Maxx Libertas Bipolar Hip Head (Bipolar Hip)

K243634 · Maxx Orthopedics, Inc. · KWY · Aug 25, 2025 · Orthopedic

Device Facts

Record IDK243634
Device NameMaxx Libertas Bipolar Hip Head (Bipolar Hip)
ApplicantMaxx Orthopedics, Inc.
Product CodeKWY · Orthopedic
Decision DateAug 25, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3390
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Maxx Libertas Bipolar Head is intended for use in combination with a Maxx Libertas Hip System femoral stem for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate: • Femoral neck and trochanteric fractures of the proximal femur; • Osteonecrosis of the femoral head; • Revision procedures where other devices or treatments for these indications have failed.

Device Story

Maxx Libertas Bipolar Hip Head is a modular orthopedic implant for hemiarthroplasty. Device consists of cobalt chromium outer shell, UHMWPE liner, and UHMWPE retention ring. Operates as a bipolar prosthesis; outer shell articulates with patient's acetabulum; inner liner articulates with 28mm femoral head of Maxx Libertas femoral stem. Retention ring prevents femoral head dislocation. Used in surgical settings by orthopedic surgeons. Provides stable articulation for hip joint; restores joint function; reduces pain in patients with femoral neck fractures or osteonecrosis.

Clinical Evidence

Bench testing only. Mechanical testing included Range of Motion, static push-out, and static lever-out tests. Engineering analysis performed for impingement scenarios. No clinical data provided.

Technological Characteristics

Cobalt chromium outer shell; UHMWPE liner; UHMWPE retention ring. Bipolar articulation design. Dimensions accommodate 28mm femoral head. Standalone mechanical implant. No software or energy source.

Indications for Use

Indicated for patients requiring primary or revision hemiarthroplasty of the hip due to femoral neck/trochanteric fractures of the proximal femur, osteonecrosis of the femoral head, or failure of previous treatments.

Regulatory Classification

Identification

A hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis is a two-part device intended to be implanted to replace the head and neck of the femur. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a snap-fit acetabular component made of an alloy, such as cobalt-chromium-molybdenum, and ultra-high molecular weight polyethylene. This generic type of device may be fixed to the bone with bone cement (§ 888.3027) or implanted by impaction.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION August 25, 2025 Maxx Orthopedics, Inc. Donald Guthner Director, Regulatory Affairs 2460 General Armistead Ave Suite 100 Norristown, Pennsylvania 19403 Re: K243634 Trade/Device Name: Maxx Libertas Bipolar Hip Head (Bipolar Hip) Regulation Number: 21 CFR 888.3390 Regulation Name: Hip Joint Femoral (Hemi-Hip) Metal/Polymer Cemented Or Uncemented Prosthesis Regulatory Class: Class II Product Code: KWY Dated: July 25, 2025 Received: July 25, 2025 Dear Donald Guthner: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243634 - Donald Guthner Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K243634 - Donald Guthner Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Limin Sun -S Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K243634 | | | Device Name Maxx Libertas Bipolar Hip Head (Bipolar Hip) | | | Indications for Use (Describe) The Maxx Libertas Bipolar Head is intended for use in combination with a Maxx Libertas Hip System femoral stem for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate: • Femoral neck and trochanteric fractures of the proximal femur; • Osteonecrosis of the femoral head; • Revision procedures where other devices or treatments for these indications have failed. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} 510(k) #: K243634 510(k) Summary Prepared on: 2025-08-22 # Contact Details | Applicant Name | Maxx Orthopedics, Inc. | | --- | --- | | Applicant Address | 2460 General Armistead Ave Suite 100 Norristown PA 19403 United States | | Applicant Contact Telephone | 646.460.2984 | | Applicant Contact | Mr. Donald Guthner | | Applicant Contact Email | don.guthner@Maxxortho.com | # Device Name | Device Trade Name | Maxx Libertas Bipolar Hip Head (Bipolar Hip) | | --- | --- | | Common Name | Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented | | Classification Name | Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis | | Regulation Number | 888.3390 | | Product Code(s) | KWY | # Legally Marketed Predicate Devices | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | --- | --- | --- | | K100761 | BioPro Bipolar Head | KWY | # Device Description Summary | The Maxx Libertas Bipolar Head consists of three factory-assembled parts: a cobalt chromium outer shell, a UHMWPE liner, and a UHMWPE retention ring. The outer shell has a highly polished spherical outer surface for articulating with the acetabular joint socket. The liner has an hemispherical inner surface for articulating with the spherical head component of the Libertas femoral stem (28 mm). The retention ring provides a locking function to resist dislocation of the femoral head from the bipolar head. | | --- | # Intended Use/Indications for Use | The Maxx Libertas Bipolar Head is intended for use in combination with a Maxx Libertas Hip System femoral stem for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate: • Femoral neck and trochanteric fractures of the proximal femur; • Osteonecrosis of the femoral head; • Revision procedures where other devices or treatments for these indications have failed. | | --- | # Indications for Use Comparison | Indications for Use are the same | | --- | # Technological Comparison | Technological Comparison | | --- | {5} Subject device is identical to the Predicate Device in design, material, chemical composition, principle of operation. # Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Mechanical testing was performed - Range of Motion, Static push-out test and Static Lever-out test. Engineering analysis for impingement scenarios was performed. The device performed identical to the predicate device which supports substantial equivalence.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...