The StimTrial System is indicated for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for the StimRouter Neuromodulation System’s permanent (long term) implant indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as an adjunct to other modes of therapy used in a multidisciplinary approach and not intended to treat pain in the craniofacial region.
Device Story
StimTrial Neuromodulation System provides temporary trial stimulation to assess patient candidacy for permanent peripheral nerve stimulation. System components: percutaneous lead, lead adaptor cable, external stimulator, and disposable hydrogel patch electrode. Clinician uses programmer software to configure stimulation parameters; patient uses mobile app to adjust therapy levels. External stimulator generates electrical pulses; delivered via lead to target peripheral nerve; interrupts pain signals to induce paresthesia. Used in clinical settings for implantation; patient manages therapy at home. Benefits include pain reduction and efficacy verification prior to permanent implant surgery.
Clinical Evidence
No human clinical trial data provided. Evidence consists of bench testing (electrical safety, EMC, wireless co-existence), software verification/validation, usability testing, and animal studies (acute and long-term porcine models). Biocompatibility, sterility, and shelf-life testing were also conducted.
Technological Characteristics
System uses percutaneous lead (Platinum-Iridium 90:10, silicone insulation) and external stimulator (Tritan copolyester housing). Stimulation: charge-balanced, biphasic, rectangular pulses (1-200 Hz, 100-500 μsec). Connectivity: Bluetooth Low Energy (BLE). Power: rechargeable Li-ion polymer battery. Sterilization: Ethylene Oxide. Software: Clinician Programmer (Windows tablet) and Mobile App (iOS/Android).
Indications for Use
Indicated for trial stimulation (max 30 days) in adults with severe intractable chronic pain of peripheral nerve origin to determine efficacy for permanent StimRouter implantation. Adjunct to other therapies; excludes craniofacial pain.
Regulatory Classification
Identification
An implanted peripheral nerve stimulator for pain relief is a device that is used to stimulate electrically a peripheral nerve in a patient to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed around a peripheral nerve and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.
Predicate Devices
Nalu Neurostimulation System for Peripheral Nerve Stimulation (K183579)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 16, 2025
Bioventus LLC
Sageev George
Director, Regulatory Affairs
4721 Emperor Boulevard
Suite 100
Durham, North Carolina 27703
Re: K243782
Trade/Device Name: StimTrial Neuromodulation System
Regulation Number: 21 CFR 882.5870
Regulation Name: Implanted Peripheral Nerve Stimulator For Pain Relief
Regulatory Class: Class II
Product Code: GZF
Dated: June 16, 2025
Received: June 16, 2025
Dear Sageev George:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243782 - Sageev George
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243782 - Sageev George
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Amber T. Ballard -S
Amber Ballard, PhD
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243782 | |
| Device Name StimTrial Neuromodulation System | |
| Indications for Use (Describe) The StimTrial System is indicated for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for the StimRouter Neuromodulation System’s permanent (long term) implant indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as an adjunct to other modes of therapy used in a multidisciplinary approach and not intended to treat pain in the craniofacial region. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
bioventus
Bioventus LLC
4721 Emperor Blvd., Suite 100
Durham, NC 27703
USA
1-919-474-6700
1-800-396-4325
www.Bioventus.com
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
## I. Submitter Information
**Company:**
Bioventus LLC
4721 Emperor Boulevard
Suite 100
Durham, NC 27703 USA
Phone Number: (661) 378-8036
Email: Sageev.George@Bioventus.com
**Contact:**
Sageev George
Director, Regulatory Affairs Manager
Phone Number: (661) 378-8036
Email: Sageev.George@Bioventus.com
Kellie Stefaniak
Vice-President, Global Regulatory and Quality
Phone Number: (919) 566-1224
Email: Kellie.Stefaniak@Bioventus.com
**Date Prepared:** July 9, 2025
## II. Name of Device
**Device Trade Name:** StimTrial Neuromodulation System
**Classification Name:** Implantable peripheral nerve stimulator for pain relief
**Common Name:** Implantable Neurostimulator
**Product Code:** GZF
**Regulation Number:** 21 CFR §882.5870
**Device Class:** Class II
**Panel Identification:** Neurology
## III. Predicate Devices
**Predicate #1 Manufacturer:** Nalu Medical Inc.
**Predicate #1 Trade Name:** Nalu Neurostimulation System for Peripheral Nerve Stimulation
**Predicate #1 510(k):** K183579
**Predicate #2 Manufacturer:** Bioness Inc.
**Predicate #2 Trade Name:** StimRouter Neuromodulation System
**Predicate #2 510(k):** K211965
{5}
bioventus
Bioventus LLC
4721 Emperor Blvd., Suite 100
Durham, NC 27703
USA
1-919-474-6700
1-800-396-4325
www.Bioventus.com
Predicate #3 Manufacturer: SPR Therapeutics
Predicate #3 Trade Name: Smartpatch PNS System
Predicate #3 510(k): K161154
## IV. Device Description
The StimTrial Neuromodulation System is to be used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) implant for a system indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as an adjunct to other modes of therapy used in a multidisciplinary approach and not intended to treat pain of craniofacial nerve origin.
The StimTrial Neuromodulation System is intended to help determine patient candidacy for a permanent implant to help manage pain of peripheral nerve origin. The StimTrial System works by sending electrical impulses from an external stimulator to a lead that is percutaneously placed next to a target nerve. These impulses are intended to interrupt or change the pain signals, inducing the feeling of tingling or numbness (paresthesia), and possibly reducing or replacing the feeling of pain.
The StimTrial Neuromodulation System consists of three main parts - the percutaneous StimTrial Lead, the StimTrial Lead Adaptor Cable, and the StimTrial External Stimulator. The StimTrial Lead is percutaneously placed with the distal, stimulating end located at or near the targeted peripheral nerve and with the proximal end remaining outside of the body for the trial duration. The StimTrial Lead Adaptor Cable is an external cable that directly connects the StimTrial Lead to the StimTrial External Stimulator providing a direct electrical pathway from the External Stimulator to the Lead. The StimTrial External Stimulator is a device which, when connected to the Lead Adaptor Cable and the StimRouter Electrode (a hydrogel patch electrode) attached to the skin near the implant site, generates electrical stimulation pulses that travel to the StimTrial Lead. Accessories for the StimTrial System include the StimRouter Electrode (a disposable electrode patch, cleared most recently in K211965), the StimTrial Clinician Programmer with Software (CPS) and the optional StimTrial Mobile Application (MAPP) installed on a Smartphone.
The StimTrial System incorporates both commercially available and specially designed components. The materials used in the StimTrial Lead have a long history of use in implanted devices; the insertion tools are also constructed from materials with a long history of surgical use. Materials in the external accessories are commonly used in both medical and non-medical applications. The powered components of the StimTrial System use commercially available IEC and UL approval rechargeable batteries. There are no components which are plugged into a wall socket during the use of the system by the patient.
## V. Indications for Use
The StimTrial System is indicated for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for the StimRouter Neuromodulation
{6}
3
bioventus
Bioventus LLC
4721 Emperor Blvd., Suite 100
Durham, NC 27703
USA
1-919-474-6700
1-800-396-4325
www.Bioventus.com
System’s permanent (long term) implant indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as an adjunct to other modes of therapy used in a multidisciplinary approach and not intended to treat pain in the craniofacial region.
## VI. Comparison of Technological Characteristics
The table below compares the principles of operation and operational/technological characteristics for the StimTrial System and its predicate devices. These comparisons confirm that all of these PNS systems have very similar overall principles of operation and operational/technological characteristics, thereby supporting the substantial equivalence of the StimTrial System to these previously cleared predicates. Comparison is listed per following aspects:
- System Characteristics
- Leads
- Externally Worn Devices
- Clinician Programmer
- Patient Remote Control
{7}
bioventus
Bioventus LLC
4721 Emperor Blvd., Suite 100
Durham, NC 27703
USA
1-919-474-6700
1-800-396-4325
www.Bioventus.com
Comparison of Subject Device and Predicate Device Systems
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | Tertiary Predicate Device | Equivalency discussion |
| --- | --- | --- | --- | --- | --- |
| 510(k) Number | K243782 | K183579 | K211965 | K161154 | |
| Device Name | StimTrial Neuromodulation System | Nalu Neuromodulation System for Peripheral Nerve Stimulation | StimRouter Neuromodulation System | Smartpatch PNS System | |
| Regulation and Product Codes | 21 CFR 882.5870
GZF | 21 CFR 882.5870
GZF | 21 CFR 882.5870
GZF | 21 CFR 882.5890
NHI | |
| Indications for Use | The StimTrial System is indicated for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for the StimRouter Neuromodulation System’s permanent (long term) implant indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as an adjunct to other modes of therapy used in a multidisciplinary approach and not intended to treat pain in the craniofacial region. | This system is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as the sole mitigating agent or as an adjunct to other modes of therapy used in a multidisciplinary approach. The system is not intended to treat pain in the craniofacial region. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device. | The StimRouter Neuromodulation System is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as an adjunct to other modes of therapy (e.g., medications). The StimRouter is not intended to treat pain in the craniofacial region. | The Smartpatch Peripheral Nerve Stimulation (PNS) System is indicated for up to 30 days in the back and/or extremities for:
- Symptomatic relief of chronic, intractable pain, post-surgical and post-traumatic acute pain;
- Symptomatic relief of post-traumatic pain;
- Symptomatic relief of post-operative pain.
The Smartpatch PNS System is not intended to treat pain in the craniofacial region. | |
| Implant Site | Subcutaneous, stimulating electrode(s) of the lead are at peripheral nerves, excluding craniofacial region. | Peripheral nerves, excluding craniofacial region. | Not publicly available | Percutaneous electrode placed in proximity to a target peripheral nerve. | Similar. The subject device and predicates target the peripheral nerves. |
| Mode of action | Wearable external stimulator delivering electrical stimulation to a percutaneous electrode placed in proximity to a target peripheral nerve. | Not publicly available | Not publicly available | External stimulator delivers stimulation therapy to percutaneous electrode placed in proximity to a target peripheral nerve. | Similar. All devices deliver energy to implanted receiver. |
{8}
5
bioventus
Bioventus LLC
4721 Emperor Blvd., Suite 100
Durham, NC 27703
USA
1-919-474-6700
1-800-396-4325
www.Bioventus.com
Stimulation Parameters
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | Tertiary Predicate Device | Equivalency discussion |
| --- | --- | --- | --- | --- | --- |
| Stimulator battery type | One Lithium ion Polymer, rechargeable | One Lithium ion Polymer, rechargeable | One Lithium ion Polymer, rechargeable | Replaceable batteries | The differences do not affect safety and effectiveness of intended use |
| Current / Voltage regulated | Current | Current | Current | Not publicly available | The differences do not affect safety and effectiveness of intended use |
| Amplitude of Stimulator current (max) | 10mA | 10.2 mA
(300Ω load) | 30 mA | 20 mA | Subject device within range of predicates |
| RMS value of Stimulator current (max) | 3.16 mA | 7.2 mA
(300Ω load)
Assuming continuous stimulation, positive phase rectangle, negative is decaying exponential | Not publicly available | Not publicly available | The differences do not affect safety and effectiveness of intended use |
| Pulse frequency | 1 – 200 Hz | 2 – 1,500 Hz | 1 – 200 Hz | 5 – 100 Hz | Subject device within range of predicates |
| Pulse width (positive phase) | 100 – 500 μsec | 12 – 1,000 μsec | 100 – 500 μsec | Up to 200 μsec | Subject device within range of predicates |
| Maximum charge per pulse | 5.0 μC | 10.2 μC
(300Ω load) | Not publicly available | Not publicly available | The differences do not affect safety and effectiveness of intended use |
| Maximum charge density (at stimulating electrode) | 26.5 μC /cm² | 83.3 μC /cm² (300Ω load) | Not publicly available | 40 μC /cm² | Subject device within range of predicates |
{9}
bioventus
Bioventus LLC
4721 Emperor Blvd., Suite 100
Durham, NC 27703
USA
1-919-474-6700
1-800-396-4325
www.Bioventus.com
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | Tertiary Predicate Device | Equivalency discussion |
| --- | --- | --- | --- | --- | --- |
| Maximum current density at (stimulating electrode) | 52.9 mA/cm² | 83.3 mA/cm² (300Ω load) | Not publicly available | Not publicly available | Subject device within range of predicates. |
| Net charge | 0 μC | 0 μC | Not publicly available | Not publicly available | Subject device within range of predicates. |
| Max average phase power | 0.030 W / phase (300Ω load) | 0.031 W / phase (300Ω load) | Not publicly available | Not publicly available | Subject device within range of predicates |
| Max average phase power density | 0.16 W/cm² / phase (300Ω load) | 0.25 W/cm² / phase (300Ω load) | Not publicly available. | Not publicly available. | Subject device within range of predicates |
| Waveform | Charge balanced, biphasic, rectangular, symmetrical or asymmetrical | Charge balanced (delayed) biphasic asymmetrical | Charge balanced, biphasic, rectangular, symmetrical or asymmetrical | Biphasic rectangular with a positive and negative phase that is charge balanced | The differences do not affect safety and effectiveness of intended use |
| Pulse shape | Rectangular | Decaying exponential | Not publicly available | Rectangular | The differences do not affect safety and effectiveness of intended use. |
| Stimulation modality | Monopolar | Bipolar | Not publicly available | 1 pair (1 percutaneous electrode and 1 surface electrode) | The differences do not affect safety and effectiveness of intended use. |
| Main materials in long term contact with intact skin (see biocomp report for complete list) | Santoprene® 8281-75MED PELLETHANE 2363-80A TRITAN MX731 | PC ABS housing. Textile material of belt / cuff may be worn over clothing. Hydrocolloid adhesive applied to skin. | Santoprene® 8281-75MED PELLETHANE 2363-80A TRITAN MX731 | Not publicly available | The differences do not affect safety and effectiveness of intended use. |
| Wireless Communications | Bluetooth Low Energy (BLE) | Bluetooth Low Energy (BLE) | Bluetooth Low Energy (BLE) | Not publicly available | The differences do not affect safety and effectiveness of intended use. |
6
{10}
7
bioventus
Bioventus LLC
4721 Emperor Blvd., Suite 100
Durham, NC 27703
USA
1-919-474-6700
1-800-396-4325
www.Bioventus.com
Comparison of Subject Device and Predicate Device Leads
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | Tertiary Predicate Device | Equivalency discussion |
| --- | --- | --- | --- | --- | --- |
| Number of stimulating electrodes | 1 (divided into 3 segments) | 8 | 1 (divided into 3 segments) | 1 pair (1 percutaneous electrode and 1 surface electrode) | The differences do not affect safety and effectiveness of intended use. |
| Electrode shape | Cylindrical | Cylindrical | Cylindrical | Straight with tine | The differences do not affect safety and effectiveness of intended use. |
| Individual Electrode surface area | 6.3 mm² | 12.25 mm² | 6.3 mm² | Not publicly available. | StimTrial individual electrode surface is the same as StimRouter predicate. Differences with Nalu predicate do not affect safety and effectiveness of the intended use. |
| Electrode array length | 5 mm | 52 mm | Not publicly available | Not publicly available. | Differences with predicates do not affect safety and effectiveness of the intended use. |
| Individual electrode length | 1 mm | 3 mm | Not publicly available. | 15.2mm | Differences with predicates do not affect safety and effectiveness of the intended use. |
| Electrode spacing | Spacing between electrode segments 1mm | 4 mm | Not publicly available. | Not publicly available. | Differences with predicates do not affect safety and effectiveness of the intended use. |
| Electrode material | Platinum – Iridium (90:10) | Platinum – Iridium (90:10) | Platinum – Iridium (90:10) | 316L Stainless Steel | The differences do not affect safety and effectiveness of intended use. |
| Lead body insulation material | Silicone | Pellethane 2363-55D | Silicone | Not publicly available. | The differences do not affect safety and effectiveness of intended use. |
| Wire features | Multi-strand, non-insulated, helical | Multilumen tube | Not publicly available. | Fine wire | The differences do not affect safety and effectiveness of intended use. |
{11}
8
bioventus
Bioventus LLC
4721 Emperor Blvd., Suite 100
Durham, NC 27703
USA
1-919-474-6700
1-800-396-4325
www.Bioventus.com
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | Tertiary Predicate Device | Equivalency discussion |
| --- | --- | --- | --- | --- | --- |
| Anchor | No anchor.
Lead is secured at skin level with a wound bandage. | Optional molded silicone anchor with Ti locking mechanism | Not publicly available. | Tine | The differences do not affect safety and effectiveness of intended use. |
| Lead + Lead Adaptor length | ~45 cm | 40cm, 60cm | Not publicly available. | Lead is either 20cm or 40cm with 10cm introducer needle | The differences do not affect safety and effectiveness of intended use. |
| Lead body diameter | 1.2 mm | 1.3 mm | 1.2 mm | Not publicly available. | The differences do not affect safety and effectiveness of intended use. |
| Biocompatibility | Yes | Yes | Yes | Yes | Substantially equivalent |
| Sterilization method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Provided sterile. Sterilization method not publicly available | The differences do not affect safety and effectiveness of intended use. |
{12}
bioventus
Bioventus LLC
4721 Emperor Blvd., Suite 100
Durham, NC 27703
USA
1-919-474-6700
1-800-396-4325
www.Bioventus.com
Comparison of Subject Device and Predicate Device Externally Worn Devices
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | Tertiary Predicate Device | Equivalency discussion |
| --- | --- | --- | --- | --- | --- |
| Name | External Stimulator | Trial Therapy Disc | StimRouter External Electrical Field Conductor (E-EFC) | Not publicly available. | N/A |
| Electronics | A printed circuit board Assembly (PCBA) that generates therapeutic stimulation currents and buttons for power and adjusting parameter settings as needed by the user. | A printed circuit board Assembly (PCBA) that generates therapeutic stimulation currents and buttons for power and adjusting parameter settings as needed by the user. | A Printed Circuit Board Assembly (PCBA) that generates therapeutic stimulation currents and buttons for power and adjusting parameter settings as needed by the user. | Not publicly available. | The differences do not affect safety and effectiveness of intended use. |
| User interface | Integrated controls and visual and audio indicators that allow the user to turn the device on/off, increase or decrease therapy levels, select from configured therapy profiles and monitor device status. | Integrated controls and indicators that allow the user to turn the device on/off, increase or decrease therapy levels, select from configured therapy profiles and monitor device status | Integrated controls and visual and audio indicators that allow the user to turn the device on/off, increase or decrease therapy levels, select from configured therapy profiles and monitor device status. | Button controls for turning device on/off and for decreasing/increasing therapy intensity. Text prompts provided on LCD screen | The differences do not affect safety and effectiveness of intended use. |
| Delivery of energy | External Stimulator delivers electrical stimulation directly (via connectors and cables) to the percutaneously implanted trial lead. | Trial Therapy Disc delivers electrical stimulation directly (via connectors and cables) to the percutaneously implanted trial lead. | Not publicly available. | Wearable external stimulator delivers stimulation therapy to the percutaneous electrode | The differences do not affect safety and effectiveness of intended use. |
| Area of hydrogel skin electrode for return. | 28 cm² | N/A (not used for Nalu System) | Not publicly available. | Not publicly available. | The differences do not affect safety and effectiveness of intended use. |
| Max current density (at hydrogel electrode) | 0.226 mA/cm² (RMS) | N/A | Not publicly available. | Not publicly available. | The differences do not affect safety and effectiveness of intended use. |
{13}
bioventus
Bioventus LLC
4721 Emperor Blvd., Suite 100
Durham, NC 27703
USA
1-919-474-6700
1-800-396-4325
www.Bioventus.com
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | Tertiary Predicate Device | Equivalency discussion |
| --- | --- | --- | --- | --- | --- |
| Wearing | External Stimulator is plugged into hydrogel electrode patch. The patch is attached to the skin. The subcutaneous lead is connected through the Lead adaptor to the USB output of Stimulator. Hydrogel patch serves as a mechanical attachment to the skin and as an electrical conductor for monopolar return to close the circuit between stimulating electrode of the implanted trial lead. | Trial Therapy Disc is positioned over IPG via two options:
• Adhesive clip (hydrocolloid adhesive)
• Elastic Belt / Cuff | StimRouter E-EFC is plugged into a hydrogel electrode patch. The patch is positioned on the skin over the implanted receiver. Hydrogel patch serves as a mechanical attachment to the skin and as an electrical conductor to deliver transcutaneous stimulation. | Hand-held | The differences do not affect safety and effectiveness of intended use. |
| Size / Weight | External Stimulator: 5.7cm x 3.7cm x 1.2cm 28g
Hydrogel patch: 10.6 x 3.4 x 0.26cm 6g | Trial Therapy Disc: ~1.5cm thick, 7.5cm diameter
Weight: ~80g | StimRouter E-EFC: 5.7 x 3.7 x 1.2cm 28g | 24g | The differences do not affect safety and effectiveness of intended use. |
| Externally contacting materials | External Stimulator: Housing - Tritan™ Copolyester MX731
Button overlay: Polycarbonate polyform
Snap connector: CDA 260 Brass 200 Series S.S
Patch: Hydrogel, Non- woven fabric (MI) | PC ABS housing.
Textile material of belt / cuff may be worn over clothing.
Hydrocolloid adhesive applied to skin. | StimRouter E-EFC: Housing - Tritan™ Copolyester MX731
Button overlay: Polycarbonate polyform
Snap connector: CDA 260 Brass 200 Series S.S | Not publicly available. | The differences do not affect safety and effectiveness of intended use. |
| Battery charging | Medical grade USB charger with a USB-A to USB-C cable plugging to the External Stimulator. | Electrically isolated cradle charger. | E-EFC uses USB-C charging port | Not publicly available. | The differences do not affect safety and effectiveness of intended use. |
10
{14}
bioventus
Bioventus LLC
4721 Emperor Blvd., Suite 100
Durham, NC 27703
USA
1-919-474-6700
1-800-396-4325
www.Bioventus.com
# Substantial Equivalence Table – Clinician Programmer
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | Tertiary Predicate Device | Equivalency discussion |
| --- | --- | --- | --- | --- | --- |
| Configuration | Software installed on a Windows tablet. | Software installed on an Android tablet. | Software installed on a Windows tablet. | Not publicly available. | The differences do not affect safety and effectiveness of intended use. |
| Purpose | Allows healthcare provider to set desired therapy levels and device settings across External Stimulator and optional Mobile App for Remote Control. | Allows healthcare provider to set desired therapy levels and device settings across Trial Therapy Disc and Patient Remote Control. | Allows healthcare provider to set desired therapy levels and device settings across StimRouter E-EFC and optional Mobile App for Remote Control. | Not publicly available. | The differences do not affect safety and effectiveness of intended use. |
| Communication | Bluetooth Low Energy (BLE) to External Stimulator. | Bluetooth Low Energy (BLE) to Trial Therapy Disc and Patient Remote Control. | Bluetooth Low Energy (BLE) to E-EFC. | Not publicly available. | The differences do not affect safety and effectiveness of intended use. |
# Substantial Equivalence Table – Patient Remote Control
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | Tertiary Predicate Device | Equivalency discussion |
| --- | --- | --- | --- | --- | --- |
| Name | Patient Mobile App (MAPP) | Patient Remote Control | Patient Mobile App (MAPP) | None | The differences do not affect safety and effectiveness of intended use. |
| Remote Control Required? | Optional | Optional | Optional | None | The differences do not affect safety and effectiveness of intended use. |
| Configuration | Software app installed on a mobile device (iOS or Android) providing wireless selection among preconfigured options and status readout for paired External Stimulator. | Software app installed on a mobile device (Android / iOS) providing wireless selection among preconfigured options and status readout for paired Therapy Disk. | Not publicly available. | None | The differences do not affect safety and effectiveness of intended use. |
| Communication | Bluetooth Low Energy (BLE) | Bluetooth Low Energy (BLE) | Bluetooth Low Energy (BLE) | None | The differences do not affect safety and effectiveness of intended use. |
{15}
bioventus
Bioventus LLC
4721 Emperor Blvd., Suite 100
Durham, NC 27703
USA
1-919-474-6700
1-800-396-4325
www.Bioventus.com
# VII. Performance Testing
The StimTrial System was qualified through the following labeling, software, electrical, usability, bench, functional and animal testing. Additionally, sterility, shelf life and biocompatibility testing were performed.
- Labeling Validation
- Software verification and validation (for External Stimulator, MAPP and CPS software devices)
- EMC, Wireless Co-Existence and Electrical Safety
- Usability testing (for Implanting Physician, Treating Clinician and Patient)
- Bench testing
- Tests of the StimTrial Lead
- Tests of the External Stimulator
- StimTrial System Testing
- Animal testing (Acute and Long-term studies in porcine animal model)
- Sterilization and Shelf Life (StimTrial Surgical Kit)
- Biocompatibility (for StimTrial Lead, Insertion Tools and External Stimulator)
# VIII. Conclusion.
Bioventus concludes that the StimTrial System is substantially equivalent to the Nalu Neurostimulation System for Peripheral Nerve Stimulation predicate device and does not raise any new issues or concerns of safety or effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.