K243560 · Remex Medical Corp. · OLO · Dec 13, 2024 · Neurology
Device Facts
Record ID
K243560
Device Name
Remex Spine Surgery Navigation Instrument
Applicant
Remex Medical Corp.
Product Code
OLO · Neurology
Decision Date
Dec 13, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The Remex Spine Surgery Navigation instrument is designed for use with Anatase Spine Surgery Navigation System(K230783) and Remex Spine Surgery Navigation System II(K233513). The Anatase Spine Surgery Navigation System and Remex Spine Surgery Navigation System II are indicated for precise positioning of surgical instruments or spinal implants during general spine surgery when reference to a rigid anatomical structure, such as the vertebra, can be identified relative to a patient's fluoroscopic or CT imagery. It is intended as a planning and intraoperative guidance system to enable open or percutaneous image guided surgery by means of registering intraoperative 2D fluoroscopic projections to pre-operative 3D CT imagery. Example procedures include: Posterior-approach spinal implant procedures, such as pedicle screw placement, with the lumbar region.
Device Story
Remex Spine Surgery Navigation Instrument comprises Nav Awl, Nav Drill Guide, Nav Tap, Nav Screwdriver, and Nav Handle; used with Anatase Spine Surgery Navigation System (K230783) and Remex Spine Surgery Navigation System II (K233513). System registers intraoperative 2D fluoroscopic projections to pre-operative 3D CT imagery to provide intraoperative guidance. Surgeon uses instruments to create entry points, drill holes, tap threads, and drive screws into vertebrae. Operates in clinical setting; physician-operated. Provides visual guidance on system display to assist precise positioning of spinal implants. Benefits include improved accuracy and guidance during open or percutaneous spinal procedures.
Clinical Evidence
Bench testing only. Verification and validation activities performed in accordance with IEC 62304 (software), ISO 14971 (risk management), and ISO 10993-1 (biocompatibility). Design verification confirmed outputs meet all input requirements. No clinical data presented.
Technological Characteristics
Navigation instruments (Awl, Drill Guide, Tap, Screwdriver, Handle) for use with existing navigation systems. Materials evaluated per ISO 10993-1. Connectivity via integration with navigation system hardware. Software-based guidance using 2D-3D image registration. Sterilization method consistent with predicate devices.
Indications for Use
Indicated for patients undergoing general spine surgery, specifically posterior-approach spinal implant procedures such as pedicle screw placement in the lumbar region, where navigation relative to rigid anatomical structures (vertebrae) via fluoroscopic or CT imagery is required.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
Remex Spine Surgery Navigation System II (K233513)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 13, 2024
Remex Medical Corp. Cheng-Hsiung Wang Quality Representative I 4F, No.9, Jingke Road, Nantun District Taichung, 408209 Taiwan
Re: K243560
Trade/Device Name: Remex Spine Surgery Navigation Instrument Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: October 11, 2024 Received: November 18, 2024
Dear Cheng-Hsiung Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
{2}------------------------------------------------
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shumaya Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K243560
Device Name Remex Spine Surgery Navigation Instrument
#### Indications for Use (Describe)
The Remex Spine Surgery Navigation instrument is designed for use with Anatase Spine Surgery Navigation System(K230783) and Remex Spine Surgery Navigation System II(K233513). The Anatase Spine Surgery Navigation System and Remex Spine Surgery Navigation System II are indicated for precise positioning of surgical instruments or spinal implants during general spine surgery when reference to a rigid anatomical structure, such as the vertebra, can be identified relative to a patient's fluoroscopic or CT imagery. It is intended as a planning and intraoperative guidance system to enable open or percutaneous image guided surgery by means of registering intraoperative 2D fluoroscopic projections to pre-operative 3D CT imagery. Example procedures include: Posterior-approach spinal implant procedures, such as pedicle screw placement, with the lumbar region.
Type of Use (Select *one* or *both*, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "REMEX" in a sans-serif font. The letters are gray, except for a green parallelogram in the lower-left corner of the "R". The letters are large and bold, taking up most of the image space. The background is white.
## 510(k) Summary
| Submitted By | REMEX MEDICAL CORP.<br>4F., No.9 Jingke Rd., Nantun District, Taichung, TW 408209<br>Tel. +886-4-23595336<br>Fax.+886-4-23598875 | |
|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Cheng-Hsiung Wang<br>Kevin816@mail.intai.com.tw | |
| Date Prepared | October 11th, 2024 | |
| Device Name | Remex Spine Surgery Navigation Instrument | |
| Classification Name | Stereotaxic instrument | |
| Regulation Number | 882.4560 | |
| Product Codes | OLO | |
| Device Class | Class II | |
| Predicate Devices<br>Information | K233513, Remex Spine Surgery Navigation System II (primary)<br>K230783, Anatase Spine Surgery Navigation System | |
| Device Description | The Remex Spine Surgery Navigation Instrument is designed for use with<br>Anatase Spine Surgery Navigation System (K230783) and Remex Spine Surgery<br>Navigation System II (K233513).<br>The Remex Spine Surgery Navigation Instrument consists of the Nav Awl, Nav<br>Drill Guide, Nav Tap, Nav Screwdriver and Nav Handle.<br>1. Nav Awl: The subject device is intended to support the surgeon to create an<br>entry point before implanting the screw.<br>2. Nav Drill Guide: The subject device is intended to assist in drilling holes to a<br>desired depth and ensuring proper orientation.<br>3. Nav Tap: The subject device is intended to create screw threads.<br>4. Nav Screwdriver: The subject device is intended to driver the screw into the<br>vertebra.<br>Nav Handle: The subject device is intended to connected with Nav Awl, Nav Tap<br>or Nav Screwdriver for navigation tracking purpose. | |
| Indications for Use | The Remex Spine Surgery Navigation Instrument is designed for use with<br>Anatase Spine Surgery Navigation System (K230783) and Remex Spine Surgery<br>Navigation System II (K233513).<br>The Anatase Spine Surgery Navigation System and Remex Spine Surgery<br>Navigation System II, are indicated for precise positioning of surgical<br>instruments or spinal implants during general spinal surgery when reference to a<br>rigid anatomical structure, such as the vertebra, can be identified relative to a<br>patient's fluoroscopic or CT imagery. It is intended as a planning and<br>intraoperative guidance system to enable open or percutaneous image guided | |
| surgery by means of registering intraoperative 2D fluoroscopic projections to | | |
| pre-operative 3D CT imagery. | | |
| Example procedures include: | | |
| Posterior-approach spinal implant procedures, such as pedicle screw placement, | | |
| within the lumbar region. | | |
| Technological<br>Characteristics | The subject devices have the same intended use, same indications for use, same<br>sterilization method, same fundamental technology, and similar materials, similar<br>design as the predicate devices, Anatase Spine Surgery Navigation System<br>(K230783) and Remex Spine Surgery Navigation System II (K233513). | |
| | The difference between the subject devices and the predicate devices are the<br>upgrade of the system software version, and new instruments for corresponding<br>compatible devices. However, these modifications share same function and<br>fundamental technology with the predicate devices. | |
| Performance Data | Verification and validation activities have been completed to provide sufficient<br>assurance that the subject device meets the performance requirements under its<br>indications for use conditions. Below is a summary of all performance tests which<br>should carried out on the subject device to demonstrated that the subject device<br>performs as safely and effectively as the predicate device. | |
| | Test | Description |
| | Software | Software is verified and validated in accordance with FDA<br>guidance for the content of premarket submissions for<br>software contained in medical devices and IEC 62304. |
| | Risk Assessment | The effectiveness of all risk control measures is verified in<br>accordance with ISO 14971. |
| | Design<br>Verification | The design output fulfills all design input requirements. |
| | Biocompatibility<br>Evaluation | The materials that contact with human tissue, blood or<br>bone must be considered for biocompatibility to ensure no<br>rejection, allergic and adverse reactions occur. The subject<br>device is evaluated in accordance with FDA guidance for<br>the use of international standard ISO 10993-1, "Biological<br>evaluation of medical devices - Part 1: Evaluation and<br>testing within a risk management process" and ISO 10993-<br>1. |
| | The intended use and fundamental technology of the device are identical to the<br>predicate device, and could share the same testing methods as the predicate<br>devices. All the results are passed. Therefore, REMEX believes the testing<br>demonstrated that the subject devices are as safe and effective as the predicate<br>devices. | |
| Conclusion | The Remex Spine Surgery Navigation Instrument and the predicate devices have<br>the same intended use and have the same or similar technological characteristics.<br>Moreover, the information contained in this submission demonstrates that any<br>differences in their technological characteristics do not raise new issues of safety | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "REMEX" in a sans-serif font. The letters are gray, except for a green leaf-like shape that replaces the bottom left portion of the "R". The word is presented in a clear and legible manner, with a modern and professional aesthetic.
Remex Medical Corporation 4F, No. 9, Jingke Road, Nantun District, Taichung City408209, Taiwan
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "REMEX" in a sans-serif font. The letters are gray, except for a green leaf-like shape in the lower-left corner of the "R". The text is centered and appears to be a logo or brand name.
Remex Medical Corporation 4F, No. 9, Jingke Road, Nantun District, Taichung City408209, Taiwan
| or effectiveness. Hence, The Remex Spine Surgery Navigation Instrument is |
|---------------------------------------------------------------------------|
| substantially equivalent to the predicate devices. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.