K243753 · Roche Molecular Systems, Inc. · OZZ · Nov 20, 2025 · Microbiology
Device Facts
Record ID
K243753
Device Name
cobas liat Bordetella panel nucleic acid test
Applicant
Roche Molecular Systems, Inc.
Product Code
OZZ · Microbiology
Decision Date
Nov 20, 2025
Decision
SESE
Submission Type
Dual Track
Regulation
21 CFR 866.3980
Device Class
Class 2
Indications for Use
The cobas liat Bordetella panel nucleic acid test (cobas liat Bordetella panel) is an automated real-time polymerase chain reaction (PCR) test intended for the simultaneous qualitative detection and differentiation of Bordetella pertussis (Bp), Bordetella parapertussis (Bpp), and Bordetella holmesii (Bh) nucleic acid in human nasopharyngeal swabs taken from patients with suspected pertussis respiratory infection. The test is meant to be used in conjunction with other clinical and epidemiological information and laboratory findings. When clinical factors suggest that B. pertussis, B. parapertussis or B. holmesii may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines. Negative results do not preclude Bp, Bpp, or Bh infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out co-infection with other bacteria or viruses. The agent detected may not be the definite cause of disease.
Device Story
Automated multiplex real-time PCR assay for qualitative detection/differentiation of B. pertussis, B. parapertussis, and B. holmesii DNA in nasopharyngeal swabs. System comprises cobas liat analyzer and single-use disposable assay tube. Operator scans specimen/assay tube barcodes, adds sample to tube; system automates nucleic acid extraction, purification, amplification, and real-time detection using TaqMan probes. Results delivered in ~15 minutes. Used in near-patient settings (emergency rooms, outpatient clinics) by CLIA-waived users. Output provides qualitative detection status for each target; aids clinical decision-making alongside other clinical/epidemiological data. Benefits include rapid turnaround time for respiratory infection diagnosis.
Clinical Evidence
Multi-site prospective study (n=823) in U.S. (July 2023–Feb 2024) plus archived and contrived samples. Comparator: FDA-cleared NAAT and bi-directional sequencing. Overall PPA: 99.2% (Bp), 99.3% (Bpp), 100% (Bh). Overall NPA: 99.8% (Bp), 99.9% (Bpp), 99.7% (Bh). Analytical studies confirmed LoD, inclusivity, and lack of cross-reactivity/interference.
Technological Characteristics
Automated real-time PCR; TaqMan probe-based detection with fluorescent dyes. Single-use disposable assay tube with frangible seals for reagent storage. Integrated system with thermal, loading, and motion units. CLIA-waived. Software-controlled via assay-specific package (script).
Indications for Use
Indicated for patients with suspected pertussis respiratory infection. Used for simultaneous qualitative detection and differentiation of B. pertussis, B. parapertussis, and B. holmesii nucleic acid in nasopharyngeal swabs. Not for use as sole basis for treatment decisions; does not rule out co-infection.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
Predicate Devices
Verigene Respiratory Pathogens Flex Nucleic Acid Test (Verigene RP Flex) (K143653)
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K243753
B Applicant
Roche Molecular Systems, Inc.
C Proprietary and Established Names
cobas liat Bordetella panel nucleic acid test
Common name: cobas liat Bordetella panel
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| OZZ | Class II | 21 CFR 866.3980 - Respiratory Viral Panel Multiplex Nucleic Acid Assay | MI - Microbiology |
| OOI | Class II | 21 CFR 862.2570 – Instrumentation for clinical multiplex test systems | CH - Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
The purpose of this submission is to obtain a substantial equivalence determination for the cobas liat Bordetella panel nucleic acid test on the cobas Liat System.
B Measurand:
Bordetella pertussis, Bordetella parapertussis, and Bordetella holmesii nucleic acids
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
C Type of Test:
This assay is a multiplex nucleic acid assay for the qualitative detection and differentiation of Bordetella pertussis, B. parapertussis, and B. holmesii DNA through nucleic acid extraction, amplification, and detection using real-time PCR. All steps of the assay are automated within the cobas Liat System, after scanning the specimen ID barcode, scanning the assay tube barcode, and the manual addition of sample into the assay tube.
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The cobas liat Bordetella panel nucleic acid test (cobas liat Bordetella panel) is an automated real-time polymerase chain reaction (PCR) test intended for the simultaneous qualitative detection and differentiation of Bordetella pertussis (Bp), Bordetella parapertussis (Bpp), and Bordetella holmesii (Bh) nucleic acid in human nasopharyngeal swabs taken from patients with suspected pertussis respiratory infection.
The test is meant to be used in conjunction with other clinical and epidemiological information and laboratory findings. When clinical factors suggest that B. pertussis, B. parapertussis or B. holmesii may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines.
Negative results do not preclude Bp, Bpp, or Bh infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out co-infection with other bacteria or viruses. The agent detected may not be the definite cause of disease.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
For use with the cobas Liat system
IV Device/System Characteristics:
A Device Description:
The cobas liat Bordetella panel nucleic acid test (cobas liat Bordetella panel) is an automated multiplex real-time polymerase chain reaction (PCR) assay for the rapid in vitro qualitative detection and differentiation of B. pertussis (Bp), B. parapertussis
K243753 - Page 2 of 19
{3}
(Bpp), and B. holmesii (Bh) DNA in human nasopharyngeal swabs taken from patients with suspected pertussis respiratory infection.
The different fluorescent dye designs enable the specific detection and differentiation of the three microorganisms (Bp, Bpp, and Bh) independently in a multiplex system. The system automates all nucleic acid amplification test sample processing steps, including inhibitor removal, nucleic acid extraction, purification, amplification, real-time detection, and result interpretation in a rapid manner. The test is designed for use in near-patient settings to deliver results in approximately 15 minutes.
The cobas liat system is comprised of the cobas liat analyzer (analyzer) hardware with integrated cobas liat system software (analyzer software + liat assay specific package (script)) for running tests and analyzing the results, and a single-use disposable cobas liat assay tube (assay tube).
## Reagents and Controls:
- Cobas liat Bordetella panel
- Cobas liat Bordetella panel control kit
## Additional materials required but not provided:
- Nasopharyngeal swab collection kit
- Flexible minitip FLOQSwab with Universal Transport Media from Copan Diagnostics or BD Universal Viral Transport 3-mL collection kit with a flocked flexible minitip swab
## B Principle of Operation:
The test is performed on the cobas liat analyzer which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using real-time PCR assays. The assay targets the IS110-like element IS1663 family transposase of Bordetella pertussis, the ISL3-like element IS1001 family transposase of Bordetella parapertussis, and the ISL3 family transposase of Bordetella holmesii. An Internal Control (IC) is also included. The IC is present to control for adequate processing of the target bacteria through steps of sample purification, nucleic acid amplification, and to monitor the presence of inhibitors in the PCR processes.
## C Instrument Description:
The cobas liat system can automatically perform qualitative in vitro nucleic acid amplification tests. The cobas liat system is comprised of the cobas liat analyzer hardware with integrated cobas liat system software and a single-use disposable cobas liat assay tube. The cobas liat analyzer is a system component that consists of one software subsystem and three hardware units: a) Infrastructure unit which consists of the hardware and embedded software (firmware); b) thermal, loading and motion unit which is the processing module that interacts physically with the assay tube during the assay execution, and c) Detection unit consisting of the photodetectors that is used for the fluorescence detection during the PCR reaction. The assay script provides a set of instructions to the analyzer hardware and software for assay tube processing, PCR, result calculation and interpretation and result reporting. The assay script can be installed on the
K243753 - Page 3 of 19
{4}
analyzer independently of the analyzer software. The cobas liat Bordetella panel nucleic acid test is supported with a Liat Assay Specific Package (LASP).
The assay tube holds all reagents needed for sample preparation and PCR processes. Collectively with an assay tube, the analyzer performs reagent preparation, target enrichment, inhibitor removal, nucleic acid extraction, polymerase chain reaction (PCR) amplification, real-time detection and result interpretation to automate the detection of nucleic acid targets in the biological sample.
V Substantial Equivalence Information:
A Predicate Device Name(s):
Verigene Respiratory Pathogens Flex Nucleic Acid Test (RP Flex)
B Predicate 510(k) Number(s):
K143653
C Comparison with Predicate(s):
| Device & Predicate Device | K243753 | K143653 |
| --- | --- | --- |
| Device Trade Name | cobas liat Bordetella Panel nucleic acid test | Verigene Respiratory Flex Panel |
| General Device Characteristic Similarities | | |
| Regulation name | 866.3980 | 866.3980 |
| Product code | OZZ
OOI | OZZ, OOI, OCC, OEM, OEP, OOU, OZE |
| Intended Use | The cobas liat Bordetella panel nucleic acid test (cobas liat Bordetella panel) is an automated real-time polymerase chain reaction (PCR) test intended for the simultaneous qualitative detection and differentiation of Bordetella pertussis (Bp), Bordetella parapertussis (Bpp), and Bordetella holmesii (Bh) nucleic acid in human nasopharyngeal swabs taken from patients with suspected pertussis respiratory infection.
The test is meant to be used in conjunction with other clinical and epidemiological | The Verigene Respiratory Pathogens Flex Nucleic Acid Test (RP Flex) is a multiplexed qualitative test intended for the simultaneous detection and identification of multiple viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infection. The test is performed on the automated Verigene System utilizing reverse transcription (RT), polymerase chain reaction (PCR), and microarray hybridization to detect gene sequences of the following organism types and subtypes: |
K243753 - Page 4 of 19
{5}
K243753 - Page 5 of 19
| Device & Predicate Device | K243753 | K143653 |
| --- | --- | --- |
| | information and laboratory findings. When clinical factors suggest that B. pertussis, B. parapertussis or B. holmesii may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines.
Negative results do not preclude Bp, Bpp, or Bh infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule out co-infection with other bacteria or viruses. The agent detected may not be the definite cause of disease. | Viruses:
Adenovirus
Human
Metapneumovirus
Influenza A
Influenza A (Subtype H1)
Influenza A (Subtype H3)
Influenza B
Parainfluenza 1
Parainfluenza 2
Parainfluenza 3
Parainfluenza 4
Respiratory Syncytial Virus A
Respiratory Syncytial Virus B
Rhinovirus
Detecting and identifying specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory infection, if used in conjunction with other clinical and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or patient management decisions. Negative results in the presence of a respiratory illness do not preclude respiratory infection and may be due to infection with pathogens that are not detected by this test or lower respiratory tract infection that is not detected by an NPS specimen. Conversely, positive results do not rule-out infection or co-infection with organisms not detected by RP Flex. The agent(s) detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation may be necessary to establish a final diagnosis of respiratory infection. Clinical evaluation indicates a lower sensitivity specific to RP Flex for the |
{6}
K243753 - Page 6 of 19
| Device & Predicate Device | K243753 | K143653 |
| --- | --- | --- |
| | | detection of Rhinovirus. If infection with Rhinovirus is suspected, negative samples should be confirmed using an alternative method. Performance characteristics for Influenza A were established when Influenza A/H1 (2009 Pandemic) and A/H3 were the predominant Influenza A viruses in circulation. RP Flex may not detect novel Influenza A strains. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions used specifically for novel virulent influenza viruses and sent to appropriate health authorities for testing. Viral culture should not be attempted in these cases unless a biosafety level (BSL) 3+ facility is available to receive and culture specimens. |
| Sample Type | Nasopharyngeal swab (NPS) | Same |
| Ancillary Collection Kits | Flocked swab with Universal Transport Media (UTM) | Same |
| Amplification Technology | Real-time PCR | Same |
| Controls Used | Sample processing control (IC) and Positive and negative control | Same |
| **General Device Characteristic Differences** | | |
| Analyte Targets | Bordetella pertussis (BP)
Bordetella parapertussis (BPP)
Bordetella holmesii (BH) | Viruses:
Adenovirus
Human
Metapneumovirus
Influenza A
Influenza A (Subtype H1)
Influenza A (Subtype H3)
Influenza B
Parainfluenza 1
Parainfluenza 2
Bacteria:
Bordetella parapertussis/bronchiseptica,
Bordetella holmesii,
Bordetella pertussis |
{7}
| Device & Predicate Device | K243753 | K143653 |
| --- | --- | --- |
| | | Parainfluenza 3
Parainfluenza 4
Respiratory Syncytial
Virus A
Respiratory Syncytial
Virus B
Rhinovirus |
| Time to result | ~15 minutes | ~ 2 hours |
| Detection chemistry | TaqMan probes with fluorescent dyes | Microarray hybridization, Gold nanoparticles probe with light scattering |
| Instrumentation | cobas liat system | Verigene System |
VI Standards/Guidance Documents Referenced:
ISO 14971 – Application of risk management to medical devices.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
A reproducibility study was conducted to assess the total variability of the cobas liat Bordetella panel assay across operators, study sites, testing days, cobas liat analyzers, and cobas liat assay tube lots. The reproducibility study was conducted across three CLIA waived sites using a testing panel of three sample types: low positive (1-2x LoD), moderate positive (3-5x LoD), and negative samples. Each sample was run in triplicate on three analyzers across five different days with three different reagent lots. The study was performed by two operators/site resulting in approximately 270 test results/panel member or 810 total test results (3 panel members × 3 replicates × 2 operators × 5 days × 3 sites × 3 lots).
The reproducibility panel samples were prepared by spiking different concentrations of one strain each of Bp, Bpp and Bh bacteria into a UTM-based human clinical matrix. The panels were provided to the sites with coded sample identification numbers to reduce bias. Each sample was processed according to the cobas liat Bordetella panel instructions for use. Analysis of the Ct signal variability for the positive panel members is presented below in Table 1.
K243753 - Page 7 of 19
{8}
Table 1: Reproducibility Results of Positive Panel Members
| Panel Member | Target Analyte | Mean Ct | Between-Site | | Between-Lot | | Between-Day | | Between-Run (Operator) | | Repeatability | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% |
| Low Positive (1-2x LoD) | B. pertussis | 33.2 | 0.05 | 0.1 | 0.35 | 1.1 | 0.23 | 0.7 | 0.15 | 0.5 | 0.59 | 1.8 | 0.74 | 2.2 |
| | B. parapertussis | 31.9 | 0.17 | 0.5 | 0.31 | 1.0 | 0.15 | 0.5 | 0.00 | 0.0 | 0.57 | 1.8 | 0.68 | 2.1 |
| | B. holmesii | 27.8 | 0.00 | 0.0 | 0.27 | 1.0 | 0.14 | 0.5 | 0.11 | 0.4 | 0.56 | 2.0 | 0.65 | 2.3 |
| Moderate Positive (3-5x LoD) | B. pertussis | 32.0 | 0.20 | 0.6 | 0.49 | 1.5 | 0.00 | 0.0 | 0.22 | 0.7 | 0.45 | 1.4 | 0.73 | 2.3 |
| | B. parapertussis | 30.5 | 0.21 | 0.7 | 0.29 | 1.0 | 0.14 | 0.5 | 0.05 | 0.2 | 0.51 | 1.7 | 0.64 | 2.1 |
| | B. holmesii | 26.8 | 0.14 | 0.5 | 0.29 | 1.1 | 0.04 | 0.1 | 0.14 | 0.5 | 0.41 | 1.5 | 0.54 | 2.0 |
Ct: cycle threshold, CV%: percent coefficient of variation, LoD: limit of detection, SD: standard deviation.
The total Ct CV% ranged from 2.0 – 2.3 across the target panel members tested. These results indicate that the reproducibility of the cobas liat Bordetella panel assay on the liat system is acceptable in NPS samples. For all positive panel members, the repeatability/within-run factor (i.e., random error) followed by the between lot reproducibility was the largest contributor to total variability. Percent agreement across the three testing sites is summarized in Table 2.
Table 2: Reproducibility Result Summary Across Sites
| Panel Member | Target Analyte | Total number of valid test runs | % Agreement (n Agreement/N Tested) [95% CI] | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | Site A | Site B | Site C | Overall |
| Negative | N/A | 264 | 100.0 (87/87) | 100.0 (89/89) | 100.0 (88/88) | 100.0 (264/264) [98.6, 100.0] |
| Low Positive (1-2× LoD) | B. pertussis | 258 | 100.0 (85/85) | 100.0 (85/85) | 98.9 (87/88) | 99.6 (257/258) [97.8, 99.9] |
| | B. parapertussis | 258 | 100.0 (85/85) | 100.0 (85/85) | 98.9 (87/88) | 99.6 (257/258) [97.8, 99.9] |
| | B. holmesii | 258 | 100.0 (85/85) | 100.0 (85/85) | 98.9 (87/88) | 99.6 (257/258) [97.8, 99.9] |
| Moderate Positive (3-5× LoD) | B. pertussis | 265 | 100.0 (88/88) | 100.0 (89/89) | 100.0 (88/88) | 100.0 (265/265) [98.6, 100.0] |
| | B. parapertussis | 265 | 100.0 (88/88) | 100.0 (89/89) | 100.0 (88/88) | 100.0 (265/265) [98.6, 100.0] |
| | B. holmesii | 265 | 100.0 (88/88) | 100.0 (89/89) | 100.0 (88/88) | 100.0 (265/265) [98.6, 100.0] |
The cobas liat Bordetella panel assay demonstrated $100\%$ agreement for the negative panel members and for all target analytes tested at the moderate positive concentration across the three testing sites. For low positive panel members, the assay yielded $99.6\%$ agreement for all target analytes (see Table 2 above). Notably, there was one negative test result for all 3 target analytes when tested at 1-2x LoD, occurring at site C, with one operator on day two. Overall, the total agreement of $99.6\%$ for low positive panel member is acceptable, since the analyte concentration between 1-2x LoD is expected to yield a $\geq 95\%$ detection rate.
2. Linearity:
Not applicable; this is a qualitative assay.
K243753 - Page 8 of 19
{9}
# 3. Analytical Specificity/Interference:
# i. Cross-reactivity and Microbial Interference
A cross-reactivity and microbial interference study was conducted to verify the analytical specificity of the cobas liat Bordetella panel by demonstrating that microorganisms closely related to Bp, Bpp and Bh as well as common respiratory flora, pathogens and other clinically relevant microorganisms that could be present in nasopharyngeal swab specimens, do not cause false positive results due to cross-reactivity or false negative results due to interference. Forty-four different microorganisms listed in Table 3 representing closely related or found in the upper respiratory tract were individually mixed with pooled negative clinical matrix with and without co-formulated Bp, Bpp and Bh isolates at $\sim 3\mathrm{x}$ LoD concentration. Samples that contained co-formulated Bp, Bpp and Bh assessed potential interference with detection of one or all of the target analytes, while samples without Bp, Bpp and Bh assessed potential cross-reactivity.
Each potentially interfering microorganism was tested at $1\mathrm{E} + 06$ CFU/mL for bacterial and yeast organisms and $1.0\mathrm{E} + 05$ TCID of copies/mL for viruses or at the highest concentration possible. Results from the study indicated that none of the non-target organisms generated false positive or false negative results due to cross-reactivity or interference.
The in-silico analysis for possible cross-reactivity was conducted by mapping binding regions of the primers and probes in the cobas assay to sequences available from the NCBI database. In silico analysis revealed two potential cross-reactants for the Bpp target, $B$ . bronchiseptica and $A$ . denitrificans (i.e., greater than or equal to $80\%$ homology for the Bpp primers and probe). However, no cross-reactivity was observed in wet testing using three different strains of these organisms listed in Table 3.
Table 3: List of Microorganisms tested for Analytical Specificity or Interference
| Micro-organism name, Bacteria/Fungi | | Micro-organism name/Virus |
| --- | --- | --- |
| Bordetella bronchiseptica* | Legionella pneumophila | Cytomegalovirusa |
| Bordetella avium | Moraxella catarrhalis | Epstein-Barr Virus |
| Bordetella hinzii | Mycobacterium tuberculosis | Influenza A |
| Bordetella petrii | Mycoplasmoides pneumoniae | Influenza B |
| Bordetella trematum | Neisseria gonorrhoeae | Human Rhinovirus |
| Achromobacter denitrificans** | Neisseria meningitidis | Respiratory Syncytial Virus Type A |
| Chlamydia pneumoniae | Pseudomonas aeruginosa | Adenovirus |
| Chlamydia trachomatis | Methicillin-resistant Staphylococcus aureus | Human Coronavirus |
| Corynebacterium diphtheria | Staphylococcus epidermidis | Herpes Simplex Virus Type 1 |
K243753 - Page 9 of 19
{10}
| Micro-organism name, Bacteria/Fungi | | Micro-organism name/Virus |
| --- | --- | --- |
| Escherichia coli | Stenotrophomonas maltophilia | Herpes Simplex Virus Type 2 |
| Haemophilus influenzae | Streptococcus pneumoniae | Measles |
| Klebsiella pneumoniae | Streptococcus pyogenes | Mumps |
| Lactobacillus acidophilus | Streptococcus salivarius | SARS-WA1 |
| Lactiplantibacillus plantarum | Candida albicans (fungus) | Human Metapneumovirus |
| -- | -- | Human Bocavirusa |
| -- | -- | Enterovirus*** |
* Three strains of B. bronchiseptica were tested: ATCC 785, Z339 (NCTC 452), Z341 (NCTC 13252)
** Three strains of A. denitrificans were tested: ATCC 337, ATCC 13138, ATCC 15173
*** Three strains of Enterovirus were tested: ATCC 836, ATCC 1824, ATCC 1432
${}^{a}$ Tested at highest concentration possible per stock concentration
# ii. Exogenous/endogenous Interference
Potentially interfering exogenous/endogenous substances that are commonly encountered in nasopharyngeal swab specimens were tested at the concentrations found in Table 4. Each substance was tested in replicates of five both with and without co-formulated Bp, Bpp and Bh bacterial culture (each at $3\mathrm{x}$ LoD) spiked into negative clinical matrix. Internal controls remained valid for all samples and replicates. One false negative result was reported with Bpp and $10\%$ Flonase; however, reduction to $5\%$ Flonase resolved the interference. All other substances tested did not interfere with the performance of the cobas liat Bordetella panel assay.
Table 4: Exo/Endogenous Substances Interference Results Summary
| Interferent | Concentration | Positive Sample # Positive/ # Total (%Agreement) | | | Negative Sample # Positive/ # Total (% Agreement) | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Bp | Bpp | Bh | Bp | Bpp | Bh |
| Human cell (PBMC) | 1E+06 cell/mL | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Mucin Bovine Type 1-S | 10 mg/mL | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Human Whole Blood | 10% (v/v) | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Afrin Pump Mist | 10.0% (v/v) | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Flonase | 5.0% (v/v) | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| | 10% (v/v) | 5/5 (100%) | 4/5 (80%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Chloraseptic Sore Throat Lozenges | 10.0 mg/mL | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Mupirocin | 10.0 mg/mL | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
K243753 - Page 10 of 19
{11}
K243753 - Page 11 of 19
| Interferent | Concentration | Positive Sample # Positive/# Total (%Agreement) | | | Negative Sample # Positive/# Total (% Agreement) | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Bp | Bpp | Bh | Bp | Bpp | Bh |
| Relenza | 10.0 mg/mL | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Tamiflu | 7.5 mg/mL | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Tobramycin | 0.6 mg/mL | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Albuterol Sulfate | 5.0% (w/v) | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Sucrets Sore Throat & Cough – Vapor Cherry | 5.0% (w/v) | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Ampicillin powder | 10.0 mg/mL | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Azithromycin tablet | 0.25 mg/mL | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Beconase AQ Nasal Spray | 8.5 mg/mL | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Children’s Dimetapp Cold and Cough | 5.0% (v/v) | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Ciprofloxacin | 1.3 mg/mL | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Erythromycin | 0.47 mg/mL | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Neo-Synephrine Nasal Spray | 15.0% (v/v) | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Rifampicin | 3.75 mg/mL | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Robitussin Cough + Chest Congestion DM liquid gel capsule | 10.0% (v/v) | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Ocean Saline Nasal Spray (sodium chloride) | 10.0% (v/v) | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
| Flumist | 6.3% (v/v) | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) | 0/5 (100%) | 0/5 (100%) | 0/5 (100%) |
iii. Competitive Inhibition
A competitive interference study was conducted to evaluate the potential interference of on-panel organisms when one or more targets are present at high concentrations relative to the other targets. Contrived mixtures of low (3x LoD) and high (1.0E+07 CFU/mL) concentration targets in negative clinical matrix were tested with the cobas liat Bordetella panel assay to assess competitive inhibition between Bp, Bpp and Bh analytes. Five replicates of each co-formulated sample were utilized. Test results as shown in Table 5 indicate that when two target microorganisms were present at high concentrations, no interference was observed for microorganisms that were present at low concentrations.
Table 5: Competitive Inhibition Result Summary
| Concentration | | | # Detected (%) | | |
| --- | --- | --- | --- | --- | --- |
| Bp | Bpp | Bh | Bp | Bpp | Bh |
| Low | Low | Low | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) |
| High | High | Low | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) |
{12}
K243753 - Page 12 of 19
| Concentration | | | # Detected (%) | | |
| --- | --- | --- | --- | --- | --- |
| Bp | Bpp | Bh | Bp | Bpp | Bh |
| High | Low | High | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) |
| Low | High | High | 5/5 (100%) | 5/5 (100%) | 5/5 (100%) |
4. Assay Reportable Range:
Not applicable; this is a qualitative assay
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
i. Specimen Stability
Pooled nasopharyngeal swab (NPS) clinical specimens were stored at various times at each of three storage temperatures: ≤ -70°C, 2-8°C, and 30°C using three different media types: UTM, M4RT and eSwab. For each media type, two replicates each of five unique specimen pools negative for all assay targets (Bp/Bpp/Bh) and five unique specimen pools positive for all assay targets at 3x LoD were tested, giving a total of 10 positive and 10 negative results for each media at each storage temperature and at each time point as shown in Table 6 below.
Table 6: Conditions Evaluated for Specimen Stability
| Sample* | Temperature | Timepoint | No. of Replicates |
| --- | --- | --- | --- |
| Negative | 30°C, Sample | 4hr | 10 |
| 5hr | 10 |
| 30°C, Tube | 4hr | 10 |
| 5hr | 10 |
| 2-8°C | 0 | 10 |
| 1d | 10 |
| 2d | 10 |
| 3d | 10 |
| 4d | 10 |
| -80°C, 2 F/T | 3d | 10 |
| 4d | 10 |
| -80°C, 3 F/T | 3d | 10 |
| 4d | 10 |
| 103d# | 10 |
| Positive | 30°C, Sample | 4hr | 10 |
| 5hr | 10 |
| 30°C, Tube | 4hr | 10 |
| 5hr | 10 |
| 2-8°C | 0 (Baseline) | 10 |
| 1d | 10 |
| 2d | 10 |
| 3d | 10 |
| 4d | 10 |
| -80°C, 2 F/T | 3d | 10 |
| 4d | 10 |
| -80°C, 3 F/T | 3d | 10 |
| 4d | 10 |
| 103d# | 10 |
*Positive and negative Samples were tested with UTM, M4RT and ESwab media types
#This condition was tested only for samples in UTM media.
The data supported specimen stability for use with cobas liat Bordetella panel is as follows:
{13}
- NPS specimens (UTM, M4RT and ESwab) are stable for up to 3 days when stored at 2-8°C
- NPS specimens (UTM, M4RT and ESwab) are stable for up to 4 hours when stored at 30°C in both the specimen collection tube and the assay tube.
- NPS specimens (M4RT and ESwab) are stable for up to 3 days when stored at ≤-70°C with three freeze thaw cycles.
- NPS specimens (UTM) are stable for up to 4 days when stored at ≤-70°C with three freeze thaw cycles.
## ii. Assay Controls
The cobas liat Bordetella panel nucleic acid test for use on the cobas Liat System contains two sets of controls: an internal process control and an external assay quality control. The Internal Process Control (IPC) monitors the performance of the cobas Bordetella panel assay sample processing and PCR amplification/detection. The external positive and negative controls monitor the assay process and potential reagent/assay tube failure.
## 6. Detection Limit
The limit of detection of the cobas liat Bordetella panel nucleic acid test was determined by analyzing a dilution series of two representative strains of *B. pertussis* (Strain A639 and E431), *B. parapertussis* (Strain E838 and A747), and *B. holmesii* (Strain F061 and ATCC 51541). The culture stocks were co-formulated to an intermediate level and serially diluted in pooled negative nasopharyngeal swab (NPS) clinical specimens collected in UTM to seven (7) concentration levels. All levels were tested with at least 20 replicates per concentration tested across three unique lots of reagents, over the course of three days. A summary of the LoD as determined by the hit rate for each specimen type, lot and strain is shown below in Table 7. The highest concentration detected at ≥ 95% from testing of the two strains across three lots of the assay is the final LoD and is highlighted in the below table.
Table 7: Limit of Detection Reported Based on 95% Hit Rate
| Target | Strain | Lot 1 | Lot 2 | Lot 3 |
| --- | --- | --- | --- | --- |
| | | CFU/mL | | |
| Bp | A639 | 43 | 43 | 43 |
| | E431 | 18 | 36 | 36 |
| Bpp | E838 | 32 | 32 | 32 |
| | A747 | 9 | 36 | 18 |
| Bh | F061 | 5 | 34 | 9 |
| | 51541 | 21 | 21 | 11 |
To demonstrate that presence of all three *Bordetella* strains did not impact the LoD for each individual strain, the LoD for single analyte samples were compared to the co-formulated samples containing all on-panel *Bordetella* strains. The LoDs were considered equal if the single LoD and co-formulated LoD was no more than 3-fold different between results. Co-formulated samples were derived from culture stocks and spiked into clinical negative nasopharyngeal swab matrix. Three dilution levels were prepared for each co-formulated panel, one above, one below and the LoD set at the hit rate of ≥ 95%. Twenty replicates of
K243753 - Page 13 of 19
{14}
each concentration were tested using one lot of the cobas liat Bordetella assay tubes, across 21 cobas liat analyzers. The results of this analysis demonstrate that the co-formulated LoD is comparable to the single LoD (see Table 8).
Table 8: Limit of Detection Comparison for Single and Co-formulated Panels
| Analyte | Target Concentration (CFU/mL) | Formulation | |
| --- | --- | --- | --- |
| | | Co-formulated | Single |
| | | Hit Rate | Hit Rate |
| Bp | 126 | 100% | 100% |
| | 63 | 85% | 100% |
| | 21 | 60% | 65% |
| Bpp | 120 | 100% | 100% |
| | 60 | 100% | 100% |
| | 20 | 75% | 90% |
| Bh | 30 | 100% | 100% |
| | 15 | 100% | 100% |
| | 5 | 95% | 85% |
The results of the study met the acceptance criteria, verifying that the LoD with co-formulated target panels is comparable to single formulated target panels when tested with the cobas liat Bordetella panel. The results demonstrate that co-formulated samples are representative of single-formulated samples with respect to assay performance. Further analytical studies were conducted using co-formulated samples, unless otherwise specified.
# 7. Analytical Reactivity:
An inclusivity study was conducted using various strains of Bp, Bpp and Bh by diluting these cultures into negative clinical matrix pool using one reagent lot. Three replicates were tested for each dilution level for each strain. The strains tested and the lowest concentration detected are listed in Table 9.
Table 9: Inclusivity Testing Result Summary
| Analyte | Strains | Detection Level (CFU/mL) |
| --- | --- | --- |
| Bp | ATCC 8467 | 0.12 |
| | ATCC BAA-589 | 0.006 |
| | ATCC 51445 | 0.12 |
| | ATCC 53894 | 0.05 |
| | ATCC 9340 | 0.04 |
| | ATCC 10380 | 0.48 |
| | ATCC BAA-1335 | 0.3 |
| | ATCC 9306 | 0.3 |
K243753 - Page 14 of 19
{15}
| Analyte | Strains | Detection Level (CFU/mL) |
| --- | --- | --- |
| Bpp | ATCC 12743 | 0.12 |
| | ATCC 9797 | 0.12 |
| | ATCC 12742 | 5.0 |
| | ATCC 8478 | 40.0 |
| Bh | ATCC 15311 | 2.5 |
| | E595 | 40.0 |
| | ATCC 15237 | 10.0 |
| | ATCC BAA-587 | 15.0 |
| | ATCC 9305 | 11.9 |
| | C510 | 47.6 |
| Bh | ATCC 700053 | 1.75 |
| | ATCC 700052 | 0.48 |
| | C690 | 12.5 |
8. Assay Cut-Off:
The assay cutoffs for Bp, Bpp and Bh and internal control were initially evaluated using past remnant samples from other clinical study samples, contrived cultures into clinical negative nasopharyngeal matrix, and neat UTM. This data (3,843 valid results, including 1,350 neat UTM samples) were used to confirm adequate assay cut-off and separation between the latest Ct value observed in the analyte positive specimens and negative samples. Assay cut-off values were verified during clinical performance evaluation.
The final Ct cutoffs were set at 40 for Bp, 34.8 for Bpp and 38 for Bh.
9. Accuracy (Instrument):
N/A
10. Carry-Over:
A carry-over study evaluating the cobas liat system was conducted with the first assay cleared for use with the cobas liat system (K153544) and demonstrated that there was no carry over or cross-contamination observed. The cobas liat Bordetella panel utilizes the same instrumentation. Therefore, additional no carry-over studies were conducted in this submission.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not Applicable
2. Matrix Comparison:
Not Applicable
K243753 - Page 15 of 19
{16}
# C Clinical Studies:
The clinical performance of the cobas liat Bordetella panel was evaluated in a multi-site prospective study in the U.S. between July 2023 - February 2024. The study enrolled 823 subjects suspected of pertussis respiratory infection and nasopharyngeal specimens were prospectively collected from these subjects presenting to point-of-care settings (e.g., emergency rooms, outpatient clinics, etc.). Out of 823 subjects, 54 subjects were excluded due to protocol deviations or indeterminate or invalid comparator results. Twenty-six operators representative of CLIA-waived users at eight different external study sites were involved in sample collection and testing with the cobas liat Bordetella panel. The demographic summary of the prospective clinical subjects is shown in Table 10 below.
Table 10: Demographics of Evaluable Subjects from Prospective Clinical Study
| Age Category | No. of Specimens | Percentage |
| --- | --- | --- |
| < 1 | 27 | 3.51% |
| 1 to <5 | 89 | 11.57% |
| 5 to <12 | 69 | 8.97% |
| 12 to <18 | 32 | 4.16% |
| 18 to <40 | 219 | 28.48% |
| 40 to <65 | 270 | 35.11% |
| >=65 | 63 | 8.19% |
| Sex at Birth | No. of specimens | Percentage |
| Male | 338 | 43.95% |
| Female | 431 | 56.05% |
| Total | 769 | 100% |
Archived clinical specimens and contrived samples were also evaluated to establish device performance due to the very low prevalence observed in the prospective clinical study. The archived positive specimens previously collected from symptomatic patients were characterized using an FDA-cleared NAAT (Nucleic Acid Amplification Test) and included in the study along with additional negative archived specimens to avoid potential bias. Contrived samples were prepared in pooled negative clinical matrix by spiking $2\mathrm{x}$ , $3\mathrm{x}$ , $5\mathrm{x}$ , $10\mathrm{x}$ and $20\mathrm{x}$ LoD of each of the target. All archived and contrived specimens were randomized and distributed to the study site for testing on the cobas liat Bordetella panel. Prospective and archived specimens were then sent to a laboratory for comparator method testing per respective IFU and if required a validated sequencing method was preformed to confirm the comparator result.
K243753 - Page 16 of 19
{17}
The clinical performance of the cobas liat Bordetella panel was assessed by comparing results to FDA-cleared target-specific NAAT. For Bp, the reference method was a composite of an FDA cleared NAAT and a validated bi-directional sequencing method (PCR-BDS) to confirm the presence of Bp. For Bpp and Bh, NAAT was used as the reference comparator method.
The clinical performance of the cobas liat Bordetella panel in terms of Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) versus the comparator reference method is shown in Table 11 below. For prospective and archived specimens, the reference method is the respective comparator results and for contrived specimens the reference method is the expected result.
Table 11: Overall Clinical Performance Summary of cobas liat Bordetella Panel
| Target | Specimen Type | Total (N) | PPA | PPA 95% CI | NPA | NPA 95% CI |
| --- | --- | --- | --- | --- | --- | --- |
| Bp | Prospective | 743 | NC | NC | 100.0% (743/743) | 99.5 – 100% |
| | Archived | 160 | 100% (42/42) | 91.6 – 100% | 99.2% (117/118) | 95.4 – 99.9% |
| | Contrived | 327 | 98.8% (80/81) | 93.3 – 99.8% | 99.6% (245/246) | 97.7 – 99.9% |
| | Overall | 1230 | 99.2% (122/123) | 95.5 – 99.9% | 99.8% (1105/1107) | 99.3 - 100% |
| Bpp | Prospective | 743 | 0.0% (0/1) | 0.0 –79.3% | 100.0% (742/742) | 99.5 - 100% |
| | Archived | 170 | 100.0% (28/28) | 87.9 – 100% | 100.0% (142/142) | 97.4 - 100% |
| | Contrived | 327 | 100.0% (108/108) | 96.6 – 100% | 99.5% (218/219) | 97.5 – 99.9% |
| | Overall | 1240 | 99.3% (136/137) | 96.0 – 99.9% | 99.9% (1102/1103) | 99.5 - 100% |
| Bh | Prospective | 702 | NC | NC | 99.9% (701/702) | 99.2 - 100% |
| | Contrived | 328 | 100.0% (139/139) | 97.3 – 100% | 98.9% (187/189) | 96.2 – 99.7% |
| | Overall | 1030 | 100.0% (139/139) | 97.3 – 100% | 99.7% (888/891) | 99.0 – 99.9% |
Abbreviations: NC- Not Calculable; CI: Confidence Interval
The percent agreement of contrived positive clinical specimens included in the study are provided in Table 12 by strain and LoD level for Bp, Bpp and Bh targets.
K243753 - Page 17 of 19
{18}
Table 12: Contrived Specimens Percent Agreement for Bp, Bpp and Bh
| Strain | LoD | BP Target Percent Agreement (n / N) | BPP Target Percent Agreement (n / N) | BH Target Percent Agreement (n / N) |
| --- | --- | --- | --- | --- |
| A | 2x | 100.0% (4/4) | 100.0% (5/5) | 100.0% (9/9) |
| A | 3x | 100.0% (4/4) | 100.0% (5/5) | 100.0% (9/9) |
| A | 5x | 100.0% (3/3) | 100.0% (3/3) | 100.0% (8/8) |
| A | 10x | 100.0% (2/2) | 100.0% (3/3) | 100.0% (5/5) |
| A | 20x | 100.0% (1/1) | 100.0% (2/2) | 100.0% (4/4) |
| B | 2x | 100.0% (4/4) | 100.0% (5/5) | 100.0% (9/9) |
| B | 3x | 75.0% (3/4) | 100.0% (4/4) | 100.0% (9/9) |
| B | 5x | 100.0% (3/3) | 100.0% (3/3) | 100.0% (8/8) |
| B | 10x | 100.0% (2/2) | 100.0% (3/3) | 100.0% (5/5) |
| B | 20x | 100.0% (1/1) | 100.0% (2/2) | 100.0% (4/4) |
| C | 2x | 100.0% (4/4) | 100.0% (5/5) | 100.0% (9/9) |
| C | 3x | 100.0% (4/4) | 100.0% (5/5) | 100.0% (9/9) |
| C | 5x | 100.0% (3/3) | 100.0% (3/3) | 100.0% (8/8) |
| C | 10x | 100.0% (2/2) | 100.0% (3/3) | 100.0% (5/5) |
| C | 20x | 100.0% (1/1) | 100.0% (2/2) | 100.0% (4/4) |
| D | 2x | 100.0% (4/4) | 100.0% (5/5) | 100.0% (9/9) |
| D | 3x | 100.0% (4/4) | 100.0% (6/6) | 100.0% (9/9) |
| D | 5x | 100.0% (3/3) | 100.0% (3/3) | 100.0% (8/8) |
| D | 10x | 100.0% (2/2) | 100.0% (3/3) | 100.0% (5/5) |
| D | 20x | 100.0% (1/1) | 100.0% (2/2) | 100.0% (4/4) |
| E | 2x | 100.0% (4/4) | 100.0% (5/5) | N/A |
| E | 3x | 100.0% (3/3) | 100.0% (5/5) | N/A |
| E | 5x | 100.0% (3/3) | 100.0% (3/3) | N/A |
| E | 10x | 100.0% (2/2) | 100.0% (3/3) | N/A |
| E | 20x | 100.0% (1/1) | 100.0% (2/2) | N/A |
| F | 2x | 100.0% (4/4) | 100.0% (5/5) | N/A |
| F | 3x | 100.0% (3/3) | 100.0% (5/5) | N/A |
| F | 5x | 100.0% (3/3) | 100.0% (3/3) | N/A |
| F | 10x | 100.0% (1/1) | 100.0% (3/3) | N/A |
| F | 20x | 100.0% (1/1) | 100.0% (2/2) | N/A |
K243753 - Page 18 of 19
{19}
| Strain | LoD | BP Target Percent Agreement (n / N) | BPP Target Percent Agreement (n / N) | BH Target Percent Agreement (n / N) |
| --- | --- | --- | --- | --- |
| Positive Percent Agreement (95% CI) | | 98.8% (80/81: 93.3%, 99.8%) | 100.0% (108/108: 96.6%, 100.0%) | 100.0% (139/139: 97.3%, 100.0%) |
Note: n is the number of positive results and N is the number of valid results.
Strains used for BP contrived specimen: A=A639, B=E431, C=ATCC 51445, D=ATCC 9797, E=ATCC 8467, F=ATCC 9306; for BPP: A=E838, B=A747, C=ATCC 15311, D=ATCC 15237, E=BAA-587, F=E595; for BH: A=F061, B=ATCC 51541, C=ATCC 700053, D=ATCC 700052.
## D Clinical Cut-Off:
Not Applicable
## E Expected Values/Reference Range:
The observed positivity for Bordetella in the prospective population during the cobas liat Bordetella panel clinical studies conducted between July 2023 – February 2024 is 0% (0/1) for Bordetella parapertussis and not calculable for Bordetella pertussis and Bordetella holmesii due to the lack of any positive prospective samples.
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K243753 - Page 19 of 19
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.