HemoSphere Advanced Monitor (HEM1); HemoSphere Technology Module (HEMTOM10); HemoSphere ClearSight Module (HEMCSM10); Smart Pressure Controller (PC1Q); Acumen IQ Plus Finger Cuff (AIQCA2); HemoSphere Pressure Cable (HEMPSC100); HemoSphere Vita Monitor (HEMVITA1); HemoSphere Vita Technology Module (HEMVTOM1); HemoSphere VitaWave module (HEMVWM1); VitaWave Plus finger cuff (VWCA2)
Applicant
Edwards Lifesciences
Product Code
DQK · Cardiovascular
Decision Date
Jul 23, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Hypotension Prediction Index
—
—
—
—
—
—
—
Assisted Fluid Management
—
—
—
—
—
—
—
Indications for Use
The HemoSphere advanced monitoring platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. The Viewfinder remote mobile application can be used for supplemental near real-time remote display of monitored hemodynamic parameter data as well as faults, alerts and notifications generated by the HemoSphere advanced monitoring platform. The HemoSphere advanced monitoring platform is intended for use with compatible Edwards Swan-Ganz and oximetry catheters, Swan-Ganz Jr catheters, FloTrac sensors, FloTrac Jr sensors, Acumen IQ sensors, TruWave disposable pressure transducers, ForeSight/ForeSight Jr sensors, Acumen IQ fluid meter, and ClearSight/ClearSight Jr/Acumen IQ/Acumen IQ Plus/VitaWave/VitaWave Plus finger cuffs.
Device Story
HemoSphere Advanced Monitoring Platform is a modular hemodynamic monitoring system for hospital critical care (OR, ICU, ER). It processes inputs from various sensors (Swan-Ganz catheters, oximetry cables, pressure cables, finger cuffs, ForeSight sensors) to compute hemodynamic parameters (CO, CI, SV, SVR, MAP, etc.). The system includes the Smart Pressure Controller for noninvasive blood pressure measurement, correcting for hydrostatic height differences between finger and heart. Software features include Acumen Hypotension Prediction Index (HPI) for predicting hypotensive events and Assisted Fluid Management (AFM) for fluid therapy suggestions. Clinicians view data on the HemoSphere monitor or via the Viewfinder remote app. Output provides physiological insight to support clinical decision-making regarding fluid administration and hemodynamic optimization, potentially benefiting patients by enabling goal-directed therapy.
Clinical Evidence
Bench testing only. Verification and validation activities included system integration, mechanical testing, electrical safety (IEC 60601 series), EMC, and software verification. Usability study conducted per FDA guidance confirmed primary operating functions and critical tasks can be performed without usability issues leading to patient/user harm.
Technological Characteristics
Modular platform; components include HemoSphere monitors, technology modules, pressure cables, oximetry cables, and finger cuffs. Sensing principles: thermodilution, oximetry, pressure transduction, and plethysmography. Connectivity: networked/standalone. Software: embedded firmware with GUI. Complies with IEC 60601-1, 60601-1-2, 60601-1-8, 62304, 62366-1, 60601-2-34, 60601-2-57, 60601-2-49, and 80601-2-49.
Indications for Use
Indicated for adult and pediatric critical care patients in hospital environments requiring continuous assessment of cardiac function, fluid status, vascular resistance, and pressure. Specific modules support cardiac output (CO/iCO), venous oxygen saturation (SvO2/ScvO2), tissue oxygen saturation (StO2), and noninvasive blood pressure monitoring. Acumen HPI feature provides physiological insight into likelihood of future hypotensive events for surgical/non-surgical patients. Acumen AFM feature provides fluid therapy response suggestions for surgical patients ≥18 years. ForeSight cable indicated for regional hemoglobin oxygen saturation in ischemic-risk individuals. ClearSight/finger cuffs indicated for patients >18 years.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Acumen Hypotension Prediction Index (HPI) algorithm (K230057)
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 23, 2025
Edwards Lifesciences
Varad Raghuwanshi
Director, Regulatory Affairs
One Edwards Way
Irvine, California 92614
Re: K243781
Trade/Device Name: HemoSphere Advanced Monitor (HEM1); HemoSphere Technology Module (HEMTOM10); HemoSphere ClearSight Module (HEMCSM10); Smart Pressure Controller (PC1Q); Acumen IQ Plus Finger Cuff (AIQCA2); HemoSphere Pressure Cable (HEMPSC100); HemoSphere Vita Monitor (HEMVITA1); HemoSphere Vita Technology Module (HEMVTOM1); HemoSphere VitaWave module (HEMVWM1); VitaWave Plus finger cuff (VWCA2)
Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable Diagnostic Computer
Regulatory Class: Class II
Product Code: DQK, DQE, QAQ, MUD, DXN, DSB, FLL, QMS
Dated: December 7, 2024
Received: December 9, 2024
Dear Varad Raghuwanshi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K243781 - Varad Raghuwanshi
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K243781 - Varad Raghuwanshi
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen C. Browning -S
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K243781
Device Name
HemoSphere Advanced Monitor (HEM1); HemoSphere Technology Module (HEMTOM10); HemoSphere ClearSight Module (HEMCSM10); Smart Pressure Controller (PC1Q); HemoSphere Pressure Cable (HEMPSC100); Acumen IQ Plus Finger Cuff (AIQCA2); HemoSphere Vita Technology Module (HEMVTOM1); HemoSphere VitaWave module (HEMVWM1); VitaWave Plus fin
Indications for Use (Describe)
HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module
The HemoSphere advanced monitor when used with the HemoSphere Swan-Ganz module and Edwards Swan-Ganz catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output (continuous [CO] and intermittent [iCO]) and derived hemodynamic parameters in a hospital environment. Pulmonary artery blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter and Swan-Ganz Jr catheter indications for use statements for information on target patient population specific to the catheter being used. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere Oximetry Cable
The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry Cable and Edwards oximetry catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere Pressure Cable
The HemoSphere advanced monitor when used with the HemoSphere pressure cable is indicated for use in adult and pediatric critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring of hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac sensor, FloTrac Jr sensor, Acumen IQ sensor, and TruWave disposable pressure transducer indications for use statements for information on target patient populations specific to the sensor/transducer being used.
The Edwards Acumen Hypotension Prediction Index software feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
FORM FDA 3881 (8/23)
Page 1 of 3
PSC Publishing Services (301) 443-6740
{5}
HemoSphere Advanced Monitor with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor:
The Acumen Assisted Fluid Management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients $\geq 18$ years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy. Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the Assisted Fluid Management suggestions.
The Acumen Assisted Fluid Management software feature may be used with the Acumen AFM Cable and Acumen IQ fluid meter.
HemoSphere Advanced Monitor with HemoSphere Technology Module and ForeSight Oximeter Cable
The non-invasive ForeSight oximeter cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced-flow or no flow ischemic states. The ForeSight Oximeter Cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight Oximeter Cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere advanced monitor.
- When used with large sensors, the ForeSight Oximeter Cable is indicated for use on adults and transitional adolescents $\geq 40\mathrm{kg}$.
- When used with medium sensors, the ForeSight Oximeter Cable is indicated for use on pediatric subjects $\geq 3\mathrm{kg}$.
- When used with small sensors, the ForeSight Oximeter Cable is indicated for cerebral use on pediatric subjects $< 8\mathrm{kg}$ and non-cerebral use on pediatric subjects $< 5\mathrm{kg}$.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere ClearSight Module
The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller or Smart Pressure Controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with comorbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.
The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
Indication for Acumen IQ Plus and VitaWave Plus finger cuffs:
The Acumen IQ Plus and VitaWave Plus finger cuff adult indicated for patients over 18 years of age to continuously blood pressure and associated hemodynamic parameters when used with a compatible Edwards monitoring platform.
Smart Pressure Controller:
FORM FDA 3881 (8/23)
PSC Publishing Services (301) 443-6740
{6}
The Smart Pressure Controller is intended for use with an Edwards compatible noninvasive monitoring system - composed of compatible monitor, pressure source (pump), compatible Edwards finger cuff(s) and pressure controller - for continuous noninvasive measurement of blood pressure and associated hemodynamic parameters.
Refer to the operator's manual of the compatible Edwards monitor being used for specific information on the intended use environment and patient population.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 3 of 3
PSC Publishing Services (301) 443-6740
{7}
K243781
Edwards
# 510(k) Summary – HemoSphere Alta Advanced Monitoring Platform
## I. Submitter:
**Sponsor:** Edwards Lifesciences LLC
One Edwards Way
Irvine, CA 92614
**Establishment Registration Number:** 2015691
**Contact Person:** Varad Raghuwanshi
Director, Regulatory Affairs
One Edwards Way
Irvine, CA 92614
Telephone: (213) 479-4688
**Date Prepared:** July 22, 2025
## II. Device Information:
**Platform Name (Name of the Device):** HemoSphere Advanced Monitoring Platform
**Trade Name:** HemoSphere Advanced Monitor- HEM1
HemoSphere Vita Monitor- HEMVITA1
HemoSphere ClearSight Module- HEMCSM10
HemoSphere VitaWave Module- HEMVWM1
HemoSphere Technology Module- HEMTOM10
HemoSphere Vita Technology Module- HEMVTOM1
HemoSphere Pressure Cable- HEMPSC100
Smart Pressure Controller- PC1Q
Acumen IQ Plus Finger cuff- AIQCA2
VitaWave Plus Finger cuff- VWCA2
**Common Name:** Programmable Diagnostic Computer, Fiber Optic Oximeter Catheter, Adjunctive predictive Cardiovascular Indicator, Medium-Term Adjunctive Predictive Cardiovascular Indicator, Adjunctive Open Loop Fluid Therapy Recommender, and Noninvasive blood pressure measurement system
**Classification Name:** Programmable Diagnostic Computer
Fiberoptic Oximeter Catheter
21 CFR 870.1425
21 CFR 870.1230
{8}
K243781
Edwards
Adjunctive Predictive Cardiovascular Indicator 21 CFR 870.2210
Oximeter, Tissue Saturation 21 CFR 870.2700
Noninvasive blood pressure measurement system 21 CFR 870.1130
Impedance plethysmograph 21 CFR 870.2770
Thermometer, Electronic, Clinical 21 CFR 880.2910
Adjunctive Open Loop Fluid Therapy Recommender 21 CFR 870.5600
Product Code and Regulatory Class:
DQK, Class II
DQE, Class II
QAQ, Class II
MUD, Class II
DXN, Class II
DSB, Class II
FLL, Class II
QMS, Class II
## III. Predicate Device
### Primary Predicate Device:
- HemoSphere Advanced Monitoring Platform manufactured by Edwards Lifesciences, K223865, cleared June 9th, 2023. It is the base device on which the various hardware and software modifications have been implemented. It has been utilized for substantial equivalence in terms of the graphical user interface (GUI) used, indication for use, intended use, technological characteristics, accessories, basic device functionality, new configuration Pressure controller (PC1Q).
### Additional Predicate Devices:
- HemoSphere Alta Advanced Monitoring Platform (K242451, Cleared Dec 9th, 2024) is being utilized as a secondary predicate for Fluid Meter Only Mode, modification to provide an option to use the existing pulse rate instead of heart rate input for calculating associated parameters, and additional miscellaneous GUI updates for user convenience such as HPI Yellow display and Cardiac Power Output (CPO) as a key parameter. It has been utilized for substantial equivalence in terms of the intended use, and graphical user interface (GUI) used.
{9}
K243781
Edwards
- Acumen IQ finger cuff (AIQCA) by Edwards Lifesciences, K230919, cleared October 24th, 2023, is being utilized for the Acumen IQ Plus finger cuff (AIQCA2). It has been utilized for substantial equivalence in terms of the indication for use, intended use, and technological characteristics
- Acumen Hypotension Prediction Index (HPI) algorithm by Edwards Lifesciences, K230057, cleared on June 8th, 2023, utilized for the HPI Adjustable MAP (Hypotension threshold) feature.
## IV. Device Description
### Device Description:
The HemoSphere Advanced Monitor was designed to simplify the customer experience by providing one platform with modular solutions for all hemodynamic monitoring needs. The user can choose from available optional sub-system modules or use multiple sub-system modules at the same time. This modular approach provides the customer with the choice of purchasing and/or using specific monitoring applications based on their needs. Users are not required to have all of the modules installed at the same time for the platform to function.
## V. Indications for Use:
### HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module
The HemoSphere advanced monitor when used with the HemoSphere Swan-Ganz module and Edwards Swan-Ganz catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output (continuous [CO] and intermittent [iCO]) and derived hemodynamic parameters in a hospital environment. Pulmonary artery blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter and Swan-Ganz Jr catheter indications for use statements for information on target patient population specific to the catheter being used. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
{10}
K243781
Edwards
# HemoSphere Advanced Monitor with HemoSphere Oximetry Cable
The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry Cable and Edwards oximetry catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
# HemoSphere Advanced Monitor with HemoSphere Pressure Cable
The HemoSphere advanced monitor when used with the HemoSphere pressure cable is indicated for use in adult and pediatric critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring of hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac sensor, FloTrac Jr sensor, Acumen IQ sensor, and TruWave disposable pressure transducer indications for use statements for information on target patient populations specific to the sensor/transducer being used.
The Edwards Acumen Hypotension Prediction Index software feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
# HemoSphere Advanced Monitor with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor:
The Acumen Assisted Fluid Management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response
{11}
K243781
Edwards
to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients ≥18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy. Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the Assisted Fluid Management suggestions.
The Acumen Assisted Fluid Management software feature may be used with the Acumen AFM Cable and Acumen IQ fluid meter.
# HemoSphere Advanced Monitor with HemoSphere Technology Module and ForeSight Oximeter Cable
The non-invasive ForeSight oximeter cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced-flow or no flow ischemic states. The ForeSight Oximeter Cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight Oximeter Cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere advanced monitor.
- When used with large sensors, the ForeSight Oximeter Cable is indicated for use on adults and transitional adolescents ≥40 kg.
- When used with medium sensors, the ForeSight Oximeter Cable is indicated for use on pediatric subjects ≥3 kg.
- When used with small sensors, the ForeSight Oximeter Cable is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5 kg.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
# HemoSphere Advanced Monitor with HemoSphere ClearSight Module
The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller or Smart Pressure Controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the
{12}
K243781
Edwards
noninvasive system is indicated for use in patients with comorbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.
The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
## Indication for Acumen IQ Plus and VitaWave Plus finger cuffs:
The Acumen IQ Plus and VitaWave Plus finger cuff adult indicated for patients over 18 years of age to continuously blood pressure and associated hemodynamic parameters when used with a compatible Edwards monitoring platform.
## Smart Pressure Controller:
The Smart Pressure Controller is intended for use with an Edwards compatible noninvasive monitoring system - composed of compatible monitor, pressure source (pump), compatible Edwards finger cuff(s) and pressure controller - for continuous noninvasive measurement of blood pressure and associated hemodynamic parameters.
Refer to the operator's manual of the compatible Edwards monitor being used for specific information on the intended use environment and patient population.
{13}
K243781
Edwards
# Intended Use:
The HemoSphere advanced monitoring platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting.
The Viewfinder remote mobile application can be used for supplemental near real-time remote display of monitored hemodynamic parameter data as well as faults, alerts and notifications generated by the HemoSphere advanced monitoring platform.
The HemoSphere advanced monitoring platform is intended for use with compatible Edwards Swan-Ganz and oximetry catheters, Swan-Ganz Jr catheters, FloTrac sensors, FloTrac Jr sensors, Acumen IQ sensors, TruWave disposable pressure transducers, ForeSight/ForeSight Jr sensors, Acumen IQ fluid meter, and ClearSight/ClearSight Jr/Acumen IQ/Acumen IQ Plus/VitaWave/VitaWave Plus finger cuffs
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere Swan-Ganz module are listed below. Only iCO, iCI, iSVR, and iSVRI are available to the pediatric patient population.
| Parameter | Description | Sub-system technology used | Patient Population | Hospital Environment |
| --- | --- | --- | --- | --- |
| CO | continuous cardiac output | HemoSphere Swan-Ganz module | Adult only | Operating Room, Intensive Care Unit, Emergency Room |
| sCO | STAT cardiac output | | | |
| CPO | Cardiac Power Output | | | |
| CI | continuous cardiac index | | | |
| sCI | STAT cardiac index | | | |
| EDV | right ventricular end diastolic volume | | | |
| sEDV | STAT right ventricular end diastolic volume | | | |
| EDVI | right ventricular end diastolic volume index | | | |
| sEDVI | STAT right ventricular end diastolic volume index | | | |
| HRavg | averaged heart rate | | | |
| LVSWI | left ventricular stroke work index | | | |
| PVR | pulmonary vascular resistance | | | |
| PVRI | pulmonary vascular resistance index | | | |
| RVEF | right ventricular ejection fraction | | | |
| sRVEF | STAT right ventricular ejection fraction | | | |
| RVSWI | right ventricular stroke work index | | | |
| SV | stroke volume | | | |
| SVI | stroke volume index | | | |
| SVR | systemic vascular resistance | | | |
{14}
K243781
Edwards
| SVRI | systemic vascular resistance index | | | |
| --- | --- | --- | --- | --- |
| iCO | intermittent cardiac output | | Adult and Pediatric | |
| iCI | intermittent cardiac index | | | |
| iSVR | intermittent systemic vascular resistance | | | |
| iSVRI | intermittent systemic vascular resistance index | | | |
A comprehensive list of parameters available for adult and pediatric patient populations while monitoring with the HemoSphere advanced monitor and a connected HemoSphere oximetry cable are listed below:
| Parameter | Description | Sub-system technology used | Patient Population | Hospital Environment |
| --- | --- | --- | --- | --- |
| SvO2 | Mixed Venous Oxygen Saturation | HemoSphere oximetry cable | Adult and Pediatric | Operating Room, Intensive Care Unit, Emergency Room |
| ScvO2 | Central Venous Oxygen Saturation | | | |
A comprehensive list of parameters available for adult and pediatric patient populations while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere Swan-Ganz module and oximetry cable are listed below:
| Parameter | Description | Sub-system technology used | Patient Population | Hospital Environment |
| --- | --- | --- | --- | --- |
| DO2 | Oxygen Delivery | HemoSphere Swan-Ganz module and HemoSphere oximetry cable | Adult and Pediatric | Operating Room, Intensive Care Unit, Emergency Room |
| DO2I | Oxygen Delivery Indexed | | | |
| VO2 | Oxygen Consumption | | | |
| VO2e | Estimated Oxygen Consumption when ScvO2 is being monitored | | | |
| VO2I | Oxygen Consumption Index | | | |
| VO2Ie | Estimated Oxygen Consumption Index when ScvO2 is being monitored | | | |
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere Swan-Ganz module and pressure cable are listed below:
| Parameter | Description | Sub-system technology used | Patient Population | Hospital Environment |
| --- | --- | --- | --- | --- |
| CO20s | 20-second cardiac output | HemoSphere Swan-Ganz module and | Adult only | Operating room, intensive care unit, emergency room |
| CI20s | 20-second cardiac index | | | |
| SV20s | 20-second stroke volume | | | |
{15}
K243781
E
Edwards
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere pressure cable are listed below:
| Parameter | Description | Sub-system technology used | Patient Population | Hospital Environment |
| --- | --- | --- | --- | --- |
| CO | Continuous Cardiac Output1 | HemoSphere pressure cable | Adult and Pediatric ≥ 12 years of age | Operating Room, Intensive Care Unit, Emergency Room |
| CI | Continuous Cardiac Index1 |
| CPO | Cardiac Power output |
| CPI | Cardiac Power index |
| DIAART | Systemic arterial diastolic blood pressure |
| MAP | Mean Arterial Pressure |
| PPV | pulse pressure variation1 |
| PR | Pulse rate |
| SV | Stroke Volume1 |
| SVI | Stroke Volume Index1 |
| SVR | Systemic Vascular Resistance1 |
| SVRI | Systemic Vascular Resistance1 Index |
| SVV | Stroke Volume Variation1 |
| SYSAART | Systolic Blood Pressure |
| CVP | Central Venous Pressure | Adult only |
| DIAPAP | pulmonary artery diastolic blood pressure |
| dP/dt | Systolic slope2 |
| Eadyn | Dynamic Arterial Elastance2 |
| HPI | Acumen Hypotension Prediction Index |
| MPAP | Mean Pulmonary Arterial Pressure |
| SYSPAP | Systolic pulmonary artery blood pressure |
| 1FloTrac parameters are available when using a FloTrac/Acumen IQ sensor and if the FloTrac feature is enabled. 2HPI parameters are available when using an Acumen IQ sensor and if the HPI feature is activated. |
A list of Acumen Assisted Fluid Management (AFM) outputs available for surgical patients $\geq 18$ years of age while monitoring with the HemoSphere advanced monitor and a connected HemoSphere pressure cable are listed below:
| AFM output | Sub-system technology used | Patient Population | Hospital Environment |
| --- | --- | --- | --- |
| Fluid Bolus Suggested | | | |
{16}
K243781
E
Edwards
| Test Bolus Suggested | HemoSphere pressure cable | ≥18 years of age only | Operating room |
| --- | --- | --- | --- |
| Fluid Not Suggested | | | |
| Suggestions Suspended | | | |
| Bolus In Progress... | | | |
| Bolus Complete | | | |
| Bolus Complete; Analyzing Hemodynamic Response | | | |
| Tracked Case Vol. | | | |
| Flow Rate | | | |
| Bolus Volume | | | |
| AFM outputs are available when using an Acumen IQ sensor and if the AFM feature is activated. Flow rate and Bolus Volume are visible when using automatic fluid tracking mode. | | | |
A comprehensive list of parameters available for adult and pediatric patient populations while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere pressure cable and oximetry cable are listed below:
| Parameter | Description | Sub-system technology used | Patient Population | Hospital Environment |
| --- | --- | --- | --- | --- |
| DO2 | Oxygen Delivery | HemoSphere pressure cable and HemoSphere oximetry cable | Adult only | Operating Room, Intensive Care Unit, Emergency Room |
| DO2I | Oxygen Delivery Indexed | | | |
| VO2 | Oxygen Consumption | | | |
| VO2e | Estimated Oxygen Consumption when ScvO2 is being monitored | | | |
| VO2I | Oxygen Consumption Index | | | |
| VO2Ie | Estimated Oxygen Consumption Index when ScvO2 is being monitored | | | |
Tissue oxygen saturation, StO2, can be monitored with the HemoSphere advanced monitor, a connected HemoSphere technology module, and the ForeSight oximeter cable as listed below.
| Parameter | Description | Sub-system technology used | Patient Population | Hospital Environment |
| --- | --- | --- | --- | --- |
| StO2 | Tissue oxygen saturation | ForeSight oximeter cable and HemoSphere technology module | Adult and Pediatric | Operating Room, Intensive Care Unit, Emergency Room |
| ΔctHb | relative change in total hemoglobin | | | |
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere ClearSight module are listed below:
| Parameter | Description | Sub-system technology used | Patient Population | Hospital Environment |
| --- | --- | --- | --- | --- |
| CO | Continuous Cardiac Output | HemoSphere | Adult and | Operating |
{17}
K243781
E
Edwards
| CPO | Cardiac Power Output | ClearSight module | pediatric ≥ 12 | Room, Intensive Care Unit, Emergency Room |
| --- | --- | --- | --- | --- |
| CI | Continuous Cardiac Index |
| DIAART | arterial diastolic blood pressure |
| SYSAART | Systolic Blood Pressure |
| MAP | Mean Arterial Pressure |
| PPV | pulse pressure variation |
| PR | Pulse rate |
| SV | Stroke Volume |
| SVI | Stroke Volume Index |
| SVR | Systemic Vascular Resistance |
| SVRI | Systemic Vascular Resistance Index |
| SVV | Stroke Volume Variation |
| dP/dt | systolic slope | Adult only |
| Eadyn | Dynamic Arterial Elastance1 |
| HPI | Acumen Hypotension PredictionIndex1 |
| 1HPI parameters are available when using an Acumen IQ or Acumen IQ Plus finger cuff and heart reference sensor (HRS), if applicable, and if the HPI feature is activated. Activation is only available in certain areas. Please contact your local Edwards representative for more information on enabling this advanced feature. NOTE: CO/CI and SV/SVI are measured using a reconstructed brachial arterial waveform. All other monitored parameters use a reconstructed radial arterial waveform. SVR/SVRI are derived from CO/CI and MAP along with an entered or monitored CVP value. |
A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere ClearSight module and oximetry cable are listed below:
| Parameter | Description | Sub-System Module Used | Patient Population | Hospital Environment |
| --- | --- | --- | --- | --- |
| DO2 | Oxygen Delivery | HemoSphere ClearSight Module and HemoSphere Oximetry Cable | Adult only | Operating Room, Intensive Care Unit |
| DO2I | Oxygen Delivery Indexed | | | |
| VO2 | Oxygen Consumption | | | |
| VO2e | Estimated Oxygen Consumption when ScvO2 is being monitored | | | |
| VO2I | Oxygen Consumption Index | | | |
| VO2Ie | Estimated Oxygen Consumption Index when ScvO2 is being monitored | | | |
{18}
K243781
Edwards
# VI. Comparison of Technological Characteristics with the Predicate Devices:
The subject and predicate devices are based on the following same technological elements:
- Hardware: The subject device uses the same platform as the predicate (K223865, cleared June 9th, 2023). Additionally, the updated hardware configurations for the components: Acumen IQ Plus finger cuff and Smart Pressure Controller remain similar in terms of the functionality, intended use and indication to their respective predicate devices.
- Indication and Intended Use: The subject device has the same intended use and indication as the predicates.
- Technological characteristics: The new features included in the subject device have the same technological characteristics as their respective predicate devices.
- Accessories/Components: The subject and the predicate device both use the same accessories and components.
The following technological differences exist between the subject and predicate devices:
- Hardware: The subject device includes new hardware configuration of the Pressure Controller and finger cuff: Acumen IQ/VitaWave Plus finger cuff and Smart Pressure Controller. The updated configuration includes modifications to provide correction for the hydrostatic height difference between the finger and heart.
- Software: The subject device includes updated software to support the new features.
- Graphical User Interface (GUI): The subject device includes an updated graphical user interface to accommodate the new features.
A brief description of the key highlights of the subject HemoSphere Advanced Monitoring Platform and a comparative overview with the predicate HemoSphere Advanced Monitoring Platform is provided below:
{19}
K243781
Edwards
Software modifications to the subject HemoSphere Advanced Monitoring Platform (K223865, Cleared June 9th, 2023):
- Software modification to the HemoSphere Advanced Monitor (HEM1): The existing software components of HemoSphere Advanced Monitor (HEM1-SHM and DBM) have been updated to support graphical user interface (GUI) associated with the new features introduced to the HemoSphere Advanced Monitoring Platform.
The GUI updates include the following features:
- Fluid meter only Mode: For user convenience, the automatic bolus tracking feature can now also be used independently in "Fluid Meter Only" mode to track fluid administration. In this mode, users can track fluid delivery via an existing Acumen fluid meter instead of manually tracking and logging. During this mode, the AFM algorithm will be disabled, and the user will not get any bolus suggestions.
- Modification to provide an option to use the existing pulse rate instead of heart rate input for calculating associated parameters: The HemoSphere Advanced Monitoring Platform currently requires a heart rate (HR) signal to be computed to calculate parameters such as EDV, EDVI, RVEF, sEDV, sEDVI, SV and SVI parameters. Modifications have been made to allow users to have the option to use HR or alternatively utilize existing pulse rate (PR) for the parameter calculation.
- Miscellaneous Graphical User Interface (GUI) updates: The HemoSphere Advanced Monitoring Platform GUI has been updated to support the features listed below and to provide user convenience.
- Integration of the existing Acumen Hypotension Prediction Index (HPI) Algorithm: The previously cleared HPI Adjustable MAP target algorithm that was cleared in K230057 on June 8, 2023, is being integrated to the HemoSphere Advanced Monitoring Platform. There are no changes to the Acumen HPI algorithm from what was cleared in K230057. The same feature is now integrated
{20}
K243781
Edwards
under the name “Hypotension threshold” into the HemoSphere Advanced Monitoring Platform.
The existing HemoSphere Pressure Cable (HEMPSC100) and HemoSphere ClearSight Module (HEMCSM10) software have been updated to support integration of the HPI Adjustable MAP target algorithm.
- Software modification to the HemoSphere ClearSight/VitaWave Module (HEMCSM10/HEMVWM1): The existing HemoSphere ClearSight Module software has been updated to support a new configuration of the Pressure Controller.
Hardware modifications to the subject HemoSphere Advanced Monitoring Platform (original K223865, Cleared June 9th, 2023):
- Smart Pressure Controller: A new hardware configuration of the existing accessory, Pressure controller (PC2/HEMPC2) most recently cleared in K223865, is being introduced to provide correction for the hydrostatic height difference between the finger and heart under the tradename “Smart Pressure Controller” (Model- PC1Q).
Hardware modifications to the subject Acumen IQ finger cuff, Model: AIQCA (K230919, Cleared October 24th, 2023):
- Acumen IQ Plus finger cuff (AIQCA2): A new configuration of the Acumen IQ finger cuff is being introduced to provide continuous noninvasive measurement of blood pressure and hemodynamic monitoring under the tradename of Acumen IQ Plus finger cuff (AIQCA2) and VitaWave Plus finger cuff (VWCA2).
Labelling Modifications to the HemoSphere Advanced Monitoring Platform and Acumen IQ finger Cuff to address the device modifications
New labelling is being created for the updated Intended use and Indications for use for Acumen IQ/VitaWave Plus finger cuff.
{21}
K243781
Edwards
- Indication update to include Acumen Hypotension Prediction Index (HPI) modification cleared in K230057 on June 8, 2023: The Acumen Hypotension Prediction Index (HPI) feature indication has been updated to include the Adjustable MAP targets and patient population to include non-surgical patients during non-invasive monitoring cleared in K230057 on June 8, 2023.
## Performance Data:
The following verification activities were performed in support of a substantial equivalence determination.
## Usability Study
Usability study was conducted per FDA’s guidance document “Applying Human Factors and Usability Engineering to Medical Devices” to investigate primary operating functions and critical tasks of the system for any usability issues regarding the HemoSphere Advanced Monitoring Platform that may lead to patient or user harm.
The usability study demonstrated that the intended users can perform primary operating functions and critical tasks of the system without any usability issues that may lead to patient or user harm.
## System Verification (Non-Clinical Performance):
Completion of all verification and validation activities demonstrated that the subject devices meet their predetermined design and performance specifications. Verification activities performed confirmed that the differences in the design and materials used did not adversely affect the safety and effectiveness of the subject device.
Measured and derived parameters were tested using a bench simulation. Additionally, system integration and mechanical testing was successfully conducted to verify the safety and effectiveness of the device. All tests passed.
## Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the subject HemoSphere Advanced Monitoring Platform, consisting of the HemoSphere Monitor, HemoSphere Vita Monitor, HemoSphere Swan-Ganz Module, HemoSphere Oximetry Cable, HemoSphere Pressure Cable, HemoSphere Technology Module, ForeSight Oximeter Cable, ClearSight Module, VitaWave Module, Pressure Controller, Heart Reference Sensor, Acumen AFM Cable, and Smart Pressure Controller. The system complies with the IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8,
{22}
K243781
E
Edwards
IEC 62304, IEC 62366-1, IEC 60601-2-34, IEC 60601-2-57, IEC 60601-2-49 and IEC 80601-2-49. All tests passed.
## Software Verification
Software verification testing was conducted, and documentation was provided per FDA’s Guidance for Industry and FDA Staff, “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices”. All tests passed.
## Conclusion
The technological characteristics of the subject and predicate devices are identical. The HemoSphere Advanced Monitoring Platform has successfully passed functional and performance testing, including software verification and validation, system integration, mechanical, electrical, human factor usability, and safety testing. The testing performed demonstrates that the HemoSphere Advanced Monitoring Platform is substantially equivalent to the legally marketed predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.