DCM is a small worn activity monitor designed for documenting physical movement associated with applications in physiological monitoring. The device is intended to monitor the activity associated with movement during sleep. DCM can be used to analyze circadian rhythms and assess activity in any instance where quantifiable analysis of physical motion is desirable. DCM is indicated for monitoring of adult patients only.
Device Story
DCM is a wrist-worn wearable biosensor (PW010) that continuously records high-resolution 3-axis linear acceleration data. The device is prescribed by healthcare professionals or researchers for use during sleep or normal daily activities over days to weeks. The system includes a mobile app (PW400) on an off-the-shelf mobile device (PW030) for data collection, cloud-based storage/processing (PW100), and an investigator dashboard (PW500) for data review. The patient does not interact with the device to control collection. Data is transmitted to the cloud, where it is post-processed based on timestamp and magnitude of acceleration. Healthcare professionals use the dashboard to view activity data, which assists in clinical decision-making regarding circadian rhythms and physical motion analysis. The device benefits patients by providing objective, quantifiable activity data without requiring patient interaction.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including acceleration accuracy (5% at 1g), timing accuracy (±10s/hour), data storage/transmission reliability, usability (IEC 62366-1), EMC (IEC 60601-1-2), wireless coexistence (ANSI C63.27), and electrical safety (IEC 60601-1/1-11).
Technological Characteristics
Wrist-worn triaxial MEMS accelerometer; sensitivity 0.061mg/LSB; dynamic range +/-2g; frequency 12.5Hz. Materials: PC/ABS plastic biosensor, PET foam/acrylate EKG electrode patch. Powered by CR2032 lithium coin-cell. Connectivity: Bluetooth/mobile app to cloud. Non-sterile. Software developed per IEC 62304.
Indications for Use
Indicated for monitoring of adult patients only. Intended for documenting physical movement, monitoring activity during sleep, analyzing circadian rhythms, and assessing activity where quantifiable analysis of physical motion is desirable.
Regulatory Classification
Identification
A biofeedback device is an instrument that provides a visual or auditory signal corresponding to the status of one or more of a patient's physiological parameters (e.g., brain alpha wave activity, muscle activity, skin temperature, etc.) so that the patient can control voluntarily these physiological parameters.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter when it is a prescription battery powered device that is indicated for relaxation training and muscle reeducation and prescription use, subject to § 882.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 16, 2025
PneumoWave, Ltd
Stuart Kelly
Director of Engineering & Chief Technical Officer
Suites 3 and 4, Second Floor, Maxim 3
Maxim Office Park, Parklands Avenue
Eurocentral, ML1 4WQ
United Kingdom
Re: K243513
Trade/Device Name: DCM (PW-DCM)
Regulation Number: 21 CFR 882.5050
Regulation Name: Biofeedback Device
Regulatory Class: Class II
Product Code: LEL
Dated: March 14, 2025
Received: March 17, 2025
Dear Stuart Kelly:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243513 - Stuart Kelly
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243513 - Stuart Kelly
Page 3
Sincerely,
Jay R. Gupta -S
Jay Gupta
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243513 | |
| Device Name DCM (PW-DCM) | |
| Indications for Use (Describe) DCM is a small worn activity monitor designed for documenting physical movement associated with applications in physiological monitoring. The device is intended to monitor the activity associated with movement during sleep. DCM can be used to analyze circadian rhythms and assess activity in any instance where quantifiable analysis of physical motion is desirable. DCM is indicated for monitoring of adult patients only. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
DCM 510(k) Summary March 2025
# 510(k) Summary
March 14, 2025
## Submitter Information:
PneumoWave Limited
Suites 3 and 4,
Second Floor, Maxim 3
Maxim Office Park
Parklands Ave
Eurocentral ML1 4WQ
United Kingdom
## Contact:
Stuart Kelly
Director of Engineering and CTO
+44 (0)7505 710 276
## Device Information:
Proprietary Name: DCM
Common Name: Movement Assessment Device
Classification Name: Biofeedback Device
Classification Codes: LEL
Regulation Number: 21 CFR §882.5050
## Predicate Devices:
Actigraph LEAP (K231532)
Actigraph, LLC
## Indications for Use:
DCM is a small worn activity monitor designed for documenting physical movement associated with applications in physiological monitoring.
The device is intended to monitor the activity associated with movement during sleep.
DCM can be used to analyze circadian rhythms and assess activity in any instance where quantifiable analysis of physical motion is desirable.
1 of 7
{5}
DCM is indicated for monitoring of adult patients only.
## Device Description:
DCM is a wrist-worn wearable device intended to continuously record high resolution digital acceleration data associated with a patient's physical movement.
In practice, a healthcare professional or researcher can prescribe the device to collect physiological data from patients during sleep and in applications where quantifiable analysis of physical motion is desirable.
The device is set up to collect data by the healthcare professional then placed on the subject's wrist. The device is designed to be worn during normal activities and/or during sleep over a period of days to weeks. The patient does not need to interact with the device to control data collection.
The data stored on the device can be transmitted to the cloud for storage, and made accessible to healthcare professionals or researchers for further analysis. Downloaded data can be post-processed based on the timestamp and magnitude of acceleration along each axis.
The DCM system comprises a system of components:
- wearable biosensor (PW010)
- off the shelf mobile device (PW030) running the DCM mobile app (PW400)
- cloud-based data storage and data processing (PW100) (back-end)
- investigator dashboard (PW500) accessed through a web browser (front-end)
## Comparison to the Predicate:
The table below compares the DCM device to the Actigraph LEAP (K231532).
DCM 510(k) Summary March 2025
{6}
| | Subject Device: DCM | Predicate: Actigraph Leap (K231532) | Comments |
| --- | --- | --- | --- |
| Indications for use | DCM is a small worn activity monitor designed for documenting physical movement associated with applications in physiological monitoring.
The device is intended to monitor the activity associated with movement during sleep.
DCM can be used to analyze circadian rhythms and assess activity in any instance where quantifiable analysis of physical motion is desirable.
DCM is indicated for monitoring of adult patients only. | The ActiGraph LEAP™ is a small worn activity monitor designed for documenting physical movement associated with applications in physiological monitoring.
The device is intended to monitor the activity associated with movement during sleep.
The ActiGraph LEAP™ can be used to analyze circadian rhythms and assess activity in any instance where quantifiable analysis of physical motion is desirable. | Equivalent |
| Intended patient population | Adult | Adult | Equivalent
Not intended for paediatric use |
| Prescription or OTC | Prescription | Prescription | Equivalent |
| Classification & Product code | 882.5050, LEL | 882.5050, LEL | Equivalent |
| Where device is worn (anatomical site) | Wrist | Wrist | Equivalent |
| What the device measures | Physical activity | Physical activity | Equivalent |
DCM 510(k) Summary March 2025
{7}
| | Subject Device: DCM | Predicate: Actigraph Leap (K231532) | Comments |
| --- | --- | --- | --- |
| Use during rest/motion | During motion and at rest | During motion and at rest | Equivalent |
| Spot checking / continuous measurement | Continuous measurement | Continuous measurement | Equivalent |
| Real time monitoring / retrospective evaluation | Retrospective | Retrospective | Equivalent |
| Operating principle | Measurement of linear accelerations in 3-axes associated with the patient's physical movement. | Measurement of linear accelerations in 3-axes associated with the patient's physical movement. | Equivalent |
| Biosensor sensor type | Triaxial MEMS accelerometer | Triaxial MEMS accelerometer | Equivalent |
| Biosensor sensitivity/resolution | 0.061mg/LSB | 3 mg/LSB | Equivalent: Similar range |
| Biosensor dynamic range | +/-2 g | +/-8 g | Equivalent: Similar range |
| Biosensor frequency | 12.5Hz | 32-256Hz | Equivalent: Similar range |
| Battery | Lithium coin-cell battery [CR2032] | Lithium rechargeable battery, max 4.2V | Equivalent: Both lithium batteries |
| Mounting components | Biosensor is mounted on an EKG electrode (Vermed A10005SG (Nissha Medical) or Skintact FS-40 (Leonhard Lang GmbH) which can be placed directly on the wrist, or attached to a wrist strap | Armband/strap | Equivalent: Biosensor can be mounted in the same way as predicate. |
DCM 510(k) Summary March 2025
{8}
DCM 510(k) Summary March 2025
| | Subject Device: DCM | Predicate: Actigraph Leap (K231532) | Comments |
| --- | --- | --- | --- |
| Skin contacting material | Biosensor: PC/ABS plastic
EKG Electrode fixation patch: PET foam with acrylate skin adhesive (off the shelf EKG electrode) | Housing: Plastic polyamide (Chendong PA+50%GF CP9021 F60LW BK3027F)
Lens: Hardened glass (Caihong CG01pro+ high aluminosilicate glass)
Strap (band) and buckle: Silicon (Midgold GF2041E) and Plastic polyamide (Chendong PA+50%GF CP9021 F60LW BK3027F)
Bottom case lens: Nylon | Equivalent: All skin contacting materials are biocompatible polymers. |
| Software and firmware | Mobile app is pre-loaded onto an off the shelf mobile device. The mobile app sends data to the cloud. The health care professional accesses a web-based dashboard for data transfer. | Embedded firmware for microcontroller which does not connect to the internet. Communication software for data transfer. | Equivalent |
| User interface | Mobile app screen shows battery level, functional status and time of day. No data measures or feedback to patient (Health care professionals can access the web-based dashboard). | Device screen shows battery level, functional status, and time of day. No data measures or feedback to patient. | Equivalent |
| Sterility | Supplied non-sterile | Supplied non-sterile | Equivalent |
| Respiratory rate measurement | None | None | Equivalent |
| Apnea claims | None | None | Equivalent |
| Alarms / Alerts | None | None | Equivalent |
{9}
# Summary of Testing
The 510(k) summary of the primary predicate device does not include detailed information on testing that would allow for comparison. However, other biofeedback devices such as MotionWatch (K132764) include testing information similar to the table below that summarizes tests performed to establish the safety and effectiveness of DCM.
| Requirement | Test Method | Pass/Fail Criteria | Result |
| --- | --- | --- | --- |
| The system shall measure acceleration with accuracy of 5% or better (at 1g), in 3 orthogonal directions with sensitivity to at least 0.005g | Bench testing with the biosensor in a range of orientations. | Acceleration recorded by the device meets the requirements.
Accelerometer accuracy to be tested across extended duration data collection runs to confirm that there is no sensor drift. | PASS |
| The system shall capture sensor data with timing accuracy within ±10 seconds per hour | Bench testing with mobile app paired to biosensor for the length of a typical session, timestamps inspected. | Data is transmitted to the cloud and the timestamps are visible and accurate within the requirements when viewed in the Investigator Dashboard. | PASS |
| In case of connectivity issues, data is stored on the device until it can be transferred. | Bench testing with mobile app paired with biosensor, manual interruption and restoration of connectivity, followed by inspection of data. | Data is stored on the biosensor when connection to the mobile device is interrupted and transferred when connection is restored.
Data is stored on the mobile device when connection to the cloud platform is interrupted and transferred when connection is restored. | PASS |
| Usability | Usability complies with the requirements of IEC 62366-1 in relation to safety | Usability activities are conducted according to the IEC 62366-1 process and demonstrates that the usability of the medical device is acceptable with regard to safety | PASS |
DCM 510(k) Summary March 2025
{10}
| Requirement | Test Method | Pass/Fail Criteria | Result |
| --- | --- | --- | --- |
| Packaging | Testing to confirm that the packaging design and construction is suitable to protect the system components during exposure to typical shipping stresses and rough handling. | Device meets visual inspection criteria and passes functional tests following exposure to typical shipping stresses and rough handling | PASS |
| EMC | EMC testing per IEC 60601-1-2 and 47 CFR Part 15 Subpart B. | Device meets requirements for emissions (Class B) and immunity | PASS |
| Wireless Coexistence | Wireless coexistence testing per ANSI C63.27 | No interruption to wireless data connections. | PASS |
| Radio Frequency | Radiated Spurious Emissions testing per 47 CFR 15.247 | Device meets requirements for spurious emissions. | PASS |
| Electrical Safety | Testing to verify applicable clauses of IEC 60601-1 and IEC 60601-1-11 safety standards | Device meets applicable requirements for electrical, mechanical and thermal safety, for healthcare and home use environments. | PASS |
| Software verification and validation | Software developed and tested per IEC 62304 | Software developed and maintained in accordance with the IEC 62304 lifecycle process, and all verification and validation test passed | PASS |
## Clinical Testing
DCM did not require clinical studies to support substantial equivalence to the predicate device. Non-clinical and human factors testing results indicate that the device meets user needs and that it is safe and effective for its intended use.
## Conclusions
DCM is substantially equivalent to the predicate device with respect to intended use, design features, technological characteristics, and performance criteria. All predetermined pass/fail criteria were met.
DCM 510(k) Summary March 2025
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.