← Product Code [HXA](/productcode/HXA) · K243477

# Universal Tendon Spacer (K243477)

_Askorn Medical · HXA · Jan 10, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K243477

## Device Facts

- **Applicant:** Askorn Medical
- **Product Code:** [HXA](/productcode/HXA.md)
- **Decision Date:** Jan 10, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3025
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

Universal Tendon Spacer is a medical device used in the reconstruction of the flexor and extensor tendons of the hand. The device is implantable for 2 to 6 months and intended for single use. This device is used for: Adherent or mutilated tendons following trauma or a primary repair which has failed. Absence of the tendon sheath. Non-functioning, mutilated or adherent tendon pulleys. Ruptured tendons. The patient must be in good health with sufficient tissue cover. Sufficiently healthy joints, bone marrow and adequate vascular and neuro-muscular conditions are also necessary to facilitate a definite functional improvement using this reconstruction technique. The patient must be cooperative, especially concerning the need for rehabilitation measures.

## Device Story

Universal Tendon Spacer (UTS) is a temporary, single-use, ovoid-shaped silicone rod implanted for 2-6 months during two-stage tendon reconstruction surgery. It acts as a passive prosthesis to create a synovial neo-sheath in the digital canal, allowing for subsequent tendon grafting. The device features a single rod containing multiple sizes. It is implanted by a surgeon in a clinical setting. The rod is radio-opaque due to added barium sulfate, allowing visualization via X-ray. By facilitating the formation of a smooth neo-sheath, the device enables the future implanted tendon to slide, restoring finger function. Benefits include improved range of motion and functional recovery for patients with mutilated or adherent tendons.

## Clinical Evidence

Bench testing only. Mechanical strength (tensile strength, elongation, strain at break) tested per NF EN ISO 527-2; results equivalent to predicate. Biocompatibility testing (cytotoxicity, sensitization, irritation, pyrogenicity) performed per ISO 10993-1 and USP Class VI. Packaging integrity and 5-year shelf-life validated per ISO 11607-1/2 and NF EN 868-5.

## Technological Characteristics

Material: Medical-grade silicone elastomer with barium sulfate (USP Class VI). Design: Ovoid-shaped rod, single-use, sterile. Principle: Passive tendon prosthesis for two-stage reconstruction. Manufacturing: Injection-molding. Sterilization: Gamma irradiation. Connectivity: None. Software: None.

## Regulatory Identification

A passive tendon prosthesis is a device intended to be implanted made of silicon elastomer or a polyester reinforced medical grade silicone elastomer intended for use in the surgical reconstruction of a flexor tendon of the hand. The device is implanted for a period of 2 to 6 months to aid growth of a new tendon sheath. The device is not intended as a permanent implant nor to function as a replacement for the ligament or tendon nor to function as a scaffold for soft tissue ingrowth.

## Predicate Devices

- Single Size Tendon Spacer ([K000019](/device/K000019.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 10, 2025

Askorn Medical Denis Pichon General Manager 44c rue de Bray Cesson-Sévigné, 35510 France

Re: K243477

Trade/Device Name: Universal Tendon Spacer Regulation Number: 21 CFR 888.3025 Regulation Name: Passive Tendon Prosthesis Regulatory Class: Class II Product Code: HXA Dated: November 5, 2024 Received: November 8, 2024

Dear Denis Pichon:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Christopher Ferriera, MS Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

Submission Number (if known)

### K243477

Device Name

Universal Tendon Spacer

### Indications for Use (Describe)

Universal Tendon Spacer is a medical device used in the reconstruction of the flexor and extensor tendons of the hand. The device is implantable for 2 to 6 months and intended for single use.

This device is used for:

Adherent or mutilated tendons following trauma or a primary repair which has failed. Absence of the tendon sheath. Non-functioning, mutilated or adherent tendon pulleys. Ruptured tendons. The patient must be in good health with sufficient tissue cover. Sufficiently healthy joints, bone marrow and adequate vascular and neuro-muscular conditions are also necessary to facilitate a definite functional improvement using this reconstruction technique. The patient must be cooperative, especially concerning the need for rehabilitation measures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

he-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

| 510(k) #: | K243477 | 510(k) Summary | Prepared on: 2024-11-08 |
|-----------|---------|----------------|-------------------------|
|-----------|---------|----------------|-------------------------|

| Contact Details | 21 CFR 807.92(a)(1) |
|-----------------|---------------------|
|-----------------|---------------------|

| Applicant Name              | Askorn Medical                              |
|-----------------------------|---------------------------------------------|
| Applicant Address           | 44c rue de Bray Cesson-Sévigné 35510 France |
| Applicant Contact Telephone | +33223355335                                |
| Applicant Contact           | Mr. Denis Pichon                            |
| Applicant Contact Email     | d.pichon@askorn.bzh                         |

| Device Name | 21 CFR 807.92(a)(2) |
|-------------|---------------------|
|-------------|---------------------|

| Device Trade Name   | Universal Tendon Spacer     |
|---------------------|-----------------------------|
| Common Name         | Passive tendon prosthesis   |
| Classification Name | Prosthesis, Tendon, Passive |
| Regulation Number   | 888.3025                    |
| Product Code(s)     | HXA                         |

| Legally Marketed Predicate Devices |                                                          | 21 CFR 807.92(a)(3) |
|------------------------------------|----------------------------------------------------------|---------------------|
| Predicate #                        | Predicate Trade Name (Primary Predicate is listed first) | Product Code        |
| K000019                            | Single Size Tendon Spacer                                | HXA                 |

| Device Description Summary |  | 21 CFR 807.92(a)(4) |
|----------------------------|--|---------------------|
|----------------------------|--|---------------------|

Universal Tendon Spacer (UTS) is a medical device made of silicone for temporary implantation (two to six months) for use in the twostage reconstruction of the flexor and extensor tendons of the hand. When implanted, UTS is in contact with the tissues and bones of the fingers for a long-term contact duration (>30 days).

This medical device contributes to the rehabilitation of the digital canal by creating a neo-sheath which should be as large as possible depending on the patient's anatomy. It is the body with the silicone which makes it possible to obtain the synovial neosheath which will receive, during the second operative time, a tendon of the formation of this neo-sheath the tendon will be able to slide again in the sheath and will make the finger of the patient work.

The Universal Tendon Spacer is a single size ovoid-shaped rod whose over its entire length. The rule is for the larger diameter to always be twice that of the smaller diameter.

The principle is the same as those of rival man difference being that the four different sizes generally provided are all available on one single rod. The UTS is made of long-term nigh performance implantable silicone to which barium sulfate has been added to make the rod radio-opaque for X-ray examination. Raw materials of Universal Tendon Spacer meet USP Class VI test requirements.

Only one reference exists for this device, and it is not intended to be assembled with any other device.

The intended patient population is adults and children. The device is for single use only and is provided sterile.

{5}------------------------------------------------

# Intended Use/Indications for Use

Universal Tendon Spacer is a medical device used in the flexor and extensor tendons of the hand. The device is implantable for 2 to 6 months and intended for single use.

This device is used for:

Adherent or mutilated tendons following trauma or a primary repair which has failed. Absence of the tenctioning, mutilated or adherent tendon pulleys. Ruptured tendons.

The patient must be in good health with sufficiently healthy joints, bone marrow and adequate vascular and neuro-muscular conditions are also necessary to facilitate a definite functional improvement using the The patient must be cooperative, especially concerning the need for rehabilitation measures.

# Indications for Use Comparison

The subject device has the exact same indications for use than the predicate device.

# Technological Comparison

The Universal Tendon Spacer is technology substantially equivalent to the predicate device in material, design features, size and mechanical strength.

The subject and predicate device are based on the exact same technological characteristics:

- Function and principle of operation: designed for the two-step tendon reconstruction surgery of the hand
- Design: unique size implant with an oval cross-section, exact same design and dimensions
- Material: same medical-grade silicone elastomer and barium sulfate from the same raw material supplier

- Manufacturing process same injection-molding process (same mold used), same final cleaning and packaging process, same sterilization process (gamma irradiation)

- Mechanical strength: mechanical testing was performed, and results indicate that the Universal Tendon Spacer manufactured by Askorn is equivalent to the predicate device.
- Both devices are provided sterile, are for single use only, and are biocompatible as per ISO 10993-1

Both the proposed and predicate designs are intended to function as a temporary implant for the two-stage tendon of the hand and are manufactured from the same materials with the same manufacturing process.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Mechanical strength test was performed to compare the Universal Tendon Spacer strength with the predicate device, according to NF EN ISO 527-2. Tensile strength, elongation and strain at break were found to be equivalent to the predicate device.

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing:

The biocompatibility evaluation for the Universal Tendon Spacer was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:

- · Cytotoxicity
- · Sensitization
- · Irritation
- · Pyrogenicity

The Universal Tendon Spacer is considered as an implant tissue and bone contacting for a duration of more than 30 days. The UTS materials conforms to USP Class VI requirements.

Packaging testing:

Package tests were performed to validate package integrity and shelf-life claims (5 years). The following test were carried out after realtime and accelerated aging:

- · Visual inspection of the packaging
- · Dye penetration (integrity test)
- · Seal strength test

# 21 CFR 807.92(a)(6)

21 CFR 807.92(a)(5)

### 21 CFR 807.92(a)(5)

{6}------------------------------------------------

The results show that the packaging type double blister for Universal Tendon Spacer has correct integrity, resistance and peelability proprieties after sterilization and 5 years aging, according to the relevant chapters of ISO 11607-1, ISO 11607-2 and NF EN 868-5 standards.

From the above provided information and the history of the subject device, Universal Tendon Spacer is a ssafe, as effective, and performs as well as the legally marketed device.

#### CONCLUSION:

The design characteristics of the device do not raise any new types of questions of safety or effectiveness. The non-clinical data support the safety of the device. The Universal Tendon Spacer is substantially equivalent to the predicate marketed device based on a comparison of intended use, indications for use, design and technological characteristics. From the evidence submitted in this 510(k) Premarket Notification, the device is as safe, as effective and performs at least as well as the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K243477](https://fda.innolitics.com/device/K243477)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
