The JETi Registry provided clinical evidence on the safety, performance, and clinical benefits of the JETi HTS system for the treatment of thrombosis in the peripheral vasculature, specifically supporting the updated Indications for Use for DVT.
Registry; DVT; Peripheral vasculature; Standard of care
260 subjects treated for thrombosis in the peripheral vasculature (arterial, venous, or arteriovenous); 116 subjects treated for lower extremity (LE) DVT; Sample Size: 260 total subjects (116 in LE DVT analysis population); Number of Sites: 27 sites in the US and Europe
Not applicable for this study
Primary effectiveness: >=75% reduction in modified Marder Score; Primary safety: composite of JETi-related major adverse events up to 30 days
Indications for Use
The JETi™ Hydrodynamic Thrombectomy System is intended to remove/aspirate fluid and break-up soft emboli and thrombus from the peripheral vasculature, including deep vein thrombosis (DVT), and to subselectively infuse/deliver diagnostics or therapeutics with or without vessel occlusion.
Device Story
Hydro-mechanical aspiration system for intravascular thrombus removal. System components: catheter, pump set, saline drive unit (SDU), accessory cart, suction tubing, non-sterile canister set. Operation: continuous aspiration of thrombotic material into catheter; high-pressure saline stream at catheter tip macerates thrombus during aspiration. Used in clinical settings by physicians. Output: aspirated fluid/thrombus; infusion of diagnostics/therapeutics. Clinical benefit: thrombus reduction, improved vessel patency, reduced leg pain, improved quality of life. Affects clinical decision-making by providing minimally invasive treatment for peripheral vascular thrombosis.
Clinical Evidence
Prospective, single-arm, multicenter registry (JETi Registry) of 260 subjects (116 with LE DVT). Primary effectiveness: 84.5% (93/110) limbs achieved ≥75% reduction in modified Marder Score (p < 0.0001 vs 64% performance goal). Primary safety: 1.7% (2/116) rate of JETi-related major adverse events at 30 days (95% CI: 0.21%–6.09%), meeting <10% acceptance criterion. Improvements observed in vessel patency, Villalta score, pain, and quality of life.
Technological Characteristics
Hydro-mechanical aspiration system. Components: catheter, pump set, saline drive unit, accessory cart, suction tubing, canister set. Principle: high-pressure saline jet for clot maceration combined with aspiration. Connectivity: standalone system. Sterilization: not specified. Software: not specified.
Indications for Use
Indicated for removal/aspiration of fluid and break-up of soft emboli and thrombus from peripheral vasculature, including deep vein thrombosis (DVT), and subselective infusion/delivery of diagnostics or therapeutics with or without vessel occlusion in adult patients.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 4, 2025
Abbott Medical
Fiona Pu
Manager, Regulatory Affairs
3200 Lakeside Drive
Santa Clara, California 95054
Re: K243549
Trade/Device Name: JETi™ Hydrodynamic Thrombectomy System
Regulation Number: 21 CFR 870.5150
Regulation Name: Embolectomy catheter
Regulatory Class: Class II
Product Code: QEZ, KDQ, FOX
Dated: March 4, 2025
Received: March 4, 2025
Dear Fiona Pu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K243549 - Fiona Pu
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K243549 - Fiona Pu
Page 3
Sincerely,
GREGORY W. O'CONNELL -S
Digitally signed by
GREGORY W. O'CONNELL -S
Date: 2025.04.04 14:49:22
-04'00'
Gregory O'Connell
Assistant Director
DHT2C: Division of Coronary and Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
| Submission Number (if known) |
| --- |
| K243549 |
| Device Name |
| JETi™ Hydrodynamic Thrombectomy System |
| Indications for Use (Describe) |
| The JETi™ Hydrodynamic Thrombectomy System is intended to remove/aspirate fluid and break-up soft emboli and thrombus from the peripheral vasculature, including deep vein thrombosis (DVT), and to subselectively infuse/deliver diagnostics or therapeutics with or without vessel occlusion. |
| Type of Use (Select one or both, as applicable) |
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{5}
Abbott
ATTACHMENT 011-04
# 510(K) SUMMARY
| 510(k) Summary Per 21 CFR §807.92 | |
| --- | --- |
| 510(k) Number | K243549 |
| Date Prepared | November 15, 2024 |
| Submitter Name & Address | Abbott Medical 3200 Lakeside Drive Santa Clara, CA 95054 |
| Contact Person | Fiona Pu 408-748-4990 |
| Alternative Contact Person | Claire Elkins 408-845-0625 |
| Proprietary/Trade Name | JETi™ Hydrodynamic Thrombectomy System |
| Common/Usual Name | Embolectomy/Thrombectomy Catheter |
| Classification Name | Embolectomy Catheter (21 CFR 870.5150, Product Code QEZ) Infusion stand (21 CFR 880.6990, Product Code FOX) Vacuum-powered body fluid suction apparatus (21 CFR 880.6740, Product Code KDQ) |
| Predicate Device | Predicate device: ZelanteDVT Thrombectomy System, K202218 Reference device: JETi Hydrodynamic Thrombectomy System, K232458 |
| Device Description | The JETi™ Hydrodynamic Thrombectomy System (JETi HTS) is a hydro-mechanical aspiration system intended for the removal of intravascular thrombus. The system is comprised of the JETi™ Catheter, JETi™ Pump Set, JETi™ Saline Drive Unit (SDU), JETi™ Accessory Cart, JETi™ Suction Tubing, and JETi™ Non-sterile Canister Set. The JETi Hydrodynamic Thrombectomy System is designed to continuously aspirate thrombotic material into the JETi Catheter, where a high-pressure stream of saline within the catheter tip macerates the thrombus as it is aspirated. |
| Indications for Use/Intended Use | The JETi™ Hydrodynamic Thrombectomy System is intended to remove/aspirate fluid and break-up soft emboli and thrombus from the peripheral vasculature, including deep vein thrombosis (DVT), and to sub-selectively infuse/deliver diagnostics or therapeutics with or without vessel occlusion. |
| Comparison to Predicate Device | The JETi Hydrodynamic Thrombectomy System (JETi HTS) is substantially equivalent to the predicate ZelanteDVT Thrombectomy |
JETi™ Hydrodynamic Thrombectomy System
Traditional 510(k) Pre-market Notification
{6}
Abbott
ATTACHMENT 011-04
| | System (K202218). Although the Indications for Use statement of the subject device is not identical to the predicate device, the differences do not alter the intended therapeutic use of the device and do not affect the safety and effectiveness of the device relative to the predicate. Both the JETi HTS and predicate device have the same intended use: removal of thrombi or emboli from and infusion of fluids into the peripheral vasculature including DVT.
The technological principle for both the subject and predicate device is based on a clot fragmentation mechanism, which uses aspiration and a jet with fluid delivery features. Although JETi HTS and predicate device have different clot fragmentation mechanism and catheter size offering, the different technological characteristics between the subject device and predicate do not raise different questions of safety and effectiveness. |
| --- | --- |
| Summary on Non-Clinical Testing | Non-clinical testing was not required to support the update to the Indications for Use. The JETi HTS (K232458) is used as a reference device to account for differences in the technological features between the predicate ZelanteDVT Thrombectomy System and the JETi HTS. JETi HTS (K232458) is identical in design, technological, and performance characteristics to the subject device. |
| Summary on Clinical Testing | The JETi Registry collected real-world data on the safety, performance, and clinical benefits of the JETi HTS System for the treatment of thrombosis in the peripheral vasculature. It is a prospective, single-arm, multicenter study that registered 260 subjects treated for thrombosis in the peripheral vasculature from 27 sites in the United States (US) and Europe. Subjects were treated for arterial, venous, or arteriovenous thromboses and are followed for 12 months, with data collection at baseline, during the procedure, discharge, 30-day, and 12-month visits. This was a standard of care all-comers registry, and there were no restrictions on presenting thrombus location in the peripheral vasculature or onset of signs and symptoms.
A total of 116 registered subjects were treated for lower extremity (LE) DVT. Subject mean age was 56.6 ± 18.5 years, with the youngest subject being 18 and oldest subject being 85 years old. The majority, 55.2% (64/116), were under 65 years old. The split between male 50.0% (58/116) and female 50.0% (58/116) subjects was even. Of the analysis population, 50.9% (59/116) identified as Caucasian, 6.9% (8/116) identified as Black; 1.7% (2/116) identified as American Indian or Alaskan Native; 31.9% (37/116) of subjects declined to disclose or did not disclose their race information due to Europe’s |
JETi™ Hydrodynamic Thrombectomy System
Traditional 510(k) Pre-market Notification
Page 2 of 3
{7}
Abbott
ATTACHMENT 011-04
| | local regulation. Self-identification of ethnicity of the subjects were 10.3% (12/116) as Hispanic or Latino, 59.5% (69/116) as non-Hispanic or Latino, 30.2% (35/116) as declined or non-disclosed due to Europe’s local regulations.
Treatment strategy, including the use of adjunctive therapy, was per investigator’s discretion. The majority of patients in the registry (66.1% of limbs) were not treated with any adjunctive interventional device(s) (CDT or other thrombectomy device).
The primary effectiveness endpoint, defined as at least 75% reduction in modified Marder Score from pre-JETi venogram to final venogram, per core laboratory assessment, was achieved in 84.5% (93/110) limbs, with a one-sided 97.5% lower confidence limit of 76.4% comparing to a performance goal of 64% with a p value of < 0.0001. The primary effectiveness endpoint of the JETi Registry was met. Primary effectiveness endpoint was comparable when the JETi system was used with or without adjunctive interventional device(s); with at least 75% thrombus reduction achieved in 86.8% and 83.3% of limbs, respectively.
The primary safety endpoint, defined as the composite of JETi-related major adverse events up to 30 days post-JETi per CEC adjudication, was 1.7% (2/116) (95% CI, 0.21%, 6.09%). The rate of 1.7% (2/116) is less than the acceptance criterion of 10%. The primary safety endpoint of the JETi Registry was met.
Treated vessel patency and compressibility, Villalta post-thrombotic syndrome score, leg pain, and quality of life were all improved at 30 days. |
| --- | --- |
| Statement of Equivalence | The proposed changes to the Indications for Use, namely adding the specific use for DVT, is supported by the clinical data presented and thus, has been shown to not raise new or different questions of safety and effectiveness relative to the predicate, therefore, it can be concluded that the JETi HTS is substantially equivalent to the predicate device. |
JETi™ Hydrodynamic Thrombectomy System
Traditional 510(k) Pre-market Notification
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.