoramic (PANΟ)=185kg(407.9lb) ±<br>10%<br>2) Computed Tomography (CT) +<br>Panoramic(PANO) + Ceph (Scan<br>type)= 212.5kg (468.5lb) ± 10%<br>3) Computed Tomography (CT) +<br>Panoramic(PANO) + Ceph (One<br>shot type, installed in Standard<br>size)= 211kg (465.2lb) ± 10%<br>4) Computed Tomography (CT) +<br>Panoramic(PANO) + Ceph (One<br>shot type, installed in Large size)<br>211kg (465.2lb) ± 10% |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of installation | Same as predicate device #1 | Wall or floor mount |
| Patient position | Same as predicate device #1 | Standing / Wheelchair |
| Applicable Standards | Same as predicate device #1 | IEC 60601-1<br>IEC 60601-1-3<br>IEC 60601-2-63<br>IEC 60601-1-2 |
The product is principally just the same as in the previous 510(k) #K213226
The complete of differences of the subject device to the predicate device #K213226 is as follows
- Delete of the One-shot Ceph Option
- Detector (using Scan Ceph)
- The magnification of CT and Panorama has been changed from 1.44 and 1.31 to 1.55 and 1.44
- The CT, Scan ceph scan time has been changed from 14sec and 20sec to 20sec and 17.6sec
- Change of the Image viewing software
However, X-ray voltage, X-ray current, and CT, Pano detector was identified in #K213226.
The 510(k) for the existing detector used in our equipment is provided below.
| Modality | Detector Model | Cleared | 510(k) No. |
|-----------|----------------|---------|------------|
| CT | Jupi0606X1 | No PMA | K213226 |
| Pano | Jupi0606X1 | No PMA | K213226 |
| Scan Ceph | Pluto0900X | No PMA | K211159 |
{10}------------------------------------------------
### 10. Safety and Effectiveness Information
The RCT600 system described in this 510(k) is similar to the predicate device in terms of indications for use, materials, safety characteristics and X-ray source.
The following information further substantiates the substantial equivalence between the subject device and predicate device. The fundamental technological characteristics of the subject and predicate devices are similar. The imaging modes are similar; PANO, CEPH (Optional), CBCT. The sponsor tested the subject device in a laboratory and provided a non-clinical performance report. The same test protocol was used to test the performance of the subject and predicate devices for comparison. The sponsor certifies that adequate design and development controls (according to 21 CFR 820.30) were in place for manufacturing the subject device.
The differences are as follows.
- Delete of the One-shot Ceph Option
- Detector (using Scan Ceph).
- The magnification of CT and Panorama has been changed from 1.44 and 1.31 to 1.55 and 1.44
- The CT, Scan ceph scan time has been changed from 14sec and 20sec to 20sec and 17.6sec
- Change of the Image viewing software
However, X-ray voltage, X-ray current, and CT, Pano detector was identified in #K213226.
Electrical, mechanical and environmental safety testing according to the standard of IEC 60601-1:2005/AMD1:2020 (3.2 Edition), IEC 60601-1-3:2008/AMD1:2013/AMD2:2021 (2.2 Edition), IEC 60601-1-6:2010/AMD1:2013/AMD2:2020 (3.2 Edition), and IEC 60601-2-63:2012/AMD1:2017/AMD2:2021 (1.2 Edition) were performed.
EMC testing was conducted in accordance with the standard IEC 60601-1-2:2014+AMD1:2020 (Edition 4.1).
The software of RCT600 saves patient and image data and offers an inquiry function. In addition, it supports the image generate function intended to obtain images using the RCT600 equipment and various sensors for diagnosis. That has been validated according to the FDA Guidance for the "Content of Premarket Submissions for Device Software Functions" and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" to assure substantial equivalence. The software for this device was considered a "Basic Documentation Level" of concern since a failure or latent flaw in the software would not directly result in serious injury or death to the patient or operator.
As a result, we identified the level of concern associated with a new device and provided documentation consistent with that level. Based on our risk analysis of software, the difference does not affect its safety and effectiveness.
Bench testing was conducted according to the FDA Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices". Bench testing is used to assess whether the x-ray exposure parameters are adequate to generate images of diagnostic quality.
{11}------------------------------------------------
Testing for Imaging Performance was conducted according to the recognized standards IEC 61223-3-4 and IEC 61223-3-7. Key imaqe quality performance metrics such as MTF, DQE, and NPS demonstrated that there are no significant differences between the two devices. All test results are satisfactory and they indicate safety and effectiveness for the dental system RCT600.
Non-clinical considerations were conducted in accordance with the FDA Guidance "Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices." Because the subject device uses the same detector as the predicate device, there are no significant differences between the two devices as a result of non-clinical testing.
Clinical testing was conducted according to the FDA Document "Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices". Sample clinical images were submitted for each imaging dental modality, and they provide evidence to show that the RCT600 works as intended while generating images of diagnostic quality.
The features of RCT600 (including CBCT, panoramic, and cephalometric acquisitions) were clinically tested and approved by an independent, licensed practitioner/clinician DDS/ABOMR). Clinical images were gathered from all detectors of the RCT600 system using recognized protocols for image acquisition. A licensed practitioner reviewed the sample clinical images and deemed them to be of acceptable quality for the device intended use.
#### 11. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Ray Co., Ltd. concludes that the newly RCT600 is safe, effective and substantially equivalent to the predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.