K243842 · Centerline Biomedical, Inc. · DQK · Mar 6, 2025 · Cardiovascular
Device Facts
Record ID
K243842
Device Name
Intra-Operative Positioning System (IOPS®)
Applicant
Centerline Biomedical, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Mar 6, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Indications for Use
The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.
Device Story
IOPS is a surgical navigation system using electromagnetic tracking to display real-time position/orientation of sensor-equipped catheters and guidewires. System maps patient vascular anatomy from previously acquired CT scan data; superimposes navigation data onto vascular map on a monitor. Used in clinical settings by physicians during endovascular interventions as an adjunct to fluoroscopy. Provides real-time visual guidance to assist in catheter/guidewire placement; does not perform diagnosis. Benefits include improved navigation accuracy and potential reduction in radiation exposure by supplementing fluoroscopic imaging.
Clinical Evidence
No clinical or animal testing performed. Evidence consists of bench testing, specifically software verification and design validation (usability testing). Validation included physician targeting of aortic arch branches, distal descending aorta branches, and peripheral vasculature, plus technician tracking pad placement scenarios. Results support the expanded indications for use.
Technological Characteristics
Electromagnetic tracking technology; mobile cart (MC-1) form factor. Accessories include sterile (EtO) single-use guidewires, catheters, fiducial tracking pads, and guidewire handles. Software version 1.5.81. System is non-contact, multi-patient use. No hardware changes from predicate.
Indications for Use
Indicated for adult patients undergoing endovascular interventions in the peripheral, aortic, and aortic side branch vasculature. Used as an adjunct to fluoroscopy for real-time tip positioning and navigation of sensor-equipped catheters and guidewires. Not for diagnostic use.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
Intra-Operative Positioning System (IOPS) (K190106)
Intra-Operative Positioning System (IOPS®) (K230309)
Intra-Operative Positioning System (IOPS®) Viewpoint (K241243)
Intra-Operative Positioning System (IOPS®) (K242133)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION" below it.
March 6, 2025
Centerline Biomedical, Inc. Kevin Reed Director, Quality & Regulatory 4535 Renaissance Pkwy Cleveland, Ohio 44128
Re: K243842
Trade/Device Name: Intra-Operative Positioning System (IOPS®) Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK, DQY, DQX Dated: December 13, 2024 Received: December 13, 2024
### Dear Kevin Reed:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{2}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# MARCO CANNELLA -S
for
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
Submission Number (if known)
K243842
Device Name
Intra-Operative Positioning System (IOPS®)
Indications for Use (Describe)
The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### SUBMITTER
Centerline Biomedical, Inc. 4535 Renaissance Pkwy Cleveland, OH 44128
Phone: 330-577-5033 Email: kevin.reed@centerlinebiomedical.com
Contact Person: Kevin Reed Date Prepared: 21 FEB 2023
### DEVICE
| Trade Name: | Intro-Operative Positioning System (IOPS®) |
|----------------------|--------------------------------------------|
| Common Name: | Programmable diagnostic computer |
| Classification Name: | Computer, diagnostic, programmable |
| Regulatory Class: | II |
| Product Codes: | DQK, DQY, DQX |
| Regulation Number: | 870.1425 |
# PREDICATE DEVICE
The legally marketed device to which equivalence is being claimed [807.92(a)(3)] is as follows:
| 510(k) Number | Name | Product Code |
|---------------|------------------------------------------------------|--------------|
| K190106 | Intra-Operative Positioning System (IOPS) | DQK |
| K230309 | Intra-Operative Positioning System (IOPS®) | DQK |
| K241243 | Intra-Operative Positioning System (IOPS®) Viewpoint | DQK |
| K242133 | Intra-Operative Positioning System (IOPS®) | DQK |
Primary predicate is listed first. The predicates have not been the subject of any design-related recalls.
### DEVICE DESCRIPTION
The Intra-Operative Positioning System (IOPS) consists of a surgical navigation technology and a number of associated accessories. The navigation technology is a non-contact reusable multi-patient use device. The associated accessories are single use devices provided sterile (EtO).
The IOPS displays the position and orientation of sensor equipped catheters, guidewires, and tracking pad utilizing electromagnetic tracking technology. The system enables mapping of the patient's vascular system utilizing previously acquired CT scan data. IOPS registers the location and orientation of the sensors in real time, superimposing navigation of the catheters and guidewires to the patient's vascular map displayed on a monitor. The system is for use as an adjunct to fluoroscopy and does not make a diagnosis.
The associated accessories include:
- Guidewire
- Catheters
- Fiducial Tracking Pad
- · Guidewire Handle
{6}------------------------------------------------
## INDICATIONS FOR USE
The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Indications for Use statement for the IOPS and its associated accessories is not identical to the predicate device; however, the differences do not alter the intended use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have the same intended use for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data and for real time tip positioning and navigation using sensor equipped compatible catheters and guidewires.
All technological characteristics of the IOPS and its associated accessories are the same as the predicate devices except for IOPS software version.
Table 1 lists the differences between the subject device and the predicate device.
| Table 1: Summary Comparison | | | |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Predicate Device<br>(K190106, K230309,<br>K2414243, K242133)* | Discussion |
| Trade Name | Intra-Operative Positioning<br>System (IOPS) | Intra-Operative Positioning<br>System (IOPS) | Same |
| Model<br>Name/Number | Mobile Cart<br>MC-1 | Mobile Cart<br>MC-1 | Same |
| Product Code | DQK | DQK | Same |
| Class | II | II | Same |
| Classification<br>Identification | 870.1425 | 870.1425 | Same |
| Device<br>Description | The IOPS displays the<br>position and orientation of<br>sensor equipped guidewires<br>and catheters utilizing<br>electromagnetic tracking<br>technology. The system<br>enables mapping of the<br>patient's vascular system<br>utilizing previously acquired<br>CT scan data. IOPS registers<br>the location and orientation<br>of the sensors in real time<br>superimposing navigation of<br>the catheters and guidewires<br>to the patient's vascular<br>map. | The IOPS displays the<br>position and orientation of<br>sensor equipped guidewires<br>and catheters utilizing<br>electromagnetic tracking<br>technology. The system<br>enables mapping of the<br>patient's vascular system<br>utilizing previously acquired<br>CT scan data. IOPS registers<br>the location and orientation<br>of the sensors in real time<br>superimposing navigation of<br>the catheters and guidewires<br>to the patient's vascular<br>map. | Same |
| Device<br>Accessories | • IOPS Guidewire (ATW-2)<br>• IOPS Viewpoint Simple<br>Curve Catheters | • IOPS Guidewire (ATW-2)<br>• IOPS Viewpoint Simple<br>Curve Catheters | Similar |
| | (C00751; C01251)<br>IOPS Viewpoint Double<br>Curve Catheters<br>(C00752; C01252) IOPS Fiducial Tracking<br>Pad (T02111) IOPS Guidewire Handle<br>(H01035) | (C00751; C01251)<br>IOPS Viewpoint Double<br>Curve Catheters<br>(C00752; C01252) IOPS Fiducial Tracking<br>Pad (T02111;TP-1) IOPS Guidewire Handle<br>(H01035) | subject device in this<br>submission is not intended to<br>be used with the first<br>generation IOPS Tracking Pad<br>(TP-1). |
| Intended<br>Use/Indications<br>for Use | The IOPS (Intra-Operative<br>Positioning System) is<br>intended for the evaluation<br>of vascular anatomy as<br>captured via 3D modeling<br>from previously acquired<br>scan data. It is intended for<br>real time tip positioning and<br>navigation using sensor-<br>equipped compatible<br>catheters and guidewires<br>used in endovascular<br>interventions in the<br>peripheral, aortic and aortic<br>side branch vasculature. The<br>system is indicated for use as<br>an adjunct to fluoroscopy.<br>The IOPS does not make a<br>diagnosis. | The IOPS (Intra-Operative<br>Positioning System) is<br>intended for the evaluation<br>of vascular anatomy as<br>captured via 3D modeling<br>from previously acquired<br>scan data. It is intended for<br>real time tip positioning and<br>navigation using sensor<br>equipped compatible<br>catheters and guidewires<br>used in endovascular<br>interventions in the<br>descending aorta. The<br>system is indicated for use as<br>an adjunct to fluoroscopy.<br>The IOPS does not make a<br>diagnosis. | The Indications for Use<br>statement for the IOPS and<br>its associated accessories is<br>not identical to the predicate<br>device; however, the<br>differences do not alter the<br>intended use of the device<br>nor do they affect the safety<br>and effectiveness of the<br>device relative to the<br>predicate. Both the subject<br>and predicate devices have<br>the same intended use for<br>the evaluation of vascular<br>anatomy as captured via 3D<br>modeling from previously<br>acquired scan data and for<br>real time tip positioning and<br>navigation using sensor<br>equipped compatible<br>catheters and guidewires.<br>The subject device has been<br>evaluated through design<br>validation. |
| Intended Use<br>Population | Adults Only | Adults Only | Same |
| Software Version | 1.5.81 | 1.0.5056 | The difference in the<br>software version does not<br>raise new questions of safety<br>or effectiveness. The subject<br>device has been evaluated<br>through software verification<br>testing |
Table 1: Summary Comparison
{7}------------------------------------------------
* The predicate device was originally cleared in K190106 and cleared with updated labeling (changed Applied Part designations from Type BF to Defibrillation-Proof Type BF) in K230309. IOPS Viewpoint Catheters were cleared in K241243. IOPS Fiducial Tracking Pad and Guidewire Handle were cleared in K242133.
# PERFORMANCE DATA
### Bench Testing
Software verification was conducted to establish equivalency to the predicate device in safety and effectiveness. There are no hardware changes to IOPS and its associated accessories and therefore Electrical, Mechanical and Thermal Safety Testing is not needed.
Design Validation, including usability testing, was performed to support the inclusion of all peripheral,
{8}------------------------------------------------
aortic and aortic side branch vasculature vessels in the indications for use statement and to establish equivalency to the predicate device in safety and effectiveness. These studies included physician targeting of three regions for cannulation: branches of the aortic arch, distal branches of the descending aorta, and branches of the peripheral vasculature. In addition, technicians were given test case scenarios and asked to place the tracking pad in each of the three proposed expanded indications regions.
#### Animal Testing
No animal testing was completed as a part of this submission.
#### Clinical Testing
No clinical testing was completed as a part of this submission.
#### CONCLUSIONS
The change in the indications for use does not raise new or different questions of safety or effectiveness. The successful completion of non-clinical testing demonstrates that IOPS performs as intended and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.