The Mako System is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. The Mako System is indicated for use in total hip arthroplasty procedures in which: - the use of stereotactic surgery may be appropriate, and - where reference to rigid anatomical bony structures can be identified relative to a CT-based model of the anatomy; including revision total hip arthroplasty procedures where: - the primary acetabular component is well-fixed to the pelvic anatomy, and - the acetabular component can be revised with either a cup or cup and wedge augment construct. These procedures include: Total Hip Arthroplasty (THA)
Device Story
Mako System is a stereotactic instrument for orthopedic surgery; includes robotic arm, optical detector, camera, computer, and planning laptop. Inputs: patient CT data. Operation: surgeon uses CT-based model for pre-surgical planning and intraoperative navigation; robotic arm acts as intelligent tool holder/guide; provides haptic guidance by constraining tool movement within software-defined spatial boundaries. Output: haptic feedback and visual navigation for implant placement. Used in OR by surgeons. Benefits: assists in precise implant placement and anatomical preparation. Subject device adds revision THA capabilities, including screw/augment planning, pelvic registration for revision, and compatibility with Mako 4 hardware platform.
Clinical Evidence
No clinical data. Evidence consists of non-clinical performance testing: software testing, bone registration accuracy verification, bone resection accuracy verification, system verification, and simulated-use cadaveric surgeon validation.
Technological Characteristics
Stereotactic instrument; robotic arm with haptic guidance; optical tracking; CT-based imaging. Components: Stryker Q-Guidance System, robotic arm, cutting system, planning laptop. Software: CRISIS robotic arm software, Treatment Design Application (TDA) v1.10, Flex OS. Connectivity: Networked system components. Sterilization: Reusable and disposable instruments.
Indications for Use
Indicated for patients undergoing total hip arthroplasty (THA) or revision THA where stereotactic surgery is appropriate and rigid anatomical bony structures can be identified relative to a CT-based model. Revision indication requires the primary acetabular component to be well-fixed to the pelvic anatomy.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. The logo is simple and professional, and it is easily recognizable.
March 5, 2025
Mako Surgical Corp. Rita Koremblum Senior Staff Regulatory Affairs Specialist 3365 Enterprise Avenue Weston, Florida 33331
Re: K243751
Trade/Device Name: Mako Total Hip Application 5.0 Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: December 4, 2024 Received: December 5, 2024
Dear Rita Koremblum:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
{3}------------------------------------------------
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shumaya Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K243751
Device Name Mako Total Hip Application 5.0
Indications for Use (Describe)
The Mako System is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
The Mako System is indicated for use in total hip arthroplasty procedures in which:
- · the use of stereotactic surgery may be appropriate, and
- · where reference to rigid anatomical bony structures can be identified relative to a CT-based model of the anatomy;
including revision total hip arthroplasty procedures where:
- · the primary acetabular component is well-fixed to the pelvic anatomy, and
- · the acetabular component can be revised with either a cup or cup and wedge augment construct.
These procedures include: Total Hip Arthroplasty (THA)
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## K243751 510(k) SUMMARY
| Sponsor: | Mako Surgical Corp.<br>3365 Enterprise Ave<br>Weston, FL 33331 |
|------------------------|------------------------------------------------------------------------------------------------------------|
| Contact Person: | Rita Koremblum<br>Senior Staff Regulatory Affairs Specialist<br>rita.koremblum@stryker.com<br>201-565-6636 |
| Date Prepared: | |
| Proprietary Name: | Mako Total Hip Application |
| Common Name: | Total Hip Application (THA) |
| Regulation Name: | Stereotaxic Instrument |
| Regulation Number: | 21 CFR Section 882.4560 |
| Device Classification: | Class II |
| Product Code: | OLO |
## Substantial Equivalence Claimed To:
The subject device, the Mako Total Hip Application, is substantially equivalent to the predicate device, Mako Total Hip Application, cleared via K193128.
## Device Modifications:
The following changes have been made to the Mako Total Hip Application:
Indications For Use: A revision indication has been added to the Mako Total Hip Application Indications for Use statement.
Integration with the updated Mako 4 Hardware: The subject Mako Total Hip software application has been integrated with the new Mako 4 hardware platform, recently cleared in K241011.
## Clinical Software Application:
- New implant system compatibility for revision indication .
{6}------------------------------------------------
- Screw planning and intraoperative screw trajectory
- Augment planning, preparation and plane capture
- . Additional cup, augment and stem planning features
- Pelvic registration for revision indication
- Bone mapping
- . Pre-impaction and post-impaction clocking
## Additional Software Changes:
- Update to robotic arm software CRISIS to enable compatibility with the updated . Mako 4 hardware operating system (Flex OS).
- MGo: Minor updates to MGo were required for compatibility with the Flex 0S. .
- Treatment Design Application (TDA): A newer version (1.10) of TDA has been . released that supports segmentation of Total Hip Application (THA) revision cases in TDA. Specifically, for revision cases, primary implant surfaces can be segmented in addition to the pelvis, operative femur, and non-operative femur.
## Description:
The Mako System with the subject Mako Total Hip Application is a stereotactic instrument that includes a robotic arm, an integrated cutting system, an optical detector, camera, a computer, dedicated instrumentation, operating software, a planning laptop, and tools and accessories.
The system's architecture is designed to support total and partial knee procedures and total hip procedures. With application specific hardware and software, the system provides haptic guidance during orthopedic surgical procedures by using patient CT data to assist a surgeon with pre-surgical planning, implant placement and interpretive / intraoperative navigation of the patient's anatomy.
Once configured for a specific application, the Mako robotic-arm can serve as the surgeon's "intelligent" tool holder or tool guide by passively constraining the preparation of an anatomical site for an orthopedic implant with software-defined spatial boundaries.
## Summary of Technological Characteristics Compared to Predicate Device:
The technological characteristics of the Mako Total Hip Application compared to the predicate device are listed below:
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are black and the background is white. The letters are closely spaced together, and the word is horizontally oriented.
| Technological<br>Characteristics | Mako Total Hip Application | Mako Total Hip Application –<br>K193128 |
|----------------------------------|--------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Major Components | Stryker Q-Guidance System, robotic<br>arm, cutting system, preoperative<br>planning laptop | Robotic Arm, Guidance Module,<br>Camera Stand, cutting system,<br>preoperative planning laptop |
| Tools/accessories | Various reusable and disposable<br>instruments | Various reusable and disposable<br>instruments |
| Image Use | CT | CT |
## Intended Use
The subject device has the same intended use as that specified in the cleared 510(k) premarket notification for the predicate device listed in this 510(k) premarket notification.
## Indications for Use: Mako Total Hip Application
The Mako System is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
The Mako System is indicated for use in total hip arthroplasty procedures in which:
- the use of stereotactic surgery may be appropriate, and
• where reference to rigid anatomical bony structures can be identified relative to a CT-based model of the anatomy;
including revision total hip arthroplasty procedures where:
- the primary acetabular component is well-fixed to the pelvic anatomy, and
- the acetabular component can be revised with either a cup or cup and wedge augment construct.
These procedures include: Total Hip Arthroplasty (THA)
Performance Data – The modified Mako System with the subject Mako Total Hip Application has been evaluated through the following non-clinical performance testing:
- Software testing
- Bone registration accuracy verification
- Bone resection accuracy verification ●
- System verification
- Simulated-Use Cadaveric Surgeon Validation
#### Conclusions:
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are black and the background is white. The word is written in all lowercase letters.
The hardware and software of the subject Mako System with Mako Total Hip Application is similar to the hardware and software of the predicate Mako System with Mako Total Hip Application. The software modifications being made do not impact the intended use or the fundamental technology of the device. Furthermore, performance testing demonstrates that the characteristics of the subject Mako Total Hip Application are equivalent to the characteristics of the predicate device. The subject device is also as safe and as effective as the predicate device and does not raise different questions of safety and effectiveness. Therefore, the performance testing supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.