Retrospective analysis of real-world data; Systematic literature review
The sponsor utilized existing real-world data and literature reviews from the primary predicate device (K231926) to demonstrate the safety and effectiveness of the device for the adjunctive treatment of MDD in adolescent patients (15-21).
Adolescent MDD; Retrospective analysis; Literature review; Adjunctive treatment
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective analysis of real-world data and systematic literature review (referenced from K231926); Retrospective analysis and systematic literature review
Adolescent patients (15-21) suffering from MDD
Antidepressant therapy alone
Reduction in depression; safety and tolerability
Indications for Use
The Apollo TMS Therapy System is indicated for the treatment of major depressive disorder (MDD) in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode and as an adjunct for the treatment of MDD in adolescent patients (age 15-21).
Device Story
Electromagnetic device delivering non-invasive, rapidly pulsed magnetic fields to cerebral cortex; activates neurons without inducing seizure. Components: user interface, main unit, stimulation coil, coil positioning system. Operator uses 'Stimware' software to manage treatment protocols and individual stimulation doses based on motor threshold. Coil positioned to left dorsolateral prefrontal cortex (DLPFC). Used in clinical settings by trained operators. Output: magnetic pulses; affects clinical decision-making by providing non-pharmacological treatment for MDD. Benefits: potential reduction in depression symptoms for patients unresponsive to medication.
Clinical Evidence
No new clinical data generated. Safety and effectiveness for adolescent MDD adjunct treatment supported by retrospective analysis of real-world data and systematic literature review provided for primary predicate (K231926).
Technological Characteristics
Electromagnetic stimulation system; figure-of-eight coil (pCool/aCool); biphasic waveform. Standards: ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62304, ISO 14971, IEC 81001-5-1. Coil positioning via landmark-aided placement or TMS cap. Software-controlled stimulation parameters (10 Hz or iTBS protocols).
Indications for Use
Indicated for treatment of Major Depressive Disorder (MDD) in adults failing prior antidepressant therapy and as an adjunct for MDD in adolescents (age 15-21).
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
{0}
Funded by U.S. Food & Drug Administration
Build Correspondence
Convert to PDF
September 4, 2025
MAG & More GmbH
Juliane Rieß
Regulatory Affairs Manager
Machtlfinger Straße 13
Munich, 81379
Germany
Re: K243700
Trade/Device Name: Apollo TMS Therapy System
Regulation Number: 21 CFR 882.5805
Regulation Name: Repetitive transcranial magnetic stimulation system
Regulatory Class: Class II
Product Code: OBP
Dated: November 29, 2024
Received: November 29, 2024
Dear Juliane Rieß:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243700 - Juliane Rieß
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243700 - Juliane Rieß
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
PAMELA D. SCOTT -S
2025.09.04 16:54:24
-04'00'
Pamela D. Scott
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
Device Name
Apollo TMS Therapy System
Indications for Use (Describe)
The Apollo TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode and as an adjunct for the treatment of MDD in adolescent patients (age 15-21).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(k) Summary
Apollo TMS Therapy System
MAG & MORE
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
| 510(k) Owner: | MAG & More GmbH
Machtlfinger Straße 13
81379 Munich, Germany
Phone: +49 (0)89 998 292 300
Fax: +49 (0)89 998 292 330 |
| --- | --- |
| Primary Contact: | Juliane Rieß |
| Date Prepared: | 03-Jul-2025 |
| Proprietary Name: | Apollo TMS Therapy System |
| Common/Usual Name: | Repetitive Transcranial Magnetic Stimulation (rTMS) System |
| Classification Name: | Repetitive Transcranial Magnetic Stimulation Device
21 CFR 882.5805, Product Code OBP |
| Predicate Device: | NeuroStar Advanced Therapy System (K231926)
Apollo TMS Therapy System (K180313, K232639) |
# Intended Use
The Apollo TMS Therapy System is indicated for the treatment of major depressive disorder (MDD) in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode and as an adjunct for the treatment of MDD in adolescent patients (age 15-21).
# Device Description and Function
The Apollo TMS Therapy System is an electromagnetic device that non-invasively delivers a rapidly pulsed magnetic field to the cerebral cortex in order to activate neurons within a limited volume without inducing a seizure. This method of cortical stimulation by application of brief magnetic pulses to the head is known as Transcranial Magnetic Stimulation (TMS).
The Apollo TMS Therapy System is comprised of the following principal components:
- User Interface
- Main Unit (with or without housing)
- Stimulation Coil
- Coil Positioning System
510(k) Summary
Page 1 of 8
{5}
510(k) Summary
Apollo TMS Therapy System
MAG & MORE
The operator controls the Apollo TMS Therapy System via the user interface (application software "Stimware"). "Stimware" is a treatment and data management software that administers treatment protocols and the patient's individual stimulation dose determined by the patient's individual motor threshold. Stimulation is applied via the stimulation coil. For the treatment of major depressive disorder, the stimulation coil is positioned to the left dorsolateral prefrontal cortex (DLPFC) by means of the coil positioning system.
## Performance Standards
The Apollo TMS Therapy System conforms to the following recognized consensus standards:
- ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-6 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 62304 Medical device software - Software life cycle processes
## Further Applicable Standards
The Apollo TMS Therapy System complies with applicable requirements of the following additional standards:
- ISO 14971 Medical devices - Application of risk management to medical devices
- ISO 15223 Medical devices - Symbols to be used with medical device labels, labeling and information to be supplied - Part 1: General requirements
- IEC 81001-5-1 Health software and health IT systems safety, effectiveness and security - Part 5-1: Security - Activities in the product life cycle
## Non-Clinical and Clinical Performance Data
Apollo TMS Therapy System and its technological characteristics remain identical to that cleared within K180313 and K232639. No hardware changes have been made to the subject device, and no change to the treatment protocols was made.
The contents of this 510(k) complies with the FDA Guidance Document: "Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems - Guidance for Industry and Food and Drug Administration Staff." Non-clinical performance testing was performed according to the standards listed above, including electrical safety, electromagnetic compatibility and biocompatibility and already cleared by the FDA earlier in K180313 and K232639. To establish substantial equivalence between the predicate device and the subject device comparative performance testing was carried out. The data provided in this submission, such as the comparison of the electric and magnetic field output of the device, demonstrate that expanding the indications to include the adjunct treatment of major depressive disorder (MDD) in adolescent patients (aged 15-
510(k) Summary
{6}
510(k) Summary
Apollo TMS Therapy System
MAG & MORE
21) does not raise any different questions regarding safety and effectiveness. The effectiveness and safety of the adjunctive treatment of adolescent patients (15-21) suffering from MDD has been demonstrated for the primary predicate device (K231926) by a large-scale retrospective analysis of real-world data and a systematic literature review. The data demonstrated a substantially equivalent treatment effect of TMS therapy as an adjunct to antidepressant therapy over antidepressant therapy alone in reducing depression in adolescents. Furthermore, The authors of the studies included in the literature review concluded that TMS is well tolerated and safe for adolescents. No additional clinical data is necessary to demonstrate safety and effectiveness.
## Software Verification and Validation Testing
Software verification and validation testing was conducted and documented in accordance with IEC 62304 and internal quality procedures considered a well-established method as already used in K180313 and K232639. The documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." No software changes have been made to the subject device.
## Risk Management
Risk assessment is applied throughout the product development lifecycle process in accordance with the requirements set forth under the Agency recognized consensus standards ISO 14971 and IEC 62304. The results of the comprehensive risk analysis for the device indicate that there are no new hazards, harms, or safety risks introduced when compared to the predicate devices.
The proposed changes for the subject device for the adjunctive treatment of adolescent patients (15-21) suffering from major depressive disorder (MDD) are limited to updates to the IFU. These changes have been made to enable the safe and effective treatment of adolescent patients (15-21) suffering from MDD and do not introduce any new safety or effectiveness considerations.
## Substantial Equivalence
The Apollo TMS Therapy System described here is identical to the one already cleared in K180313 and K232639 for the treatment of major depressive disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
The Apollo TMS Therapy System is substantially equivalent to its predicate devices, the NeuroStar Advanced Therapy System (K231926) and Apollo TMS Therapy System (K180313, K232639). The intended use and indications for use for the Apollo TMS Therapy System and predicate devices are equivalent. The Apollo TMS Therapy System and the secondary predicate device have identical system components, consisting of a main unit (available in two variants 918001 Apollo and 918010 Apollo Light) including the stimulator, a stimulation coil, a coil positioning system, and the application software. The basic operational procedure is identical, and consists of system setup, patient preparation, coil positioning, determination of patient's motor threshold, and administration of treatment at predefined treatment stimulation parameters. Technological characteristics of the subject device are substantially equivalent to the primary predicate device NeuroStar Advanced Therapy System (K231926).
510(k) Summary
Page 3 of 8
{7}
510(k) Summary
Apollo TMS Therapy System
MAG & MORE
# Conclusions
The indication for use, the target population, the treatment procedure, and the treatment spot are identical for the Apollo TMS Therapy System and the predicate devices NeuroStar Advanced Therapy System (K231926) and Apollo TMS Therapy System (K180313, K232639).
The comparative non-clinical performance testing, such as the comparison of the electric and magnetic field output of the device, demonstrate that the Apollo TMS Therapy System is substantially equivalent to the predicate devices. All identified differences between the subject device, the Apollo TMS Therapy System, and its primary predicate device are minor and do not raise different questions of safety or effectiveness. This demonstrates that the Apollo TMS Therapy System is as safe and effective for use as an adjunct for treatment of adolescent patients (15-21) suffering from major depressive disorder as the primary predicate device.
Therefore, the Apollo TMS Therapy System is substantially equivalent to its predicate devices NeuroStar Advanced Therapy System (K231926) and Apollo TMS Therapy System (K180313, K232639).
510(k) Summary
Page 4 of 8
{8}
510(k) Summary
Apollo TMS Therapy System
MAG & MORE
Substantial Equivalence Comparison
| Characteristic | Subject Device | Predicate Device | Secondary Predicate |
| --- | --- | --- | --- |
| 510(k) Number | N/A | K231926 | K180313, K232639 |
| Device Trade Name | Apollo TMS Therapy System | NeuroStar Advanced Therapy System | Apollo TMS Therapy System |
| 510(k) submitter | MAG & More GmbH
Machtlfinger Straße 13
81379 Munich, Germany | Neuronetics, Inc.
3222 Phoenixville Pike
Malvern, PA 19355 | MAG & More GmbH
Machtlfinger Straße 13
81379 Munich, Germany |
| Indications for use | The Apollo TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode and as an adjunct for the treatment of MDD in adolescent patients (age 15-21). | NeuroStar Advanced Therapy is indicated as an adjunct for the treatment of Major Depressive Disorder (MDD) in adolescent patients (15-21). | The Apollo TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode. |
| Target population | Adult and Adolescent patients age 15-21 | Adolescent patients age 15-21 | Adult patients |
| Product Code | OBP | OBP | OBP |
| Classification | 21 CFR 882.5805 | 21 CFR 882.5805 | 21 CFR 882.5805 |
| Standard Treatment Stimulation Parameters | | | |
| Area of brain to be stimulated | Left DLPFC | Left DLPFC | Left DLPFC |
510(k) Summary
{9}
510(k) Summary
Apollo TMS Therapy System
MAG & MORE
| Characteristic | Subject Device | Predicate Device | Secondary Predicate |
| --- | --- | --- | --- |
| Stimulation intensity | 120% MT with allowable adjustments | 120% MT with allowable adjustments | 120% of MT |
| %MT range | 50% to 150% MT | 25% to 140% MT | 50% to 150% MT |
| Stimulation frequency | 10 Hz | 10 Hz | 10 Hz |
| Pulse train duration | 4 sec | 4 sec | 4 sec |
| Inter-train interval | 11 – 26 sec | 11 – 26 sec | 11 – 26 sec |
| Trains per session | 75 | 75 | 75 |
| Max No. of Pulses | 3,000 | 3,000 | 3,000 |
| **Stimulation parameter ranges** | | | |
| Amplitude in SMT units | 0 – 2.0 SMT | 0.22 – 2.08 SMT | 0 – 2.0 SMT |
| Pulse width (± accuracy) | pCool coils: 162 μs ± 1.9 μs
aCool coils: 163 μs ± 2.5 μs | 185 μs | pCool coils: 162 μs ± 1.9 μs
aCool coils: 163 μs ± 2.5 μs |
| Pulse train duration (sec) | 0.04 - 2000 | 1 - 600 | 0.04 - 2000 |
| ITI range (sec) | 1 - 65 | 0 - 600 | 1 - 65 |
| Max # of pulses per session | 75,000 | 5,000 | 75,000 |
510(k) Summary
Page 6 of 8
{10}
510(k) Summary
Apollo TMS Therapy System
MAG & MORE
| Characteristic | Subject Device | | Predicate Device | Secondary Predicate | |
| --- | --- | --- | --- | --- | --- |
| Stimulation Coil Parameters | | | | | |
| Coil | pCool coils: HANS-pCool PMD70-pCool | aCool coils: HANS-aCool PMD70-aCool | Ferromagnetic Coil | pCool coils: HANS-pCool PMD70-pCool | aCool coils: HANS-aCool PMD70-aCool |
| Configuration | figure-of-eight coil | | figure-of-eight coil | figure-of-eight coil | |
| Output waveform | biphasic | | biphasic | biphasic | |
| Cooling | passive cooling | active air-cooling | air | passive cooling | active air-cooling |
| E-Field at 1.0 SMT | 130 V/m | | 135 V/m | 130 V/m | |
| Coil Positioning System | Integrated into Head-and-Neck support system, landmark-aided coil placement or TMS cap for standardized 10-20-EEG positioning with coil positioning arm | | Integrated into Head Support System Laser-aided coil placement | Integrated into Head-and-Neck support system, landmark-aided coil placement or TMS cap for standardized 10-20-EEG positioning with coil positioning arm | |
| iTBS Treatment Protocol | | | | | |
| Area of brain to be stimulated | Left DLPFC | | Left DLPFC | Left DLPFC | |
| Stimulation Intensity | 80 - 120% MT with allowable adjustments | | 80 - 120% MT with allowable adjustments | 120% of the MT | |
| Repetition Rate | 50 Hz (5 pulses per sec) | | 50 Hz (5 pulses per sec) | 50 Hz (5 pulses per sec) | |
| Train Duration | 2 sec | | 2 sec | 2 sec | |
| Inter-train Interval | 8 sec | | 8 sec | 8 sec | |
510(k) Summary
{11}
510(k) Summary
Apollo TMS Therapy System
MAG & MORE
| Characteristic | Subject Device | Predicate Device | Secondary Predicate |
| --- | --- | --- | --- |
| Burst Pulses | 3 | 3 | 3 |
| Bursts | 200 | 200 | 200 |
| Inter Pulse interval | 20 ms | 20 ms | 20 ms |
| Number of trains | 20 | 20 | 20 |
| Number of Pulses per Session | 600 | 600 | 600 |
NOTE: Only the standard and iTBS treatment protocol parameters have been cleared. Other stimulation parameters have not been cleared or evaluated for safety and effectiveness.
510(k) Summary
Page 8 of 8
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.