The Compressor Nebulizer system is intended to provide air to the pneumatic nebulizer in order to aerosolize medications for inhalation by the patient for respiratory disorders. The nebulizer is driven by the integral air compressor. The system is designed for use with pediatric (ages 2 years and above) and adult patients in the home, hospital, and sub-acute settings.
Device Story
Compressor Nebulizer system (models NB-1100, NB-1101, NB-1102, NB-1103) provides compressed air to a pneumatic nebulizer kit to aerosolize liquid medications for inhalation. Operates on Venturi principle: motor-driven compressed air creates high-speed flow through a small pipe; negative pressure draws liquid to a barrier; high-speed impact converts liquid into mist particles. Components include compressor (motor, pump, air filter, PCB in NB-1102/1103) and nebulizer kit (cup, air tube, mouthpiece, adult/child masks). Used in home, hospital, and sub-acute settings by patients or caregivers. Output is medical aerosol for respiratory treatment; helps manage respiratory disorders by delivering medication directly to airways.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (ANSI AAMI ES60601-1), EMC (IEC 60601-1-2), biocompatibility (ISO 10993 series, ISO 18562 series), and comparative particle size characterization (MMAD, GSD, respirable dose/fraction) against the predicate using Albuterol Sulfate, Ipratropium Bromide, and Budesonide.
Technological Characteristics
Pneumatic jet nebulizer; AC-powered compressor. Components: shell, motor, pump, air filter, PCB (NB-1102/1103). Standards: ANSI AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11. Biocompatibility per ISO 10993 and ISO 18562. Dimensions vary by model (e.g., 267.9x157.3x104.6mm to 286x154x134.5mm).
Indications for Use
Indicated for pediatric (ages 2+) and adult patients with respiratory disorders requiring aerosolized medication delivery in home, hospital, or sub-acute settings.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 2, 2025
JOYTECH Healthcare Co., Ltd.
% Jie Yang
Consultant
Chonconn Consulting Co., Ltd.
Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District
Shenzhen, Guangdong 518067
China
Re: K243468
Trade/Device Name: Compressor Nebulizer (NB-1100); Compressor Nebulizer (NB-1101); Compressor Nebulizer (NB-1102); Compressor Nebulizer (NB-1103)
Regulation Number: 21 CFR 868.5630
Regulation Name: Nebulizer
Regulatory Class: Class II
Product Code: CAF
Dated: June 6, 2025
Received: June 6, 2025
Dear Jie Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243468 - Jie Yang
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243468 - Jie Yang
Page 3
Sincerely,
John S. Bender-S
2025.07.02 14:09:29 -04'00'
for Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243468 | |
| Device Name Compressor Nebulizer (NB-1100); Compressor Nebulizer (NB-1101); Compressor Nebulizer (NB-1102); Compressor Nebulizer (NB-1103) | |
| Indications for Use (Describe) The Compressor Nebulizer system is intended to provide air to the pneumatic nebulizer in order to aerosolize medications for inhalation by the patient for respiratory disorders. The nebulizer is driven by the integral air compressor. The system is designed for use with pediatric (ages 2 years and above) and adult patients in the home, hospital, and sub-acute settings. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
Prepared Date: 2025/6/4
1. Submission sponsor
Name: JOYTECH Healthcare Co., Ltd.
Address: No. 365. Wuzhou Road, Yuhang Economic Development Zone, Hangzhou City, 311100 Zhejiang, China.
Contact person: Cong Jing
Title: Regulatory manager
E-mail: jingc@sejoy.com
Tel: +86-571-81957767
2. Submission correspondent
Name: Chonconn Consulting Co., Ltd.
Address: No. Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P.R. China 518067
Contact person: Yang Jie
E-mail: yangjie@chonconn.com
Tel: +86-755 33941160
3. Subject Device Information
| Trade/Device Name | Compressor Nebulizer |
| --- | --- |
| Model | NB-1100, NB-1101, NB-1102, NB-1103 |
| Common Name | Nebulizer |
| Regulatory Class | Class II |
| Classification | 21CFR §868.5630 / Nebulizer |
| Submission type | Traditional 510(K) |
4. Predicate Device Information
K110860, NE-C801 Nebulizer Compressor System, Omron Healthcare, Incorporated
5. Device Description
The compressor Nebulizer is mainly composed of the compressor and nebulizer kit(Optional). The compressor is mainly composed of shell, compressed motor, pump, fuse wire, air filter, power cord and plug (NB-1100,NB-1101,NB-1102,NB-1103 applicable), PCB (only model NB-1102,NB-1103applicable). The device is equipped with a nebulizer kit including Nebulizer Cup, Air Tube, Mouthpiece(Optional), Adult Mask(Optional), Child Mask(Optional) to easily delivery the medical aerosol. The Nebulizer kit is available in three models/configurations as below.
1 / 10
{5}
NK-101: Nebulizer Cup, Air Tube, Adult Mask and/or Child Mask, Mouthpiece;
NK-301: Nebulizer Cup, Air Tube, Adult Mask and/or Child Mask;
NK-501: Nebulizer Cup, Air Tube, Mouthpiece
This device operates on the Venturi principle. According to the principle of Venturi, the compressed air of the motor is used to form a high-speed air flow through the small pipe mouth. The negative pressure generated drives the liquid or other fluid to spray together on the barrier, and under the high-speed impact makes the droplets turn into mist particles to spray from the outlet trachea.
## 6. Intended use & Indication for use
The Compressor Nebulizer system is intended to provide air to the pneumatic nebulizer in order to aerosolize medications for inhalation by the patient for respiratory disorders. The nebulizer is driven by the integral air compressor.
The system is designed for use with pediatric (ages 2 years and above) and adult patients in the home, hospital, and sub-acute settings.
## 7. Comparison to the Predicate Device
| ITEM | Proposed Device
Compressor Nebulizer
NB-1100, NB-1101, NB-1102, NB-1103
K243468 | Predicate Device
Omron Compressor
Nebulizer Systems NE-C801
K110860 | Comparison Result |
| --- | --- | --- | --- |
| Manufacture | JOYTECH Healthcare Co., Ltd. | Omron Healthcare, Incorporated | / |
| Product code | CAF | CAF | Same |
| Indications for Use | The Compressor Nebulizer system is intended to provide air to the pneumatic nebulizer in order to aerosolize medications for inhalation by the patient for respiratory disorders. The nebulizer is driven by the integral air compressor.
The system is designed for use with pediatric (ages 2 years and above) and adult patients in the home, hospital, and sub-acute settings. | The NE-C801 Nebulizer Compressor System is intended to provide air to the pneumatic nebulizer in order to aerosolize medications for inhalation by the patient for respiratory disorders. The system is designed for use with pediatric (defined by the prescribed medication) and adult patients in the home, hospital, and sub-acute settings. | Same |
{6}
| Principle of Operation | Yes Pneumatic (gas powered) jet nebulizer | Yes Pneumatic (gas powered) jet nebulizer | Same |
| --- | --- | --- | --- |
| Compressed gas source | Nebulizer compressor | Nebulizer compressor Wall air / oxygen with flow rate control | Equivalent |
| Aerosolization | Continuous during inhalation and exhalation | Continuous during inhalation and exhalation | Same |
| Typical flow rate | ≥6.5L/min | 8 lpm | Similar |
| Reservoir size | 8 ml | 7 ml | Similar |
| Power Source | AC | AC | Same |
| Components available in kit with nebulizer | Mouthpiece or mask | Mouthpiece or mask | Same |
| Component / Accessories intended use | All are single patient, multi-use | All are single patient, multi-use | Same |
| Nebulizer components cleanable | Yes | Yes | Same |
| Standards met | ANSI AAMI ES 60601-1
IEC 60601-1-2
IEC 60601-1-11 | IEC 6060 1-1
IEC 60601-1-2 | Equivalent |
| Operating conditions | +5°C to +40°C (+41°F to +104°F)/15% to 90% RH/ 860 hPa to 1060 hPa | 10°C to 40°C
30% to 85% RH | Similar |
| Storage conditions | -20°C to +55°C (-4°F to +131°F)/5% to 93% RH/ 700 hPa to 1060 hPa | -20°C to 60°C
10% to 95% RH | Similar |
| Dimensions (mm) | 267.9 mm × 157.3 mm × 104.6 mm
170.2 mm × 144.4 mm × 117.9 mm
246.3 mm × 121.6 mm × 148.5 mm
286 mm × 154 mm × 134.5 mm | 142(W) x 72(D) x 98(H) | Similar |
| Weight(kg) | 1.554kg
1.238kg
1.413kg
1.57kg | 0.270kg | Similar |
3 / 10
{7}
| Aerosol performance | Table: Comparative particle test comparison | Table: Comparative particle test comparison | Equivalent |
| --- | --- | --- | --- |
As evidenced by the above table, both the subject and the predicate devices have same intended use, but the subject and predicate devices have different technological characteristics. However, performance testing was conducted on the subject device, and it was established that the differences in technological characteristics between the subject and the predicate does not raise different questions of safety or effectiveness.
{8}
Table: Comparative particle test comparison
Adult condition
| Test items | Drug | Subject Device (JOYTECH NB-1103) with adult mask | Predicate device (OMRON NE-C801) with adult mask | Subject Device (JOYTECH NB-1103) with Mouthpiece | Predicate device (OMRON NE-C801) with Mouthpiece |
| --- | --- | --- | --- | --- | --- |
| MMAD(μm) | Albuterol Sulfate (5 mg/2.5 ml) | 3.466±0.173 | 4.173±0.126 | 3.761±0.06 | 3.7±0.248 |
| | Ipratropium Bromide (0.5 mg/2 ml) | 3.413±0.061 | 4.227±0.383 | 3.954±0.032 | 4.067±0.16 |
| | Budesonide (0.5 mg/2 ml) | 4.781±0.102 | 6.06±0.352 | 5.344±0.025 | 5.704±0.124 |
| GSD | Albuterol Sulfate (5 mg/2.5 ml) | 1.988±0.033 | 1.845±0.046 | 2.013±0.016 | 1.993±0.029 |
| | Ipratropium Bromide (0.5 mg/2 ml) | 1.979±0.029 | 1.843±0.077 | 1.979±0.02 | 1.898±0.055 |
| | Budesonide (0.5 mg/2 ml) | 1.764±0.01 | 1.605±0.105 | 1.81±0.018 | 1.72±0.087 |
| Respirable Dose(μg) | Albuterol Sulfate (5 mg/2.5 ml) | 1628.633±99.49 | 1239.14±150.674 | 1556.481±102.075 | 1138.953±108.149 |
| | Ipratropium Bromide (0.5 mg/2 ml) | 153.331±4.765 | 147.57±11.505 | 145.246±2.803 | 132.881±16.413 |
| | Budesonide (0.5 mg/2 ml) | 82.049±1.341 | 63.292±10.329 | 78.617±1.532 | 63.24±4.57 |
| Respirable fraction | Albuterol Sulfate (5 mg/2.5 ml) | 67.273±2.066 | 58.136±2.583 | 60.726±0.999 | 63.741±3.979 |
| | Ipratropium Bromide (0.5 mg/2 ml) | 67.958±0.561 | 56.616±5.676 | 58.302±0.731 | 60.095±2.068 |
| | Budesonide (0.5 mg/2 ml) | 49.433±1.385 | 31.781±4.934 | 40.028±0.211 | 36.912±1.149 |
{9}
| Total delivered Dose (μg) | Albuterol Sulfate (5 mg/2.5 ml) | 2418.382±90.219 | 2131.073±210.711 | 2562.202±150.856 | 1788.603±272.194 |
| --- | --- | --- | --- | --- | --- |
| | Ipratropiμm Bromide (0.5 mg/2 ml) | 225.607±6.35 | 260.75±14 | 249.133±3.845 | 221.123±26.449 |
| | Budesonide (0.5 mg/2 ml) | 166.098±3.573 | 199.13±5.564 | 196.392±3.13 | 171.383±17.566 |
| Total delivered Dose fraction (%) | Albuterol Sulfate (5 mg/2.5 ml) | 46.227±1.75 | 40.801±4.143 | 49.063±2.955 | 34.316±5.245 |
| | Ipratropiμm Bromide (0.5 mg/2 ml) | 41.066±1.176 | 48.1±3.044 | 45.427±0.695 | 40.534±4.349 |
| | Budesonide (0.5 mg/2 ml) | 31.225±0.606 | 37.472±1.072 | 37.024±0.537 | 32.058±3.425 |
| Coarse Particle Fraction (%)(>4.7μm) | Albuterol Sulfate (5 mg/2.5 ml) | 35.924±2.212 | 45.607±2.486 | 42.39±0.981 | 39.467±4.121 |
| | Ipratropiμm Bromide (0.5 mg/2 ml) | 35.232±0.613 | 47.06±5.886 | 44.926±0.698 | 43.487±2.044 |
| | Budesonide (0.5 mg/2 ml) | 54.669±1.372 | 72.054±5.274 | 63.702±0.144 | 67.192±0.865 |
| Fine Particle Fraction (%) (<4.7μm) | Albuterol Sulfate (5 mg/2.5 ml) | 64.076±2.212 | 54.393±2.486 | 57.61±0.981 | 60.533±4.121 |
| | Ipratropiμm Bromide (0.5 mg/2 ml) | 64.77±0.612 | 52.94±5.886 | 55.074±0.698 | 56.513±2.044 |
| | Budesonide (0.5 mg/2 ml) | 45.333±1.373 | 27.946±5.274 | 36.298±0.144 | 32.808±0.865 |
| Ultra-Fine Particle Fraction (%) (<1.0μm) | Albuterol Sulfate (5 mg/2.5 ml) | 5.891±0.795 | 3.207±0.351 | 4.806±0.133 | 4.26±0.435 |
| | Ipratropiμm Bromide (0.5 mg/2 ml) | 5.093±0.416 | 2.456±1.207 | 4.006±0.252 | 2.932±0.23 |
| | Budesonide (0.5 mg/2 ml) | / | / | / | / |
6 / 10
{10}
Pediatric condition
| Test items | Drug | Subject Device (JOYTECH NB-1103) with child mask | Predicate device (OMRON NE-C801) with child mask | Subject Device (JOYTECH NB-1103) with Mouthpiece | Predicate device (OMRON NE-C801) with Mouthpiece |
| --- | --- | --- | --- | --- | --- |
| MMAD(μm) | Albuterol Sulfate (5 mg/2.5 ml) | 3.651±0.126 | 4.313±0.355 | 3.707±0.113 | 4.475±0.275 |
| | Ipratropium Bromide(0.5 mg/2 ml) | 3.727±0.045 | 4.464±0.365 | 3.846±0.076 | 4.521±0.36 |
| | Budsonide (0.5 mg/2 mel) | 5.249±0.036 | 5.82±0.411 | 5.757±0.029 | 6.248±0.373 |
| GSD | Albuterol Sulfate (5 mg/2.5 ml) | 2.253±0.023 | 2.066±0.052 | 2.361±0.028 | 2.141±0.011 |
| | Ipratropium Bromide(0.5 mg/2 ml) | 2.257±0.037 | 2.093±0.045 | 2.381±0.021 | 2.129±0.027 |
| | Budesonide (0.5 mg/2 ml) | 2.082±0.032 | 1.946±0.039 | 2.139±0.02 | 1.999±0.023 |
| Respirable Dose(μg) | Albuterol Sulfate (5 mg/2.5 ml) | 999.897±53.773 | 981.97±130.406 | 1298.691±62.27 | 1008.827±116.221 |
| | Ipratropium Bromide(0.5 mg/2 ml) | 90.589±7.122 | 113.967±14.919 | 128.632±3.083 | 106.836±13.229 |
| | Budesonide (0.5 mg/2 ml) | 51.307±3.217 | 55.097±4.08 | 56.223±0.77 | 50.723±8.793 |
| Respirable fraction | Albuterol Sulfate (5 mg/2.5 ml) | 64.296±1.681 | 57.134±4.819 | 62.503±1.297 | 54.422±3.494 |
| | Ipratropium Bromide(0.5 mg/2 ml) | 63.278±0.395 | 55.054±4.774 | 60.641±1.078 | 54.084±4.865 |
| | Budesonide (0.5 mg/2 ml) | 45.429±0.443 | 37.724±4.958 | 39.103±0.225 | 33.385±3.354 |
| Total delivered Dose (μg) | Albuterol Sulfate (5 mg/2.5 ml) | 1554.763±66.546 | 1721.173±331.733 | 2081.438±132.873 | 1855.027±279.931 |
| | Ipratropium Bromide (0.5 mg/2 ml) | 143.129±10.902 | 207.123±28.684 | 212.202±5.762 | 197.53±16.536 |
7 / 10
{11}
| | Budesonide (0.5 mg/2 ml) | 112.976±7.384 | 146.303±20.144 | 143.774±1.357 | 151.83±13.266 |
| --- | --- | --- | --- | --- | --- |
| Total delivered Dose fraction (%) | Albuterol Sulfate (5 mg/2.5 ml) | 29.758±1.286 | 32.967±6.141 | 39.856±2.599 | 35.502±5.346 |
| | Ipratropium Bromide(0.5 mg/2 ml) | 26.156±1.969 | 37.898±5.327 | 38.586±1.084 | 35.898±3.328 |
| | Budesonide (0.5 mg/2 ml) | 21.305±1.372 | 27.55±3.912 | 27.236±0.343 | 28.845±2.019 |
| Coarse Particle Fraction (%)(>4.7μm) | Albuterol Sulfate (5 mg/2.5 ml) | 38.54±1.74 | 46.27±4.824 | 40.247±1.363 | 48.871±3.508 |
| | Ipratropium Bromide(0.5 mg/2 ml) | 39.612±0.434 | 48.266±4.606 | 42.12±0.969 | 49.191±4.684 |
| | Budesonide (0.5 mg/2 ml) | 58.453±0.481 | 66.472±4.87 | 64.547±0.227 | 70.463±3.126 |
| Fine Particle Fraction (%)(<4.7μm) | Albuterol Sulfate (5 mg/2.5 ml) | 61.461±1.741 | 53.73±4.824 | 59.753±1.363 | 51.129±3.508 |
| | Ipratropium Bromide(0.5 mg/2 ml) | 60.389±0.433 | 51.734±4.606 | 57.881±0.968 | 50.809±4.684 |
| | Budesonide (0.5 mg/2 ml) | 41.547±0.481 | 33.528±4.87 | 35.454±0.227 | 29.537±3.126 |
| Ultra-Fine Particle Fraction (%)(<1.0μm) | Albuterol Sulfate (5 mg/2.5 ml) | / | / | / | / |
| | Ipratropium Bromide(0.5 mg/2 ml) | / | / | / | / |
| | Budesonide (0.5 mg/2 ml) | / | / | / | / |
Note: "/" means that the raw data is below the detection limit LOD and cannot be calculated
{12}
8. Non-Clinical tests summary
The following performance data were provided in support of the substantial equivalence determination.
Electrical Safety and Electromagnetic Compatibility (EMC):
Electrical safety and EMC testing were conducted on the device. Testing established that, with respect to electrical safety, the device meets the applicable requirements of:
- ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
- IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-11 Edition 2.1 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
Particle Size characterization testing:
Particle characterization testing was performed on the subject device with both proposed patient interface and on the predicate per FDA guidance “Reviewer Guidance for Nebulizer, Metered Dose Inhalers, Spacers and Actuators” (FDA/CDRH – 1993).
Biocompatibility testing
According to FDA's 2023 Guidance entitled, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", the following biological compatibility test need to be executed as follow:
In Vitro Cytotoxicity Test -ISO10993-5
Skin Sensitization Test(polar and non polar) -ISO10993-10
Intracutaneous Reactivity Test-ISO10993-23
Acute Systemic Toxicity Test (polar and non polar) -ISO10993-11
Material-mediated pyrogens Test -ISO10993-11
Bacterial Reverse Mutation Assay --ISO10993-3
In Vitro Mammalian Chromosomal Aberration test --ISO10993-3
Subacute systemic Toxicity Test (polar and non polar) -ISO10993-11
Subcutaneous Implantation Test -ISO10993-6
Chemical characterization study- ISO10993-18
Toxicological risk assessment study-ISO10993-17
9 / 10
{13}
Biocompatibility evaluation- ISO 18562-1
Test for emissions of particulate matter - ISO 18562-2
Test for emissions of VOCs -ISO 18562-3
## Device life testing
Device life were tested with simulation method, the device meets the applicable requirements.
## 9. Clinical Tests
Not applicable.
## 10. Conclusion
Substantial equivalence comparisons, performance testing and compliance with voluntary standards demonstrate that the proposed subject device is substantially equivalent to the predicate device.
10 / 10
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.