K243558 · Apr 11, 2025
Canvas Dx
Cognoa, Inc.
Real-world data from the intended use population (patient-use data, caregiver/provider questionnaires, and video analysis)
The device includes an authorized Predetermined Change Control Plan (PCCP) that allows for future model retraining, threshold optimization, and feature addition based on real-world data collected during routine clinical use.
PCCP; Machine Learning; Post-market optimization; Model retraining
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
PCCP-based Model Updates; Prospective real-world data collection for model optimization; Follow-up/Duration: Ongoing; Study Period: Post-clearance
Patients ages 18 months through 72 months at risk for developmental delay
Not applicable for this study
Optimization of decision thresholds, model retraining, and feature addition to maintain or improve PPV, NPV, and determinate rate
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Autism Spectrum Disorder diagnosis
Machine learning algorithm modeled after standard medical evaluation methodologies
Prospective, double-blinded, single-arm clinical validation study performed at 14 sites
3 (clinical specialists)
Indications for Use
Canvas Dx is intended for use by healthcare providers as an aid in the diagnosis of Autism Spectrum Disorder (ASD) for patients ages 18 months through 72 months who are at risk for developmental delay based on concerns of a parent, caregiver, or healthcare provider. The device is not intended for use as a stand-alone diagnostic device but as an adjunct to the diagnostic process.
Device Story
Canvas Dx is a prescription Software as a Medical Device (SaMD) used by healthcare providers to aid in ASD diagnosis. It integrates inputs from three sources: a caregiver questionnaire (mobile app), patient video analysis (video analyst portal), and a healthcare provider questionnaire (HCP portal). A machine learning algorithm processes these inputs to generate one of three outputs: 'Positive for autism', 'Negative for autism', or 'Indeterminate'. The system includes backend infrastructure for data management, HIPAA-compliant encryption, and security. The device is used in clinical settings to support the diagnostic process, helping providers make more informed clinical decisions. The submission includes a Predetermined Change Control Plan (PCCP) allowing for future model retraining, threshold optimization, and feature addition based on real-world data to maintain performance and generalizability.
Clinical Evidence
No clinical testing was performed in support of this submission. The device relies on clinical data previously established under DEN200069.
Technological Characteristics
SaMD and MDDS components. Software compliant with ISO 62304. Machine learning algorithm architecture. Connectivity via mobile app, web portals, and backend cloud infrastructure. Security includes HIPAA-compliant encryption.
Indications for Use
Indicated for patients ages 18 months through 72 months at risk for developmental delay based on concerns of parent, caregiver, or healthcare provider, as an aid in the diagnosis of Autism Spectrum Disorder (ASD). Not for use as a stand-alone diagnostic device.
Regulatory Classification
Identification
A pediatric Autism Spectrum Disorder diagnosis aid is a prescription device that is intended for use as an aid in the diagnosis of Autism Spectrum Disorder in pediatric patients.
Special Controls
In combination with the general controls of the FD&C Act, the pediatric Autism Spectrum Disorder diagnosis aid is subject to the following special controls:
- (1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including an evaluation of sensitivity, specificity, positive predictive value, and negative predictive value using a reference method of diagnosis and assessment of patient behavioral symptomology.
- (2) Software verification, validation, and hazard analysis must be provided. Software documentation must include a detailed, technical description of the algorithm(s) used to generate device output(s), and a cybersecurity assessment of the impact of threats and vulnerabilities on device functionality and user(s).
- (3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
- (4) Labeling must include:
- (i) Instructions for use, including a detailed description of the device, compatibility information, and information to facilitate clinical interpretation of all device outputs:
- (ii) A summary of any clinical testing conducted to demonstrate how the device functions as an interpretation of patient behavioral symptomology associated with Autism Spectrum Disorder. The summary must include the following:
- (A) A description of each device output and clinical interpretation;
- (B) Any performance measures, including sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV);
- (C) A description of how the cut-off values used for categorical classification of diagnoses were determined; and
- (D) Any expected or observed adverse events and complications.
(iii)A statement that the device is not intended for use as a stand-alone diagnostic.
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including an evaluation of sensitivity, specificity, positive predictive value, and negative predictive value using a reference method of diagnosis and assessment of patient behavioral symptomology.
(2) Software verification, validation, and hazard analysis must be provided. Software documentation must include a detailed, technical description of the algorithm(s) used to generate device output(s), and a cybersecurity assessment of the impact of threats and vulnerabilities on device functionality and user(s).
(3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
(4) Labeling must include:
(i) Instructions for use, including a detailed description of the device, compatibility information, and information to facilitate clinical interpretation of all device outputs; and
(ii) A summary of any clinical testing conducted to demonstrate how the device functions as an interpretation of patient behavioral symptomology associated with Autism Spectrum Disorder. The summary must include the following:
(A) A description of each device output and clinical interpretation;
(B) Any performance measures, including sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV);
(C) A description of how the cutoff values used for categorical classification of diagnoses were determined; and
(D) Any expected or observed adverse events and complications.
(iii) A statement that the device is not intended for use as a stand-alone diagnostic.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 11, 2025
Cognoa, Inc.
% Alexia Haralambous
Senior Principal
RQM+
2790 Mosside Blvd, #800
Monroeville, Pennsylvania 15146
Re: K243558
Trade/Device Name: Canvas Dx
Regulation Number: 21 CFR 882.1491
Regulation Name: Pediatric Autism Spectrum Disorder Diagnosis Aid
Regulatory Class: Class II
Product Code: QPF
Dated: March 12, 2025
Received: March 12, 2025
Dear Alexia Haralambous:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243558 - Alexia Haralambous
Page 2
required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
{2}
K243558 - Alexia Haralambous
Page 3
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay R. Gupta -S
Jay Gupta
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243558
Device Name
Canvas Dx
Indications for Use (Describe)
Canvas Dx is intended for use by healthcare providers as an aid in the diagnosis of Autism Spectrum Disorder (ASD) for patients ages 18 months through 72 months who are at risk for developmental delay based on concerns of a parent, caregiver, or healthcare provider.
The device is not intended for use as a stand-alone diagnostic device but as an adjunct to the diagnostic process.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
Traditional 510(k) Premarket Notification
Canvas Dx
# 510(k) Summary
Page 1 of 5
{5}
Canvas Dx
# 510(K) SUMMARY
## DATE PREPARED
March 10, 2025
## MANUFACTURER AND 510(k) OWNER
Cognoa, Inc.
2185 Park Blvd.
Palo Alto, CA 94306
Telephone: +1 (650) 785-2624
Official Contact: Sharief Taraman, MD, Chief Executive Officer
## REPRESENTATIVE/CONSULTANT
Alexia Haralambous, MS, RAC
Allison Komiyama, Ph.D., RAC
RQM+
Telephone: +1 (267) 778-5114
Email: aharalambous@rqmplus.com
## DEVICE INFORMATION
Proprietary Name/Trade Name: Canvas Dx
Common Name: Pediatric Autism Spectrum Disorder diagnosis aid
Regulation Number: 21 CFR 882.1491
Class: II
Product Code: QPF
Premarket Review: OHT5/DHT5A
Review Panel: Neurology
## PREDICATE DEVICE IDENTIFICATION
Canvas Dx is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
| --- | --- | --- |
| DEN200069 | Cognoa ASD Diagnosis Aid / Cognoa, Inc. | ✓ |
## DEVICE DESCRIPTION
Canvas Dx is a prescription diagnostic aid for healthcare professionals (HCP) considering the diagnosis of Autism Spectrum Disorder (ASD) in patients 18 months through 72 months of age at risk for developmental delay. The subject device is identical to the Cognoa ASD Diagnosis Aid which was authorized under DEN200069 and was renamed Canvas Dx shortly thereafter. Canvas Dx consists of Software as a Medical Device (SaMD) together with several medical device data system (MDDS) components. The SaMD components consist of the following:
- Device inputs:
- Device Input 1: The answers to the Caregiver Questionnaire
- Device Input 2: Patient Video Analysis
- Device Input 3: The answers to the Healthcare Provider Questionnaire
- A machine learning (ML) algorithm ('Algorithm') modeled after standard medical evaluation methodologies and drives the device outputs.
- Device outputs:
{6}
Canvas Dx
- ‘Positive for autism’
- ‘Negative for autism’
- ‘Indeterminate’
The MDDS components that are compatible with the SaMD components include the following:
- A caregiver facing mobile application, which provides Device Input 1;
- A video analyst system, which provides Device Input 2;
- A healthcare provider portal, which provides Device Input 3;
- Several supporting software and backend services and infrastructure, including privacy and security encryption and infrastructure in compliance with HIPAA and other best practices.
The subject of this submission is the inclusion of a Predetermined Change Control Plan (PCCP) that allows updates to the Canvas Dx model and performance thresholds.
## INDICATIONS FOR USE
Canvas Dx is intended for use by healthcare providers as an aid in the diagnosis of Autism Spectrum Disorder (ASD) for patients ages 18 months through 72 months who are at risk for developmental delay based on concerns of a parent, caregiver, or healthcare provider.
The device is not intended for use as a stand-alone diagnostic device but as an adjunct to the diagnostic process.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Cognoa, Inc. believes that the Canvas Dx device is substantially equivalent to the predicate devices based on the information summarized here:
| | Subject Device: Canvas Dx | Primary Predicate: Cognoa ASD Diagnosis Aid (DEN200069) |
| --- | --- | --- |
| Manufacturer | Cognoa, Inc. | Cognoa, Inc. |
| Regulation Number | 21 CFR 882.1491 | 21 CFR 882.1491 |
| Regulation Name | Pediatric Autism Spectrum Disorder diagnosis aid | Pediatric Autism Spectrum Disorder diagnosis aid |
| Regulatory Class | Class II | Class II |
| Product Code | QPF | QPF |
| Indications for Use | Canvas Dx is intended for use by healthcare providers as an aid in the diagnosis of Autism Spectrum Disorder (ASD) for patients ages 18 months through 72 months who are at risk for developmental delay based on concerns of a parent, caregiver, or healthcare provider.
The device is not intended for use as a stand-alone diagnostic device but as an adjunct to the diagnostic process. | The Cognoa ASD Diagnosis Aid is intended for use by healthcare providers as an aid in the diagnosis of Autism Spectrum Disorder (ASD) for patients ages 18 months through 72 months who are at risk for developmental delay based on concerns of a parent, caregiver, or healthcare provider.
The device is not intended for use as a stand-alone diagnostic device but as an adjunct to the diagnostic process. |
| Data Collection | A caregiver facing mobile application, which provides Device Input 1 (Caregiver Questionnaire) | A caregiver facing mobile application, which provides Device Input 1 (Caregiver Questionnaire) |
{7}
Canvas Dx
| | • A video analyst portal, which provides Device Input 2 (Patient Video Analysis)
• A healthcare provider portal, which provides Device Input 3 (HCP Questionnaire) | • A video analyst portal, which provides Device Input 2 (Patient Video Analysis)
• A healthcare provider portal, which provides Device Input 3 (HCP Questionnaire) |
| --- | --- | --- |
| **Outputs** | • ‘Positive for Autism’
• ‘Negative for Autism’
• ‘Indeterminate’ | • ‘Positive for ASD
• ‘Negative for ASD
• ‘No Result’ |
| **Software** | Compliant to ISO 62304 | Compliant to ISO 62304 |
| **Performance** | The Canvas Dx performance thresholds will be updated based on improvements to the Algorithm per the terms of the PCCP presented in this 510(k). | The following thresholds are based on the initial performance of the device under the initial clinical validation study (prospective, double-blinded, single-arm) performed at 14 sites, compared to the clinical reference standard, in support of the DEN200069 submission. These values are based on review by 3 clinical specialists.
• PPV: 80.77% (CI: 70.27-88.82%)
• NPV: 98.25% (CI: 90.61-99.96%)
• Determinate Rate (percent of users who received a positive or negative result): 31.76% (CI: 63.58%-87.67%)
• Sensitivity: 98.44% (CI: 91.6%-99.96%)
• Specificity: 78.87% (CI: 67.56%-87.67%) |
## SUMMARY OF NON-CLINICAL TESTING
Testing was performed in accordance with IEC 62304:2006 to demonstrate safety based on current industry standards. The results of these tests indicate that Canvas Dx is substantially equivalent to the predicate device.
## SUMMARY OF CLINICAL TESTING
Clinical testing was not performed in support of this submission
## PREDETERMINED CHANGE CONTROL PLAN
Canvas Dx includes an authorized Predetermined Change Control Plan (PCCP) that allows for planned updates and modifications based on real-world data within the cleared indications for use. These modifications are related to:
1) Quantitative measures of machine learning-enabled device software functions (ML-DSF) performance specifications. This includes re-training the ML model based on new data within the intended use population from the same type and range of input signal. The re-training may result in internal decision trees with different weights or overall structure
2) Device inputs to the ML-DSF. This may involve expanding the algorithm to include new inputs of the same signal type.
Page 4 of 5
{8}
Canvas Dx
The three modifications are summarized below.
1. **Optimization of Decision Thresholds**: Adjusting the predefined decision thresholds based on new data without modifying the existing model architecture, parameters, or structure. Optimizing decision thresholds can yield a higher determinate rate while maintaining or improving performance on PPV and NPV.
2. **Model Retraining**: Retraining the model with new data while maintaining the same set of input features and model architecture. Retraining may change feature weights and decision nodes based on the new data. Retraining the model based on new data enables the Device to adapt, improving generalizability and maintaining performance across evolving clinical populations.
3. **Feature Addition**: Adding neurodevelopmental/neurobehavioral features as inputs from the parent, video, or healthcare provider questionnaires, while maintaining the original set of input features and model architecture. Any new features will be limited to those that can be directly observed or reported through existing questionnaires and will not require additional separate testing or assessments. The addition of features may improve Device generalizability and performance, allowing for higher reliability and accuracy of the Device overall.
All modifications are subject to performance evaluation and update procedures, including software verification and validation. All modifications follow the same data management practices, as outlined in the PCCP Modification Protocol.
## CONCLUSION
As the intended use and technological features of the subject Canvas Dx device have not changed since the original De Novo authorization of the Cognoa ASD Diagnosis Aid under DEN200069, it can be concluded that the subject Canvas Dx device does not raise new issues of safety or effectiveness compared to the predicate device. The Canvas Dx model and performance thresholds will be updated based on improvements to the Algorithm per the terms of the PCCP presented in this 510(k). Therefore, the subject Canvas Dx device is substantially equivalent to DEN200069.
Page 5 of 5
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.