The Nano-Check Influenza+COVID-19 Dual Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab (ANS) samples from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2 and influenza can be similar. All negative results are presumptive and should be confirmed with a molecular assay, if necessary, for patient management. Negative results do not rule out infection with influenza or SARS-CoV-2 and should not be used as the sole basis for treatment or patient management decisions. Positive results do not rule out bacterial infection or co-infection with other viruses.
Device Story
Lateral flow immunochromatographic assay for qualitative detection/differentiation of influenza A/B nucleoprotein and SARS-CoV-2 nucleocapsid antigens. Input: anterior nasal swab (ANS) samples. Operation: specimen extraction into buffer, application to test cassette sample well; viral antigens form complexes with colloidal gold-conjugated monoclonal antibodies; captured by streptavidin or specific anti-Flu A/B antibodies on test lines. Output: visual red lines on test cassette (C/A/B lines) indicating presence of specific antigens; control line (CON) confirms proper wicking. Used in clinical settings by healthcare providers. Results interpreted visually within 15-20 minutes. Negative results are presumptive; positive results do not rule out co-infections. Assists clinicians in differentiating respiratory viral infections to guide patient management.
Clinical Evidence
Prospective multi-center clinical study (n=1,969) at 6 CLIA-waived sites. Compared to FDA-cleared RT-PCR. SARS-CoV-2: 87.6% PPA, 99.8% NPA. Influenza A: 86.9% PPA, 99.6% NPA. Influenza B: 86.8% PPA, 99.7% NPA. Bench testing included precision, reproducibility, cross-reactivity, interference, and LoD studies.
Technological Characteristics
Lateral flow immunochromatographic assay. Components: sample pad, reagent pad (colloidal gold-conjugated monoclonal antibodies), biotin pad (biotin-conjugated mAb), reaction membrane (streptavidin and anti-Flu A/B antibodies), absorbent pad. Energy source: none (passive capillary flow). Storage: 2-30°C. Qualitative visual readout.
Indications for Use
Indicated for qualitative detection and differentiation of influenza A, influenza B, and SARS-CoV-2 antigens in anterior nasal swab samples from symptomatic individuals. For prescription use.
Regulatory Classification
Identification
A multi-analyte respiratory virus antigen detection test is an in vitro diagnostic device intended for the detection and/or differentiation of respiratory viruses directly from respiratory clinical specimens. The device is intended to be performed at the site of sample collection, does not involve sample storage and/or transport.
Predicate Devices
Healgen Rapid Check® COVID-19/Flu A&B Antigen Test (DEN240029)
Submission Summary (Full Text)
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K243561
B Applicant
Nano-Ditech Corporation
C Proprietary and Established Names
Nano-Check Influenza+COVID-19 Dual Test
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| SCA | Class II | 21 CFR 866.3987 - Multi-Analyte Respiratory Virus Antigen Detection Test | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain 510(k) clearance for the Nano-Check Influenza+COVID-19 Dual Test.
B Measurand:
Influenza A and B Nucleoprotein antigens and SARS-CoV-2 nucleocapsid protein antigens
C Type of Test:
Qualitative lateral flow immunoassay
## III Intended Use/Indications for Use:
A Intended Use(s):
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
See Indications for Use below.
## B Indication(s) for Use:
The Nano-Check Influenza+COVID-19 Dual Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab (ANS) samples from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2 and influenza can be similar.
All negative results are presumptive and should be confirmed with a molecular assay, if necessary, for patient management. Negative results do not rule out infection with influenza or SARS-CoV-2 and should not be used as the sole basis for treatment or patient management decisions.
Positive results do not rule out bacterial infection or co-infection with other viruses.
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
IVD - For In Vitro Diagnostic Use Only
## D Special Instrument Requirements:
Not Applicable
## E Device/System Characteristics:
### 1. Device Description:
The Nano-Check Influenza+COVID-19 Dual Test is an in vitro lateral flow immunochromatographic assay intended for rapid, simultaneous qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigens. This device is intended for prescription use only with anterior nasal swab (ANS) samples from individuals with signs and symptoms of respiratory infection within the first four (4) days of symptom onset.
The assay kit consists of 25 test cassette devices, 25 reagent tubes, 25 ampules containing extraction buffer, 25 anterior nasal specimen collection swabs, one positive control swab (contains noninfectious influenza A, influenza B, and SARS-CoV-2 recombinant antigen), one negative control swab (without recombinant antigen), one Instructions for Use (IFU), and one Quick Reference Instruction (QRI).
Test strips are enclosed in a cassette housing comprised of the following components: sample pad, reagent pad, biotin pad, reaction membrane, and absorbent pad. The reagent pad contains colloidal gold conjugated with monoclonal antibodies (mAb) specific for SARS-CoV-2, influenza A, and influenza B target proteins. The biotin pad contains biotin
K243561 - Page 2 of 25
{3}
conjugated with mAb specific for SARS-CoV-2. The reaction membrane contains the secondary antibodies for the proteins of influenza A and influenza B, and streptavidin for the biotinylated SARS-CoV-2 antibody. The whole strip is fixed inside a plastic cassette. The schematic of the test cassette device is shown below.

Fig 1: Test Device of Nano-Check Influenza+COVID-19 Dual Test Strip
## 2. Principle of Operation:
The Nano-Check Influenza+COVID-19 Dual Test is designed to detect the extracted nucleoprotein antigen specific to influenza (Flu) A, and Flu B and the extracted nucleocapsid antigen specific to SARS-CoV-2 in ANS specimens directly collected from patients exhibiting signs or symptoms of a respiratory infection.
When the specimens are extracted and added to the sample well of the test device, Flu A, Flu B nucleoproteins and/or SARS-CoV-2 nucleocapsid antigen is present in the specimen, a complex form between the anti-Flu A/Flu B/ SARS-CoV-2 conjugate and the viral antigen will be captured by the streptavidin or specific anti-Flu A/Flu B mAb coated on the test line region (C/A/B line). The absence of the test line (C/A/B line) is interpreted as a negative result. To serve as a procedural control, a red line will always appear in the control line region (CON) indicating that proper volume of sample has been added and membrane wicking has occurred. Any result without this control line is invalid.
## IV Substantial Equivalence Information:
### A Predicate Device Name(s):
Healgen Rapid Check COVID-19/Flu A&B Antigen Test
### B Predicate 510(k) Number(s):
DEN240029
### C Comparison with Predicate(s):
| Device & Predicate Device(s): | K243561 | Predicate: DEN240029 |
| --- | --- | --- |
| Device Trade Name | Nano-Check Influenza+COVID-19 Dual | Healgen Rapid Check COVID-19/Flu A&B |
K243561 - Page 3 of 25
{4}
K243561 - Page 4 of 25
| Device & Predicate Device(s): | K243561 | Predicate: DEN240029 |
| --- | --- | --- |
| | Test | Antigen Test |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The Nano-Check Influenza+COVID-19 Dual Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab (ANS) samples from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2 and influenza can be similar.
All negative results are presumptive and should be confirmed with a molecular assay, if necessary, for patient management. Negative results do not rule out infection with influenza or SARS-CoV-2 and should not be used as the sole basis for treatment or patient management decisions.
Positive results do not rule out bacterial infection or co-infection with other viruses. | The Healgen Rapid Check COVID-19/Flu A&B Antigen Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar.
This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens. Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up |
{5}
K243561 - Page 5 of 25
| Device & Predicate Device(s): | K243561 | Predicate: DEN240029 |
| --- | --- | --- |
| | | care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens, and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up. |
| Regulation Number | 21 CFR 866.3987 | Same |
| Assay Principle (Technology) | Lateral flow immune chromatographic assay | Same |
| Specimen Type | Anterior nasal swab specimens | Same |
| Analyte | Influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen | Same |
| Usage | Single use test | Same |
| Test Result | Qualitative | Same |
| Result Interpretation | Visually Read | Same |
| Reading Time | 15 – 20 minutes | 15 - 20 minutes |
| Storage Temperature | 2°-30°C | Same |
| General Device Characteristic Differences | | |
| Patient Use Setting | For prescription use | Over the counter use |
V Standards/Guidance Documents Referenced:
| Document Number | Title | Publishing Organization | Purpose |
| --- | --- | --- | --- |
| 21 CFR 866.3987 | Special controls for multi-analyte respiratory virus antigen detection test, an in vitro diagnostic device intended for the detection and/or differentiation of respiratory viruses directly from respiratory clinical specimens. The device is intended to be performed at the site of sample collection, does not involve sample storage and/or transport. | FDA/CDRH | General Use |
| EP12-A2 | User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition. | CLSI | General Use |
{6}
K243561 - Page 6 of 25
| EP05-A3
(Reaffirmed: September 2019) | Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition | CLSI | General Use |
| --- | --- | --- | --- |
| EP37 1st Edition | Supplemental tables for Interference Testing in Clinical Chemistry | CLSI | General Use |
| 14971 Third Edition 2019-12 | Medical Devices – Application of risk management to medical devices | ISO | General Use |
## VI Performance Characteristics (if/when applicable):
### A Analytical Performance:
#### 1. Precision/Reproducibility:
##### a. Within-lab Precision/Single-lot Repeatability Study:
The repeatability study was conducted by two operators with each operator conducting 24 test runs over a 12-day period, resulting in a total of 96 replicates per concentration and negative samples. The study included blind sample panels comprising of true negative (TN), low positive (LP, 1x LoD) and moderate positive (MP, 3x LoD) samples for each analyte (SARS-CoV-2, Flu A and Flu B). The study results are summarized in Table 1 below and shows 100% agreement for all samples.
Table 1: Overall Results from Single-lot Repeatability Study
| Analyte | SARS-CoV-2 | | Flu A | | Flu B | | TN |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample Level | MP (3× LoD) | LP (1× LoD) | MP (3× LoD) | LP (1× LoD) | MP (3× LoD) | LP (1× LoD) | |
| # Correct/Total | 96/96 | 96/96 | 96/96 | 96/96 | 96/96 | 96/96 | 96/96 |
| All operators Total % Agreement | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
| All operators Total 95% CI | 96.2 - 100% | 96.2 - 100% | 96.2 - 100% | 96.2 - 100% | 96.2 - 100% | 96.2 - 100% | 96.2 - 100% |
##### b. Lot-to-lot Precision Study:
The lot-to-lot consistency evaluation was conducted over a 12-day period to assess the precision of test results across different production lots using a contrived sample panel consisting of true negative, a low positive (C90) and a moderate positive sample (3x LoD). The study was conducted by two on-site operators for over 12 non-consecutive days with two replicate tests performed per sample per day per device lot, resulting in a total of 24 independent replicates per sample per device lot. The summary of lot-to-lot precision study results are shown in Table 2 below and are acceptable.
{7}
Table 2: Summary of Between-lot and Between-operator Variability
| Analyte | Test line | Between lot | | | | | | Between operator | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Lot 1 | | Lot 2 | | Lot 3 | | Operator 1 | | Operator 2 | |
| | | Positive | %* | Positive | %* | Positive | %* | Positive | %* | Positive | %* |
| Negative | COVID-19 | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| | Flu A | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| | Flu B | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| C90 COVID-19 | COVID-19 | 20/24 | 83.3% | 23/24 | 95.8% | 21/24 | 87.5% | 31/36 | 86.1% | 33/36 | 91.7% |
| | Flu A | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| | Flu B | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| C90 Flu A | COVID-19 | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| | Flu A | 20/24 | 83.3% | 21/24 | 87.5% | 23/24 | 95.8% | 32/36 | 88.9% | 32/36 | 88.9% |
| | Flu B | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| C90 Flu A | COVID-19 | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| | Flu A | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| | Flu B | 22/24 | 91.7% | 21/24 | 87.5% | 24/24 | 100% | 33/36 | 88.9% | 34/36 | 94.4% |
| 3X LoD COVID-19 | COVID-19 | 24/24 | 100% | 24/24 | 100% | 24/24 | 100% | 36/36 | 100% | 36/36 | 100% |
| | Flu A | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| | Flu B | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| 3X LoD Flu A | COVID-19 | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| | Flu A | 24/24 | 100% | 24/24 | 100% | 24/24 | 100% | 36/36 | 100% | 36/36 | 100% |
| | Flu B | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| 3X LoD Flu B | COVID-19 | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| | Flu A | 0/24 | 100% | 0/24 | 100% | 0/24 | 100% | 0/36 | 100% | 0/36 | 100% |
| | Flu B | 24/24 | 100% | 24/24 | 100% | 24/24 | 100% | 36/36 | 100% | 36/36 | 100% |
*Percent Agreement
# c. Multi-site Reproducibility Study:
A multi-site reproducibility study was performed to assess the performance of the candidate device using a contrived sample panel comprised of a true negative (TN), a high negative sample (HN, 0.1x LoD), a low positive (LP, 1x LoD) and a moderate positive (MP, 5x LoD) sample for each analyte. Contrived swab samples were prepared by spiking pooled human nasal wash solution (confirmed SARS-CoV-2, Flu A, and Flu B negative by RT-PCR) with SARS-CoV-2 (Omicron variant), Flu A (H1N1), and Flu B (Victoria Lineage) to concentrations of 5x LoD, 1x LoD or 0.1x LoD. Each diluted sample $(50~\mu \mathrm{L})$ was directly applied onto the sample collection swab head. True negative swab samples were prepared by applying fifty $(50)$ $\mu \mathrm{L}$ of negative pooled human nasal wash directly onto the sample collection swab head.
The contrived sample swabs were randomized and blinded to each operator at three (3) discrete CLIA-waived sites and an in-house site. Eight (8) operators at the CLIA-waived sites and three (3) additional trained operators at the internal site conducted testing for over five non-consecutive days. Each operator tested 150 encoded samples, consisting of 15 samples each for various analyte concentrations.
The results in Table 3 below showed complete concordance, with $100\%$ agreement (95% CI: 97.7-100%) among the eleven (11) operators for TN HN for COVID-19 and Flu A, LP for COVID-19 and Flu B, as well as all MP samples. The results between HN-Flu B and LP-Flu A exhibited agreement of $99.4\%$ (95% CI: 96.7-99.9%). These outcomes all met the predefined acceptance criteria and generated no significant difference between sites.
K243561 - Page 7 of 25
{8}
Table 3: Summary Results of Multi-site Reproducibility Study
| Sample | No. of Positive Results/No. of Total Tests (Positive Rate %) | | | | All Sites | |
| --- | --- | --- | --- | --- | --- | --- |
| | CLIA-Waived Site 1 (2 operators) | CLIA-Waived Site 2 (3 operators) | CLIA-Waived Site 3 (3 operators) | In-house Site 4 (3 operators) | Agreement | 95% CI |
| True Negative NCM | 0/30 (0%) | 0/45 (0%) | 0/45 (0%) | 0/45 (0%) | 165/165 (100%) | 97.7-100.0 |
| 0.1x LoD SARS CoV-2 | 0/30 (0%) | 0/45 (0%) | 0/45 (0%) | 0/45 (0%) | 165/165 (100%) | 97.7-100.0 |
| 0.1x LoD Flu A | 0/30 (0%) | 0/45 (0%) | 0/45 (0%) | 0/45 (0%) | 165/165 (100%) | 97.7-100.0 |
| 0.1x LoD Flu B | 0/30 (0%) | 0/45 (0%) | 1/45 (2.2%) | 0/45 (0%) | 164/165 (99.4%) | 96.7-99.9 |
| 1x LoD SARS CoV-2 | 30/30 (100%) | 45/45 (100%) | 45/45 (100%) | 45/45 (100%) | 165/165 (100%) | 97.7-100.0 |
| 1x LoD Flu A | 30/30 (100%) | 45/45 (100%) | 44/45 (97.8%) | 45/45 (100%) | 164/165 (99.4%) | 96.7-99.9 |
| 1x LoD Flu B | 30/30 (100%) | 45/45 (100%) | 45/45 (100%) | 45/45 (100%) | 165/165 (100%) | 97.7-100.0 |
| 5x LoD SARS CoV-2 | 30/30 (100%) | 45/45 (100%) | 45/45 (100%) | 45/45 (100%) | 165/165 (100%) | 97.7-100.0 |
| 5x LoD Flu A | 30/30 (100%) | 45/45 (100%) | 45/45 (100%) | 45/45 (100%) | 165/165 (100%) | 97.7-100.0 |
| 5x LoD Flu B | 30/30 (100%) | 45/45 (100%) | 45/45 (100%) | 45/45 (100%) | 165/165 (100%) | 97.7-100.0 |
2. Linearity:
This is a qualitative assay and linearity is not applicable.
3. Analytical Specificity/Interference:
a. Cross-reactivity and Microbial Interference:
Cross-reactivity and microbial interference studies were conducted to determine if other respiratory pathogens/microbial flora that may be present in the nasal swab samples could cause a false positive test result or interfere with a true positive test result. A total of 50 potential cross-reactive or interfering pathogens comprising of bacteria (19), fungus (1), and viruses (28) and negative matrix (2) were tested in three (3) replicates in the absence (cross-reactivity) or presence (microbial interference) of 2x LoD concentrations of SARS-CoV-2, Flu A and Flu B antigens. These pathogens were diluted using the pooled human nasal fluid in 1X PBS to the target concentration, higher than $1 \times 10^{6} \mathrm{CFU} / \mathrm{mL}$ or IFU/mL for bacteria and fungi, or higher than $1 \times 10^{5} \mathrm{pfu} / \mathrm{mL}$ ( $\mathrm{TCID}_{50} / \mathrm{mL}$ , $\mathrm{CEID}_{50} / \mathrm{mL}$ , or $\mathrm{cp} / \mathrm{mL}$ ) for viruses. Samples were tested in randomized and blinded manner by three operators.
All samples containing potential cross-reactive or interfering pathogens yielded the anticipated test results, as described in Table 4 below. No instances of false negatives or false positives were observed during the study. Consequently, this study affirms that the Nano-Check Influenza +
K243561 - Page 8 of 25
{9}
COVID-19 Dual Test exhibits no reactivity with, nor interference from, the listed pathogens, high prevalence disease agents, and normal or pathogenic flora that are reasonably likely to be encountered in clinical specimens.
Table 4: Results of Cross-reactivity/Microbial Interference
| Microorganism (Strain) | Concentration Tested | Positive Sample (# Positive/# Tested) | Negative Sample (# Positive/# Tested) | Cross-Reactivity/Microbial Interference |
| --- | --- | --- | --- | --- |
| Bordetella pertussis | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Candida albicans | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Chlamydophila pneumoniae | 1.0×106IFU/mL | 3/3 | 0/3 | No |
| Corynebacterium diphtheriae | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Escherichia coli | 1.0×106IFU/mL | 3/3 | 0/3 | No |
| Haemophilus influenzae B | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Lactobacillus acidophilus | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Legionella Pneumophila subsp. Pneumophila | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Moraxella catarrhalis | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Mycobacterium tuberculosis | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Mycoplasma pneumoniae | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Neisseria meningitidis | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Neisseria mucosa | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Neisseria subflava | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Pneumocystis jirovecii -S. cerevisiae | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Pseudomonas aeruginosa | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Staphylococcus aureus | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Staphylococcus epidermidis | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Streptococcus pneumoniae | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Streptococcus pyogenes | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Streptococcus salivarius | 1.0×106CFU/mL | 3/3 | 0/3 | No |
| Epstein-Barr Virus | 1.0×105cp/mL | 3/3 | 0/3 | No |
| Enterovirus 71 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Enterovirus D 68 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human Herpesvirus 5 | 8.0×104TCID50/mL | 3/3 | 0/3 | No |
| Human Adenovirus 1 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human Adenovirus 2 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human Adenovirus 7 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human Coronavirus, 229E | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human Coronavirus, NL63 | 7.0×104TCID50/mL | 3/3 | 0/3 | No |
| Human Coronavirus, OC43 | 4.5×104TCID50/mL | 3/3 | 0/3 | No |
| Human Metapneumovirus 3, B1, Peru2-2002 | 1.95×104TCID50/mL | 3/3 | 0/3 | No |
| Human Metapneumovirus, TN/83-1211 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human Parainfluenza Virus 1 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human Parainfluenza Virus 2 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human Parainfluenza Virus 3 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human Parainfluenza Virus 4B | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human RSV, A Long | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human RSV, A 9320 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human RSV, A2 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human RSV, B 18537 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human RSV, B WV/14617/85 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
{10}
| Microorganism (Strain) | Concentration Tested | Positive Sample (# Positive/# Tested) | Negative Sample (# Positive/# Tested) | Cross-Reactivity/Microbial Interference |
| --- | --- | --- | --- | --- |
| Human RSV, B1 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Human Rhinovirus 1A | 1.0×105PFU/mL | 3/3 | 0/3 | No |
| Measles Virus | 1.7×104TCID50/mL | 3/3 | 0/3 | No |
| MERS-CoV | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Mumps Virus | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| Rhinovirus 20, 15-CV19 | 1.0×105TCID50/mL | 3/3 | 0/3 | No |
| SARS-CoV | 1.0×105PFU/mL | 3/3 | 0/3 | No |
| Pooled Human Nasal Wash | N/A | 3/3 | 0/3 | No |
| Pooled Human Nasal Fluid | N/A | 3/3 | 0/3 | No |
| Note: Coronavirus HKU1 was not tested for cross-reactivity due to lack of availability. 20 clinical samples that are positive for Coronavirus HKU1 were tested, and all resulted as negative, however, the viral load/concentration of each sample is unknown. | | | | |
# b. Endogenous/Exogenous Substances Interference Study:
The interference study aimed to evaluate potential endogenous and exogenous interference with the Nano-Check Influenza+COVID-19 Dual Test. Thirty-four (34) substances, including whole blood, various over-the-counter (OTC) products, household items, and common chemicals, were assessed for cross-reactivity or interference with SARS-CoV-2, Flu A, and Flu B detection. Each substance was spiked onto swab heads in the presence or absence of 2x LoD concentrations of the antigens. Samples were tested in randomized and blinded manner by three operators.
All samples, including those containing potentially interfering substances, yielded expected results without false negatives or false positives as shown in Table 5 below. Therefore, the identified substances are unlikely to interfere with the ability of the Nano-Check Influenza+COVID-19 Dual Test to detect SARS-CoV-2, Flu A, or Flu B.
Table 5: Endogenous/Exogenous Interference Test Results
| Endogenous/Exogenous Substance Test concentration | Spiked Virus (Concentration) | No. of Positive Results / No. of Total replicates | | |
| --- | --- | --- | --- | --- |
| | | SARS CoV-2 | Flu A | Flu B |
| Nasal Spray (Flonase) Test conc. = 15% v/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Nasal Spray (Samilisan) Test conc. = 15% v/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Nasal Spray (CVS) Test conc. = 15% v/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Nasal Spray 4 (Afrin) Test conc. = 15% v/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
K243561 - Page 10 of 25
{11}
| Endogenous/Exogenous Substance Test concentration | Spiked Virus (Concentration) | No. of Positive Results / No. of Total replicates | | |
| --- | --- | --- | --- | --- |
| | | SARS CoV-2 | Flu A | Flu B |
| Budesonide Nasal Spray Test conc. = 15% v/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| NASONEX 24 hr. Allergy Test conc. = 15% v/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Nasacort Allergy 24 hr. Test conc. = 15% v/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Sore Throat (Oral Pain Reliever spay) Test conc. = 15% v/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| ZICAM Oral mist Test conc. = 15% v/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Sore Throat Lozenges (Cepacol) Test conc. = 15% w/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Zinc Cold Therapy (Walgreens) Test conc. = 15% w/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Homeopathic Allergy Nasal Spray (Alkalol) Test conc. = 15% v/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| NasoGEL (Gel Spray) Test conc. = 15% v/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Nasalcrom Nasal Allergy spray Test conc. = 15% v/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Histaminum 30C Test conc. = 15% w/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
K243561 - Page 11 of 25
{12}
| Endogenous/Exogenous Substance Test concentration | Spiked Virus (Concentration) | No. of Positive Results / No. of Total replicates | | |
| --- | --- | --- | --- | --- |
| | | SARS CoV-2 | Flu A | Flu B |
| Skin relief hand cream (Aveeno) Test conc. = 1% w/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Hand Soap Fresh Breeze Scent Test conc. = 1% w/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Antibacterial liquid Hand Soap (Dial Complete) Test conc. = 1%, 5%, 10% and 15% w/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Hand Sanitizer Gel Test conc. = 1%, 5%, 10% and 15% w/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Hand Sanitizer Lotion Test conc. = 15% w/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 0 / 3* |
| Disinfectant Spray (Lysol) Test conc. = 1% v/v | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Mucin from bovine submaxillary glands-Type I-S Test conc. = 5 mg/mL | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Purified Human Neutrophils Test conc. = 5×10^6 cells/mL | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Whole Blood Test conc. = 2.5% | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Acetylsalicylic acid Test conc. = 3.00×10^3 μg/mL | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Dexamethasone Test conc. = 1.20×10^3 μg/mL | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Mometasone furoate | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
K243561 - Page 12 of 25
{13}
| Endogenous/Exogenous Substance Test concentration | Spiked Virus (Concentration) | No. of Positive Results / No. of Total replicates | | |
| --- | --- | --- | --- | --- |
| | | SARS CoV-2 | Flu A | Flu B |
| Test conc. = 4.50×10-4μg/mL | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Mupirocin Test conc. = 1.50×100μg/mL | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Oseltamivir phosphate (Tamiflu) Test conc. = 3.99×10-1μg/mL | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Tobramycin Test conc. = 3.30×101μg/mL | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Beclomethasone dipropionate Test conc. = 5.04 μg/mL | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Flunisolide Test conc. = 870 μg/mL | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Molnupiravir Test conc. = 3.29 mg/mL (10 mM) | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Remdesivir Test conc. = 240 μg/mL | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
| Zanamivir Test conc. = 30 mg/mL | Negative | 0 / 3 | 0 / 3 | 0 / 3 |
| | SARS-CoV-2 (2x LoD) | 3 / 3 | 0 / 3 | 0 / 3 |
| | Influenza A (2x LoD) | 0 / 3 | 3 / 3 | 0 / 3 |
| | Influenza B (2x LoD) | 0 / 3 | 0 / 3 | 3 / 3 |
* Hand Sanitizer Lotion tested at diluted concentrations of 1%, 5%, and 10% w/v showed no interference in test performance for Flu B.
c. Biotin Interference:
The Nano-Check Influenza+COVID-19 Dual Test utilizes a sandwich assay format that incorporates a biotin-streptavidin complex. This study was designed to ascertain whether the test is affected by free biotin at levels up to $3500\mathrm{ng / mL}$ in samples in the presence of 2x LoD concentrations of SARS-CoV-2, Flu A, and Flu B antigens, spiked individually and were tested in triplicate. As summarized in Table 6 below, the Nano-Check Influenza+COVID-19 Dual Test demonstrated no interference up to $3,500\mathrm{ng / mL}$ of biotin in test samples.
K243561 - Page 13 of 25
{14}
Table 6: Biotin Interference Study Results
| Sample | Biotin Concentration | (No. of Positive results / Total replicates) | | |
| --- | --- | --- | --- | --- |
| | | COVID | Flu A | Flu B |
| SARS-CoV-2 2x LoD Positive Sample | 3,500 ng/mL | 3 / 3 | 0 / 3 | 0 / 3 |
| | 2,400 ng/mL | 3 / 3 | 0 / 3 | 0 / 3 |
| | 1,200 ng/mL | 3 / 3 | 0 / 3 | 0 / 3 |
| | 600 ng/mL | 3 / 3 | 0 / 3 | 0 / 3 |
| | 300 ng/mL | 3 / 3 | 0 / 3 | 0 / 3 |
| | 150 ng/mL | 3 / 3 | 0 / 3 | 0 / 3 |
| Influenza A 2x LoD Positive Sample | 3,500 ng/mL | 0 / 3 | 3 / 3 | 0 / 3 |
| | 2,400 ng/mL | 0 / 3 | 3 / 3 | 0 / 3 |
| | 1,200 ng/mL | 0 / 3 | 3 / 3 | 0 / 3 |
| | 600 ng/mL | 0 / 3 | 3 / 3 | 0 / 3 |
| | 300 ng/mL | 0 / 3 | 3 / 3 | 0 / 3 |
| | 150 ng/mL | 0 / 3 | 3 / 3 | 0 / 3 |
| Influenza B 2x LoD Positive Sample | 3,500 ng/mL | 0 / 3 | 0 / 3 | 3 / 3 |
| | 2,400 ng/mL | 0 / 3 | 0 / 3 | 3 / 3 |
| | 1,200 ng/mL | 0 / 3 | 0 / 3 | 3 / 3 |
| | 600 ng/mL | 0 / 3 | 0 / 3 | 3 / 3 |
| | 300 ng/mL | 0 / 3 | 0 / 3 | 3 / 3 |
| | 150 ng/mL | 0 / 3 | 0 / 3 | 3 / 3 |
| Negative sample (NCM) | 3,500 ng/mL | 0 / 3 | 0 / 3 | 0 / 3 |
| | 2,400 ng/mL | 0 / 3 | 0 / 3 | 0 / 3 |
| | 1,200 ng/mL | 0 / 3 | 0 / 3 | 0 / 3 |
| | 600 ng/mL | 0 / 3 | 0 / 3 | 0 / 3 |
| | 300 ng/mL | 0 / 3 | 0 / 3 | 0 / 3 |
| | 150 ng/mL | 0 / 3 | 0 / 3 | 0 / 3 |
# d. Competitive Interference:
The competitive study interference was performed to assess the ability of the Nano-Check Influenza+COVID-19 Dual Test to detect multiple analytes simultaneously without inhibiting each other's detection. Strains of SARS-CoV-2 (Omicron lineage BA.5), Flu A (H1N1 and H3N2) and Flu B (Victoria and Yamagata lineage) were utilized. Each analyte was prepared at a high concentration (at least $10^{5}\mathrm{TCID}_{50} / \mathrm{mL}$ or $\mathrm{CEID}_{50} / \mathrm{mL}$ ) and low concentration (3x LoD).
Table 7 below summarizes the results of the competitive interference study. For each condition tested, all three replicates tested at the low target analyte condition tested positive in the presence of other target analytes at high concentrations. Thus, this indicates the capability of the Nano-Check Influenza+COVID-19 Dual Test to detect low levels of Flu A and Flu B in the presence of a high level of SARS-CoV-2, as well as low levels of Flu A and SARS-CoV-2 in the presence of a high level of Flu B and vice versa. No false positive results were observed for analytes that are not present in the sample.
K243561 - Page 14 of 25
{15}
Table 7: Competitive Interference Results
| SARS-CoV-2 | | Influenza A (H1N1) | | Influenza B (Victoria) | |
| --- | --- | --- | --- | --- | --- |
| Concentration | % Agreement | Concentration | % Agreement | Concentration | % Agreement |
| Low | 100% | High | 100% | - | 100% |
| - | 100% | High | 100% | Low | 100% |
| Low | 100% | High | 100% | Low | 100% |
| High | 100% | Low | 100% | Low | 100% |
| High | 100% | - | 100% | Low | 100% |
| Low | 100% | - | 100% | High | 100% |
| - | 100% | Low | 100% | High | 100% |
| Low | 100% | Low | 100% | High | 100% |
| High | 100% | Low | 100% | - | 100% |
| SARS-CoV-2 | | Influenza A (H3N2) | | Influenza B (Victoria) | |
| Concentration | % Agreement | Concentration | % Agreement | Concentration | % Agreement |
| Low | 100% | High | 100% | - | 100% |
| - | 100% | High | 100% | Low | 100% |
| - | 100% | Low | 100% | High | 100% |
| High | 100% | Low | 100% | - | 100% |
| Low | 100% | High | 100% | Low | 100% |
| Low | 100% | Low | 100% | High | 100% |
| High | 100% | Low | 100% | Low | 100% |
| SARS-CoV-2 | | Influenza A (H1N1) | | Influenza B (Yamagata) | |
| Concentration | % Agreement | Concentration | % Agreement | Concentration | % Agreement |
| - | 100% | High | 100% | Low | 100% |
| - | 100% | Low | 100% | High | 100% |
| High | 100% | - | 100% | Low | 100% |
| Low | 100% | High | 100% | Low | 100% |
| Low | 100% | Low | 100% | High | 100% |
| High | 100% | Low | 100% | Low | 100% |
| Low | 100% | - | 100% | High | 100% |
| SARS-CoV-2 | | Influenza A (H3N2) | | Influenza B (Yamagata) | |
| Concentration | % Agreement | Concentration | Concentration | % Agreement | Concentration |
| - | 100% | High | 100% | Low | 100% |
| - | 100% | Low | 100% | High | 100% |
| Low | 100% | High | 100% | Low | 100% |
| Low | 100% | Low | 100% | High | 100% |
4. Assay Reportable Range:
Not applicable; the device is a binary qualitative assay that is visually read.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
a. Internal Control
The Nano-Check Influenza+COVID-19 Dual Test contains built in internal assay control. The appearance of the control line on the test ensures that sufficient flow of the sample occurred during the assay. Any result without this control line is invalid.
b. External Quality Controls
The Nano-Check Influenza+COVID-19 Dual Test contains one positive external control swab and one negative external control swab that allows for monitoring of the performance of the assay. The positive control swab contains recombinant SARS-CoV-2 nucleocapsid, recombinant
K243561 - Page 15 of 25
{16}
Flu A nucleoprotein, and recombinant Flu B nucleoprotein whereas the negative control swab does not contain recombinant antigen.
A precision study was conducted for multiple lots of the Nano-Check Influenza+COVID-19 Dual Test's external controls. The aim was to assess the precision and repeatability of the external control swab in a blinded manner. Ten (10) individual swabs from three (3) different lots of positive and negative external control swabs were randomized and distributed blind to one (1) assigned operator within the laboratory, who tested the control swabs in accordance with the Instructions for Use of the test. The collected data were evaluated by calculating the percent agreement between test results and sample properties.
The results of this study, as detailed in Table 8 below indicated 100% agreement, meeting the acceptance criteria for the external control precision study.
Table 8: Lot-to-lot Precision Study Results for External Controls
| External control | Lot | Test Results (No. of Positive Results / No. of Total Replicates) | | | |
| --- | --- | --- | --- | --- | --- |
| | | COVID | Flu A | Flu B | Agreement |
| Positive External Control | Lot 1 | 10 / 10 | 10 / 10 | 10 / 10 | 100% |
| | Lot 2 | 10 / 10 | 10 / 10 | 10 / 10 | 100% |
| | Lot 3 | 10 / 10 | 10 / 10 | 10 / 10 | 100% |
| Negative External Control | Lot 1 | 0 / 10 | 0 / 10 | 0 / 10 | 100% |
| | Lot 2 | 0 / 10 | 0 / 10 | 0 / 10 | 100% |
| | Lot 3 | 0 / 10 | 0 / 10 | 0 / 10 | 100% |
c. Stability
i) Specimen Stability
Two test samples were prepared for the specimen stability study: a negative sample (consisting of pooled human nasal fluid with no analyte) and a low positive sample (containing a diluted SARS-CoV-2, Flu A, and Flu B sample at 2x LoD in negative pooled human nasal fluid). Swabs were spiked with 50 µL of each sample and subjected to four different temperature conditions: room temperature (23.5°C), high room temperature (30 °C), 2°C to 8°C, and frozen (-20°C). The samples were stored under each condition for various time intervals including 0 hours (baseline), 1 hour, 2 hours, 4 hours, 8 hours, 24 hours, and 48 hours. Subsequently, the exposed sample swabs were tested using 5 replicates for each exposure time. The results showed that the nasal swab samples were stable for up 48 hours under all temperature conditions, ranging from -20°C to 30°C. Therefore, freshly collected specimen swabs are recommended to be processed no later than one hour after specimen collection when kept at room temperature (15°C - 30°C) or within 24 hours when stored at 2°C to 8°C.
ii) Real Time Stability
Three lots of Nano-Check Influenza+COVID-19 Dual Test were stored at three different temperature conditions: 30°C, ambient room temperature (23.5°C), and 2-8°C. The study involved positive and negative external control swabs, as well as positive and negative contrived samples. Positive samples were prepared by diluting the SARS-CoV-2 (Omicron variant), Flu A (H1N1), and Flu B (Yamagata lineage) strains to a 3x LoD concentration into pooled human nasal fluid,
K243561 - Page 16 of 25
{17}
while negative contrived swab samples were prepared using pooled human nasal fluid confirmed to be negative for SARS-CoV-2, Flu A, and Flu B by RT-PCR. Testing was performed at time 0 (baseline) and monthly thereafter up to 19 months. All study data are 100% concordant with expected results and support a shelf-life of up to 18 months at the time of 510(k) clearance when stored between 2-30°C.
### ii) Transport Stability
The transport simulation study was performed to evaluate the robustness of the Nano-Check Influenza+COVID-19 Dual Test kit under fluctuating transport and storage conditions. Positive and negative external control swabs, as well as positive and negative contrived swab samples, were included in the study. Positive contrived swab samples were prepared by diluting the SARS-CoV-2 (Omicron variant), Flu A (HlNl), and Flu B (Yamagata lineage) strains to a 3x LoD concentration into pooled human nasal fluid, while negative contrived swab samples were prepared using pooled human nasal fluid confirmed to be negative for SARS-CoV-2, Flu A, and Flu B by RT-PCR. The study was scheduled over 15-month period, with assessments carried out on Day 0, Day 14, and every three months from Month 3 to Month 12. Results indicated that the Nano-Check Influenza+COVID-19 Dual Test kit demonstrated stability and remained unaffected by temperature excursions for 12 months.
### iv) Open Kit Stability Study:
Open kit stability study was performed to assess the stability of the test devices after opening the sealed pouch using ten replicates of positive samples (2x LoD concentration of SARS-CoV-2, Flu A and Flu B viruses spiked individually) and 5 replicates of negative samples (pooled human fluid confirmed to be SARS-CoV-2, Flu A and B negative by RT-PCR). The test devices were exposed to ambient conditions for various durations: 0 min, 30 min, 1 hr, 2 hrs, 3 hrs and 4 hrs. The results showed that all positive samples had positive results for up to 4 hours and negative samples did not produce any false positive results meeting the acceptance criteria.
### v) Short-term Stability:
The short-term reagent stability study for the Nano-Check Influenza+COVID-19 Dual Test aimed to assess the stability of the Nano-Check Influenza+COVID-19 Dual Test kit components when all kit components are stored at extremely low temperature (-20°C) and high temperature (45°C). The Nano-Check Influenza+COVID-19 Dual Test kit was stable for 16 days when it was stored at low temperature (-20°C) and for 14 days at a high temperature (45°C). However, it is recommended that the test kit should be stored at 2°C - 30°C in the original sealed pouch.
### 6. Detection Limit:
#### a. Single-analyte LoD:
The LoD of the device was performed to determine the lowest detectable concentration of SARS-CoV-2, Flu A and Flu B at which at least 95% of all true positive replicates are consistently detected as positive. The LoD was assessed for each analyte in two parts, a preliminary range finding study, followed by a confirmatory LoD study. A preliminary LoD was determined by first testing serial ten-fold dilutions of live Flu A and B, and inactivated SARS-CoV-2 virus stocks diluted into pooled negative swab fluid in 3 replicates per dilution using one device lot. Single analyte virus dilutions (50 μL/swab) were each spiked onto dry sterile swabs and tested per the IFU. The preliminary LoD results for each individual virus strain are shown in below tables.
K243561 - Page 17 of 25
{18}
Table 9: Preliminary LoD - SARS-CoV-2
| Virus strain | Sample conc. (TCID_{50}/mL) | Sample conc. (TCID_{50}/swab) | # Positive/# Total |
| --- | --- | --- | --- |
| USA/MD HP20874/2021 (Omicron, B.1.1.529) | 3.89 × 10^{3} | 1.95 × 10^{2} | 3 / 3 |
| | 3.89 × 10^{2} | 1.95 × 10^{1} | 3 / 3 |
| | 3.89 × 10^{1} | 1.95 | 0 / 3 |
| | 1.95 × 10^{2} | 9.75 | 3 / 3 |
| | 9.73 × 10^{1} | 4.87 | 1 / 3 |
| USA/COR-22-063113/2022 (Omicron, BA.5) | 2.53 × 10^{5} | 1.27 × 10^{4} | 3 / 3 |
| | 2.53 × 10^{4} | 1.27 × 10^{3} | 3 / 3 |
| | 2.53 × 10^{3} | 1.27 × 10^{2} | 0 / 3 |
| | 1.27 × 10^{4} | 6.35 × 10^{2} | 3 / 3 |
| | 6.35 × 10^{3} | 3.18 × 10^{2} | 0 / 3 |
Table 10: Preliminary LoD - Flu A
| Virus strain | Sample conc. (TCID_{50}/mL or CEID_{50}/mL) | Sample conc. (TCID_{50}/swab or CEID_{50}/swab) | # Positive/# Total |
| --- | --- | --- | --- |
| A/California/04/2009 (H1N1) | 2.8 × 10^{5} | 1.4 × 10^{4} | 3 / 3 |
| | 2.8 × 10^{4} | 1.4 × 10^{3} | 3 / 3 |
| | 2.8 × 10^{3} | 1.4 × 10^{2} | 3 / 3 |
| | 2.8 × 10^{2} | 1.4 × 10^{1} | 0 / 3 |
| | 1.4 × 10^{3} | 7 × 10^{1} | 0 / 3 |
| A/Victoria/361/2011 (H3N2) | 2.8 × 10^{7} | 1.4 × 10^{6} | 3 / 3 |
| | 2.8 × 10^{6} | 1.4 × 10^{5} | 3 / 3 |
| | 2.8 × 10^{5} | 1.4 × 10^{4} | 3 / 3 |
| | 2.8 × 10^{4} | 1.4 × 10^{3} | 0 / 3 |
| | 1.4 × 10^{5} | 7 × 10^{3} | 3 / 3 |
| | 7.0 × 10^{4} | 3.5 × 10^{3} | 1 / 3 |
Table 11: Preliminary LoD - Flu B
| Virus strain | Sample conc. (TCID_{50}/mL or CEID_{50}/mL) | Sample conc. (TCID_{50}/swab or CEID_{50}/swab) | # Positive/# Total |
| --- | --- | --- | --- |
| B/Hong Kong/330/2001 (Victoria) | 1.8 × 10^{6} | 9 × 10^{4} | 3 / 3 |
| | 1.8 × 10^{5} | 9 × 10^{3} | 1 / 3 |
| | 9.0 × 10^{5} | 9 × 10^{2} | 3 / 3 |
| | 4.5 × 10^{5} | 2.25 × 10^{4} | 3 / 3 |
| | 2.25 × 10^{5} | 1.13 × 10^{4} | 3 / 3 |
| | 1.13 × 10^{5} | 5.6 × 10^{3} | 0 / 3 |
| B/Phuket/3073/13 (Yamagata) | 4.17 × 10^{4} | 2.09 × 10^{3} | 3 / 3 |
| | 4.17 × 10^{3} | 2.09 × 10^{2} | 3 / 3 |
| | 4.17 × 10^{2} | 2.09 × 10^{1} | 3 / 3 |
K243561 - Page 18 of 25
{19}
| Virus strain | Sample conc. (TCID_{50}/mL or CEID_{50}/mL) | Sample conc. (TCID_{50}/swab or CEID_{50}/swab) | # Positive/# Total |
| --- | --- | --- | --- |
| | 4.17 × 10^{1} | 2.09 | 0 / 3 |
| | 2.09 × 10^{2} | 1.04 x 10^{1} | 3 / 3 |
| | 1.04 × 10^{2} | 5.2 | 3 / 3 |
| | 5.20 × 10^{1} | 2.6 | 1 / 3 |
LoD confirmatory testing was performed individually for each virus by testing 20 replicates at the virus’s preliminary 1x LoD concentration, as determined above using three device lots. For the LoD to be confirmed, at least 95% of the replicates (≥57/60) needed to be positive. Results of the LoD confirmation testing for each virus are summarized in the table below.
Table 12: Confirmatory LoD
| Virus | Isolate (Lineage) | Strain | LoD concentration (TCID_{50}/mL or CEID_{50}/mL) | LoD concentration (TCID_{50}/Swab or CEID_{50}/Swab) | # Positive/# Total |
| --- | --- | --- | --- | --- | --- |
| SARS-CoV-2 | Omicron (B.1.1.529) | USA/MD-HPA20874/2021 (Heat Inactivated) | 1.95 × 10^{2} | 9.75 | 59/60 |
| | Omicron (BA.5) | USA/COR-22-063113/2022 (Heat Inactivated) | 1.27 × 10^{4} | 6.35 x 10^{2} | 60/60 |
| Flu A | H1N1 | A/California/04/2009 | 2.8 × 10^{3} | 1.4 x 10^{2} | 59/60 |
| | H3N2 | A/Victoria/361/2011 | 1.4 × 10^{5} | 7 x 10^{3} | 59/60 |
| Flu B | Victoria | B/Hong Kong/330/2001 | 2.25 × 10^{5} | 1.13 × 10^{4} | 57/60 |
| | Yamagata | B/Phuket/3073/13 | 1.04 × 10^{2} | 5.2 | 60/60 |
b. NIBSC 21/368 – International Standard:
The LoD of the candidate device was determined using the NIBSC 21/368 International Standard for SARS-CoV-2 spiked into pooled negative matrix (nasal fluid). A preliminary LoD test was performed by spiking 50 µL of each diluted sample onto the sample collection swab head in three replicates. The preliminary LoD concentration was determined to be 1,333 IU/mL (or 67 IU/swab).
The LoD confirmatory study was performed using additional 17 replicates at the preliminary LoD concentration totaling 20 replicates. All 20 replicates resulted in positive results. Additional dilutions were investigated bracketing the confirmed LoD concentration with 20 replicates at each dilution level. The lowest concentration at which a minimum of 95% of results were positive was confirmed to be 667 IU/mL (or 33.5 IU/swab) as shown in Table 13 below.
K243561 - Page 19 of 25
{20}
Table 13: Summary Results of LoD with NIBSC 21/368 International Standard
| Preliminary LoD | | | | Confirmatory LoD | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample dilution | Concentration | | # Positive / # total replicates | Sample dilution | Concentration | | # Positive / # total replicates |
| | IU/mL | IU/swab | | | IU/mL | IU/swab | |
| 1:5 (Master Stock) | 4,000 | 200 | 3/3 | 1:7.5 | 2,666 | 134 | 20/20 |
| 1:15 | 1,333 | 67 | 20/20 | 1:22.5 | 889 | 45 | 20/20 |
| 1:45 | 444 | 22 | 1/3 | 1:30 | 667 | 33.5 | 19/20 |
| 1:135 | 148 | 7.4 | 0/3 | 1:37.5 | 533 | 27 | 12/20 |
| 1:405 | 49 | 2.5 | 0/3 | | | | |
7. High Dose Hook Effect Study:
A high-dose hook effect study was conducted to determine the concentration at which false negative results might occur when very high levels of the target antigen are present in a testing sample. In this study, 50μL of the highest concentration possible for inactivated SARS-CoV-2 virus stock and for live Flu A and Flu B virus stocks were spiked onto sterile swabs for triplicate measurements, and swabs were tested on the device per IFU of the candidate device. No hook effect for SARS-CoV-2, Flu A, and Flu B were observed up to the concentrations listed in Table 14 below.
Table 14: Summary of High Dose Hook Effect Study
| Virus | Isolate/Lineage | Strain | Virus Concentration | | # of Positive/# Total tested | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | TCID_{50}/mL or CEID_{50}/mL | TCID_{50}/swab or CEID_{50}/swab | COVID | Flu A | Flu B |
| SARS-CoV-2 | Omicron (B.1.1.529) | USA/MD-HPA20874/2021 | 3.89 x 10^{4} | 1.95 x 10^{3} | 3/3 | 0/3 | 0/3 |
| | Omicron (BA.5) | USA/COR-22-063113/2022 | 2.53 x 10^{6} | 1.27 x 10^{5} | 3/3 | 0/3 | 0/3 |
| Flu A | H1N1 | A/California/04/2009 | 2.8 x 10^{6} | 1.4 x 10^{5} | 0/3 | 3/3 | 0/3 |
| | H3N2 | A/Victoria/361/2011 | 2.8 x 10^{8} | 1.4 x 10^{7} | 0/3 | 3/3 | 0/3 |
| Flu B | Victoria | B/Hong Kong/330/2001 | 1.8 x 10^{7} | 9 x 10^{6} | 0/3 | 0/3 | 3/3 |
| | Yamagata | B/Phuket/3073/13 | 4.17 x 10^{5} | 2.08 x 10^{4} | 0/3 | 0/3 | 3/3 |
8. Analytical Reactivity Study:
a. Inclusivity – Selected Strains
An inclusivity study was conducted involving three (3) operators encompassing various variants, lineages and clades of the SARS-CoV-2 virus and Flu viruses. The study involved 14 strains of SARS-CoV-2, thirty (30) strains of Flu A, and sixteen (16) strains of Flu B. The virus stocks were diluted at the predetermined concentration in the pooled negative nasal fluid matrix and 50μL were pipetted onto the swabs. The swabs were tested in 3 replicates in this study. The
K243561 - Page 20 of 25
{21}
lowest concentration of each strain that resulted in $100\%$ detection (3/3) is presented in Table 15 below.
Table 15: Summary of Inclusivity Results
| Virus | Virus Strains | Concentration | Units |
| --- | --- | --- | --- |
| SARS-CoV-2(B.1.1.7, Alpha) | USA/CA/CDC/5574/20201) | 6.77E+06 | GE /mL |
| | USA/CA/CDC/5574/20202) | 2.39E+04 | TCID50/mL |
| | England/204820464/2020 | 7.19E+03 | TCID50/mL |
| SARS-CoV-2(B.1.1.351, Beta) | USA/MD-HP01542/20213) | 7.20E+04 | GC/mL |
| | USA/MD-HP01542/20214) | 3.80E+06 | GC/mL |
| | South Africa/KRISP-K0053 25/2020 | 1.90E+04 | TCID50/mL |
| SARS-CoV-2(B.1.617.2, Delta) | USA/MD-HP05285/20215) | 7.20E+07 | GC/mL |
| | USA/MD-HP05285/20216) | 5.00E+06 | GC/mL |
| | USA/PHC658/2021 | 5.21E+02 | TCID50/mL |
| SARS-CoV-2(P.1, Gamma) | Japan/TY7-503/2021 | 1.58E+04 | TCID50/mL |
| | USA/NY-Wadsworth-21033 899-01/ 2021 | 7.85E+03 | TCID50/mL |
| SARS-CoV-2(B.1.617.1, Kappa) | USA/CA-Stanford-15_S02/ 2021 | 8.48E+04 | TCID50/mL |
| SARS-CoV-2(B.1.1.529, Omicron) | USA/GA-EHC-2811C/2021 | 2.11E+07 | GC/mL |
| SARS-CoV-2(JN.1, Omicron) | USA/New York/PV96109/20 23 | 3.14E+03 | TCID50/mL |
| Flu A H1N1 | A/Puerto Rico/8/34 | 8.00E+06 | CEID50/mL |
| | A/Brisbane/59/2007 | 4.45E+06 | CEID50/mL |
| | A/Denver/1/57 | 4.00E+05 | CEID50/mL |
| | A/San Diego/1/2009 pdm09 | 2.80E+04 | TCID50/mL |
| | A/Tijuana/4/09 | 2.45E+01 | TCID50/mL |
| | A/Solomon Islands/3/2006 | 4.45E+05 | CEID50/mL |
| | A/NWS/33 | 1.23E+05 | CEID50/mL |
| | A/FM/1/47 | 4.25E+05 | CEID50/mL |
| | A/New Jersey/8/76 | 1.70E+04 | CEID50/mL |
| | A/New Caledonia/20/1999 | 4.00E+05 | CEID50/mL |
| | A/Hawaii/66/2019 | 1.28 | HA |
| | A/Hawaii/66/2019 X-345A | 1.28 | HA |
| | A/Guangdong-Maonan/1536/ 2019 | 1.28 | HA |
| | A/Guangdong-Maonan/1536/ 2019 CNIC-1909 | 0.64 | HA |
| | A/Victoria/4897/2022(pdm09) | 1.58E+07 | EID50/mL |
| | A/Victoria/2570/2019 (pdm09) | 9.98E+05 | EID50/mL |
| Flu A H1N2 | A/Swine/Ohio/09SW1477/2009 | 2.30E+04 | TCID50/mL |
| Flu A H3N2 | A/Hong Kong/8/1968 | 1.40E+05 | CEID50/mL |
| | A/Aichi/2/1968 | 4.00E+05 | CEID50/mL |
| | A/Wisconsin/67/2005 | 7.00E+05 | CEID50/mL |
| | A/Hong Kong/4801/2014 | 9.60E+05 | CEID50/mL |
| | A/Netherlands/22/2003 | 8.00E+02 | TCID50/mL |
| | A/Netherlands/823/1992 | 1.44E+01 | TCID50/mL |
K243561 - Page 21 of 25
{22}
| Virus | Virus Strains | Concentration | Units |
| --- | --- | --- | --- |
| | A/Brisbane/10/2007 | 1.38E+05 | CEID50/mL |
| | A/Wisconsin/15/2009 | 5.00E+03 | CEID50/mL |
| | A/Sydney/5/97 | 4.45E+04 | CEID50/mL |
| | A/Port Chalmers/1/73 | 2.00E+05 | CEID50/mL |
| | A/Victoria/3/75 | 4.00E+05 | CEID50/mL |
| | A/Perth/16/2009 x A/Puerto Rico/8/19 34, NIB-64 | 2.80E+06 | CEID50/mL |
| | A/Singapore/INFIMH-16-0019 /16 | 2.51E+03 | TCID50/mL |
| Flu A H5N1 | A/mallard/Wisconsin/2576/2009 | 5.25E+05 | GE/mL |
| Flu B (Victoria lineage) | B/Brisbane/60/2008 | 9.00E+04 | CEID50/mL |
| | B/Malaysia/2506/2004 | 1.12E+06 | CEID50/mL |
| | B/New York/1056/2003 | 3.20E+03 | TCID50/mL |
| | B/Brisbane/60/2008 | 9.00E+04 | CEID50/mL |
| Flu B (Yamagata lineage) | B/Florida/78/2015 | 2.80E+04 | TCID50/mL |
| | B/Texas/06/2011 | 4.45E+08 | CEID50/mL |
| | B/New York/1061/2004 | 8.00E+02 | TCID50/mL |
| | B/Christchurch/33/2004 | 8.00E+02 | TCID50/mL |
| | B/Sydney/507/2006 | 1.60E+05 | TCID50/mL |
| | B/Wisconsin/1/2010 | 1.80E+06 | CEID50/mL |
| | B/Florida/4/2006 | 3.50E+05 | CEID50/mL |
| Flu B (Non-Victoria/Yamagata) | B/Colorado/6/17 | 1.78E+02 | TCID50/mL |
| | B/Taiwan/2/1962 | 4.45E+03 | CEID50/mL |
| | B/Lee/1940 | 9.00E+04 | CEID50/mL |
| | B/GL/1739/54 | 5.00E+04 | CEID50/mL |
| | B/Great Lakes/1739/1954 | 3.20E+04 | CEID50/mL |
1) Source: Bei Resources (Cat. #: NR-55245, Lot#:70043111), 2) Source: ZeptoMetrix (Cat. #: 0810612CFHI, Lot#:328055), 3) Source: Bei Resources (Cat. #: NR-553651, Lot#:70045299), 4) Source: Bei Resources (Cat. #: NR-55350, Lot#:70045608), 5) Source: Bei Resources (Cat. #: NR-56128, Lot#:70048021), 6) Source: ATCC (Cat. #: VR-3342HK, Lot#:70048932
# 9. Assay Cut-Off:
Not applicable as this is a qualitative visually read assay without numeric data.
# B Comparison Studies:
# 1. Method Comparison with Predicate Device:
Please refer to section VI.C (Clinical Studies) below for the clinical validation regarding the method comparison studies.
# 2. Matrix Comparison:
The Nano-Check Influenza+COVID-19 Dual Test is intended only for the qualitative detection of nucleocapsid protein antigen from SARS-CoV-2 and the nucleoprotein from Flu A and Flu B in direct anterior nasal swab specimens. As no other sample types are claimed herein, a matrix comparison study is not applicable.
K243561 - Page 22 of 25
{23}
# C Clinical Studies:
The clinical performance of the Nano-Check Influenza+COVID-19 Dual Test was evaluated in a multi-center, prospective clinical study in the U.S. between November 2022 and February 2025. The study only enrolled subjects who presented with symptoms of respiratory infection. A total of 2,223 subjects were initially enrolled however, 254 samples were excluded due to subjects having symptoms longer than 4 days, invalid comparator results, or samples lost during shipment. The remaining 1,969 subjects were consecutively enrolled and tested by fourteen (14) operators representative of CLIA-waived users (i.e., not formally trained in a laboratory setting) across six (6) different clinical CLIA-waived sites. Two anterior nasal swabs (ANS) were collected from each study subject during the same visit. The first ANS specimen for the comparator method was collected by the operators from both sides of the nose. The comparator ANS specimens were stored in UVT media and packaged in dry ice for aero-transport, sent to a reference laboratory, and tested with an FDA cleared RT-PCR method as per the cleared instructions for use. The second ANS specimen was collected from both sides of the nose using the provided swab and was tested immediately using the Nano-Check Influenza+COVID-19 Dual Test by each operator using the QRI only and with no training specifically with the candidate test.
The summary of demographics of the clinical subjects are shown in Table 16 below.
Table 16: Subject Demographics
| Age Group | Female | | Male | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | No. of Sample | % | No. of Sample | % | No. of Sample | % |
| ≤ 5 years | 201 | 19.8 | 215 | 22.5 | 416 | 21.1 |
| 6 to 21 years | 401 | 39.5 | 462 | 48.5 | 863 | 43.8 |
| 22 to 60 years | 313 | 30.8 | 204 | 21.4 | 517 | 26.3 |
| ≥ 61 years | 101 | 9.9 | 72 | 7.6 | 173 | 8.8 |
| Total | 1016 | 100 | 953 | 100 | 1969 | 100 |
| | Female: 51.6% (1016/1969) | | Male: 48.4% (953/1969) | | | |
The clinical performance of the Nano-Check Influenza+COVID-19 Dual Test compared to the highly sensitive RT-PCR comparator is shown below.
Table 17: SARS-CoV-2 Clinical Performance
| Nano-Check Influenza+ COVID-19 Dual Test | RT-PCR Comparator | | Total |
| --- | --- | --- | --- |
| | Positive | Negative | |
| Positive | 232 | 3 | 275 |
| Negative | 33 | 1701 | 1734 |
| Total | 265 | 1704 | 1969 |
| Positive Percent Agreement (PPA) = 87.6% (232/265, 95% CI: 83.0% - 91.0%) | | | |
| Negative Percent Agreement (NPA) = 99.8% (1701/1704, 95% CI: 99.5% - 99.9%) | | | |
K243561 - Page 23 of 25
{24}
Results of SARS-CoV-2 were also analyzed stratified by the days post symptom onset (DPSO) and are presented in the table below.
Table 18: Clinical Performance for Detection of SARS-CoV-2 stratified by DPSO
| Days Post Onset | Comparator Positive | Candidate Test Positive | PPA % | 95%CI |
| --- | --- | --- | --- | --- |
| 1 | 82 | 74 | 90.2 | 81.9-95.0 |
| 2 | 98 | 85 | 86.7 | 78.6-92.1 |
| 3 | 64 | 56 | 87.5 | 77.2-93.5 |
| 4 | 21 | 17 | 81.0 | 60.0-92.3 |
| Total | 265 | 232 | 87.6 | 83.0-91.0 |
The clinical performance for the detection of Influenza A and Influenza B are shown in Table 19 and Table 20 respectively.
Table 19: Clinical Performance for Detection of Influenza A
| Nano-Check Influenza+ COVID-19 Dual Test | RT-PCR Comparator | | Total |
| --- | --- | --- | --- |
| | Positive | Negative | |
| Positive | 417 | 6 | 423 |
| Negative | 63 | 1483 | 1546 |
| Total | 480 | 1489 | 1969 |
| Positive Percent Agreement (PPA) = 86.9% (417/480, 95% CI: 83.6% - 89.6%) | | | |
| Negative Percent Agreement (NPA) = 99.6% (1483/1489, 95% CI: 99.1% - 99.8%) | | | |
Table 20: Clinical Performance for Detection of Influenza B
| Nano-Check Influenza+ COVID-19 Dual Test | RT-PCR Comparator | | Total |
| --- | --- | --- | --- |
| | Positive | Negative | |
| Positive | 99 | 5 | 104 |
| Negative | 15 | 1850 | 1865 |
| Total | 114 | 1855 | 1969 |
| Positive Percent Agreement (PPA) = 86.8% (99/114, 95% CI: 79.4% - 91.9%) | | | |
| Negative Percent Agreement (NPA) = 99.7% (1850/1855, 95% CI: 99.4% - 99.9%) | | | |
1. Clinical Sensitivity:
Please refer to Section VI.C (clinical Studies) above for the clinical validation. The PPA for the test for each analyte is as follows:
SARS-CoV-2: 87.6% (232/265) – 95% CI: 83.0% - 90.0%
Flu A: 86.9% (417/480) – 95% CI: 83.6% - 89.6%
Flu B: 86.8% (99/114) – 95% CI: 79.4% - 91.9%
K243561 - Page 24 of 25
{25}
K243561 - Page 25 of 25
2. Clinical Specificity:
Please refer to Section VI.C (Clinical Studies) above for the clinical validation. The NPA for the test for each analyte is as follows:
SARS-CoV-2: 99.8% (1701/1704) – 95% CI: 99.5% - 99.9%
Flu A: 99.6% (1483/1489) – 95% CI: 99.1% - 99.8%
Flu B: 99.7% (1850/1855) – 95% CI: 99.4% - 99.9%
D Clinical Cut-Off:
The test is a qualitative test with a binary positive/negative signal and there is no clinical cut-off for the test.
E Expected Values/Reference Range:
When the test result is valid, it produces binary values, positive or negative for SARS-CoV-2, influenza A and influenza B antigens. Therefore, the reference range is not applicable.
VII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
VIII Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.