← Product Code [HTN](/productcode/HTN) · K243624

# Acu-Sinch Knotless Mini (K243624)

_Acumed, LLC · HTN · Jan 23, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K243624

## Device Facts

- **Applicant:** Acumed, LLC
- **Product Code:** [HTN](/productcode/HTN.md)
- **Decision Date:** Jan 23, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Acu-Sinch Knotless Mini, when used for fixation of bone-to-bone or soft-tissue to bone, is intended as a fixation post, distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, Acu-Sinch Knotless Mini is indicated for Carpal Metacarpal (CMC) joint arthroplasty as an adjunct in the healing process of hematoma distraction arthroplasty by providing stablization at the base of the first and second metacarpal when the trapezium has been excised due to osteoarthritis.

## Device Story

Acu-Sinch Knotless Mini is a bone fixation system consisting of two titanium alloy buttons and a UHMWPE suture strand. Used during trapeziectomy procedures to stabilize the first and second metacarpals. The device acts as a fixation post or distribution bridge to manage suture tension during ligament or tendon repair. It is implanted by a surgeon in an operating room setting. The system provides mechanical stabilization to support the healing process of hematoma distraction arthroplasty. The device is sterile and for single use.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including static and dynamic fatigue testing, MR environment testing (ASTM F2052, F2119, F2182, F2213), endotoxin testing, and shelf-life verification (5-year accelerated and 2-year real-time).

## Technological Characteristics

Buttons: Titanium alloy (Ti-6AL-4V ELI) per ASTM F136. Suture: Ultra High Molecular Weight Polyethylene (UHMWPE). System: Two-button and suture construct. Sterile, single-use. Mechanical fixation device.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Arthrex CMC Mini Tightrope ([K140328](/device/K140328.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 23, 2025

Acumed LLC Marina Bull Regulatory Affairs Specialist 5885 NE Cornelious Pass Hillsboro, Oregon 97124

Re: K243624

Trade/Device Name: Acu-Sinch Knotless Mini Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HTN Dated: November 22, 2024 Received: November 25, 2024

Dear Marina Bull:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# RYAN TROMBETTA -S

For: Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K243624

Device Name

Acu-Sinch Knotless Mini

#### Indications for Use (Describe)

Acu-Sinch Knotless Mini, when used for fixation of bone-to-bone or soft-tissue to bone, is intended as a fixation post, distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, Acu-Sinch Knotless Mini is indicated for Carpal Metacarpal (CMC) joint arthroplasty as an adjunct in the healing process of hematoma distraction arthroplasty by providing stablization at the base of the first and second metacarpal when the trapezium has been excised due to osteoarthritis.

Type of Use (Select one or both, as applicable)

IX.Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE F NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

'en agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 1 of 1

{4}------------------------------------------------

| 510(k) #:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | K243624                                                  |                                                                             |  |              | Prepared on: 2024-11-23 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|-----------------------------------------------------------------------------|--|--------------|-------------------------|
| Contact Details                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                          |                                                                             |  |              | 21 CFR 807.92(a)(1)     |
| Applicant Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                          | Acumed LLC                                                                  |  |              |                         |
| Applicant Address                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                          | 5885 NE Cornelious Pass Hillsboro OR 97124 United States                    |  |              |                         |
| Applicant Contact Telephone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                          | 970-846-6282                                                                |  |              |                         |
| Applicant Contact                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                          | Ms. Marina Bull                                                             |  |              |                         |
| Applicant Contact Email                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                          | marina.bull@acumed.net                                                      |  |              |                         |
| Device Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                          |                                                                             |  |              | 21 CFR 807.92(a)(2)     |
| Device Trade Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Acu-Sinch Knotless Mini                                  |                                                                             |  |              |                         |
| Common Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                          | Washer, Bolt, Nut                                                           |  |              |                         |
| Classification Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                          | Single/multiple component metallic bone fixation appliances and accessories |  |              |                         |
| Regulation Number                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                          | 888.3030                                                                    |  |              |                         |
| Product Code(s)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                          | HTN                                                                         |  |              |                         |
| Legally Marketed Predicate Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                          |                                                                             |  |              | 21 CFR 807.92(a)(3)     |
| Predicate#                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Trade Name (Primary Predicate is listed first) |                                                                             |  | Product Code |                         |
| K140328                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Arthrex CMC Min Tightrope                                |                                                                             |  | HTN          |                         |
| Device Description Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                          |                                                                             |  |              | 21 CFR 807.92(a)(4)     |
| The Acu-Sinch Knotless Mini is intended to provide fixation during the healing process of a trapeziectomy. Acu-Sinch Knotless Mini consists of two buttons, a suture strand, and instruments to aid in insertion. The buttons are manufactured from titanium alloy conforming to ASTM FI 36 (Ti-6AL-4V ELI). The suture is manufactured from Ultra High Molecular Weight Polyethylene (UHMWPE). The devices are sterile and for single use.                                                                                              |                                                          |                                                                             |  |              |                         |
| Intended Use/Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                          |                                                                             |  |              | 21 CFR 807.92(a)(5)     |
| Acu-Sinch Knotless Mini, when used for fixation of bone-to-bone or soft-tissue to bone, is intended as a fixation post, distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, Acu-Sinch Knotless Mini is indicated for Carpal Metacarpal (CMC) joint arthroplasty as an adjunct in the healing process of hematoma distraction arthroplasty by providing stablization at the base of the first and second metacarpal when the trapezium has been excised due to osteoarthritis. |                                                          |                                                                             |  |              |                         |
| Indications for Use Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                          |                                                                             |  |              | 21 CFR 807.92(a)(5)     |

{5}------------------------------------------------

The Acu-Sinch Knotless Minj, has been compared to the predicate device, The Arthrex OMC Mini Tightrope within the 51 0(k) submission. The basis of substantial equivalence for the subject device is their similarities in intended use, material technology, operating principles, anatomical site, for implantation, performance and design. The analysis of difference between the subject device and predicate device supports substantial equivalence as the differences do not constitute a new intended use/indications for use, and the information the submission demonstrated that the subject device is comparable to the predicate device and does not raise different questions of safety or effectiveness.

The Acu-Sinch Knotless Mini has the same intended use/indications for use as the Arthrex CMC Mini Tightrope. Both have two buttons and a suture when used for bone or soft-tisue to bone, is intended as a fixation post, distribution bridge, or for distributing suture tension over areas of ligament or tendon repair.

## Technological Comparison

The Subject Device has the same technological characteristics as the Predicate Device and Predicate Devices make use of a suture to provide fixation between the buttons, located on the thumb metacarpal and index metacarpal.

The button and suture system is used for distributing suture tension over areas of ligament or tendon repair or in the suspensory fration of the thumb metacarpal at the base of the second metacarpal when the trapezium is removed.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Static and Dynamic fatigue testing was conducted to demonstrate that the proposed Acu-Sinch Knotless Mini performs statistically equivalent to the Predicate Device; Arthrex CMC Mini Tightrope cleared under KI 40328. Acu-Sinch Knotless Mini performed as well as, or better than the predicate device. " 1 1 1 8

WN force, torque, and image artifact testing were conducted in accordance with ASTM F2052 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment, ASTM F2119 Standard Test Method for Evaluation of In Image Artifacts from Passive Implants, ASTM F2182 Standard Test Method for Measurement of Measurement of Radio Frequency Induced Heating Near Passive Implants During Maging and ASTM F2213 Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment.

Endotoxin testing (LAL and MMP) was conducted and met the requirement of <20 EU/device.

Device performance was verified following 5-year accelerated and 2-vear real time testing and passed.

Clinical testing was not required to support substantial equivalence (Not Applicable).

Based on the results of the nonclinical bench testing described above, it was concluded that the Subject and Predicate Devices are equivalent in performance specifically for the intended use, hence was proven to be safe and effective for the indication.

Page 2 of 2

## 21 CFR 807.92(a)(6)

---

**Source:** [https://fda.innolitics.com/device/K243624](https://fda.innolitics.com/device/K243624)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
